Upper Arm Blood Pressure Monitor SC 6800. Instructions for use



Similar documents
Automatic Wrist Blood Pressure Monitor with

INSTRUCTION MANUAL. 7 series Blood Pressure Monitor with ComFit TM Cuff. Model BP760 ESPAÑOL ENGLISH

INSTRUCTION MANUAL. Ultra Premium Blood Pressure Monitor featuring ComFit TM Cuff. Model HEM-775

INSTRUCTION MANUAL. 10 series Blood Pressure Monitor with ComFit TM Cuff. Model BP785 ESPAÑOL ENGLISH

Affordable ABPM and strictly guideline-based SBPM for accurate out-of-office blood pressure measurement.

INSTRUCTION MANUAL. 10 series Blood Pressure Monitor with ComFit TM Cuff. Model BP785CAN FRANÇAIS ENGLISH

Name: OMRON M3 HEM-7200-E Upper Arm Blood Pressure Monitor

microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el

How to measure blood pressure using digital monitors

INSTRUCTION MANUAL HEM-6052

Device Equivalence Evaluation Form

Alarm Clock USER GUIDE

INSTRUCTION MANUAL. 10 series Plus Blood Pressure Monitor with ComFit TM Cuff. Model BP791IT ESPAÑOL ENGLISH

AMBULATORY BLOOD PRESSURE MONITORING QUICK START GUIDE

Device Equivalence Evaluation Form

DR 70 Portable Digital DAB+ and FM Radio Receiver User Manual English Please read carefully before using this product

Device Equivalence Evaluation Form

Operating Manual Remote Cost Control Monitor

TECHNI-TOUCH TECHNI-TOUCH

Digital Blood Pressure Monitor. Model UA-767

User Manual Color video door phone

INSTRUCTION MANUAL Contactless Clinical Infrared Thermometer GT-101

EVAPORATIVE AIR COOLER INSTRUCTION MANUAL

Atomic Alarm Clock model 13131W2

Instruction Manual. Image of SP-3015 & SP Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

ILLUMINATING EYE MASSAGER

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

Runtastic rechargeable battery for iphone 4, 4S

User's Guide. Pinless Moisture Meter. Model MO257

Operating Manual Electronic Wheelchair Scale with cable display M50610

OEM MAXNIBP Frequently Asked Questions

Time. Choose a location 6 feet or more from electronics such as cordless phones, wireless gaming systems, televisions,

Intelli-Time Alarm Clock model 13027

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

WRIST BLOOD PRESSURE MONITOR

Fox 40 Whistle Watch USER MANUAL

Rousseau 10. User Guide

THE NEW GENERATION IN TECHNOLOGY. NI-3103A Alarm Clock w/fm Radio ipod & iphone Docking. Operating Instructions

August International Ltd United Kingdom Telephone: +44 (0)

Integrating Sound Level Meter and Datalogger

MICA HEATER INSTRUCTION MANUAL Model No: UHM V 50Hz 2200W

Phoenix Body Fat Scale Manual

GB Instruction Manual.

Important Safeguards

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CHCSCS

Operating Instructions

Magic Wand Portable Scanner

Instructional & Operating Manual

Dehumidifier Users manual. For Models: DH45S DH65S

DAB+ / FM Tuner Model: TU-201

Contents. Product Code Page 1

Installation and Operating Manual p. 23. Radio push button 2 channel: HM-PB-2-WM55-2

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Portable Bluetooth Speaker. Quick Start Guide. Model: BTS201

Hyundai Video Pen VP 02 HD8

Model: Manual DC: WIRELESS COLOR FORECAST STATION

REMOTE TERMOMETER TE 219 LED USER MANUAL

Weather Radio Alarm Clock

Leak Detection System

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Wireless Internet Keyboard and Laser Mouse

Rechargeable Hearing Aid Battery Charger

Wake-up Light. Always here to help you. HF3510. User manual Manuel d utilisation. Register your product and get support at

Instant Ear Thermometer

User Guide Nokia Portable Wireless Charging Plate DC-50

P R E S S U R E P O I N T S S E R I E S : Measuring your blood pressure at home

Important Safety Instructions

CappController Pipette Controller. Operation Manual Version:

Operating instructions Platform/floor scales

DX-395. FM/AM/SW1/SW2 PLL Synthesized Receiver OWNER S MANUAL. Cat. No Please read before using this equipment.

Table of Contents. Stanley TLM330 1

FT-100 Fire Training Fog Machine User Manual English

Use and Care Manual. Model CPA12KH AIR CONDITIONER

Pocket Tach 99 (PT99) Non-Contact Tachometer

VOICE RECORDER NOTE CORDER DP-211 INSTRUCTIONS. Register your product on and get additional benefits from Olympus!

Nokia Wireless Keyboard (SU-8W) User Guide

CAD-05 Kit GSM Auto Dialer. Owner s Manual

AD-01 Slave Auto Dialer. Owner s Manual

VACUUM CLEANER INSTRUCTION MANUAL

Getting started with

The FreeStyle Lite Meter and How It Works

Register your product and get support at LFH0645 LFH0648. EN User manual

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR İ-COLD Rev. 00

G1 GPS Sensor User Manual

400 Amp Rechargeable Jump Start System RAC-HP082

reflecta Super 8 Scanner

PORTABLE AIR CONDITIONER (LOCAL)

Manual FlashLED Li-Ion. Imprint: GIFAS-ELECTRIC GmbH CH-9424 Rheineck Switzerland Technical changes and errors reserved.

Single Day Alarm Clock with. Bluetooth and USB Play-through

CAUTION OPC-LM1-IL. Option Card for Encoder of Line Driver Output. Instruction Manual

Wolverine Data PASS Portable Scanner User Manual V1.0

Owner s Manual RBC-AMS41E. Remote controller with weekly timer. Owner s Manual

BLOOD COLLECTION MIXER

Wireless Indoor/ Outdoor Thermometer

How To Use The Visifault Visual Fault Locator

Owner s manual. Free 5 year parts, 1 year labour gurantee Register your guarantee at

CONTIENTS. 1. SPECIFICATIONS (mm) CAUTIONS INSTALLATION INSTRUCTION

USER S MANUAL. Duct Type Series. Free Joint Multi Air Conditioner (Cooling and Heating) FUEA Series E S F I P D G DB A(1) ENGLISH ESPAÑOL

Transcription:

Upper Arm Blood Pressure Monitor SC 6800 Instructions for use

Contents page Introduction... 3 Features of the SC 6800. 3 Safety instructions... 4 WHO-Interpretation of blood pressure readings... 4 Device description... 5 The LC display... 6 Instructions for measuring the blood pressure... 7 How to put the SC 6800 into operation / Changing the batteries... 7 Setting the clock... 8 The blood pressure measurement... 9 Aborting a measurement... 9 The memory function... 10 Delete a blood pressure value from the memory... 10 Delete all blood pressure values in the memory... 10 Troubleshooting and remedies... 11 Care and maintenance... 12 Periodic re-calibration... 12 Referenced norms... 13 Disposal note... 13 Warranty... 13 Specifications... 14 Customer service SCALA Electronic GmbH SES GmbH Ruhlsdorfer Straße 95 Ruhlsdorfer Straße 95 14532 Stahnsdorf 14532 Stahnsdorf Explanation of symbols Rev. 03/12/2013/GB

Specifications Measuring method: Oscillometric, fuzzy logic Measuring range: Pressure: 20 to 300 mmhg Heart rate: 40 to 200 / min Measuring accuracy: stat. pressure: +/- 3 mmhg Heart rate: +/- 5 % Measuring value memory: 85 memories for systolic / diastolic / heart rate, date and time Display: LC display Pressure pick-up: Semi-conductor pressure sensor Automatic switch-off: after approx. 2 minutes of inactivity Voltage supply: 4 x LR6 (alkali batteries) Working temperature range: +10 C to +40 C Storage temperature range: - 10 C to +50 C Cuff length: for upper arm circumference from 22 cm to 32 cm Specifications may change as technical improvements take place Introduction thank you for deciding for the blood pressure monitor SC 6800 from SCALA Electronic GmbH. To be able to utilise all advantages of the blood pressure monitor to the fullest, you should carefully read these instructions before using the monitor. Keep the instruction leaflet in a safe place. Features of the SC 6800 - The blood pressure monitor has fuzzy logic to identify the pumping pressure. If the pumping pressure set in the factory is not sufficient, the device identifies this and continues pumping. - The device has 85 memory locations - The date and time are displayed and stored together with every blood pressure value. - WHO-Interpretation of blood pressure readings with colored band (See description at page 4, 5 and 6). - If the battery icon is displayed, the batteries are discharged and the device cannot be used until new batteries are inserted. - The cuff is suitable for upper arms with a circumference between 22 and 32 cm. 14 3

Safety instructions The blood pressure monitor SC 6800 enables you to monitor your blood pressure regularly. However, you should never change the treatment which your doctor has prescribed Self-measurement means monitoring, it does not imply diagnosis or treatment. Discuss unusual values with your doctor. Do not change the medication prescribed by your doctor. The heart rate is not a useful indicator of the frequency of a cardiac pace maker. Patients with cardiac arrhythmia should use the blood pressure monitor only after consulting a doctor. WHO*-Interpretation of blood pressure readings 29.12.2007 Meaning Systolic / mmhg Diastolic / mmhg optimal < 120 < 80 normal < 130 < 85 high-normal 130-139 85-89 mildy hypertonic 140-159 90-99 hypertonic, 160-179 100-109 medium severity severly > 180 > 110 hypertonic The World Health Organization (WHO) recommends medical treatment, when the blood pressure value exceeds 140/90 mmhg. 4 Referenced norms The device has been certified in accordance with DIRECTIVE 93/42/EU of 14th June 1993 relating to medicinal products. For registration, a defined number of blood pressure monitors was subjected to stringent clinical testing. These monitors used the same program for blood pressure measurement as your monitor uses. Thus, this program is also regarded as clinically tested. The manufacture of the blood pressure monitor complies with the European norms for blood pressure monitors: EN 60601-1, EN 1060-1 to EN 1060-4. The blood pressure monitor complies with the requirements of European norm EN 60601-1- 2 and EN 60601-1: 2007. Disposal note The disposal of batteries is regulated. Spent batteries are collected at collection points provided by battery dealers and the communities. Dispose of the blood pressure monitor as prescribed, i.e., at a collection point for electrical scrap. Warranty We grant a warranty of two years for the blood pressure monitor SC 6800. Within this period, defects of the BP monitor will be repaired free of charge. You can also have your BP monitor repaired by our customer service after the warranty period. We will provide you with a free cost estimate for the repair. The warranty does not cover damage due to improper handling, accident, natural wear and tear (cuff), non-compliance with the instructions for use or unauthorized modifications of the monitor. Not covered by the warranty are the packaging, batteries and the instructions for use. 13

wrong application of the cuff, use of wrong batteries or use of a rechargeable battery. Repeat a measurement not earlier than after 5 minutes. Care and maintenance Avoid extreme temperatures, moisture, dust and the direct exposure to sunlight. The cuff incorporates a sensitive, air-tight bubble. Please avoid any stress by turning or bending it. Clean the instrument with a soft, dry cloth. Never use petrol, thinner, or similar solvents. Stains on the cuff can be gently removed with a wet cloth and soap. Do not wash the cuff! Do not drop the instrument or treat it violently in any other way. Never attempt to open the instrument, as the calibration set by the manufacturer will become invalid. Periodic re-calibration We recommend re-calibrating the pressure display regularly every 2 years. Device description Cuff Air plug WHO indicator LC display Memory button O/I START button 12 5

The LC display Troubleshooting and remedies Possible errors which can occur during a measurement are listed in the table below. Message Error description Correction Memory icon Battery icon WHO indicator Heartbeat icon Date Time Systolic blood pressure Diastolic blood pressure Heart rate per minute or no display Pressure is not build up in the cuff. Cuff not applied tightly enough or leaky. Pressure is built up very quickly at the sensor. The cuff is too tight or the bubble sticks. Pump defective or pumping error The pressure can not be measured due to signal noise. The pumping pressure is higher than 300mmHg. Low batteries. Apply cuff again and measure again. Consult customer service if necessary Apply cuff again and measure again. Consult customer service if necessary Contact customer service Please take the measurement on the left upper arm. Make sure of the correct seat of the cuff. Repeat measurement. Contact customer service Please check and replace the batteries if necessary. Do not use Ni-Cd or NiMH rechargeable batteries From many years of experience we know that most faults are the consequence of wrong use. The most frequent causes of faults are excessive movement during the measurement, 6 11

The memory function The SC 6800 saves up to 85 measurements. Together with each measurement (systolic, diastolic and heart rate), the date and time are also saved. The measurement with the longest residence time in the memory is assigned number 1. All other measurements increment by one until number 85. Then the memory is full. The next measurement occupies memory location 85. The measurement formerly at location 85 changes to location 84 (etc., 83 82... 2 1). The oldest measurement formerly at location 1 is lost. Delete a blood pressure value from the memory Press the MEMORY button to view the blood pressure value you want to delete. Pressing the 0/I-START button instantly (and holding it for at least 4 secs.) displays del (for "delete") and the number of the memory location. Press the 0/I-START button instantly and the value will be deleted. Delete all blood pressure values in the memory Remove the batteries for about 5 minutes to delete all memory entries, the time and date settings. Or: Press the MEMORY button to display any blood pressure value. Pressing the 0/I-START button instantly (and holding it for at least 4 secs.) displays del (for "delete") and the number of the memory location Press the MEMORY button instantly and the LCD displays del (for "delete") and ALL (for all memory locations). Press the 0/I-START button instantly and all values will be deleted. Time and date are not deleted. Instructions for measuring the blood pressure In order to obtain comparable values, always measure the blood pressure under the same conditions, i.e. in a relaxed atmosphere, not after the consumption of alcohol, not during a meal and not after being exposed to physical or nervous stress. If possible, make the measurement while being seated at a table and only with your left upper arm stripped. How to put the SC 6800 into operation / Changing the batteries Put 4 LR6 batteries in the compartment as indicated on the back of the device. Preferably, alkali batteries (LR6) should be used. If the batteries are flat, an empty battery will be displayed at the LC display above:. The battery needs urgent replacement because you cannot use the monitor with depleted batteries. Caution: Do not dispose of depleted batteries with the household waste. Also see disposal note. 10 7

Setting the clock After you have inserted the two batteries, you should set the clock. It starts with year 20XX, 1/ 1 (for the 1st month / 1st day) and 12:00 o clock after every battery change. To set the clock, the device must be OFF. Hold the MEMORY button pressed for about 5 seconds. The year 20XX blinks in the top left corner of the display. Set the year with the 0/I-START button. When the correct year is set, press the MEMORY button again. Now the 1 blinks in the top left corner of the LCD. It stands for the month. Set the month with the 0/I-START button. Another press of the MEMORY button sets another 1 blinking. It stands for the first day in the month. Set the day with the 0/I-START button. Another press of the MEMORY button takes you to the clock setting the hours blink. As earlier, set the clock with the 0/I-START button. Pressing the MEMORY button advances to the minutes setting. Set the minutes with the 0/I-START button. A final press of the MEMORY button completes the setting of the date and time. 8 The blood pressure measurement Now connect the cuff with the instrument by plugging the air plug at the tube of the cuff into the socket on the back side of the instrument (see Figs. 1 and 2). Put the cuff around your left upper arm in such a way that the tube is directed towards the bend of the elbow, with the lower cuff edge ending about 2-3 cm above the bend of the elbow. The cuff should be reasonably tight. Pressing the 0/I-START button starts the automatic measurement. The cuff inflates to approx. 190 mmhg pressure. The automatic deflation starts the measurement. After about 3 seconds, a heart icon starts blinking in the LC display. If a Korotkoff s sound is detected already at that time, it means a systolic pressure of more than 150 mmhg. Then the monitor re-inflates (to about 220 mmhg). At the end of the measurement, the result is displayed in the LCD and the cuff deflates and the measuring value is entered in the memory area. Aborting a measurement You can stop the measurement at any time by pressing the 0/I-START button. The BP monitor deflates the cuff instantly. 9