Soft Contact Lens Material Properties Author: Dr Alison Alderson, PhD MCOptom

Size: px
Start display at page:

Download "Soft Contact Lens Material Properties Author: Dr Alison Alderson, PhD MCOptom"

Transcription

1 Soft Contact Lens Material Properties Author: Dr Alison Alderson, PhD MCOptom Contact lens manufacturers are constantly striving to develop the perfect contact lens material, one which satisfies both the patient and practitioner and can be produced cost effectively at a rate which supplies demand. Patient satisfaction, the physiological response of the lens on the eye, reproducibility, predictability and the commercial aspects of the lens are priorities from a practitioner s perspective. From a patient s point of view the lens is required to be comfortable, give excellent vision, be easy to handle, convenient and affordable. The aim of this article is to provide an overview of soft contact lens material properties and also highlight the more recent developments in this area. Hydrogel lenses The first soft lens monomer hydroxyethyl methacrylate (HEMA), was developed in the late 1950s by Otto Wichterle, however, it wasn t until the 1970s that the first commercial soft contact lens became available. During soft contact lens production, monomers such as HEMA, with the aid of a catalyst, undergo polymerization to produce a sequence of repeating units termed a polymer. These polymers are bound to each other by the process of cross-linking, creating pores in the material into which fluid can enter. If more than one monomer is used in the material it is termed a copolymer [1]. A polymerised HEMA (phema) lens has a water content of about 38%, however, by varying the monomers used, materials with varying water contents, refractive indices, hardness, strength and oxygen permeabilities can be created. Silicone hydrogel lenses Silicone hydrogel (SiH) lenses were launched in 1999 by Bausch and Lomb and Ciba Vision. Silicone elastomers had already been used to create high oxygen permeability lenses but they had poor wettability and comfort, attracted deposits and caused problems with corneal adherence. SiH lenses combined silicone with conventional hydrogel monomers. Silicone is extremely oxygen permeable, even more so than water. The more silicone a lens contains the more permeable it will be. However, silicone is not without its problems; it is hydrophobic and relatively stiff. This created a challenge for the early SiH lens manufacturers as it was necessary to incorporate hydrophilic coatings on the surface of the lenses to give patients the levels of comfort and wettability they desired. Contact lens material classification Contact lens materials are classified using the ISO system. BS EN ISO : 2006/DAM1 sets out the new international standard method for the classification of contact lens materials [2]. Each material is classified by a 6 part code as shown in Table 1. Oxygen performance The passage of oxygen through a contact lens can be quantified in a number of ways. Most commonly quoted by contact lens manufacturers is oxygen transmissibility (Dk/t). The Dk of a material represents its oxygen permeability. In the past this was measured in CGS(FATT) units, however, the International Standard for Dk measurement now recommends using International System Units (SI) instead i.e. changing from the imperial measurements of mmhg to metric measurements of hpa. The units are (cm2/s).[ml.o2/(ml.hpa)]. Generally Dk is referred to as being in ISO Dk units with values in the order of The Dk/t of a lens is the Dk of its

2 material divided by the thickness of the contact lens in cm. FATT units can be converted to ISO units for both Dk and Dk/t by multiplying the value by Generally contact lens manufacturers quote the Dk/t at the centre of a -3D lens. Although this gives practitioners a single number to compare, it tends to overestimate both the Dk/t at the periphery of the lens and also at the centre of lenses of different powers [3]. This is an important factor, as a lower Dk/t, particularly in the periphery of the lens, has been linked to an increase in limbal hyperaemia [4]. It may be more appropriate to use an average thickness when calculating Dk/t and it has been proposed that the mean harmonic thickness may be the best standard to use [5]. Mean harmonic thickness averages the radial thicknesses at the centre of a lens and the periphery of at least five annular zones extending to the edge of the optic zone [6, 7]. It takes into account the thickness of the lens when calculating these zones. There are two main techniques for measuring the Dk/t of a lens in vitro; polarographic measurement and coulometric measurement. Measurements are dependent on environmental conditions and should be made at 35⁰C ± 1⁰. Errors can occur due to the edge effect and lens dehydration. For this reason, measurements taken by different laboratories must be viewed with caution as they may not be directly comparable. The Dk/t for a number of commercially available lenses are shown in Figure 1. Figure 1 The normal cornea relies on oxygen from the atmosphere to maintain metabolic activity, and a contact lens on the eye acts as a barrier to this. There is, however, a significant variation in the amount of oxygen an individual cornea requires [8].The laboratory methods for measuring the passage of oxygen through a lens do not give us any information regarding the amount of oxygen reaching the anterior surface of the cornea per unit time, or how much oxygen is available for the cornea to metabolise. It has been suggested that using oxygen flux (the amount of oxygen which actually passes through a lens on the eye) [9] may be a better way of representing the oxygen performance of a contact lens. So how much oxygen is enough? A number of studies have produced recommendations as to the amount of oxygen the average cornea requires for daily wear and extended wear of contact lenses [3, 4, 10, 11]. These are listed in Table 2 [10, 11]. The results vary depending on the methodology and the indicator used to determine the critical oxygen value, however, it is generally accepted that a level of ISO Dk units in the open eye and ISO Dk units in the closed eye are necessary.

3 The oxygen permeability of a conventional hydrogel material increases exponentially with increased water content [12]. As oxygen transmissibility is linked to lens thickness as well as Dk, the thinner the lens the higher the Dk/t, therefore for optimum Dk/t a hydrogel lens would be very thin with a high water content. In practice, however, this creates a fragile lens that dehydrates quickly and would therefore not be acceptable from a patient s point of view. As a compromise, lenses have tended to be manufactured from mid water content materials (55-60%) with centre thicknesses of around 0.08 mm. This, however, does not provide a high enough Dk/t to meet the critical oxygen requirements of the cornea. The recent development of a high water content, bio inspired material (Hypergel) with a Dk/t of 42 ISO units, aims to tackle both the oxygen transmissibility and dehydration issues. SiH lenses also contain water, however, the Dk of the material is dependent on the silicone content of the material, as silicone transports oxygen better than water. If water is removed from the material and replaced with silicon the Dk increases [12]. The introduction of SiH materials has reduced concern about corneal hypoxia, particularly in daily wear lenses, as these lenses all meet the critical oxygen requirements of the cornea both centrally and peripherally. It appears, however, that high Dk/t alone does not guarantee lens comfort or corneal integrity, as these factors are influenced by a number of other material properties such as modulus, wettability, biocompatibility, friction and deposit resistance. Modulus The modulus of a material describes its ability to keep its shape and resist deformation. The modulus of a contact lens is measured using Young s modulus (E), measured in megapascals (MPa) and is described as the ability of a lens material to align to the ocular surface and resist deformation under tension. It is the ratio of stress to strain: stress being the force per unit area required to change the shape of a solid, and strain being the deformation undergone by the sample in the direction of the force applied during testing [13]. Modulus is an inherent property of the material itself, and the stiffness of a contact lens is also dependant on its design and thickness. The modulus values for some of the currently available SiH contact lens materials are shown for comparison in Figure 2, however, at present there is no standardised method of measurement, which means comparison of figures from different sources must be made with caution. Figure 2

4 The early SiH lens materials (lotrafilcon A; Night and Day and balafilcon A; PureVision), have a modulus two or three times higher than that of phema. This higher modulus makes the lenses easier to handle (an advantage for older, presbyopic patients), however, evidence suggests a relationship between some of these high oxygen transmissibility, high modulus materials [14] and some non-inflammatory ocular complications such as superior epithelial arcuate lesions (SEALs) (Figure 3), mucin balls and contact lens induced papilliary conjunctivitis (CLPC) (Figure 4). As stiffer lenses tend not to drape the cornea in the same way as a conventional hydrogel, fluting has been seen around the edges of these high modulus lenses suggesting more practitioner skill and additional parameters may be needed to achieve a successful fit. Anecdotal evidence suggests that patients are initially more aware of SiH lenses than conventional hydrogels and a correlation between increased modulus and decreased comfort has been indicated [15], however, it appears that the majority of hydrogel lens wearers can be refitted with a SiH material with no comfort issues [16]. Surprisingly, there is evidence that the increased Dk/t of SiH lens materials has not reduced the incidence of microbial keratitis and inflammatory events as compared to conventional hydrogel lenses, even on a daily wear basis [17]. This suggests the benefits of high Dk SiH materials may be offset by the relatively high modulus [18] with corneal insult possibly being a precursor to microbial infection. Figure 3 SEAL Figure 4 contact lens induced papilliary conjunctivitis (CLPC)

5 Later SiH lens materials such as galyfilcon A (Acuvue Advance) and comfilcon A (Biofinity) have a lower modulus which may make them more acceptable to the patient, with greater initial comfort and less lens awareness. In theory there is also the possibility of less ocular insult and therefore a lower rate of inflammatory events and microbial infection, however, at present there is little scientific evidence to support this. Contact lens surface characteristics The surface characteristics of a contact lens, such as wettability, biocompatibility and lubricity, and the interaction they have with the ocular surface and eyelids are extremely important when considering patient comfort. Placing a lens on the cornea interferes with the normal tear film as it separates it into two layers; the pre-lens and post-lens tear films. Ideally the contact lens needs to mimic the wettability of the cornea enabling the formation of a good, stable pre and post-lens tear film which in turn will give clear vision and good comfort throughout the day. Wettability is the ability of a liquid to adhere to a solid surface, and is determined by a balance between the adhesive and cohesive forces acting on the surface [19]. Cohesion is the attraction between molecules that are the same i.e. the force which holds a drop of liquid together. Adhesion is the attraction between dissimilar substances i.e. the contact lens surface and the tears. A drop of liquid is created because the surface molecules are pulled inwards by the adhesive forces within the liquid. These forces are unopposed which means the outer molecules have excess potential energy; known as surface tension. The lower the surface tension of a liquid the more likely it is to spread across a solid surface. The surface tension of tears can be reduced by using wetting drops during lens wear. When a drop of liquid comes in to contact with a solid surface the excess potential energy that the molecules at the surface of the liquid have is reduced by attraction to molecules at the surface of the solid. This new energy value is the interfacial tension, and the lower this is the greater the attraction between the liquid and solid will be, therefore improving wettability. Using wetting solution on lens insertion will reduce the interfacial tension between the tears and lens. Finally increasing the surface tension of a solid improves wettability and this can be achieved by surface treatment of the contact lens. The wettability of a material can be measured in vitro using the contact angle which is formed when a drop of liquid is placed on a solid surface (Figure 5). Its size can be used to provide an indication of how wettable a surface is. There are various methods of measuring contact angle; the three most common being the Sessile drop, Captive bubble and Wilhelmy balance techniques. The smaller the contact angle the more wettable the surface, however, the method used affects the resulting measurement [20]. Manufacturers quoted measurements of wettability must therefore be compared with caution, although independent research has shown them to be reasonably accurate at room temperature [12]. Figure 5 Measuring wettability

6 These in vitro measures do not necessarily relate to the in vivo performance of a contact lens. In vivo wettability is much more difficult to measure. It is often done in practice using slit lamp biomicroscopy or keratometer mires, however, these are relatively subjective techniques. Researchers are constantly striving for more quantitative methods of measuring in vivo contact lens wettability and a method of measuring the contact angle on the eye has recently been developed [21]. The difficulty in acquiring consistent, quantitative, on eye measurements may be the reason studies have found that symptoms of dryness do not necessarily appear to correlate with signs [22, 23]. Two properties of the material itself can affect the wettability of hydrogel lenses [19]. The first is that they dehydrate throughout the day therefore becoming less wettable, and the second is that the polymer chains making up the materials have both hydrophilic and hydrophobic groups. These polymer chains are mobile and the hydrophilic groups can rotate towards the lens surface if exposed to water, but away from the lens surface if they come in to contact with a hydrophobic surface such as air or lipid from the tear film. Lubricity is a relatively new concept in the world of contact lenses. It is also referred to as the coefficient of friction (CoF) of a lens and is quantitatively measured by microtribometry. It refers to how smooth and slippery the surface of a lens is, which in turn affects the friction between the lens and eyelid. A low CoF allows the eyelid to flow over the surface of the lens and improves patient comfort [24]. Figure 6 shows the CoF for a number of contact lenses on the market [25]. At present there is no standard method of measurement for CoF making comparison between measurements from different sources difficult. Figure 6 Various techniques have been used by the contact lens industry to improve the wettability, biocompatibility and lubricity of conventional hydrogel lenses. Biocompatibility describes the ability of the lens to be in harmony with the eye and have little or no adverse effects on the ocular surface. Biomimisis is the development of artificial biomaterials that mimic natural forms and effects, and is the concept behind the material omafilcon A (Proclear). This material contains a synthetic analogue of a naturally occurring phospholipid (phosphorycholine) which encourages moisture retention. Etafilcon A (1 Day Acuvue Moist), contains a hydrophilic polymer which is embedded in the matrix of the material. In nelifilcon A (Dailies AquaComfort Plus), blink action is used to release a moisturising agent, Polyvinyl Alcohol (PVA) throughout the day. The majority of the PVA is bound within the matrix of the lens, however, a small amount is free and is released

7 throughout the day. Ciba Vision have gone a step further and added hydroxypropylmethyl cellulose (HPMC) to the saline storage solution for comfort on insertion, and the hydrophilic wetting agent Polyethylene Glycol (PEG) into the lens to provide comfort early on in the day and to bind to the PVA to slow its release. There is evidence this technology produces a more stable pre-lens tear film [26]. New materials are constantly being developed to address the issue of biocompatibility and wettability. A new daily lens has been developed from a bio inspired material (Hypergel) which has the same water content as the cornea (78%). The outer surface is designed to mimic the lipid layer of the tear film to prevent dehydration and maintain constant optics. The lens does not contain silicone, however, it has a Dk/t of 42 ISO units at the centre of a -3.00D lens, which is adequate for oxygen delivery to the cornea in open eye conditions by delivering an oxygen flux percentage of 93%. Silicone is a naturally hydrophobic material and the challenge faced by the lens manufacturers was to develop methods of making SiH lenses more wettable and therefore more acceptable to the ocular surface. One method of achieving this is surface treatment, which was the technique used by the manufacturers of the early SiH lenses [27]. Two different technologies were used in an effort to create a hydrophilic surface. The first was a permanent, ultrathin (25 nm), uniform, high refractive index plasma coating on the surface of the lenses (lotrafilcon A; Night and Day). The second method was a plasma oxidation surface treatment resulting in glassy silicate islands (balafilcon A; PureVision). The gap between the islands is so small that their wettability bridges the hydrophobic balafilcon. The flow of oxygen and fluids is not impeded by these surface treatments, however, studies have shown that these methods have only been partially effective, resulting in reduced wettability and increased deposition [28]. With the introduction of later SiH materials, the issue of wettability has been approached differently by using internal wetting agents such as polyvinylpyrolidone (PVP) (senofilcon A; Acuvue Oasys and galyfilcon A; Acuvue Advance), silicone macromers to produce a naturally wettable lens (comifilcon A; Biofinity) and a packaging solution containing a moisturising surfactant (polaxamine) designed to improve comfort on insertion (balafilcon A; PureVision). A daily disposable lens which has recently been released into the market in Europe (Dailies Total1) aims to solve the problem of SiH wettability. It is a water gradient SiH material with the water content increasing from 33% at the core of the lens to over 80% at the surface. Figure 7 Lenses dehydrating outdoors in a dehydration test at a recent CET meeting

8 Are SiH lens materials less prone to dehydration than conventional hydrogels? SiH materials are by their nature less wettable than conventional hydrogel lenses, however, lens manufacturers are striving to remedy this. SiH materials have a lower water content than conventional hydrogel lenses, which may possibly make them less prone to dehydration. At present the answer is equivocal and further investigation is necessary. There are a number of studies which suggest the biocompatible material omafilcon A (Proclear) dehydrates less than some SiH lenses, however, studies which involved refitting conventional hydrogel patients with SiH lenses have suggested they report reduced dryness [16] and longer wearing times [29]. Lens Depositition Deposition of tear film components on a contact lens surface causes a reduction in the overall performance of the lens and an increase in inflammatory responses (Figure 8). There are various factors which influence deposition on contact lens materials such as tear film composition, the ionicity of the material, water content, pore size and hydrophobicity. A large proportion of the protein in the tear film is lysozyme, and group IV materials (ionic, high water content) tend to attract this protein to a greater extent than the other groups. Group II (non-ionic, high water content) lens materials have a tendency to attract lipids from the tear film. Some SiH materials have been found to deposit considerably less protein and more lipid than conventional hydrogels, however, the pattern of deposition appears to depend on the formula of the material and whether it is surface treated [28, 30]. There is also the possibility that a higher percentage of denatured proteins accumulate on some SiH lenses as the protein is not able to penetrate the matrix of the lens due to surface treatment. This may contribute to the increase in CLPC which has been reported with some SiH materials [14]. Figure 8 Deposits on a soft contact lens Antimicrobial lens surfaces and materials SiH lens materials have reduced practitioners concerns regarding the supply of oxygen to the cornea, however, studies have shown the risk of infection to be similar for SiH lenses as it is for conventional hydrogels [17]. This risk is highest when lenses are worn on an extended wear basis and lowest for daily disposables. If this risk is to be reduced, further developments in contact lens materials are necessary. During contact lens wear bacteria adhere to biofilms which build up on the lens surface. Surface techniques are being developed with the aim of preventing biofilm formation and acting to kill the bacteria. Silver is an

9 antimicrobial agent with a low toxicity to human tissue and is already used to reduce microbial contamination in contact lens cases (AQuify, Ciba Vision). A glass powder is added to the case material during manufacture, and on contact with the moisture from the solution, silver ions with their antimicrobial properties are released. Clinical studies have shown there to be a significant reduction in lens contamination using this technology [31]. There are also a number of other technologies being investigated for their use with medical devices which could possibly be applied to contact lens materials [32]. These include polyquats, polymeric pyridinum compounds, free radical-producing agents, quorum-sensing blockers and antiinfective agents. This type of technology could revive interest in overnight wear of lenses as extended wear has the highest risk factor for corneal infection regardless of lens material. It would also be of benefit if applied to daily wear materials as contact lens patients are not always compliant with their care regimes and extra protection against microbial infection would be provided. An antimicrobial lens would need to be compatible with the ocular surface, not elicit an allergic response, be wettable and not attract deposits and be compatible with the lens care regime. Ease of manufacture and cost effective production must also be taken into consideration. Conclusion The developmentof silicone hydrogel contact lens materials, which reduced concerns regarding oxygen supply to the cornea during daily contact lens wear, was a huge step forward for the contact lens industry. A silicone containing material, however, does not necessarily produce a perfect lens or provide the best wearing experience for the patient. Properties such as visual quality, lubricity, modulus, wettability, de-hydration and antimicrobial surface treatment have now become the focus of research. Contact lens manufacturers are rising to the challenge and over the next few years practitioners will no doubt see a number of advances in material technologies to meet the needs of the wearer. 1. Bennett ES, Henry VA. Soft Material Selection/Fitting and Evaluation. In Clinical Manual of Contact Lenses, 2009, Lippincott, Williams & Wilkins: Philadelphia. 2. Kerr KR, Rushton D. ed. The ACLM Contact Lens Year Book Morgan PB et al. Central and Peripheral Oxygen Transmissibility Thresholds to Avoid Corneal Swelling During Open Eye Soft Contact Lens Wear. Journal of Biomedical Materials Research Part B-Applied Biomaterials, B(2): Papas E. On the relationship between soft contact lens oxygen transmissibility and induced limbal hyperaemia. Experimental Eye Research, (2): Benjamin WJ. Defining hypertransmissibility of a contact lens. Contact lens spectrum, August Ophthalmic Optics - Contact lenses - Part 1: Vocabulary, classification system and recommendations for labelling specifications, in BS EN ISO : Ophthalmics - Contact lenses - Standard terminology, tolerances, measurements and physiochemical properties, in ANSI Z Holden BA. et al. The minimum precorneal oxygen-tension to avoid corneal edema. Investigative Ophthalmology & Visual Science, (4): Brennan NA. Corneal oxygenation during contact lens wear: comparison of diffusion and EOP-based flux models. Clinical & experimental optometry : journal of the Australian Optometrical Association, (2): Holden BA, and Mertz GW. Critical oxygen levels to avoid corneal edema for daily and extended wear contact-lenses. Investigative Ophthalmology & Visual Science, (10):

10 11. Harvitt DM Bonanno JA. Re-evaluation of the oxygen diffusion model for predicting minimum contact lens Dk/t values needed to avoid corneal anoxia. Optometry and Vision Science, (10): Efron N. et al. Oxygen permeability and water content of silicone hydrogel contact lens materials. Optometry and Vision Science, (4): French K. Contact Lens Material Properties. Part 2 - Mechanical behaviour and modulus. The Optician, : Dumbleton K. Noninflammatory silicone hydrogel contact lens complications. Eye & contact lens, (1 Suppl):S186-9; discussion S190-1, S Ozkan JW. The Effect of Lens Modulus on Insertion Comfort with Silicone Hydrogel Lenses. in ARVO, Florida. 16. Dumbleton K. et al. Objective and subjective responses in patients refitted to daily-wear silicone hydrogel contact lenses. Optometry and Vision Science, (10): Stapleton F. et al. The incidence of contact lens-related microbial keratitis in Australia. Ophthalmology, (10): Efron N, Morgan PB. Is it the modulus or the method that matters? - Response. Clinical and Experimental Optometry, (6): French K. Contact Lens Material Properties. Part 1 - Wettability. The Optician, : Read ML. et al. Measurement Errors Related to Contact Angle Analysis of Hydrogel and Silicone Hydrogel Contact Lenses. Journal of Biomedical Materials Research Part B-Applied Biomaterials, B(2): Haddad M. et al. A Novel On-Eye Wettability Analyzer for Soft Contact Lenses. Optometry and Vision Science, (10):E1188-E Young G. et al. Soft Contact Lens-Related Dryness with and without Clinical Signs. Optometry and Vision Science, (8): Fonn D. Targeting contact lens induced dryness and discomfort: What properties will make lenses more comfortable. Optometry and Vision Science, (4): Sulley A. The Best CL for my Patient. The Optician, : Roba M. et al. Friction Measurements on Contact Lenses in Their Operating Environment. Tribology Letters, (3): Wolffsohn JS. et al. Objective clinical performance of comfort-enhanced daily disposable soft contact lenses. Contact Lens & Anterior Eye, (2): Tighe BJ. Silicone Hydrogels: structure, properties and behaviour. In Silicone Hydrogels: continuous-wear contact lenses, 2004, Butterworth-Heinemann: Oxford. 28. Boone A. et al. Ex Vivo Protein Deposition on Bi-Weekly Silicone Hydrogel Contact Lenses. Optometry and Vision Science, (11): Dumbleton KA. et al. Comfort and adaptation to silicone hydrogel lenses for daily wear. Eye & Contact Lens-Science and Clinical Practice, (4): Jones L. et al. Lysozyme and lipid deposition on silicone hydrogel contact lens materials. Eye & contact lens, (1 Suppl):S75-9; discussion S83-4, S Amos CF, George MD. Clinical and laboratory testing of a silver-impregnated lens case. Contact lens & anterior eye : the journal of the British Contact Lens Association, (5): Weisbarth RE. et al. Creating antimicrobial surfaces and materials for contact lenses and lens cases. Eye & contact lens, (6 Pt 2):

The ABCs of Fitting Soft Contact Lenses

The ABCs of Fitting Soft Contact Lenses The ABCs of Fitting Soft Contact Lenses What is a Soft Contact Lens? Pliable material Large diameter Excellent memory Bulk manufacturing Successful Soft Contact Lenses Top 3 considerations Wettability

More information

Contact Lens Wear during Prolonged Military Operations, Is it Safe and Effective or is Refractive Surgery a Better Option?

Contact Lens Wear during Prolonged Military Operations, Is it Safe and Effective or is Refractive Surgery a Better Option? Is it Safe and Effective or is Refractive Surgery a Better Option? Dr. A.J.P. Rouwen Col. Surg. Ministry of Defence MGFB Central Military Hospital PO Box 90000 3509 AA Utrecht THE NETHERLANDS A.P.J.Rouwen@mindef.nl

More information

The Truth About Silicone-hydrogel Contact Lenses

The Truth About Silicone-hydrogel Contact Lenses The Truth About Silicone-hydrogel Contact Lenses a production of studymyhealth.wordpress.com This packet is an effort of studymyhealth.wordpress.com. It is intended to provide a first-hand look at the

More information

CODE OF PRACTICE FOR THE SUPPLY OF CONTACT LENSES AND CONTACT LENS CARE PRODUCTS

CODE OF PRACTICE FOR THE SUPPLY OF CONTACT LENSES AND CONTACT LENS CARE PRODUCTS CODE OF PRACTICE FOR THE SUPPLY OF CONTACT LENSES AND CONTACT LENS CARE PRODUCTS Published by the NZAO; Wellington; May 2001 Reviewed March 2013 NZAO Code of Practice for the Supply of Contact Lenses and

More information

Hioxifilcon D SOFT CONTACT LENS FOR DAILY WEAR

Hioxifilcon D SOFT CONTACT LENS FOR DAILY WEAR 1 of 5 PACKAGE INSERT FOR Hioxifilcon D SOFT CONTACT LENS FOR DAILY WEAR CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A LICENSED PRACTITIONER Important: Please read carefully

More information

MOST FREQUENTLY ASKED CLIENT QUESTIONS

MOST FREQUENTLY ASKED CLIENT QUESTIONS 6/2 Victor Crescent Narre Warren VIC 3805 Ph: (03) 8794 8255 Fax: (03) 8794 8256 www.ortho-k.com.au Sleep your way to great vision MOST FREQUENTLY ASKED CLIENT QUESTIONS What is ortho-k? Ortho-k (Orthokeratology)

More information

Vision Glossary of Terms

Vision Glossary of Terms Vision Glossary of Terms EYE EXAMINATION PROCEDURES Eyeglass Examinations: The standard examination procedure for a patient who wants to wear eyeglasses includes at least the following: Case history; reason

More information

A Clinical Study of a Hydrogel Multifocal Contact Lens

A Clinical Study of a Hydrogel Multifocal Contact Lens A Clinical Study of a Hydrogel Multifocal Contact Lens By Peter Walker, BOptom TPA Background Despite the increasing number of presbyopic patients within the population, this group remain one of the greatest

More information

WHAT TO DO IF YOU HAVE DRY EYES...

WHAT TO DO IF YOU HAVE DRY EYES... WHAT TO DO IF YOU HAVE DRY EYES... Frank J. Holly, PhD Dry Eye Institute Yantis, Texas Based on a presentation given in Spanish at the patient session of the VIIIth Triennial Meeting of the International

More information

Course # Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism

Course # Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism Course # 772 Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism Intra Corneal Ring Segments Contact Lens Management of Irregular Astigmatism Financial Disclosure I do not own

More information

THE ART OF BIFOCAL CONTACT LENS FITTING Eef van der Worp BOptom PhD FAAO

THE ART OF BIFOCAL CONTACT LENS FITTING Eef van der Worp BOptom PhD FAAO Introduction THE ART OF BIFOCAL CONTACT LENS FITTING Eef van der Worp BOptom PhD FAAO Bifocal contact lenses have high potential in contact lens practice. The contact lens wearer who turns presbyopic is

More information

By Kimberly A. Millard, MS; Daniel Hook, PhD; Andrew Hoteling, PhD & Katarzyna Wygladacz, PhD

By Kimberly A. Millard, MS; Daniel Hook, PhD; Andrew Hoteling, PhD & Katarzyna Wygladacz, PhD A One-Step Hydrogen Peroxide-based Contact Lens Solution Ms. Millard is a senior research scientist in Vision Care at Bausch + Lomb, a division of Valeant Pharmaceuticals. She has more than 18 years of

More information

Dispersion of Synthetic Fibers in Wet-Lay Nonwovens

Dispersion of Synthetic Fibers in Wet-Lay Nonwovens Dispersion of Synthetic Fibers in Wet-Lay Nonwovens This article was written by James M. Keith, retired Director of Technical Services for MiniFIBERS, and published in the Tappi Journal, Vol. 77, No. 6,

More information

Corneal Collagen Cross-Linking (CXL) With Riboflavin

Corneal Collagen Cross-Linking (CXL) With Riboflavin Dr. Paul J. Dubord, MD, FRCSC Clinical Professor Department of Ophthalmology and Visual Sciences University of British Columbia Patient Information Guide Corneal Collagen Cross-Linking (CXL) With Riboflavin

More information

CHAPTER 1 INTRODUCTION

CHAPTER 1 INTRODUCTION CHAPTER 1 INTRODUCTION 1.1 Research Background Diabetes mellitus is a disease in which the body cannot produce sufficient insulin in their pancreas to adequately control the level of glucose in their blood

More information

MAke A difference in someone s life

MAke A difference in someone s life MAke A difference in someone s life Fitting guide En del av Multilens contents Introduction 1 Applications 2 characteristics 2 Fitting philosophy 3 Fitting process 4 Ideal fit 5 Comfort vs lens awarness

More information

Eye Injuries. The Eyes The eyes are sophisticated organs. They collect light and focus it on the back of the eye, allowing us to see.

Eye Injuries. The Eyes The eyes are sophisticated organs. They collect light and focus it on the back of the eye, allowing us to see. Eye Injuries Introduction The design of your face helps protect your eyes from injury. But injuries can still damage your eyes. Sometimes injuries are severe enough that you could lose your vision. Most

More information

Polymers in everyday things contact lenses

Polymers in everyday things contact lenses HOTOCOY olymers in everyday things contact lenses (Background information) olymers are a part of everyday life and examples can be found almost anywhere. Many people think of polymers simply as plastics

More information

Consent for LASIK (Laser In Situ Keratomileusis) Retreatment

Consent for LASIK (Laser In Situ Keratomileusis) Retreatment Consent for LASIK (Laser In Situ Keratomileusis) Retreatment Please read the following consent form very carefully. Please initial at the bottom of each page where indicated. Do not sign this form unless

More information

WOOD WEAR TESTING USING TRIBOMETER

WOOD WEAR TESTING USING TRIBOMETER WOOD WEAR TESTING USING TRIBOMETER Prepared by Duanjie Li, PhD 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials. 2015 NANOVEA INTRO

More information

1.1.2. thebiotutor. AS Biology OCR. Unit F211: Cells, Exchange & Transport. Module 1.2 Cell Membranes. Notes & Questions.

1.1.2. thebiotutor. AS Biology OCR. Unit F211: Cells, Exchange & Transport. Module 1.2 Cell Membranes. Notes & Questions. thebiotutor AS Biology OCR Unit F211: Cells, Exchange & Transport Module 1.2 Cell Membranes Notes & Questions Andy Todd 1 Outline the roles of membranes within cells and at the surface of cells. The main

More information

Basic Properties and Application of Auto Enamels

Basic Properties and Application of Auto Enamels Basic Properties and Application of Auto Enamels Composition of Ceramic Automotive Glass Enamels Ceramic automotive glass colours are glass enamels that fire on to the glass during the bending process

More information

ORTHOKERATOLOGY AND THE OPHTHALMIC ASSISTANT Elliott M. Rosengarten, OD

ORTHOKERATOLOGY AND THE OPHTHALMIC ASSISTANT Elliott M. Rosengarten, OD ORTHOKERATOLOGY AND THE OPHTHALMIC ASSISTANT Elliott M. Rosengarten, OD This article and accompanying quiz are worth.5 JCAHPO Group A continuing education credit. CONTINUING EDUCATION CREDITS ARE SUBJECT

More information

Advanced Ortho-K Technology. Proudly brought to you by. Australian Contact Lenses

Advanced Ortho-K Technology. Proudly brought to you by. Australian Contact Lenses Proudly brought to you by Australian Contact Lenses Advanced Ortho-K Technology Are your patients looking for crystal-clear vision without surgery? Are you looking to grow your practice and enhance patient

More information

Puzio Eye Care Associates

Puzio Eye Care Associates Puzio Eye Care Associates DENNIS HARWICH 38, Route 134 Route 137 & Route 39, Suite 119B P.O. Box 1412 P.O. Box 1661 South Dennis MA 02660 East Harwich, MA 02645 Voice: 508-394-221 Voice: 508-432-3444 Fax:

More information

Figure 1. A schematic representation of a PEO coatings: mushroom conformation on the left hand side and brush conformation on the right hand side.

Figure 1. A schematic representation of a PEO coatings: mushroom conformation on the left hand side and brush conformation on the right hand side. 7 General discussion Chapter 7 Microbial adhesion onto the surface of medical implants, like hip prostheses and urinary catheters, followed by biofilm formation, often leads to infections on these implants.

More information

PARAGON CRT CERTIFICATION TRAINING INFORMATION

PARAGON CRT CERTIFICATION TRAINING INFORMATION PARAGON CRT CERTIFICATION TRAINING INFORMATION Thank you for choosing Paragon CRT, the world leader in contact lens corneal reshaping. By reviewing the basic fitting dynamics of Paragon CRT and taking

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

How To Know If You Can See Without Glasses Or Contact Lense After Lasik

How To Know If You Can See Without Glasses Or Contact Lense After Lasik The LASIK experience I WHO CAN HAVE LASIK? To be eligible for LASIK you should be at least 21 years of age, have healthy eyes and be in good general health. Your vision should not have deteriorated significantly

More information

ALTERNATIVES TO LASIK

ALTERNATIVES TO LASIK EYE PHYSICIANS OF NORTH HOUSTON 845 FM 1960 WEST, SUITE 101, Houston, TX 77090 Office: 281 893 1760 Fax: 281 893 4037 INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) INTRODUCTION This information

More information

Paragon RG-4 Fitting Reference Guide

Paragon RG-4 Fitting Reference Guide PARAGON VISION SCIENCES Paragon RG-4 Fitting Reference Guide* & Certification Manual 2008 *This document is an abbreviated version and does not replace the FDA approved professional fitting guide. Lenses

More information

Biomaterials in tissue engineering

Biomaterials in tissue engineering Biomaterials in tissue engineering S. Swaminathan Director Centre for Nanotechnology & Advanced Biomaterials School of Chemical & Biotechnology SASTRA University Thanjavur 613 401 Tamil Nadu Page 1 of

More information

Make a difference in someone s life FITTING GUIDE

Make a difference in someone s life FITTING GUIDE Make a difference in someone s life FITTING GUIDE CONTENTS Introduction 3 Applications 4 Characteristics 4 Fitting philosophy 5 Mid-peripheral / limbal zone 6 Only 3 fitting variables to determine 7 Ideal

More information

LASER VISION CORRECTION SURGERY A GUIDE FOR PATIENTS. Professional care for your eye health

LASER VISION CORRECTION SURGERY A GUIDE FOR PATIENTS. Professional care for your eye health LASER VISION CORRECTION SURGERY A GUIDE FOR PATIENTS Professional care for your eye health Contents About Dr John Males... 1 COMMON QUESTIONS How does an eye work?... 2 What is Myopia (short sightedness)...

More information

Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology

Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology Eye Associates Custom LASIK With IntraLASIK Correction Of Nearsightedness, Farsightedness, and Astigmatism Using IntraLase TM Technology INDICATIONS AND PROCEDURE This information is being provided to

More information

IVALON Ophthalmic. IVALON Ophthalmic Products

IVALON Ophthalmic. IVALON Ophthalmic Products IVALON Ophthalmic IVALON Ophthalmic Products IVALON Ophthalmic products are constructed using a highly absorbent, unique hydroxylated polyvinyl acetal (PVAc) sponge. Several densities of sponge are available

More information

Chemical Basis of Life Module A Anchor 2

Chemical Basis of Life Module A Anchor 2 Chemical Basis of Life Module A Anchor 2 Key Concepts: - Water is a polar molecule. Therefore, it is able to form multiple hydrogen bonds, which account for many of its special properties. - Water s polarity

More information

Contact lenses. DRAFT VERSION Not for distribution

Contact lenses. DRAFT VERSION Not for distribution Contact lenses Overview Most people who need to wear glasses can wear contact lenses. There are two main types of contact lenses: soft lenses which are floppy and mould to the shape of your eye rigid gas

More information

Contact Lens Fitting. Soft toric contact lens fitting was previously regarded as a speciality only to be undertaken by experienced practitioners.

Contact Lens Fitting. Soft toric contact lens fitting was previously regarded as a speciality only to be undertaken by experienced practitioners. Essential Contact Lens Practice Medical Ltd Soft Toric Contact Lens Fitting key points Key Points Improvements in toric lens speed of orientation, lens stability, corneal oxygen supply and reproducibility

More information

Power Diameter Base curve SOFT TORIC CONTACT LENS PROF.DR.SAMIHA ABOULMAGD 1

Power Diameter Base curve SOFT TORIC CONTACT LENS PROF.DR.SAMIHA ABOULMAGD 1 Fitting Soft Contact Lens Prof. Dr. Samiha Aboulmagd MD Head of fophthalmology l Department Al-Azhar University 2010 1 Fitting of contact lens Power Diameter Base curve 2 PROF.DR.SAMIHA ABOULMAGD 1 Fitting

More information

Designed for comfort and built to last.

Designed for comfort and built to last. Designed for comfort and built to last. Inter IKEA Systems B.V. 2015 Whether we work in an office or at home, many of us spend the majority of the day at a desk. A comfortable workspace is therefore essential

More information

ORTHOKERATOLOGY CONTACT LENSES

ORTHOKERATOLOGY CONTACT LENSES ORTHOKERATOLOGY CONTACT LENSES A Guide to Trouble-Free Contact lens Wear (Instructions for Use) Consult an Eye Care Professional for suitability of wear www.oculusgp.com Page 1 of 9 Product Type Contents

More information

Biological cell membranes

Biological cell membranes Unit 14: Cell biology. 14 2 Biological cell membranes The cell surface membrane surrounds the cell and acts as a barrier between the cell s contents and the environment. The cell membrane has multiple

More information

INFORMED CONSENT FOR PHAKIC IMPLANT SURGERY

INFORMED CONSENT FOR PHAKIC IMPLANT SURGERY INFORMED CONSENT FOR PHAKIC IMPLANT SURGERY INTRODUCTION This information is being provided to you so that you can make an informed decision about having eye surgery to reduce or eliminate your nearsightedness.

More information

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant INTRODUCTION Except for unusual situations, a cataract operation is indicated only when you cannot function satisfactorily

More information

That s a BIG lens!! Scleral Contact Lenses. Average Corneal diameter = 11.5mm. Size matters. Soft Contact Lens 9/24/2014. Average RGP lens = 9.

That s a BIG lens!! Scleral Contact Lenses. Average Corneal diameter = 11.5mm. Size matters. Soft Contact Lens 9/24/2014. Average RGP lens = 9. That s a BIG lens!! Scleral Contact Lenses MOA Convention 10.4.2014 Chris DeRose, OD Size matters Average RGP lens = 9.5 mm Average Corneal diameter = 11.5mm Soft Contact lens = 14.0 to 14.5mm Mini scleral

More information

Engine Bearing Materials

Engine Bearing Materials Engine Bearing Materials Dr. Dmitri Kopeliovich (Research & Development Manager) The durable operation of an engine bearing is achieved if its materials combine high strength (load capacity, wear resistance,

More information

Soft Contact Lens Seeing a brighter future

Soft Contact Lens Seeing a brighter future Soft Contact Lens Seeing a brighter future Dane Callow Javier Garcia Teruel Ziv Gelfer Brandon Lee Yoko Warita Yaron Zimmerman Soft Contact Lens Seeing a brighter future 1 Contents 1 Introduction... 3

More information

IntraLase and LASIK: Risks and Complications

IntraLase and LASIK: Risks and Complications No surgery is without risks and possible complications and LASIK is no different in that respect. At Trusted LASIK Surgeons, we believe patients can minimize these risks by selecting a highly qualified

More information

LASIK EPILASIK FEMTOSECOND LASER. Advantages

LASIK EPILASIK FEMTOSECOND LASER. Advantages LASIK EPILASIK FEMTOSECOND LASER Advantages There are many advantages to having laser vision correction. Laser vision correction gives most patients the freedom to enjoy their normal daily activities without

More information

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK)

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) Lasik Center 2445 Broadway Quincy, IL 62301 217-222-8800 INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) INTRODUCTION This information is being provided to you so that you can make an informed

More information

Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.

Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery. Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction Scout Vessel Guard 2

More information

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE TECHNOLOGY

INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE TECHNOLOGY EYE PHYSICIANS OF NORTH HOUSTON 845 FM 1960 WEST, SUITE 101, Houston, TX 77090 Office: 281 893 1760 Fax: 281 893 4037 INFORMED CONSENT FOR LASER IN-SITU KERATOMILEUSIS (LASIK) USING INTRALASE TM BLADE-FREE

More information

Plasma Electronic is Partner of. Tailor-Made Surfaces by Plasma Technology

Plasma Electronic is Partner of. Tailor-Made Surfaces by Plasma Technology Precision Fair 2013 Stand 171 Plasma Electronic is Partner of Tailor-Made Surfaces by Plasma Technology Dr. J. Geng, Plasma Electronic GmbH Modern Surface Technology in 1900 Overview A short introduction

More information

PRACTITIONER S FITTING GUIDE. Keratoconus Irregular Cornea Post Graft

PRACTITIONER S FITTING GUIDE. Keratoconus Irregular Cornea Post Graft PRACTITIONER S FITTING GUIDE Keratoconus Irregular Cornea Post Graft Four lens designs to fit all corneal shapes... One simple systematic approach to fitting Featuring Easy-to-fit using a simple systematic

More information

Fig. 1. Background. Name: Class: Date:

Fig. 1. Background. Name: Class: Date: Background Bubbles make a great stand in for cell membranes. They re fluid, flexible, and can self-repair. Bubbles and cell membranes are alike because their parts are so similar. If you could zoom down

More information

Scout Vessel Guard. A hydrogel membrane with exceptional handling characteristics.

Scout Vessel Guard. A hydrogel membrane with exceptional handling characteristics. . A hydrogel membrane with exceptional handling characteristics. Pliable Permanent Pore size < 0.08μm Instruments and implants approved by the AO Foundation . A hydrogel membrane with exceptional handling

More information

It is an important tool to assess factors that affect the bioavailability of a drug from a solid preparartion.

It is an important tool to assess factors that affect the bioavailability of a drug from a solid preparartion. Quality control of tablets Dissolution It is an important tool to assess factors that affect the bioavailability of a drug from a solid preparartion. To ensure that the preparation comply with product

More information

A Compendium of Clinical and Laboratory Studies Multi-Purpose Disinfecting Solutions (MPDS)

A Compendium of Clinical and Laboratory Studies Multi-Purpose Disinfecting Solutions (MPDS) A Compendium of Clinical and Laboratory Studies Multi-Purpose Disinfecting Solutions () Contact Lens Case Contamination Disinfection and FDA/ISO Standards Disinfection and Clinical Isolates Acanthamoeba

More information

INFORMED CONSENT FOR PHAKIC LENS IMPLANT SURGERY

INFORMED CONSENT FOR PHAKIC LENS IMPLANT SURGERY INTRODUCTION INFORMED CONSENT FOR PHAKIC LENS IMPLANT SURGERY This information is being provided to you so that you can make an informed decision about having eye surgery to reduce or eliminate your nearsightedness.

More information

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM 1 BoydVision TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM Risks and Side Effects... 2 Risks Specific to PRK... 3 Risks Specific to LASIK... 4 Patient Statement of Consent... 5 Consent for Laser Eye

More information

BIOLOGICAL MEMBRANES: FUNCTIONS, STRUCTURES & TRANSPORT

BIOLOGICAL MEMBRANES: FUNCTIONS, STRUCTURES & TRANSPORT BIOLOGICAL MEMBRANES: FUNCTIONS, STRUCTURES & TRANSPORT UNIVERSITY OF PNG SCHOOL OF MEDICINE AND HEALTH SCIENCES DISCIPLINE OF BIOCHEMISTRY AND MOLECULAR BIOLOGY BMLS II / B Pharm II / BDS II VJ Temple

More information

Fitting Guide. Dallas 800-366-3933 Denver 800-362-4233 Tulsa 800-685-5367

Fitting Guide. Dallas 800-366-3933 Denver 800-362-4233 Tulsa 800-685-5367 Fitting Guide Dallas 800-366-3933 Denver 800-362-4233 Tulsa 800-685-5367 Sagittal Depth Fitting the Jupiter lens design is based upon sagittal depth. Relative sagittal depth is not just the steepness of

More information

FITTING & PATIENT MANAGEMENT GUIDE. For ACUVUE Brand BIFOCAL (etafilcon A) Contact Lenses Visibility Tinted With UV Blocker

FITTING & PATIENT MANAGEMENT GUIDE. For ACUVUE Brand BIFOCAL (etafilcon A) Contact Lenses Visibility Tinted With UV Blocker FITTING & PATIENT MANAGEMENT GUIDE For ACUVUE Brand BIFOCAL (etafilcon A) Contact Lenses Visibility Tinted With UV Blocker PRESCRIBING QUESTIONS? CALL THE VISTAKON, DIVISION OF JOHNSON & JOHNSON VISION

More information

Ch 2 Properties of Fluids - II. Ideal Fluids. Real Fluids. Viscosity (1) Viscosity (3) Viscosity (2)

Ch 2 Properties of Fluids - II. Ideal Fluids. Real Fluids. Viscosity (1) Viscosity (3) Viscosity (2) Ch 2 Properties of Fluids - II Ideal Fluids 1 Prepared for CEE 3500 CEE Fluid Mechanics by Gilberto E. Urroz, August 2005 2 Ideal fluid: a fluid with no friction Also referred to as an inviscid (zero viscosity)

More information

LAB 24 Transpiration

LAB 24 Transpiration Name: AP Biology Lab 24 LAB 24 Transpiration Objectives: To understand how water moves from roots to leaves in terms of the physical/chemical properties of water and the forces provided by differences

More information

STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER. Static COF 0.00 0.0 0.5 1.0 1.5 2.0. Time(min) Prepared by Duanjie Li, PhD

STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER. Static COF 0.00 0.0 0.5 1.0 1.5 2.0. Time(min) Prepared by Duanjie Li, PhD STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER 0.20 Static COF Coefficient of Friction 0.15 0.10 0.05 0.00 0.0 0.5 1.0 1.5 2.0 Time(min) Prepared by Duanjie Li, PhD 6 Morgan, Ste156, Irvine

More information

Orthokeratology F.A.Q.!

Orthokeratology F.A.Q.! Orthokeratology F.A.Q. Why orthokeratology? There are many reasons why people choose to correct their vision with orthokeratology. The main reason is the freedom from glasses and contact lenses. Orthokeratology

More information

Optical Customisation of Touch Screens and Display Front Surfaces

Optical Customisation of Touch Screens and Display Front Surfaces Optical Customisation of Touch Screens and Display Front Surfaces Society for Information Display March 2013 Rob Bennett Europe Summary Performance characteristics that can be altered : contrast and readability

More information

SMILE SURGERY A GUIDE FOR PATIENTS. Professional care for your eye health

SMILE SURGERY A GUIDE FOR PATIENTS. Professional care for your eye health SMILE SURGERY A GUIDE FOR PATIENTS Professional care for your eye health Contents About Dr John Males... 1 Our Commitment to Our Patients... 2 COMMON QUESTIONS How Does an Eye Work?... 3 What is Myopia

More information

Automotive Base Oil Presentation

Automotive Base Oil Presentation Automotive Base Oil Presentation What is a Base Oil? The refined petroleum mineral or synthetic material that is produced by a refinery to a required set of specifications. A lubricant s quality can depend

More information

WEARER S INSTRUCTION GUIDE

WEARER S INSTRUCTION GUIDE WEARER S INSTRUCTION GUIDE FOR THE EXTREME H 2 O MONTHLY (hioxifilcon D) Soft Contact Lens for Daily Wear CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A LICENSED PRACTITIONER

More information

MAZAHERI LASIK METHOD FOR VISUAL ENHANCEMENT TECHNICAL FIELD OF THE INVENTION. [0001] The present invention is directed, in general, to

MAZAHERI LASIK METHOD FOR VISUAL ENHANCEMENT TECHNICAL FIELD OF THE INVENTION. [0001] The present invention is directed, in general, to MAZAHERI LASIK METHOD FOR VISUAL ENHANCEMENT TECHNICAL FIELD OF THE INVENTION [0001] The present invention is directed, in general, to a surgical procedure and, more particularly, to surgical procedure

More information

...You Need to know about

...You Need to know about What......You Need to know about LASIK Our Eyes Eyes are the windows to our world. They are so important to us that for many years we have looked for better ways to fix visual problems and improve our

More information

Preferential and Non-Preferential Reasons for Opting Contact Lens

Preferential and Non-Preferential Reasons for Opting Contact Lens Preferential and Non-Preferential Reasons for Opting Contact Lens Ravipati Sarath 1, *Jannitha PA 1 and Pavan KV 2 1 School of Medical Sciences, University of Hyderabad, Hyderabad, Andhra Pradesh, India

More information

Fundamental Shift in Use of Vision. Fundamental Shift in Use of Vision. Visual Fatigue Syndrome. Widespread Computer Use. Work, Home, School:

Fundamental Shift in Use of Vision. Fundamental Shift in Use of Vision. Visual Fatigue Syndrome. Widespread Computer Use. Work, Home, School: Visual Fatigue Syndrome Fundamental Shift in Use of Vision The Price of Today s Near and Intermediate World Past: Primarily used Distance vision Today: Primarily use Intermediate/Near Dr. Kris Kerestan

More information

Fluid Mechanics: Static s Kinematics Dynamics Fluid

Fluid Mechanics: Static s Kinematics Dynamics Fluid Fluid Mechanics: Fluid mechanics may be defined as that branch of engineering science that deals with the behavior of fluid under the condition of rest and motion Fluid mechanics may be divided into three

More information

Description of mechanical properties

Description of mechanical properties ArcelorMittal Europe Flat Products Description of mechanical properties Introduction Mechanical properties are governed by the basic concepts of elasticity, plasticity and toughness. Elasticity is the

More information

Life just looks better without reading glasses

Life just looks better without reading glasses Life just looks better without reading glasses TM Enjoy life without depending on reading glasses. The KAMRA inlay treatment: Restores everyday vision so you can see text messages, a computer screen and

More information

The Clinic You see - at Sjonlag Eye Center, we care about your eyes

The Clinic You see - at Sjonlag Eye Center, we care about your eyes The Clinic You see - at Sjonlag Eye Center, we care about your eyes Welcome to Sjonlag Eye Center, a total refractive surgery center. Established in 2001, the Center has grown into a fully staffed eye

More information

(A5) Brakes Sample Questions and Answers

(A5) Brakes Sample Questions and Answers (A5) Brakes Sample Questions and Answers Answers to these questions are found beginning on page 4 of this document 1. Two technicians are discussing a problem where the brake pedal travels too far before

More information

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK?

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK? 2014 2015 LASIK What is LASIK? LASIK (laser in situ keratomileusis) is a type of refractive surgery. This kind of surgery uses a laser to treat vision problems caused by refractive errors. You have a refractive

More information

PATIENT INSTRUCTION GUIDE

PATIENT INSTRUCTION GUIDE PATIENT INSTRUCTION GUIDE 1-DAY ACUVUE TruEye Brand Contact Lenses with HYDRACLEAR 1 Technology (narafilcon A) Visibility Tinted with UV Blocker for Daily Wear Single Use Only TABLE OF CONTENTS Introduction...

More information

FPO. Coefficient of Friction New Method, New Requirements - Introducing the DCOF AcuTest SM. Katelyn Simpson

FPO. Coefficient of Friction New Method, New Requirements - Introducing the DCOF AcuTest SM. Katelyn Simpson A major change is underway regarding the coefficient of friction for tile, the frequently cited and specified measurement of a tile s frictional resistance and the parameter most closely related to traction

More information

This FDA approved procedure is available from Dr. Russell and is performed at the owder prings and arie a ye linic ain o ce loca ons.

This FDA approved procedure is available from Dr. Russell and is performed at the owder prings and arie a ye linic ain o ce loca ons. Gregg Russell, OD, FAAO Dr. Russell and arie a ye linic o er the latest technology to correct your nearsightedness and reduce your dependence on contact lenses or glasses without having to undergo eye

More information

PATIENT INSTRUCTION GUIDE

PATIENT INSTRUCTION GUIDE PATIENT INSTRUCTION GUIDE DISPOSABLE & FREQUENT REPLACEMENT ACUVUE OASYS Brand Contact Lenses with HYDRACLEAR Plus ACUVUE OASYS Brand Contact Lenses for ASTIGMATISM with HYDRACLEAR Plus ACUVUE OASYS Brand

More information

Cylinder Head Gasket Contact Pressure Simulation for a Hermetic Compressor

Cylinder Head Gasket Contact Pressure Simulation for a Hermetic Compressor Purdue University Purdue e-pubs International Compressor Engineering Conference School of Mechanical Engineering 2006 Cylinder Head Gasket Contact Pressure Simulation for a Hermetic Compressor Pavan P.

More information

Managing Irregular Astigmatism

Managing Irregular Astigmatism Non-scleral Lens Designs for Irregular Corneas Michael A. Ward, MMSc, FAAO, FCLSA Emory University Jessica Bezner Mannas, OD Koffler Vision Group Michael A Ward Potential COI Disclosures previous 12 months

More information

COLUMBIA UNIVERSITY HEALTH SCIENCES DIVISION

COLUMBIA UNIVERSITY HEALTH SCIENCES DIVISION COLUMBIA UNIVERSITY HEALTH SCIENCES DIVISION ENVIRONMENTAL HEALTH & SAFETY Working Safely with Ultraviolet Radiation Policy and Procedures Purpose: To provide information and guidelines for the safe use

More information

Materials & Solutions. Product Guide

Materials & Solutions. Product Guide Materials & Solutions Product Guide Materials & Solutions Product Guide Mission Statement Through constant innovation and continual improvement of our business processes, we will strive to be the worldwide

More information

Visibility Tinted Contact Lenses PATIENT INFORMATION BOOKLET FOR TRADITIONAL OR FREQUENT/PLANNED REPLACEMENT WEAR SL7489 8023203

Visibility Tinted Contact Lenses PATIENT INFORMATION BOOKLET FOR TRADITIONAL OR FREQUENT/PLANNED REPLACEMENT WEAR SL7489 8023203 38 Visibility Tinted Contact Lenses PATIENT INFORMATION BOOKLET FOR TRADITIONAL OR FREQUENT/PLANNED REPLACEMENT WEAR CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a

More information

Laser sintering of greens compacts of MoSi 2

Laser sintering of greens compacts of MoSi 2 Laser sintering of greens compacts of MoSi 2 G. de Vasconcelos 1, R. Cesar Maia 2, C.A.A.Cairo 3, R. Riva 2, N.A.S.Rodrigues 2, F.C.L.Mello 3 Instituto de Estudos Avançados 1, Instituto Tecnológico de

More information

CHAPTER 5.1 5.2: Plasma Membrane Structure

CHAPTER 5.1 5.2: Plasma Membrane Structure CHAPTER 5.1 5.2: Plasma Membrane Structure 1. Describe the structure of a phospholipid molecule. Be sure to describe their behavior in relationship to water. 2. What happens when a collection of phospholipids

More information

Naue GmbH&Co.KG. Quality Control and. Quality Assurance. Manual. For Geomembranes

Naue GmbH&Co.KG. Quality Control and. Quality Assurance. Manual. For Geomembranes Naue GmbH&Co.KG Quality Control and Quality Assurance Manual For Geomembranes July 2004 V.O TABLE OF CONTENTS 1. Introduction 2. Quality Assurance and Control 2.1 General 2.2 Quality management acc. to

More information

and LUMINOUS CHEMICAL VAPOR DEPOSITION INTERFACE ENGINEERING HirotsuguYasuda University of Missouri-Columbia Columbia, Missouri, U.S.A.

and LUMINOUS CHEMICAL VAPOR DEPOSITION INTERFACE ENGINEERING HirotsuguYasuda University of Missouri-Columbia Columbia, Missouri, U.S.A. LUMINOUS CHEMICAL VAPOR DEPOSITION and INTERFACE ENGINEERING HirotsuguYasuda University of Missouri-Columbia Columbia, Missouri, U.S.A. MARCEL MARCEL DEKKER. NEW YORK DEKKER Contents Preface iii Part I.

More information

Chapter 8. Movement across the Cell Membrane. AP Biology

Chapter 8. Movement across the Cell Membrane. AP Biology Chapter 8. Movement across the Cell Membrane More than just a barrier Expanding our view of cell membrane beyond just a phospholipid bilayer barrier phospholipids plus Fluid Mosaic Model In 1972, S.J.

More information

ATOMS AND BONDS. Bonds

ATOMS AND BONDS. Bonds ATOMS AND BONDS Atoms of elements are the simplest units of organization in the natural world. Atoms consist of protons (positive charge), neutrons (neutral charge) and electrons (negative charge). The

More information

Class Im MD A novel device for Automated Analysis of Tear Film Dynamics. Michael Ring, DI(FH) Upper Austria University of Applied Sciences

Class Im MD A novel device for Automated Analysis of Tear Film Dynamics. Michael Ring, DI(FH) Upper Austria University of Applied Sciences Class Im MD A novel device for Automated Analysis of Tear Film Dynamics Michael Ring, DI(FH) Upper Austria University of Applied Sciences Aim of the Research Develop a method to give information on the

More information

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

Geometric Optics Converging Lenses and Mirrors Physics Lab IV Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The

More information

Excimer Laser Refractive Surgery

Excimer Laser Refractive Surgery Excimer Laser Refractive Surgery In the field of ophthalmology has achieved great technological advances and, undoubtedly, the most representative have focused on refractive surgery, which aims to eliminate

More information