A comparison of 18 different x-ray detectors currently used in dentistry
|
|
|
- Blaze Sutton
- 9 years ago
- Views:
Transcription
1 A comparison of 18 different x-ray detectors currently used in dentistry Allan G. Farman, BDS, PhD, DSc, MBA, a and Taeko T. Farman, DMD, PhD, MS, b Louiville, Ky THE UNIVERSITY OF LOUISVILLE Purpose. There has been a proliferation of available dental x-ray detectors over the recent past. The purpose of this short technical report is to provide a basic comparison of spatial resolution, contrast perceptibility, and relative exposure latitudes of 18 current dental x-ray detectors, including solid-state systems (CCD and CMOS), photostimulable phosphors, and analog film. Methods. Spatial resolution was measured using a mm Pb phantom test grid with a measurement range from 1.5 to 20 lp/mm. For contrast perceptibility, a 7-mm thick aluminum perceptibility test device with wells of mm depth at 0.1 mm intervals and 1 defect of 1.5 mm was used. The relative exposure latitude was determined by expert consensus using clear discrimination of the enamel-dentin junction as the lower limit and pixel blooming or unacceptable levels of cervical burnout as the upper limit. Results. The highest spatial resolution was found with Kodak InSight film, RVG-ui (CCD), and RVG 6000 (CMOS) detectors, each of which achieved 20 lp/mm, followed by the Planmeca Dixi2 v3 at $ 16 lp/mm. Contrast resolution was at least to 0.2 mm through 7 mm aluminum for all 18 detectors, with the best results being found for the Visualix HDI, RVG-ui, RVG 5000, and RVG 6000 detectors and the Schick CDR wired and wireless systems. The greatest exposure ranges were found with photostimulable phosphors and the Kodak RVG 6000 and RVG 5000 detectors. Conclusions. Most current x-ray detectors generally perform well in terms of spatial and contrast resolutions, and in terms of exposure latitude. These findings were independent of the modality applied. (Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2005;99:485-9) Essential properties for any digital imaging system should include the following: (1) the image produced is of diagnostic quality; (2) the radiation dose is equal or reduced compared to film; (3) that digital radiology techniques are compatible with conventional x-ray generators; (4) that lossless archiving is allowed in an image file format that promotes interoperability within the DICOM standard; and (5) the time required for the total procedure should be equal to or less than with film. 1 The detector need not be identical to film in all image properties; however, it should provide sufficient information for necessary diagnoses. The information needed might be task dependent: eg, high spatial resolution could well be needed to accurately measure the lengths of fine endodontic instruments, but would not necessarily be required in the detection of proximal surface dental caries. At this time there is a surprising dearth of information on the resolution and contrast requirements for detection of different dental structures a Professor of Radiology and Imaging Sciences, Department of Surgical and Hospital Dentistry, The University of Louisville Louisville, Ky. b Adjunct Professor of Radiology and Imaging Sciences, Department of Surgical and Hospital Dentistry, The University of Louisville Louisville, Ky. Received for publication Dec 22, 2003; accepted for publication Apr 4, /$ - see front matter Ó 2005 Elsevier Inc. All rights reserved. doi: /j.tripleo and disease processes both for analog film radiographs and for their digital counterparts. It should be remembered that the ideal detector should be able to multitask and therefore both sufficient spatial resolution and good contrast discrimination are needed. The production of a radiographic image can be considered to follow an imaging chain. This chain starts with the production of x-rays by a generator and their selective attenuation by the tissues to be studied. A detector with high sensitivity to x-radiation requires the use of a generator with a high accuracy at low dosages, particularly given a narrow exposure latitude. On the other hand, excessive latitude could results in unnecessarily high radiation dosages being applied to the patient if care is not taken to avoid this eventuality. At the level of the detector the imaging chain differs depending on the technology involved. For film radiography, the intraoral package needs to be unpacked in a dark/safelight environment and processed chemically. The quality of film images is greatly dependent upon appropriate time and temperature and strength of the chemistry applied. Optimal film radiograph results are only attained given appropriate processing conditions. For photostimulable phosphors, the latent images are highly light sensitive. 2 The exposed plates need to be kept out of direct normal light prior to processing or they can lose much of the x-ray signal. This is not always apparent to the dentist as images can be optimized by virtue of the acquisition and display software; however, the results of poor handling are likely to be measurable 485
2 486 Farman and Farman April 2005 physically. The processing of a photostimulable phosphor plate requires scanning by a laser and the photomultiplication and analog to digital conversion of the resulting signal. This process takes from seconds to minutes. The time is greater when high spatial resolution is desired as this requires a smaller scanning pitch be applied. The quality of the image from a photostimulable phosphor is dependent not only on the quality of the phosphor plate itself, but also on the scanning mechanism, the electronics, and the acquisition and display software. Solid-state detectors vary in construction; presently, however, they all consist of a charge-coupled device (CCD) or complementary metal oxide semiconductor (CMOS) chip that is sensitive to light and a scintillator layer that converts x-rays to light. There may be an intervening fiberoptic or the scintillator can be applied directly to the chip. The quality of an image produced by a solid-state detector is dependent upon not only the chip pixel dimensions, but also upon the type and configuration of the scintillator layer, the electronics including analog to digital conversion, and the acquisition and display software. Contrast resolution is an important characteristic of the diagnostic image. Some available systems can produce grayscale images in 16-bit depth making it possible to have up to 65,536 gray levels recorded. However, standard computer monitors at this time can display only a depth of 8-bit or 256 gray levels, but the remaining information need not be wasted as it is possible to select different viewing windows. Care needs to be made to differentiate between the detector bit-depth capability and the actual information that is stored and can be retrieved. Often digital detectors are capable of acquiring images of a greater bit-depth than is stored, and specific bits are discarded to reduce noise. The last steps in the imaging chain are display, viewing conditions, and observation. All diagnostic radiographs should be viewed under subdued ambient lighting, whether analog or digital. For digital imaging it is very important to have a high-quality display and to set the display to the recommended settings. Sixteen-bit color or less is not sufficient to display digital grayscale images at a contrast resolution of 256 gray levels (8-bit grayscale) irrespective of the system being used. Further, using the wrong resolution settings can impair apparent image spatial resolution if a 1:1 image size is not possible. Getting the most out of any imaging system requires training and experience of the observer. There has been a proliferation of available dental x-ray detectors over the past decade; however, there has been no recent study that independently compares these devices. This short technical paper briefly compares the spatial resolution, contrast perceptibility, and diagnostically acceptable exposure ranges of 18 currently marketed detectors including solid-state (CCD and CMOS), photostimulable phosphors, and analog film modalities. The newly released high-fill CMOS RVG 6000 and RVG 5000 detectors produced by Trophy Radiology are reported for the first time. METHODS The detectors studied were the following: CMOS devices including CDR wired and wireless systems (Schick Technologies, Inc, Long Island City, NY), RVG 6000 and RVG 5000 (Eastman Kodak, Rochester, NY); CCD-devices including CygnusRay MPS (Cygnus Technologies, Scottsdale, Ariz), Dexis (Provision Dental Supplies, Atlanta, Ga), Dixi@ v3 (Planmeca Oy, Helsinki, Finland), DSX 730 USB and DSX 730 Évolution (Owandy/Julie RadioVision, Gragny, France), Sigma (GE/Instrumentarium Imaging, Tuusula, Finland), Sidexis (Sirona, Bensheim, Germany), RSV (Visiodent, St Denis, France); RVG-ui (Formerly Trophy Radiology, now Kodak), ViperRay (Integra Medical, American Fork, Utah), and Visualix (Gendex, Des Plaines, Ill; photostimulable phosphor systems including DenOptix (Gendex), and ScanX (Air Techniques Inc, Hicksville, NY); and analog InSight silver halide film (Eastman Kodak, Rochester, NY). (See Fig 1 and Table I.) The AC x-ray generator used was an Irix with CCX timer (Kodak) operated at 70 kvp, 8 ma, with a nominal focal spot size of 0.7 mm and 2.5 mm Al equivalent filtration. For the CDR Wireless system, the detector to radio frequency receiver distance was kept to 1 m with the battery level and signal strengths maintained in the 75% to 90 % levels. The analog film used was InSight (Lot No , Expiry date ). All films were processed in a darkroom under safelight conditions (GBX-II, Kodak) using an automatic processor (Air Techniques AT 2000) with freshly primed ReadyMatic processing solutions (Kodak) at 288C. Exposures for all modalities were measured using an Unfors 512 Mult-O-Meter (Unfors Instument, Billdal, Sweden). Film radiographs were evaluated in a darkened room using transmitted light from a viewbox, with all extraneous light masked. The digital radiographic images were displayed on a 17-inch CRT screen (Samsung, Korea) in the same a darkened room. The 4 evaluators worked independently. Outcomes were decided by consensus when there were discrepancies in ratings. Spatial resolution As a guide to image sharpness, spatial resolution was measured in line pairs per millimeter (lp/mm). The test grid employed was a Pb mm (Nr , Nuclear Associates, Carle Place, NY) phantom providing
3 Volume 99, Number 4 Farman and Farman 487 Fig 1. Detectors used in this study. a measurement range from 1.5 to 20 lp/mm. In accordance with Nuclear Associates instruction manual, the focus-to-detector distance was set at 1 meter where possible. The one exception was the CDR Wireless TM System (Schick Technologies) where a distance of 75 cm was necessitated to achieve image reception. Given pixel symmetry the results were confirmed at 45 degrees to minimize aliasing. For the ScanX used in high-resolution modes the derived pixels are asymmetrical, necessitating making separate vertical and horizontal assessments. Determination of the limiting resolution by inspecting the finished radiographs for all digital systems was achieved with the assistance of software zoom provided by the manufacturers. For analog film radiographs, limiting resolution was assessed using a 37 power magnifying glass. Contrast perceptibility An aluminum perceptibility test device was manufactured under special order by Rogers Machine Co, Inc (Louisville, Ky). The device measures 25 mm 3 14 mm and is 7 mm thick with wells of mm depth at 0.1-mm intervals and one of 1.5 mm. The focusto-detector distance was set at 25 cm throughout. Practical exposure range The practically achievable exposure range to produce an image of diagnostic quality determines the exposure latitude available to the dentist when making radiographs. The wider the latitude the less likelihood of remakes being necessitated by variations in patient bone density and soft tissue bulk. An optical bench was used to assure reproducible projection geometry and to accommodate human tooth specimens and the various x-ray detectors. Seventeen mm of acrylic was placed to simulate soft tissue radiation scatter. The focus to imaging plate was set at 25 cm. No variation in distance or filtration was applied. The acceptable exposure range is subject to variation depending on definition and the specimen employed so the results should be considered to be relative rather than absolute. For this study the exposure latitude was determined by expert consensus using clear discrimination of the enamel-dentin junction as the criterion
4 488 Farman and Farman April 2005 Table I. Detector specifications provided by manufacturers Detector Pixel size (mm) Technology Software CDR CMOS CDR for DICOM Windows CDR Wireless CMOS CDR for DICOM Windows CygnusRay MPS CCD CygnusMedia 3.0 Dexis CCD DEXIS Software 3.01 Dixi 2 v CCD Dimaxis Pro DSX USB CCD Owandy/Julie RV2000 DSX Évolution CCD Owandy/Julie RV2000 Sigma CCD CliniView 5.1 Sidexis CCD Sirona Sidexis XG RVG-ui CCD Trophy Windows 5.05 RVG CMOS Kodak Windows RVG CMOS Kodak Windows ViperRay M CCD Vipersoft 4.0 Visualix HDI CCD VixWin 2000 Visiodent RSV CCD RSV Imaging XP DenOptix Scan Pitch PIP VixWin 2000 ScanX Scan Pitch PIP EagleSoft 9.10 InSight NA Silver halide NA Table II. Measured spatial resolution (lp/mm) CDR 9 CDR Wireless 9 CygnusRay MPS 8 Dexis 11 Dixi2 v3 High resolution mode $ 16 Dixi2 v3 Normal resolution mode 11 DSX DSX 730 Évolution High resolution mode 9 DSX 730 Évolution Low resolution mode 8 Sigma 11 Sidexis \10 RVG-ui High resolution mode $ 20 RVG-ui High sensitivity mode 12 RVG 6000 $ 20 RVG 5000 $ 14 ViperRay 7 Visualix HDI 11 Visiodent RSV 6 DenOptix 600 dpi scan 11 DenOptix 300 dpi scan 7 DenOptix 150 dpi scan 5 ScanX Very high resolution scan 9 horiz; 13 vertical ScanX High resolution scan 8 horiz; 10 vertical ScanX Standard resolution scan 7 InSight film $ 20 for the minimum acceptable exposure and cervical burnout, pixel blooming, or, depending on which came first, the maximum exposure attainable with the x-ray generator as dictating the upper limit. RESULTS Spatial resolution Spatial resolution measured in lp/mm is listed for each of the systems in Table II. The highest spatial resolutions ($ 20 lp/mm) were found for InSight film, the RVG-ui Table III. Practical exposure ranges for human tooth specimens with 17-mm acrylic attenuation to simulate soft tissues Detector Mode (if applicable) Acceptable range (mgy) Latitude ratio CDR Min-585* [4.5:1 CDR Wireless Min-429* [3.3:1 CygnusRay MPS Min-509* [4:1 Dexis :1 Dixi2 v3 HR Min-920* [7:1 Normal Min-920* [7:1 DSX USB :1 DSX Évolution High :1 Low :1 Visualix HDI Min-669 [5.2:1 Visiodent RSV :1 Sigma Min-458* [3.6:1 Sidexis Min-458* [3.6:1 RVG-ui HR Min-1308* [10.2:1 HS Min-882* [6.9:1 RVG 6000 Min-2648* [20.7:1 RVG * 19.8:1 ViperRay :1 DenOptix 300 dpi :1 ScanX IVS :1 IS 442->5720 y [12.9:1 InSight :1 *Minimum exposure using standard x-ray generator produced diagnostic image. Actual minimal exposure not determinable under experimental conditions. y Image not saturated at maximum exposure possible from x-ray generator. used in high-resolution (HR) mode, and the RVG 6000 detectors. The Planmeca Dixi2 v3 (high-resolution mode) achieved $ 16 lp/mm using a detector and software supplied and installed by the manufacturer (and was substantially superior to a previously tested Dixi2 v2). Resolution of 10 lp/mm or better was also determined for
5 Volume 99, Number 4 Farman and Farman 489 the Dexis, Dixi2 v3 (normal mode), DSX 730-USB (but not Évolution), Sigma, Kodak RVG-ui in high sensitivity mode, Kodak RVG 5000, Visualix HDI, DenOptix (600 dpi scan), and ScanX (very high resolution scan). Contrast perceptibility Without added highlight or color filters, only the CDR wired and wireless detectors, the Kodak RVG-ui, RVG 6000 and RVG 5000, and the Visualix HDI proved capable of detecting wells of 0.1 mm depth through 7 mm of aluminum. With added highlight filtration this was also possible using the Dixi2 v3. With added pseudocolor, this was also achieved by the AT ScanX in standard (lowresolution) mode and by the DSX 730 Évolution (lowresolution mode). All other detectors were able to detect wells of 0.2 mm or better within 7 mm of aluminum. Practical exposure range Table III lists the practically achievable ranges in acceptable exposures for each system tested using an unmodified commercially available standard dental x-ray generator. The mean minimum entrance exposure achieved under these conditions was 128 mgy. The widest ranges (and highest acceptable dosages permitting a diagnostic image) were found with the photostimulable phosphor systems and the RVG 6000 and RVG 5000 CMOS detectors. DISCUSSION Essential properties for a imaging system should include the following: (1) the image produced is of diagnostic quality; (2) the radiation dose is equal or reduced compared to that needed to expose analog film; (3) that digital radiology techniques are compatible with conventional x-ray generators; (4) that lossless archiving is allowed in an image file format that promotes interoperability within the DICOM standard; and (5) the time required for the total procedure should be equal to or less than with film. The present investigation was limited to the first 3 of these ideal property ingredients. The detector need not be identical to film in all image properties; however, it must provide sufficient information for the diagnostic challenges at hand. The information needed might be task dependent: for example, high spatial resolution could well be needed to accurately measure the lengths of fine endodontic instruments, but might not be critical to the detection of proximal surface dental caries. At this time there is a surprising dearth of information on the resolution and contrast requirements for detection of different dental structures and disease processes both for analog film radiographs and for their digital counterparts. It should be remembered that the ideal detector should be able to multitask and therefore both sufficient spatial resolution and good contrast discrimination are needed. Irrespective of the stored image bit-depth (which varied from 8 to 16), all 18 detectors studied showed excellent contrast resolution in terms of perceptibility of low-contrast details. However, the image display for the digital images was a standard 8-bit computer monitor. Substantial variations were determined for measured spatial resolution; however, the task-relevance of the measured spatial resolutions remains to be determined by carefully designed scientific trials. Only the Kodak RVGui, RVG 6000 CMOS detector, and InSight film proved capable of demonstrating 20 lp/mm or finer spatial resolution, with Planmeca Dixi2 v3 closely following at $ 16 lp/mm. It should be noted that some manufacturers advertised claims of theoretical resolution based on pixel dimensions were not achieved in practice. Theoretical spatial resolution estimates based on picture element size do not provide consistently useful information regarding the actual spatial resolution of the system. The question of whether newer detectors are compatible with existing dental x-ray generators can be answered both in the affirmative and in the negative. Yes, an acceptable diagnostic image could be achieved by all of the detectors with an unmodified dental x-ray generator. However, without increasing filtration or distance the minimum exposure for many solid-state detectors was not determinable. This suggests that additional radiation dosage savings might well be possible using generators capable of achieving lower exposures than commonly in present usage in combination with digital detectors. CONCLUSIONS Most current x-ray detectors generally perform well in terms of spatial and contrast resolutions, and in terms of exposure latitude. These findings were independent of the modality applied. Dental x-ray generators should be modified to provide lower exposure options when using digital x-ray detection systems for dentistry. REFERENCES 1. Farman AG. Fundamentals of image acquisition and processing in the digital era. Orthod Craniofacial Res 2003;6(Suppl 1): Molteni R. Effect of visible light on photo-stimulated-phosphor imaging plates. In: Lemke HU, Vannier MW, Inamura K, Farman AG, Doi K, Reiber JHC, editors. Computer-assisted radiology and surgery. Proceedings of the 17th Congress and Exposition; 2003 Jun London, UK. Amsterdam: Elsevier; p Reprint requests: Allan G. Farman, BDS, PhD, DSc, MBA Professor of Radiology and Imaging Sciences School of Dentistry The University of Louisville Louisville, KY [email protected]
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Integration of multiple direct digital imaging sources in a picture archiving and communication system (PACS)
(2003) 32, 337 342 q 2003 The British Institute of Radiology http://dmfr.birjournals.org TECHNICAL REPORT Integration of multiple direct digital imaging sources in a picture archiving and communication
Dental Radiography Core Subject. Digital Radiography
Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
Volume 3, Issue 4 US $6.00
Volume 3, Issue 4 US $6.00 Editor: Allan G. Farman, BDS, PhD (odont.), DSc (odont.), Diplomate of the American Board of Oral and Maxillofacial Radiology, Professor of Radiology and Imaging Sciences, Department
Digital Imaging & Communications in Dentistry: DICOM and Interoperability
2 nd International Congress, Budapest, 2005 Digital Imaging & Communications in Dentistry: DICOM and Interoperability Allan G. Farman ADA Representative to the DICOM Standard Committee Co-Chair Chair DICOM
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging
Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging Gail F. Williamson, RDH, MS Continuing Education Units: 4 hours Online Course: www.dentalcare.com/en-us/dental-education/continuing-education/ce350/ce350.aspx
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational
In Practice: How Going Digital Will Affect the Dental Office -- Farman et al. 139 (3): 14S -- The Journal of the American Dental Association
QUICK SEARCH: Author: Keyword(s): [advanced] HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS Year: Vol: Page: J Am Dent Assoc, Vol 139, No suppl_3, 14S-19S. 2008 American Dental Association
I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit
I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete
X-RAY IMAGING Emerging Digital Technology - CMOS Detectors
Application Note Case Study Technology Primer White Paper X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Image Sensors X-Ray DETECTORS Scanners Image Processing Custom Solutions In all domains
An Quantitative Evaluation of the Canadian Forces New Radiology Inspection Systems for the Detection of Water Ingress in CF188 Flight Control Surfaces
17th World Conference on Nondestructive Testing, 25-28 Oct 2008, Shanghai, China An Quantitative Evaluation of the Canadian Forces New Radiology Inspection Systems for the Detection of Water Ingress in
Quality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
EVA. Digital X-Ray System. Exceptional Image Quality Ideal For Any Practice Full Integration Best Supported Sensor Great Value
EVA Digital X-Ray System Exceptional Image Quality Ideal For Any Practice Full Integration Best Supported Sensor Great Value ImageWorks Generations of Imaging www.imageworkscorporation.com Generations
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
Digital radiography conquers the veterinary world
Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing
RVG 6500. Innovation set free.
RVG 6500 Innovation set free. Several steps ahead Introducing the RVG 6500 system the first RVG sensor to offer Wi-Fi compatibility while maintaining the same best-in-class image quality you trust. The
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
RVG 6200. Full range imaging for diagnostic confidence ROBUST DESIGN SUPERIOR ERGONOMICS NEW IMAGE PROCESSING TOOLS ENHANCED WORKFLOW
ROBUST DESIGN SUPERIOR ERGONOMICS NEW IMAGE PROCESSING TOOLS RVG 6200 ENHANCED WORKFLOW BROAD EXPOSURE RANGE SERVICE MADE SIMPLE Full range imaging for diagnostic confidence WITH TECHNOLOGY THAT WORKS
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
QUALITY ASSURANCE IN DENTAL RADIOGRAPHY. Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000
QUALITY ASSURANCE IN DENTAL RADIOGRAPHY Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000 Quality Assurance (QA) Ensure consistently adequate diagnostic information whilst keeping doses
Systematic X-ray diagnostics as flexible and individual as your dental practice. XIOS PLUS the intraoral sensor system
C A D / C AM SYSTEMS INSTRUMENT S HYGIENE SYSTEMS TRE ATMENT CENTERS IM AGING SYSTEMS XIOS PLUS the intraoral sensor system Systematic X-ray diagnostics as flexible and individual as your dental practice.
X-Mind. Instinct for perfection
X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in
DR Retrofit Kit 1417 WL Configuration Integrated PACS
DR Retrofit Kit 1417 WL Configuration Integrated PACS Description: The DR Retrofit Kit 1417WL - Configuration Integrated PACS makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films,
Remote RadEye Product Family
Remote RadEye Product Family Overview The Remote RadEye x-ray camera provides the ultimate flexibility in design and product options for your complex imaging applications. Our unique camera design separates
DENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
Digital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
Digital Dental X-ray. Midmark Animal Health Products
Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
Integrated Management & Educational Consultancy
Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:[email protected]
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
ProOne Digital X-ray Imaging Hardware Diagrams and System Requirements
Digital X-ray Imaging Hardware Diagrams and System Requirements Digital X-Ray Note: An Uninterrupted Power Supply (UPS) / Power Conditioner is recommended for the X-Ray. Install a new Planmeca Power Conditioner
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS
I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number
EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
Oftentimes, as implant surgeons, we are
CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the
Digital panoramic X-ray for practical diagnostics
ORTHOPHOS XG 3 instruments treatment centers Imaging systems cad/cam systems ORTHOPHOS XG 3 Digital panoramic X-ray for practical diagnostics System concept Standard panoramic X-ray with proven technology.
Checklist for the Qualification of Digital Detector Array Systems
Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction
We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement
MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
Medical imaging monitors specification guidelines
Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis David Bernard & Steve Ainsworth Dage Precision Industries Abstract Non-destructive testing during the manufacture of
Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.
Veraview IC5 HD High definition, digital imaging excellence Thinking ahead. Focused on life. Thinking ahead. Focused on life. 2 3 IC5 HD Super High-Speed with High Definition Clarity For dental radiology
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
Radiation safety in dental radiography
Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum
NON-DESTRUCTIVE TESTING OF PIPELINES. L. Annila. Abstract
NON-DESTRUCTIVE TESTING OF PIPELINES L. Annila Abstract This paper shall present different, contemporarily available non-destructive testing (NDT) methods of pipelines and compare them to each other from
EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams
EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully
WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter?
WHITE PAPER www.basler-ipcam.com Are More Pixels Better? The most frequently asked question when buying a new digital security camera is, What resolution does the camera provide? The resolution is indeed
AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy
Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both
The Trade-off between Image Resolution and Field of View: the Influence of Lens Selection
The Trade-off between Image Resolution and Field of View: the Influence of Lens Selection I want a lens that can cover the whole parking lot and I want to be able to read a license plate. Sound familiar?
Systematic X-ray diagnostics as flexible and individual as your dental practice.
CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS XIOS PLUS THE INTRAORAL SENSOR SYSTEM Systematic X-ray diagnostics as flexible and individual as your dental practice. T h
HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950
HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple
Scanners and How to Use Them
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types
Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose!
Put Your Panoramic Imaging on Steroids While Reducing the Patient s Dose! Dale A. Miles BA, DDS, MS, FRCD(C) Diplomate, American Board of Oral and Maxillofacial Radiology It used to be that panoramic images
Patient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
The Convenient and Compact CR System for Veterinary Practices
The Convenient and Compact CR System for Veterinary Practices Fire CR+ 20 / 40 / 60 / 80 vet with a throughput of 20 to 70 cassettes per hour and the option of a space - saving wall mount The small and
KODAK Dental Radiography Series. Quality Assurance in Dental Radiography DENTAL PRODUCTS
KODAK Dental Radiography Series Quality Assurance in Dental Radiography DENTAL PRODUCTS Quality Assurance in Dental Radiography A complete quality assurance program is easy to establish and maintain. Quality
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
WHO CARES ABOUT RESOLUTION?
WHO CARES ABOUT RESOLUTION? At Poli we certainly do not care about resolution as a concept! However we do care deeply about resolution when used as a measure of quality. Resolution is a very dry and technical
www.siemens.com/agile Luminos Agile A RADical new way in fluoroscopy Answers for Life
www.siemens.com/agile Luminos Agile A RADical new way in fluoroscopy Answers for Life 2 Easy patient transfer with table height adjustment from 65 cm -112 cm (26 "- 44 ") True dual-use capability for fluoroscopy
Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices
IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists
DAISY. A single solution for your practice. DAISY 1-800-368-6401. www.dmcdental.com
Practice management software and business services for the modern dentist. www.dmcdental.com 1-800-368-6401 A single solution for your practice. is a product of Dentists Management Corporation 10505 S.E.
Data. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
DryView CHROMA Imaging System
DryView CHROMA Imaging System VERSATILE PRINTING ON FILM OR PAPER One of the most trusted names in medical printing now offers a highly versatile printer that empowers physicians and other healthcare professionals
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
The convenient and compact CR system for veterinary practices
Divario CR-Fvet with dicom PACS DX-R Software CR Systems for the Future The convenient and compact CR system for veterinary practices Divario CR-F20/ F40/ F60/ F80 vet Divario CR-F20/ F40/ F60/ F80 vet
European Academy of DentoMaxilloFacial Radiology
European Academy of DentoMaxilloFacial Radiology Framework for Specialist Training in Dental and Maxillofacial Radiology Background The scope of DentoMaxilloFacial Radiology DMFR (Dental and Maxillofacial
X-ray 100KV. Speed as compared to that of type IX100 used as a standard 100 under each exposure conditions.
DATA SHEET INDUSTRIAL RADIOGRAPHIC SYSTEMS RADIOGRAPHIC FILM IX20, IX25, IX50, IX80, IX, IX, IX29 AND IX59 Fujifilm industrial X-ray films feature revolutionary new film technology. The combination of
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
16 th IOCCG Committee annual meeting. Plymouth, UK 15 17 February 2011. mission: Present status and near future
16 th IOCCG Committee annual meeting Plymouth, UK 15 17 February 2011 The Meteor 3M Mt satellite mission: Present status and near future plans MISSION AIMS Satellites of the series METEOR M M are purposed
Economic Evaluation of Conventional Radiography with Film and Computed Radiography: Applied at BMC
Advances in Computed Tomography, 2012, 1, 23-29 http://dx.doi.org/10.4236/act.2012.13006 Published Online December 2012 (http://www.scirp.org/journal/act) Economic Evaluation of Conventional Radiography
Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral
Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
AxioCam HR The Camera that Challenges your Microscope
Microscopy from Carl Zeiss AxioCam HR The Camera that Challenges your Microscope Documentation at the edge of the visible The Camera for Maximum Success: AxioCam HR Low light fluorescence, live cell imaging,
VistaScan Perio Plus Image plate scanner more time for patient care
VistaScan Perio Plus Image plate scanner more time for patient care Compressed air Suction Imaging Dental care Hygiene Efficiency for work in the dental practice Up to 8 images at once Digital X-ray with
The professional for your diagnostics.
CAD/CAM SYSTEMS INSTRUMENTS HYGIENE SYSTEMS TREATMENT CENTERS IMAGING SYSTEMS DIGITAL PANORAMIC X-RAY MADE SIMPLE The professional for your diagnostics. System concept ORTHOPHOS XG 5 completes the next
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2
Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for
White Paper. "See" what is important
Bear this in mind when selecting a book scanner "See" what is important Books, magazines and historical documents come in hugely different colors, shapes and sizes; for libraries, archives and museums,