Dimethylamylamine: A Drug Causing Positive Immunoassay Results for Amphetamines *
|
|
|
- Ralf Harrington
- 10 years ago
- Views:
Transcription
1 Dimethylamylamine: A Drug Causing Positive Immunoassay Results for Amphetamines * Technical Note Shawn P. Vorce, Justin M. Holler, Brian M. Cawrse, and Joseph Magluilo, Jr. Division of Forensic Toxicology, Armed Forces Medical Examiner System, Armed Forces Institute of Pathology, 1413 Research Boulevard, Building 102, Rockville, Maryland Abstract The Department of Defense (DoD) operates six forensic urine drug-testing laboratories that screen close to 5 million urine samples for amphetamines yearly. Recently, the DoD laboratories have observed a significant decrease in the confirmation rates for amphetamines because of specimens screening positive by two separate immunoassays and confirming negative by gas chromatography mass spectrometry (GC MS). Previous studies conducted by the Division of Forensic Toxicology, Armed Force Institute of Pathology (AFIP) utilizing a GC MS basic drug screen and a designer drug screen revealed no common compound or compound classes as to the cause of the immunoassay-positive results. Additional information obtained from an immunoassay vendor suggested the anorectic compound dimethylamylamine (DMAA) may be the cause of the false-positive screens. An additional 134 false-positive samples were received and analyzed using liquid chromatography tandem mass spectrometry (LC MS MS) for DMAA. LC MS MS analysis revealed the presence of DMAA in 92.3% of the false-positive samples at a concentration of approximately 6.0 mg/l DMAA, causing a positive screen on both immunoassay kits. ephedrine (2). The nutritional supplements may list DMAA in the ingredients as any of the previously mentioned names, or as geranium oil extract, or as a proprietary blend. DMAA is available in over-the-counter party pills sold in New Zealand, and in November 2009, the government moved to restrict sales of those pills. Their use as a drug of abuse became prevalent in New Zealand after 1-benzylpiperazine (BZP) became a scheduled drug (3). The Department of Defense (DoD) currently employs a three test system to report a positive result for a urine specimen. The tests are composed of two qualitative immunoassays and one confirmatory test, usually gas chromatography mass spectrometry (GC MS). For the amphetamines class, the DoD uses two different immunoassay reagents to improve selectivity and decrease over-the-counter medication positives from being ex- Introduction Dimethylamylamine (DMAA) is a straight chain aliphatic amine (Figure 1) found naturally in geranium flowers. It is also referred to as forthane, methylhexaneamine, 1,3-dimethylpentylamine, and geranamine. DMAA was originally used as a nasal decongestant for its vasoconstrictor action on the nasal mucosa (1). Today, it can be found in nutritional and bodybuilding energy supplements such as Jack3d and OxyELITE Pro that are available online and at health supplement suppliers such as General Nutrition Center (GNC). One manufacturer refers to DMAA as being a low-side-effect alternative to * Disclaimer: The opinions or assertions contained herein are the private views of the authors and are not to be construed as official or as reflecting the views of the Department of Defense, the Army, Navy, or Air Force. Author to whom correspondence should be addressed. [email protected]. Figure 1. Structure of dimethylamylamine (DMAA) as related to amphetamine (AMP), methamphetamine (METH), 3,4-methoxyamphetamine (MDA), and 3,4-methylenedioxymethamphetamine (MDMA). Reproduction (photocopying) of editorial content of this journal is prohibited without publisher s permission. 183
2 tracted for confirmation. In July 2009, the confirmation rates began to decrease at several laboratories. The confirmation rates at DoD drug-testing laboratories for amphetamines were 82.3 and 81.2%, respectively, for fiscal years 2008 and To date (through June 2010), the confirmation rate for amphetamines in fiscal year 2010 is 50.4%. One laboratory reported confirmation rates as low as 23%. Some laboratories have implemented a third screening assay, which has shown improvement in confirmation rates (all three immunoassays must be positive for confirmation analysis to proceed). Low confirmation rates cost the DoD laboratories time, money, and material, as well as challenge DoD turnaround time requirements. Testing at the Division of Forensic Toxicology (DFT), The Armed Forces Medical Examiner System (AFMES) was conducted in late 2009 on 52 specimens that screened positive and confirmed negative for amphetamines. The specimens were analyzed by an alkaloid drug screen and a designer drug screen. The results did not indicate a common denominator as to the cause of the positive immunoassay results. At the annual DoD drug-testing program meeting, a representative from Siemens stated DMAA may be the cause of the positive immunoassay results. The DFT requested additional specimens that screened positive and confirmed negative to analyze specifically for DMAA. Experimental Reagents and materials All organic solvents were high-performance liquid chromatography (HPLC) grade and purchased from Fisher Scientific (Pittsburgh, PA). Potassium hydroxide pellets were also purchased from Fisher Scientific. Formic acid and DMAA were purchased from Aldrich (Milwaukee, WI). A methanolic standard of amphetamine-d 8 was purchased from Cerilliant (Round Rock, TX). Immunoassay A Roche/Hitachi Modular P automated screening instrument (Indianapolis, IN) was used to screen urine samples for amphetamines. The kits used were Roche Amphetamines KIMS assay and Siemens Syva EMIT II Plus Amphetamines assay (Newark, DE). Each kit was calibrated on the Modular P analyzer using d-amphetamine spiked at 500 ng/ml with certified standards purchased from Cerilliant. Negative (75% cutoff concentration) and positive (125% cutoff concentration) controls were included in the initial calibration. Standards and calibrators preparation A stock solution of DMAA was prepared at target concentration of 1.0 mg/ml, in ethanol and stored at 20 C. A stock solution of the internal standard amphetamine-d 8 was prepared in amber glass at a target concentration of mg/ml and refrigerated. Working solutions of DMAA were prepared by serial dilution with ethanol at concentrations of 0.01 and mg/ml. Calibrators for DMAA were spiked into certified drug-free negative urine at 25, 50, 100, 250, and 500 ng/ml. DMAA sample preparation and extraction To 1 ml of urine, 100 μl of the stock internal standard solution was added for a final concentration of 100 ng/ml, 3 drops of concentrated potassium hydroxide and 3 ml of ethyl acetate were also added. The samples were mixed for 5 min and centrifuged for 5 min at 3000 rpm. The upper organic layer was transferred to clean conical tubes and evaporated at 40 C under nitrogen at 5 psi after the addition of 25 μl of 10% methanolic HCl. The samples were reconstituted in 200 μl mobile phase (3:2 0.1% formic acid/methanol), transferred to properly labeled autosampler vials, and capped. DMAA instrumental analysis The LC MS MS analysis of DMAA was performed using an Agilent 1100 series HPLC system (Palo Alto, CA) coupled with an Applied Biosystems/MDS SCIEX 3200 QTRAP (Foster City, CA) equipped with a Turbo V source. Analyst 1.5 software was used for data acquisition and analysis. Chromatographic separation was performed on an Agilent Zorbax XDB C 18 column ( mm, 3.5 μm). The column compartment was maintained at 35 C, and the injection volume was set at 2 μl. The mobile phase was set at a constant flow of 800 μl/min and consisted of 0.1% formic acid in deionized water (A) and methanol with 5% acetonitrile and 0.1% formic acid (B). A gradient elution was used as follows: pre-injection equilibration with 65% A for 3.0 min, hold at 65% A for 2.0 min after injection, ramp to 40% A over 4.0 min, and hold until 6.0 min. The MS was operated in positive electrospray ionization mode (+ESI). The analysis of DMAA and amphetamine-d 8 (ISTD) was operated in multiple reactions monitoring (MRM) acquisition mode. Two MRM transitions (* denotes quantitation transition) were monitored for both DMAA (m/z 116.1/57.0*, m/z 116.1/99.1) and amphetamine-d 8 (m/z 144.2/ 97.1*, m/z 144.1/127.2). The source-dependent parameters for the MS MS analysis of DMAA were determined by the flow injection analysis (FIA). The optimized source-dependent parameters for the analysis were as follows: GS1 gas (nebulizer) was set to 60 psi, GS2 gas Table I. Compound-Dependent MS MS Parameters MRM Compound-Dependent Parameter (volts) Transition Compound (Da/Da) DP EP CEP CE CXP DMAA 116.1/ / ISTD 144.2/ /
3 (turbo) was set to 70 psi, CUR (curtain gas) was set to 40 psi, TIS (TurboIonSpray voltage) was set to 1500V, and the source temperature was set at 550 C. The compound-dependent parameters for the MS MS analysis of DMAA were determined by direct infusion. An integrated infusion pump delivered a 10 mg/l standard solution at a constant flow (10 μl/min) directly into the TIS source. The auto-optimization process determined the optimal parameters for each MRM transition. The following parameters were optimized during the process: DP (declustering potential), EP (entrance potential), CEP (collision cell entrance potential), CE (collision energy), and CXP (collision cell exit potential). Table I lists the optimal compound-dependent parameters for DMAA and internal standard. Identification of DMAA in the random specimens was based on the ratio of MRM2/MRM1 transitions being within 20% of the average ratios and the relative retention time being within ±2% of the averages measured from the calibrators. Results and Discussion The screening results are presented in Table II. Most speci- Table II. Immunoassay Screening and Quantitation Results DMAA DMAA DMAA Roche Siemens Quant. Roche Siemens Quant. Roche Siemens Quant. Sample Screening Screening (ng/ml) Sample Screening Screening (ng/ml) Sample Screening Screening (ng/ml) T ND* T G T T G T T G T T G T T ND G T T G T T ND G T ND T G T T ND G T T G T T G T T G T T G T T G T ND T G T T G T B G T B G T B G T B G T B G T B G T B G T B G T B G T ND B G T B G T B G T B G T B G T ND B G T B G T B G T B G T B G T B G T B G T B G T B G T B G T G G T G G T G ND G T G G * None detected. 185
4 mens screened positive with both immunoassays, although some did not, which was probably due to sample degradation. The confirmation results for DMAA are also presented in Table II. The confirmation method was developed solely to detect and quantitate DMAA. Overall, 92.3% of the specimens contained DMAA at or above 2.5 mg/l. Figure 2 illustrates the screening results from the two immunoassay kits. The charts plot DMAA concentrations versus immunoassay screening response for each immunoassay. It was of interest to determine what concentration of DMAA alone would cause a positive immunoassay result. Certified negative drug-free urine was spiked with DMAA to determine the lowest concentration that yields a positive immunoassay result. The Roche Amphetamines KIMS assay and Siemens Syva EMIT II Plus Amphetamines assay gave positive responses at 7500 and 3125 ng/ml, respectively. The concentrations determined experimentally correlate fairly well to the real-life specimens analyzed in the study. The DMAA concentrations in the analyzed specimens range from 2.5 to 67.0 mg/l with 6.9 mg/l being the lowest concentration to give two positive immunoassay results using real urine samples. The specimens that confirmed negative for DMAA were subjected to a basic drug screen analysis to determine if there were any drugs present that could explain the positive immunoassay results. Each of the nine urine samples that confirmed negative for DMAA contained compounds known to cross-react with amphetamine immunoassays used by the DoD laboratories. Specifically, six of the negative DMAA urine samples contained phentermine, two contained bupropion and its metabolites, and one contained a high concentration of pseudoephedrine. The cross-reactivity of immunoassay kits is affected by the specific coupling sites for the protein used in the assay. For amphetamines, the protein could be coupled to either the aromatic ring or the nitrogen, or a combination of both. The Siemens and Roche kits both target d-methamphetamine and d-amphetamine with some cross-reactivity to MDMA and MDA. Figure 1 illustrates the structural similarity between DMAA and the sympathomimetic amines that the immunoassay kits are designed to detect. Because DMAA is a small molecule with structural similarities to the amphetamines, it is likely to cross-react with any immunoassay targeting amphetaminetype compounds. Previous studies conducted at AFIP, DFT using GC MS full scan analysis failed to detect DMAA. The Figure 2. Comparison of the screening responses for the two immunoassay kits used in the study. The charts plot the concentration of DMAA (ng/ml) versus the immunoassay response with 100 as the cutoff response. All samples that are plotted above the cutoff line are false-positive amphetamine screens. Figure 3. GC MS splitless injection of a DMAA reference standard. DMAA is not detected because it is buried in the solvent front (A). GC MS split injection (100:1) of a DMAA reference standard. DMAA is detected as a double peak because it is a racemic mixture (B). LC MS MS of DMAA and the amp-d 8 internal standard. DMAA is detected as a double peak (C). 186
5 standard alkaline full scan GC MS drug screen has a 4.0-min solvent delay to protect the life of the filament and electron multiplier. However, it was discovered that DMAA eluted in 2.0 min along with the solvent peak. A study was performed varying the GC inlet and oven parameters to determine the effects on DMAA detection. A series of injections performed while varying the inlet temperature from 120 to 270 C revealed no discernable thermal degradation of DMAA in the injection port. On a J&W DB-5MS column (20 m 0.18 mm 0.18 μm), DMAA is only retained 2.0 min at 50 C. In order to perform a GC MS analysis, the solvent delay must be set before 2.0 min, and the split vent must be greater than 50:1. This eliminates most of the solvent before it gets onto the column and will allow detection of the DMAA peak. GC MS analysis is feasible, but great care must be taken during the method development process for the initial temperature, injection parameters, and the solvent delay. The GC MS and LC MS MS analyses of DMAA result in a double chromatographic peak (Figure 3). DMAA has two chiral centers, which will result in four possible stereoisomers: (R,S)-, (S,R)-, (S,S)-, and (R,R)-1,3-dimethylpentylamine. The (S,S) and (R,R) stereoisomers are enantiomers (optical isomers) that have the same chemical and physical properties and cannot be separated. The same is true for (R,S) and (S,R). The [(S,S), (R,R)] and [(R,S), (S,R)] isomers are diastereomers that differ in some physical properties and can be separated. The smaller the distance between the optical centers, the better the chromatographic separation. For DMAA, the two methyl groups are at the C1 and C3 carbon positions, which result in a double chromatographic peak that is almost baseline resolved. The first peak is from [(S,S), (R,R)] and the second peak is from [(R,S), (S,R)] isomers (4 7). Both the reference standard and the positive urine specimens had the double peak, indicating that DMAA is made and sold as a racemic compound. The quantitative results were calculated on the total area of both peaks as compared to a standard curve for DMAA. Conclusions Products containing DMAA have become an issue for the DoD drug-testing program. The cost for confirmation testing of the specimens containing DMAA can and has become quite substantial. DMAA is a straight chain amine with a fairly simple structure that shows reactivity with the antibodies currently employed in some commercially available immunoassays. The overall safety of DMAA should be explored to determine if this is a safer alternative to other sympathomimetic amines. Highthroughput urine drug-screening laboratories need to be aware of the impact DMAA can have on their testing efficiency. Acknowledgments This work was funded in part by the American Registry of Pathology, Washington, D.C The authors would like to acknowledge the help of Josh Seither, Jeff Chmiel, and Laura Regester. The authors would also like to thank Mr. Robert Privon of Siemens for suggesting DMAA as a potential source of the screen-positive results. Additionally, the authors acknowledge the DoD drug-testing laboratories United States Army Forensic Toxicology Drug Testing Laboratory Tripler, Navy Drug Screening Laboratory Great Lakes, and Air Force Drug Testing Laboratory Brooks for their assistance in obtaining the 134 urine specimens. References 1. New and nonofficial remedies: methylhexamine; forthane. J. Am. Med. Assoc. 143(13): 1156 (1950) Government restricts sales of party pill drug DMAA J.W. Westley, B. Halpern, and B.L. Karger. Effect of solute structure on separation of diastereoisomeric esters and amides by gasliquid chromatrography. Anal. Chem. 40: (1968). 5. J. Gal. Stereochemistry of metabolism of amphetamines: use of ( )-alpha-methoxy-alpha-(trifluoromethyl)phenylacetyl chloride for GLC resolution of chiral amines. J. Pharm. Sci. 66: (1977). 6. B.D. Paul, J. Jemionek, D. Lesser, A. Jacobs, and D.A. Searles. Enantiomeric separation and quantitation of (±)-amphetamine, (±)-methamphetamine, (±)-MDA, (±)-MDMA, and (±)-MDEA in urine specimens by GC EI-MS after derivatization with (R)-( )- or (S)-(+)-α-methoxy-α-(trifluoromethyl)phenylacetyl chloride (MTPA). J. Anal. Toxicol. 28(6): (2004). 7. J.M. Holler, S.P. Vorce, T.Z. Bosy, and A. Jacobs. Quantitative and isomeric determination of amphetamine and methamphetamine from urine using a nonprotic elution solvent and R( )-α-methoxyα-trifluoromethylphenylacetic acid chloride derivatization. J. Anal. Toxicol. 29(7): (2005). Manuscript received September 13, 2010; revision received October 4,
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504
Validation of Analysis of Amphetamines, Opiates, Phencyclidine, Cocaine, and Benzoylecgonine in Oral Fluids by Liquid Chromatography Tandem Mass Spectrometry Subbarao V. Kala*, Steve E. Harris, Tom D.
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS
SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS Ingrid J. BOSMAN 1, Douwe DE BOER 2, Robert A. A. MAES 1 1 Department of Human
Navy Drug Screening Laboratory Jacksonville Screening News
Navy Drug Screening Laboratory Jacksonville Screening News July 2011 Volume 2, Number 4 This Issue CO s Desk In Focus Discrepancies of the Month Drug Facts Ask the Expert CO s Desk Presently, more than
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
Opiates in Urine by SAMHSA GC/MS
application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Supplementary Materials and Methods (Metabolomics analysis)
Supplementary Materials and Methods (Metabolomics analysis) Metabolite extraction and mass spectrometry analysis. 12 Vldlr-/- knock out and 12 wild type (WT) retinas were separately frozen at -80 ºC in
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in uman Serum Technical verview Clinical Research Authors Anthony Macherone, Ph.D. Agilent Technologies,
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine
Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine Application Note Authors Peter JW Stone and Kevin McCann Agilent Technologies, Inc. Santa Clara,
Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis
Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis 1.0 Purpose - This procedure specifies the required elements for the solid phase extraction of acidic,
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS *
MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS * Ingrid J. BOSMAN 1, Douwe DE BOER 2, Edwin B. SIDERIUS 1, Lesseps J. A. L. DOS REYS 2, Robert A. A. MAES 1 1
AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT
AppNote 9/2013 Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides Oscar G. Cabrices Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS
AppNote 1/2012 Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS Oscar G. Cabrices, Fred D. Foster, John R. Stuff, Edward
Analysis of Liquid Samples on the Agilent GC-MS
Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >
CORESTA RECOMMENDED METHOD N 72
CORESTA RECOMMENDED METHOD N 72 DETERMINATION OF TOBACCO-SPECIFIC NITROSAMINES IN SMOKELESS TOBACCO PRODUCTS BY LC-MS/MS (February 216). INTRODUCTION The CORESTA Smokeless Tobacco Sub-Group studied various
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
TECHNICAL NOTE Toxicology Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT
AppNote 7/2009 Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS Fred D. Foster, John R. Stuff, Edward A. Pfannkoch Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum,
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS
htslabs.com [email protected] 800 541.4792 760 757.8080 760 757.9367 Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS Nadine Koenig 2,
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection LGC/R/2011/181 LGC Limited 2011 GC11/0005A A feasibility study of the determination of B-group
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
AppNote 6/2011. Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS KEYWORDS ABSTRACT
AppNote 6/2011 Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS Fred D. Foster, John R. Stuff, Edward A. Pfannkoch Gerstel, Inc., 701 Digital Dr. Suite
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
This is the written version of our Hot Topic video presentation available at: MayoMedicalLaboratories.com/hot-topics
This is the written version of our Hot Topic video presentation available at: MayoMedicalLaboratories.com/hot-topics Welcome to Mayo Medical Laboratories Hot Topics. These presentations provide short discussion
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT
AppNote 1/2014 Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine Fred D. Foster, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J,
Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
New drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
NUVISAN Pharma Services
NUVISAN Pharma Services CESI MS Now available! 1st CRO in Europe! At the highest levels of quality. LABORATORY SERVICES Equipment update STATE OF THE ART AT NUVISAN CESI MS Now available! 1st CRO in Europe!
Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
APPLICATIONS MANUAL. Ifosfamide in blood serum... 2. Ingredients in blood serum... 3. Organophosphorus pesticides in tea leaf... 5
1 APPLICATIONS MANUAL CONTENTS Ifosfamide in blood serum... 2 Ingredients in blood serum... 3 Organophosphorus pesticides in tea leaf... 5 Sudan i ii iii iv in chilli sauce... 6 Pah in water... 7 Phenols
Instrumentation. Sample pretreatment
Analysis of Propofol (2,6-Diisopropylphenol) and its Metabolites in One Injection using the DuoSpray Ion Source AB SCIEX QTRAP 500 LC-MS/MS System and Gerstel, Inc. MultiPurpose Sampler Adrian M. Taylor,
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*
Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites
Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract
Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester
Quantitation of Drugs in Dried Bloodstains Thomas Meyer Anna Gomenyuk Nadiah Lester Objective To find the quantitative value of drugs in dried bloodstains Currently Cloth With Dried Bloodstain Containing
Urine Drug Testing. Why drug test? Set the Standard!
Urine Drug Testing Urine drug testing is a valuable tool for the physician when making treatment decisions. Patients and physicians benefit from safe and effective treatment of pain but there are always
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit Fast, simple and cost effective analysis Many areas of biochemical research and
SUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
Confirmation by Triple Quadrupole LC/MS/MS for HHS-compliant Workplace Urine Drug Testing
Confirmation by Triple Quadrupole LC/MS/MS for HHS-compliant Workplace Urine Drug Testing A complete applications solution for forensic toxicology laboratories John Hughes Agilent Technologies Pleasanton,
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.
Q-TOF User s Booklet Enter your username and password to login. After you login, the data acquisition program will automatically start. ~ 1 ~ This data acquisition window will come up after the program
How To Develop A Recipe Card
TSNA Method Development Dr Michael Intorp CORESTA Cooperation Center for Scientific Research Relative to Tobacco FDA Center for Tobacco Products Scientific Workshop Tobacco Product Analysis 30 31 July
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS
Application note Explosives by Direct Injection LC/MS/MS Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS API 5000 LC/MS/MS System
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Chemical analysis service, Turner s Green Technology Group
Chemical analysis service, Turner s Green Technology Group Hourly costs, Academic collaborations leading to co-publications: 627 SEK/hr (excl. VAT) Hourly costs, Industry/Academic collaborations, no publications:
Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2 1 OpAns LLC, Durham, NC, USA; 2 Ionics Mass Spectrometry
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns
Hydrophilic-Interaction Chromatography (HILIC) for LC-MS/MS Analysis of Monoamine Neurotransmitters using XBridge BEH Amide XP Columns Jonathan P. Danaceau, Kenneth J. Fountain, and Erin E. Chambers Waters
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS Mikaël LEVI Shimadzu France Welcome Seminar Barcelona 2013 Interest in Steroids assay by LC-MS/MS Steroids measurement is essential
Central Government Laboratory Department of Health and Environment
Central Government Laboratory Department of Health and Environment Overview who we are and what we do What is available and what is protocol for good science Sample collection Screening Tests Confirmation
Discovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan