Spectrophotometer - Milton Roy Spectronic 21D or equivalent.



Similar documents
NITRIC OXIDE and NITROGEN DIOXIDE 6014

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Measuring Protein Concentration through Absorption Spectrophotometry

Lab 2. Spectrophotometric Measurement of Glucose

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

The Determination of an Equilibrium Constant

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter

Phenolphthalein-NaOH Kinetics

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Running protein gels and detection of proteins

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

HS 1003 Part 2 HS 1003 Heavy Metals Test

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA

Quaternary Ammonium Compounds DOC

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

THE ACTIVITY OF LACTASE

Chem 131A: Absorbance of Riboflavin

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

ab Protein Sumoylation Assay Ultra Kit

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Phosphorus, Reactive (Orthophosphate)

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

DR/4000 PROCEDURE. CHLORINE, Free

Chromatin Immunoprecipitation

HiPer Ion Exchange Chromatography Teaching Kit

Determination of calcium by Standardized EDTA Solution

International Journal of Research and Reviews in Pharmacy and Applied science.

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

TOTAL PROTEIN FIBRINOGEN

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Uscn Life Science Inc. Wuhan

Color, True and Apparent

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #:

6 H2O + 6 CO 2 (g) + energy

Color, True and Apparent

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

ENZYME KINETICS ENZYME-SUBSTRATE PRODUCTS

ph and Acidity in Wine and Fruit Juice

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1

U.S. Department of Labor Occupational Safety & Health Administration

Lab #11: Determination of a Chemical Equilibrium Constant

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

A Potentiometric Analysis of Fluoride Ion in Toothpaste

Human Free Testosterone(F-TESTO) ELISA Kit

ph and Acidity in Tomato Sauce

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

SUCRALOSE. White to off-white, practically odourless crystalline powder

MTT Cell Proliferation Assay

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer

Catalase. ***You will be working with hot water, acids and bases in this laboratory*** ****Use Extreme Caution!!!****

18 Conductometric Titration

Transformation Protocol

Quantifying Bacterial Concentration using a Calibrated Growth Curve

Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit

PROTEINS (LOWRY) PROTOCOL

Determination of the Mass Percentage of Copper in a Penny. Introduction

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Acid Base Titrations

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

INTERNATIONAL OLIVE COUNCIL

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

KODAK Developer System Cleaner and Neutralizer

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

MLX BCG Buccal Cell Genomic DNA Extraction Kit. Performance Characteristics

The Determination of an Equilibrium Constant

MOISTURE (Karl Fischer, Buffered)

Hydrogen Peroxide Cell-Based Assay Kit

Stage 1 Desired Results

TECHNICAL BULLETIN. FluoroTag FITC Conjugation Kit. Product Number FITC1 Storage Temperature 2 8 C

Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate

Method 202 Dry Impinger Method for Determining Condensable Particulate Emissions from Stationary Sources. By: Paul Ottenstein

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit

A Beer s Law Experiment

TECHNICAL REPORT STUDY CHEMICAL PROPERTIES OF RATTAN SHOOT FROM PLANTATION IN THAILAND

NCL Method ITA-6. Leukocyte Proliferation Assay

Measurement by Ion Selective Electrodes

Classic Immunoprecipitation

Solubility Product Constants

EFFECT OF SALT ON CELL MEMBRANES

Experiment 7: Titration of an Antacid

ExpressArt Bacterial H-TR cdna synthesis kit. With extreme selectivity against rrnas

Qualification Study CHO 360-HCP ELISA (Type A to D)

All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &

Transcription:

3M COMPANY OCCUPATIONAL HEALTH AND ENVIRONMENTAL SAFETY DIVISION DETERMINATION OF FORMALDEHYDE VAPORS IN AIR USING 3M 3721 FORMALDEHYDE MONITORS May, 2002 SCOPE This procedure covers the method of collecting and analyzing samples to determine the amount of formaldehyde present in the air using the 3M 3721 Formaldehyde Monitor. The calibration curve covers the 0-15 microgram range. SUMMARY OF THE METHOD In the monitor, formaldehyde vapors are adsorbed on bisulfite impregnated paper and desorbed with formaldehyde-free distilled water. Aliquots are reacted with chromotropic acid in the presence of sulfuric acid to form a purple monocationic chromogen. 1 The absorbance of the colored solution is read in a spectrophotometer at 580 nanometers (nm) and is proportional to the amount of formaldehyde in the solution. APPARATUS Spectrophotometer - Milton Roy Spectronic 21D or equivalent. REAGENTS/SUPPLIES The following or equivalent reagents and supplies are used: 3M 3721 Formaldehyde Monitor Chromotropic Acid Solution - Dissolve 0.25 g of 4,5-dihydroxy-2,7-naphthalenedisulfonic acid disodium salt dihydrate (Aldrich 21,327-6 or equivalent) in 25 ml of formaldehyde-free distilled water. Make solution fresh each day. Previously used Acros 40525-0100, formerly Kodak 230, is no longer available, but existing supplies may be used. Sulfuric Acid - Concentrated, reagent grade. Distilled Water - Formaldehyde-free. 1% Sodium Bisulfite Solution - Dissolve 2.5 g of sodium bisulfite (Baker 3556-01 or equivalent) in 250 ml of distilled water. Standard Formaldehyde Solution - 25 ul of 37% formaldehyde (Aldrich 25,254-9 or equivalent) is diluted to 10 ml with 1% sodium bisulfite solution. This solution is used to prepare the calibration curve - each microliter is equivalent to one microgram of formaldehyde.

SAMPLING The monitor is removed from the package and assembled according to the instructions in the package. The following information should be recorded: 1. Monitor number 2. Date of exposure 3. Employee or area identification 4. Temperature and relative humidity 5. Start and stop times 6. Any potential interferences, such as phenol 7. Any comments or unusual circumstances When personal sampling is being performed, the monitor should be attached near the breathing zone. For area sampling, the monitor should be placed so that at least a 25 fpm face velocity is maintained. The monitor should not be placed in a corner or along a wall where stagnant air may exist. The 3721 monitor is designed for 8-hour sampling and is not recommended for 15-minute STEL sampling. After sampling, the retaining ring and white barrier film are removed and discarded. The two monitor sections are separated and the bottom section with the clip is discarded. The clear elution cap and primary cup are snapped onto the top section. The ports should be securely sealed with the plugs. A blank control sample should be prepared at the monitoring site and submitted with the samples. SAMPLE ANALYSIS (a) Elution Both ports of the elution cap are opened and 3 ml of formaldehyde-free distilled water is added to each monitor through the center port using a repipet or syringe. The ports are immediately resealed and each monitor allowed to elute for 30 minutes with occasional gentle agitation. Transfer a 2 ml aliquot of the eluate to a test tube or vial for color development. (The amount of the aliquot may be varied to be sure that each sample absorbance will be within the calibration curve. Dilute the aliquot to 2 ml with 1% sodium bisulfite solution.) (b) Color Development Add 1.0 ml of chromotropic acid solution to each sample and mix well. Carefully add 5 ml of concentrated sulfuric acid slowly with mixing. CAUTION: Take proper safety precautions such as goggles, gloves and apron when handling concentrated sulfuric acid. Allow the samples to cool to room temperature and measure the absorbance at 580 nm (medium sensitivity setting) using 1 cm cells or 1 in test tube cuvettes. Use distilled water in the reference cell. A reagent blank is carried through all the steps of the sample analysis. Subtract the absorbance of the reagent blank from that of the standards and samples and refer to the calibration curve to determine the micrograms of formaldehyde present. CALIBRATION CURVE 2

To a series of test tubes or vials, carefully add 1.0, 3.0, 5.0, 10.0 and 15.0 microliters of standard formaldehyde solution equivalent to 1.0, 3.0, 5.0, 10.0, and 15.0 micrograms of formaldehyde. Adjust the volumes to 2 ml with 1% sodium bisulfite solution. Develop the color and measure the absorbance as described above. Carry a reagent blank through all the steps and subtract its absorbance from all the standards as well as the samples. Prepare a calibration curve by plotting the corrected absorbance (x axis) vs. micrograms of formaldehyde (y axis). Determine the slope and intercept for the calibration curve. Appropriate quality control samples are run with each set of standards and samples. CALCULATIONS (a) W = [I + S(A s - A b )](3/A) where: W = micrograms of formaldehyde found A s = absorbance for sample A b = absorbance for reagent blank S = slope of the calibration curve I = intercept of the calibration curve A = aliquot taken in ml (usually 2 ml) (b) where: (W - B)(MV)(1000) C = -------------------------- (SR)(r)(t)(MW) C = concentration of formaldehyde in air in ppm W = micrograms of formaldehyde found B = micrograms of formaldehyde in the monitor blank MV= molar volume at given temperature and pressure (24.45 L/mole at 25 o C and 760 mm Hg) MW= molecular weight of formaldehyde (30 amu) SR = sampling rate for formaldehyde (61.4 cm 3 /min) r = recovery (1.0) t = sampling time in minutes The above expressions calculate the time-weighted-average concentrations at a sampling temperature of 25 0 C and pressure of 760 mm Hg. When sampling at other environmental conditions, the concentration need only be corrected for variations in temperature. If the temperature correction is desired, the time-weighted-average concentration can be calculated by mutiplying the concentration calculated above by the temperature correction factor (CFt.) Sampling Temperature Correction Factor ( 0 C) ( 0 F) (CFt) 44 111.97 3

37 99.98 31 88.99 25 77 1.00 19 66 1.01 13 55 1.02 7 45 1.03 2 36 1.04-3 27 1.05-8 18 1.06 As shown in the above table, every 10-11 degrees above or below 77 0 F requires a one percent correction to the calculated time-weighted-average concentration. For every 10 0 F increment above 77 0 F, there is a decrease of about 1% in the time-weighted-average concentration. Below 77 0 F, there is an increase in the concentration. LIMIT OF QUANTITATION The routine limit of quantitation (LOQ) is about 1 ug. This results in an air concentration of 0.03 ppm for an 8-hour sample. STORAGE Analysis of spiked monitors after 13 weeks of storage at room temperature showed a recovery of 100.5%. DISCUSSION The precision of the analytical method is reported to be +/-3%. 2 The precision of the sampling and analytical method has been shown to be +/- 6% in our laboratory. The bias from the midget impinger method has been found to be -3% with a resulting accuracy of 15%. Both the metatrioxane and paraformaldehyde as well as dimethoxymethane represent positive interferences. Large amounts of phenols cause a negative interference which may completely mask color formation. For monitor samples containing phenol, color development may be accomplished by using 1 ml of 10% chromotropic acid instead of 1% chromotropic acid to decrease the phenol interference, however, the monitor does not collect phenol efficiently and therefore any potential interference is expected to be minimal. 3 REFERENCES 1. Methods of Air Sampling and Analysis, Third Edition, J. P. Lodge, Ed., 274 (1989). 4

2. NIOSH Manual of Analytical Methods, Method No. 3500, Fourth Edition, 1994. 3. Johnson, G. D., 3M Corporate Research Laboratory, Personal Communication. 5