PPM Formaldemeter 400. Formaldehyde Monitor. Operation Manual

Similar documents
RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

el-720 Carbon Dioxide Gas Leak Detector User s Guide

ExStik TM ph Waterproof Meters

Instruction Manual HI Refractometer for Refractive Index and Brix Measurements

RT-1261A Ear and Forehead Thermometer Owner s Manual

Mini-led spotlight with magnetic base

WS-9005TWC 915MHz WIRELESS RAIN GAUGE INSTRUCTION MANUAL

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

Wireless Indoor/ Outdoor Thermometer

EZ Hold USER MANUAL EZ HOLD. Music On Hold. Models M2060, EZM480. EZ ADVANTAGE CORP., 221, Broad Street, MILFORD, CT U.S.A.

Installation and Operation Manual. Digital Remote Meter for Monitoring System Performance. Version: RM-1

Milwaukee USER MANUAL

Important Operating Instructions and Warranty Information On Your New Electronic AMSEC Safe

Instruction Manual HI Refractometer for Sucrose Measurements

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Periodic Verification System Model 775PVS

Operator Quick Guide EC SENSOR

Mold Alert... Operation Manual P0270

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

100 Series Maxi Internal Filters. Instruction Manual

e I N S T RUC TION M ANU A L

Ozone Generator Manual

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / Instruction Manual. Page 1 of 9

Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions

Automatic Wrist Blood Pressure Monitor with

ADVANCED INTERNET DOWNLOADABLE POSTAL & SHIPPING SCALE

Digital Fingerprint safe

E&P HYDRAULICS Remote control EPRC-01 Level System Camper

Intelli-Time Alarm Clock model 13027

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

Leak Detection System

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

Wireless Home Security Alarm System AM 500

BroadBand PowerShield. User Manual

FARMEX HT-PRO OPERATING INSTRUCTIONS PORTABLE MOISTURE AND TEMPERATURE TESTER FOR BALED HAY

Important Operating Instructions and Warranty Information On Your New Electronic AMSEC Safe

ROTOR LOADER OWNER S MANUAL

Wolverine Data PASS Portable Scanner User Manual V1.0

Instruction Manual Temp 4/5/6 Temperature Meter Series

Optical Power Meter. Specification & User Manual

User's Guide. Integrating Sound Level Datalogger. Model Introduction

ECO SENSORS, INC. OZONE SENSOR AND MONITOR

Alarm Clock USER GUIDE

Indoor Remote Controlled Power Points

Sip/Puff Switch User Guide

2-3 SAS/SATA II HDD Canister USER S MANUAL XC-23D1-SA10-0-R. Document number: MAN A

Tympanic Thermometer

User Manual. Humidity-Temperature Chart Recorder. Model RH520

IR Repeater. Operation Manual P8620

COMBUSTIBLE GAS AND HC REFRIGERANT LEAK DETECTOR OPERATING INSTRUCTIONS DETECTS ALL HYDROCARBON & OTHER COMBUSTIBLE GASES JB INDUSTRIES

Please read and keep these instructions. Run water through your machine before first use

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model #

Mini Power Bank and Charger

XTEND 900 MHZ WIRELESS MODEM For use with VEEDER-ROOT EMR³ DATALINK INSTALLATION INSTRUCTIONS Manual , Rev. B

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

User's Guide Integrating Sound Level Datalogger Model

Thermocouple Thermometer Models

Thermometer model 00754

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL

INFRARED QUARTZ WALL HEATER

EzFlex Combustible Gas Leak Detector Model EZ40

DICKSON PW4 DICKSON. Pressure Chart Recorders. Applications & Specifications. Product. Getting Started, Charts & Accessories.

COASTAL AMUSEMENTS INC.

Get Me Cool Now Portable Air Conditioner User s Manual

DX-395. FM/AM/SW1/SW2 PLL Synthesized Receiver OWNER S MANUAL. Cat. No Please read before using this equipment.

ROUTER TABLE INSERT ASSEMBLY AND OPERATING INSTRUCTIONS

We reserve the right to make technical changes.

Dual Cushion User Manual. Specialist Suppliers to the Healthcare Industry

3 WATT LED SPOTLIGHT Model No. SLM

NFC Communication Tray Model: HHX-IT3-Z

THERMONET UNDERFLOOR HEATING STANDARD THERMOSTAT KIT STOCK CODE 5260

digital fever alert ear thermometer suitable from birth instruction manual

User Guide. Temperature and Humidity Datalogger. Model 42280

High Temperature InfraRed Thermometer with Laser Pointer

19 LED Tube Controller ORDERCODE 41003

Instructions & Safety Information Models A220-20D and A220-20L Version 2

INSTALLATION AND OPERATING INSTRUCTIONS

LCD BacPac. User Manual + Warranty Info

ThermoScan Instant Thermometer IRT 1020 Type 6005 Home Model

Single Port T1/E1 Tap

AN500T, AN1000, AN1000T, AN1500, AN1500T AN2000, AN2000T

ClearView 100. Instruction Manual

DAB+ / FM Tuner Model: TU-201

Ultrasonic Distance Meter. Operation Manual

Messager USB Digital Messaging System

LCD5500Z / PKP-LCD v3.x Installation Instructions

AM/FM PLL Clock/Table Radio

Use & Care. of your Iron. All about the TA B L E O F C O N T E N T S. USA and Canada

Model ATD Refrigerant Gas Leak Detector. User Manual. Detects all CFC, HFC, HCFC, HFO Refrigerants including blends

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

Wave/PC Interactive System USB Adapter Kit. Installation Guide

tire inflator with pressure gauge

Air vacuum pump with r134a & r12/22

IMPORTANT NOTE: Please select the timing function you wish to use by pressing button B. Then, proceed as described in these instructions.

High Efficiency Vaporizing System and Highly Durable Heaters Ensure 25% Increased Output.

Electrical Stability Tester, Model 23E Instruction Manual Manual No , Revision B Instrument No

Light Meter (Foot Candles)

CompacVideo Otoscope. Model (NTSC) and 23120P (PAL) Operating Instruction Manual

WTSL-SN-1 Alarm Control Unit with Wi-Safe 2 Installation and User Guide

Transcription:

PPM Formaldemeter 400 Formaldehyde Monitor Operation Manual NOTICE: PPM Technology Limited shall not be liable for errors that may appear herein or for incidental or consequential damages in connection with or arising from the use of this material.

TABLE OF CONTENTS 1 INTRODUCTION 3 1.1 Initial Receipt of the PPM Formaldemeter 400 Kit 4 1.2 General Description 4 1.3 Instrument Features 6 2 TECHNICAL INFORMATION 8 2.1 Principle of Operation 8 2.2 Interferants 9 2.3 Sensor Background Reading 10 3 USING THE PPM FORMALDEMETER 400 10 3.1 Taking a Sample 10 3.2 Sensor Recovery Period 11 4 CALIBRATION CHECK AND ADJUSTMENT 12 4.1 General Information 12 4.2 The Formaldemeter 400 Calibration Standard 12 4.3 Calibration Check Procedure 13 4.4 Calibration Adjustment Procedure 15 5 MAINTENANCE 17 5.1 Instrument Battery 18 5.2 Cleaning the Formaldemeter 400 18 5.3 Storage 18 WARRANTY 19 2

1 INTRODUCTION The PPM Formaldemeter 400 is an easy-touse, hand held direct reading instrument designed for the rapid measurement of airborne formaldehyde levels. Please read these instructions carefully and familiarise yourself with the instrument before use. This operating manual will provide you with all the necessary information for the correct use of your Formaldemeter 400. 3

1.1 Initial Receipt of the PPM Formaldemeter 400 Kit Your PPM Formaldemeter 400 kit has been packaged carefully and includes all the components necessary for full operation. Immediately upon receipt, please examine the kit contents carefully to ensure that you have received the following items in good condition. Component list The instrument kit contains: Formaldemeter 400 instrument with battery installed Formaldehyde calibration standard Thermometer Vial of phenol filters (10) Ball point pen Certificate of calibration Operation manual Warranty registration card Please report any missing items to your dealer. Damage Inspect all items carefully. Any damage must be reported immediately to both the carrier and your dealer. 1.2 General Description Unlike other formaldehyde monitoring devices such as colour stain tubes and badges, the Formaldemeter 400 is capable of measuring many samples consecutively without the need for inconvenient ancillary equipment. Being ultra compact and battery operated, the instrument is truly field portable. The Formaldemeter 400 is extremely simple to use 4

and provides immediate, semi-quantitative readings of atmospheric formaldehyde concentration displayed on a digital read-out. The instrument is designed to measure the concentration of formaldehyde in snatch (discrete) samples of air and should be employed primarily as a screening device. Important points It is important that you are aware of the following points when using the instrument The Formaldemeter 400 is temperature compensated to operate most accurately in the range 5 40 C. The results obtained with the PPM Formaldemeter 400 are instantaneous spot readings. A single reading is not necessarily representative of long term personal exposure. A series of readings taken at short intervals is preferable to infrequent tests. Avoid smoking in the environment to be analysed tobacco smoke contains formaldehyde. Care must be taken to ensure that fluid or dust is not drawn into the instrument. This could permanently damage the sensor. The Formaldemeter 400 has been designed to be sufficiently robust for everyday field use. However, should the unit sustain a severe physical shock, the operation and calibration of the instrument should be checked using the supplied formaldehyde calibration standard. 5

1.3 Instrument Features 6

Sampling Port Display Display Key The brass sampling port is the inlet through which the sample is drawn into sensor. The liquid crystal display (LCD) shows the formaldehyde concentration of the sample in 0.01 ppm increments. The display also shows text messages during certain operations. sensor recovering sensor ready sampling taking calibration sample set calibration level replace battery instrument not calibrated Keypad Battery compartment The instrument is operated by four buttons, some of which have multiple functions. The button switches are located beneath the membrane and are operated by pressing firmly where indicated. The battery compartment is located beneath a cover at the bottom rear of the instrument. To gain access to the battery, simply remove the cover by unscrewing the two screws. A flat battery is indicated by on the display. See section 5.1 for guidance on battery replacement. Serial port This socket can be used for connecting your Formaldemeter 400 to the PPM Base Unit or PC for automated continuous monitoring applications. 7

2 TECHNICAL INFORMATION 2.1 Principle of Operation Electrochemical formaldehyde sensor The PPM Formaldemeter 400 uses proven electrochemical sensing technology for determining the concentration of formaldehyde in air samples. The instrument contains an electrochemical formaldehyde sensor comprising two noble metal electrodes and a suitable electrolyte. When air is drawn into the sensor by means of the internal sampling system, a small voltage is generated which is directly proportional in magnitude to the concentration of formaldehyde in the sample. This voltage is produced as a result of the electrooxidation of formaldehyde at one of the catalytically active electrodes. The signal is fed to a precision electronic amplifier and output on the instrument s display, when calibrated, as formaldehyde concentration in ppm (parts-per-million by volume). All the electronic systems are based on modern, integrated circuitry employing the latest surface mount technology to ensure that the Formaldemeter 400 is an exceptionally robust and reliable instrument. 8

2.2 Interferants Phenol & resorcinol Phenol filter life Alcohols & aldehydes The presence of phenol in the air can give a reading on the Formaldemeter 400. When monitoring formaldehyde in situations where phenolic resins are used, the phenol filters provided should be used. These fit on to the sampling port of the instrument. The filters will completely remove phenols from the sample even at concentrations in excess of 1000 ppm without affecting the formaldehyde reading. Each filter should be used no more than five times and then discarded. Partially used filters should not be stored in the vial with unused filters. Replacement filters can be ordered from PPM through your local distributor. As is found with other portable detection equipment, the meter is not totally specific to formaldehyde alone, being susceptible to a degree of interference from a small range of other chemicals. Other aldehydes and alcohols such as methanol and ethanol in the atmosphere can cause cross-interference effects. A list of common interferants is available on request. 9

2.3 Sensor Background Reading Due to the high sensitivity of the sensor and the widespread occurrence of formaldehyde in the indoor environment, a background reading of up to 0.05 ppm can often be produced even when sampling in an atmosphere considered to be free of formaldehyde or other contaminants. 3 USING THE PPM FORMALDEMETER 400 3.1 Taking a Sample Power on Press the ON-OFF button once. The instrument will display: flashing for 3 seconds as the instrument checks the sensor, followed by: indicating that the instrument is ready to take a sample. Sampling Hold the instrument in the atmosphere to be sampled. Press and release the SAMPLE button. The display will show and the internal pump should be heard running for almost two seconds as it samples the air. 10

Display reading As the sample is analysed, the display will show: flashing After approximately 10 seconds, a value will be displayed on the LCD and held until the instrument is switched off. This indicates the formaldehyde concentration in ppm. Power off Switch the instrument off by pressing the ON- OFF button once. If you forget to switch the instrument off after a test, the PPM Formaldemeter 400 will automatically switch itself off after 5 minutes. 3.2 Sensor Recovery Period Sensor clearing Between samples, the instrument should be left switched off for a few minutes to allow the sensor to clear of any residual formaldehyde. As a general rule, the higher the reading obtained, the longer it takes for the sensor to clear. If the instrument is switched on before the sensor has cleared, the display will show: flashing and the sampling pump will not operate. The cell is clear only when the display shows: on. permanently 3 seconds after power The instrument is now ready to take the next sample. 11

4 CALIBRATION CHECK AND ADJUSTMENT PLEASE READ THIS SECTION THOROUGHLY BEFORE ATTEMPTING TO CHECK OR ADJUST CALIBRATION. USERS ARE STRONGLY ADVISED TO FAMILIARISE THEMSELVES WITH THE INSTRUMENT BEFORE ATTEMPTING TO ADJUST THE CALIBRATION AND SHOULD FOLLOW THE INSTRUCTIONS CAREFULLY. 4.1 General Information Check calibration regularly Sensor sensitivity can change very gradually with time, so periodic recalibration may be required. It is advisable to check calibration regularly to ensure that the instrument is functioning correctly. A quick calibration check can be carried out by drawing a formaldehyde vapour sample of known concentration into the instrument s sensor and noting whether the displayed reading agrees with the expected concentration value. The PPM Formaldemeter 400 is supplied complete with a formaldehyde calibration standard and a thermometer, which are essential components for checking and adjusting calibration. 4.2 The Formaldemeter 400 Calibration Standard The PPM Formaldehyde Calibration Standard consists of formaldehyde absorbed on a solid substrate in a glass tube from which a headspace vapour sample can be drawn. Each standard is carefully manufactured to a high tolerance. 12

Effect of temperature The concentration of formaldehyde vapour generated in the calibration tube varies with temperature and for this reason, a thermometer and temperature/concentration table is supplied. Handle the calibration standard as little as possible, and only by the yellow end caps to avoid heating the tube. Temperature equilibration Before using the standards, they should always be allowed to stabilise within the recommended temperature range of 15-29 C for at least 1 hour. Standard life Each standard has a capacity for 100 samples or a life of six months (whichever occurs sooner) and it must not be used after the indicated expiry date. New calibration standards can be ordered from PPM or through your local distributor. 4.3 Calibration Check Procedure Temperature equilibration Sensor check Place the instrument, thermometer and calibration standard together in a place where the temperature is stable for at least one hour before commencing the calibration check procedure to allow thermal equilibration. Before carrying out a calibration check, the sensor must be clear of formaldehyde vapour from any previous samples. When the sensor is clear, the display will show: 3 seconds after power on. 13

If the display shows flashing for longer than 5 seconds, then the sensor is not ready to take a sample. Switch the instrument off and allow it to recover for a few minutes. Procedure 1. Place the standard with the thermometer on a work surface. Handle the calibration standard as little as possible to avoid heating the tube, holding it by the yellow end caps. Remove both end plugs. 2. Switch the instrument on by pressing the ON-OFF button once. Wait for the display to show: 3. Place the instrument on the work surface and insert the nozzle into the sampling end of the tube (as indicated by the black arrow). Ensure a good seal around the instrument nozzle by pushing the standard tube firmly against the instrument when taking a sample. 4. Press the SAMPLE button and wait until the internal sampling pump stops before removing the standard from the instrument. Replace the end plugs securely. 5. The display will initially show: flashing Observe the final reading displayed (after approximately 10 seconds). 14

6. Note the reading obtained and temperature 7. Refer to the Temperature/Concentration look-up table on the standard tube. If the reading is within 10% of the value shown in the table, then no recalibration is required. If recalibration is required, follow the procedure in Section 4.4. Leave the instrument switched off for approximately 5 minutes to recover before commencing another atmospheric analysis or calibration adjustment. 4.4 Calibration Adjustment Procedure PLEASE READ THIS SECTION THOROUGHLY BEFORE ATTEMPTING TO ADJUST CALIBRATION. USERS ARE STRONGLY ADVISED TO FAMILIARISE THEMSELVES WITH THE INSTRUMENT BEFORE ATTEMPTING TO ADJUST THE CALIBRATION AND SHOULD FOLLOW THE INSTRUCTIONS CAREFULLY. Under normal operating conditions, instrument calibration should require only minimal periodic adjustment. To see whether a full recalibration is required, perform a calibration check first as described in section 4.3. Temperature equilibration Procedure Leave the instrument and calibration standard in a room where the temperature is constant for at least one hour before calibrating to allow thermal equilibration. 1. Place the standard with the thermometer on a work surface. Handle the calibration standard as little as possible to avoid heating the tube, holding it only by the yellow end caps. 15

2. Read the temperature on the thermometer and determine the required concentration reading from the lookup table on the calibration tube. for example: Temperature: 23 C Reading: 3.15 ppm 3. Remove the yellow plugs from both ends of the calibration standard tube. 4. Press the ON-OFF button once to switch the instrument on. 5. Wait for the display to show 6. Insert the instrument nozzle into the sampling end of the calibration standard (indicated by the black arrow). To maintain airtight contact between the nozzle and the standard, push the standard firmly against the instrument. 7. Simultaneously depress and release both Cal and buttons. 8. will appear on the LCD, followed by the sound of the pump drawing a vapour sample from the tube. 9. When the pump stops, remove the calibration tube and replace both yellow end plugs. 16

10. As the instrument analyses the sample, the display will show: flashing for approximately 10 seconds, followed by: alternating with 11. Now use the Cal and buttons to adjust the display reading to the required concentration. 12. Press the SAMPLE button to store this calibration value. The display will show: followed by The Formaldemeter 400 will then switch off automatically. The instrument has now been recalibrated. 5 MAINTENANCE With the exception of the battery, your PPM Formaldemeter 400 has no user-serviceable components. It is important that no attempt is made to open the instrument other than to replace the battery. Any evidence of tampering with the instrument will invalidate the warranty. If you find that your instrument requires service or repair, please return it to the factory or an authorised PPM Service Centre. 17

5.1 Instrument Battery Low battery indicator If appears on the display when the instrument is switched on, then the battery voltage is too low and needs replacing Battery Replacement The instrument requires a 9V PP3 / MN1604 / 6LR61 type alkaline battery. To replace the battery, simply remove the battery compartment cover at the bottom rear of the instrument by unscrewing the two screws. Clip a new battery to the connector and place the battery in the compartment. Secure the cover in place again with the two screws. 5.2 Cleaning the Formaldemeter 400 Take care during cleaning On no account should the Formaldemeter 400 be immersed in liquid. Any fluid entering the instrument will destroy the sensor and the electronic circuitry. If the instrument s enclosure requires cleaning, it should be wiped with a damp cloth. Never use abrasive or solvent based cleaning agents. 5.3 Storage When not in use, your Formaldemeter 400 should be stored in the supplied carrying case in a clean, dry environment and away from extremes of temperature. 18

WARRANTY The PPM Formaldemeter 400 is warranted to be free of defects in materials and workmanship under proper and normal use and service for a period of 1 year from the date of purchase. This warranty is limited to repair or replacement (at our option) of any part that proves defective in material or workmanship under normal use and service, provided the product is returned to PPM Technology Limited, shipment charges prepaid. Damage due to defacement, misuse, tampering, lack of prescribed maintenance or use in violation of the instructions furnished by PPM Technology Limited is not covered. This warranty is in lieu of all other warranties, express or implied, including but not limited to merchantability or fitness for a particular purpose. In no event shall we be liable for any incidental or consequential damages of any nature. PPM Technology Limited reserves the right to make changes at any time to this document and to the design, construction, appearance and specifications of its products without notice. No part of this document may be reproduced in any form without the prior written consent of PPM Technology Limited. Copyright 2000 PPM Technology Limited Ver 1.1 Nov 2001 Part# OM/FMD400 19

NOTES 20