Sub-100 nm calibration for size and number



Similar documents
ME 538: Experiment # 1 Flow measurement and particle generation instruments

Current and Future Methods for Calibration of Black Carbon Real-Time Mass Instruments

The ultimate Nano Particle Counters and Sizers

Particle Number Counter Calibration Procedures. Report to the Department for Transport

Scanning Mobility Particle Sizer

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Critical Radius and Supersaturation. Condensed Water Molecules. Cloud Droplet Nucleation. Bubbles. Clouds. Change in number of liquid molecules

Particulate Emission from Internal Combustion Engines

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

OLIVÉR BÁNHIDI 1. Introduction

Atmospheric Monitoring of Ultrafine Particles. Philip M. Fine, Ph.D. South Coast Air Quality Management District

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

Solutions to the questions from chapter 1 and 2 in GEF Cloud Physics

ISO A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners

PHYSICAL-CHEMICAL PROCESSES OF CLOUD ACTIVATION STUDIED WITH A DESKTOP CLOUD MODEL

Typically the braking system relies on friction between an iron disc (rotor) and a brake pad made of friction material bonded to a backing plate.

1.0 What Are the Purpose and Applicability of Performance Specification 11?

SCION SQ Series GC-MS

POLLUTED EMISSION TREATMENTS FROM INCINERATOR GASES


PROCESS PARTICLE COUNTER (PPC) SENSOR/CONTROLLER FOR OPTIMIZING POWER RECOVERY EXPANDER AND GAS TURBINE PERFORMANCE

The Basic Principles of Sieve Analysis

Technical Considerations Detecting Transparent Materials in Particle Analysis. Michael Horgan

Physical & Chemical Properties. Properties

Fault codes DM1. Industrial engines DC09, DC13, DC16. Marine engines DI09, DI13, DI16 INSTALLATION MANUAL. 03:10 Issue 5.0 en-gb 1

Particle Surface Area

Supporting document to NORSOK Standard C-004, Edition 2, May 2013, Section 5.4 Hot air flow

Use of OpenFoam in a CFD analysis of a finger type slug catcher. Dynaflow Conference 2011 January , Rotterdam, the Netherlands

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering

AEROSOL STATISTICS LOGNORMAL DISTRIBUTIONS AND dn/dlogd p

Estimating Aerosol Surface Area from Number and Mass Concentration Measurements

Measuring Laser Power and Energy Output

L-LAS-TB-CL serie. laser light curtains for inline measuring tasks

REPRESENTATIONS OF CLOUD DROPLET ACTIVATION PROPERTIES FOR SURFACE ACTIVE ORGANIC AEROSOL. Kuopio, Finland

Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System

Getting the most from this book...4 About this book...5

Kompressoren. Compressed Air Filters PF/DF/MF/SF/AF

MILESTONE H E L P I N G C H E M I S T S

Measuring of optical output and attenuation

Method 202 Dry Impinger Method for Determining Condensable Particulate Emissions from Stationary Sources. By: Paul Ottenstein

Type Testing Procedure for Crankcase Oil Mist Detection and Alarm Equipment

Ice nucleation of desert dust and other mineral aerosols: processes and parameterisations

DUST EMISSION MONITORING SYSTEM

ELECTRICITE ET MAGNETISME.

MATRIX TECHNICAL NOTES

QGA Quantitative Gas Analyser

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

Application Note Noise Frequently Asked Questions

Assessments of LEDs IESNA LM-79, LM-80 & LM-82

THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS

S300 Series Dust Monitors

Hellma. Where precision becomes an art. Calibration standards for spectrophotometers by Hellma

TRIAL CHEMICAL CLEANING OF FOULED APH BASKETS

Type Testing Procedure for Crankcase Oil Mist Detection and Alarm Equipment

REAL-TIME DUST AND AEROSOL MONITORING

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Air Pollution Control Division. Technical Services Program

Near Real-time Monitoring of Diesel Exhaust Particulate (DEP)

Computational Fluid Dynamics (CFD) and Multiphase Flow Modelling. Associate Professor Britt M. Halvorsen (Dr. Ing) Amaranath S.

LVDS Technology Solves Typical EMI Problems Associated with Cell Phone Cameras and Displays

This article appeared in a journal published by Elsevier. The attached copy is furnished to the author for internal non-commercial research and

AR200. Refractometer User s Guide. Page 1

Chapter 13 - Solutions

Continuous process of sodium bicarbonate production by Solvay method

DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements

Particle motion, particle size, and aerodynamic diameter

Passive Remote Sensing of Clouds from Airborne Platforms

Power Amplifier Gain Compression Measurements

A software for calculation of optimum conditions for cotton, viscose, polyester and wool based yarn bobbins in a hot-air bobbin dryer

HS 1003 Part 2 HS 1003 Heavy Metals Test

Nederland België / Belgique

Experiment #5: Qualitative Absorption Spectroscopy

IB Chemistry. DP Chemistry Review

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Factors Affecting Precipitation of Calcium Carbonate

UVC LEDs for. Environmental Monitoring

Certification system of air filter quality (CAF mark)

Experiment #10: Liquids, Liquid Mixtures and Solutions

Fig.1. The DAWN spacecraft

Solidification, Crystallization & Glass Transition

Calibration of the passive cavity aerosol spectrometer probe for airborne determination of the size distribution

CFD Modelling of Wet Flue Gas Desulphurization (WFGD) Unit: A New Era of Process System Control and Optimization

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS

Fiscal Measurement Natural Gas

Common Defects in Digital Printing. Paul Geldenhuys & Amir Shapira January, 2009

Engine Heat Transfer. Engine Heat Transfer

Laminar Flow Elements

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

Energy and Flow Measurement for Hydronic Systems

Transcription:

Calibration of DMS Series Fast Particulate Spectrometers Contents Introduction...1 Sub-100 nm calibration for size and number...2 Supra-100 nm calibration for size...3 Certification...4 Agglomerate Aerosols: Calibration for Diesel...5 Calibration for 2.5 µm and switchable instruments...5 Appendix A: Reasons For Not Using a CPC...5 Appendix B: Sulphuric Acid Aerosol Generation....6 Introduction This document outlines how a DMS500 or DMS50 is currently calibrated for particle size and number. The fundamental base calibration arises from a detailed computer model of the instrument, which takes into account the aerosol and sheath flows, the unipolar diffusion charger characteristics, and classifier column voltage, pressure, temperature and geometry. The probability distribution of electrical charges on a particle of given size is calculated, and this is used to predict which rings such particles are likely to land on, given the conditions of the classifier. This model produces a complete transfer function for the instrument, mapping particle size to ring currents (Figure 1). Figure 1: DMS500 Transfer Function After the DMS has been checked for correct operation, with its internal flows within tolerance, the computer generated transfer function is empirically adjusted with real Cambustion 2007-08. Version 3. Page 1 of 6

aerosols. The lognormal data parameterisation (user interface 2.01 et seq.) is used for this (as opposed to the discrete spectrum) as it provides a direct reading of the mean diameter of individual aerosol modes, and improves the spectral resolution. Sub-100 nm calibration for size and number This procedure is closely related to the calibration procedure generally used by CPC manufacturers, using a standard electrometer, whilst taking into account the size information provided by the DMA. Number calibration against a CPC is not performed for the reasons discussed in Appendix A. The aerosols used are: Nucleated Liquid Sulphuric Acid (~ 15nm). See appendix B for the generation method. Atomised and dried solid Sodium Chloride (~ 50 & 100 nm) These aerosols are size selected with a TSI 3080 electrostatic classifier (with 3081 DMA). The monodisperse aerosol stream is split into two, with half passing through a radioactive neutraliser into the DMS, and the other half passing through a Faraday cage electrometer system. The size and concentration are measured by the DMS, and compared with the size set by the DMA and the concentration measured by the electrometer. The setup is shown in Figure 2. air Regulator Collison Nebuliser H 2 SO 4 (aq) or NaCl (aq) Diffusion Dryer spill Path for H 2SO 4 Regulator Nucleation Apparatus Pump DMS MFC Neutraliser Filter in Faraday cage 3081 DMA spill in Neutraliser (bipolar charger) Computer 617 Electrometer Figure 2: Sub-100 nm Calibration The electrometer system is based around a filter enclosed by a conductive shell. By Gauss theorem, charged particles arriving within the shell (i.e. collected by e.g. Liu, B.Y.H and Pui, D.Y.H. (1974) A Submicron Aerosol Standard and the Primary Absolute Calibration of the Condensation Nuclei Counter, J. Colloid Interface Science, 47, 155. Page 2 of 6

the filter) will induce a charge in the shell, which will produce a measurable current when connected to a sensitive electrometer (Keithley model 617). A pump draws a comparable flow to that of the DMS through the filter, metered and controlled by an Aalborg mass flow controller. The number concentration (in cm -3 ) is then given by: I N = eq where Q is the flow through the filter (in cm 3 s -1 ), e is the electronic charge (in C) and I is the measured current (in A). This calculation assumes that the aerosol leaving the DMA is singly charged. In practice this is achieved by ensuring that the original aerosol has very few particles larger than the required DMA size, as these particles might be doubly charged yet still appear to have the same electrical mobility as singly charged particles of the required size. This is ensured by (a) the H 2 SO 4 aerosol from the nucleation apparatus being essentially monodisperse and (b) carefully choosing the NaCl(aq) concentration so that DMA size selection occurs from the largest particle end of the polydisperse aerosol distribution. Supra-100 nm calibration for size NIST Traceable Polystyrene Latex (PSL) spheres of certified size are used. These are nebulised in aqueous suspension, then dried and neutralised before being sampled by the DMS. Particles produced by the nebuliser which have been dried will be very highly charged, so they must be neutralised before DMS measurement, or both measured size and number will be affected. The hardware setup is shown in Figure 3. air Regulator Collison Nebuliser Diffusion Dryer Neutraliser DMS PSL suspension spill Figure 3: PSL Aerosol Generation Page 3 of 6

dn/dlogdp /cc 299nm PSL Spheres - Standard DMS500 Spectral Output 1.20E+06 1.00E+06 Standard Lognormal 8.00E+05 6.00E+05 4.00E+05 SURFACTANT PEAK PSL PEAK 2.00E+05 0.00E+00 1.00 10.0 100 1000 Dp (nm) Figure 4: 299 nm PSL sampled with the DMS500 Figure 4 shows an example of a PSL spectrum. The lognormal data inversion gives a much higher spectral resolution than the discrete spectrum. Note the broad mode at small sizes, caused by surfactants and impurities. PSL is unsuitable for use below 100 nm, as this surfactant mode can mask or distort the PSL peak for smaller PSL sizes. Also, it important to dilute the PSL suspension sufficiently so that no more than one PSL particle is present per water droplet, or particle aggregation can occur upon drying, giving an inaccurate mobility size. This problem is less significant for larger PSL sizes. The PSL point at 100 nm provides a useful cross-check of the calibration of the DMA used for the sub-100 nm region. Due to the issue of multiple charging, it is not yet practical to provide number certification for larger particle sizes, so the original computer model is currently relied upon here. Certification After empirical adjustment of the transfer function, a new.dmd file containing the new matrix is loaded, and the instrument is then checked against similar aerosols. Certification is dependent upon: The size being within 5% of the standard in the region of 5 300 nm. The size being within 10% of the standard for diameters > 300 nm. The number concentration being within 10% of the standard. Page 4 of 6

Agglomerate Aerosols: Calibration for Diesel Because highly fractal aerosols acquire more charge in this type of instrument, a separate calibration is provided for use with Diesel engines. In particular, this ensures greater accuracy when compared with a Particle Measurement Programme (PMP) number measurement system. The Diesel.dmd calibration file supplied with each DMS uses this calibration for the accumulation mode of the lognormal output only; the nucleation mode and the continuous red spectrum use the standard calibration and therefore one can expect the accumulation mode lognormal peak to indicate a lower concentration and larger size than the continuous spectrum. The calibration procedure is similar to the sub-100nm calibration for size and number descried above, except rather than being NaCl or H 2 SO 4 the aerosol source is soot from a propane fuelled CAST device. The soot is pre size-selected with a second DMA to be monodisperse thus reducing the likelihood of multiple charging affecting the gain calibration. Certification by this process is at 50, 75, 100 and 200 nm. The instrument is expected to meet the standard within 10% for both size and number. To allow for the same level of accuracy in mass calculation as was achieved in Symonds et al. (2007), since the introduction of the agglomerate calibration, the diameter to mass rule has been revised for this calibration to be Mass = 2.2 10-15 D p 2.65, and the Diesel.dmd file now uses this rule. The GDi.dmd file uses the standard spherical calibration for both nucleation and accumulation modes as this gives the best results for particle number and mass, see application note DMS08 for more details. Calibration for 2.5 µm and switchable instruments This is similar to the standard spherical calibration, except that monodisperse DEHS (oil) aerosol from a Sinclair-LeMer type condensation aerosol generator is used for DMA based size/number calibration between 100 and 500 nm, and PSL spheres are used beyond this size. Appendix A: Reasons For Not Using a CPC. The practical minimum concentration for such a measurement with a DMS type instrument (> 10 4 cm -3 ) would mean that most CPCs would be operating in photometric mode, (as opposed to direct count mode), which is based upon bulk optical properties of the aerosol and is dependent upon manufacturer calibration. It is Symonds, J.P.R., & Reavell, K.St.J. (2007). Calibration of a Differential Mobility Spectrometer, European Aerosol Conference, Salzburg, T02A034 Symonds, J.P.R., Reavell, K.St.J., Olfert, J.S., Campbell, B.W., & Swift, S.J. (2007). Diesel Soot Mass Calculation In Real-time With A Differential Mobility Spectrometer, Journal of Aerosol Science, 38 52 68 Page 5 of 6

this mode which CPC manufacturers calibrate against an electrometer as the primary standard, so the use of a CPC would be a secondary standard in this case. Figure 5 shows a comparison of the DMS500 with a TSI 3022 CPC. A sodium chloride aerosol with a GMD of ~ 55 nm and a GSD of ~ 1.9 was used. The DMS500 and CPC were fed with the same aerosol, and a DMS500 analogue input was used to log the CPC. Large scale concentration fluctuations in the aerosol are due to deliberate changes in the dilution flow, small scale fluctuations are due to the nebuliser s air regulator. It can clearly be seen from Figure 5 that the DMS500 and CPC are in excellent agreement on total number whilst the CPC is in count mode, but as soon as the CPC enters photometric mode an offset is seen on the CPC s output. However, these data are rather close to the noise floor of a DMS, and so to use a CPC in count mode for certification would require an additional diluter, which in itself would need calibration and would be an additional source of error. An SMPS (DMA + CPC) is a complex system, reliant upon its own calibration and complex data inversion, so this is not used either. 5.00E+04 Comparison of DMS500 and 3022 CPC 4.50E+04 4.00E+04 3.50E+04 3.00E+04 DMS CPC N/cc 2.50E+04 2.00E+04 1.50E+04 1.00E+04 5.00E+03 CPC PHOTOMETRIC MODE CPC COUNT MODE 0.00E+00 0.00 100.00 200.00 300.00 400.00 500.00 600.00 700.00 800.00 900.00 Time / s Figure 5: Comparison of DMS500 and 3022 CPC Appendix B: Sulphuric Acid Aerosol Generation. Sulphuric acid is nebulised and then partially dried before passing into the nucleation apparatus. This consists of a heated pipe (~ 120 C) to vaporise some of the nebulised H 2 SO 4, followed by a cold pipe into which dilution air is added to cause re-nucleation of H 2 SO 4 and hence the generation of very small particles. This aerosol has two characteristic peaks, the nucleation mode at < 20 nm and the remaining un-evaporated acid mist at >20 nm. The apparatus can be tuned in size and in concentration ratio of the two peaks by adjusting the temperature and flows; this is to ensure the >20 nm peak is negligible in concentration compared with the nucleation peak and that the nucleation peak is smaller than 15 nm in size for selection by the DMA. For more details on this method of aerosol generation, see Cambustion Application Note DMS02. Page 6 of 6