Renee Scott, Brad S. Karon, Laurie Griesmann, Sandra C. Bryant, Jeffrey A. DuBois, Terry L. Shirey, Steven Presti, and Paula J.
|
|
- Anis Williamson
- 7 years ago
- Views:
Transcription
1 ABSTRACT We evaluated four hospital-based glucose meter technologies for accuracy, precision, and analytical interferences likely to be encountered in critically ill patients. Precision of all four glucose meters, both within run and day-to-day, instruments. Accuracy (bias) of the meters and analytical interference were evaluated by comparing results obtained on whole blood specimens to plasma samples obtained from these whole blood specimens run on a hexokinase reference method. There were significant differences in the degree to which the meters correlated with a reference hexokinase method. Ascorbic acid showed significant interference with 3 of the 4 meters. Hematocrit also affected the correlation between whole blood and plasma hexokinase glucose on 3 of the 4 glucose meters tested, with the magnitude of this interference also varied by glucose meter technology. The use of glucose meters to maintain tight glycemic control in critically ill patients may be impacted by the accuracy and hematocrit dependence of some glucose meter technologies. GLUCOSE METHODS USED Accu-Chek Inform (Roche Diagnostics, Indianapolis, IN) Glucose dehydrogenase based amperometric strip Precision PCx (Abbott Diabetes, Alameda, CA) Glucose oxidase amperometric detection SureStepFlexx (LifeScan, Malpitas, CA) Photometric glucose oxidase detection StatStrip (Nova Biomedical, Waltham, MA) Modified glucose oxidase based amperometric test system Roche Integra 400 Analyzer (Roche Diagnostics, Indianapolis, IN) Hexokinase method TABLE 1 Within Run and Day-to-Day precision for glucose meters, in percent coefficient of variation (CV), n= number of replicates tested Within Run Precision, Day-to-Day Precision, CV (%) n = 20 CV (%) n = 30 Meter Low Medium High Low High Technology Glucose Glucose Glucose Glucose Glucose mg/dl mg/dl mg/dl mg/dl mg/dl StatStrip Accu-Chek PCx SureStep The coefficient of variation (CV) obtained from both within-run and day-to-day precision experiments was less than 5% for all meters tested except day-to-day precision at low glucose on PCx which was 5.1% STUDY DESIGN Method correlation was performed by analyzing 133 whole blood specimens and 52 spiked whole blood patient specimens on the 4 glucose meters, compared to plasma obtained from those specimens and run on the Integra (reference method). Percent bias (meter result minus plasma result/plasma result) plotted as a function of hematocrit for the 133 whole blood specimens. Interference studies were performed using ascorbic acid added to whole blood at 4 different glucose levels. Six strips tested on each meter at each concentration of interferant, mean used for data analysis. Clinically significant interference defined as > 10 mg/dl(for glucose concentration < 100 mg/dl) or > 10% (for glucose concentration > 100 mg/dl) change from baseline. Hematocrit interference was tested using 3 glucose concentrations over a 25-60% hematocrit range, using six strips for each meter at each hematocrit range tested. Clinically significant effect of hematocrit defined as > 10 mg/dl(for glucose concentrations < 100 mg/dl) or > 10% (for glucose concentrations > 100 mg/dl) change between lowest and highest hematocrit tested. TABLE 2 Correlation data for glucose meters vs reference plasma hexokinase method, n=185 or 133 Meter Slope Intercept (mg/dl) r 2 Median bias (mg/dl) Technology N=185 N=133 N=185 N=133 N=185 N=133 N=185 N=133 StatStrip Accu-Chek PCx SureStep n=133 data set, whole blood samples n=185 data set, 133 whole blood samples and 52 spiked whole blood samples StatStrip and Accu-Chek meters correlate better with the reference hexokinase method, as measured by slope and intercept Median bias was lower for the StatStrip and SureStep meters 2007 Mayo Foundation for Medical Education and Research 11 1
2 12
3 13
4 Glucose meter percent bias was plotted against hematocrit values for the 133 patient sample set (no sample manipulation). There is a clear trend for negative bias associated with increasing hematocrit for the PCx and SureStep meters. The Accu-Chek and StatStrip were largely unaffected by hematocrit. 14
5 CONCLUSION Glucose meters vary in the degree of correlation to the hexokinase reference glucose methods. Most glucose meters continue to be affected by analytical interferences, especially hematocrit. The new StatStrip glucose meter technology correlated closely with a reference hexokinase glucose method and did not demonstrate clinically significant interference from any of the substances tested. 15
SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)
WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory
More informationUSING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
More informationBlood Glucose Monitoring: The Facts about Accuracy
Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined
More informationAn important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and
An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and CoaguChek systems Cardiovascular disease: the #1 killer in
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from
More informationFDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM
More informationA comparison of blood glucose meters in Australia
Diabetes Research and Clinical Practice 71 (2006) 113 118 www.elsevier.com/locate/diabres A comparison of blood glucose meters in Australia Matthew Cohen *, Erin Boyle, Carol Delaney, Jonathan Shaw International
More informationSKUP Scandinavian evaluation of laboratory equipment for primary health care
SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Aviva Meter and test strips designed for glucose self-measurement Manufactured by Roche Diagnostics GmbH Report from
More informationMethod Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
More informationAccuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care
Accuracy Test Omnitest Blood Glucose Monitoring System B. Braun Diabetes Care Accuracy Test Omnitest blood glucose monitoring system according to ISO 97 The accuracy study of Omnitest blood glucose monitoring
More informationHow To Test For Creatinine
Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed
More informationBlood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff
Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff GUIDANCE guidance. Submit written comments to the Division of
More informationPROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.
Department Of Pathology POC.514.01- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 1 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.01 LBH 1 OF 7 Printed copies are
More informationApplying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven
More informationData Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial
Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few
More informationBlood Sugar Monitoring
Sugar Monitoring Know Your Sugar Numbers Taking control of your diabetes and blood glucose (blood sugar) can make you feel better and stay healthy. Research shows that maintaining your blood glucose level
More informationPractice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013
Practice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013 For the person with diabetes, blood glucose monitoring provides immediate feedback on the effects of daily activities such as taking
More informationContent Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
More informationPerformance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers
Journal of Diabetes Science and Technology Volume 7, Issue 1, January 2013 Diabetes Technology Society ORIGINAL ARTICLE Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems:
More informationTitle: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:
More informationValidation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
More information510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2
Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
More informationUS Diabetic Devices Industry Research Report: KenResearch
US Diabetic Devices Industry Research Report: KenResearch by KenResearch - Monday, February 24, 2014 https://www.kenresearch.com/blog/2014/02/us-diabetic-devices-industry-research-report-kenresearch/ The
More informationCenter for Clinical Standards and Quality/Survey & Certification Group
DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality/Survey
More informationQuick Reference Guide
Quick Reference Guide Look inside for step-by-step instructions on how to: >> Transfer readings from your OneTouch Meter >> View and share any of the 11 available reports >> Add information to a patient
More informationGuidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
More informationCountry-Specific Glucose Monitor List
Country-Specific Glucose Monitor List Country Name: United States Important Information This is a non-comprehensive list, current as of March 2016. Absence of a specific glucose monitor or test strips
More informationVALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
More informationApplication Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO
Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban
More informationA Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved
A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved Overview and Objectives This module provides an introduction to
More informationComparison of Glucose Determinations on Blood Samples Collected in Three Types of Tubes
278 Available online at www.annclinlabsci.org Comparison of Glucose Determinations on Blood Samples Collected in Three Types of Tubes Geling Li 1,2, Michael Cabanero 1,2, Zhenglong Wang 1,2, Huiying Wang
More informationBuyers guide. Blood glucose systems CEP 08008
Buyers guide Blood glucose systems CEP 08008 May 2008 Contents 2 Introduction... 3 Technical considerations... 5 Operational considerations... 9 Economic considerations... 14 Purchasing considerations...
More informationHEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.
More informationBiochemistry Validation Form
Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August
More informationA guidebook for people with diabetes
A guidebook for people with diabetes This booklet is designed to supplement, not replace, your doctor s advice. Please consult your doctor if you have any questions about what you read. You ll learn how
More informationIT Risk Management. Agenda. Roche at a glance OSI Layers above 7 What we did What we plan to do how to not get shoot
IT Risk Management 30. 2006 ISACA Switzerland Chapter After Hours Seminar Mario Walter Agenda Roche at a glance OSI Layers above 7 What we did What we plan to do how to not get shoot 2006 ISACA After Hours
More informationWorldwide Markets and Emerging Technologies for Point-of-Care Testing, 2013-2019. February 2015
Worldwide Markets and Emerging Technologies for Point-of-Care Testing, 2013-2019 February 2015 INTELAB CORPORATION MARKET AND TECHNOLOGY REPORTS Report 530 Copyright 2015 Worldwide Markets and Emerging
More informationAnalytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods
Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness
More informationPRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL
More informationICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
More informationMouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma
Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use
More informationQuantifying measurement error from digital instruments
Quantifying measurement error from digital instruments W. BLAKE LAING AND SEAN BRYANT SOUTHERN ADVENTIST UNIVERSITY CHAT TANOOGA, TN What I m doing HELPING STUDENTS LEARN TO CONSTRUCT KNOWLEDGE First lab:
More informationValidation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
More informationSample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
More informationInform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM
Inform II BLOOD GLUCOSE MONITORING SYSTEM Quick Reference Guide for Healthcare Professionals 2 Table of Contents Important... 4 General Information... 6 ACCU-CHEK Inform II Meter...10 How to Perform Patient
More informationLead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
More informationTest strips and meter designed for glucose self-measurement manufactured by Abbott. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care Precision Xtra Plus (G3c) Precision Xceed Test strips and meter designed for glucose self-measurement manufactured by Abbott
More informationCorrelation. What Is Correlation? Perfect Correlation. Perfect Correlation. Greg C Elvers
Correlation Greg C Elvers What Is Correlation? Correlation is a descriptive statistic that tells you if two variables are related to each other E.g. Is your related to how much you study? When two variables
More informationHow to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
More informationTextbook: Chapter 12 pages 351-353, Chapter 13 pages 385-388
EXERCISE 11: BEDSIDE GLUCOSE TESTING Textbook: Chapter 12 pages 351-353, Chapter 13 pages 385-388 Skills: Objectives: 15 points 1. List the limitations of the bedside glucose test procedure. 2. State the
More informationANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
More informationAccu-Chek Inform II system. Professional glucose testing for the wireless age
Accu-Chek Inform II system Professional glucose testing for the wireless age The Accu-Chek Inform II system Real time control The Accu-Chek Inform II System overview Quality Advanced chemistry with multiple
More informationBasic Lessons in Laboratory Quality Control
Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and
More informationfor patient monitoring
Thermo Scientific Indiko Reagents for Immunosuppressant Drug Testing Complete solutions for patient monitoring The cost-effective Thermo Scientific Indiko benchtop analyzer and bar-coded reagent is designed
More informationStep-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
More informationSymmetry in Diagnostics HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. TOSOH BIOSCIENCE www.tosohbioscience.us
Symmetry in Diagnostics HPLC ANALYZER Gold Standard Accuracy by Ion-Exchange HbA1c TOSOH BIOSCIENCE www.tosohbioscience.us HPLC ANALYZER HbA1c analysis in 1.6 minutes Direct determination of stable HbA1c
More informationFor people with diabetes. A blood sugar diary
For people with diabetes A blood sugar diary A circle of help to live a healthy life YOU are at the center of a healthy life with diabetes. All the elements of good care begin and end with YOU. The importance
More informationEffects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose Analyzer
Clinical Chemistry / DRUG EFFECTS AND POINT-OF-CARE GLUCOSE TESTING Effects of Drugs on Glucose Measurements With Handheld Glucose Meters and a Portable Glucose Analyzer Zuping Tang, MD, Xiaogu Du, MD,
More informationSAMPLE. Effects of Different Sample Types on Glucose Measurements
1st Edition POCT06 Effects of Different Sample Types on Glucose Measurements This report provides information to assist the clinical and point-ofcare staff in result and measurement procedure comparisons
More informationD-Facts. Vitamin D Competitive Evaluation
D-Facts Vitamin D Competitive Evaluation Abstract 101 patient samples collected across sites in Germany were used to compare the performance of the Siemens ADVIA Centaur Vitamin D Total (VitD), Abbott
More information1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number
1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression
More informationAlignment and Preprocessing for Data Analysis
Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise
More informationQuality Assurance/Quality Control in Acid Deposition Monitoring
Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial
More informationBuilding Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs
Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs Elizabeth Herron URI Watershed Watch/ National Facilitation of CSREES Volunteer Monitoring Ingrid Harrald Cook
More informationOneStep Fecal Occult Blood RapiDip InstaTest. Cat # 13020-1
CORTEZ DIAGNOSTICS, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: onestep@rapidtest.com Web site: www.rapidtest.com See external label
More informationThe Monitoring of Metabolic Control for Patients With Diabetes Mellitus
your lab focus interpretation [chemistry hematology generalist] The Monitoring of Metabolic Control for Patients With Diabetes Mellitus Anthony L. McCall, MD, PhD, FACP, 1,2 Nancy Allison, RN, ANP, CDE,
More informationWilliamson County Diabetes Care Plan. Member Guide
Williamson County Diabetes Care Plan Member Guide Administered by Allegiance Benefit Plan Management, Inc. providing the Cigna Open Access Network and CVS CareMark Pharmacy Services Welcome Williamson
More informationMeans, standard deviations and. and standard errors
CHAPTER 4 Means, standard deviations and standard errors 4.1 Introduction Change of units 4.2 Mean, median and mode Coefficient of variation 4.3 Measures of variation 4.4 Calculating the mean and standard
More information18.6.1 Terms concerned with internal quality control procedures
18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable
More informationFor In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date
More informationNova Max Blood Glucose Monitor Owner s Guide
Nova Max Blood Glucose Monitor Owner s Guide Nova Biomedical 200 Prospect Street Waltham, MA 02454-9141 U.S.A. Telephone: 1-800-681-7390 Web site: www.novacares.com Made in the USA by Nova Biomedical Corporation
More informationFeasibility of Engaging Underserved Diabetes Patients in a Web-based Personal Health Record to Facilitate Care Outcomes:
Feasibility of Engaging Underserved Diabetes Patients in a Web-based Personal Health Record to Facilitate Care Outcomes: Michelle Magee, MD Carine Nassar, RD, MS, CDE MedStar Diabetes, Research and Innovation
More informationSection 14 Simple Linear Regression: Introduction to Least Squares Regression
Slide 1 Section 14 Simple Linear Regression: Introduction to Least Squares Regression There are several different measures of statistical association used for understanding the quantitative relationship
More informationCalcium. Table 1: Difference between method means in percent
Calcium Measurement of total calcium is widely used for both the diagnosis and the monitoring of a range of conditions related to the bones, heart, nerves, and kidneys. Total calcium measurements include
More informationReport. European Co-ordinated Post Market Surveillance Operation under COEN authority. Blood Glucose Meters
1 v 3 Report European Co-ordinated Post Market Surveillance Operation under COEN authority Blood Glucose Meters Afssaps: France Irish Medicines Board: Ireland MHRA: United Kingdom 2 Summary I - INTRODUCTION...
More informationSMF Awareness Seminar 2014
SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD
More informationAccu-Chek Connect Online diabetes management system. Setup Guide for Healthcare Providers
Accu-Chek Connect Online diabetes management system Setup Guide for Healthcare Providers Setup Accu-Chek Connect Online diabetes management system Welcome to the Accu-Chek Connect Online diabetes management
More informationKetones & diabetes; Reduce your risk!
Ketones & diabetes; Reduce your risk! Booklet includes Sick Day Wallet Card Ketones 101 Do I really need to know this? by Helen Jones RN, MSN, CSE YES! The word Ketone is important for you to know if you
More informationOnboard Quality Controls for the CoaguChek XS System
Onboard Quality Controls for the CoaguChek XS System The new CoaguChek XS Anticoagulation Monitoring System incorporates a new onboard quality control system that eliminates the need for liquid controls
More informationPrognostic impact of uric acid in patients with stable coronary artery disease
Prognostic impact of uric acid in patients with stable coronary artery disease Gjin Ndrepepa, Siegmund Braun, Martin Hadamitzky, Massimiliano Fusaro, Hans-Ullrich Haase, Kathrin A. Birkmeier, Albert Schomig,
More informationSamsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD
More informationDevelopment and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
More information5. Multiple regression
5. Multiple regression QBUS6840 Predictive Analytics https://www.otexts.org/fpp/5 QBUS6840 Predictive Analytics 5. Multiple regression 2/39 Outline Introduction to multiple linear regression Some useful
More informationCORRELATION ANALYSIS
CORRELATION ANALYSIS Learning Objectives Understand how correlation can be used to demonstrate a relationship between two factors. Know how to perform a correlation analysis and calculate the coefficient
More informationThis package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System.
GLUCOSE 3L82-20 and 3L82-40 30-3941/R4 GLUCOSE This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes Highlighted NOTE: This
More information1/27/2013. PSY 512: Advanced Statistics for Psychological and Behavioral Research 2
PSY 512: Advanced Statistics for Psychological and Behavioral Research 2 Introduce moderated multiple regression Continuous predictor continuous predictor Continuous predictor categorical predictor Understand
More informationRemote Accessibility to Diabetes Management and Therapy in Operational Healthcare Networks REACTION (FP7 248590) D3-10
Remote Accessibility to Diabetes Management and Therapy in Operational Healthcare Networks REACTION (FP7 248590) D3-10 Survey on commercially available CGM devices and insulin pumps Survey Report Document
More informationGUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
More informationAssay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff
Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding
More informationWelcome to Penzberg. One of Europe s largest biotech facilities
Welcome to Penzbeg One of Euope s lagest biotech facilities The Roche Goup s lagest biotech facility The Penzbeg site Pofile Role Diagnostics Phamaceuticals Futue Whee it all began The Penzbeg site in
More informationSpectrophotometer - Milton Roy Spectronic 21D or equivalent.
3M COMPANY OCCUPATIONAL HEALTH AND ENVIRONMENTAL SAFETY DIVISION DETERMINATION OF FORMALDEHYDE VAPORS IN AIR USING 3M 3721 FORMALDEHYDE MONITORS May, 2002 SCOPE This procedure covers the method of collecting
More informationReal-time PCR: Understanding C t
APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence
More informationWhat do you know about HbA1C
Did you know? Essential advice for people with diabetes from Accu-Chek HbA1C is a measure of your average blood glucose level over 3 months Your HbA1C target should be under 53mmol/mol It is important
More informationCourse outline. Code: MLS211 Title: Medical Biochemistry
Course outline Code: MLS211 Title: Medical Biochemistry Faculty of: Science, Health, Education and Engineering Teaching Session: Semester 2 Year: 2015 Course Coordinator: Dr Mark Holmes Tel: 5430 2844
More informationAdvisors: Masahiko Saito, Mathematics and Statistics Scott Campbell, Chemical & Biomedical Engineering. Problem Suggested By: Scott Campbell
Undergraduate Journal of Mathematical Modeling: One + Two Volume 3 2011 Spring Issue 2 Article 12 Blood Glucose Levels Carlos Estela University of South Florida Advisors: Masahiko Saito, Mathematics and
More informationBlood Glucose Management
Blood Glucose Management What Influences Blood Sugar Levels? There are three main things that influence your blood sugar: Nutrition Exercise Medication What Influences Blood Sugar Levels? NUTRITION 4 Meal
More informationHow To Run Statistical Tests in Excel
How To Run Statistical Tests in Excel Microsoft Excel is your best tool for storing and manipulating data, calculating basic descriptive statistics such as means and standard deviations, and conducting
More informationThe sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C:
The sensitive marker for glomerular filtration rate (GFR) Estimation of GFR from Serum Cystatin C: The good correlation allows close estimation of GFR Cystatin C GFR GFR in serum estimated* measured* n
More informationI know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6
I know my value CoaguChek XS Plus system Smart INR monitoring at your practice 1 of 6 CoaguChek XS Plus system A better choice for your patients and your practice There has been an extraordinary increase
More informationBasic Statistics and Data Analysis for Health Researchers from Foreign Countries
Basic Statistics and Data Analysis for Health Researchers from Foreign Countries Volkert Siersma siersma@sund.ku.dk The Research Unit for General Practice in Copenhagen Dias 1 Content Quantifying association
More information