Lesson 4 ECG 2 Electrocardiography

Size: px
Start display at page:

Download "Lesson 4 ECG 2 Electrocardiography"

Transcription

1 Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 4 ECG 2 Electrocardiography PC running Windows XP or Mac OS X Lesson Revision BIOPAC Systems, Inc. 42 Aero Camino, Goleta, CA (805) , Fax (805) info@biopac.com

2 Page 2 Biopac Science Lab Lesson 4 ELECTROCARDIOGPAHY 2 I. SCIENTIFIC PRINCIPLES Recall from Lesson 3, ECG 1, that each cardiac cycle, or heartbeat, begins with a bioelectric impulse generated by the sinoatrial (SA) node, the pacemaker of the heart. The bioelectric impulse is conducted through the heart by the pacemaker system and travels as a wave of depolarization followed in a few thousandths of a second by a wave of repolarization. Ventricular muscle depolarization and repolarization are represented in the electrocardiogram (ECG) as the QRS complex and the T wave, respectively. The body contains fluids with ions that allow for electric conduction. This makes it possible to use electrodes on the surface of the skin to detect electrical activity in and around the heart and use an electrocardiograph to record the activity. Conveniently, the legs and arms act as simple extensions of points in the torso, allowing the recording and ground electrodes to be placed on the wrists and ankles (Fig. 4.1). - Lead I Lead II + Lead III + right arm left arm right leg (ground) left leg Fig. 4.1 Electrode placement options Fig. 4.2 Einthoven s triangle and vectorgraph The ECG is a record of the overall spread of electric current through the heart as a function of time in the cardiac cycle. The direction of polarity (+ or -) of the recorded waveforms depends upon the location of the recording electrodes on the surface of the body and whether the electrical activity during the cardiac cycle is coming toward or going away from the surface electrode. In general, as a wave of depolarization approaches a positive electrode, a positive voltage is seen by that electrode. If the wave of depolarization is traveling toward a negative electrode, a negative voltage will be seen. The term lead is defined as a spatial arrangement of two recording electrodes on the body. One lead is labeled positive (+) and the other negative (-). The electrode placements define the recording direction of the lead, which is called the lead axis or angle. The axis is determined by the direction when going from the negative to positive electrode. The electrocardiograph computes the voltage difference (magnitude) between the positive and negative electrodes and displays the changes in voltage difference with time. The standard bipolar limb leads are: Lead I = right arm (-), left arm (+) Lead II = right arm (-), left leg (+) Lead III = left arm (-), left leg (+) The three standard bipolar limb leads may be used to construct an equilateral triangle, called Einthoven s triangle, at the center of which lies the heart (Fig. 4.2, left). Each side of the triangle represents one of the bipolar limb leads, and each lead forms a 60 angle with the two opposite leads. By placing Lead I through the center of the heart and moving the axes of Lead II and Lead III horizontally until they intersect Lead I at the center of the heart, a vectorgraph can be constructed (Fig. 4.2, right). The vectorgraph is used to plot vectors and aid in visualizing the electrical axes of the heart. At any given moment during the cardiac cycle (during the QRS complex, for example), the net electrical activity seen by a lead may be represented by a vector. A vector is an entity that can be represented by an arrow of specific length pointing in a specific direction. A vector has both magnitude and direction. The length of the arrow, divided equally into 0.1 mv units, represents the vector s magnitude, and is expressed as a mean potential. The direction of the vector, in degrees, is represented by its position relative to the horizontal 0º to 180º line (Lead I) on the vectorgraph. The dominant waveform in all three bipolar limb leads is the QRS complex signifying ventricular depolarization. The mean axis of ventricular depolarization is called the mean QRS axis. The mean QRS axis is diagnostically important

3 Lesson 4: ECG 2 Page 3 because it represents the majority of the heart s electrical activity occurring during a cardiac cycle. When the bipolar limb leads are used to obtain data for plotting the vector, the mean QRS axis lies in the frontal plane because the electrical picture seen by the recording electrodes placed on the extremities lies in the frontal plane. In normal persons, the mean QRS axis is between -30 and +90 (Fig. 4.2). In the frontal plane, the mean QRS axis is often between 0 and +90. A mean QRS axis between +90 and +180 indicates right axis deviation (RAD). RAD occurs, for example, in persons with right ventricular hypertrophy (over-sized right ventricle). A mean QRS axis between -30 and -90 indicates left axis deviation (LAD). LAD is found in patients with left anterior blockage in the conduction system of the ventricle. To define the mean QRS axis precisely, the cardiologist defines it in three dimensions (X, Y and Z planes). This is done in practice by using a standardized set of 12 leads. In this lesson, you will sequentially record two of the bipolar limb leads (Lead I and Lead II) and extract the data from the ECG necessary to determine the mean QRS axis in the frontal plane. A simple way to approximate the mean QRS axis in the frontal plane is to plot the magnitude of the R wave as measured from Lead I and Lead II (Fig. 4.3). magnitude of R wave from leads 0,0 I Mean Electrical Axis III Mean Electrical Axis Fig. 4.3 Mean electrical axis In this lesson, you will record data from Lead I and Lead III connections with the Subject lying down and sitting up. II. EXPERIMENTAL OBJECTIVES 1) To record ECG from Leads I and III in the following conditions: lying down, sitting up, and breathing deeply while sitting. 2) To compare the direction of the QRS complex (+ or ) with the direction of the lead axis. 3) To estimate the mean ventricular potential. 4) To estimate the mean electrical axis of the QRS complex using two methods. III. MATERIALS Computer system (running Windows XP or Mac OS X) Biopac Science Lab system (MP40 and software) Electrode lead set (40EL lead set) Disposable vinyl electrodes (EL503), four electrodes per subject Cot or lab pillow Protractor Two different color pens or pencils

4 Page 4 Biopac Science Lab IV. EXPERIMENTAL METHODS A. Set Up EQUIPMENT SUBJECT - Lead I + right forearm WHITE lead left forearm RED lead Fig. 4.4 right leg BLACK lead (Ground) Fig. 4.5 Electrodes for Leads I & III/Lead I configuration FAST TRACK 1. Turn the computer ON. 2. Set the MP40 dial to OFF. 3. Plug the equipment in as follows: Electrode leads (40EL) MP Attach four electrodes to the Subject as shown in Fig Connect the electrode leads (40EL) for a Lead I configuration, matching lead color to electrode position as shown above. Details Attach four electrodes to the Subject to allow Lead I and Lead III configurations, as shown in Fig. 4.5: one electrode on the right leg, just above the ankle bone one electrode on the left leg, just above the ankle bone one electrode on the left wrist (same side of arm as the palm of hand) one electrode on the right wrist (same side of arm as the palm of hand). For optimal electrode response, place electrodes on the skin at least five minutes before starting the Check section. Lead I will be recorded first, so connect the leads for a Lead I configuration: white: right wrist; red: left wrist; black: right ankle. IMPORTANT Clip each electrode lead color to its specified electrode position. 6. Start the Biopac Science Lab software. 7. Choose lesson L04-ECG-2 and click OK. 8. Type in a unique file name. No two people can have the same file name, so use a unique identifier, such as the Subject s nickname or student ID#. 9. Click OK. This ends the Set Up procedure.

5 Lesson 4: ECG 2 Page 5 B. Check MP40 Check FAST TRACK Details 1. Set the MP40 dial to ECG/EOG. 2. Press and hold the Check pad on the MP Click when the light is flashing. 4. Wait for the MP40 Check to stop. 5. Let go of the Check pad. 6. Click Continue. Signal Check 7. Click Check Signal. 8. Wait for the Signal Check to stop. 9. Review the data. If correct, go to the Record section. Continue to hold the pad down until prompted to let go. The MP40 check procedure will last five seconds. The light should stop flashing when you let go of the Check pad. When the light stops flashing, click Continue. Lead I configuration is used for the signal check. Subject should be supine, relaxed, and still. The electrocardiograph is very sensitive to small changes in voltage caused by contraction of skeletal muscles, and the Subject s arms and legs need to be relaxed so that the muscle (EMG) signal does not corrupt the ECG signal. The eight-second Signal Check recording should resemble Fig If incorrect, click. Fig. 4.6 If the recording does not show a recognizable ECG waveform with R-wave peaks and no large baseline drift, click Redo Signal Check.

6 Page 6 Biopac Science Lab C. Record FAST TRACK Details 1. Prepare for the recording and have the Subject lie down and relax. SEGMENT 1 Supine, Lead I Watch the Help menu videos to prepare for the recording. You will record two segments for each lead configuration, one with the Subject lying down, and another with the Subject sitting up. In order to work efficiently, read this entire section so you will know what to do before recording. Check the last line of the journal and note the total amount of time available for the recording. Stop each recording segment as soon as possible so you don t use an excessive amount of time (time is memory). Tips for obtaining optimal data To minimize muscle (EMG) corruption of the ECG and baseline drift: a) The Subject should not talk or laugh during any of the recording segments. b) When lying down or sitting up, the Subject should be completely relaxed. c) When sitting up, the Subject s arms should be supported on an armrest (if available). d) The recording should be suspended before the Subject prepares for the next recording segment. e) The Subject should breathe normally during the recording to minimize EMG from the chest area. f) Make sure electrodes do not peel up. 2. Click. When you click Record, the recording will begin and an append marker labeled Supine, Lead I will automatically be inserted. 3. Record for 30 seconds. Subject should be supine, relaxed, and still (seconds 0-30). 4. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 5. Review the data. Your data should resemble Fig If correct, go to Step 6. If incorrect, click Redo. Recording continues Fig. 4.7 Segment 1 (Supine, Lead I) A little baseline drift when the Subject breathes in and out does not indicate incorrect data. The data is incorrect if: a. The Suspend button was pressed prematurely b. An electrode peeled up, giving a large baseline drift c. The Subject has too much muscle (EMG) artifact If the data is incorrect, click Redo and repeat Steps 1-5; the last data segment you recorded will be erased.

7 Lesson 4: ECG 2 Page 7 SEGMENT 2 Supine, Lead III 6. Disconnect the leads from the Lead I configuration and reconnect to establish a Lead III configuration (Fig. 4.8). - Lead III + left forearm WHITE lead white: left wrist red: left ankle black: right ankle right leg BLACK lead (Ground) left leg RED lead Fig. 4.8 Lead III Configuration 7. Click Resume. When you click Resume, the recording will begin and an append marker labeled Lying down, Lead III will automatically be inserted. 8. Record for 30 seconds. Subject should be supine, relaxed, and still (seconds 30-60). 9. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 10. Review the data on the screen. If correct, go to Step 11. Your data should resemble Fig If incorrect, click Redo. Fig. 4.9 Segment 2 (Supine, Lead III) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a. The Suspend button was pressed prematurely b. An electrode peeled up, giving a large baseline drift c. The Subject has too much muscle (EMG) artifact If data was incorrect, click Redo and repeat Steps 7-10; the last data segment you recorded will be erased. Recording continues

8 Page 8 Biopac Science Lab SEGMENT 3 Seated, Lead III 11. Have the Subject quickly get up and sit in a chair, with arms relaxed. 12. Click Resume as soon as possible after the Subject sits up. 13. After 10 seconds of recording, the Subject should breathe in and out once such that the breath is audible Note Do not insert markers. Subject should sit with arms relaxed, preferably in a chair with armrests. In order to capture the heart rate variation, it is important that you resume recording as quickly as possible after the Subject sits. However, it is also important that you do not click Resume while the Subject is in the process of sitting or you will capture motion artifact. When you click Resume, the recording will continue and an append marker labeled Seated, Lead III will be automatically inserted. After 10 seconds of recording, the Director should instruct the Subject to breathe in and out once such that the inhale and exhale are audible. Note Do not insert markers in this segment because the automatic processing for this lesson splits the recorded segments into separate channels (after Done is clicked), which makes the time base for markers meaningless so they are not retained. The recording should run for 30 seconds (seconds 60-90). 14. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data and prepare for the next recording segment. 15. Review the data on the screen. If correct, go to Step 13. Your data should resemble Fig If incorrect, click Redo. Fig Segment 3 (Seated, Lead III) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a) The Suspend button was pressed prematurely. b) An electrode peeled up, giving a large baseline drift. c) The Subject has too much muscle (EMG) artifact. If data was incorrect, click Redo and repeat Steps 11-15; the last data segment you recorded will be erased. Recording continues

9 Lesson 4: ECG 2 Page 9 SEGMENT 4 Seated, Lead I 16. Disconnect the leads from the Lead III configuration and reconnect to establish a Lead I configuration (Fig. 4.11) while Subject remains seated and relaxed. - Lead I + right forearm WHITE lead left forearm RED lead white: right wrist red: left wrist black: right ankle right leg BLACK lead (Ground) Fig Lead 1 configuration 17. Click Resume. When you click Resume, the recording will continue and an append marker labeled Seated, Lead I will be automatically inserted. 18. After 10 seconds of recording, the Subject should breathe in and out once such that the breath is audible. Note Do not insert markers. After about 10 seconds of recording, the Director should instruct the Subject to breathe in and out once such that the inhale and exhale are audible. Note Do not insert markers in this segment because the automatic processing for this lesson splits the recorded segments into separate channels (after Done is clicked), which makes the time base for markers meaningless so they are not retained. The recording should run for 30 seconds (seconds ). 19. Click Suspend. When you click Suspend, the recording will halt, giving you time to review the data. 20. Review the data. If correct, go to Step 21. Your data should resemble Fig If incorrect, click Redo. Recording continues Fig Segment 4 (Seated, Lead I) A little baseline drift when the Subject breathes in and out is normal and does not indicate incorrect data. The data would be incorrect if: a) The Suspend button was pressed prematurely. b) An electrode peeled up, giving a large baseline drift. c) The Subject has too much muscle (EMG) artifact. If data was incorrect, click Redo and repeat Steps 17-20; the last data segment you recorded will be erased.

10 Page Optional: Click Resume to record additional segments. 22. Click Done. 23. Click Yes. 24. Choose an option and click OK. 25. Remove the electrodes. END OF RECORDING Biopac Science Lab Optional: You can record additional segments by clicking Resume instead of Done. A time marker will be inserted at the start of each added segment. A pop-up window with options will appear. Click Yes (or No if you want to redo the last segment). When you click Yes, a dialog with options will be generated. Make your choice, and click OK. If you choose Analyze current data file, go to the Analyze section for directions. Unclip the electrode lead connectors and peel off the electrodes. Throw out the electrodes.

11 Lesson 4: ECG 2 Page 11 V. ANALYZE In this section, you will examine ECG components of cardiac cycles and measure amplitudes (mv) and durations (msecs) of the ECG components. Note: Interpreting an ECG is a skill that requires practice to distinguish between normal variation and those arising from medical conditions. Do not be alarmed if your ECG is different than the examples shown or from the tables and figures in Section I (Scientific Principles). Because ECGs are widely used, basic elements have been standardized to simplify reading ECGs. ECGs have standardized grids of lighter, smaller squares and, superimposed on the first grid, a second grid of darker and larger squares (Fig. 4.13). The smaller grid always has time units of 0.04 seconds on the x-axis, and the darker vertical lines are spaced 0.2 seconds apart. The horizontal lines represent amplitude in mv. The lighter horizontal lines are 0.1 mv apart, and the darker grid lines represent 0.5 mv. Fig Standard ECG grid FAST TRACK Details 1. Enter the Review Saved Data mode and choose the correct data file. Note Channel Number (CH) designations: To review saved data, choose Analyze current data file from the Done dialog after recording data, or choose Review Saved Data from the Lessons menu and browse to the required file. Channel Displays CH1 CH2 CH3 CH4 Supine, Lead I Supine, Lead III Seated, Lead III Seated, Lead I Important For this lesson analysis, segments that were recorded sequentially have been split into channels for relative comparison and do not share a time base. 2. Set up measurement boxes for: Channel SC Measurement Max Fig IMPORTANT: Channels do not share a time base The measurement boxes are above the marker region in the data window. Each measurement has three sections: channel number, measurement type, and result. The first two sections are pull-down menus that are activated when you click them. SC SC Delta Mean Max: finds the maximum amplitude value within the selected area (including the endpoints) by the I-Beam tool. Delta: computes the difference in amplitude between the last point and the first point of the selected area. It is particularly useful for taking ECG measurements because the baseline does not have to be at zero to obtain accurate, quick measurements. Note The polarity of Delta is based on (first point-last point). First point is where you begin your selection; last point is where you release the mouse to end your selection. 3. For each of the leads, note whether the R wave is positive or negative. A This is just a visual check as to whether the R wave goes up (positive) or down (negative).

12 Page 12 Biopac Science Lab Analysis continues 4. Set up the display window for optimal viewing of one cardiac cycle in the supine channels. 5. Using the I-Beam tool, select and measure Max from a cycle in the center of each channel. B Fig The following tools help you adjust the data window: Autoscale Horizontal Autoscale Waveforms Zoom Your window should resemble Fig Horizontal (Time) Scroll Bar Vertical (Amplitude) Scroll Bar Zoom Previous/Back You ll use the recorded measurements to construct the graphical estimate of the Mean QRS Potential in the Data Report. 6. Place an event marker to indicate where the QRS measurement was taken you will add four markers. Insert the event marker directly above the selected R wave. To place an event marker after the data has been recorded, click the cursor in the event marker region under the marker label bar (above the top data channel). An inverted triangle should appear, which, when selected, will be yellow. Type marker labels: Max Lead I down, Max Lead I up, Max Lead III down, and Max Lead III up. 7. Repeat Steps 5 and 6 for a cardiac cycle in the Seated channel. B 8. Select the breathe in area of a cardiac cycle in the CH3 and repeat Steps 5 and 6. B 9. Select the breathe in area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B Breathe in was recorded in the Seated, Lead I and Seated, Lead III segments, now displayed as CH3 and CH4. The breathe change should occur as an upward spike in the data. 10. Select the breathe out area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B Breathe out was recorded in the Seated, Lead I and Seated, Lead III segments, now displayed as CH3 and CH4. The breathe change should occur as a downward spike in the data. 11. Select the breathe out area of a cardiac cycle in the CH4 and repeat Steps 5 and 6. B 12. Go back to the Max Lead I down marker created in Step 6. This is the same QRS region you measured and marked in Steps 5 and 6. It represents a QRS cycle in the supine, relaxed

13 Lesson 4: ECG 2 Page Measure and record the amplitudes of the Q, R, and S waves individually for both Lead I and Lead III for each segment. C state. You will be taking a total of six measurements. You may wish to look at one channel at a time by hiding the other channel(s) using the channel boxes. To measure a peak, go from the baseline (isoelectric line) to the peak of the wave. Fig Sample measurement of the R-wave peak 14. Save or print the data file. You may save the data, save notes that are in the journal, or print the data file. 15. Exit the program. 16. Set the MP40 dial to Off. END OF LESSON 4 Complete the Lesson 4 Data Report that follows.

14 Page 14 Biopac Science Lab The Data Report starts on the next page.

15 Biopac Science Lab Page 15 Lesson 4 ELECTROCARDIOGRAPHY 2 DATA REPORT Student s Name: Lab Section: Date: These are sample questions. You should amend, add, or delete questions to support your curriculum objectives. I. Data and Calculations Subject Profile Name Gender: Male / Female Height Weight Age Direction of R-Waves for Different Leads A. Put a check in the column to indicate whether the R-wave is positive (+) or negative ( ) for each lead configuration: Table 4.1 Lead Lead I Lead III R-wave + Mean QRS Potential and Axis Graphical Estimate Using R-wave Amplitude B. Use Table 4.2 to record measurements from the Data Analysis section: Table 4.2 QRS Potential and QRS Axis Condition Potential QRS Supine Seated Breathe In Breathe Out Lead I Max Lead III Max Mean QRS Potential Mean QRS Axis

16 Page 16 Biopac Science Lab A simple way to approximate the mean QRS axis in the frontal plane is to plot the magnitude of the R wave as measured from Lead I and Lead III (Fig. 4.3, repeated below from Section 1). magnitude of R wave from leads 0,0 I Mean Electrical Axis III Mean Electrical Axis Fig ) Determine the R-wave amplitude (mv) in Lead I and Lead III. 2) Plot the Lead I R-wave amplitude value on the Lead I axis. Draw an arrow from the center of the vectorgraph to the amplitude value. 3) Plot the Lead III R-wave amplitude value on the Lead III axis. Draw an arrow from the center of the vectorgraph to the amplitude value. 4) Draw a line perpendicular to the tip of each arrowhead 5) Determine the point of intersection of these two perpendicular lines. 6) Draw a new vector from point 0,0 at the center of the vectorgraph to the point where the perpendicular lines intersect with one another. The direction (in degrees) of this resulting vector approximates the mean QRS axis. The length (mv) of this vector approximates the mean QRS potential. To determine Mean QRS Potential from the plotted vector, subdivide it into 0.1 mv increments exactly as the leads are in the vectorgraph. Alternatively, the length of the vector can be measured (with a compass, string, notepaper, etc.) and then transferred to a lead line to determine the mv value. One way to approximate the mean electrical axis in the frontal plane is to plot the magnitude of the R wave from Lead I and Lead III, as shown in Fig 4.2 in Section I (Scientific Principles). 1. Draw a perpendicular line from the ends of the vectors (right angles to the axis of the Lead) using a protractor or right angle guide. 2. Determine the point of intersection of these two perpendicular lines. 3. Draw a new vector from point 0,0 to the point of intersection. The direction of this resulting vector approximates the mean electrical axis of the heart. The length of this vector approximates the mean potential of the heart.

17 Lesson 4: ECG 2 Page 17 Create two plots on each of the following graphs, using data from Table 4.2. Use a different color pencil or pen for each plot. Graph 1: Lead I Lying Down and Sitting Up 0,0 mv Lead I mv Lead III Lead II From the above graph, find the following values: Condition Mean QRS Potential Mean QRS Axis Lying Down Sitting Up Explain the difference (if any) in Mean QRS Potential and Axis under the two conditions:

18 Page 18 Biopac Science Lab Graph 2: Breathing In /Breathing Out or Lead III Lying Down/Sitting Up 0,0 mv Lead I mv Lead III Lead II From the above graph, find the following values: Condition Mean QRS Potential Mean QRS Axis Lying Down Sitting Up Explain the difference in Mean QRS Potential and Axis for the Lying Down data in this plot (Graph 2) and the first plot (Graph 1). C. Mean QRS Potential and Magnitude More Accurate Approximation Add the Q, R, and S potentials to obtain net potentials. Lying down: Lead I Q R S Lead III Q R S QRS Net 1 QRS Net 2

19 Lesson 4: ECG 2 Page 19 II. Data Summary and Questions D. Define ECG. E. Define Einthoven s Triangle. F. What factors affect the orientation of the mean electrical axis? G. Refer to Table 4.2: How did the amplitudes of Lead I and Lead III change between inhalation and exhalation? Did the heart s axis and magnitude change? H. What factors affect the amplitude of the R wave recorded on the different leads? I. Compare the mean electrical axis and magnitude obtained when i. using just the amplitude measurement of the R wave vs. net potentials ii. lying down vs. sitting up End of Biopac Science Lab Lesson 4 Data Report

20 Biopac Science Lab Page 20 VI. ACTIVE LEARNING LAB Design a new experiment to test or verify the scientific principle(s) you learned in the Biopac Science Lab recording and analysis segments. For this lesson, you might examine how gender, duration of exercise, and changing body position influence the ECG and/or time to return to baseline. Design Your Experiment Use a separate sheet to detail your experiment design, and be sure to address these main points: A. Hypothesis Describe the scientific principle to be tested or verified. B. Materials List the materials will you use to complete your investigation. C. Method Describe the experimental procedure be sure to number each step to make it easy to follow during recording. See the Set Up section or Help > About Electrodes for electrode placement guidelines. Run Your Experiment D. Setup Set up the equipment and prepare the subject for your experiment. E. Recording Use the Record, Resume, and Suspend buttons in the Biopac Science Lab program to record as many segments as necessary for your experiment. Click Done when you have completed all of the segments required for your experiment. Analyze Your Experiment F. Set measurements relevant to your experiment and record the results in a Data Report.

Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Introduction. Rev. 06132012

Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Introduction. Rev. 06132012 42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 6 ELECTROCARDIOGRAPHY (ECG) II Introduction Rev. 06132012 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana University

More information

Electrocardiography I Laboratory

Electrocardiography I Laboratory Introduction The body relies on the heart to circulate blood throughout the body. The heart is responsible for pumping oxygenated blood from the lungs out to the body through the arteries and also circulating

More information

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle

Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Biology 347 General Physiology Lab Advanced Cardiac Functions ECG Leads and Einthoven s Triangle Objectives Students will record a six-lead ECG from a resting subject and determine the QRS axis of the

More information

#AS148 - Automated ECG Analysis

#AS148 - Automated ECG Analysis BIOPAC Systems, Inc. 42 Aero Camino Goleta, Ca 93117 Ph (805)685-0066 Fax (805)685-0067 www.biopac.com info@biopac.com #AS148 - Automated ECG Analysis An electrocardiogram (ECG) is a graphical recording

More information

Activity 4.2.3: EKG. Introduction. Equipment. Procedure

Activity 4.2.3: EKG. Introduction. Equipment. Procedure Activity 4.2.3: EKG The following is used with permission of Vernier Software and Technology. This activity is based on the experiment Analyzing the Heart with EKG from the book Human Physiology with Vernier,

More information

Biopac Student Lab BASIC TUTORIAL. BIOPAC Systems, Inc. Version 3.0.3

Biopac Student Lab BASIC TUTORIAL. BIOPAC Systems, Inc. Version 3.0.3 Biopac Student Lab Version 3.0.3 PC running Windows or Macintosh BASIC TUTORIAL Running the Biopac Student Lab page 3 Locating files in Review Saved Data mode page 4 Important Concepts pages 6, 7, 10 Display

More information

ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM

ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM ELECTROCARDIOGRAPHY (I) THE GENESIS OF THE ELECTROCARDIOGRAM Scridon Alina, Șerban Răzvan Constantin 1. Definition The electrocardiogram (abbreviated ECG or EKG) represents the graphic recording of electrical

More information

Exercise, the Electrocardiogram, and Peripheral Circulation

Exercise, the Electrocardiogram, and Peripheral Circulation iworx Physiology Lab Experiment Experiment HH-3 Exercise, the Electrocardiogram, and Peripheral Circulation Note: The lab presented here is intended for evaluation purposes only. iworx users should refer

More information

Analyses: Statistical Measures

Analyses: Statistical Measures Application Note 129 APPLICATION NOTE Heart Rate Variability 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 05.22.14 Analyses: Statistical Measures

More information

Electrophysiology Introduction, Basics. The Myocardial Cell. Chapter 1- Thaler

Electrophysiology Introduction, Basics. The Myocardial Cell. Chapter 1- Thaler Electrophysiology Introduction, Basics Chapter 1- Thaler The Myocardial Cell Syncytium Resting state Polarized negative Membrane pump Depolarization fundamental electrical event of the heart Repolarization

More information

Evaluation copy. Analyzing the Heart with EKG. Computer

Evaluation copy. Analyzing the Heart with EKG. Computer Analyzing the Heart with EKG Computer An electrocardiogram (ECG or EKG) is a graphical recording of the electrical events occurring within the heart. In a healthy heart there is a natural pacemaker in

More information

Electrocardiogram and Heart Sounds

Electrocardiogram and Heart Sounds Electrocardiogram and Heart Sounds An introduction to the recording and analysis of electrocardiograms, and the sounds of the heart. Written by Staff of ADInstruments Introduction The beating of the heart

More information

Understanding the Electrocardiogram. David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem

Understanding the Electrocardiogram. David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem Understanding the Electrocardiogram David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem Overview 1. History 2. Review of the conduction system 3. EKG: Electrodes and Leads 4. EKG: Waves and Intervals

More information

Data Analysis 1. Click the 2-Cursor icon (Figure HH-7-L1) so that two blue vertical lines appear over the recording window.

Data Analysis 1. Click the 2-Cursor icon (Figure HH-7-L1) so that two blue vertical lines appear over the recording window. Experiment HH-7: Electrocardiograms from the Six Chest Leads Exercise 1: ECG from the Chest Leads of a Resting Subject Aim: To record electrocardiograms from the six chest leads and determine the transition

More information

Monitoring EKG. Evaluation copy

Monitoring EKG. Evaluation copy Monitoring EKG Computer 28 An electrocardiogram, or EKG, is a graphical recording of the electrical events occurring within the heart. A typical EKG tracing consists of five identifiable deflections. Each

More information

Reflex Response (Patellar Tendon) Using BIOPAC Reflex Hammer Transducer SS36L

Reflex Response (Patellar Tendon) Using BIOPAC Reflex Hammer Transducer SS36L Updated 7.31.06 BSL PRO Lesson H28: Reflex Response (Patellar Tendon) Using BIOPAC Reflex Hammer Transducer SS36L This PRO lesson describes basic reflex exercises and details hardware and software setup

More information

Electromagnetic Induction: Faraday's Law

Electromagnetic Induction: Faraday's Law 1 Electromagnetic Induction: Faraday's Law OBJECTIVE: To understand how changing magnetic fields can produce electric currents. To examine Lenz's Law and the derivative form of Faraday's Law. EQUIPMENT:

More information

iworx Sample Lab Experiment HN-2: Stretch Receptors and Reflexes

iworx Sample Lab Experiment HN-2: Stretch Receptors and Reflexes Experiment HN-2: Stretch Receptors and Reflexes Exercise 1: Achilles Tendon Reflex Aim: To determine conduction time from tendon tap to response of the gastrocnemius muscle in the Achilles tendon reflex

More information

12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P

12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P 12-Lead EKG Interpretation Judith M. Haluka BS, RCIS, EMT-P ECG Grid Left to Right = Time/duration Vertical measure of voltage (amplitude) Expressed in mm P-Wave Depolarization of atrial muscle Low voltage

More information

Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth

Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2005 Experiment 5: Magnetic Fields of a Bar Magnet and of the Earth OBJECTIVES 1. To examine the magnetic field associated with a

More information

Project 4.2.1: Heart Rate

Project 4.2.1: Heart Rate Project 4.2.1: Heart Rate Introduction Even before you were born, one of the first things your doctor did when you went for an office visit was listen to your heart. Your heart rate, the number of times

More information

Experiment MP-1: Electromyogram (EMG) Activity and Muscle Strength

Experiment MP-1: Electromyogram (EMG) Activity and Muscle Strength Experiment MP-1: Electromyogram (EMG) Activity and Muscle Strength Calibrating the Hand Dynamometer 1. Collect 5 textbooks or use the barbell weight. Weigh the stack of books on the bathroom scale. Record

More information

Lab 3: Introduction to Data Acquisition Cards

Lab 3: Introduction to Data Acquisition Cards Lab 3: Introduction to Data Acquisition Cards INTRODUCTION: In this lab, you will be building a VI to display the input measured on a channel. However, within your own VI you will use LabVIEW supplied

More information

12/15-Lead ECG Protocol approved October 2005 revised October 2008

12/15-Lead ECG Protocol approved October 2005 revised October 2008 approved October 2005 revised October 2008 Preamble Rapid diagnosis of acute myocardial infarction is essential to initiating appropriate treatment and improving outcomes. In selected practice environments

More information

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law. 260 17-1 I. THEORY EXPERIMENT 17 QUALITATIVE STUDY OF INDUCED EMF Along the extended central axis of a bar magnet, the magnetic field vector B r, on the side nearer the North pole, points away from this

More information

ADINSTRUMENTS. making science easier. LabChart 7. Student Quick Reference Guide

ADINSTRUMENTS. making science easier. LabChart 7. Student Quick Reference Guide ADINSTRUMENTS making science easier LabChart 7 Student Quick Reference Guide How to use this guide The LabChart Student Quick Reference Guide is a resource for users of PowerLab systems in the classroom

More information

Licensed Functionality: Remote Monitoring

Licensed Functionality: Remote Monitoring Appendices 473 Chapter 24 Licensed Functionality: Remote Monitoring Remote Monitoring functionality is available through an optional license available with AcqKnowledge 4.2 or above. The license must be

More information

Movement artifacts electrical noise caused by the movement of the sensor surface with respect to the skin surface

Movement artifacts electrical noise caused by the movement of the sensor surface with respect to the skin surface BIOHARNESS 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 12.15.08 Field Comparison: Zephyr BioHarness / Polar HR Monitor Introduction The Zephyr

More information

CONSTRUCTING SINGLE-SUBJECT REVERSAL DESIGN GRAPHS USING MICROSOFT WORD : A COMPREHENSIVE TUTORIAL

CONSTRUCTING SINGLE-SUBJECT REVERSAL DESIGN GRAPHS USING MICROSOFT WORD : A COMPREHENSIVE TUTORIAL CONSTRUCTING SINGLE-SUBJECT REVERSAL DESIGN GRAPHS USING MICROSOFT WORD : A COMPREHENSIVE TUTORIAL PATRICK GREHAN ADELPHI UNIVERSITY DANIEL J. MORAN MIDAMERICAN PSYCHOLOGICAL INSTITUTE This document is

More information

Fixplot Instruction Manual. (data plotting program)

Fixplot Instruction Manual. (data plotting program) Fixplot Instruction Manual (data plotting program) MANUAL VERSION2 2004 1 1. Introduction The Fixplot program is a component program of Eyenal that allows the user to plot eye position data collected with

More information

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.

More information

Lancashire Teaching Hospitals Cardio-Respiratory Department. Lead systems. Paula Hignett

Lancashire Teaching Hospitals Cardio-Respiratory Department. Lead systems. Paula Hignett Lancashire Teaching Hospitals Cardio-Respiratory Department Lead systems Paula Hignett Objectives Understand the terminology and theory of the 12 views of the heart with reference to: Unipolar and Bipolar

More information

Human ECG Laboratory Experiment By

Human ECG Laboratory Experiment By Human ECG Laboratory Experiment By Brittany Baierlein, Alison L., Thurow, Rachel Holsinger and Robin L. Cooper Department of Biology, University of KY, Lexington, KY 40506-0225, USA. (Some text taken directly

More information

Excel 2007 Basic knowledge

Excel 2007 Basic knowledge Ribbon menu The Ribbon menu system with tabs for various Excel commands. This Ribbon system replaces the traditional menus used with Excel 2003. Above the Ribbon in the upper-left corner is the Microsoft

More information

SW43W. Users Manual. FlukeView Power Quality Analyzer Software Version 3.20 onwards

SW43W. Users Manual. FlukeView Power Quality Analyzer Software Version 3.20 onwards SW43W FlukeView Power Quality Analyzer Software Version 3.20 onwards Users Manual 4822 872 30778 December 2005, Rev.1, 04/06 2005 Fluke Corporation, All rights reserved. All product names are trademarks

More information

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5 Physics 161 FREE FALL Introduction This experiment is designed to study the motion of an object that is accelerated by the force of gravity. It also serves as an introduction to the data analysis capabilities

More information

SA-9600 Surface Area Software Manual

SA-9600 Surface Area Software Manual SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The

More information

Intermediate PowerPoint

Intermediate PowerPoint Intermediate PowerPoint Charts and Templates By: Jim Waddell Last modified: January 2002 Topics to be covered: Creating Charts 2 Creating the chart. 2 Line Charts and Scatter Plots 4 Making a Line Chart.

More information

Introduction to Electrocardiography. The Genesis and Conduction of Cardiac Rhythm

Introduction to Electrocardiography. The Genesis and Conduction of Cardiac Rhythm Introduction to Electrocardiography Munther K. Homoud, M.D. Tufts-New England Medical Center Spring 2008 The Genesis and Conduction of Cardiac Rhythm Automaticity is the cardiac cell s ability to spontaneously

More information

Intro to Excel spreadsheets

Intro to Excel spreadsheets Intro to Excel spreadsheets What are the objectives of this document? The objectives of document are: 1. Familiarize you with what a spreadsheet is, how it works, and what its capabilities are; 2. Using

More information

Basic AutoSketch Manual

Basic AutoSketch Manual Basic AutoSketch Manual Instruction for students Skf-Manual.doc of 3 Contents BASIC AUTOSKETCH MANUAL... INSTRUCTION FOR STUDENTS... BASIC AUTOSKETCH INSTRUCTION... 3 SCREEN LAYOUT... 3 MENU BAR... 3 FILE

More information

Capacitive Touch Lab. Renesas Capacitive Touch Lab R8C/36T-A Family

Capacitive Touch Lab. Renesas Capacitive Touch Lab R8C/36T-A Family Renesas Capacitive Touch Lab R8C/36T-A Family Description: This lab will cover the Renesas Touch Solution for embedded capacitive touch systems. This lab will demonstrate how to setup and run a simple

More information

Tutorial for Tracker and Supporting Software By David Chandler

Tutorial for Tracker and Supporting Software By David Chandler Tutorial for Tracker and Supporting Software By David Chandler I use a number of free, open source programs to do video analysis. 1. Avidemux, to exerpt the video clip, read the video properties, and save

More information

Gestation Period as a function of Lifespan

Gestation Period as a function of Lifespan This document will show a number of tricks that can be done in Minitab to make attractive graphs. We work first with the file X:\SOR\24\M\ANIMALS.MTP. This first picture was obtained through Graph Plot.

More information

SupcoLog Logger Interface Software

SupcoLog Logger Interface Software SupcoLog Logger Interface Software USER S GUIDE Table Of Contents Quick Start... 5 Overview... 5 Connecting the Logger to Your PC... 5 Setting Up the Logger To Record Data... 6 Custom Interval Setup...

More information

Efficient Heart Rate Monitoring

Efficient Heart Rate Monitoring Efficient Heart Rate Monitoring By Sanjeev Kumar, Applications Engineer, Cypress Semiconductor Corp. Heart rate is one of the most frequently measured parameters of the human body and plays an important

More information

Updated 09.09.09. Application Note AS-199 Impedance Cardiography and Pre-ejection Period

Updated 09.09.09. Application Note AS-199 Impedance Cardiography and Pre-ejection Period Updated 09.09.09 Application Note AS-199 Impedance Cardiography and Pre-ejection Period This Application Note guides you through the analysis process only, based on the following signals: ECG, Zo and dz/dt.

More information

Real Time Monitor. A Real-Time Windows Operator Interface. DDE Compliant. (for remote data display)

Real Time Monitor. A Real-Time Windows Operator Interface. DDE Compliant. (for remote data display) Real Time Monitor A Real-Time Windows Operator Interface DDE Compliant (for remote data display) TABLE OF CONTENTS 1. INTRODUCTION...1 1.1 INSTALLATION...2 1.2 FIRST START UP - DDE CONFIGURE...2 1.3 AUTO-STARTUP...2

More information

Cardiac Conduction System (1) ECG (Electrocardiogram) Cardiac Conduction System (2) The ECG (1) The ECG (1) The ECG (1) Achmad Rizal BioSPIN

Cardiac Conduction System (1) ECG (Electrocardiogram) Cardiac Conduction System (2) The ECG (1) The ECG (1) The ECG (1) Achmad Rizal BioSPIN ECG (Electrocardiogram) Cardiac Conduction System (1) Achmad Rizal BioSPIN ARL-EL4703-Instrumentasi Biomedis 2 Cardiac Conduction System (2) The ECG (1) ARL-EL4703-Instrumentasi Biomedis 3 ARL-EL4703-Instrumentasi

More information

Data Analysis Methods: Net Station 4.1 By Peter Molfese

Data Analysis Methods: Net Station 4.1 By Peter Molfese Data Analysis Methods: Net Station 4.1 By Peter Molfese Preparing Data for Statistics (preprocessing): 1. Rename your files to correct any typos or formatting issues. a. The General format for naming files

More information

QRS Complexes. Fast & Easy ECGs A Self-Paced Learning Program

QRS Complexes. Fast & Easy ECGs A Self-Paced Learning Program 6 QRS Complexes Fast & Easy ECGs A Self-Paced Learning Program Q I A ECG Waveforms Normally the heart beats in a regular, rhythmic fashion producing a P wave, QRS complex and T wave I Step 4 of ECG Analysis

More information

R Series Quick Reference Guide

R Series Quick Reference Guide R Series Quick Reference Guide 2010 ZOLL Medical Corporation. All rights reserved. CPR Index, OneStep, R Series, Real CPR Help, See-Thru CPR, statpadz, and ZOLL are trademarks or registered trademarks

More information

Cabri Geometry Application User Guide

Cabri Geometry Application User Guide Cabri Geometry Application User Guide Preview of Geometry... 2 Learning the Basics... 3 Managing File Operations... 12 Setting Application Preferences... 14 Selecting and Moving Objects... 17 Deleting

More information

Formulas, Functions and Charts

Formulas, Functions and Charts Formulas, Functions and Charts :: 167 8 Formulas, Functions and Charts 8.1 INTRODUCTION In this leson you can enter formula and functions and perform mathematical calcualtions. You will also be able to

More information

Understand the Sketcher workbench of CATIA V5.

Understand the Sketcher workbench of CATIA V5. Chapter 1 Drawing Sketches in Learning Objectives the Sketcher Workbench-I After completing this chapter you will be able to: Understand the Sketcher workbench of CATIA V5. Start a new file in the Part

More information

SIP Trunk Installation & User Guide

SIP Trunk Installation & User Guide PRILINK http://www.prilink.com Tel: 905-882-4488 1-866-261-0649 Fax: 905-597-1139 Sales@prilink.com Support@prilink.com Table of Contents Overview...3 Important Terms...3 Units of Digital Information...3

More information

SMART Board Software for Macintosh FAQ

SMART Board Software for Macintosh FAQ SMART Board Software for Macintosh FAQ NOTE: The following FAQs relate to the latest SMART Board Software for Macintosh release and apply only to Macintosh computers. SMART Board Software for Macintosh

More information

Scatter Plots with Error Bars

Scatter Plots with Error Bars Chapter 165 Scatter Plots with Error Bars Introduction The procedure extends the capability of the basic scatter plot by allowing you to plot the variability in Y and X corresponding to each point. Each

More information

Chapter 1. Creating Sketches in. the Sketch Mode-I. Evaluation chapter. Logon to www.cadcim.com for more details. Learning Objectives

Chapter 1. Creating Sketches in. the Sketch Mode-I. Evaluation chapter. Logon to www.cadcim.com for more details. Learning Objectives Chapter 1 Creating Sketches in Learning Objectives the Sketch Mode-I After completing this chapter you will be able to: Use various tools to create a geometry. Dimension a sketch. Apply constraints to

More information

How to make a line graph using Excel 2007

How to make a line graph using Excel 2007 How to make a line graph using Excel 2007 Format your data sheet Make sure you have a title and each column of data has a title. If you are entering data by hand, use time or the independent variable in

More information

Microsoft Excel 2013 Tutorial

Microsoft Excel 2013 Tutorial Microsoft Excel 2013 Tutorial TABLE OF CONTENTS 1. Getting Started Pg. 3 2. Creating A New Document Pg. 3 3. Saving Your Document Pg. 4 4. Toolbars Pg. 4 5. Formatting Pg. 6 Working With Cells Pg. 6 Changing

More information

2 ECG basics. Leads and planes. Leads. Planes. from different perspectives, which are called leads and planes.

2 ECG basics. Leads and planes. Leads. Planes. from different perspectives, which are called leads and planes. 558302.qxp 3/14/12 10:47 PM Page 12 2 ECG basics One of the most valuable diagnostic tools available, an electrocardiogram (ECG) records the heart s electrical activity as waveforms. By interpreting these

More information

Video in Logger Pro. There are many ways to create and use video clips and still images in Logger Pro.

Video in Logger Pro. There are many ways to create and use video clips and still images in Logger Pro. Video in Logger Pro There are many ways to create and use video clips and still images in Logger Pro. Insert an existing video clip into a Logger Pro experiment. Supported file formats include.avi and.mov.

More information

SDC. Schroff Development Corporation WWW.SDCACAD.COM PUBLICATIONS. MultiMedia CD by Jack Zecher

SDC. Schroff Development Corporation WWW.SDCACAD.COM PUBLICATIONS. MultiMedia CD by Jack Zecher MultiMedia CD by Jack Zecher An audioi/visual presentation of the tutorial exercises SDC PUBLICATIONS Schroff Development Corporation WWW.SDCACAD.COM AutoCAD 2002 Tutorial 2-1 Lesson 2 Geometric Construction

More information

Experiment #8: Magnetic Forces

Experiment #8: Magnetic Forces Experiment #8: Magnetic Forces Purpose: To study the nature of magnetic forces exerted on currents. Equipment: Magnet Assembly and Stand Set of Current Loop PC oards Triple-Arm Pan alance 0 15 V dc Variable

More information

VisTablet Manual. For 12 VT Original and Mini Mouse

VisTablet Manual. For 12 VT Original and Mini Mouse VisTablet Manual For 12 VT Original and Mini Mouse I n d e x 1. Before using your VisTablet ------------------------------------------------------------------------------- Page#3 2. How to install Pen

More information

The serial port is another kind of transmission mode in some outdated PC as a supplementary in failure of USB transmission.

The serial port is another kind of transmission mode in some outdated PC as a supplementary in failure of USB transmission. Page 1 of 21 OWON Oscilloscope Installation and Use Guide Welcome to use Oscilloscope analysis software. The kind of communication software is use to acquire, store, analyze and display the data. The instruction

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

5. Tutorial. Starting FlashCut CNC

5. Tutorial. Starting FlashCut CNC FlashCut CNC Section 5 Tutorial 259 5. Tutorial Starting FlashCut CNC To start FlashCut CNC, click on the Start button, select Programs, select FlashCut CNC 4, then select the FlashCut CNC 4 icon. A dialog

More information

GENERAL SCIENCE LABORATORY 1110L Lab Experiment 6: Ohm s Law

GENERAL SCIENCE LABORATORY 1110L Lab Experiment 6: Ohm s Law GENERAL SCIENCE LABORATORY 1110L Lab Experiment 6: Ohm s Law OBJECTIVES: To verify Ohm s law, the mathematical relationship among current, voltage or potential difference, and resistance, in a simple circuit.

More information

ABOUT THIS DOCUMENT ABOUT CHARTS/COMMON TERMINOLOGY

ABOUT THIS DOCUMENT ABOUT CHARTS/COMMON TERMINOLOGY A. Introduction B. Common Terminology C. Introduction to Chart Types D. Creating a Chart in FileMaker E. About Quick Charts 1. Quick Chart Behavior When Based on Sort Order F. Chart Examples 1. Charting

More information

Tutorial: Biped Character in 3D Studio Max 7, Easy Animation

Tutorial: Biped Character in 3D Studio Max 7, Easy Animation Tutorial: Biped Character in 3D Studio Max 7, Easy Animation Written by: Ricardo Tangali 1. Introduction:... 3 2. Basic control in 3D Studio Max... 3 2.1. Navigating a scene:... 3 2.2. Hide and Unhide

More information

Introduction Configuring Your HP Mouse Batteries Replacing the Batteries Mouse Sensor Technology Navigating the HP Mouse Control Center

Introduction Configuring Your HP Mouse Batteries Replacing the Batteries Mouse Sensor Technology Navigating the HP Mouse Control Center Introduction Congratulations on your purchase of an HP mouse. This mouse is the latest in a wide range of computer peripherals manufactured by HP. It is designed to provide you with the highest level of

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

Experiment: Series and Parallel Circuits

Experiment: Series and Parallel Circuits Phy203: General Physics Lab page 1 of 6 Experiment: Series and Parallel Circuits OBJECTVES MATERALS To study current flow and voltages in series and parallel circuits. To use Ohm s law to calculate equivalent

More information

Renewable Energy Monitor User Manual And Software Reference Guide. sales@fuelcellstore.com (979) 703-1925

Renewable Energy Monitor User Manual And Software Reference Guide. sales@fuelcellstore.com (979) 703-1925 Renewable Energy Monitor User Manual And Software Reference Guide sales@fuelcellstore.com (979) 703-1925 1 Introducing the Horizon Renewable Energy Monitor The Renewable Energy Monitor is an educational

More information

Exploring Magnetism. DataQuest

Exploring Magnetism. DataQuest Exploring Magnetism Magnetism is the force of attraction or repulsion between a magnet and something else. Magnets attract materials made of iron, nickel, or cobalt. Can you think of five things to which

More information

SIP Trunk Installation & User Guide

SIP Trunk Installation & User Guide PRILINK http://www.prilink.com Tel: 905-882-4488 1-866-261-0649 Fax: 905-597-1139 Sales@prilink.com Support@prilink.com Table of Contents Overview...3 Important Terms...3 Units of Digital Information...3

More information

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of

More information

Making Basic Measurements. Publication Number 16700-97020 August 2001. Training Kit for the Agilent Technologies 16700-Series Logic Analysis System

Making Basic Measurements. Publication Number 16700-97020 August 2001. Training Kit for the Agilent Technologies 16700-Series Logic Analysis System Making Basic Measurements Publication Number 16700-97020 August 2001 Training Kit for the Agilent Technologies 16700-Series Logic Analysis System Making Basic Measurements: a self-paced training guide

More information

Statgraphics Getting started

Statgraphics Getting started Statgraphics Getting started The aim of this exercise is to introduce you to some of the basic features of the Statgraphics software. Starting Statgraphics 1. Log in to your PC, using the usual procedure

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2006 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Building a Simulink model for real-time analysis V1.15.00. Copyright g.tec medical engineering GmbH

Building a Simulink model for real-time analysis V1.15.00. Copyright g.tec medical engineering GmbH g.tec medical engineering GmbH Sierningstrasse 14, A-4521 Schiedlberg Austria - Europe Tel.: (43)-7251-22240-0 Fax: (43)-7251-22240-39 office@gtec.at, http://www.gtec.at Building a Simulink model for real-time

More information

Magnetic Fields and Their Effects

Magnetic Fields and Their Effects Name Date Time to Complete h m Partner Course/ Section / Grade Magnetic Fields and Their Effects This experiment is intended to give you some hands-on experience with the effects of, and in some cases

More information

An ECG Primer. Quick Look. I saw it, but I did not realize it. Elizabeth Peabody

An ECG Primer. Quick Look. I saw it, but I did not realize it. Elizabeth Peabody 4 An ECG Primer Quick Look Cardiac Monitoring System - p. 64 ECG Paper - p. 73 Lead Polarity and Vectors - p. 77 Basic ECG Components - p. 79 Heart Rate and Pulse Rate - p. 91 Summary - p. 94 Chapter Quiz

More information

STATIC AND KINETIC FRICTION

STATIC AND KINETIC FRICTION STATIC AND KINETIC FRICTION LAB MECH 3.COMP From Physics with Computers, Vernier Software & Technology, 2000. INTRODUCTION If you try to slide a heavy box resting on the floor, you may find it difficult

More information

Dash 18X / Dash 18 Data Acquisition Recorder

Dash 18X / Dash 18 Data Acquisition Recorder 75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

First Level: 2D Fundamentals. by Randy H. Shih Oregon Institute of Technology

First Level: 2D Fundamentals. by Randy H. Shih Oregon Institute of Technology AutoCAD 2008 Tutorial First Level: 2D Fundamentals by Randy H. Shih Oregon Institute of Technology MultiMedia CD by Jack Zecher Indiana University Purdue University Indianapolis SDC PUBLICATIONS Schroff

More information

2. Event Initiated ECG Sweep Recording (Typically used in ECG exams)

2. Event Initiated ECG Sweep Recording (Typically used in ECG exams) Reviewer Program Description The Vmed Reviewer program is included with each Vmed wireless monitoring product and is also available free to any interested person in order to facilitate ECG interpretation

More information

Creating a Poster Presentation using PowerPoint

Creating a Poster Presentation using PowerPoint Creating a Poster Presentation using PowerPoint Course Description: This course is designed to assist you in creating eye-catching effective posters for presentation of research findings at scientific

More information

The following is an overview of lessons included in the tutorial.

The following is an overview of lessons included in the tutorial. Chapter 2 Tutorial Tutorial Introduction This tutorial is designed to introduce you to some of Surfer's basic features. After you have completed the tutorial, you should be able to begin creating your

More information

NEONATAL & PEDIATRIC ECG BASICS RHYTHM INTERPRETATION

NEONATAL & PEDIATRIC ECG BASICS RHYTHM INTERPRETATION NEONATAL & PEDIATRIC ECG BASICS & RHYTHM INTERPRETATION VIKAS KOHLI MD FAAP FACC SENIOR CONSULATANT PEDIATRIC CARDIOLOGY APOLLO HOSPITAL MOB: 9891362233 ECG FAX LINE: 011-26941746 THE BASICS: GRAPH PAPER

More information

Leaf River Outdoor Products Vibrashine, Inc. P.O. Box 557 Taylorsville, MS 39168 (601) 785-9854 FAX (601) 785-9874 www.vibrashine.

Leaf River Outdoor Products Vibrashine, Inc. P.O. Box 557 Taylorsville, MS 39168 (601) 785-9854 FAX (601) 785-9874 www.vibrashine. MULTI-CARD PICTURE VIEWER Model PV-2.5 Leaf River Outdoor Products Vibrashine, Inc. P.O. Box 557 Taylorsville, MS 39168 (601) 785-9854 FAX (601) 785-9874 www.vibrashine.com 2 THANKS and WELCOME What does

More information

Microsoft Word 2010. Revising Word Documents Using Markup Tools

Microsoft Word 2010. Revising Word Documents Using Markup Tools Microsoft Word 2010 Revising Word Documents Using Markup Tools Preface Word provides several markup tools that make document collaboration easy. Color coding, highlighting, and the ability maintain multiple

More information

SMART Ink 1.5. Windows operating systems. Scan the following QR code to view the SMART Ink Help on your smart phone or other mobile device.

SMART Ink 1.5. Windows operating systems. Scan the following QR code to view the SMART Ink Help on your smart phone or other mobile device. SMART Ink 1.5 Windows operating systems User s guide Scan the following QR code to view the SMART Ink Help on your smart phone or other mobile device. Trademark notice SMART Ink, SMART Notebook, SMART

More information

Navigator Software. Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3.

Navigator Software. Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3. Navigator Software Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3.2 SYSTEM SET UP 3.3 LOAD DATA FILE 3.3.1 LOADING PARTIAL FILES 3.4 DATA TABLE WINDOW

More information

The EasySense unit can detect that the Smart Q Heart Rate Sensor is connected and the range it is set to.

The EasySense unit can detect that the Smart Q Heart Rate Sensor is connected and the range it is set to. Heart Rate Sensor Heart Rate Sensor (Product No PC-3147) Pulse rate Range: 0 to 200 bpm Resolution: 1 bpm Waveform Range: -2000 to 2000 mv Resolution: 1 mv Introduction The Smart Q Heart Rate Sensor monitors

More information

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI 12 Experiment 12 : Examination of the heart using ECG and PCG I. Background theory. 1. Construction and functioning

More information

Dash 8Xe / Dash 8X Data Acquisition Recorder

Dash 8Xe / Dash 8X Data Acquisition Recorder 75 Dash 8Xe / Dash 8X Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 2.0 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

Basic 2D Design Be sure you have the latest information!

Basic 2D Design Be sure you have the latest information! Basic 2D Design mastercam x getting started tutorials Basic 2D Design December 2011 Be sure you have the latest information! Information might have been changed or added since this document was published.

More information