Performance and advantages of qnmr measurements

Similar documents
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

4.2 Bias, Standards and Standardization

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials

Analytical Standards and CRMs for the Petroleum Industry

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Validation and Calibration. Definitions and Terminology

Guidance for Industry

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Alignment and Preprocessing for Data Analysis

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

ALACC Frequently Asked Questions (FAQs)

Indoor Air Comfort Test Report

Pesticide Analysis by Mass Spectrometry

Annex 3. General guidelines for the establishment, maintenance and distribution of chemical reference substances. Introduction.

PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS

Certifying Reference Materials to ISO Guide 34

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

IMPURITIES IN NEW DRUG PRODUCTS

OLIVÉR BÁNHIDI 1. Introduction

Stimuli to the Revision Process

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Working with ICH Quality Guidelines - the Canadian Perspective

ph: Measurement and Uses

Exp 13 Volumetric Analysis: Acid-Base titration

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

C 5. chemical development contract research custom synthesis cgmp API manufacturing commercial production. Welcome to

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications

Trans Fats. What is a trans fat? Trans fatty acids, or trans fats as they are known, are certain

SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT

Proton Nuclear Magnetic Resonance Spectroscopy

Qualitative NIR Analysis for Ingredients in the Baking Industry

A commitment to quality and continuous improvement

Waters Core Chromatography Training (2 Days)

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Calibration Service Selection Guide

Melting Range 1 Experiment 2

There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.

How To Test For Contamination In Large Volume Water

PHARMACEUTICAL REFERENCE STANDARDS

Contents. Contributors. 1 Quality control and regulation 1 C.J. MOORES

Supporting Information

Dissolved and precipitated oxalate

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Carolina s Solution Preparation Manual

TRACEABILITY. Commercially available single-element solutions are used. of Single-Element Calibration Solutions

Procedure for Equipment Calibration and Maintenance

Optimizing the Post-implementation Monitoring of a LC-MS/MS Method

LC-MS/MS, the new reference method for mycotoxin analysis

13C NMR Spectroscopy

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER , CAS NUMBER ] IRON ORE PELLETS

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

1 Quality Assurance and Quality Control Project Plan

LC-MS/MS for Chromatographers

NMR Chemical Shifts of Common Laboratory Solvents as Trace Impurities

CHE334 Identification of an Unknown Compound By NMR/IR/MS

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.

Quality. Now Certified to ISO 9001:2008

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

EL724. Biodegradable Resin Products [EL /2/ ]

1 st day Basic Training Course

Observations on the determination of gold in gold alloys cupellation method (Fire Assay)

LifeTein in Industrial Production of Therapeutic Peptides. Phil Moore, PhD Director of Business Development LifeTein LLC, NJ, USA

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

Performing Calculatons

AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

Lecture #7 (2D NMR) Utility of Resonance Assignments

Nuclear Magnetic Resonance Spectroscopy

and Drug Administration/Office of Regulatory Affairs (FDA/ORA) laboratories.

NEW CHEMICAL ENTITIES

NMR Spectroscopy in Notre Dame

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

Background Information

PURIFICATION TECHNIQUES

Volume 22 Number 22. C43-A ISBN ISSN Gas Chromatography/Mass Spectrometry (GC/MS) Confirmation of Drugs; Approved Guideline

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

Experis Gas Mixtures. When accuracy, stability and reliability are a must

Analyzing Small Molecules by EI and GC-MS. July 2014

Advanced Solutions for Gas Monitoring

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

A Complete Solution for Method Linearity in HPLC and UHPLC

P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Annex 1 WHO good practices for pharmaceutical quality control laboratories

Transcription:

Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended for use as internal standards in qnmr, while in the next issue of Analytix we will present a first series of certified organic standards for chromatography The quantification of content or purity of organic substances has up to now often been performed with chromatographic techniques such as HPLC or GC. Over the last years, quantitative NMR (qnmr) has evolved not only in the pharmaceutical industry but also in many other fields. In the R&D lab of Sigma-Aldrich Buchs (Switzerland) a new Bruker Avance III 600 MHz NMR spectrometer was installed recently for high-resolution qnmr measurements of organic substances. By the end of 2009, the new NMR lab was fully accredited under both ISO/IEC 17025 [1] and also ISO Guide 34 [2] for the certification of organic reference materials using 1HqNMR. The combination of both accreditations represents the highest achievable reliability level and earns the label of gold standard among CRM producers. A set of calibration standards for use as internal standards in qnmr is being launched as a fi rst series of new CRMs. These CRMs are comprehensively characterised by qnmr and other techniques, leading to the highest accuracy and very low uncertainties of the certified values. All of these qnmr reference materials are traceable to internationally accepted references from NIST and a detailed certifi cate is delivered with each CRM. Performance and advantages of qnmr measurements Quantitative NMR shows many advantages over other analytical techniques with regard to quantifi - cation or purity determination of organic substances. The most outstanding attribute of qnmr is that it is a relative primary method: the signal intensity is in direct proportionality with the number of protons contributing to the resonance. The structures of the chemical substances are therefore irrelevant. In addition, no significant empirical factors or unknown biases contribute to the ratio of signals. The signal ratio of two diff erent protons can therefore be measured with tremendous precision and the only signifi - cant contribution to the measurement uncertainty is the integration of the signals. In other words, the direct response of a qnmr experiment is of the greatest accuracy. Of course, a few basic requirements must be met in order to obtain good qnmr results. To begin with, the weighing of the reference and the sample prior to the qnmr measurements must be done with utmost accuracy, since this is the essential sample preparation step in qnmr analysis. It is selfevident that the availability of a suitable reference material for internal calibration is mandatory (see next section). Furthermore, the reference substance and the sample must not react with each other or with the solvent. The relevant signals which are selected for the measurement must be clearly separated from each other and also from other signals. Appropriate instrument settings are required so that no intensity is lost through incomplete relaxation. Requirements to qnmr calibration standards

Since NMR signal intensity is fully independent of the nature of the substance, many organic substances could potentially serve as an internal reference for qnmr. In reality, the number of suitable candidates for qnmr references is rather limited, since they must conform to a series of the following specifi c requirements: only substances with a very limited number of signals and of highest organic purity are suitable as a qnmr reference. The candidates must be stable in solution and should not be chemically reactive. They should not contain residual water, since the presence of water can lead to line broadening or baseline distortion in many cases. Of course, to ensure proper weighing results, a solid qnmr reference should not be hygroscopic and a liquid qnmr reference should not be volatile. The number of isochronic proton nuclei be optimum, that is, neither too high nor too low in relation to the molecular mass of the qnmr reference substance, since a 1:1 signal intensity ratio of sample and internal standard is targeted. Otherwise, the amount of substance needed for one measurement would be either very low (leading to bad weighing result) or too high (causing solubility problems). The signals of a qnmr reference should not cover areas in the spectrum where most often the analyte signals are expected to be. Finally, a qnmr reference substance should be neither extremely toxic nor carcinogenic or mutagenic. It is obvious that many different qnmr standards must be available to cover the broad variety of analytes and deuterated solvents. We have developed a series of qnmr reference substances whereby every reference substance has its individual solubility behaviour and chemical shift. The selection of new qnmr calibration standards is given in Figure 1 (water soluble standards) and Figure 2 (standards soluble in apolar organic solvents).

Figure 1 1 H NMR spectra of water soluble qnmr standards. The residual signal of HDO is partly suppressed

Figure 2 1 H NMR spectra of qnmr standards which are soluble in organic solvents Certification of calibration standards for qnmr To achieve all the above-mentioned requirements, the new class of certifi ed qnmr reference substances undergo a series of pre-tests and extensive certifi cation measurements in order to provide compliance with ISO/IEC 17025 and ISO Guide 34 requirements. Only high-purity materials are carefully sourced or purifi ed and extensively characterised by chromatography, LC-MS and other standard techniques before they undergo the qnmr certifi cation procedure. Pre-testing: the substance is tested for hygroscopy or volatility, stability in different deuterated solvents, and reactivity against NIST certified reference substances (>24 hour period). Stability testing: Replicate tests after storage at higher temperature are performed to ensure that even extreme transportation conditions do not affect the certified value (>1 month period). Data on ageing studies at the recommended storage temperature are used to define the shelf life of the products (>3 year period). Signal suitability test: Since only pure signals are allowed for NMR quantification (no overlapping and no underlying impurity signals), extensive 2D experiments are required (H- H COSY and C-H correlated experiments in some cases). With H-H COSY experiments,

hidden signals from impurities can easily be detected with a limit of detection below 0.1% (see example in Figure 3). This suitability check is performed before the certification measurements are started. Homogeneity and traceability: 10 samples are taken from different places within the batch. The NMR samples are then prepared by weighing equivalents of the qnmr substance and the NIST reference together in one vial, using an ultramicro balance. The samples are dissolved, measured and analysed further under standardised measurement settings on a Bruker Avance III 600 MHz NMR instrument. Following the analysis of variances, a detailed uncertainty budget is calculated and all the data is summarised in a comprehensive certificate. Figure 3 2D COSY of 3,5-dinitrobenzoic acid: the cross peaks of the impurity 3-nitrobenzoic acid can be identified although the relative content of this impurity is below 0.1%. By this quality procedure, most qnmr reference substances are obtained with purities 99.5% and have typical uncertainties 0.2%. All qnmr standards are packed in brown-glass bottles and closed with a sealed cap. A detailed certifi cate in accordance with ISO Guide 31 is delivered with each CRM. The certificate contains a comprehensive documentation tabulating lot-specifi c values, traceability, uncertainty calculation, expiration date and storage information. Since qnmr off ers several advantages over other analytical techniques, Sigma-Aldrich will launch a new series of certified organic standards for chromatography whereby the certifi ed value of these CRM will not be measured by HPLC or GC only but also by qnmr.

back to top References 1. ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories. 2. ISO Guide 34, General requirements for the competence of reference material producers. back to top