COMD 3700 Lesson 10 Speech Audiometry II

Similar documents
SPEECH Biswajeet Sarangi, B.Sc.(Audiology & speech Language pathology)

Hearing Tests And Your Child

Hearing Tests And Your Child

Audiometry and Hearing Loss Examples

Your Hearing ILLUMINATED

PURE TONE AUDIOMETRY Andrew P. McGrath, AuD

The Disability Tax Credit Certificate Tip sheet for Audiologists

Audio Examination. Place of Exam:

Understanding Hearing Loss

Pure Tone Hearing Screening in Schools: Revised Notes on Main Video. IMPORTANT: A hearing screening does not diagnose a hearing loss.

Noise at the Work Site

Noise at work a guide for health and safety representatives

Functional Auditory Performance Indicators (FAPI)

Dr. Abdel Aziz Hussein Lecturer of Physiology Mansoura Faculty of Medicine

Portions have been extracted from this report to protect the identity of the student. RIT/NTID AURAL REHABILITATION REPORT Academic Year

8.Audiological Evaluation

Parents views: A survey about speech and language therapy

The neurobiology of energy healing...a hands on approach.

S-822. Noise and Hearing Conservation RISK MANAGEMENT

Unilateral (Hearing Loss in One Ear) Hearing Loss Guidance

Hearcentres Guide to Hearing Aid Terminology

Assessing Speaking Performance Level B2

Getting Started Kei Te Timata

7.9 Hearing Loss and Hearing Aids

Macroaudiology a Working Model of Hearing Presented at XXI International Congress of Audiology Morioka, Japan, 1992 R.

Introduction Bone Anchored Implants (BAI), Candidacy and Pre-Operative Testing for Adult Patients

What are the causes of presbycusis? What can be done? How can I communicate with someone who has a hearing loss? How does hearing work?

So you ve had your hearing tested. What s next?

ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING

California State University, Fresno Department of Communicative Disorders and Deaf Studies CDDS 128. Observations in Audiology

So, how do we hear? outer middle ear inner ear

Hearing Loss in Geriatric Primary Care Mary Ann Forciea MD Josh Uy MD

Presbycusis. What is presbycusis? What are the symptoms of presbycusis?

Education Adjustment Program (EAP) Handbook

HEARING SCREENING: PURE TONE AUDIOMETRY

Once your baby has good head control and can turn towards something interesting, a more advanced behavioural procedure can be used.

Grade 8 English Language Arts 90 Reading and Responding, Lesson 9

COMPUTER SUPPORT STAFF TIPS FOR DISABILITY ETIQUETTE

Independent samples t-test. Dr. Tom Pierce Radford University

Link: University of Canberra

The NAL Percentage Loss of Hearing Scale

AM TRANSMITTERS & RECEIVERS

REGULATIONS FOR THE DEGREE OF MASTER OF SCIENCE IN AUDIOLOGY (MSc[Audiology])

Information for teachers about online TOEIC Listening and Reading practice tests from

The Chord Book - for 3 string guitar

Addressing Quietness on Units Best Practice Implementation Guide. A quiet environment is a healing environment

Noise and Hearing Protection

Everybody has the right to

Identifying dyslexia and other learning problems using LASS

The Problem with Faxing over VoIP Channels

Hearing Loss and Older Adults

highlight some of the problems a deafblind or sensory impaired person might have in getting tested;

NEW YORK STATE MEDICAID PROGRAM HEARING AID/ AUDIOLOGY SERVICES PROCEDURE CODES

The Top 5 Things You Should Know Before Buying Hearing Aids

Samuel R. Atcherson, Ph.D.

There are many reasons why reading can be hard. This handout describes

Learners Who are Deaf or Hard of Hearing Kalie Carlisle, Lauren Nash, and Allison Gallahan

Ten Strategies to Encourage Academic Integrity in Large Lecture Classes

The Doctor-Patient Relationship

Teaching Public Speaking Online

CONVENTIONAL AND DIGITAL HEARING AIDS

BBC Learning English Talk about English Business Language To Go Part 2 - Induction

Hear Better With FM. Get more from everyday situations. Life is on.

Very simple: the phrase 'can I' followed by the verb. But what verb form comes after the phrase 'can I'? Listen to these two examples of asking.

OAK GROVE SCHOOL DISTRICT. Hearing Conservation Program

DEMENTIA SEVERITY RATING SCALE (DSRS)

A Hearing Aid Primer

Barriers to Communication

Adolescents and Hearing Impairment

NEW YORK STATE MEDICAID PROGRAM HEARING AID/ AUDIOLOGY SERVICES PROCEDURE CODES

Noise: Impact on Hearing; Regulation

COCHLEAR NERVE APLASIA : THE AUDIOLOGIC PERSPECTIVE A CASE REPORT. Eva Orzan, MD Pediatric Audiology University Hospital of Padova, Italy

PURE TONE AUDIOMETER

Songwriting. Geniuses. for. 25 Tips for the Genius in Everyone. By Gene Burnett

Guide for families of infants and children with hearing loss

62 Hearing Impaired MI-SG-FLD062-02

MINUTE TAKING. All material copyright of Lindsay Wright This pack is for sample purposes only, and not for re-use

Stanford University. Hearing Conservation Program

Cell Phones and Cell Phone Plans for Seniors

Summary Our science project is an investigation into the varying levels of damage that different types of earphones and mp3 players cause to hearing.

Audiology as a School Based Service. Purpose. Audiology (IDEA 2004) Arkansas SPED Regulations. IDEA 2004 Part B

SOUTHERN UTAH UNIVERSITY S WRITTEN HEARING CONSERVATION PROGRAM September 24, 2015

Nonverbal Factors in Influence

Veterans UK Leaflet 10. Notes about War Pension claims for deafness

GMAT SYLLABI. Types of Assignments - 1 -

RIGHTS OF DEAF AND HARD OF HEARING PEOPLE IN MEDICAL SETTINGS

HEARING. With Your Brain

WHAT IS NLP: A MODEL OF COMMUNICATION AND PERSONALITY

Section 4. Hearing loss and hearing tests

How to become a successful language learner

The Brain, Prosody, and Reading Fluency

Directions for Administering the Graded Passages

HYPERSENSITIVITY OF HEARING (Hyperacusis, misophonia, phonophobia and recruitment - decreased sound tolerance)

Confirmation of Diagnosis of Disability (To determine eligibility under section 18(2)(b) of the Income Tax Act, 1962 (as amended))

STEP 5: Giving Feedback

Audiology Services. Carolyn Dando Audiology Services Manager South Warwickshire NHS

DAILY BIOLOGIC CHECK OF AUDIOMETERS

EMILY WANTS SIX STARS. EMMA DREW SEVEN FOOTBALLS. MATHEW BOUGHT EIGHT BOTTLES. ANDREW HAS NINE BANANAS.

Laugh at your fear of Public Speaking ~ Give an Icebreaker Speech!

Expanding Performance Leadership in Cochlear Implants. Hansjuerg Emch President, Advanced Bionics AG GVP, Sonova Medical

Transcription:

COMD 3700 Lesson 10 Speech Audiometry II COMD 3700 for Distance Education. This is lesson 10 on speech audiometry. This is the second lesson on speech audiometry. This lesson will cover pages 139-142, 147 (recording speech-recognition test results) and 149-160 in Chapter 5 of your textbook. As we discussed in lesson 9, there are 5 Factors in Speech Audiometry: Test Equipment, Test Environment, The Patient, The Clinician and The Test Procedure. We will continue to discuss the testing procedure in this lesson. The speech testing we will discuss in this lesson are the MCL, UCL and WRS or speech recognition testing, not to be confused with the speech recognition threshold. Let's discuss most comfortable loudness level. The purpose of the MCL is to determine the best level for the patient to hear and understand speech. The MCL gives important information both with regard to the patient and to the fitting of a hearing instrument. In a patient with normal hearing, most comfortable loudness level is somewhere from 40 to 55 db HL. The stimulus for finding this is cold running speech. As discussed in lesson 9, cold running speech is an informative but non-emotional series of sentences, recorded at a constant level. Cold running speech carries a lot of redundant information. The patient hears and understands the subject without realizing that each individual word may not be clear. Because presentation of this speech may be comfortable at any sensation level above SRT, a bracketing method determines a more accurate MCL. The reason for this is that the MCL is a little hard for a patient to identify. It may be hard to define. A person could be 5 or 10 db on the high or low side. We'd like to narrow that a bit and try to get a finite db level that represents MCL. To accommodate this, we use the bracketing method.

First, we instruct the client how to behave and respond. The instructions to the client may be something like this, Now you are going to hear continuous speech. You do not have to repeat any of the words. I want to find the most comfortable level for you to listen. I will increase the volume slowly, so you can listen at different levels. You can help me by pointing your thumb up if you want it louder, or down if you want it quieter. Please let me know when we find the volume you prefer. Do you understand? So we start our bracketing level by first presenting a stimulus to them that we know will be below their comfortable listening level or too soft. This will be at a hearing level slightly above the SRT. Some audiologists start at a 40 db sensation level or 40dB above SRT. Once you start speaking, the patient will raise their thumb that the sound needs to be increased. We're not going to increase it by five db or ten db and then say a word like we did in finding SRT. Instead we'll continue with cold running speech and quickly increase in 5 db steps until the speech is being presented above their threshold but just below their uncomfortable loudness level. They will put their thumb down for softer. Then decrease the speech in 10 db steps until they say it is comfortable. You can go up 5 or down 10 until you find the comfortable level for them. The forced choice method will help determine an accurate MCL. To do this you always advance the attenuator dial to a point a little above where the patient reports speech as comfortable. While really listening quietly, realization may come to the patient that speech can be louder and perhaps more comfortable. Continue to increase slowly in 5 db steps until the patient decides that speech was better lower, then return to that level. This level is noted on the audiogram as the MCL. Have patient think about how they d listen to the TV. Do they want the volume louder or softer?

Next we will determine the uncomfortable listening level or UCL of the patient. The UCL is not the threshold of pain; it is a level where speech becomes uncomfortably loud. The purpose is to determine the point where no further amplification will be permitted in a hearing instrument or accepted by the patient. For patient with normal hearing, UCL is usually 100-110 db. 110 db is usually the maximum level we deliver to any client. Levels exceeding 110 db have the power to damage hearing. Usually cold running speech is used to find UCL. We talked about running speech as being something along "Mary had a little lamb. Its fleece was white as snow. And everywhere that Mary went, her lamb was sure to go." An important point that I want you to remember is that hearing losses do not necessarily extend the UCL. That means you'd assume if a person had normal hearing and the UCL is 100-110 db, you'd assume someone with a hearing loss would have a UCL of 120 to 130 db. This is not necessarily true. People with sensorineural hearing losses in fact have many times the opposite condition where the uncomfortable loudness level is lowered. So, instead of being able to tolerate 100-110 db, they may be able to only tolerate 80-90 db so their ceiling is lowered. The UCL is also known as: TD: Threshold of Discomfort, Tolerance Level and LDL: Loudness Discomfort Level. So an example of patient instructions would be, Now I want to find a level where sound starts to bother you, not where it is just a little too loud. I want to find where it is uncomfortable for you, but not painful. I don t want you to hold your ears in pain. Please raise you hand or stay stop whenever the sound starts to bother you to the point where you couldn t listen to it or when exceeding the loudness level would be intolerable. We begin by delivering the speech at a comfortable loudness level. We'd then begin extending the db or the intensity, making the cold running speech more and more intense. They'd raise their hand eventually. You don't just stop and turn it off when they say that's as loud as they tolerate. You would then take the hearing level dial back down and increase it again back to an uncomfortable level, noting the highest level they could tolerate and record that as their uncomfortable loudness level.

Another use for the UCL is to help define the dynamic range of the patient. This is also known as the range of comfortable loudness (RCL). The dynamic range is the range of usable hearing, between the SRT and UCL. This is calculated by subtracting the SRT from the UCL. So, for example for patients with normal hearing, their dynamic range may be 100 db, or 120dB (UCL) minus 20 db (SRT). Usable hearing takes place within this dynamic range, or area of comfortable loudness. If the frequency and intensity of a sound are inside this area, we readily hear it. Any sound that falls below or outside this area remains unheard. So if you present the speech at a stimulus lower than the SRT, you're not going to function very well. If sounds are more intense than the UCL you will be uncomfortable. So, a person with a dynamic range of 50 db has no problem listening to recorded music, since most albums are produced at about a 35 db range from the faintest to loudest sound. The volume can be set at a comfortable level where the soft sounds can be heard and not bothered by the louder ones. This person has more difficulty with a live orchestra, where the dynamic range would be 70 db or greater. Sitting in the back, soft sounds would be missed, whereas by moving closer, the person would be bothered by the loud sounds. Now we will discuss speech recognition testing. There's a distinction between a threshold for speech and ability to understand the speech that is heard. Like I mentioned before there's a lot of complaints by patients with hearing impairments saying, I can hear speech but I can't understand it or Words are not clear, speech sounds muffled or distorted or I mistake one word for another. The common theme here is that the speech heard by the patient is lacking in intelligibility. This problem is experienced in terms of inaccurately receive messages or reduced clarity. So that's why we have speech recognition testing.

Speech recognition testing is also referred to as Word Recognition Testing. This to me is easier to understand. But the book refers to it as speech recognition testing. We have discussed the speech recognition threshold testing so I want you to realize that there's a difference between the SRT and what is referred to as speech recognition testing. This type of testing can also referred to as Word discrimination testing or Speech Discrimination (SD) testing. The term discrimination testing is older, but still popular. You will often hear an audiologist refer to this testing as discrim testing, as in I did SRT and discrim. The measurement of speech discrimination is referred to as: Speech recognition score (SRS), Word recognition score (WRS) or the Speech discrimination score. The purpose of speech recognition testing is the quantitative determination of a patient s ability to discriminate speech. We need to determine how well the patient hears and understands speech. This helps the audiologist in several ways: It determine the extent of the speech-recognition difficulty, it aids in diagnosis of the site of the disorder in the auditory system, it assists in the determination of the need for and proper selection of hearing instruments and it helps the clinician make a prognosis for the outcome of treatment efforts.

So although there are many terms to define this testing, the speech intelligibility testing can be tested clinically in a very straightforward manner. The most common approach is to present the patient with a list of phonetically balanced (PB) test words. Phonetically balanced just means that the percent of time any given sound appears on the list is equal to its occurrence in the English language. There are several different lists that are available for testing. Most commonly used are the CID Auditory Test W-22 and the NU-6 lists. CID stands for Central Institute for the Deaf. Hirsch and others brought these lists together in 1952. They put fifty words in a list and had four lists. There are six versions of each list with the words in a different order. These lists provided an adequate number of words for finding word recognition score for a person. So each list contain 50 phonetically balanced words. The NU-6 lists were developed at Northwestern University. There are 4 50-word lists scrambled into 4 randomizations. Either list is acceptable. For example, one Speech Discrimination list includes words such as: "are," "bar," "clove," "fraud," "hive," "nook," "pants," "rat," "slip," "there," "use," "wheat" and 38 others. There are other lists that are used for specific purposes, such as testing children, but we will not be discussing them in this course. All of the words are one-syllable words. They are designed to be common words that should be within the vocabulary of the average person. There are a variety of speech recognition test materials, including CID Auditory Test W-22 and NU-6 provided in the Professional Resources section on the CD included with your textbook. Although it is recommended that the full 50-word list is used in testing. Sometimes only a ½ list is used. It is permissible to use a half list if an individual's score is greater than 90% on the first 25 words (misses no more than 2 out of 25 words). Sometimes it may be necessary to administer more than one list. The audiologist does NOT familiarize the person with the words. A new list of words has to be used each time the test is conducted (on the same day). The use of a carrier phrase is advisable. The goal is for the words to have natural inflection. The test item should actually be slightly lower in intensity than the say the word carrier phrase portion. The inflection should drop slightly at the end. For instance, say out loud This is a book. Your pitch and loudness probably dropped slightly when saying book. This same intonation pattern should be used when you say, Say the word carve. You don t want to use a rising, singsong pitch when you present the test item. This is unnatural and will create poorer scores if the patient has a high-frequency hearing loss. It also sounds unprofessional. Because live voice testing by individual audiologists creates variability in the test presentation, it is preferable to use recorded materials. When recorded materials are used, if patient performance differences from one test date to another, it s easier to interpret that those difference are because of the patient rather than because of the person administering the test.

The test presentation level can vary based on what the audiologist is trying to determine. The standard method is to set the presentation level at a 40 db sensation level (SL) above the SRT. Some clinics may also use 30 db SL. Remember that the sensation level will vary from patient to patient depending on the degree of their hearing loss. There was some confusion about the sensation level after the last exam, so let me explain it again. Let's take three patients. Patient #1 has an SRT of 10 db. So the softest speech they could hear was 10 db. Patient #2 has an SRT of 20 db. Patient #3 has an SRT of 30 db. So we have three patients here with three different speech reception threshold levels. We want to deliver a word recognition score to each of these patients that will be perceived at the same level. We use 40 db SL as the presentation level for our clinic. So when a person comes in to our clinic, we'll administer our word recognition level at 40 db SL. So when we administer the test to a patient #1 (with an SRT of 10 db) the test presentation level would be 50 db HL. Then patient #2 (with SRT of 20 db), we'd deliver the test to them at 60 db. We would present the words to patient #3 (SRT of 30 db) at 70 db HL. We are adding 40 db to the SRT of 30 db. That is why we administer the test at 70 db. So although it is measured at 70 db HL, to the patient it is perceived as 40 db or 40 db sensation level. Once the level has been established for each patient, the intensity is not changed during the test. The intensity is left at that SL level. What happens is we now have each patient receiving the test at the same 40 db SL so each patient has no advantage over any other. So hopefully you now understand how we found the 40 db SL for each patient. Some patients with moderate sensorineural losses or worse probably have significant recruitment (can t tolerate loud sounds), and would find even 30 db SL too loud. If these levels are uncomfortably loud for the individual then testing should be performed at the most comfortable listening level (MCL) of the patient. The textbook also discusses performing the testing at a minimum of two levels. The first level is presented at 5 to 10 db above the patient s MCL. The second level is presented at a higher intensity (90 db). You can read more about this on page 151. If the audiologist wishes to know how well a person understands words when they are presented at a conversational loudness, they may present the test items at a level of 45 to 50 db HL. (All words are presented at one intensity.) If the audiologist wants to know how well the person can do when speech is made comfortably loud, the testing can be conducted at MCL.

The patient instructions are very straightforward and may be something like, You are going to hear a series of words. Please repeat each of the words you are asked to say. If you have difficulty understanding the words, it helps me if you guess. I have a better idea what part of the word was difficult for you. If the volume is not comfortable for you, or you would like the words louder or quieter, please let me know. Do you understand? So far we have been expressing the speech scores and speech test results as finite db levels such as: SRT is 30 db, the UCL is 90 db, the dynamic range is 55 db, the MCL is 35 db. With word recognition we'll talk in terms of percentage. So we'll report an individual's word recognition score as 48% or 92%. The percentage of the test words correctly repeated by the patients is recorded. Each correct response is worth 2% if using a 50-word list and 4% if using a 25-word list. If live voice testing is used then you have a form where you record next to each word if they got it correct or if they missed it. When using recorded material on a computerized audiometer, then the list of words is on the computer and after each word you indicate whether or not the patient responded correctly. This is the picture on the slide. So, If 50 words are given, and the patient misses 11 items, what is the percent correct score? 78% because each error is worth 2 points when 50 words are given. If 25 words are given, and the patient misses 6 of the words, what is the percent correct score? 76% because 25 words are given, each error counts as 4% off (since 4 X 25 = 100). The list used and the level at which the testing was performed is also noted on the audiogram.

Once the word recognition testing is complete, the results provide a relative indicator of the difficulty the person has understanding speech. The WRS suggests the approximate degree of difficulty the individual will experience understanding speech for each ear in a quiet listening condition. So, what is a "good" score? How do you interpret a word recognition score? The chart shows the general word recognition ability based on the percentage of words that the patient repeated correctly: 92-100% Excellent 82-90% Good 72-80% Fair 52-70% Poor 22-50% Very Poor 0-20% Extremely Poor However, there are some things to keep in mind. Normal scores are near 100%, if the stimulus is presented at 30 db SL or above, at MCL or above. The hearing impaired probably will not have a normal score, but will miss some items. When interpreting the word recognition score, it is important to consider the presentation intensity. A normal hearer would score poorly if tested at 5 db SL. A score of 80% would not mean the person hears 80% of words correctly in everyday life. If the testing had been conducted at the patient s MCL of 80 db HL (which is much louder than conversational speech), they may have done much better. So, the intensity of presentation must always accompany the report of the percent correct. It is simply not true that a score of X% means that a person hears that percentage of speech. Even if the testing was conducted at a conversational loudness, speech understanding is different from understanding single words presented in quiet in a test room. Speech has linguistic redundancy the listener does not need to hear each word in order to understand the message. Word recognition testing uses single words, not contextual speech. Also, note that testing is conducted in a quiet test booth. The everyday world is usually rather noisy. Speech is usually only 8 db louder than the background noise! So speech recognition tests are helpful in diagnosis, but are far from perfect for predicting real-world communication.

The speech recognition test results also alert the audiologist to patient who has unusually poor (or unusually good) scores. Very poor scores are a sign of possible retrocochlear involvement (which we haven t discussed yet), and further testing is needed. Low scores when testing is conducted at a comfortable level indicate that the person may have limited benefit from hearing aids. For a patient with normal hearing the score should be 90% or better. If they have a cochlear disorder the score is usually consistent with the degree of hearing loss; the greater the loss, the poorer the WRS. If they have a retrocochlear disorders the score is usually considerably poorer than expected based on the degree of hearing loss. Another use for speech recognition testing is to assess the PI function; also known as the performance intensity function and the PB max. When assessing the PI function, word recognition testing is repeated after increasing the presentation level to a higher intensity level. Normally, the word understanding would remain essentially the same, if not improve. In some cases, the word understanding will decrease and be abnormal. This is a sign of possible retrocochlear hearing loss. This means that there may be some damage beyond the cochlea, such as lesions in the higher auditory centers. When the ear with a damaged nervous system is stimulated with a very loud sound, the nerves may be unable to handle the increased signal load. Consider this analogy. If you injured your arm, you might be able to lift a small weight, but would be unable to lift a heavy weight. Similarly, the damaged VIII nerve may be able to transmit a conversational level speech signal, but tire out when faced with the demand of sending strong, sustained messages. So, we need to review the PI function chart, which is also located on page 133 of your textbook. On the chart are the db levels, the sound pressure level. The side is the word recognition scores or your performance. If you get 100 up the side along the ordinate, you'd have 100 percent word recognition or sixty percent word recognition score. Along the bottom is the intensity you received the speech material. You could have received it at 30, 40, or 60 db. This performance, that's the ordinate, intensity is along the abscissa. So look at

the results for #4. This is word recognition performance. This person has an obvious hearing loss because they're only getting about 15% of the words correct at 30 db. As we increase the intensity of the presentation level, as we go from 40 db to 50 db to 60 db, the performance is improving. But when we get to 75 db, all of a sudden, we reach a performance of 60%. If we increase the intensity to 80 db orr90 db, and on, the performance decreases. The peak we look at is the PB max or the maximum phonetically balanced score using PB word list. If you see PB max written, you know we've done a word recognition score on an individual, and they've reached a maximum level. If we continued to increase intensity, word recognition score diminishes. This person on curve four has a PB max of about 75 db. So the point where the scores begin to decrease is known as the rollover effect. This occurs in some ears when there are lesions in the higher auditory centers. We won t go into that any further in this course. But I just want you to be aware of another function of word recognition testing. It is important to remember that each of the speech tests help the audiologist to verify the accuracy of the audiogram. For example, the SRT verifies the PTA. So when testing, you need to ensure that each of your tests are supporting the audiogram, the type of loss, the case history, the tympanogram and the visual inspection of the ear. All of these tools are used to ensure that the correct results have been found. I would recommend that you review the audiograms with the speech results found on page 157-159 of your textbook to ensure that you understand the speech testing results. The summary of tests used in speech audiometry found on page 160 is also a helpful review.