Basics of Real Time 3D Echocardiography
|
|
|
- Jordan Cain
- 9 years ago
- Views:
Transcription
1 Basics of Real Time 3D Echocardiography J.S. McGhie W.B. Vletter F. Meijboom R. Frowijn H. Van de Zwaan No disclosures
2 Outline Image aquisition and manipulation Introduction 3D imaging Data set manipulation
3 Outline Image aquisition and manipulation Introduction 3D imaging Data set manipulation
4 Ultrasound imaging modalities in Cardiology
5 xmatrix array technology utilizes 2400 fully-sampled elements for 360-degree focusing and steering xmatrix X3-1 transducer
6 RT3DE transducers ie33 Purewave technology wider bandwith higher sensitivity X3-1 X7-2 X7-2t 24 x 15 TTE adult >20 kg TTE pediatric <20 kg 24 x 20 TEE adult 24 x 15 Footprint 15 x 24mm Footprint 15 x 20mm
7 Image aquisition and manipulation Outline Introduction 3D imaging Data set manipulation
8 3D Echo Protocol Imaging modes LIVE 3D small sector - thin slice acquisition zoom mode (enlargement of a subsegment of this sector) xplane (lateral tilt image)
9 3D Echo Protocol Imaging modes LIVE 3D small sector - thin slice acquisition zoom mode (enlargement of a sub segment of this sector) xplane (lateral tilt image)
10 Small sector Thin slice mode The current display format suffers from attempting to show 3D images on a 2D display
11 Small sector Thin slice mode Real-time datase (sector 30 x 60 ) Full-volume dataset not real-time (sector 101 x 104 )
12 LIVE 3D 3D Echo Protocol Imaging modes small sector- thin slice acquisition zoom mode (enlargement of a subsegment of this sector) xplane (lateral tilt image)
13 LIVE 3D Zoom mode (TEE)
14 LIVE 3D Zoom mode (TEE)
15 LIVE 3D 3D Echo Protocol Imaging modes small sector- thin slice acquisition zoom mode (enlargement of a subsegment of this sector) xplane (lateral tilt 2D image)
16 xplane (a cross section and its orthogonal plane)
17 xplane (a cross section and its orthogonal plane) Lateral tilting
18 2D to 3D full volume
19 Full Volume not real-time multiple subvolumes are necessary because of limited maximal scan sector Maximal pyramidal dimensions X7: 92 x 83 X4: 84 x 84 X3-1: 101 x 104
20 3D Echo Protocol: Full Volume Patient and machine preparation 1. Good ECG signal with clear R-wave (3D Full volume triggering) 2. Adjust machine settings (follow the rules as for 2D) for the best 3D resolution: Harmonics fundamental, adjust gain setting (clear blood-tissue border - minimize noise) Region of interest between the dotted lines Minimize sector ( angle, depth, density) Maximise number of subvolumes Increase overall gain before recording
21 3D aquisition work flow
22 3D aquisition work flow: Gainsetting Overall gain TGC
23 3D Analysis Cardiac Function Cardiac morpholgy
24 Image optimisation for the LV and RV Priority: Good quality 2D image Frame rate as high as possible - acquisition beats: 7 - optimise depth setting Complete ventricle in dataset - optimise density setting Extreem dilatation - acquisition beats: 5 (wide angle) - notice: drop in frame rate
25 Frame Rate 3D Volume Data Set Ultrasound speed : 1500 m/sec Depth of 16cm : transmitting + receiving time : 220 micro sec 2D ( 90 scanlines ) requires 19,8 msec FR : 50 Hz 3D ( 2400 scanlines ) requires 528 msec FR : 1,9 Hz How to improve the frame rate Receive multiple beams for each transmit event (parallel processing) Decrease image size either laterally or in depth Acquire data over multiple cardiac cycles ( 4 7 ) and build a composite volume data set
26 Depth: frame rate 35 Hz Depth 13 cm 30 Hz Depth 16 cm
27 Number of subvolumes frame rate 20 Hz 35 Hz
28 Wide angle frame rate 35 Hz 18 Hz Wide angle
29 Subvolumes Reasons to use less subvolumes Arrhythmias / severe bradycardia Breathhold problems Big heart: use wide angle / 5 subvolumes
30 Line density frame rate 35 Hz Low 35 Hz High
31 Line density sector size High density Low density
32 2D images within the 3D data set Multiplane reconstruction
33 What do I trace? Endocardial boarder detection using contrast
34 LV volume and ejection fraction
35 Echo drop out in full volume data set Sometimes unavoidable!
36 2D: RV view before RT3DE acquisition Apical four-chamber view Laterally modified apical four-chamber view
37 Full volume RT3DE RV: acquisition dataset RV volumes and ejection fraction. For the first time possible with echo!
38 What do I trace? RV endocardial boarder detection using contrast
39 Real-time 3D echo: analyse
40 3D Analysis Cardiac Function Cardiac morpholgy
41 Dotted lines: ROI = mitral valve 90º Bi-plane view mitral valve
42 Normal Mitral Valve
43 Resolution: the best plane? Dimension Axial Lateral Elevation Resolution Best Medium Least ~0.5 mm ~3.0 mm ~2.5 mm
44 Ultrasound beam perpendicular to ROI
45 Full Volume Colour Parasternal LAX Parasternal SAX MV seen from the LA
46 3D Echo pitfalls: stitching artifact
47 3D Full volume data set Multiple subvolumes Increased chance of artefact One - shot
48 Enface view IAS:seen from LA Drop-out Fossa Ovalis 3D Echo pitfalls: drop out
49 Image aquisition and manipulation Outline Introduction Tips and tricks for an optimal 3D data set Data set manipulation
50 3D Multiplane reconstruction: 4Fallot The ability to move through a 3D dataset in any 2D image plane Interogate the data set on or of-line Allows precise identification and localization of abnormalities Allows quantification.
51 3D analysis the crop-box
52 3D analysis the crop-box
53 Crop till you Drop!!!
54 X5-1: the all in one transducer 2D imaging 2D colour Doppler PW and CW Doppler + 3D imaging Live 3D imaging and colour Live 3D zoom Full volume imaging and colour
55 2D imaging S5-1 X3-1 X5-1
56 3D imaging X3-1 X5-1
57 Conclusions: Image acquisition and manipulation In 3D the cardiac structures are shown in relationship to each other in all three spacial dimensions The cardiac structures can be rotated or viewed from different orientations The ability to move through a 3D data set in any 2D image plane allows better appreciation of cardiac anatomy in complex structural heart disease. 3D images are more intuitive than 2D images allowing quicker appreciation of cardiac anatomy by other health care workers. The clinical role of 3d echocardiography will continue to evolve as technology advances
58 Conclusions: Image acquisition and manipulation But remember : The limitations of 2D are also true for 3D echo 3D echo is not a stand-alone feature, but should be used next to / on top of other echo modalities.
59 Overall gain
60 Overall gain
61 Source: SA of the Heart: Wilcock/Cook/Anderson Left A-V Valve
62 Aortic valve en-face view from Aorta Bicuspid Ao Valve
63 Normal Tricupid Valve From RV
64 Normal Pulmonary Valve Only the closure line of PV is visible! Dynamic RVOT during the cardiac cycle
65 3D reconstruction: atrial septal defect ASD enface view : a dynamic structure
66 3D reconstruction: VSD by 4Fallot(baby) En face VSD from LV En face VSD from RV
67 TEE: Enface view of the IAS ASD II-SVC-Aorta En face view LA: Device insitu
Three Dimensional Ultrasound Imaging
Three Dimensional Ultrasound Imaging Hans Torp/ Sevald Berg/Kjell Kristoffersen m/flere Department of circulation and medical imaging NTNU Hans Torp NTNU, Norway Acquisition Reconstruction Filtering Collecting
RACE I Rapid Assessment by Cardiac Echo. Intensive Care Training Program Radboud University Medical Centre NIjmegen
RACE I Rapid Assessment by Cardiac Echo Intensive Care Training Program Radboud University Medical Centre NIjmegen RACE Goal-directed study with specific questions Excludes Doppler ultrasound Perform 50
GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System
GE Healthcare Vivid 7 Dimension Cardiovascular Ultrasound System New dime Introducing multi-dimensional and 4D imaging. The ability to incorporate multi-dimensional and 4D imaging into your clinical routine
Cardiology Fellowship Manual. Goals & Objectives -Cardiac Imaging- 1 Page
Cardiology Fellowship Manual Goals & Objectives -Cardiac Imaging- 1 Page 2015-2016 UNIV. OF NEBRASKA CHILDREN S HOSPITAL & MEDICAL CENTER DIVISION OF CARDIOLOGY FELLOWSHIP PROGRAM CARDIAC IMAGING ROTATION
020 // Congenital Heart Disease
020 // Congenital Heart Disease CONTENTS 188 Basics 188 Atrial Septal Defect (ASD) 191 Patent Foramen Ovale (PFO) 192 Ventricular Septal Defects (VSD) 194 Patent Ductus Arteriosus (PDA) 195 Coronary Fistulas
Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010
1 Ny teknologi: Ultralyd måler m blodstrøm Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 2 Conventional imaging methods of blood flow using ultrasound
Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality
Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL
Fellow TEE Review Workshop Hemodynamic Calculations 2013. Director, Intraoperative TEE Program. Johns Hopkins School of Medicine
Fellow TEE Review Workshop Hemodynamic Calculations 2013 Mary Beth Brady, MD, FASE Director, Intraoperative TEE Program Johns Hopkins School of Medicine At the conclusion of the workshop, the participants
Real-Time 3-Dimensional Transesophageal Echocardiography in the Evaluation of Post-Operative Mitral Annuloplasty Ring and Prosthetic Valve Dehiscence
Journal of the American College of Cardiology Vol. 53, No. 17, 2009 2009 by the American College of Cardiology Foundation ISSN 0735-1097/09/$36.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2008.12.059
Atrial Septal Aneurysm: A Study in Five Hundred Adult Patients
Premio al mejor Tema Libre Realizado por Médicos Residentes Atrial Septal Aneurysm: A Study in Five Hundred Adult Patients Olivares-Reyes Alexander; Al-Kamme Ahmad; Gonzalez Javier Cardiology Department
FUNÇÃO VENTRICULAR ESQUERDA POR ECO-3D/4D - UM NOVO PARADIGMA DE AVALIAÇÃO? LEFT VENTRICULAR FUNCTION BY 3D/4D-ECHO - A NEW PARADIGM OF ASSESSMENT?
MULTIMODALIDADE DE IMAGEM NA DOENÇA CARDÍACA ISQUÉMICA MULTIPLE MODALITIES OF IMAGING IN ISCHEMIC CARDIAC DISEASE FUNÇÃO VENTRICULAR ESQUERDA POR ECO-3D/4D - UM NOVO PARADIGMA DE AVALIAÇÃO? LEFT VENTRICULAR
Ultrasound Simulators
Ultrasound Simulators Tripp Bell, MD University of South Carolina School of Medicine Objectives Give a brief history of simulation in medicine Describe the types of ultrasound simulators Show how ultrasound
Contents. HD11 XE Quick Guide 4535 612 62661
Contents Using the HD11 XE Help................................................3 Using the HD11 XE Ultrasound System....................................4 Imaging Modes........................................................7
Normal & Abnormal Intracardiac. Lancashire & South Cumbria Cardiac Network
Normal & Abnormal Intracardiac Pressures Lancashire & South Cumbria Cardiac Network Principle Pressures recorded from catheter tip Electrical transducer - wheatstone bridge mechanical to electrical waveform
Premium performance compact ultrasound. Philips CX50 xmatrix CompactXtreme for cardiology
Premium performance compact ultrasound Philips CX50 xmatrix CompactXtreme for cardiology Philips CX50 xmatrix for cardiology While doing portable studies, I really need to be able to count on my compact
What is echo? CHAPTER 1 1.1 BASIC NOTIONS. Ultrasound production and detection
What is echo? CHAPTER 1 1.1 BASIC NOTIONS Echocardiography (echo) the use of ultrasound to examine the heart is a safe, powerful, non-invasive and painless technique. Echo is easy to understand as many
Normal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network
Normal Intracardiac Pressures Lancashire & South Cumbria Cardiac Network Principle Pressures recorded from catheter tip Electrical transducer - wheatstone bridge mechanical to electrical waveform display
Heart Sounds & Murmurs
Cardiovascular Physiology Heart Sounds & Murmurs Dr. Abeer A. Al-Masri MBBS, MSc, PhD Associate Professor Consultant Cardiovascular Physiologist Faculty of Medicine, KSU Detected over anterior chest wall
3D Ultrasound. Outline. What is 3D US? Volume Sonography. 3D Ultrasound in Obstetrics: Current Modalities & Future Potential. Alfred Abuhamad, M.D.
in Obstetrics: Current Modalities & Future Potential Outline What is 3D US? What are obvious advantages of 3D US? What is the future of 3D US? Alfred Abuhamad, M.D. Eastern Virginia Medical School 2D US
CTA OF THE EXTRACORONARY HEART
CTA OF THE EXTRACORONARY HEART Charles White MD Director of Thoracic Imaging Department of Radiology University of Maryland NO DISCLOSURES [email protected] CARDIAC CASE DISTRIBUTION Coronary CTA 30% ED chest
Since the introduction of transesophageal echocardiography
POSITION PAPER ASE/SCA Guidelines for Performing a Comprehensive Intraoperative Multiplane Transesophageal Echocardiography Examination: Recommendations of the American Society of Echocardiography Council
Twelve. Figure 12.1: 3D Curved MPR Viewer Window
Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from
Workshop B: Essentials of Neonatal Cardiology and CHD Anthony C. Chang, MD, MBA, MPH CARDIAC INTENSIVE CARE
SHUNT LESIONS NEONATAL : CONGENITAL CARDIAC MALFORMATIONS AND CARDIAC SURGERY ANTHONY C. CHANG, MD, MBA, MPH CHILDREN S HOSPITAL OF ORANGE COUNTY ATRIAL SEPTAL DEFECT LEFT TO RIGHT SHUNT INCREASED PULMONARY
Ultrasound. Your versatile performer to depend on. Philips ClearVue ultrasound family for women s health care
Ultrasound ClearVue Your versatile performer to depend on Philips ClearVue ultrasound family for women s health care Over 70 countries More than 5,000 systems Touching 25 million lives every year For one
Fall 3 Ekokardiografi 2012-41
Fall 3 Ekokardiografi 2012-41 Fall 3 1. Vilken etiologi har vi till patientens insufficiens? 2. Tillför 3D undersökningen något ytterligare? 3 4 5 6 7 8 9 10 11 12 13 14 15 När är 3D/4D eko bättre än 2D?
Company profile. Quote from the analyst of. Company Milestone. Feel the Difference. Tel: +86-755-26722860 Fax: +86-755-26722850
Company profile Since the establishment, SonoScape has committed to provide high standard medical equipments for healthcare of human beings. SonoScape specializes in the development and production of diagnostic
Electrocardiography I Laboratory
Introduction The body relies on the heart to circulate blood throughout the body. The heart is responsible for pumping oxygenated blood from the lungs out to the body through the arteries and also circulating
How To Treat A Single Ventricle And Fontan
COACH Columbus Ohio Adult Congenital Heart Disease Program The Heart Center at Nationwide Children s Hospital & The Ohio State University Single Ventricle Defects Normal Heart Structure The heart normally
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
Section Four: Pulmonary Artery Waveform Interpretation
Section Four: Pulmonary Artery Waveform Interpretation All hemodynamic pressures and waveforms are generated by pressure changes in the heart caused by myocardial contraction (systole) and relaxation/filling
Example Theory Multiple Choice Questions
Example Theory Multiple Choice Questions Q a). In an ultrasound imaging system: Sector width, sector depth and frame rate can all be controlled independently b) Frame rate falls as sector width increases
A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)
chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy
A new era in premium ultrasound. Philips EPIQ 7 ultrasound system
A new era in premium ultrasound Philips EPIQ 7 ultrasound system The new challenges in global healthcare Unprecedented advances in premium ultrasound performance can help address the strains on overburdened
Compactly Powerful. Ultrasound System UGEO PT60A CT-UGEO PT60A-CWW-CMI-0705-EN
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer, in 110
SAMSUNG ULTRASOUND RS80A
Samsung Medison is a global leading medical device company. Founded in 1985, the company sells cutting-edge diagnostic ultrasound devices around the world in various medical fields. The company has attracted
Clinical Training for Visage 7 Cardiac. Visage 7
Clinical Training for Visage 7 Cardiac Visage 7 Overview Example Usage 3 Cardiac Workflow Examples 4 Remove Chest Wall 5 Edit Chest Wall Removal 6 Object Display Popup 7 Selecting Optimal Phase 8 Thick
Pulmonary Atresia With Intact Ventricular Septum - Anatomy, Physiology, and Diagnostic Imaging
Pulmonary Atresia With Intact Ventricular Septum - Anatomy, Physiology, and Diagnostic Imaging Larry Latson MD Director of Pediatric Interventional Cardiology and Adult CHD Joe DiMaggio Children s Hospital
Diagnostic and Therapeutic Procedures
Diagnostic and Therapeutic Procedures Diagnostic and therapeutic cardiovascular s are central to the evaluation and management of patients with cardiovascular disease. Consistent with the other sections,
Philips CX50 CompactXtreme System. A New Class of Compact Echocardiography
Philips CX50 CompactXtreme System A New Class of Compact Echocardiography Insights from compact ultrasound users Diagnostic confidence is paramount Ultrasound labs are busy and workflow efficiency is crucial
Left to Right Shunts and their Calculation. Ghada El Shahed, MD
Left to Right Shunts and their Calculation Ghada El Shahed, MD Professor of Cardiology Ain Shams University Flow through systemic & pulmonary circulations is normally balanced and equal. Two circulations
Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals
Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction
Auscultation of the Heart
Review of Clinical Signs uscultation of the Heart Series Editor: Bernard Karnath, MD Bernard Karnath, MD William Thornton, MD uscultation of the heart can provide clues to the diagnosis of many cardiac
The Essential Role of CT Planning and TEE Guidance to Optimize Device Deployment
LAA Occlusion Therapy The Essential Role of CT Planning and TEE Guidance to Optimize Device Deployment John Carroll, MD University of Colorado Interventional 3-D Lab 3-D Research Lab for Interventions
Teaching Med-5 Students Point-of-Care Transthoracic Echocardiography
Teaching Med-5 Students Point-of-Care Transthoracic Echocardiography Anthony M.-H. Ho, Lester A. H. Critchley, Patricia Kan, Sylvia Au, Siu Keung Ng, Simon K. C. Chan, Philip Lam, Gordon Choi, Alex Lee,
Understanding your child s heart Atrial septal defect
Understanding your child s heart Atrial septal defect About this factsheet This factsheet is for the parents of babies and children who have an atrial septal defect (ASD). It explains, what an atrial septal
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard
How To Teach An Integrated Ultrasound
University of South Carolina School of Medicine Integrated Ultrasound Curriculum iusc Richard Hoppmann The Integrated Ultrasound Curriculum Initiated 2006 First (M1) and Second (M2) Year Medical Students
Common types of congenital heart defects
Common types of congenital heart defects Congenital heart defects are abnormalities that develop before birth. They can occur in the heart's chambers, valves or blood vessels. A baby may be born with only
Samsung Ultrasound H60
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer around
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Improve your image. ClearVue 650 ultrasound system
Improve your image ClearVue 650 ultrasound system Introducing the ClearVue 650 The Philips ClearVue 650 ultrasound system is as versatile as it is elegant. Use it in a variety of settings for many types
Minimally Invasive Mitral Valve Surgery
Minimally Invasive Mitral Valve Surgery Stanford Health Care offers leading, superior options in cardiac surgery, including the latest techniques and research for Minimally Invasive Cardiac surgery. Advanced
State-of-the-Art Technology in Cardiac CT
1 2 Next Step Evolution or Revolution? State-of-the-Art Technology in Cardiac CT Stefan Ulzheimer, PhD Global Director of Collaborations CT Siemens Medical Solutions Major Innovations in CT Head The 80
Note: The left and right sides of the heart must pump exactly the same volume of blood when averaged over a period of time
page 1 HEART AS A PUMP A. Functional Anatomy of the Heart 1. Two pumps, arranged in series a. right heart: receives blood from the systemic circulation (via the great veins and vena cava) and pumps blood
Electrocardiography Review and the Normal EKG Response to Exercise
Electrocardiography Review and the Normal EKG Response to Exercise Cardiac Anatomy Electrical Pathways in the Heart Which valves are the a-v valves? Closure of the a-v valves is associated with which heart
Epicardial Echocardiography and Epiaortic Ultrasonography
Epicardial Echocardiography and Epiaortic Ultrasonography 20 Stanton K. Shernan and Kathryn E. Glas Despite its overwhelming popularity and favorable influence on perioperative clinical decision making
Resuscitation in congenital heart disease. Peter C. Laussen MBBS FCICM Department Critical Care Medicine Hospital for Sick Children Toronto
Resuscitation in congenital heart disease Peter C. Laussen MBBS FCICM Department Critical Care Medicine Hospital for Sick Children Toronto Evolution of Congenital Heart Disease Extraordinary success: Overall
Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice
European Journal of Echocardiography (2009) 10, 1 25 doi:10.1093/ejechocard/jen303 EAE/ASE RECOMMENDATIONS Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice
Heart valve repair and replacement
16 Heart valve repair and replacement 222 Valvular heart disease can be treated in a variety of ways: valve replacement, in which an artificial (prosthetic) heart valve is implanted surgically to replace
Read It, Code It, See It
Read It, Code It, See It Richard L. Prager, M.D. University of Michigan Ann Arbor, Michigan Dorothy Latham, R.N. Port Huron Hospital Port Huron, Michigan Nothing to Disclose Disclosure Preoperative diagnosis:
Universal Fetal Cardiac Ultrasound At the Heart of Newborn Well-being
Universal Fetal Cardiac Ultrasound At the Heart of Newborn Well-being Optimizes detection of congenital heart disease (chd) in the general low risk obstetrical population Daniel J. Cohen, M.D. [email protected]
Question 1: Interpret the rhythm strip above (comment on regularity, rate, P wave, PR interval and QRS)?
It is your first month on your NICU rotation and you are prerounding on your patients. The nurse takes you aside and says she s been seeing something funny on the cardiac monitor. You walk to the bedside
12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P
12-Lead EKG Interpretation Judith M. Haluka BS, RCIS, EMT-P ECG Grid Left to Right = Time/duration Vertical measure of voltage (amplitude) Expressed in mm P-Wave Depolarization of atrial muscle Low voltage
BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY
BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY American Osteopathic Association and the American College of Osteopathic Internists Specific Requirements For Osteopathic Subspecialty Training In Cardiology
17 Endocarditis. Infective endocarditis
17 Endocarditis 234 Endocarditis refers to inflammation of the endocardium, the inner layer of the heart (including the heart valves). Endocarditis can be: infective (e.g. bacterial, fungal) non-infective
Exploring nsight Imaging a totally new architecture for premium ultrasound
Exploring nsight Imaging a totally new architecture for premium ultrasound Philips EPIQ ultrasound system Karl Thiele, Principal Scientist James Jago, Principal Scientist Robert Entrekin, Technical Fellow
Cardiac Tool User s Guide. Cardiac Tool. Version 1.0
Cardiac Tool Version 1.1 by Bernd Jung, Harald Fischer, Michael Markl Section of Medical Physics, University Hospital Freiburg, Germany Contact: Bernd Jung, Hugstetterstr. 55, D 79106 Freiburg, email:
top ten My box ECG transmits a PDF of test results to my EMR, so I am good.
2014 Q1 Diagnostics Promotion #1 COMMOn MYths Decoded Breaking the digital stereotype My box ECG transmits a PDF of test results to my EMR, so I am good. #2 A PDF is a static image and does not contain
The P Wave: Indicator of Atrial Enlargement
Marquette University e-publications@marquette Physician Assistant Studies Faculty Research and Publications Health Sciences, College of 8-12-2010 The P Wave: Indicator of Atrial Enlargement Patrick Loftis
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
Ultrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
The ultrasound system that sets new standards
Samsung Medison is one of the world's leading researchers, developers and manufacturers of ultrasound and other medical imaging products. Founded in 1985, Samsung Medison was acquired by Samsung Electronics
9th Christmas. Echo. International Course on Perioperative and Critical Care. Echocardiography. 19-20 December 2008 Courtyard by Marriott Brussels
9th Christmas Test yourself! Basic course Pre- and Post-course exam Image regognition testing Echo Advanced course Simulated exam for the European Accreditation in Adult Transesophageal Echocardiography
A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer
Presented at IMAPS 2002 Denver, Colorado September 5, 2002 (Best of Session Award) A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer Jeff Strole*, Scott Corbett*,
Doc, I Am Fine, But I Have A Cardiac Condition
Doc, I Am Fine, But I Have A Cardiac Condition Nevine Mahmoud, MD John Ludtke, MD Maj, USAFR, MC, FS RAM Class 2014 Wright State University Boonshoft School of Medicine Division of Aerospace Medicine Dayton,
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
Heart and Vascular System Practice Questions
Heart and Vascular System Practice Questions Student: 1. The pulmonary veins are unusual as veins because they are transporting. A. oxygenated blood B. de-oxygenated blood C. high fat blood D. nutrient-rich
Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.
1 12/9/2008 Page 1 Master s Program in Medical Physics Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III Chair in Faculty of Medicine Mannheim University of Heidelberg Theodor-Kutzer-Ufer
Recommendations for Evaluation of the Severity of Native Valvular Regurgitation with Two-dimensional and Doppler Echocardiography
AMERICAN SOCIETY OF ECHOCARDIOGRAPHY REPORT Recommendations for Evaluation of the Severity of Native Valvular Regurgitation with Two-dimensional and Doppler Echocardiography A report from the American
Echocardiography has become the primary imaging
Guidelines and Standards for Performance of a Pediatric Echocardiogram: A Report from the Task Force of the Pediatric Council of the American Society of Echocardiography Wyman W. Lai, MD, MPH, FASE, Tal
Dysfunction of aortic valve prostheses
Dysfunction of aortic valve prostheses Kai Andersen Oslo University Hospital Rikshospitalet, Norway Dysfunction of aortic valve prostheses Kai Andersen Oslo University Hospital Rikshospitalet, Norway No
Electrophysiology Introduction, Basics. The Myocardial Cell. Chapter 1- Thaler
Electrophysiology Introduction, Basics Chapter 1- Thaler The Myocardial Cell Syncytium Resting state Polarized negative Membrane pump Depolarization fundamental electrical event of the heart Repolarization
High-quality/High-resolution Digital Ultrasound Diagnostic Scanner
Hitachi Review Vol. 52 (2003), No. 4 207 High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Tsuyoshi Otake Toshihiko Kawano Takashi Sugiyama Tsuyoshi Mitake Shinichiro Umemura, Ph. D. OVERVIEW:
Chapter 20: The Cardiovascular System: The Heart
Chapter 20: The Cardiovascular System: The Heart Chapter Objectives ANATOMY OF THE HEART 1. Describe the location and orientation of the heart within the thorax and mediastinal cavity. 2. Describe the
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University
Principles of Medical Ultrasound Pai-Chi Li Department of Electrical Engineering National Taiwan University What is Medical Ultrasound? Prevention: actions taken to avoid diseases. Diagnosis: the process
1p36 and the Heart. John Lynn Jefferies, MD, MPH, FACC, FAHA
1p36 and the Heart John Lynn Jefferies, MD, MPH, FACC, FAHA Director, Advanced Heart Failure and Cardiomyopathy Services Associate Professor, Pediatric Cardiology and Adult Cardiovascular Diseases Associate
American Society of Echocardiography 2014 Coding and Reimbursement Newsletter
May American Society of Echocardiography Coding and Reimbursement Newsletter The ASE Coding and Reimbursement Newsletter is a resource for cardiovascular ultrasound procedures provided in the facility
Recommendations of the European Association of Echocardiography
European Journal of Echocardiography (2011) 12, 339 353 doi:10.1093/ejechocard/jer051 RECOMMENDATIONS Recommendations of the European Association of Echocardiography How to use echo-doppler in clinical
Chapter 2 Cardiac Interpretation of Pediatric Chest X-Ray
Chapter 2 Cardiac Interpretation of Pediatric Chest X-Ray Ra-id Abdulla and Douglas M. Luxenberg Key Facts The cardiac silhouette occupies 50 55% of the chest width on an anterior posterior chest X-ray
Managing Mitral Regurgitation: Repair, Replace, or Clip? Michael Howe, MD Traverse Heart & Vascular
Managing Mitral Regurgitation: Repair, Replace, or Clip? Michael Howe, MD Traverse Heart & Vascular Mitral Regurgitation Anatomy Mechanisms of MR Presentation Evaluation Management Repair Replace Clip
Anatomi & Fysiologi 060301. The cardiovascular system (chapter 20) The circulation system transports; What the heart can do;
The cardiovascular system consists of; The cardiovascular system (chapter 20) Principles of Anatomy & Physiology 2009 Blood 2 separate pumps (heart) Many blood vessels with varying diameter and elasticity
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
DETECTION OF LEFT-TO-RIGHT INTRACARDIAC SHUNTS
10/5/00 3:35 PM 9 Shunt Detection and Quantification William Grossman University of California, San Francisco, School of Medicine; Division of Cardiology, University of California, San Francisco Medical