Installation, Operation and Performance Qualification of a Thermoguard Installation

Size: px
Start display at page:

Download "Installation, Operation and Performance Qualification of a Thermoguard Installation"

Transcription

1 Installation, Operation and Performance Qualification of a Thermoguard Installation Thermoguard

2 Qualification in Accordance with GMP To operate Thermoguard hardware and software within a regulated environment in accordance with the guidelines of Good Manufacturing Practice, GMP, harmonized by the Pharmaceutical Inspection Convention, some formal qualifying steps are required. Qualifications according to GMP DQ: Design Qualification IQ: Installation Qualification OQ: Operational Qualification PQ: Performance Qualification The system owner should document his user requirement specifications, URS, for the monitoring system in a Design Qualification document. Thermoguard provides a template DQ document with a full set of requirements. A documented check of the installation (Installation Qualification) and an assessment of the operation and performance (Operational and Performance Qualification) of the system should follow. Due to the technical nature of the system, the operational and performance check are combined. The procedures described below should be run with results within given limits before using the installation. Execution and success of these tests should be documented using this form and approved by the relevant equipment owner and qualified person according to GMP. Installation Qualification The installation qualification is proved by the documentation of the hardware and software installation. The documentation qualifications includes the device list of all installed temperature and humidity sensors as well as the configuration list of the program. The device list can be achieved by exporting the sensor inventory by using the context menu under "Configuration/Sensor Configuration/export sensor inventory". The configuration list is obtained by printing the file "TGuard.ini" from the installation directory. Thermoguard Installation, Operation and Performance Qualification Page 2 of 4 Subject to changes.

3 Operational Qualification The Operational qualification consists of two tests to check the functionality of the software as well as the sensor controllers and sensors: Test 1: After proper installing of the Thermoguard software (including the license file and complete configuration of all the sensor controllers and sensors) the function "Poll all" is used to check the communication between the software and every sensor. The result of the "Poll all" function is displayed in the "Activity log" window and stored in the log file. The correct function is proved by printing the log file. Acceptance criteria: The log window and the log file may not contain an error. The log file should be printed and signed by the equipment owner and qualified person according to GMP. Test 2: Test the alarm functions mail and SMS by using the button "send test mail(s)". A printout of the delivered mail is used as a proof of the successful alarm function. Acceptance criteria: Both test mail and SMS should be delivered successfully. Thermoguard Installation, Operation and Performance Qualification Page 3 of 4 Subject to changes.

4 Performance Qualification Thermoguard network monitoring can be classified according to the United States Pharmacopoeia (USP), Chapter 1058, as Class B equipment with a simplified performance qualification by performing a calibration. Calibration of Thermoguard network sensors can be done in several ways: Option 1: Traceable calibration certificates for temperature and humidity sensors show the correct function of these sensors. Thermoguard sensorcontrollers have a precise internal reference resistor of known accuracy and require no external calibration. Option 2: Calibration of temperature sensors including the sensorcontroller at 0.01 degree Celsius with an ice/water mixture (triple point according to the international temperature scale ITS 90). Calibration of digital humidity sensors using humidity reference salt cells according to NIST. Option 3: Running a comparative measuring during a defined period of time, using a traceable calibrated reference thermometer (e.g., Thermoguard temperature calibration set TGCSTx) respective a traceable calibrated moisture meter (such as a Thermoguard humidity calibration set TGCSrF). Note: Since pressure sensors (e.g. Thermoguard HM1) are difficult to calibrate in the field, pressure readings might be marked as uncalibrated and should be used for informational purposes only. Acceptance criteria: Measurement results should be within the specifications of the sensors and controllers (given by data sheets or norms), independent from measuring principle. Note The installation, operation and performance qualification tests do not include tests for a full qualification of all functions of the computer, data security, access protection or other requirements according the IT within a regulated environment. These requirements should be addressed in a framework of an overall, enterprise-wide IT policy. Release of the Installation After successfully finishing of all qualifying steps, the installation should be released for usage with date and signature by the equipment owner and qualified person according to GMP. Thermoguard Installation, Operation and Performance Qualification Page 4 of 4 Subject to changes.

5 Checklist: Qualification Documents Design qualification: User requirement specifications documented and released? Design qualification: Data sheets added to documentation? Installation qualification: Equipment binder prepared and equipment owner trained? Installation qualification: Device list exported, printed, marked with date and signed? Installation qualification: Configuration list ("TGuard.ini") printed, marked with date and signed? Operational qualification: All sensors successfully polled? Corresponding log file printed, marked with date and signed? Operational qualification: Test mail successfully send, printed, marked with date and signed? Performance qualification: All temperature and humidity sensors successfully and traceable calibrated? Thermoguard Installation, Operation and Performance Qualification Annex Subject to changes

6 Identification of the Thermoguard Installation Owner of the installation: Location of the installation: Date of Installation: Equipment owner: Release of the Thermoguard Installation The Thermoguard Installation described above fulfills all requirements of the installation, operation and performance qualification. Therefore, the installation is released for its intended use. Prepared: Date Signature Approved: Date Signature Released: Date Signature Thermoguard Installation, Operation and Performance Qualification Annex Subject to changes

Monitoring manufacturing, production and storage environments in the pharmaceutical industry

Monitoring manufacturing, production and storage environments in the pharmaceutical industry Application Description AD/RandC/005-EN Monitoring manufacturing, production and storage environments in the pharmaceutical industry - Provides independent verification and validation of the manufacture,

More information

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known

More information

LabChip GX/GXII with LabChip GxP Software

LabChip GX/GXII with LabChip GxP Software Regulatory Compliance LabChip GX/GXII with LabChip GxP Software Supporting Regulatory Compliance Caliper LabChip GX/GXII suite of instruments provides automated electrophoresis to analyze quality, size,

More information

OPERATIONAL STANDARD

OPERATIONAL STANDARD 1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.

More information

Contamination Control and Environmental Monitoring Systems

Contamination Control and Environmental Monitoring Systems Life Science Contamination Control and Environmental Monitoring Systems Hospital Electronics Brookhaven Instruments deliver, install and validate particle counters, Microbial Air Samplers and Monitoring

More information

Testing Automated Manufacturing Processes

Testing Automated Manufacturing Processes Testing Automated Manufacturing Processes (PLC based architecture) 1 ❶ Introduction. ❷ Regulations. ❸ CSV Automated Manufacturing Systems. ❹ PLCs Validation Methodology / Approach. ❺ Testing. ❻ Controls

More information

MHRA GMP Data Integrity Definitions and Guidance for Industry January 2015

MHRA GMP Data Integrity Definitions and Guidance for Industry January 2015 MHRA GMP Data Integrity Definitions and Guidance for Industry Introduction: Data integrity is fundamental in a pharmaceutical quality system which ensures that medicines are of the required quality. This

More information

Complete Document & Process Management for Life Sciences on SharePoint 2010

Complete Document & Process Management for Life Sciences on SharePoint 2010 TM ComplianceSP TM on SharePoint 2010 Complete Document & Process Management for Life Sciences on SharePoint 2010 Overview With increasing pressure on costs and margins across Life Sciences, the Industry

More information

User Guide. Temperature and Humidity Datalogger. Model 42280

User Guide. Temperature and Humidity Datalogger. Model 42280 User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod

More information

Guidance on Qualification of existing facilities, systems, equipment and utilities

Guidance on Qualification of existing facilities, systems, equipment and utilities QUALIFICATION_EXISTING_EQUIPMENT_FINAL page 1 / 16 1. Acknowledgement...3 2. Introduction...3 3. Scope...4 4. Regulatory requirements...4 5. Guidance...4 5.1 Risk Assessment... 4 5.2 Procedure... 7 5.3

More information

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,

More information

GAMP 5 and the Supplier Leveraging supplier advantage out of compliance

GAMP 5 and the Supplier Leveraging supplier advantage out of compliance GAMP 5 and the Supplier Leveraging supplier advantage out of compliance Paul Osborne Performance PharmaTech Ltd. Overview This document is designed to assist suppliers who wish to sell computer based equipment

More information

Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice

Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 Working Party on Control

More information

Computerised Systems. Seeing the Wood from the Trees

Computerised Systems. Seeing the Wood from the Trees Computerised Systems Seeing the Wood from the Trees Scope WHAT IS A COMPUTERISED SYSTEM? WHY DO WE NEED VALIDATED SYSTEMS? WHAT NEEDS VALIDATING? HOW DO WE PERFORM CSV? WHO DOES WHAT? IT S VALIDATED -

More information

This interpretation of the revised Annex

This interpretation of the revised Annex Reprinted from PHARMACEUTICAL ENGINEERING The Official Magazine of ISPE July/August 2011, Vol. 31 No. 4 www.ispe.org Copyright ISPE 2011 The ISPE GAMP Community of Practice (COP) provides its interpretation

More information

Wireless Data Logger Monitoring Systems

Wireless Data Logger Monitoring Systems Wireless Data Logger Monitoring Systems Table of Contents About MadgeTech, Inc. Wireless Systems & Components Wireless Data Loggers 2000A Wireless Data Logger Models MadgeTech Cloud Services MadgeTech

More information

Qualification, calibration and maintenance of stability chambers. Dr. Mathias Kramer

Qualification, calibration and maintenance of stability chambers. Dr. Mathias Kramer Qualification, calibration and maintenance of stability chambers Dr. Mathias Kramer Agenda 1. Qualification of stability chambers 2. Calibration of equipment of stability chambers 3. Maintenance of stability

More information

(WHO Technical Report Series, No. 961, 2011), Annex 9. Temperature and humidity monitoring systems for fixed storage areas 107

(WHO Technical Report Series, No. 961, 2011), Annex 9. Temperature and humidity monitoring systems for fixed storage areas 107 Annex 5 Technical supplements to Model guidance for the storage and transport of timeand temperature-sensitive pharmaceutical products (WHO Technical Report Series, No. 961, 2011), Annex 9 1. The technical

More information

To purchase a full bound copy of this document contact

To purchase a full bound copy of this document contact 29570 SUB Aqua Plus IQOQ document Unique Product Serial number label placed here To purchase a full bound copy of this document contact Grant Instruments (Cambridge) Ltd Shepreth Cambridgeshire SG8 6GB

More information

MHRA GMP Data Integrity Definitions and Guidance for Industry March 2015

MHRA GMP Data Integrity Definitions and Guidance for Industry March 2015 MHRA GMP Data Integrity Definitions and Guidance for Industry Introduction: Data integrity is fundamental in a pharmaceutical quality system which ensures that medicines are of the required quality. This

More information

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories

More information

OECD DRAFT ADVISORY DOCUMENT 16 1 THE APPLICATION OF GLP PRINCIPLES TO COMPUTERISED SYSTEMS FOREWARD

OECD DRAFT ADVISORY DOCUMENT 16 1 THE APPLICATION OF GLP PRINCIPLES TO COMPUTERISED SYSTEMS FOREWARD OECD DRAFT ADVISORY DOCUMENT 16 1 THE APPLICATION OF GLP PRINCIPLES TO COMPUTERISED SYSTEMS FOREWARD 1. The following draft Advisory Document will replace the 1995 OECD GLP Consensus Document number 10

More information

GAMP5 - a lifecycle management framework for customized bioprocess solutions

GAMP5 - a lifecycle management framework for customized bioprocess solutions GE Healthcare Life Sciences GAMP5 - a lifecycle management framework for customized bioprocess solutions imagination at work GE Healthcare s engineering department, Customized Bioprocess Solutions (CBS),

More information

FAT Factory Acceptance Testing SAT Site Acceptance Testing

FAT Factory Acceptance Testing SAT Site Acceptance Testing FAT Factory Acceptance Testing SAT Site Acceptance Testing Dr. Ingrid Walther May 2015, Page 1 Factory Acceptance Testing - FAT Site Acceptance Testing - SAT FAT and SAT - Important Milestones in the Project

More information

Unit 7: Electrical devices LO2: Understand electrical sensors and actuators Sensors temperature the thermistor

Unit 7: Electrical devices LO2: Understand electrical sensors and actuators Sensors temperature the thermistor Unit 7: Electrical devices LO2: Understand electrical sensors and actuators Sensors temperature the thermistor Instructions and answers for teachers These instructions should accompany the OCR resource

More information

EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union

EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union EUROPEAN COMMISSION HEALTH AND CONSUMERS DIRECTORATE-GENERAL Public Health and Risk Assessment Pharmaceuticals Brussels, SANCO/C8/AM/sl/ares(2010)1064599 EudraLex The Rules Governing Medicinal Products

More information

Training Course Computerized System Validation in the Pharmaceutical Industry Istanbul, 16-17 January 2003. Change Control

Training Course Computerized System Validation in the Pharmaceutical Industry Istanbul, 16-17 January 2003. Change Control Training Course Computerized System Validation in the Pharmaceutical Industry Istanbul, 16-17 January 2003 Change Control Wolfgang Schumacher Roche Pharmaceuticals, Basel Agenda Change Control Definitions

More information

Qualification of temperature-controlled storage areas

Qualification of temperature-controlled storage areas WHO Vaccine Qualification of temperature-controlled storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport of time and temperature

More information

Monitoring the autoclaving process in the pharmaceutical industry

Monitoring the autoclaving process in the pharmaceutical industry Application Description AD/RandC/006-EN Monitoring the autoclaving process in the pharmaceutical industry - Provides independent verification and validation monitoring of the autoclaving process - Enables

More information

Guidelines for new users of the Union Registry ECAS account, registration and enrolment

Guidelines for new users of the Union Registry ECAS account, registration and enrolment Guidelines for new users of the Union Registry ECAS account, registration and enrolment An ECAS account and registration are required for a new user to be added as an account representative Before a new

More information

NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING

NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING Instructions to the Assessor: This checklist addresses specific accreditation criteria prescribed in NIST Handbook 150-5, Construction Materials

More information

ENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013

ENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013 ENERGY STAR Program Requirements Product Specification for Data Center Storage Draft 2 Test Method March 2013 1 2 3 4 1 OVERVIEW The following test method shall be used for determining compliance with

More information

The Central Monitoring System we prove it.

The Central Monitoring System we prove it. - Reliable - Flexible - Compliant The Central Monitoring System we prove it. 2 intro intro 3 The Central Monitoring System A networked LAN wired or wireless sensor system Reliable There are several attributes

More information

4H Online Re-Enrollment

4H Online Re-Enrollment 4H Online Re-Enrollment This instruction sheet if for returning 4H members only! 1. Navigate to the site ne.4honline.com. You will see changes have been made to the site since last time you logged in.

More information

Foreign Pharmaceutical Manufacturing Company Registration Guideline for Afghanistan

Foreign Pharmaceutical Manufacturing Company Registration Guideline for Afghanistan Islamic Republic of Afghanistan Ministry of Public Health General Directorate of Pharmaceutical Affairs Foreign Pharmaceutical Manufacturing Company Registration Guideline for Afghanistan 2015 Islamic

More information

CSZ s EZ-View Software

CSZ s EZ-View Software www.cszindustrial.com CSZ s EZ-View Software CSZ EZ-View for simpli ied chamber control, monitoring & data management CSZ s EZ-View software provides chamber control, monitoring, and data logging for up

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (08) 69 3R Full document title and reference Document type VALIDATION OF COMPUTERISED SYSTEMS Legislative basis - CORE DOCUMENT

More information

testo Saveris 21CFR Part 11 Software Instruction manual

testo Saveris 21CFR Part 11 Software Instruction manual testo Saveris 21CFR Part 11 Software Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 2 Specifications... 4 2.1. Intended purpose... 4 2.2. 21 CFR Part 11 and terminology used... 5 3 First steps...

More information

Using the ISPE s GAMP Methodology to Validate Environmental Monitoring System Software

Using the ISPE s GAMP Methodology to Validate Environmental Monitoring System Software Using the ISPE s GAMP Methodology to Validate Environmental Monitoring System Software TEST DOCUMENTS DETAILED DOCUMENTS FUNCTIONAL SPECS USER REQUIREMENTS Introduction Continuous Monitoring Systems (CMS)

More information

Applicant Workflow Hiring Managers

Applicant Workflow Hiring Managers Applicant Workflow Hiring Managers Below is a visual diagram for the staff applicant flows and statuses in the Santa Clara University Online Recruitment System. The applicant workflow process: 1. The Applicant

More information

DataView DVTH Data Logger for Temperature and Humidity

DataView DVTH Data Logger for Temperature and Humidity DataView DVTH Data Logger for Temperature and Humidity Logs temperature, humidity, and dew point Built-in real time text and graphical display Stand-alone or PCconnected operation Deep sample memory Battery-powered

More information

Medical Device Labeling Overview

Medical Device Labeling Overview Medical Device Labeling Overview Labeling Task Group IOPP Medical Device Packaging Committee October 22, 2003 Opening Slide Forces that drive the move toward bar codes Device Manufacturers use bar codes

More information

endless One platform possibilities for DNA, RNA, protein and cell analysis Agilent 2100 Bioanalyzer Our measure is your success.

endless One platform possibilities for DNA, RNA, protein and cell analysis Agilent 2100 Bioanalyzer Our measure is your success. One platform endless possibilities for DNA, RNA, protein and cell analysis Agilent 2100 Bioanalyzer Our measure is your success. products applications software services Agilent 2100 bioanalyzer Applied

More information

Audit Report in the framework of the APIC Audit Programme

Audit Report in the framework of the APIC Audit Programme Audit Report in the framework of the APIC Audit Programme Company Location Address Country Subject of Audit Audit Date Auditor (lead) Co-Auditor(s) Participants Signature of Auditor(s) Lead Auditor Co-Auditor

More information

Frequently Asked Questions UNIFI V1.8.1 Upgrade

Frequently Asked Questions UNIFI V1.8.1 Upgrade Frequently Asked Questions UNIFI V1.8.1 Upgrade Q. What are the features and benefits of upgrading to UNIFI V1.8.1 Software? A. UNIFI V1.8.1 incorporates significant improvements across all application

More information

How To Control A Medical Device With A Pc System

How To Control A Medical Device With A Pc System Clean room monitoring and disinfection FMS 5.0, the Leading Facility Monitoring Software TSI FMS system is an advanced, reliable, and user-friendly monitoring software suite that features a truly open

More information

Pharmaceutical Wholesaler Site Inspection Checklist

Pharmaceutical Wholesaler Site Inspection Checklist Pharmaceutical Wholesaler Site Inspection Checklist Date and time of inspection: Legal business name and plant name: Location (address): Phone number: Fax number: Email address: Web site URL: Contact Person:

More information

Voting Systems Checklist

Voting Systems Checklist 4 12543 Westella, Suite 100, Houston, Texas 77077, 281-293-0101 Voting Systems Checklist Administrative Oversight 1. Have pertinent federal laws which affect voting systems been researched for an understanding

More information

Vaisala Veriteq vlog. Installation and Operational Qualification Protocol

Vaisala Veriteq vlog. Installation and Operational Qualification Protocol Vaisala Veriteq vlog Installation and Operational Qualification Protocol For Vaisala Veriteq vlog Version 4.5 Vaisala Canada Inc. Document #: M211382EN-A Date Prepared: July 22, 2011 Prepared by: Katie

More information

Temperature and humidity monitoring systems for fixed storage areas

Temperature and humidity monitoring systems for fixed storage areas WHO Vaccine Temperature and humidity monitoring systems for fixed storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport

More information

Operating Principle of the Sample Positioning Flaps

Operating Principle of the Sample Positioning Flaps Test WEIGHT, THICKNESS, DIAMETER (or LENGTH), and HARDNESS of regular tablets, odd shaped tablets (caplets, oblongs etc.) and other solid samples fully automatically. Use the WHT 3ME for the fully automated

More information

Evaluation copy. Build a Temperature Sensor. Project PROJECT DESIGN REQUIREMENTS

Evaluation copy. Build a Temperature Sensor. Project PROJECT DESIGN REQUIREMENTS Build a emperature Sensor Project A sensor is a device that measures a physical quantity and converts it into an electrical signal. Some sensors measure physical properties directly, while other sensors

More information

White paper: The 10 Golden Rules of GMP

White paper: The 10 Golden Rules of GMP White paper: The 10 Golden Rules of GMP The Good Manufacturing Practice regulations that govern pharmaceutical and medical device manufacturing can seem overwhelming. Use these ten golden rules to drive

More information

GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE VALIDATION OF COMPUTERISED SYSTEMS. Working Group on Information Technology (AGIT)

GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE VALIDATION OF COMPUTERISED SYSTEMS. Working Group on Information Technology (AGIT) GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE VALIDATION OF COMPUTERISED SYSTEMS Working Group on Information Technology (AGIT) Release Date: 14 December 2007 Version: 02 AGIT - Validation of Computerised

More information

Current Monitoring Kit QUICK START GUIDE

Current Monitoring Kit QUICK START GUIDE Current Monitoring Kit QUICK START GUIDE Pico Technology EL040 Current Monitoring Kit Quick Start Guide CONTENTS 1 Introduction... 1 2 Kit contents... 3 3 Installation... 3 3.1 Connecting the system together...

More information

IMP management at site. Dmitry Semenyuta

IMP management at site. Dmitry Semenyuta IMP management at site Dmitry Semenyuta TOP 5 FDA inspections finding 1999-2009 Center of Drug Evaluation and Research (CDER) Failure to follow the protocol Failure to keep adequate and accurate records

More information

Current Monitoring Kit

Current Monitoring Kit Current Monitoring Kit QUICK START GUIDE DO090-6 CONTENTS Issues: 1) 2.10.02 WP A4 format 2) 2.10.02 Added safety warning 3) 17.3.06 Word A5 format. S1: Removed relay modules. S2: Added MI010. S4.1: added

More information

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Service Bulletin: #001 Date: August 2012 Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Summary This Application Bulletin will give a short synopsis of the purpose of a calibration and

More information

Validation Approach and Scope for Business Process System Validation. Epitome Technologies Private Limited

Validation Approach and Scope for Business Process System Validation. Epitome Technologies Private Limited Validation Approach and Scope for Business Validation Epitome Technologies Private Limited PROPOSAL Page 2 of 8 Table Of Contents: 1.0 GENERAL... 3 2.0 SCOPE MODULES COVERED... 3 3.0 QUALIFICATION APPROACH...

More information

ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners

ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners Tony Harrison and Bob Latimer ISO 21501- A Standard Methodology to Optical Particle Counter

More information

GMP Pre-Qualification of Separators Ex Works Flexible and Safe on the GEA Pharmaceutical Test Bay

GMP Pre-Qualification of Separators Ex Works Flexible and Safe on the GEA Pharmaceutical Test Bay GMP Pre-Qualification of Separators Ex Works Flexible and Safe on the GEA Pharmaceutical Test Bay engineering for a better world GEA Mechanical Equipment Customer Specific, Economical and Safe For manufacturers

More information

MONITORING PERFORMANCE IN WINDOWS 7

MONITORING PERFORMANCE IN WINDOWS 7 MONITORING PERFORMANCE IN WINDOWS 7 Performance Monitor In this demo we will take a look at how we can use the Performance Monitor to capture information about our machine performance. We can access Performance

More information

GOOD LABORATORY PRACTICE (GLP) Working Group on Information Technology (AGIT)

GOOD LABORATORY PRACTICE (GLP) Working Group on Information Technology (AGIT) GOOD LABORATORY PRACTICE (GLP) GUIDELINES FOR THE DEVELOPMENT AND VALIDATION OF SPREADSHEETS Working Group on Information Technology (AGIT) This is the pre-peer reviewed version of the following article:

More information

PTB 311E Manual 3-in-1 Tablet Testing Instrument

PTB 311E Manual 3-in-1 Tablet Testing Instrument PTB 311E Manual 3-in-1 Tablet Testing Instrument The manual tablet testing instrument PTB 311E is a dual force mode hardness test apparatus as it can be used for either linear force or linear speed increase

More information

ELECTRONIC TRACEABILITY SYSTEMS VS. PAPER BASED TRACEABILITY SYSTEMS

ELECTRONIC TRACEABILITY SYSTEMS VS. PAPER BASED TRACEABILITY SYSTEMS ELECTRONIC TRACEABILITY SYSTEMS VS. PAPER BASED TRACEABILITY SYSTEMS OYSTEIN HELLESOY 12 th May 2008 2008 FoodReg AG All rights reserved 1 Outline 1. ISO standard recommendations 2. Pros and cons of paper

More information

STD - STANDARD VERSION extended possibilities reasonable price wide range of optional equipment

STD - STANDARD VERSION extended possibilities reasonable price wide range of optional equipment ooled Incubators ooled incubators ooled incubators () are available in two versions: STD and TOP. STD STANDARD VERSION extended possibilities reasonable price wide range of optional equipment stainless

More information

RingCentral for Skype for Business USER GUIDE

RingCentral for Skype for Business USER GUIDE RingCentral for Skype for Business USER GUIDE Contents Introduction 3 About RingCentral for Skype for Business 3 About Skype for Business 3 About This Guide 3 Getting Started with RingCentral for Skype

More information

Print then Cut Calibration

Print then Cut Calibration The Print then Cut feature of Cricut Design Space for PC and Mac allows you to print your images from your home printer and then cut them out with high precision on your Cricut machine. Print then Cut

More information

Creating Rules in Outlook

Creating Rules in Outlook Technology Help Desk 412 624-HELP [4357] http://technology.pitt.edu Creating Rules in Outlook Overview Microsoft Outlook provides a way for users to manage and organize their email using rules. A rule

More information

When the template is complete, the whole Product Description document can be printed and approved.

When the template is complete, the whole Product Description document can be printed and approved. HOW TO USE THIS TEMPLATE: Introduction The template reflects the steps set out in the PRINCE2 Method and is designed to prompt the Project Manager and help in the creation of the Product Description. The

More information

NTTCT Mail Hosting Service Account Management

NTTCT Mail Hosting Service Account Management NTTCT Mail Hosting Service Account Management (Mail Hosting: NTT Communications (Thailand) Co., Ltd.) About This Document This document is intended to be a quick reference guide to follow for administrator

More information

ComplianceSP TM on SharePoint. Complete Document & Process Management for Life Sciences on SharePoint 2010 & 2013

ComplianceSP TM on SharePoint. Complete Document & Process Management for Life Sciences on SharePoint 2010 & 2013 TM ComplianceSP TM on SharePoint Complete Document & Process Management for Life Sciences on SharePoint 2010 & 2013 Overview With increasing pressure on costs and margins across Life Sciences, the industry

More information

Dräger X-dock Release Notes Software-Update 01.07.00 and Manager Software 1.7.0

Dräger X-dock Release Notes Software-Update 01.07.00 and Manager Software 1.7.0 Supported operating system (X-dock Manager) Windows 7 Windows 8 Windows Server 2008 R2 Windows Server 2012 R2 Supported firmware versions (X-dock Manager) X-am 2000: > = v6.0 X-am 2500/5x00: > = v6.6 Pac

More information

Case study Realistic and best uncertainties with humidity instruments

Case study Realistic and best uncertainties with humidity instruments Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against

More information

Process Validation for Medical Devices

Process Validation for Medical Devices Process Validation for Medical Devices Dan O Leary CBA, CQE, CRE, CSSBB, CIRM, LLC 603-209-0600 OmbuEnterprises@msn.com Copyright 2010 by, LLC Process Validation for Medical Devices 1 Instructor Introduction

More information

Computer System Validation - It s More Than Just Testing

Computer System Validation - It s More Than Just Testing Computer System Validation - It s More Than Just Testing Introduction Computer System Validation is the technical discipline that Life Science companies use to ensure that each Information Technology application

More information

A L ERT. Quick Start With

A L ERT. Quick Start With A L ERT Quick Start With I NSTAL L ATION To launch the installation of Micromedia Solution, insert the installation CD-ROM in the CD/DVD drive. If the installation program does not start automatically

More information

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to INTRODUCTION This book offers a systematic, ten-step approach, from the decision to validate to the assessment of the validation outcome, for validating configurable off-the-shelf (COTS) computer software

More information

REAL TIME LABORATORY MONITORING SYSTEM

REAL TIME LABORATORY MONITORING SYSTEM REAL TIME LABORATORY MONITORING SYSTEM MONITORING LET XILTRIX WATCH OVER YOUR LABORATORY! 20 years experience and IKS expertise makes XiltriX the total monitoring solution for any laboratory. All day,

More information

Risk management is one of the new requirements for. An Integrated Risk Assessment for Analytical Instruments and Computerized Laboratory Systems

Risk management is one of the new requirements for. An Integrated Risk Assessment for Analytical Instruments and Computerized Laboratory Systems vember 2013 Spectroscopy 28(11) 1 Focus on Quality An Integrated Risk Assessment for Analytical Instruments and Computerized Laboratory Systems A risk assessment is presented for determining the amount

More information

WinKQCL 5 Endotoxin Detection and Analysis Software We Analyze Endotoxin Data Every Day

WinKQCL 5 Endotoxin Detection and Analysis Software We Analyze Endotoxin Data Every Day Pharma&Biotech WinKQCL 5 Endotoxin Detection and Analysis Software We Analyze Endotoxin Data Every Day For Endotoxin Detection Pharma&Biotech WinKQCL 5 Software for Endotoxin Detection and Analysis WinKQCL

More information

Application Note Room Alert

Application Note Room Alert Application Note Room Alert Monitor Room Alert 24E With PageR Enterprise Instructions for Using Room Alert 24E With PageR Enterprise The Room Alert 24E ID Box is one of AVTECH Software s hardware solutions

More information

Temperature and humidity monitoring systems for fixed storage areas

Temperature and humidity monitoring systems for fixed storage areas QAS/14.598 Supplement 6 WHO Vaccine Temperature and humidity monitoring systems for fixed storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the

More information

Review and Approve Results in Empower Data, Meta Data and Audit Trails

Review and Approve Results in Empower Data, Meta Data and Audit Trails Review and Approve Results in Empower Data, Meta Data and Audit Trails 2013 Waters Corporation 1 What is an audit trail? Systematic story of the data from creation, through interpretation and final assessment

More information

Microsoft Office. Mail Merge in Microsoft Word

Microsoft Office. Mail Merge in Microsoft Word Microsoft Office Mail Merge in Microsoft Word TABLE OF CONTENTS Microsoft Office... 1 Mail Merge in Microsoft Word... 1 CREATE THE SMS DATAFILE FOR EXPORT... 3 Add A Label Row To The Excel File... 3 Backup

More information

testo EasyClimate Software Instruction manual

testo EasyClimate Software Instruction manual testo EasyClimate Software Instruction manual 2 1 Contents 1 Contents 1 Contents... 3 1.1. About this document... 5 2 Specifications... 5 2.1. Use... 5 2.2. System requirements... 6 3 First steps... 6

More information

How to use the OMEGALOG software with the OM-SQ2010/SQ2020/SQ2040 Data Loggers.

How to use the OMEGALOG software with the OM-SQ2010/SQ2020/SQ2040 Data Loggers. How to use the OMEGALOG software with the OM-SQ2010/SQ2020/SQ2040 Data Loggers. OMEGALOG Help Page 2 Connecting Your Data Logger Page 2 Logger Set-up Page 3 Download Data Page 8 Export Data Page 11 Downloading

More information

THUM - Temperature Humidity USB Monitor

THUM - Temperature Humidity USB Monitor THUM - Temperature Humidity USB Monitor The THUM is a true USB device to monitor temperature and relative humidity of an interior living, working, and storage spaces. The THUM is ideal for computer rooms,

More information

SOLAARsecurity. Administrator Software Manual. 010508 Issue 2

SOLAARsecurity. Administrator Software Manual. 010508 Issue 2 SOLAARsecurity Administrator Software Manual 9499 400 40011 010508 Issue 2 2008. All rights reserved. SOLAAR House, 19 Mercers Row, Cambridge CB5 8BZ.United Kingdom. Telephone +44 (0) 1223 347400, Fax

More information

JOFRACAL Temperature and Pressure Calibration Software

JOFRACAL Temperature and Pressure Calibration Software 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-571-8431 Visit us at www.testequipmentdepot.com Specification Sheet SS-CP-2510-US Temperature and Pressure Multiple possibilities

More information

MCP 100 MCP 150. Modular Compact Polarimeters. ::: Superior Optical Instruments

MCP 100 MCP 150. Modular Compact Polarimeters. ::: Superior Optical Instruments MCP 100 MCP 150 Modular Compact Polarimeters ::: Superior Optical Instruments Your Promise Anton Paar's Solution You have always made sure that your optically active substances meet all quality standards.

More information

Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011

Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 WHO Vaccine Temperature mapping of storage areas Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport of time and temperature sensitive

More information

InfinityQS SPC Quality System & FDA s 21 CFR Part 11 Requirements

InfinityQS SPC Quality System & FDA s 21 CFR Part 11 Requirements InfinityQS SPC Quality System & FDA s 21 CFR Part 11 Requirements www.infinityqs.com Copyright InfinityQS International Table of Contents Overview... FDA s 21 CFR Part 11 Requirements... PART 11 ELECTRONIC

More information

SHIMADZU CORPORATION

SHIMADZU CORPORATION 638-94259 Total Organ Carbon Analyzer TOC-V Administrator's Manual Read this manual carefully and keep it with the instrument for future reference. SHIMADZU CORPORATION ANALYTICAL & MEASURING INSTRUMENTS

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (08) 88 R VALIDATION OF COMPUTERISED SYSTEMS ANNEX 2: VALIDATION OF DATABASES (DB), LABORATORY INFORMATION MANAGEMENT SYSTEMS

More information

FDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry

FDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry FDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry The Rule 21 CFR Part 11 Handwritten signature means the scripted name or legal mark of an individual handwritten

More information

Created April 2009. This application note shows how to set up the TSP100LAN for kitchen ticket printing from POSERA Maitre D Software.

Created April 2009. This application note shows how to set up the TSP100LAN for kitchen ticket printing from POSERA Maitre D Software. Created April 2009 This application note shows how to set up the TSP100LAN for kitchen ticket printing from POSERA Maitre D Software. Recommended Restaurant Printer Options: Splash Proof Cover, Vertical

More information

Avira Free Android Security (version 1.2) HowTo

Avira Free Android Security (version 1.2) HowTo Avira Free Android Security (version 1.2) HowTo Table of contents 1. Introduction... 4 2. Supported Mobile Manufacturers... 4 3. Product Overview... 5 4. Installation... 5 4.1 App installation on the device...5

More information

Considerations for validating SDS Software v2.x Enterprise Edition for the 7900HT Fast Real-Time PCR System per the GAMP 5 guide

Considerations for validating SDS Software v2.x Enterprise Edition for the 7900HT Fast Real-Time PCR System per the GAMP 5 guide WHITE PAPER SDS Software v2.x Enterprise Edition Considerations for validating SDS Software v2.x Enterprise Edition for the 7900HT Fast Real-Time PCR System per the GAMP 5 guide This white paper describes

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information