Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose



Similar documents
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

Welcome to More Choice BD Analyte Specific Reagent (ASR) Catalog

Instrument Characterization and Performance Tracking for Digital Flow Cytometers

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size

Three Rules for Compensation Controls

Introduction to flow cytometry

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences

Compensation in Flow Cytometry

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

Immunophenotyping peripheral blood cells

Fluorescence Compensation. Jennifer Wilshire, PhD

Compensation Controls Data Visualization Panel Development Gating Quantum Dots

BD FC Beads 8-Color Kit for BD OneFlow Assays

Compensation: An Instrumental Perspective

Introduction to Flow Cytometry

FlowSight. Flow cytometry with vision

Introduction to Flow Cytometry

EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocols

BD FACSComp Software Tutorial

COMPENSATION MIT Flow Cytometry Core Facility

Boundary-breaking acoustic focusing cytometry

BD FACSDiva TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Let the BD FACSAria II accompany your best performance ever

FACS Laboratory. BD LSRs Operator Training. Training Guide.

BD FACSCanto Flow Cytometer. Efficient Easy Powerful

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee

BD FACSVerse. Simply brilliant

ArC Amine Reactive Compensation Bead Kit

These particles have something in common

123count ebeads Catalog Number: Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

powerful. Intuitive. Customizable. The new star of benchtop flow cytometry

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer

CyAn : 11 Parameter Desktop Flow Cytometer

Cell Cycle Tutorial. Contents

CyFlow Space Your flexible flow cytometer

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

Standard Operating Procedure

Introduction into Flow Cytometry

Principles of Flowcytometry

Below is a list of things you should be aware of before you schedule your sort.

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

APPLICATION INFORMATION

BD FACSAria Flow Cytometer. The first high-speed sorter with a fixed-alignment cuvette flow cell from BD Biosciences

Zecotek S Light Projection Network Marketing

Standardization, Calibration and Quality Control

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Advantages of the Alexa Fluor 488 secondary antibody conjugates

Establishing Optimum Baseline PMT Gains to Maximize Resolution on BD Biosciences Digital Flow Cytometers

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Getting Started with BD FACSDiva Software

Setting up and calibration of a flow cytometer for multicolor immunophenotyping

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:

1. Instrument Layout pg Summit Overview pg Data Storage and Data Files pgs Saving Data pg Start Up pg 8

CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles

A practical approach to multicolor flow cytometry for immunophenotyping

BD FACSAria Flow Cytometer. BD Biosciences Clontech Discovery Labware Immunocytometry Systems Pharmingen

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility

Application Note. Multiparametric Immunophenotyping of Human Hematopoietic Stem Cells and Progenitor Cells by Flow Cytometry.

Services. Contract. Optimize. Results. Instrument. Guidance. Reagent

HIGH PERFORMANCE FLOW CYTOMETER FOR EVERYONE

Introduction to Flow Cytometry:

Flow Cytometry. flow cytometer DNA apoptosis ph

T-Cell Research. Novel multicolor flow cytometry tools for the study of CD4 + T-cell differentiation and plasticity

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction

BD FACSCanto II Instructions For Use

BD StemFlow Human Pluripotent Stem Cell Sorting and Analysis Kit

EdU Flow Cytometry Kit. User Manual

Uses of Flow Cytometry

BD FACSAria III User s Guide

Islet Viability Assessment by Single Cell Flow Cytometry

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

BD FACSCanto II Flow Cytometer Reference Manual

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.

How To Read Flow Cytometry Data

Product Catalog. biolegend.com General Information. Technical Charts and Protocols. Mouse Immunology Antibodies

Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113

Spherotech, Inc Irma Lee Circle, Unit 101, Lake Forest, Illinois

Institut für Experimentelle Immunologie und Bildgebung

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

Flow Cytometry A Basic Overview

Chapter 12 Filters for FISH Imaging

LIVE/DEAD Fixable Dead Cell Stain Kits

Stem Cell Research Source. From Isolation to Culture and Analysis

Easy to use. Compact. Flexible. Sorting Chip

Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available!

BD Pharmingen. Apoptosis, DNA Damage and Cell Proliferation Kit. Technical Data Sheet. Product Information. Description. Material Number: Size:

Annexin V-EGFP Apoptosis Detection Kit

Transcription:

Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities: Define winning combinations 4 Tandem dyes: Watch out for degradation 5 Controls to validate your panel 6 Multicolor flow cytometry reagents Flow cytometers that detect six, eight, and more colors have spurred the development of new fluorochromes and antibody conjugates that take advantage of these capabilities. However, choosing the optimal combination of fluorochromes for your particular antibody specificities is a complex process. This application note provides some simple guidelines to help you select reagent panels for multicolor flow cytometry, avoid rounds of trial and error, and increase your chances of a successful experiment. The basics: Know your instrument Reagent selection starts with your instrument configuration. The type and number of lasers and detectors dictate whether the optical system can excite a given fluorochrome and properly detect a given combination of fluorochromes. The design of the optical system also impacts the efficiency with which particular dyes are detected, as do the instrument settings, including PMT voltages. (See the BD Cytometer Setup and Tracking [CS&T] feature in BD FACSDiva 6.x software at bdbiosciences.com/facsdiva). Finally, the choice of optical filters that are used with each detector greatly influences the effective brightness of one fluorochrome versus another. Filter selection is a give-and-take process: using a wider bandpass filter can increase the ability to detect a given fluorochrome, but may also increase the amount of spillover background contributed into that detector from other neighboring fluorochromes. A good way to visualize these effects is by virtual testing of filter combinations using a web tool such as the viewer at bdbiosciences.com/spectra. This tool illustrates predicted spillovers* for particular fluorochromes and filter combinations. * The predicted leakage in normalized percentage of emitted fluorescence, which might not necessarily correspond to the percentage used for compensation, due to the effects of varying detector gains.

W Stain Index = D/W Figure 1. The effective brightness of a reagent depends upon the difference between the positive and the negative (D) and the spread of the negative population (W). The stain index is a useful metric for normalized signal over background. D Fluorochromes: Go for the bright... Given the many differences in instrument configuration, it is impossible to universally state the best fluorochromes to use in combinations of 6, 8, or more colors. However, for a particular cytometer such as the BD LSR II instrument, it is possible to rank available dyes according to their brightness on that instrument (when configured with a specified set of lasers and filters). But how exactly do we define and measure brightness? A good functional definition of reagent brightness should include a term to weigh the effects of background contributions, or how resolvable the signal is over the unstained cell populations. Background in a particular detector is influenced by signal levels, cell autofluorescence, non-specific staining, electronic noise, and optical background from other fluorochromes (spillover). To the extent that these factors each contribute noise and/or signal, they increase the width (the variance or SD 2 ) of a negative population (see Figure 1). As such, a good normalized functional measure of reagent brightness is the stain index, 1 defined as D/W, where D is the difference between positive and negative populations, and W is equal to 2 SD of the negative population. When the same antibody is conjugated to various dyes, their stain indexes can be compared to get an idea of the relative brightness of the dyes on a particular instrument. This assumes that the conjugation chemistries for all of the reagents have been optimized. Using this assumption, Table 1 shows the stain index for a number of different CD4 (clone RPA-T4) conjugates, using the specified filters. The information in this table provides an idea of the relative brightness of different fluorochromes on this platform. This leads to the first rule of reagent selection, which is to pick the brightest available fluorochromes. Suppose you have a four-color panel consisting of reagents in FITC, PE, PerCP-Cy 5.5, and APC, and you want to add a fifth color. PE-CF594 is an obvious choice, since it is a bright fluorochrome not already in your panel. Fluorochrome Filter Stain Index PE 575/26 232 BD Horizon PE-CF594 610/20 232 PE-Cy5 670/14 216 APC 660/20 193 PE-Cy7 780/60 86 Alexa Fluor 647 660/20 85 PerCP-Cy5.5 695/40 82 Alexa Fluor 488 530/30 53 FITC 530/30 51 BD Horizon V450 450/50 42 Pacific Blue 450/50 36 Alexa Fluor 700 730/45 27 PerCP 695/40 25 APC-Cy7 780/60 24 AmCyan 525/20 23 BD Horizon V500 525/20 21 BD APC-H7 780/60 17 Table 1. Stain index of various fluorochrome conjugates run on a BD LSR II flow cytometer. For a complete list of the relative stain indices of fluorochromes offered by BD Biosciences, see bdbiosciences.com/documents/multicolor_fluorochrome_guide.pdf

Page 3...but minimize spillover Brightness, on its own, only goes so far. The stain indexes listed in Table 1 are calculated with the reagents run singly, not as part of a cocktail. As soon as other reagents are added, spectral overlap (or spillover) becomes an issue. The more colors one attempts to resolve, the more spillover between those colors that will have to be dealt with. We deal with spillover by applying compensation, which applies a correction to all signals such that, for example, a cell population fluorescing only in PE will show no FITC fluorescence, on average. While this is true for the population on average, individual cells will fall above or below the mean, and this data spread is higher when spectral overlap introduces additional noise. 1 Compensation unfortunately does not remove this noise even though the median is corrected since the data spread (variance) from photoelectron statistics remains. The effect of data spread is thus to reduce the resolution sensitivity, and therefore the stain index, in a fluorescence detector that receives spillover from other detectors. Thus, the second rule of reagent selection is to minimize the potential for spectral overlap when choosing a reagent combination. This can conflict with the first rule, which was to choose the brightest fluorochromes. For example, PE-Cy 5 is very bright with regard to stain index (Table 1), but it has considerable spillover (ie, adds background) into the APC detector. While these two fluorochromes can be used together, the resolution sensitivity in APC will be reduced compared to, for example, a combination of PerCP-Cy5.5 and APC. This is a case when you might wish to sacrifice a certain amount of brightness in one detector to avoid spillover (and loss of resolution sensitivity) in another. Colors and specificities: Define winning combinations Taking the two rules into account, we generated a set of fluorochromes that are reasonable choices to use for experiments requiring 6, 8, or more colors (Table 2). Note that these choices are based upon BD instruments (BD FACSVerse, BD LSR cell analyzer platform, BD FACSAria cell sorter platform, or BD FACS Canto flow cytometers), and upon reagents that usually can be purchased from a catalog as antibody conjugates. Once the fluorochromes to be used have been defined, you can begin to match antibody specificities to particular fluorochromes to select the actual conjugates to be used. For this purpose, brightness and spillover remain key issues, as the following example illustrates. 6-color 8-color 10-color FITC or Alexa Fluor 488 FITC or Alexa Fluor 488 FITC or Alexa Fluor 488 PE PE PE PE-Texas Red or BD Horizon PE-CF594 PerCP-Cy5.5 PerCP-Cy5.5 PerCP-Cy5.5 PE-Cy7 PE-Cy7 PE-Cy7 APC or Alexa Fluor 647 APC or Alexa Fluor 647 APC or Alexa Fluor 647 Alexa Fluor 680 or 700 APC-Cy7 or APC-H7 APC-Cy7 or APC-H7 APC-Cy7 or APC-H7 AmCyan or BD Horizon V500 AmCyan or BD Horizon V500 BD Horizon V450 or Brilliant Violet 421 Table 2. Common choices for 6, 8, and 10-color experiments BD Horizon V450 or Brilliant Violet 421

Imagine that you want to look at CD62L staining on CD8 + T cells. While CD8 is an abundant protein, and antibodies to it stain cells very brightly, CD62L is relatively dim (the protein is not abundant on the cell surface, and/or the available antibodies are of low affinity). Initially, you might consider using the brightest available fluorochrome, PE, for CD62L, while using a dimmer fluorochrome, like FITC, for CD8. But FITC has considerable spillover into the PE detector (see Figure 2). The result is that, for CD8 + cells that are highly stained in the FITC detector, resolution sensitivity in PE is compromised, to the point that CD62L resolution might be suboptimal. There are multiple possible solutions to improve this, including: 1) Move CD8 to a fluorochrome that has less spectral overlap with PE (such as PerCP-Cy5.5 or APC). 2) Move CD62L to a detector that is still relatively bright, but does not overlap with FITC (such as APC or PE-Cy5). Note that in this example, only the CD8 + cells are highly stained in the FITC detector, so only CD8 + cells contribute to data spread in the PE detector. If you were interested in CD62L staining only on CD4 + (CD8 ) cells, the original reagent combination (CD8 FITC, CD62L PE) would be fine. This example illustrates two additional rules in reagent selection: reserve the brightest fluorochromes for dim antibodies, and vice versa, but avoid spillover from bright cell populations into detectors requiring high sensitivity for those populations. Figure 2. Spectral overlap of FITC into the PE detector. Taken from bdbiosciences.com/spectra Tandem dyes: Watch out for degradation One final topic to consider is the potential for tandem dye degradation. APC-Cy7, and to a lesser extent, PE-Cy7, can degrade in the presence of light, fixation, and elevated temperatures, so that they emit in the parent dye detector (APC or PE). This process often starts with a small subpopulation of cells, leading to false positive events in APC or PE. 1 By minimizing the

Page 5 exposure of samples to light, heat, and formaldehyde-based fixatives, this problem can be largely avoided. Additionally, BD Biosciences has developed APC-H7, an analog of APC-Cy7 with enhanced stability in light, heat, and formaldehyde-based fixatives. Still, for some applications, there is an additional rule: consider the consequences of degradation of tandem dyes and whether this will compromise sensitive readouts in the APC or PE detectors. If so, a different reagent configuration might be in order. For situations in which final fixation of samples is required (eg, biohazardous samples), there is a stabilizing fixative available that helps prevent degradation of APC-Cy7 while still fixing cells (BD Biosciences Cat. No. 338036). For more stable tandem dyes, BD now offers APC-H7 conjugated antibodies. Controls to validate your panel Once all of these factors have been considered, you are ready to test a multicolor reagent cocktail. In so doing, there are two types of controls that you might wish to include in your initial testing: fidelity controls and fluorescence-minus-one (FMO) controls. 2 Fidelity controls use a given antibody by itself (or with minimal additional gating reagents), and they compare the results to the use of that antibody in a complete cocktail. From this, you can see the effect of additional reagents on the readout of interest, to be sure that the other reagents are not compromising that readout. FMO controls combine all the reagents in a given cocktail, except for one reagent of interest. They are useful to gauge the sensitivity of particular detectors in the context of the other reagents. They also may be used for routine gating of those detectors for which other means of setting gates are not possible or practical. In general, though, once a reagent panel has been well validated, it is not necessary to run all of these controls on a day-to-day basis. To summarize, this article has set forth rules for selecting reagents for multicolor flow cytometry. These rules need to be balanced to achieve the best possible results: Rule 1: Choose the brightest set of fluorochromes for your particular instrument configuration. Rule 2: Choose fluorochromes to minimize the potential for spectral overlap. Rule 3: Reserve the brightest fluorochromes for dim antibodies, and vice versa. Rule 4: Avoid spillover from bright cell populations into detectors requiring high sensitivity for those populations. Rule 5: Take steps to avoid tandem dye degradation, and consider its impact upon results. With time, proven panels of multicolor reagents will become available, both from literature sources and from manufacturers. Many users will not have to start from scratch in designing reagent panels. Nevertheless, given the huge variety of applications of multicolor flow cytometry, most researchers sometimes will find themselves having to select their own multicolor reagents. We hope that these rules will prove helpful. Please visit bdbiosciences.com/colors for more details about our multicolor flow cytometry reagents.

CD8 PE-Cy7-A CD19 APC-H7 10 2 10 3 10 4 10 5 0 10 2 10 3 10 4 10 5-10 2 0 10 2 CD8 B Cells 10 3 10 4 10 5 HLA-DR V450 CD4 Multicolor flow cytometry reagents Expand your multicolor flow with options beyond your typical five colors. BD Biosciences continues to increase our portfolio of multicolor reagents to support your cutting-edge research using the expanded fluorescent detection capabilities of our flow cytometers. We offer directly conjugated antibodies to fluorochromes that go beyond your typical five colors by providing colors for violet, red, and blue lasers. These include AmCyan, BD Horizon V500, BD Horizon V450, Brilliant Violet 421, BD Horizon PE-CF594, Alexa Fluor 700, and BD APC-H7 conjugates to support up to ten-color staining with off-theshelf reagents. BD Biosciences Custom Conjugation Program* offers more choices for your multicolor experiment design. This program offers multiple colors, as shown in Table 3. * Please contact your BD sales representative for more information about the Custom Conjugation Program. Fluorochrome Ex Max (nm) 10 2 10 3 10 4 10 5 CD4 AmCyan-A Figure 3. Example of the use of gating reagents (CD4, CD8, CD19) using non-traditional fluorochrome conjugates excited by blue, red, and violet lasers. Alexa Fluor 488 495 Alexa Fluor 594 590 Alexa Fluor 647 650 Alexa Fluor 680 679 Alexa Fluor 700 696 AmCyan 457 APC 650 APC-Cy7 650 BD APC-H7 650 BD Horizon V450 404 FITC 494 Pacific Blue 401 PE-Cy5 496, 564 PE-Cy7 496, 564 PerCP 482 PerCP-Cy5.5 482 PE-Texas Red 496, 564 Table 3. Custom conjugates available from BD Biosciences References 1. Maecker HT, Frey T, Nomura LE, Trotter J. Selecting fluorochrome conjugates for maximum sensitivity. Cytometry A. 2004;62:169. 2. Baumgarth N, Roederer M. A practical approach to multicolor flow cytometry for immunophenotyping. J. Immunol. Met. 2000;243:77.

Page 7 Table 4. BD Biosciences fluorochrome specifications Analyzers Sorters Fluorochrome Fluorescence Emission Color Ex-Max (nm) Excitation Laser Line (nm)* Em-Max (nm) Hoechst 33342 Blue 350 355, 375 461 Brilliant Violet 421 Blue 407 405 421 BD Horizon V450 Blue 404 405 448 Pacific Blue Blue 401 405 452 BD Horizon V500 Green 415 405 500 AmCyan Green 457 405 491 Alexa Fluor 488 Green 495 488 519 FITC Green 494 488 519 PE Yellow 496, 564 488, 532, 561 578 Brilliant Violet 605 Orange 407 405 602 BD Horizon PE-CF594 Orange 496, 564 488, 532, 561 612 PI Orange 351 488, 532, 561 617 7-AAD Red 543 488, 532, 561 647 APC Red 650 633, 635, 640 660 Alexa Fluor 647 Red 650 633, 635, 640 668 PE-Cy 5 Red 496, 564 488, 532, 561 667 PerCP Red 482 488, 532 678 PerCP-Cy 5.5 Far Red 482 488, 532 695 Alexa Fluor 700 Far Red 696 633, 635, 640 719 PE-Cy 7 Infrared 496, 564 488, 532, 561 785 APC-Cy7 Infrared 650 633, 635, 640 785 BD APC-H7 Infrared 650 633, 635, 640 785 BD Accuri C6 BD FACSCalibur BD FACSVerse BD FACSCanto II BD LSRFortessa and Special Order BD LSRFortessa BD FACSAria III and Special Order BD FACSAria BD Influx BD FACSJazz * The excitation laser line represents commonly used lasers that excite the fluorochrome. It does not necessarily reflect the lasers available for each particular instrument. APC and PE-Cy5 may be used together on instruments with cross-beam compensation. Sensitivity for PerCP with high-power lasers (>25 mw) is reduced and is only recommended for known highly expressed markers. Capable of detecting 8 colors simultaneously (4 blue laser, 2 red laser, 2 violet laser). BD Horizon PE-CF594 and Alexa Fluor 700 filters are available separately. Extra care must be taken to avoid spillover.

For Research Use Only. Not for use in diagnostic or therapeutic procedures. APC-Cy7: US patent 5,714,386 Brilliant Violet 421 and Brilliant Violet 605 are trademarks of Sirigen Group Ltd. Cy is a trademark of Amersham Biosciences Corp. Cy dyes are subject to proprietary rights of Amersham Biosciences Corp and Carnegie Mellon University and are made and sold under license from Amersham Biosciences Corp only for research and in vitro diagnostic use. Any other use requires a commercial sublicense from Amersham Biosciences Corp, 800 Centennial Avenue, Piscataway, NJ 08855-1327, USA. Pacific Blue is a trademark, and Alexa Fluor and Texas Red are registered trademarks of Molecular Probes, Inc. CF is a trademark of Biotium, Inc. BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2012 BD 23-9538-06 BD Biosciences bdbiosciences.com