Guideline for the flow cytometric enumeration of CD34 + haematopoietic stem cells

Size: px
Start display at page:

Download "Guideline for the flow cytometric enumeration of CD34 + haematopoietic stem cells"

Transcription

1 Clin. Lab. Haem. 1999, 21, Guideline for the flow cytometric enumeration of CD34 + haematopoietic stem cells PREPARED BY THE CD34 + HAEMATOPOIETIC STEM CELL WORKING PARTY* *D. Barnett, G. Janossy, A. Lubenko, E. Matutes, A. Newland & J.T. Reilly Members of the General Haematology Task Force of the British Committee for Standards in Haematology: J.T. Reilly (Chairman), B.J. Bain (Secretary), R. Amos, I. Cavill, C. Chapman, K. Hyde, E. Matutes, J. Parker-Williams, I.D. Walker Introduction It is well established that the 3% of cells in the bone marrow which express the CD34 antigen, a heavily glycosylated mucin-like structure, are capable of reconstituting longterm, multilineage haematopoiesis after myeloablative therapy (Berenson et al. 1988; Andrews et al. 1992). CD34 + cells are also found in the peripheral blood of normal individuals, but are extremely rare (approximately % of total nucleated cells). However, current treatment and mobilization regimes, including chemotherapy and/or haematopoietic growth factors, can significantly increase circulating CD34 + stem cell numbers in patients and healthy donors. Peripheral blood stem cells (PBSC) have now virtually replaced bone marrow as the primary source of stem cells for autologous transplantation (Gratwohl, Hermans & Baldomero 1996) and the procedure is being increasingly used for allogeneic transplantation between HLA-identical siblings (Russell, Gratwohl & Schmitz 1996). Potential advantages of PBSC transplantation include a more rapid haematopoietic reconstitution, lower risk of contamination by tumour cells, reduced hospitalization costs, increased numbers of T-lymphocytes and NK cells that may reduce post-transplant relapse, and the elimination of a need for general anaesthesia (Dreger et al. 1994; Ager et al. 1995). In addition, the PBSC product is also more suitable for ex vivo manipulation, such as CD34 + cell selection (Brugger, Henschler & Heimfeld 1994), tumour purging (Ross et al. 1995) and gene transfer (Bregni et al. 1992). Transplant centres routinely rely on CD34 + cell quantification by flow cytometry to determine the optimal timing and to confirm the adequacy of PBSC harvests (Haas et al. 1994). In contrast, assessment of haematopoietic pro- Correspondence: BCSH Secretary, British Society for Haematology, 2 Carleton House Terrace, London SW1Y 5AF, UK. Whilst the advice and information in these guidelines is believed to be true and accurate at the time of going to press, neither the authors nor the publishers can accept any legal responsibility or liability for any errors or omissions that may be made Blackwell Science Limited genitor cells (HPC) in colony-forming assays, although correlating with CD34 levels (Siena et al. 1991), is handicapped by a lack of reproducibility and the prolonged assay time. A minimum threshold dose, in adults, of between 2 and CD34 + cells/kg body weight has been observed in multiple clinical settings to result in adequate engraftment (Krause et al. 1996), although the lack of standardization of the assay prevents a more exact definition of threshold level (Bender et al. 1992). This lack of standardization with respect to reagents and gating strategies, as well as to the use of different haematology instruments to derive absolute total nucleated cell counts has contributed to the widespread inter-laboratory variation (Barnett et al. 1998). As a result, there is an urgent need for a nationally agreed protocol to enable standardization of the assay and to enable comparison of clinical and laboratory data. This guideline provides recommendations for: (1) specimen collection and frequency of CD34 testing; (2) monoclonal antibody (mab) selection; (3) cell separation, lysing and counting techniques; (4) gating strategies; (5) isotype controls; (6) determination of absolute counts; (7) instrument quality control; (8) data analysis; (9) data reporting; (10) data storage; and (11) quality assurance. Finally, this guideline will also give a brief overview of several new technologies that still require evaluation. Specimen collection and frequency of CD34 testing All peripheral blood specimens should be collected by venepuncture into either 0.34 M di- or tri-potassium ethylenediaminetetraacetic acid (K 2 EDTA or K 3 EDTA) anticoagulant. The total WBC must be obtained within 6 h of venepuncture. If a specimen is referred to a central laboratory for analysis resulting in a delay of over 6 h, then the result of a total WBC must accompany the sample. All samples must be labelled with the patient s surname and forename(s), a unique patient identifier such as the hospital reference number and date of birth as well as 301

2 302 CD34 + Haematopoietic Stem Cell Working Party the patient location and the date and time of collection. Limited data are available on the stability of the CD34 antigen at different temperatures. Thus, to minimize the influence of temperature and storage it is recommended that the specimen (peripheral blood or apheresis specimens) is stored at a constant temperature of 4 C and all processing completed within 12 h. Packaging and transportation of peripheral blood, bone marrow or apheresis samples should be in accordance with the regulations of the postal and/or courier service used. Specimen integrity must be checked upon receipt and a repeat specimen requested if there is evidence of clot formation, gross haemolysis and/or if the specimen is received more than 12 h from the time of collection. Timing and adequacy of stem cell harvest should be decided by the CD34 + cell count. Harvesting can commence when the peripheral blood CD34 count is 10 cells/ml with the aim of obtaining a yield of CD34 + cells/kg (Dreger & Schmitz 1998). Monoclonal antibody selection The CD34 antigen has epitopes that can be classified according to their sensitivity to neuraminidase and glycoprotease enzymes obtained from Pasteurella haemolytica (Sutherland, Marsh & Davidson 1992). Epitopes recognized by class I antibodies (e.g. ICH3, MY10, 12.8, Immuno133, Immuno409) are sensitive to both enzymes. Those detected by class II antibodies (e.g. QBEND10, 43.A1) are sensitive to glycoprotease only, while those detected by class III antibodies (e.g. Tük3, HPCA-2, BIRMA-K3) are resistant to both enzymes. Class I antibodies fail to detect all the glycoforms of the CD34 antigen and may fail therefore to detect CD34 expression on some leukaemic cells (Sutherland & Keating 1992). In addition, the lower avidity and reduced reactivity of conjugated class I antibodies when compared to those of class II and III, further limits their applicability. Overall, the above observations not only underline the importance of selecting an appropriate CD34 antibody clone, but also of selecting one that retains the high specificity and avidity of its binding after conjugation to the designated fluorochrome. Therefore, only class II (phycoerythrin (PE)-conjugated) or class III (any conjugate) antibodies should be used for CD34 + stem cell enumeration. The use of the most discriminating fluorochrome with an argon laser, PE, is recommended for detection of CD34 + cells (Johnsen 1995; Sutherland et al. 1996). Advantages of PE-conjugates include better demarcation between negative and positive cells, reduced binding to Fc receptors and lower non-specific binding to dead cells (Gratama et al. 1997a; Barnett et al. 1998; Gratama et al. 1998). Fluorescein isothiocyanate (FITC) conjugation, because of its induced negative charge, interferes with the binding properties of some class II antibodies, such as QBEND10, to their epitopes. Therefore, FITC-conjugated class II antibodies must not be used. Antibodies conjugated with tandem fluorochromes (i.e. PE Cy5) or allophycocyanin have only recently become readily available and data on their performance is limited. If these latter fluorochromes are used then the performance should be established in comparison to the equivalent PE-conjugate. An FITC-conjugated mab to the CD45 antigen should be used which not only detects all isoforms but also all glycoforms, e.g. anti-hle-1 (Becton- Dickinson Immunocytometry Systems, San Jose, CA, USA) or J33 (Beckman-Coulter, Miami, FL, USA). Used in combination with a PE-conjugated anti-cd34, the anti-cd45 FITC mab will thus provide an additional parameter for the identification of CD34 + cells. This antibody combination enables the identification of true CD34 + cells and simultaneously provides a useful indicator for the effectiveness of red cell lysis. It also facilitates the discrimination of cells from contaminating events, such as platelets, platelet aggregates and other debris that can bind low levels of PE-conjugated CD34 antibodies (Sutherland et al. 1996). Nucleic acid dyes (NAD) as gating reagents have recently been introduced and enable the exclusion of unlysed red cells, platelets and debris from the analysis (e.g. Laser Dye Solution (LDS)-751 and the proprietary NAD reagent provided with the ProCOUNT TM (kit)). However, NADs have a number of limitations. For example, the NAD used in ProCOUNT TM has a significant spectral overlap within the PE channel, which can reduce the resolution between CD34 + and CD34 populations. Furthermore, the detection of LDS-751 on some Coulter flow cytometers can only be achieved after the replacement of the nm band pass filter with a 650-nm long pass filter (Gratama et al. 1998). However, LDS-751 is of low cost and has negligible spectral overlap with PE. Finally, if a NAD is used in conjunction with anti-cd45 mab then an instrument capable of four colour analysis will be required for CD34 + subset analysis. All initial data must be acquired ungated because in some circumstances, such as BM regeneration and cytoreductive therapy, some CD34 + cells may express very low levels of CD45 (Gutensohn et al. 1996). Cell separation, lysis and counting techniques It should be noted that the CD34 antigen may be affected by fixatives used with specific cell labelling methods (Macey

3 Flow cytometric enumeration of CD34 + haematopoietic stem cells 303 et al. 1997). Furthermore, a number of studies have reported improved recovery using whole blood lysis when compared to cell separation techniques (Siena et al. 1991; Owens & Loken 1995). Although the optimal method for sample preparation has yet to be established, data are available suggesting that CD34 + stem cell analysis using a whole blood, lyse-no-wash, technique, with fixative-free NH 4 Cl-based lysing reagents, is the preferred approach (Menéndez et al. 1998). However, it should be stressed that certain proprietary kits, i.e. ProCOUNT TM, use a fixative-based reagent and must therefore be used under manufacturer s recommendations only. Furthermore, to ensure optimum sample condition, samples should be analysed immediately post-lysis, or kept on melting ice until analysis (Serke et al. 1998). However, all analyses must be completed within 1 h of addition of the lysing reagent. It is important when counting rare events that sufficient numbers are collected to produce a statistically valid result. Reducing the number of events analysed per test (tube) reduces the reliability of the estimation. Given the fact that the standard error of the number of positive cells per analysis is given by the square root of the number of positive cells, the larger the acquisition the lower the coefficient of variation (CV) (Wunder et al. 1992). UK NEQAS data has highlighted a marked variation in the number of CD45 + cells counted (ranging from 10,000 to 1.2 million), with a number of laboratories counting significantly less than 50,000 (Barnett et al. 1998). To achieve an intra-assay CV of 10% a minimum of 100 CD34 + events should be collected (Sutherland et al. 1996). Gating strategies A number of assay protocols for the flow cytometric enumeration of CD34 + cells have been published. The first of these was the Milan protocol (Siena et al. 1991) which was later improved upon by the Mulhouse group to incorporate direct immunofluorescence, whole blood staining and a lyse-wash method (Bender, Unverzagt & Walker 1994) and anti-cd45, whilst the SIHON protocol (Gratama et al. 1997b) evolved further to incorporate the laser dye solution (LDS) 751, anti-cd14 and anti-cd66e in order to achieve greater gating accuracy and eliminate non-specific mab binding. However, the ISHAGE (International Society for Hematotherapy and Graft Engineering) gating strategy developed by Sutherland et al. (1996) has gained increased popularity and is the most frequently used analysis method in the UK NEQAS CD34 + stem cell quantitation scheme (Barnett et al. 1998). A study by Chang & Ma (1996) reported that gating strategies are a major contributory factor in result variability and that the gating strategy described in the ISHAGE protocol gave the most reproducible results between centres. Furthermore, the gating strategy described in the ISHAGE protocol overcomes the limitations of the Milan and Bender protocols. Both these latter protocols rely upon the use of isotype controls, and may also be unsatisfactory for samples that are not in optimal condition, or contain high numbers of lysis-resistant nucleated red cells (i.e. cord blood) (Barnett et al. 1998). For these reasons, staining of CD34 + stem cells should employ the use of CD45 (FITC) vs. CD34 (PE) together with the sequential gating strategy as described within the ISHAGE protocol (Sutherland et al. 1996). The ISHAGE sequential gating strategy (Sutherland et al. 1996) takes advantage of the fact that blast cells can be identified by their dim CD45 expression and low sidescatter (SSC) (Borowitz et al. 1993). Two tubes are required, containing anti-cd45-fitc and anti-cd34-pe. Isotype controls are not required because the gating strategy excludes cells that non-specifically bind anti-cd34. An example of a specimen analysed using the ISHAGE gating strategy is shown in Figure 1. An initial gate (R1) (Figure 1: plot 1) is set on a CD45 vs. SSC dot plot, so as to contain all CD45 + events including CD45 dim and CD45 bright. This will exclude CD45 events (i.e. red blood cells, platelets and other debris). The events in gate R1 are then displayed on a CD34 vs. SSC dot plot (Figure 1: plot 2) and a second gate (R2) produced to include the cluster of CD34 + events (Figure 1: plot 3). The third plot is obtained by plotting the events that fulfil the criteria of gates R1 and R2 (i.e. sequential gating) (Figure 1: plot 3). Cells forming a cluster with characteristic low SSC and low to intermediate CD45 fluorescence are then gated on this third plot to produce a third region (R3) (Figure 1: plot 3). Ungated data are displayed on a CD34 vs. CD45 histogram to establish the lower limit of CD45 expression by the CD34 + events (Figure 1: plot 5). Finally, the events fulfilling the criteria of all three gates (R1, R2 and R3) are then displayed on a forward light scatter (FSC) vs. SSC dot plot to confirm that the selected events fall into a generic lymph-blast region (R4) (Figure 1: plot 4). This region (R4) is precisely set to include events no smaller than lymphocytes by back scattering a small number of lymphocytes from plot 1 (Figure 1: plot 6) (high CD45 staining low side-scatter). Any events falling outside region R4 are excluded from the final calculation. CD34 + cell determination must be performed in duplicate and the mean CD34 + value used. Each laboratory should establish its own acceptability criteria for the differences between the two tubes. If the two values obtained fall outside the level of acceptability the analysis

4 304 CD34 + Haematopoietic Stem Cell Working Party Figure 1. A peripheral blood sample analysed for CD34 + stem cells using the ISHAGE logical gating strategy described by Sutherland et al See text for precise logical gating strategy sequence.

5 Flow cytometric enumeration of CD34 + haematopoietic stem cells 305 should be repeated or a fresh specimen requested. Intraassay variation should also be established with an appropriate target CV of ³ 5%. The percentage of CD34 + cells is determined by using the number of CD34 + events expressed as a percentage of the CD45 + events (both values derived from the mean of the replicate tubes see above). The absolute CD34 + cell count in the sample is then calculated by multiplying this percentage value by the absolute total WBC as provided by the haematology analyser after correction to exclude nucleated red cells. To reduce inter-laboratory variation, a single platform approach should be employed (Barnett et al. 1999). Keeney et al. (1998) have recently adapted the ISHAGE strategy with the addition of a known number of Flow- Count TM fluorospheres (Beckman-Coulter) and implemented NH 4 Cl-lyse, no-wash, no-fix sample processing. These modifications convert the basic protocol into a single platform method to determine the absolute CD34 + cell count directly from a flow cytometer and form the basis of the Stem-Kit TM from Beckman-Coulter. Thus, the absolute count can be calculated from the observed ratio between the number of flow cytometrically counted beads and CD34 + cells. This guideline supports the use of such a system. Recently, Becton-Dickinson (Verwer & Ward 1997) have developed ProCOUNT TM, a single platform approach that uses a purpose-written computer software package. The system is based on bead counting using TruCOUNT TM tubes, and thresholding on nucleated cells, using a proprietary NAD. The TruCOUNT TM tubes contain a known number of lyophilized polyfluorescent microbeads retained in the bottom of the tube by a wire mesh grid. Sample and reagents are mixed in fixed amounts in the tubes, thus reconstituting the beads and enabling the absolute count to be determined. The system uses a lyse-no-wash procedure and sufficient events are collected to obtain a precision of 10%. This guideline supports the use of the ProCOUNT TM and Stem-Kit TM approaches. However, there must be operator inspection when suboptimal samples are being analysed using semi or fully automatic computer-assisted software programs. A limitation of the ProCOUNT TM approach is that all three fluorescence channels are utilized, thus preventing CD34 + subset analysis or viability assessment. Isotype controls Isotype controls are not required for the ISHAGE strategy as sequential gating excludes cells and debris responsible for non-specific staining with anti-cd34. Thus, very few events will meet the conditions of the combined gates and be falsely classified as CD34 +. Furthermore, isotype and fluorochrome-matched antibodies may not be representative of the concentration and fluorochrome protein (F:P) ratio of the anti-cd34 mab in use (Gratama et al. 1998). At present, isotype controls form an integral part of the ProCOUNT TM kit and should therefore be used in accordance with manufacturer s recommendations. Determination of absolute counts Absolute counts should be calculated on all specimens tested and expressed as cells/ml. A recent study has shown that the use of single platform analysis results in lower inter-laboratory variation (Barnett et al. 1999). Therefore, wherever possible, absolute counts should be derived from single platform analysis. The single platform approach also removes the need for correcting for nucleated red cells. However, care should be taken when using a dual-platform approach, particularly if dealing with apheresis, bone marrow or cord blood samples, since many haematology analysers over-estimate the leucocyte count in the presence of nucleated red cells. Correction of the total WBC, using validated methods should be employed in such circumstances. Reverse pipetting should always be used when using single platform systems (see below). In addition, apheresis specimens may require dilution with phosphate-buffered saline before staining. The WBC should be in the range for which the CD34 + HPC assay is linear, /l being preferable (Gratama et al. 1998). It is important for ProCOUNT TM users to refer to the manufacturer s current instructions with respect to the working WBC concentration. Extreme care is required when diluting the specimen, since errors introduced at this stage will adversely affect the final CD34 + haematopoietic cell count. The final CD34 + count is obtained following correction by the relevant dilution factor. Reverse pipetting The accuracy and reproducibility of single platform techniques are influenced by several factors with reverse pipetting being one of the most important. With conventional pipetting there is an increased risk of inclusion of air bubbles in the dispensed volume. Thus, to improve the accuracy and reproducibility of sample dispensing, reverse pipetting should be employed. With such an approach, a slight excess of sample is aspirated into the pipette and the precise volume ejected. A small amount of the sample remains in the tip, so that deviation from the required volume as a result

6 306 CD34 + Haematopoietic Stem Cell Working Party of the formation of air bubbles is reduced. It is recommended that a calibrated, electronically adjustable automated pipette be employed for the procedure. Reverse pipetting should be carried out as follows: (1) press the plunger of the pipette to the second stop; (2) gently aspirate fluid; and (3) pipette the fluid against the lower end of the wall of the tube press the plunger to the first stop. Instrument quality control Daily calibration of the instrument must be performed to ensure optimal performance and must be carried out in all laboratories performing immunophenotyping. Commercial beads and process controls are used to achieve this. Commercial beads are used in order to: (i) Monitor the light scatter and fluorescence peak channel coefficients of variation. (ii) Monitor light and fluorescence peak channel drift. (iii) Monitor instrument sensitivity. (iv) Facilitate compensation set-up to adjust for spectral overlap. All values should be logged daily, together with instrument settings. In addition, all settings should be re-established following a change in bead batch or after an instrument service. Beads only provide guidance for the final flow cytometer set-up and the optimization of settings requires the analysis of a fresh, normal specimen. The use of a process control is also recommended to enable the monitoring of reagent performance, staining, lysis and analysis. Such controls should be run at least once a week (prior to any of the week s work being processed) and preferably on a daily basis. They must also be performed if: (i) there has been a change in reagent and/or laboratory personnel; (ii) there has been an instrument service and/or instrument calibration; and/or (iii) the validity of the technique is suspect (i.e. lysis problems). Process controls should be used to test the labelling procedure and if possible to test the lysing step. They should also be stable over time (minimum of 30 days). The manufacturer should have assigned target values to these specimens. Results obtained from these reagents must be plotted on a Levy Jennings type plot, thus providing a visual indication of drift or bias over time. A fresh, previously analysed or cryopreserved specimen must not be used as a process control. Each laboratory must establish its own intra-assay variation and define acceptability criteria for the test as mentioned earlier. Data analysis Operator inspection is recommended if a laboratory uses computer-assisted analysis that automatically defines the analysis regions and calculates the percentage and absolute numbers of CD34 + cells (e.g. ProCOUNT TM ). If there is any doubt as to the validity of the analysis or if replicate CD34 + values exceed the criteria stated earlier then a repeat test must be performed. Furthermore, regardless of gating strategy, no result should be issued until all calculations have been independently checked and validated. Data reporting All results must be validated before issue (see above). To avoid clerical error (e.g. erroneous placement of decimal point, insertion of additional zero, etc.) results should be reported as cells/ml. Wherever possible, cumulative reports should be collated to allow a day-to-day check. If results are telephoned, faxed or ed then the laboratory should have in place a standard operating procedure that will facilitate an audit trail in case of errors or misunderstandings. All results should also be reported in a hard copy format. Data storage All primary data files, worksheets and copy report forms should be kept for a minimum of 6 months and ideally stored electronically. The flow cytometry data should be stored as list-mode files. Deletion of the data or destruction of paper records should be in accordance with the locally agreed guidelines and the Department of Health s document HC(89)20 Preservation, Retention and Destruction: Responsibilities of Health Authorities under the Public Records Act. Quality assurance It is expected that laboratories performing CD34 + haematopoietic cell counting are fully conversant with all procedures employed. Such laboratories should meet Clinical Pathology Accredited (CPA) (UK) Ltd. standards and be accredited as appropriate. In addition, both internal and external quality assurance should be undertaken. Satisfactory performance should be demonstrable in external proficiency testing schemes such as UK NEQAS Leucocyte Immunophenotyping Schemes. All quality assurance activities should be documented.

7 Flow cytometric enumeration of CD34 + haematopoietic stem cells 307 New technologies for haematopoietic cell analysis A number of new technologies have the potential to simplify haematopoietic analysis including the STELLer assay (Dietz et al. 1996) and haematopoietic cell measurement by haematology analyser, i.e. Sysmex SE-9000 (Takekawa et al. 1997). The STELLer system employs the IMAGN 2000 micro volume fluorimeter and measures the number of cells which are CD34 + using Cy5-conjugated anti-cd34. In contrast, the Sysmex SE-9000 system identifies haematopoietic cells based upon their phospholipid content. However, comparative data is awaited before firm recommendations can be issued. This guideline provides information for the enumeration of CD34 + stem cells using current techniques. However, several issues remain unresolved and will require further investigation before recommendations can be made, including: CD34 + subset analysis and their clinical relevance; the use of fixative-based lysing reagents; the determination of viability and finally the evaluation of the newer fluorochrome conjugates for CD34 mabs. Key points (1) The CD34 + cell count and not the change in total WBC should determine timing and adequacy of stem cell harvest. (2) Total nucleated cell count must be completed within 6 h and the CD34 + haematopoietic enumeration performed within 12 h of venepuncture. (3) All samples must be labelled with the patient s surname and forename(s), a unique patient identifier (such as the hospital reference number), date of birth, patient location, date and time of collection. (4) All samples must be fully processed within 12 h. (5) To minimize the influence of temperature, specimens should be stored at a constant temperature of 4 C. (6) Harvesting should commence when the peripheral blood CD34 + count is 10 cells/ml with the aim of having a final yield of CD34 + cells/kg. (7) The use of the brightest excitable fluorochromes with an argon laser, i.e. phycoerythrin (PE), is recommended for detection of CD34 + cells. (8) To ensure there is no loss of cells during centrifugation, a lyse-no-wash technique is recommended using a fixative-free NH 4 Cl-based erythrocyte lysis reagent. (9) Class II (PE-conjugated) or class III conjugated anti-cd34 antibodies must be used. (10) Anti-pan-FITC CD45 should be used as an additional parameter for identifying the CD34 + cell population. Pan-CD45 reagents should be used that not only detect all isoforms but also all glycoforms, e.g. anti-hle-1 (Becton-Dickinson Immunocytometry Systems) or J33 (Beckman-Coulter). (11) A minimum of 100 CD34 + events should be collected. (12) The number of CD34 + cells transplanted per kg body weight should be calculated. (13) The sequential gating strategy as described in the ISHAGE protocol is recommended. (14) CD34 + determination must be performed in duplicate and the mean CD34 + value used. (15) All initial data should be collected ungated. (16) Wherever possible absolute counts should be derived using single platform technology employing reverse pipetting. (17) Control tubes for non-specific antibody binding are not required unless using ProCOUNT TM. (18) Process controls to monitor reagent, staining and, ideally, lysis procedures should be used at least once a week, but preferably daily. (19) Flow cytometer performance should be monitored daily using unstained and fluorescent latex beads. (20) Laboratories providing a routine CD34 + haematopoietic cell enumeration service should meet Clinical Pathology Accredited (CPA) (UK) Ltd. standards and be accredited where appropriate. (21) Participation (together with satisfactory performance) in accredited external proficiency testing schemes should be demonstrable. (22) Deletion of the data or destruction of paper records should be in accordance with the locally agreed guidelines and the Department of Health s document HC(89)20 Preservation, Retention and Destruction: Responsibilities of Health Authorities under the Public Records Act. References Ager S., Scott M.A., Mahendra P. et al. (1995) Peripheral blood stem cell transplants after high-dose therapy in patients with malignant lymphoma: a retrospective comparison with autologous bone marrow transplantation. Bone Marrow Transplantation 16, Andrews R.G., Bryant E.M., Bartelmez S.H. et al. (1992). CD34 + marrow cells, devoid of T and B lymphocytes, reconstitute stable lymphopoiesis and myelopoiesis in lethally irradiated baboons. Blood 80, Barnett D., Granger V., Storie I. et al. (1998) Quality assessment of CD34 + stem cell enumeration. Experience of the united kingdom national external quality assessment scheme (UK NEQAS) using

8 308 CD34 + Haematopoietic Stem Cell Working Party a unique stable whole blood preparation. British Journal of Haematology 102, Barnett D., Granger V., Whitby L., Storie I. & Reilly J.T. (1999) Absolute CD4 + T-lymphocyte and CD34 + stem cell counts by single platform flow cytometry. The way forward. British Journal of Haematology 106, Bender J., To L.B., Williams S. & Schwartzberg L. (1992) Defining a therapeutic dose of peripheral blood stem cells. Journal of Hematotherapy 1, Bender J.G., Unverzagt K. & Walker D. (1994) Guidelines for determination of CD34 + cells by flow cytometry: application to the harvesting and transplantation of peripheral blood stem cells. In: Wunder E., Sovalat, H., Henon, P., Serke, S, eds. Hematopoietic Stem Cells The Mulhouse Manual. Alpha Med Press, Dayton, OH, Berenson R.J., Andrews R.G., Bensinger W.J. et al. (1988). Antigen CD34-positive marrow cells engraft lethally irradiated baboons. Journal of Clinical Investigation 81, Borowitz M.J., Guenther K.L., Schultz K.E. & Stelzer G.T. (1993) Immunophenotyping of acute leukaemia by flow cytometry: use of CD45 and right angle light scatter to gate on leukaemic blasts in three colour analysis. American Journal of Clinical Pathology 100, Bregni M., Magni M., Siena S., Di Nicola M., Bonadonna G. & Gianni A.M. (1992) Human peripheral blood haematopoietic progenitors are optimal targets of retroviral-mediated gene transfer. Blood 80, Brugger W., Henschler R. & Heimfeld S. (1994) Positively selected autologous blood CD34 + cells and unseparated peripheral blood progenitor cells mediate identical haematopoietic engraftment after high dose VP16, ifosamide, carboplatin and epirubicin. Blood 84, Chang A. & Ma D. (1996) The influence of flow cytometric gating strategy on the standardization of CD34 + cell quantitation: an Australian multi-centre study. Journal of Hematotherapy 5, Dietz L.J., Dubrow R.S., Manian B.S. & Sizto N.L. (1996) Volumetric capillary cytometry: a new method for absolute cell enumeration. Cytometry 23, Dreger P., Haferlach T., Eckstein V. et al. (1994) G-CSF mobilised peripheral blood progenitor cells for allogeneic transplantation: safety, kinetics of mobilisation, and composition of the graft. British Journal of Haematology 87, Dreger P. & Schmitz N. (1998) Sources of stem cells: autografts. In: Apperley J. Gluckman E. & Gratwohlthe A. eds EBMT Handbook: Blood and Marrow Transplantation. Gratama J.W., Van der Linden R., Van der Holt B., Bolhuis R.L.H. & Van de Winkel J.G.J. (1997a) Analysis of factors contributing to the formation of mononuclear cell aggregates (escapees) in flow cytometric immunophenotyping. Cytometry 29, Gratama J.W., Kraan J., Levering W., van Bockstaele D.R., Rijkers G.T. & van der Schoot C.E. (1997b) Analysis of variation in results of CD34 + haematopoietic progenitor cell enumeration in a multi-centre study. Cytometry 30, Gratama J.W., Orfao A., Barnett D. et al. (1998) Flow cytometric enumeration of CD34 + hematopoietic stem and progenitor cells. Cytometry (Communications in Clinical Cytometry) 34, Gratwohl A., Hermans J. & Baldomero H. (1996) Haemopoietic precursor cell transplant activity in Europe Bone Marrow Transplantation 17, Gutensohn K., Serke S., Cassens U. et al. (1996) Durchflu)zytometrische Analyse CD34 exprimierender häma- topoetischer Zellen in Blut und Zytafereseprodukten. Infusionsther Transfusionsmed 23 (Suppl. 2), Haas R., Mohle R., Fruehauf S. et al. (1994) Patient characteristic associated with successful mobilising and autografting of peripheral blood progenitor cells in malignant melanoma. Blood 83, Johnsen H.E. (1995) Report from a Nordic Workshop on CD34 + cell analysis: technical recommendations for progenitor cell enumeration in leukapheresis from multiple myeloma patients. Journal of Hematotherapy 4, Keeney M., Chin-Yee I., Weir K., Popma J., Nayar R. & Sutherland D.R. (1998) Single platform flow cytometric absolute CD34+ cell counts based on the ISHAGE guidelines. Cytometry (Communications in Clinical Cytometry) 34, Krause D., Fackler M.J., Civin C.I. & May W.S. (1996) CD34: structure, biology, and clinical utility. Blood 87, Macey M.G., McCarthy D.A., van Agthoven A.J. & Newland A.C. (1997) How should CD34 + cells be analysed? A study of three classes of antibody and five leucocyte preparation procedures. Journal of Immunological Methods 204, Menéndez P., Redondo O., Rodriguez A. et al. (1998) Comparison between a lyse-and-then-wash method and a lyse-non-wash technique for the enumeration of CD34 + hematopoietic progenitor cells. Cytometry 34, Owens M.A. & Loken M.P. (1995) Peripheral blood stem cell quantitation. In: Flow Cytometry Principles for Clinical Laboratory Practice. Wiley-Liss Inc, New York, 111. Ross A.A., Miller G.W., Moss T.J. et al. (1995) Immunocytochemical detection of tumour cells in bone marrow and peripheral blood stem cell collections from patients with ovarian cancer. Bone Marrow Transplant 15, Russell N.H., Gratwohl A. & Schmitz N. (1996) The place of blood stem cells in allogeneic transplantation. British Journal of Haematology 93, Serke S., van Lessen A., Pardo I. & Huhn D. (1998) Selective susceptibility of CD34-expressing cells to acquire flow cytometric features of apoptosis/necrosis on exposure to an ammonium chloride-based red blood cell lysing reagent. Journal of Hematotherapy 7, Siena S., Bregni M., Brando B. et al. (1991) Flow cytometry for the clinical estimation of circulating hematopoietic progenitors for autologous transplantation in cancer patients. Blood 77, Sutherland D.R., Anderson L., Keeney M., Nayar R. & Chin-Yee I. (1996) The ISHAGE guidelines for CD34 + cell determination by flow cytometry. Journal of Hematotherapy 5, Sutherland D.R. & Keating A. (1992) The CD34 antigen: structure, biology and potential clinical applications. Journal of Hematotherapy 1, Sutherland D.R., Marsh J.C.W. & Davidson J. (1992) Differential sensitivity of CD34 epitopes to cleavage by Pasteurella haemolytica glycoprotease: implications for purification of CD34- positive progenitor cells. Experimental Hematology 20, Takekawa K., Yamane T., Suzuki K., Hino M. & Tatsumi N. (1997) Identification of hematopoietic stem cells by the SE-9000 automated hematology analyzer in peripheral blood stem cell harvest samples. Acta Haematologica 98, Verwer B.J.H. & Ward D.M. (1997) An automated classification algorithm for ProCount flow cytometric acquisition and analysis. Journal of Hematotherapy 6, 169. Wunder E., Sovalat H., Fritsch G., Silvestri F., Henon P.R. & Serke S. (1992) Report on the European Workshop on Peripheral Blood Stem Cell Determination and Standardization Mulhouse. France. Journal of Hematotherapy 1,

Standardization, Calibration and Quality Control

Standardization, Calibration and Quality Control Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic

More information

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep

More information

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit 2 Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit Presented by Ellen Meinelt, MS MLS(ASCP) CM, Technical Applications Specialist and Calin Yuan, Product Course Developer,

More information

Enumeration of CD34-positive hematopoietic progenitor cells by flow cytometry: comparison of a volumetric assay and the ISHAGE gating strategy

Enumeration of CD34-positive hematopoietic progenitor cells by flow cytometry: comparison of a volumetric assay and the ISHAGE gating strategy Bone Marrow Transplantation, (1998) 22, 699 76 1998 Stockton Press All rights reserved 268 3369/98 $12. http://www.stockton-press.co.uk/bmt Enumeration of CD34-positive hematopoietic progenitor cells by

More information

Absolute counting using flow cytometry. Bright Count Microspheres

Absolute counting using flow cytometry. Bright Count Microspheres Absolute counting using flow cytometry Single-platform system for absolute counts I PRODUCT MANUAL [REF] IQP-391 [REF] IQP-392 s 50 tests s 100 tests [IVD] In Vitro Diagnostic medical device IQ Products

More information

APPLICATION INFORMATION

APPLICATION INFORMATION DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL

More information

The success of bone marrow or peripheral blood stem

The success of bone marrow or peripheral blood stem CD34 + Cell Enumeration by Flow Cytometry A Comparison of Systems and Methodologies Alain Ngoma, MD; Shunnichi Saito, MT; Hitoshi Ohto, MD, PhD; Kazuhiko Ikeda, MD, PhD; Hiroyasu Yasuda, PhD; Kinuyo Kawabata,

More information

A. Gajkowska 1, T. Oldak 1,2, M. Jastrzewska 1, E.K. Machaj 1,2, J. Walewski 3, E. Kraszewska 4 and Z. Pojda 1,2. Introduction

A. Gajkowska 1, T. Oldak 1,2, M. Jastrzewska 1, E.K. Machaj 1,2, J. Walewski 3, E. Kraszewska 4 and Z. Pojda 1,2. Introduction FOLIA HISTOCHEMICA ET CYTOBIOLOGICA Vol. 44, No. 1, 2006 pp. 53-60 Flow cytometric enumeration of CD34 + hematopoietic stem and progenitor cells in leukapheresis product and bone marrow for clinical transplantation:

More information

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.

More information

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes

More information

WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS

WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS FOR IN-VITRO DIAGNOSTIC USE INVITROGEN CAL-LYSE TM Lysing Solution WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS CAL-LYSE TM CATALOG No. GAS-010 250 tests 25 ml CATALOG No. GAS-010S-100

More information

CD34 numeration through the French external quality control: Impact in the implementation of new technical chapters at the European Pharmacopoeia

CD34 numeration through the French external quality control: Impact in the implementation of new technical chapters at the European Pharmacopoeia CD34 numeration through the French external quality control: Impact in the implementation of new technical chapters at the European Pharmacopoeia Panterne Béatrice Cells and Tissues Coordinator Batch release

More information

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating

More information

2. SUMMARY AND EXPLANATION

2. SUMMARY AND EXPLANATION M3630IV_C 2003-09-01 1 / 29 English Stem-Kit Reagents 1-29 1. INTENDED USE 2 2. SUMMARY AND EXPLANATION 2 3. PRINCIPLE OF TEST 3 4. CLINICAL RELEVANCE 3 5. REAGENTS AND CONTENTS 3 6. STATEMENT OF WARNING

More information

Immunophenotyping peripheral blood cells

Immunophenotyping peripheral blood cells IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune

More information

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number

More information

Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer

Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer Purpose This procedure gives instruction on minimal residual disease (MRD) detection in patients with

More information

BD Trucount Tubes IVD

BD Trucount Tubes IVD 02/2015 23-3483-07 IVD BD Trucount Tubes For determining absolute counts of leucocytes in blood 25 Tubes Catalog No. 340334 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company.

More information

Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood

Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood Application Note Primary Cell Analysis Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood Cell Viability Introduction: Bone Marrow, Cord Blood, Stem

More information

Flow cytometric analysis of stem cells derived from umbilical cord blood CD34+/CD45+ enumeration and viability

Flow cytometric analysis of stem cells derived from umbilical cord blood CD34+/CD45+ enumeration and viability Flow cytometric analysis of stem cells derived from umbilical cord blood CD34+/CD45+ enumeration and viability Doretha Thompson Dissertation in fulfillment of the requirements for the degree Magister Scientiae

More information

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the

More information

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation... Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and

More information

Direct measurement of CD34+ cord blood stem cell absolute counts, viability and WBC using flow cytometry

Direct measurement of CD34+ cord blood stem cell absolute counts, viability and WBC using flow cytometry Original Article Direct measurement of CD34+ cord blood stem cell absolute counts, viability and WBC using flow cytometry Vasiliki e. Kalodimou 1, Lefteris Bikos 2, Calliope Kourenti 3, Aikaterini Fifa

More information

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch

More information

Reference: NHS England B04/P/a

Reference: NHS England B04/P/a Clinical Commissioning Policy: Haematopoietic Stem Cell Transplantation (HSCT) (All Ages): Revised Reference: NHS England B04/P/a 1 NHS England Clinical Commissioning Policy: Haematopoietic Stem Cell Transplantation

More information

Notes on titering antibodies

Notes on titering antibodies Home Events This Week Bulletin Boards Lab Personnel Lab Policies Protocols MolBio Tetramers Proteins Flow Cytometry Immunoassays Instruments Computers Lab Publications Scientiific Literature Science on

More information

Introduction to flow cytometry

Introduction to flow cytometry Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression

More information

Fresh PBSC harvests, but not BM, show temperature-related loss of CD34 viability during storage and transport

Fresh PBSC harvests, but not BM, show temperature-related loss of CD34 viability during storage and transport Cytotherapy (2006) Vol. 8, No. 2, 158/165 Fresh PBSC harvests, but not BM, show temperature-related loss of CD34 viability during storage and transport V Antonenas 1, F Garvin 1, M Webb 1, M Sartor 2,

More information

BD FACSComp Software Tutorial

BD FACSComp Software Tutorial BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS

More information

Below is a list of things you should be aware of before you schedule your sort.

Below is a list of things you should be aware of before you schedule your sort. Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will

More information

TITLE: TRU-COUNT ENUMERATION OF T-CELLS IN WHOLE

TITLE: TRU-COUNT ENUMERATION OF T-CELLS IN WHOLE TITLE: TRU-COUNT ENUMERATION OF T-CELLS IN WHOLE PRINCIPLES OF THE PROCEDURE: When whole blood is added to the reagent, the fluorochrome labeled antibodies in the reagent bind specifically to leukocyte

More information

Uses of Flow Cytometry

Uses of Flow Cytometry Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer

More information

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University

More information

Setting up and calibration of a flow cytometer for multicolor immunophenotyping

Setting up and calibration of a flow cytometer for multicolor immunophenotyping Journal of Biological Regulators and Homeostatic Agents Setting up and calibration of a flow cytometer for multicolor immunophenotyping J. KRAAN 1, J.W. GRATAMA 1, M. KEENEY 2, J.L. D HAUTCOURT 3 1 Department

More information

ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This

More information

HAEMATOLOGY LABORATORY

HAEMATOLOGY LABORATORY HAEMATOLOGY LABORATORY Head of Unit : Dr Raudhawati Osman Phone : +603-26155281 Email : raudhaosman@yahoo.com 1. INTRODUCTION The Haematology Unit, Department of Pathology, HKL provides tertiary diagnostic

More information

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions Chapter 6 Antigen-Antibody Interactions: Principles and Applications Antigen-Antibody Properties You must remember antibody affinity (single) VS avidity (multiple) High affinity: bound tightly and longer!

More information

COMPENSATION MIT Flow Cytometry Core Facility

COMPENSATION MIT Flow Cytometry Core Facility COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,

More information

Low Cost CD4 Enumeration Using Generic Monoclonal Antibody Reagents and a Two-Color User-Defined MultiSET TM Protocol

Low Cost CD4 Enumeration Using Generic Monoclonal Antibody Reagents and a Two-Color User-Defined MultiSET TM Protocol Cytometry Part B (Clinical Cytometry) 70B:355 360 (2006) Low Cost CD4 Enumeration Using Generic Monoclonal Antibody Reagents and a Two-Color User-Defined MultiSET TM Protocol Kovit Pattanapanyasat, 1 *

More information

Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com

Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com Fluorescence Compensation Jennifer Wilshire, PhD jennifer@flowjo.com Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due

More information

Use and evaluation of leucocyte monoclonal antibodies in the diagnostic laboratory: a review

Use and evaluation of leucocyte monoclonal antibodies in the diagnostic laboratory: a review Clin. Lab. Haem. 1996, 18, 1 5 J.T. REILLY Summary Keywords BCSH REVIEW Use and evaluation of leucocyte monoclonal antibodies in the diagnostic laboratory: a review (On behalf of the General Haematology

More information

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating

More information

Standard Operating Procedure

Standard Operating Procedure 1.0 Purpose: 1.1 The characterisation of of main leukocyte subsets in peripheral blood cells from mice by flow cytometry. Reliable values of frequencies of leukocyte clusters are very much dependent on

More information

for Leucocyte Immunophenotyping Leukaemia Diagnosis Interpretation All Participants Date Issued: 08-September-2014 Closing Date: 26-September-2014

for Leucocyte Immunophenotyping Leukaemia Diagnosis Interpretation All Participants Date Issued: 08-September-2014 Closing Date: 26-September-2014 for Leucocyte Immunophenotyping Leukaemia Interpretation All Participants Participant: 4xxxx Trial No: 141502 Date Issued: 08-September-2014 Closing Date: 26-September-2014 Trial Comments This was an electronic

More information

CPT Codes for Bone Marrow Transplant January 2015 James L. Gajewski, MD

CPT Codes for Bone Marrow Transplant January 2015 James L. Gajewski, MD The blood and marrow transplant field has 15 dedicated CPT codes. These CPT codes can be categorized into three groups: 1. Collection Codes 2. Cell Processing Codes 3. Cell Infusion Codes Collection Codes

More information

Haematopoietic Chimerism Analysis after Allogeneic Stem Cell Transplantation

Haematopoietic Chimerism Analysis after Allogeneic Stem Cell Transplantation Haematopoietic Chimerism Analysis after Allogeneic Stem Cell Transplantation Dr Ros Ganderton, Ms Kate Parratt, Dr Debbie Richardson, Dr Kim Orchard and Dr Liz Hodges Departments of Molecular Pathology

More information

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor

More information

Principles of Flowcytometry

Principles of Flowcytometry Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal

More information

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection

More information

Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1

Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1 78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: daisy_kuo@bd.com BDBiosciences Application Examples

More information

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.

More information

CyAn : 11 Parameter Desktop Flow Cytometer

CyAn : 11 Parameter Desktop Flow Cytometer CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on

More information

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required. PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

Human Free Testosterone(F-TESTO) ELISA Kit

Human Free Testosterone(F-TESTO) ELISA Kit Human Free Testosterone(F-TESTO) ELISA Kit Catalog Number. MBS700040 For the quantitative determination of human free testosterone(f-testo) concentrations in serum, plasma. This package insert must be

More information

Compensation in Flow Cytometry

Compensation in Flow Cytometry Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome

More information

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that

More information

Improved Ex Vivo Expansion of Functional CD34 + Cells Using Stemline II Hematopoietic Stem Cell Expansion Medium

Improved Ex Vivo Expansion of Functional CD34 + Cells Using Stemline II Hematopoietic Stem Cell Expansion Medium Improved Ex Vivo Expansion of Functional CD34 + Cells Using Stemline II Hematopoietic Stem Cell Expansion Medium Daniel W. Allison, 1 Stacy L. Leugers, 1 Barry J. Pronold, 2 Gary Van Zant, 2 Jenny A. Harrington,

More information

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results

More information

14.0 Stem Cell Laboratory Services

14.0 Stem Cell Laboratory Services Laboratory Services Contact Information: To inquire about assisting with surgical harvesting of bone marrow, cellular therapy (CT) product processing, cryopreservation, storage, or any other lab services,

More information

ArC Amine Reactive Compensation Bead Kit

ArC Amine Reactive Compensation Bead Kit ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component

More information

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry

More information

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY 页 码,1/6 CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY From Givan, A.L. (2000), chapter in In Living Color: Protocols in Flow Cytometry and Cell Sorting (R. Diamond and S. DeMaggio, eds). Springer, Berlin,

More information

Islet Viability Assessment by Single Cell Flow Cytometry

Islet Viability Assessment by Single Cell Flow Cytometry Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing

More information

Titering Antibodies INTRODUCTION Materials Procedure for titering antibodies to extracellular antigens A. Directly conjugated antibodies

Titering Antibodies INTRODUCTION Materials Procedure for titering antibodies to extracellular antigens A. Directly conjugated antibodies C.C. Stewart, S.J. Stewart, Titering Antibodies. In: Current Protocols in Cytometry, (J.P. Robinson, Z. Darzynkiewicz, P. Dean. L. Dressler, P.Rabinovitch, C. Stewart, H. Tanke, L. Wheeless, eds.) J.Wiley

More information

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows

More information

Three Rules for Compensation Controls

Three Rules for Compensation Controls From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

flow cytometry reinvented

flow cytometry reinvented FLOW CYTOMETRY flow cytometry reinvented The Attune Acoustic Focusing Cytometer Precision and at all speeds for rare events and precious samples The Attune Acoustic Focusing Cytometer Precise high collection

More information

Recommendations to Transplant Centres Performing Cord Blood Transplants. Why Choosing the Right Thaw Method Could Save a Patient s Life

Recommendations to Transplant Centres Performing Cord Blood Transplants. Why Choosing the Right Thaw Method Could Save a Patient s Life Recommendations to Transplant Centres Performing Cord Blood Transplants Vicki Antonenas Why Choosing the Right Thaw Method Could Save a Patient s Life Lynn O Donnell (USA) 1 The process of receiving, testing

More information

DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES

DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES CLAUDIO ORTOLANI UNIVERSITY OF URBINO - ITALY SUN TZU (544 b.c. 496 b.c) SUN TZU (544 b.c. 496 b.c.) THE ART OF CYTOMETRY

More information

Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry

Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry by Debby R. Walser-Kuntz Biology Department Carleton College, Northfield, MN Taylor goes in to see her doctor, Dr. Chavez,

More information

低 剂 量 辐 射 对 脐 血 T 淋 巴 细 胞 膜 分 子 表 达 的 影 响 EFFECT OF LOW DOSE IRRADIATION ON EXPRESSION OF MEMBRANE MOLECULES OF T LYMPHOCYTES IN CORD BLOOD

低 剂 量 辐 射 对 脐 血 T 淋 巴 细 胞 膜 分 子 表 达 的 影 响 EFFECT OF LOW DOSE IRRADIATION ON EXPRESSION OF MEMBRANE MOLECULES OF T LYMPHOCYTES IN CORD BLOOD CNIC-01513 PUTH-0001 低 剂 量 辐 射 对 脐 血 T 淋 巴 细 胞 膜 分 子 表 达 的 影 响 EFFECT OF LOW DOSE IRRADIATION ON EXPRESSION OF MEMBRANE MOLECULES OF T LYMPHOCYTES IN CORD BLOOD 中 国 核 情 报 中 心 China Nuclear Information

More information

Whole Blood Flow Cytometry

Whole Blood Flow Cytometry Whole Blood Flow Cytometry y Nailin Li Department t of Medicine, i Clinical i l Pharmacology Unit Karolinska Institute/University Hospital, 171 76 Stockholm Department of Pathology & Pathophysiology Zhejiang

More information

Changes to UK NEQAS Leucocyte Immunophenotyping Chimerism Performance Monitoring Systems From April 2014. Uncontrolled Copy

Changes to UK NEQAS Leucocyte Immunophenotyping Chimerism Performance Monitoring Systems From April 2014. Uncontrolled Copy Changes to UK NEQAS Leucocyte Immunophenotyping Chimerism Performance Monitoring Systems From April 2014 Contents 1. The need for change 2. Current systems 3. Proposed z-score system 4. Comparison of z-score

More information

MANUAL WBC LAB OBJECTIVE

MANUAL WBC LAB OBJECTIVE MANUAL WBC LAB OBJECTIVE The student will be able to perform, within K20% accuracy as compared to the automated result, five manual white blood cell counts using the Unopette system. PRINCIPLE Free-flowing

More information

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility 10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

a Beckman Coulter Life Sciences: White Paper

a Beckman Coulter Life Sciences: White Paper a Beckman Coulter Life Sciences: White Paper Violet-Excited nim-da Allows Efficient and Reproducible Cell Cycle Analysis on the Gallios Flow Cytometer Authors: Valdez, Ben 1. Carr, Karen 2. Norman, John

More information

Blood-Forming Stem Cell Transplants

Blood-Forming Stem Cell Transplants Blood-Forming Stem Cell Transplants What are bone marrow and hematopoietic stem cells? Bone marrow is the soft, sponge-like material found inside bones. It contains immature cells known as hematopoietic

More information

Boundary-breaking acoustic focusing cytometry

Boundary-breaking acoustic focusing cytometry Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory

More information

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE Flow Data Analysis Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA Flow Data Analysis From Cells to Data - data measurement and standard Data display and gating - Data Display options -

More information

CAP Accreditation Checklists 2015 Edition

CAP Accreditation Checklists 2015 Edition CAP Accreditation Checklists 2015 Edition The College of American Pathologists (CAP) accreditation checklists contain the CAP accreditation program requirements, developed on more than 50 years of insight

More information

Blood & Marrow Transplant Glossary. Pediatric Blood and Marrow Transplant Program Patient Guide

Blood & Marrow Transplant Glossary. Pediatric Blood and Marrow Transplant Program Patient Guide Blood & Marrow Transplant Glossary Pediatric Blood and Marrow Transplant Program Patient Guide Glossary Absolute Neutrophil Count (ANC) -- Also called "absolute granulocyte count" amount of white blood

More information

Preparation of cord blood for infusion: bedside thaw, dilute and wash, or somewhere in between

Preparation of cord blood for infusion: bedside thaw, dilute and wash, or somewhere in between Preparation of cord blood for infusion: bedside thaw, dilute and wash, or somewhere in between Donna Wall, MD Director, Manitoba Blood and Marrow Transplant Program ISCT 2012 Disclosures: none The problem:

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Item No. 10009853 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions

More information

Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)

Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806) Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)

More information

use. 3,5 Apoptosis Viability Titration Assays

use. 3,5 Apoptosis Viability Titration Assays Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,

More information

Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin. Antibodies and reagents for 25 tests

Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin. Antibodies and reagents for 25 tests page 1 of 8 Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin Antibodies and reagents for 25 tests Key to symbols used Glycotope Biotechnology GmbH Czernyring 22 69115

More information

BD FC Beads 8-Color Kit for BD OneFlow Assays

BD FC Beads 8-Color Kit for BD OneFlow Assays BD FC Beads 8-Color Kit for BD OneFlow Assays 7/2014 23-15785-00 IVD 5 tests per kit Catalog No. 658621 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2014 BD Becton,

More information

Bone Marrow Transplantation and Peripheral Blood Stem Cell Transplantation: Questions and Answers. Key Points

Bone Marrow Transplantation and Peripheral Blood Stem Cell Transplantation: Questions and Answers. Key Points CANCER FACTS N a t i o n a l C a n c e r I n s t i t u t e N a t i o n a l I n s t i t u t e s o f H e a l t h D e p a r t m e n t o f H e a l t h a n d H u m a n S e r v i c e s Bone Marrow Transplantation

More information

Cell Cycle Tutorial. Contents

Cell Cycle Tutorial. Contents Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6

More information

Stem Cell Transplantation

Stem Cell Transplantation Harmony Behavioral Health, Inc. Harmony Behavioral Health of Florida, Inc. Harmony Health Plan of Illinois, Inc. HealthEase of Florida, Inc. Ohana Health Plan, a plan offered by WellCare Health Insurance

More information

THE His Tag Antibody, mab, Mouse

THE His Tag Antibody, mab, Mouse THE His Tag Antibody, mab, Mouse Cat. No. A00186 Technical Manual No. TM0243 Update date 01052011 I Description.... 1 II Key Features. 2 III Storage 2 IV Applications.... 2 V Examples - ELISA..... 2 VI

More information

ENUMERATION OF LYMPHOCYTE SUBSETS USING FLOW CYTOMETRY: EFFECT OF STORAGE BEFORE AND AFTER STAINING IN A DEVELOPING COUNTRY SETTING

ENUMERATION OF LYMPHOCYTE SUBSETS USING FLOW CYTOMETRY: EFFECT OF STORAGE BEFORE AND AFTER STAINING IN A DEVELOPING COUNTRY SETTING ENUMERATION OF LYMPHOCYTE SUBSETS USING FLOW CYTOMETRY: EFFECT OF STORAGE BEFORE AND AFTER STAINING IN A DEVELOPING COUNTRY SETTING Sanju Jalla*, Sunil Sazawal***, Saikat Deb***, Robert E. Black**, Satya

More information

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection CyTOF2 Mass cytometry system Unveil new cell types and function with high-parameter protein detection DISCOVER MORE. IMAGINE MORE. MASS CYTOMETRY. THE FUTURE OF CYTOMETRY TODAY. Mass cytometry resolves

More information

Cord Blood for Cellular Therapy: A Snapshot of this Evolving Market Landscape

Cord Blood for Cellular Therapy: A Snapshot of this Evolving Market Landscape GENReports: Market & Tech Analysis Cord Blood for Cellular Therapy: A Snapshot of this Evolving Market Landscape > Enal Razvi, Ph.D. Biotechnology Analyst, Managing Director SELECTBIO US enal@selectbio.us

More information

Immunophenotyping. Alberto Orfao, MD, PhD

Immunophenotyping. Alberto Orfao, MD, PhD Immunophenotyping Alberto Orfao, MD, PhD Introduction Immunophenotyping of biological samples refers to the use of immunological tools (e.g. monoclonal and/or polyclonal antibodies) for the specific detection

More information