Renee Scott, Brad S. Karon, Laurie Griesmann, Sandra C. Bryant, Jeffrey A. DuBois, Terry L. Shirey, Steven Presti, and Paula J.

Size: px
Start display at page:

Download "Renee Scott, Brad S. Karon, Laurie Griesmann, Sandra C. Bryant, Jeffrey A. DuBois, Terry L. Shirey, Steven Presti, and Paula J."

Transcription

1 ABSTRACT We evaluated four hospital-based glucose meter technologies for accuracy, precision, and analytical interferences likely to be encountered in critically ill patients. Precision of all four glucose meters, both within run and day-to-day, instruments. Accuracy (bias) of the meters and analytical interference were evaluated by comparing results obtained on whole blood specimens to plasma samples obtained from these whole blood specimens run on a hexokinase reference method. There were significant differences in the degree to which the meters correlated with a reference hexokinase method. Ascorbic acid showed significant interference with 3 of the 4 meters. Hematocrit also affected the correlation between whole blood and plasma hexokinase glucose on 3 of the 4 glucose meters tested, with the magnitude of this interference also varied by glucose meter technology. The use of glucose meters to maintain tight glycemic control in critically ill patients may be impacted by the accuracy and hematocrit dependence of some glucose meter technologies. GLUCOSE METHODS USED Accu-Chek Inform (Roche Diagnostics, Indianapolis, IN) Glucose dehydrogenase based amperometric strip Precision PCx (Abbott Diabetes, Alameda, CA) Glucose oxidase amperometric detection SureStepFlexx (LifeScan, Malpitas, CA) Photometric glucose oxidase detection StatStrip (Nova Biomedical, Waltham, MA) Modified glucose oxidase based amperometric test system Roche Integra 400 Analyzer (Roche Diagnostics, Indianapolis, IN) Hexokinase method TABLE 1 Within Run and Day-to-Day precision for glucose meters, in percent coefficient of variation (CV), n= number of replicates tested Within Run Precision, Day-to-Day Precision, CV (%) n = 20 CV (%) n = 30 Meter Low Medium High Low High Technology Glucose Glucose Glucose Glucose Glucose mg/dl mg/dl mg/dl mg/dl mg/dl StatStrip Accu-Chek PCx SureStep The coefficient of variation (CV) obtained from both within-run and day-to-day precision experiments was less than 5% for all meters tested except day-to-day precision at low glucose on PCx which was 5.1% STUDY DESIGN Method correlation was performed by analyzing 133 whole blood specimens and 52 spiked whole blood patient specimens on the 4 glucose meters, compared to plasma obtained from those specimens and run on the Integra (reference method). Percent bias (meter result minus plasma result/plasma result) plotted as a function of hematocrit for the 133 whole blood specimens. Interference studies were performed using ascorbic acid added to whole blood at 4 different glucose levels. Six strips tested on each meter at each concentration of interferant, mean used for data analysis. Clinically significant interference defined as > 10 mg/dl(for glucose concentration < 100 mg/dl) or > 10% (for glucose concentration > 100 mg/dl) change from baseline. Hematocrit interference was tested using 3 glucose concentrations over a 25-60% hematocrit range, using six strips for each meter at each hematocrit range tested. Clinically significant effect of hematocrit defined as > 10 mg/dl(for glucose concentrations < 100 mg/dl) or > 10% (for glucose concentrations > 100 mg/dl) change between lowest and highest hematocrit tested. TABLE 2 Correlation data for glucose meters vs reference plasma hexokinase method, n=185 or 133 Meter Slope Intercept (mg/dl) r 2 Median bias (mg/dl) Technology N=185 N=133 N=185 N=133 N=185 N=133 N=185 N=133 StatStrip Accu-Chek PCx SureStep n=133 data set, whole blood samples n=185 data set, 133 whole blood samples and 52 spiked whole blood samples StatStrip and Accu-Chek meters correlate better with the reference hexokinase method, as measured by slope and intercept Median bias was lower for the StatStrip and SureStep meters 2007 Mayo Foundation for Medical Education and Research 11 1

2 12

3 13

4 Glucose meter percent bias was plotted against hematocrit values for the 133 patient sample set (no sample manipulation). There is a clear trend for negative bias associated with increasing hematocrit for the PCx and SureStep meters. The Accu-Chek and StatStrip were largely unaffected by hematocrit. 14

5 CONCLUSION Glucose meters vary in the degree of correlation to the hexokinase reference glucose methods. Most glucose meters continue to be affected by analytical interferences, especially hematocrit. The new StatStrip glucose meter technology correlated closely with a reference hexokinase glucose method and did not demonstrate clinically significant interference from any of the substances tested. 15

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory

More information

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

Blood Glucose Monitoring: The Facts about Accuracy

Blood Glucose Monitoring: The Facts about Accuracy Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined

More information

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and CoaguChek systems Cardiovascular disease: the #1 killer in

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from

More information

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM

More information

A comparison of blood glucose meters in Australia

A comparison of blood glucose meters in Australia Diabetes Research and Clinical Practice 71 (2006) 113 118 www.elsevier.com/locate/diabres A comparison of blood glucose meters in Australia Matthew Cohen *, Erin Boyle, Carol Delaney, Jonathan Shaw International

More information

SKUP Scandinavian evaluation of laboratory equipment for primary health care

SKUP Scandinavian evaluation of laboratory equipment for primary health care SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Aviva Meter and test strips designed for glucose self-measurement Manufactured by Roche Diagnostics GmbH Report from

More information

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz

More information

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care Accuracy Test Omnitest Blood Glucose Monitoring System B. Braun Diabetes Care Accuracy Test Omnitest blood glucose monitoring system according to ISO 97 The accuracy study of Omnitest blood glucose monitoring

More information

How To Test For Creatinine

How To Test For Creatinine Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed

More information

Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff

Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff GUIDANCE guidance. Submit written comments to the Division of

More information

PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.01- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 1 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.01 LBH 1 OF 7 Printed copies are

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few

More information

Blood Sugar Monitoring

Blood Sugar Monitoring Sugar Monitoring Know Your Sugar Numbers Taking control of your diabetes and blood glucose (blood sugar) can make you feel better and stay healthy. Research shows that maintaining your blood glucose level

More information

Practice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013

Practice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013 Practice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013 For the person with diabetes, blood glucose monitoring provides immediate feedback on the effects of daily activities such as taking

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers

Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers Journal of Diabetes Science and Technology Volume 7, Issue 1, January 2013 Diabetes Technology Society ORIGINAL ARTICLE Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems:

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2 Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

More information

US Diabetic Devices Industry Research Report: KenResearch

US Diabetic Devices Industry Research Report: KenResearch US Diabetic Devices Industry Research Report: KenResearch by KenResearch - Monday, February 24, 2014 https://www.kenresearch.com/blog/2014/02/us-diabetic-devices-industry-research-report-kenresearch/ The

More information

Center for Clinical Standards and Quality/Survey & Certification Group

Center for Clinical Standards and Quality/Survey & Certification Group DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality/Survey

More information

Quick Reference Guide

Quick Reference Guide Quick Reference Guide Look inside for step-by-step instructions on how to: >> Transfer readings from your OneTouch Meter >> View and share any of the 11 available reports >> Add information to a patient

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

Country-Specific Glucose Monitor List

Country-Specific Glucose Monitor List Country-Specific Glucose Monitor List Country Name: United States Important Information This is a non-comprehensive list, current as of March 2016. Absence of a specific glucose monitor or test strips

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban

More information

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved Overview and Objectives This module provides an introduction to

More information

Comparison of Glucose Determinations on Blood Samples Collected in Three Types of Tubes

Comparison of Glucose Determinations on Blood Samples Collected in Three Types of Tubes 278 Available online at www.annclinlabsci.org Comparison of Glucose Determinations on Blood Samples Collected in Three Types of Tubes Geling Li 1,2, Michael Cabanero 1,2, Zhenglong Wang 1,2, Huiying Wang

More information

Buyers guide. Blood glucose systems CEP 08008

Buyers guide. Blood glucose systems CEP 08008 Buyers guide Blood glucose systems CEP 08008 May 2008 Contents 2 Introduction... 3 Technical considerations... 5 Operational considerations... 9 Economic considerations... 14 Purchasing considerations...

More information

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.

More information

Biochemistry Validation Form

Biochemistry Validation Form Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August

More information

A guidebook for people with diabetes

A guidebook for people with diabetes A guidebook for people with diabetes This booklet is designed to supplement, not replace, your doctor s advice. Please consult your doctor if you have any questions about what you read. You ll learn how

More information

IT Risk Management. Agenda. Roche at a glance OSI Layers above 7 What we did What we plan to do how to not get shoot

IT Risk Management. Agenda. Roche at a glance OSI Layers above 7 What we did What we plan to do how to not get shoot IT Risk Management 30. 2006 ISACA Switzerland Chapter After Hours Seminar Mario Walter Agenda Roche at a glance OSI Layers above 7 What we did What we plan to do how to not get shoot 2006 ISACA After Hours

More information

Worldwide Markets and Emerging Technologies for Point-of-Care Testing, 2013-2019. February 2015

Worldwide Markets and Emerging Technologies for Point-of-Care Testing, 2013-2019. February 2015 Worldwide Markets and Emerging Technologies for Point-of-Care Testing, 2013-2019 February 2015 INTELAB CORPORATION MARKET AND TECHNOLOGY REPORTS Report 530 Copyright 2015 Worldwide Markets and Emerging

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use

More information

Quantifying measurement error from digital instruments

Quantifying measurement error from digital instruments Quantifying measurement error from digital instruments W. BLAKE LAING AND SEAN BRYANT SOUTHERN ADVENTIST UNIVERSITY CHAT TANOOGA, TN What I m doing HELPING STUDENTS LEARN TO CONSTRUCT KNOWLEDGE First lab:

More information

Validation of measurement procedures

Validation of measurement procedures Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM Inform II BLOOD GLUCOSE MONITORING SYSTEM Quick Reference Guide for Healthcare Professionals 2 Table of Contents Important... 4 General Information... 6 ACCU-CHEK Inform II Meter...10 How to Perform Patient

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Test strips and meter designed for glucose self-measurement manufactured by Abbott. Report from an evaluation organised by SKUP

Test strips and meter designed for glucose self-measurement manufactured by Abbott. Report from an evaluation organised by SKUP SKUP Scandinavian evaluation of laboratory equipment for primary health care Precision Xtra Plus (G3c) Precision Xceed Test strips and meter designed for glucose self-measurement manufactured by Abbott

More information

Correlation. What Is Correlation? Perfect Correlation. Perfect Correlation. Greg C Elvers

Correlation. What Is Correlation? Perfect Correlation. Perfect Correlation. Greg C Elvers Correlation Greg C Elvers What Is Correlation? Correlation is a descriptive statistic that tells you if two variables are related to each other E.g. Is your related to how much you study? When two variables

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

Textbook: Chapter 12 pages 351-353, Chapter 13 pages 385-388

Textbook: Chapter 12 pages 351-353, Chapter 13 pages 385-388 EXERCISE 11: BEDSIDE GLUCOSE TESTING Textbook: Chapter 12 pages 351-353, Chapter 13 pages 385-388 Skills: Objectives: 15 points 1. List the limitations of the bedside glucose test procedure. 2. State the

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Accu-Chek Inform II system. Professional glucose testing for the wireless age

Accu-Chek Inform II system. Professional glucose testing for the wireless age Accu-Chek Inform II system Professional glucose testing for the wireless age The Accu-Chek Inform II system Real time control The Accu-Chek Inform II System overview Quality Advanced chemistry with multiple

More information

Basic Lessons in Laboratory Quality Control

Basic Lessons in Laboratory Quality Control Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and

More information

for patient monitoring

for patient monitoring Thermo Scientific Indiko Reagents for Immunosuppressant Drug Testing Complete solutions for patient monitoring The cost-effective Thermo Scientific Indiko benchtop analyzer and bar-coded reagent is designed

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

Symmetry in Diagnostics HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. TOSOH BIOSCIENCE www.tosohbioscience.us

Symmetry in Diagnostics HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. TOSOH BIOSCIENCE www.tosohbioscience.us Symmetry in Diagnostics HPLC ANALYZER Gold Standard Accuracy by Ion-Exchange HbA1c TOSOH BIOSCIENCE www.tosohbioscience.us HPLC ANALYZER HbA1c analysis in 1.6 minutes Direct determination of stable HbA1c

More information

For people with diabetes. A blood sugar diary

For people with diabetes. A blood sugar diary For people with diabetes A blood sugar diary A circle of help to live a healthy life YOU are at the center of a healthy life with diabetes. All the elements of good care begin and end with YOU. The importance

More information

Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose Analyzer

Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose Analyzer Clinical Chemistry / DRUG EFFECTS AND POINT-OF-CARE GLUCOSE TESTING Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose Analyzer Zuping Tang, MD, Xiaogu Du, MD,

More information

SAMPLE. Effects of Different Sample Types on Glucose Measurements

SAMPLE. Effects of Different Sample Types on Glucose Measurements 1st Edition POCT06 Effects of Different Sample Types on Glucose Measurements This report provides information to assist the clinical and point-ofcare staff in result and measurement procedure comparisons

More information

D-Facts. Vitamin D Competitive Evaluation

D-Facts. Vitamin D Competitive Evaluation D-Facts Vitamin D Competitive Evaluation Abstract 101 patient samples collected across sites in Germany were used to compare the performance of the Siemens ADVIA Centaur Vitamin D Total (VitD), Abbott

More information

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number 1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression

More information

Alignment and Preprocessing for Data Analysis

Alignment and Preprocessing for Data Analysis Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs

Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs Elizabeth Herron URI Watershed Watch/ National Facilitation of CSREES Volunteer Monitoring Ingrid Harrald Cook

More information

OneStep Fecal Occult Blood RapiDip InstaTest. Cat # 13020-1

OneStep Fecal Occult Blood RapiDip InstaTest. Cat # 13020-1 CORTEZ DIAGNOSTICS, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: onestep@rapidtest.com Web site: www.rapidtest.com See external label

More information

The Monitoring of Metabolic Control for Patients With Diabetes Mellitus

The Monitoring of Metabolic Control for Patients With Diabetes Mellitus your lab focus interpretation [chemistry hematology generalist] The Monitoring of Metabolic Control for Patients With Diabetes Mellitus Anthony L. McCall, MD, PhD, FACP, 1,2 Nancy Allison, RN, ANP, CDE,

More information

Williamson County Diabetes Care Plan. Member Guide

Williamson County Diabetes Care Plan. Member Guide Williamson County Diabetes Care Plan Member Guide Administered by Allegiance Benefit Plan Management, Inc. providing the Cigna Open Access Network and CVS CareMark Pharmacy Services Welcome Williamson

More information

Means, standard deviations and. and standard errors

Means, standard deviations and. and standard errors CHAPTER 4 Means, standard deviations and standard errors 4.1 Introduction Change of units 4.2 Mean, median and mode Coefficient of variation 4.3 Measures of variation 4.4 Calculating the mean and standard

More information

18.6.1 Terms concerned with internal quality control procedures

18.6.1 Terms concerned with internal quality control procedures 18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date

More information

Nova Max Blood Glucose Monitor Owner s Guide

Nova Max Blood Glucose Monitor Owner s Guide Nova Max Blood Glucose Monitor Owner s Guide Nova Biomedical 200 Prospect Street Waltham, MA 02454-9141 U.S.A. Telephone: 1-800-681-7390 Web site: www.novacares.com Made in the USA by Nova Biomedical Corporation

More information

Feasibility of Engaging Underserved Diabetes Patients in a Web-based Personal Health Record to Facilitate Care Outcomes:

Feasibility of Engaging Underserved Diabetes Patients in a Web-based Personal Health Record to Facilitate Care Outcomes: Feasibility of Engaging Underserved Diabetes Patients in a Web-based Personal Health Record to Facilitate Care Outcomes: Michelle Magee, MD Carine Nassar, RD, MS, CDE MedStar Diabetes, Research and Innovation

More information

Section 14 Simple Linear Regression: Introduction to Least Squares Regression

Section 14 Simple Linear Regression: Introduction to Least Squares Regression Slide 1 Section 14 Simple Linear Regression: Introduction to Least Squares Regression There are several different measures of statistical association used for understanding the quantitative relationship

More information

Calcium. Table 1: Difference between method means in percent

Calcium. Table 1: Difference between method means in percent Calcium Measurement of total calcium is widely used for both the diagnosis and the monitoring of a range of conditions related to the bones, heart, nerves, and kidneys. Total calcium measurements include

More information

Report. European Co-ordinated Post Market Surveillance Operation under COEN authority. Blood Glucose Meters

Report. European Co-ordinated Post Market Surveillance Operation under COEN authority. Blood Glucose Meters 1 v 3 Report European Co-ordinated Post Market Surveillance Operation under COEN authority Blood Glucose Meters Afssaps: France Irish Medicines Board: Ireland MHRA: United Kingdom 2 Summary I - INTRODUCTION...

More information

SMF Awareness Seminar 2014

SMF Awareness Seminar 2014 SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD

More information

Accu-Chek Connect Online diabetes management system. Setup Guide for Healthcare Providers

Accu-Chek Connect Online diabetes management system. Setup Guide for Healthcare Providers Accu-Chek Connect Online diabetes management system Setup Guide for Healthcare Providers Setup Accu-Chek Connect Online diabetes management system Welcome to the Accu-Chek Connect Online diabetes management

More information

Ketones & diabetes; Reduce your risk!

Ketones & diabetes; Reduce your risk! Ketones & diabetes; Reduce your risk! Booklet includes Sick Day Wallet Card Ketones 101 Do I really need to know this? by Helen Jones RN, MSN, CSE YES! The word Ketone is important for you to know if you

More information

Onboard Quality Controls for the CoaguChek XS System

Onboard Quality Controls for the CoaguChek XS System Onboard Quality Controls for the CoaguChek XS System The new CoaguChek XS Anticoagulation Monitoring System incorporates a new onboard quality control system that eliminates the need for liquid controls

More information

Prognostic impact of uric acid in patients with stable coronary artery disease

Prognostic impact of uric acid in patients with stable coronary artery disease Prognostic impact of uric acid in patients with stable coronary artery disease Gjin Ndrepepa, Siegmund Braun, Martin Hadamitzky, Massimiliano Fusaro, Hans-Ullrich Haase, Kathrin A. Birkmeier, Albert Schomig,

More information

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD

More information

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i

More information

5. Multiple regression

5. Multiple regression 5. Multiple regression QBUS6840 Predictive Analytics https://www.otexts.org/fpp/5 QBUS6840 Predictive Analytics 5. Multiple regression 2/39 Outline Introduction to multiple linear regression Some useful

More information

CORRELATION ANALYSIS

CORRELATION ANALYSIS CORRELATION ANALYSIS Learning Objectives Understand how correlation can be used to demonstrate a relationship between two factors. Know how to perform a correlation analysis and calculate the coefficient

More information

This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System.

This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System. GLUCOSE 3L82-20 and 3L82-40 30-3941/R4 GLUCOSE This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This

More information

1/27/2013. PSY 512: Advanced Statistics for Psychological and Behavioral Research 2

1/27/2013. PSY 512: Advanced Statistics for Psychological and Behavioral Research 2 PSY 512: Advanced Statistics for Psychological and Behavioral Research 2 Introduce moderated multiple regression Continuous predictor continuous predictor Continuous predictor categorical predictor Understand

More information

Remote Accessibility to Diabetes Management and Therapy in Operational Healthcare Networks REACTION (FP7 248590) D3-10

Remote Accessibility to Diabetes Management and Therapy in Operational Healthcare Networks REACTION (FP7 248590) D3-10 Remote Accessibility to Diabetes Management and Therapy in Operational Healthcare Networks REACTION (FP7 248590) D3-10 Survey on commercially available CGM devices and insulin pumps Survey Report Document

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding

More information

Welcome to Penzberg. One of Europe s largest biotech facilities

Welcome to Penzberg. One of Europe s largest biotech facilities Welcome to Penzbeg One of Euope s lagest biotech facilities The Roche Goup s lagest biotech facility The Penzbeg site Pofile Role Diagnostics Phamaceuticals Futue Whee it all began The Penzbeg site in

More information

Spectrophotometer - Milton Roy Spectronic 21D or equivalent.

Spectrophotometer - Milton Roy Spectronic 21D or equivalent. 3M COMPANY OCCUPATIONAL HEALTH AND ENVIRONMENTAL SAFETY DIVISION DETERMINATION OF FORMALDEHYDE VAPORS IN AIR USING 3M 3721 FORMALDEHYDE MONITORS May, 2002 SCOPE This procedure covers the method of collecting

More information

Real-time PCR: Understanding C t

Real-time PCR: Understanding C t APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence

More information

What do you know about HbA1C

What do you know about HbA1C Did you know? Essential advice for people with diabetes from Accu-Chek HbA1C is a measure of your average blood glucose level over 3 months Your HbA1C target should be under 53mmol/mol It is important

More information

Course outline. Code: MLS211 Title: Medical Biochemistry

Course outline. Code: MLS211 Title: Medical Biochemistry Course outline Code: MLS211 Title: Medical Biochemistry Faculty of: Science, Health, Education and Engineering Teaching Session: Semester 2 Year: 2015 Course Coordinator: Dr Mark Holmes Tel: 5430 2844

More information

Advisors: Masahiko Saito, Mathematics and Statistics Scott Campbell, Chemical & Biomedical Engineering. Problem Suggested By: Scott Campbell

Advisors: Masahiko Saito, Mathematics and Statistics Scott Campbell, Chemical & Biomedical Engineering. Problem Suggested By: Scott Campbell Undergraduate Journal of Mathematical Modeling: One + Two Volume 3 2011 Spring Issue 2 Article 12 Blood Glucose Levels Carlos Estela University of South Florida Advisors: Masahiko Saito, Mathematics and

More information

Blood Glucose Management

Blood Glucose Management Blood Glucose Management What Influences Blood Sugar Levels? There are three main things that influence your blood sugar: Nutrition Exercise Medication What Influences Blood Sugar Levels? NUTRITION 4 Meal

More information

How To Run Statistical Tests in Excel

How To Run Statistical Tests in Excel How To Run Statistical Tests in Excel Microsoft Excel is your best tool for storing and manipulating data, calculating basic descriptive statistics such as means and standard deviations, and conducting

More information

The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C:

The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C: The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C: The good correlation allows close estimation of GFR Cystatin C GFR GFR in serum estimated* measured* n

More information

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6 I know my value CoaguChek XS Plus system Smart INR monitoring at your practice 1 of 6 CoaguChek XS Plus system A better choice for your patients and your practice There has been an extraordinary increase

More information

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries Basic Statistics and Data Analysis for Health Researchers from Foreign Countries Volkert Siersma siersma@sund.ku.dk The Research Unit for General Practice in Copenhagen Dias 1 Content Quantifying association

More information