PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS



Similar documents
DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level

IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE

DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

DEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

Ra, 232 Th and 40 K analysis in sand samples from some beaches of Great Vitória, Espírito Santo, Brazil: Preliminary results.

THE RESTRUCTURING OF THE QUALITY MANAGEMENT SYSTEM OF CDTN

Simone Alberigi and Brigitte Roxana Soreanu Pecequilo

TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON

IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

NUCLEAR ENERGY EDUCATION SCENARIO AROUND THE WORLD

STUDY OF MAGNETISM IN GDTIN (T = NI, PD, CU) COMPOUNDS BY PERTURBED GAMMA-GAMMA ANGULAR CORRELATION SPECTROSCOPY

Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor

ASPECT OF SUCROSE AND ITS MONOMERS FROM SUGARCANE JUICE SUBMITTED TO DIFFERENT DOSES OF COBALT-60 IRADDIATION

Quality Assurance Testing of AtomLab Dose Calibrators

CFD SIMULATION OF IPR-R1 TRIGA SUBCHANNELS FLUID FLOW

STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)

PROCESSING OF AISI M2 HSS WITH ADDITION OF NbC BY MECHANICAL ALLOYING USING TWO DIFFERENT TYPES OF ATTRITOR MILLS

Public Funding and the Beginning of a New Era in Higher Education in Brazil

EVALUATION OF ANALYSIS METHOD STANDARDLESS BY WDXRF AND EDXRF OF ALUMINUM POWDER USED IN MTR TYPE FUEL

What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?

Dedicated computer software to radiation dose optimization for the staff performing nuclear medicine procedures

A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

DOSE RATES EVALUATION ON ENVIRONMENT AROUND ALMIRANTE ALVARO ALBERTO NUCLEAR POWER STATION AFTER ANGRA 2 NUCLEAR POWER PLANT S OPERATION

VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES

VIRTUAL INSTRUMENTATION ON MOBILE DEVICES FOR DEPLOYMENT IN NUCLEAR INSTALLATIONS

One approach to the use of the practices of CMMI- DEV V1.3 level 2 in a process of development of Embedded Systems

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

How To Test Water Penetration On Concrete Block And Mortar

The Calibration Chain: Role of BIPM,

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

A MULTIELEMENT ANALYSIS OF NATURAL RADIONUCLIDES IN AN AREA OF PERNAMBUCO - BRAZIL

Radium-223 Therapy: Handling & Radiation Safety Issues

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

Research Topics in Renewable Energies in Brazil

Evaluation of the image quality in computed tomography: different phantoms

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

Syringe Calibration Factors and Volume Correction Factors for the NPL Secondary Standard Radionuclide Calibrator.

A COMPUTATIONAL PHANTOM OF HEAD AND NECK - SISCODES

Lasers and Specturbility of Light Scattering Liquids - A Review

Procedure for Equipment Calibration and Maintenance

Study of the Natural Circulation Phenomenon for Nuclear Reactors

WebBee a Web-based Information Network on Bees

Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

centre for radiation protection

AP Physics 1 and 2 Lab Investigations

HEALTH PROFESSIONS ACT 56 OF 1974

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

The International classification for nursing practice: Participation of Brazilian Nurses in the Project of the International Council of Nurses*

MAJOR AND TRACE ELEMENTS IN GEOLOGICAL SAMPLES FROM ITINGUSSÚ BASIN IN COROA-GRANDE, R J.

Natural and Man-Made Radiation Sources

DATABASE IMPLEMENTATION TO A FLUIDIZED CRACKING CATALYTIC-FCC PROCESS

OPERATIONAL STANDARD

arxiv: v1 [stat.ml] 21 Jan 2013

RESULTS OBTAINED IN MEASUREMENTS OF Rn-222 WITH THE ROMANIAN STANDARD SYSTEM*

A RELAP5 STUDY TO IDENTIFY FLOW REGIME IN NATURAL CIRCULATION PHENOMENON

Lake at IPEN location

José Egidio M.Tondin 1, Mario O. de Menezes 1, Marina B.A.Vasconcellos 1 and João A.Osso Jr. 1 ABSTRACT 1. INTRODUCTION

PHYS 1624 University Physics I. PHYS 2644 University Physics II

THE EFFECT OF THE HDDR PROCESS ON THE PRODUCTION OF Pr-Fe-Co-B-Nb SINTERED MAGNETS

THE IMPACT OF EXCHANGE RATE VOLATILITY ON BRAZILIAN MANUFACTURED EXPORTS

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Improvement and storage of conventional mammography images using techniques of digital image processing

APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMMAND: PERMIT NUMBER: TYPE OF ACTION:

Antoine Henri Becquerel was born in Paris on December 15, 1852

CHARACTERIZATION OF COLOR CENTERS IN QUARTZ INDUCED BY GAMMA IRRADIATION

Validation and Calibration. Definitions and Terminology

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration

TRAINING PROGRAMS IN MEDICAL PHYSICS IN THE UNITED STATES

Principles of dosimetry The ionization chamber

Transcription:

2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS Elaine Wirney Martins and Maria da Penha A. Potiens Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor Lineu Prestes 2242 05508-000 São Paulo, SP ewmartins@ipen.br, mppalbu@ipen.br @ipen.br ABSTRACT The formation of good quality image in nuclear medicine services depends on several factors, including the radiopharmaceutical activity, which must be well determined by a specific apparatus, in perfect operating condition, called activimeter. Therefore, the establishment of a quality control program for measuring the radiopharmaceuticals radioactivity before being administered to the patient is crucial to the safe and effective use of the radiopharmaceuticals used in diagnostic and therapeutical procedures. Two activimeters, belonging by the Laboratório de Calibração do Instrumentos (LCI) of Instituto de Pesquisas Energéticas e Nucleares (IPEN), were evaluated: the secondary standard system NPL-CRC radionuclide calibrator, manufactured by Southern Scientific, and the work standard system CRC -15BT with traceability to National Institute of Standard and Technology (NIST). A set of standard sources of the radionuclides 133 Ba, 57 Co and 137 Cs was used as reference to perform the quality control tests. The precision, accuracy and repeatability tests were in agreement with those established by the CNEN-NE 3.05 Brazilian standard, which recommends the variation limit of up to 10%, 5% and 5% respectively. The results obtained for the reproducibility tests presented a variation always below 1,5%, which means it is within the IEC 61674 international standard recommendation of up to 3%. Activimeters response regarding linearity showed concordance between the measured activities and the theoretical curve for both activimeters. 1. INTRODUCTION Due to the large number of procedures performed with the use of ionizing radiation in medicine, it is necessary to use a substantial number of activity meters in order to cover all NMS, and concerns about the development of quality control programs as well as radiological safety are inevitable. These two factors have often been studied, and they are a requisite for the national and international standards, since their main objective is to ensure the precise measurement of radiopharmaceuticals activity before being administered to patients. The activimeter needs to be in perfect operation conditions, otherwise it may present values smaller or larger than the real one, resulting in unnecessary high doses to patients without any benefit, or repeated procedures leading to radiation exposure to the patient as well as individuals involved in the process, thus failing to meet the radioprotection principles (1,2).

The activimeter has a unique feature compared to other systems which use ionization chambers, the electronic circuit which allows response to be showed directly in units of activity (3). Several factors influence the accuracy of the dose given by activimeters. Before becoming available for sale, the activimeter, after its manufacture, is submitted to calibration measurements using a set of certified radiation sources. The manufacturer limits its initial accuracy to a range between +/- 1% and +/- 5%. Over time this accuracy can vary due to pressure changes in the gas of its chamber and electronic attraction. To ensure the proper operation of the instrument, due to these variations, regular tests were established to be applied at specific intervals. There are three types of tests which are part of a quality control program: (I) acceptance tests, (II) reference tests, (III) operational tests. (I) Acceptance tests should be performed after the acquisition of a new activimeter by NMS. Some of the acceptance tests are: physical inspection (display lighting, battery, support, wiring, etc), precision, accuracy, activity response linearity and background radiation. (II) To verify the apparatus performance, reference tests are periodically performed, as listed in Table 1, including accuracy, precision, linearity and background radiation tests. Those tests must conform to acceptance limits established by the CNEN standard (1). TABLE 1: Tests established by CNEN for quality control of activimeters. Test Precision Accuracy Linearity Reproducibility Frequency Bianual Bianual Bianual Anual Recommended sources Acceptance limits 57 Co, 133 Ba or 137 Cs 10% 57 Co, 133 Ba or 137 Cs 5% 99m Tc 20% 57 Co or 133 Ba 5% (III) Operational tests include reproducibility and background radiation tests (4). 2. MATERIALS AND METHODS In this study the following activimeters were tested: the reference system, the secondary standard, NPL-CRC radionuclide calibrator, and the tertiary system, Capintec basic CRC - 15BT. They consist of two parts: the console in which the electrometer is situated, and the ionization chamber, both by Capintec, as presented in Figure 1. The reference system is a secondary standard, model NPL-CRC radionuclide calibrator, serial number 111113, manufactured by Southern Scientific plc, with traceability to National Physical Laboratory (NPL). The system to be used as work standard is a model CRC -15BT,

serial number 180020, calibrated by Accredited Dosimetry Calibration Laboratory of Medical Radiation Research Center University of Wisconsin, with traceability to National Institute of Standard and Technology (NIST). (a) (b) Figure 1. a) Reference system: NPL-CRC, b) Work system: CRC -15 BT. For the quality control tests, a set of standard sources were used as reference, provided by the Centro de Tecnologia das Radiações (CTR) of IPEN. Three sealed sources: 133 Ba, 57 Co e 137 Cs, are presented in Figure 2. Figure 2: Set of standard Eckert & Ziegler sources. Each standard source is supplied in a 27 ml polyethylene vials with uniform distribution of the active element in 20 ml of epoxy resin, composed of a density of approximately 1,0 g/cm 3. These sources were manufactured by the German company Eckert & Ziegler Isotope Products (E&Z), with traceability certificates to National Institute of Standards and Technology (NIST). In Table 2 its characteristics are related. Radionuclides TABLE 2: Main characteristics of sources E&Z. Primary photon energy (kev) Activity (MBq) Uncertainty (%) T 1/2 (days) Reference date 133 Ba 81; 356 9.55 ± 3.0 3853.6 01/11/09 57 Co 122 206.8 ± 3.0 272.11 01/03/10 137 Cs 662 7.26 ± 3.0 110.15 01/11/09

Precision and accuracy tests were applied after stabilization of the device, 10 readings were performed with a 30 seconds time interval between them. The precision test was calculated using the percentage deviation (P) between individual activity measure (A i ) and the measurements mean ( A ) according to equation 1. ( A A P (%) = 100 x i ) A (1) In which: A = Arithmetic mean of the activity measures A i = Activity of the individual source The accuracy test (E) was calculated using the percentage deviation between the average of the activity measures ( A ) relative to the mean of the first ten measurements in the standard system with correction for the decay ( A ), given by the equation 2 (1,3,5,6). 10 E (%)= 100 x ( A A A 10 10 ) (2) In which: A = Arithmetic mean of the activity measurements A 10 = Arithmetic mean of the first ten measurements in the standard system with correction for the decay on the date in which the test was realized. The repeatability and reproducibility tests were applied equally in both activimeters to ensure the quality control of each apparatus regarding stability, using a 137 Cs source. For the linearity test a 99m Tc source was used, in a homogeneous mixture of 4 ml of saline solution into a penicilin vial with maximum volume of 20 ml and initial activity of 5.1 GBq. 3. RESULTS Precision: the maximum percentage deviation between the individual activity measure (A i ) with respect to the measured mean activities (Ā) was less than 0.32% for all the tests performed, not exceeding the recommended limit of ± 5%, thus showing satisfactory results with reliability of 95%. Accuracy: the acceptance limit is ± 10% with a reliability of 90% recommended by CNEN standard (1). All tests performed were satisfactory, not exceeding the recommended limit. In Table 3, it is possible to observe the results of accuracy test, that is the percentage deviation between mean of activity measured (10 measurements) and the standard source activity with the application of the decay correction factor.

TABLE 3: Results obtained for the accuracy test with reference sources in both activimeters. NPL-CRC CRC -15BT Ā (MBq) A (MBq) Accuracy (%) Ā (MBq) A (MBq) Accuracy (%) Bário -133 8.03 8.01 0.27 7.93 7.90 0.29 Césio -137 6.93 6.92 0.14 7.40 7.38 0.26 Cobalto 57 169.3 169.2 0.06 170.0 169.9 0.06 For the stability test a sequence of measurements was performed. 10 first measurements were employed to determine the value to be used as reference, the maximum coefficient of variation obtained was 1% (acceptable variation is 5%) (7). Figure 3 shows the reference and work system stability over time (standard deviation of the average), respectively. For the reference system, maximum variation of the results was 1% throughout the testing period and for the work system, 1.5%, both within the acceptance limits recommended by international protocols (8,9). ACKNOWLEDGMENTS Relative activity This template was adapted from the template for the American Nuclear Society Topical Meeting in Mathematics & Computations, Gatlinburg, TN, 2003 posted on the Internet. Acknowledge the help of colleagues, and sources of funding, if you wish. Relative activity Measurement number Measurement number Figure 3: Stability test using 137 Cs standard source. a) Reference system: NPL-CRC, b) Work system: CRC -15 BT. Linearity: activimeters response was accompanied by the decay of a 99m Tc source, with initial activity of 5,01 (GBq), as shown in Figure 4. Date and time of the measurements were registered in time intervals of one hour between them for a period of 56 hours.

Activity (Gbq) Linear (T) Linear (CRC) Linear (NPL) Time (h) Figure 4: Linearity test The reference system (NPL-CRC ) and the work system (CRC -15BT) presented results within the acceptance limit recommended by the manufacturer s manual (± 10%), so both activimeters revealed linear responses between the values of measured activity which is in accordance with the experimental values. 4. CONCLUSIONS For the precision, accuracy and reproducibility tests, the Brazilian standard for nuclear medicine services, CNEN-NE 3.05, was used as reference, which recommends variation limits of up to 10%, 5% and 5% respectively. Results obtained for repeatability presented a variation lower than 1,5%, within the recommendation of up to 3% of IEC 61674 international standard (7). Activimeters response regarding linearity showed concordance between measured activities and the theoretical curve for both activimeters. Therefore, the quality control program implemented for the LCI activimeters demonstrated that both instruments are in good operation condition and the CRC - 15 BT apparatus can be used as work standard, employing the appropriate corrections. ACKNOWLEDGMENTS The authors thank to Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), to Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), to Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) and to Ministério de Ciência e Tecnologia (MCT), Project: Instituto Nacional de Ciência e Tecnologia (INCT) em Metrologia das Radiações na Medicina, the partial financial support.

REFERENCES 1. Comissão Nacional de Energia Nuclear. CNEN-NN-3.05. Requisitos de radioproteção e segurança para serviços de medicina nuclear. Resolução CNEN-10/96. Diário Oficial da União, (1996). 2., A. C. M. Bessa, A. M. Costa, L. V. E Caldas. Levantamento do controle de qualidade de calibradores de dose de radiofármacos em serviços de medicina nuclear na cidade de São Paulo. Radiol. Bras. v. 51, p. 115-118 (2008). 3. A.M. Costa. Métodos de calibração e de intercomparação de calibradores de dose utilizados em serviços de medicina nuclear. São Paulo. Dissertação (Mestrado) Universidade de São Paulo. IPEN/USP (1999). 4. International Atomic Energy Agency. Quality control of nuclear medicine instruments. TecDoc-602, (1991). 5. S.G.P. Cecatti. Desenvolvimento de sistemas Tandem de activímetros e estabelecimento de dosimetria beta em serviços de medicina nuclear. São Paulo. Tese (Doutorado) Universidade de São Paulo. IPEN/USP (2004). 6. M.C.F Fragoso. Implantação de um programa de intercomparação de medidas de atividade de radiofármacos utilizados em serviços de medicina nuclear no Nordeste. Recife. Tese (Mestrado) Universidade Federal de Pernambuco (2010). 7. International Electrotechnical Comission. Medical electrical equipment Dosimeters with ionization chambers and or semi- conductor detectors as used in X-ray diagnostic imaging. Geneva. IEC 61674 (1997). 8. National Physical Laboratory. Protocol for establishing and maintaining the calibration of medical radionuclide calibrators and their quality control. Middlesex, United Kingdom. A National Measurement Good Practice Guide, n. 93, NPL (2006). 9. Laboratoire National Henri Becquerel. Guide d utilisation et de controle qualité dês activimètres. França. Societé française de radiopharmacie (2006).