TECHNICAL REQUIREMENTS OF ISO/IEC 17025:2005

Similar documents
Quality Management System Policy Manual

ISO/IEC QUALITY MANUAL

NIST HANDBOOK 150 CHECKLIST

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA USA

OD-2017-Ed.1.5 CHECK LIST FOR TESTING AND CALIBRATION LABORATORIES

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Procedure for Equipment Calibration and Maintenance

Checklist. Standard for Medical Laboratory

Luminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual

TITLE: Quality System Manual DOCUMENT NUMBER: QMS

Testing Facility Laboratory Manual

Internal Quality Management System Audit Checklist (ISO9001:2015) Q# ISO 9001:2015 Clause Audit Question Audit Evidence 4 Context of the Organization

NABL NATIONAL ACCREDITATION

Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing

Surgi Manufacturing Quality Manual

ISO 9001:2015 Internal Audit Checklist

ISO 9001:2000 Gap Analysis Checklist

ISO and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

ISO 9001:2000 AUDIT CHECKLIST

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

American Association for Laboratory Accreditation

REQUIREMENTS FOR AN ACCREDITED VETERINARY MEDICAL DIAGNOSTIC LABORATORY

Non-Controlled Copy. Quality Manual Revision Log. Rev # Date Description

DNV GL Assessment Checklist ISO 9001:2015

OH&S MANAGEMENT SYSTEM CHECKLIST - AS 4801:2001 (STATUS A = Acceptable; N = Not Acceptable; N/A = Not Applicable)

ISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR

CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS

EA IAF/ILAC Guidance. on the Application of ISO/IEC 17020:1998

South Coast Air Quality Management District

NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory

CHECKLIST ISO/IEC 17021:2011 Conformity Assessment Requirements for Bodies Providing Audit and Certification of Management Systems

EDUCATION AND TRAINING OF THE LABORATORY STAFF AS A PART OF THE LABORATORY COMPETENCE

GENERAL WELDING & FABRICATING, INC. 991 Maple Road Elma, New York 14059

Product Liability Risk Control Checklist

ISO 9001:2008 Audit Checklist

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

GLOBAL AEROSPACE PROCEDURE

Good Research Practice

Recognition of my Product (Testing Laboratory) Why I need recognition, what is the effect?

Section i POLICY AND AUTHORITY STATEMENT

BLOOM AND WAKE (ELECTRICAL CONTRACTORS) LIMITED QUALITY ASSURANCE MANUAL

NIST HANDBOOK CHECKLIST CONSTRUCTION MATERIALS TESTING

PROPOSED UPDATED TEXT FOR WHO GOOD MANUFACTURING PRACTICES FOR PHARMACEUTICAL PRODUCTS: MAIN PRINCIPLES (JANUARY 2013)

DEC STD ISO Quality Systems - Model for Quality Assurance in Production and Installation

Measuring equipment in a laboratory.

Software Verification and Validation

THERMAL INSULATION MATERIALS

Company Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES

ISO/IEC Inspection Standard Application Document

COMPANY NAME. Environmental Management System Manual

ISO 9001:2008 Quality Management System Requirements (Third Revision)

EffiValidation 3.0 software basic training module

Measuring instruments in industry.

ISO-9001:2000 Quality Management Systems

COMMISSION DIRECTIVE 2003/94/EC

CORPORATE QUALITY MANUAL

UNCONTROLLED COPY FOR REFERENCE ONLY

ISO 9001: 2008 Boosting quality to differentiate yourself from the competition. xxxx November 2008

Row Manufacturing Inc. Quality Manual ISO 9001:2008

QUALITY MANUAL ISO 9001:2015

Supplier Quality Management System Audit Checklist (ISO 9000:2000, TS 16949:2002)

RTP s NUCLEAR QUALITY ASSURANCE PROGRAM

OPERATIONAL STANDARD

Generic CMMS Quality Assurance Plan

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037

US EPA REGION III QUALITY MANAGEMENT PLAN REVIEW CHECKLIST

Eagle Machining, Inc. Quality Management System

CCD MARINE LTD QUALITY MANUAL PROCEDURE Q Date: Title. Revision: QUALITY MANUAL PROCEDURE Q September 2014

ISO 9001 : 2008 QUALITY MANAGEMENT SYSTEM AUDIT CHECK LIST INTRODUCTION

A Guide for Documenting the Management System for a Testing/Calibration Lab. May 2015

QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08

FMC Technologies Measurement Solutions Inc.

Validation and Calibration. Definitions and Terminology

NC SBI QUALITY ASSURANCE PROGRAM

CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES

UL Qualified Firestop Contractor Program Requirements April 2012

P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014

PRODUCT CERTIFICATION REQUIREMENTS

DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL

UL Qualified Firestop Contractor Program Management System Elements. March 13, 2013

ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan)

QUALITY MANUAL ISO Quality Management System

Procedures and General Requirements

AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189

MEAT GRADING AND CERTIFICATION BRANCH QUALITY MANUAL

Internal Audit Checklist

AS9100 Quality Manual

WORKPLACE HEALTH AND SAFETY AUDITING GUIDELINES

1 For more information T: / E: DAP@ul.com / W: ul.com/dap

Network Certification Body

SMALL BUSINESS OH&S SELF APPRAISAL

How To Manage A Quality Management System

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

EMS ISO Checklist

Transcription:

SADCAS Ref. No: TECHNICAL REQUIREMENTS OF ISO/IEC 17025:2005 Date/s of Evaluation: Assessor/s & Observers: Laboratory: Area / Field of Operation: Laboratory Representative: This report covers the following: Document Review only Implementation on Site Visit only Document Review And Site Visit Other Clause REQUIREMENTS & COMMENTS Compliance = C, Non-Compliance = NC, Not Applicable = NA Comment below on adequacy of how requirements have been addressed, documented and/or implemented. References to ISO/IEC 17025 are in italics & indicated as Std. The order of assessment need not follow the order of the checklist. Assessors are expected to know & have the standard, this worksheet is designed as guidance to prompt detailed recording of the process. REFER TO ISO/IEC 17025 FOR DETAIL AND FOR CLARIFICATION NOTES. 5 TECHNICAL REQUIREMENTS 5.1 General: 5.1.1 Factors determining correctness, reliability of tests/calibrations performed 5.1.2 Extent to which factors contributing to total uncertainty of measurement taken into account in: developing methods/procedures; training, qualification of staff; selection, calibration of equipment. C NC NA 5.2 Personnel: How are the following addressed/ implemented? 5.2.1 Note Std Competence of all who operate specific equipment, perform tests/calibrations, evaluate results, sign reports/ certificates. Qualified on basis of appropriate education, training, experience and/or demonstrated skills. Appropriate supervision of trainees. 5.2.2 Goals education, training, skills. ID training needs, provide training. Evaluate effectiveness of training. Programme relevant to tasks. 5.2.3 Employed by/under contract. Supervised, competent, work in accordance with lab s management system. 5.2.4 Current job descriptions for managerial, technical, key support personnel involved in Note Std tests/calibrations. 5.2.5 Authorize specific personnel to: perform particular types of sampling, test, calibration; issue reports certificates; give opinions, interpretations; operate particular types of equipment. Records of authorization(s), competence, educational, professional qualifications, training, skills, experience of all technical personnel, including contracted personnel. Date, readily available. Issue No.: 1 Page 1 of 12 Date of Issue: 2010-06-15

Additional / General Comments (This space may also be used to expand on comments in specific sections) 5.3 Accommodation and Environmental Conditions: How are the following addressed/ implemented? Issue No.: 1 Page 2 of 12 Date of Issue: 2010-06-15

5.3.1 Not invalidate/adversely affect results. Permanent lab/other sites. Requirements documented. 5.32 Monitor, control, record where quality of results influenced. Stop work when results jeopardized. 5.3.3 Effective separations between incompatible activities. Measures to prevent cross contamination. 5.3.4 Access to areas affecting quality of tests/calibrations controlled based on particular circumstances. 5.3.5 Measures taken to ensure good housekeeping. Special procedures prepared where necessary. 5.4 Test &Calibration Methods &Method validation: How are the following addressed/ implemented? 5.4.1 General: Appropriate for scope. Include sampling, handling, transport, storage, preparation, Note Std use/operation of equipment. Estimation of measurement uncertainty & statistic techniques for & See analysis of data, where their absence could jeopardize results. Up-to-date and readily available. 4.3 Deviations documented, technically justified, authorized, accepted by customer 5.4.2 Selection of methods: How are the following addressed/ implemented? Meet needs of customer, appropriate for tests/calibrations. Prefer international, regional, national standards. Latest valid edition. Supplements to ensure consistency. By reputable technical organizations/in relevant scientific texts/journals or specified by the manufacturer of the equipment. Laboratory developed method/s appropriate for intended use and validated. Customer informed as to the method chosen. Confirm that it can properly operate standard methods before introducing the tests/calibrations. Confirmation repeated if standard method changes. Customer informed when method proposed by customer is considered to be inappropriate or out of date. 5.4.3 Laboratory developed methods: How are the following addressed/ implemented? 5.4.4 Note Std Introduction of method planned, assigned to qualified persons equipped with adequate resources. Plans up-dated as development proceeds, effective communication amongst all persons involved. Non-standard methods: How are the following addressed/ implemented? Agreement with the customer, include clear specification of the customer s requirements and the purpose of the test and/or calibration. Validated appropriately before use. 5.4.5 5.4.5.1 5.4.5.2 Notes Std 5.4.5.3 Notes Std Validation of methods: Confirmation by examination and provision of objective evidence that particular requirements for specific intended uses are fulfilled. Validate non-standard methods; laboratory-developed methods; standard methods used outside intended scope; amplifications/modifications of standard methods. As extensive as is necessary for application/field of application. Record results, procedure, statement of fitness for intended use. Range and accuracy as assessed for the intended use relevant to the customers needs. e.g. the uncertainty, detection limit, selectivity, linearity, repeatability, reproducibility, robustness) Issue No.: 1 Page 3 of 12 Date of Issue: 2010-06-15

Additional / General Comments (This space may also be used to expand on comments in specific sections) Issue No.: 1 Page 4 of 12 Date of Issue: 2010-06-15

5.4.6 Estimation of Uncertainty of Measurement: How are the following addressed/ implemented? 5.4.6.1 Test &Calibration labs; uncertainty procedure for all calibrations and types of calibrations. 5.4.6.2 NB Notes std 5.4.6.3 Notes Std Testing laboratories, identification of all the components of uncertainty & reasonable estimation is attempted. Reporting of results do not give wrong impression of uncertainty. Estimation based on knowledge of performance of method, measurement scope and validation data. All uncertainty components of importance in the given situation taken into account using appropriate methods of analysis. 5.4.7 Control of data: How are the following addressed/ implemented? 5.4.7.1 Calculations and data transfers subject to appropriate checks in a systematic manner. (e.g. Scientific notation & rounding-off policy) 5.4.7.2 Electronic equipment for acquisition, processing, recording, reporting, storage or retrieval of data. a) Software developed by user documented in sufficient detail, suitably validated as adequate for use. b) Procedures for integrity & confidentiality of data entry, collection, storage, transmission, processing. c) Notes Std Computers/automated equipment, proper functioning, conditions to maintain the integrity of data. 5.5 Equipment: How are the following addressed/ implemented? All required equipment & its software, including that outside its permanent control available. 5.5.2 Capable of achieving accuracy required, comply with relevant specifications. Calibration programmes See 5.6 for key quantities having significant effect on results. Calibrated/checked before use. 5.5.3 Operated by authorized personnel. Up-to-date instructions use, maintenance readily available. 5.5.4 Each item significant to the result, when practicable, uniquely identified. 5.5.5 Records maintained of each item significant to the tests and/or calibrations performed. a) The identity of the item of equipment and its software. b) The manufacture s name, type identification and serial number or other unique identification. c) Checks that equipment complies with the specification (see 5.5.2) d) The current location, where appropriate. e) The manufacturers instructions, if available, or reference to their location. f) Dates, results, reports/certificates, adjustments, acceptance criteria, due date of next calibration. g) The maintenance plan, where appropriate and maintenance carried out to date. h) Any damage, malfunction, modification or repair to the equipment. Issue No.: 1 Page 5 of 12 Date of Issue: 2010-06-15

5.5.6 Notes Std Procedures for safe handling, transport, storage, use and planned maintenance of measuring equipment to ensure proper functioning and in order to prevent contamination or deterioration. 5.5.7 Defective equipment, prevented from use until shown by calibration or test to perform correctly. See 4.9 Examine effect of defect on previous tests/calibrations, control non-conforming work procedure. 5.5.8 When practicable identified to indicate calibration status, date last calibrated, re-calibration due. 5.5.9 Outside direct control, ensure function & calibration checked, satisfactory before returned to use. 5.5.10 Intermediate checks to maintain confidence in calibration status carried out to defined procedure. 5.5.11 Calibrations correction factors, procedures to ensure copies (e.g. in software) are correctly updated. 5.5.12 Equipment, hardware and software, safeguarded from adjustments, which would invalidate results. Additional / General Comments (This space may also be used to expand on comments in specific sections) Issue No.: 1 Page 6 of 12 Date of Issue: 2010-06-15

5.6 Measurement Traceability: How are the following addressed/ implemented? 5.6.1 Note Std General: All equipment including for subsidiary measurements having a significant effect on validity of results, calibrated before being put into service. Established programme and procedure for the calibration of its equipment. 5.6.2 Specific requirements 5.6.2.1 Calibration: How are the following addressed/ implemented? 5.6.2.1.1 Calibration laboratories, programme to ensure traceability to International System of Units (SI). Notes Std Stds & instruments to SI by unbroken chain of comparisons link to primary standards. See also Use laboratories demonstrating competence, measurement capability and traceability. 5.10.4.2 Certificates contain results, uncertainty/statement of compliance with identified metrological spec. 5.6.2.1.2 Calibration providing confidence in measurement by establishing traceability to standards such as: Certified reference materials provided by a competent supplier. Specified methods and/or consensus standards that are clearly described, agreed by all parties. Participation in a suitable programme of inter-laboratory comparisons where possible. 5.6.2.2 Testing: How are the following addressed/ implemented? 5.6.2.2.1 Measuring & test equipment with measuring functions apply 5.6.2.1, unless calibration contributes Note Std little to total uncertainty of test result. Equipment providing uncertainty of measurement needed. 5.6.2.2.2 SI not possible, then certified reference materials, agreed methods, consensus standards, 5.6.2.1.2 5.6.3 Reference standards and reference materials: How are the following addressed/ implemented? 5.6.3.1 Reference standards: Programme & procedure for calibration, traceable. Used for calibration only unless not validated. Calibrated before and after any adjustment. 5.6.3.2 Reference materials: Traceable to SI units/to certified reference materials. Internal checked 5.6.3.3 Intermediate checks: Checks to maintain confidence in calibration status of reference, primary, transfer/working standards, reference materials according to defined procedures and schedules. 5.6.3.4 Transport and storage: Procedures for safe handling, transport, storage and use in order to Note Std prevent contamination or deterioration and to protect their integrity. Issue No.: 1 Page 7 of 12 Date of Issue: 2010-06-15

5.7 Sampling: How are the following addressed/ implemented? 5.7.1 Notes Std Sampling plan & procedures available at the location where sampling is undertaken. Based on appropriate statistical methods, address factors to control to ensure validity of results. 5.7.2 Customer required deviations, recorded in detail with sampling data, results, communicated to all staff. 5.7.3 Records of data & operations, procedure, sampler, environmental conditions, location, statistics 5.8 Handling of test and calibration items: How are the following addressed/ implemented? 5.8.1 Transport, receipt, handling, protection, storage, retention, disposal, protect integrity/interests. 5.8.2 Identify, for life of item, not confuse, accommodate sub-division and transfer of items. 5.8.3 Departures recorded on receipt. Where doubt consult customer before proceeding and record. 5.8.4 Notes Std Avoid deterioration, loss /damage storage, handling, preparation. Follow instruction. Environmental conditions maintained, monitored, recorded. Storage secure, protect condition and integrity. Additional / General Comments (This space may also be used to expand on comments in specific sections) Issue No.: 1 Page 8 of 12 Date of Issue: 2010-06-15

5.9 Assuring the quality of test & calibration results: How are the following addressed/ implemented? 5.9.1 QC procedures to monitor validity of results, data recorded, trends, statistics. Planned & reviewed: Note Std a) Regular use of certified reference materials and/or secondary reference materials. b) Participation in inter-laboratory comparison or proficiency testing programmes. c) Replicate tests or calibrations using the same or different methods. d) Re-testing or re-calibration of retained items. e) Correlation of results for different characteristics of an item. 5.9.2 QC data analysis. Action plan for incorrect results. Preventing the reporting of incorrect results. Issue No.: 1 Page 9 of 12 Date of Issue: 2010-06-15

5.10 Reporting the results: How are the following addressed/ implemented? 5.10.1 Notes Std 5.10.2 Notes Std General Accurate, clear, unambiguous, objective, in accordance with specific instructions in methods. Information requested by client & for interpretation of results, refer 5.10.2 and 5.10.3 or 5.10.4 Internal customers/written agreement, report simplified. Information 5,10.2 to 5.10.4 readily available. Test reports and calibration certificates: How are the following addressed/ implemented? Report/certificate include at least following information unless the laboratory has valid reasons: a) A title (e.g. Test Report or Calibration Certificate ). b) Name and address of the laboratory and location of work, if different from address of laboratory c) Unique identification of report/certificate, on each page, clear identification of end. d) The name and address of the customer. e) Identification of the method used. f) Description of, the condition of, and unambiguous identification of the item(s) tested or calibrated. g) Date of receipt of item(s) where critical to validity and application of results, date(s) of performance. h) Reference to sampling plan and procedures used where relevant to validity or application of results. i) The test or calibration results with, where appropriate, the units of measurement. j) Name/s, function/s, signature/s or equivalent identification of person/s authorizing report certificate. k) Where relevant, a statement to the effect that the results relate only to the items tested/calibrated. 5.10.3 Tests reports 5.10.3.1 In addition to 5.10.2, for test reports, where necessary for interpretation of test results, include; a) Deviations, additions, exclusions from test method, specific test conditions, such as environmental b) Where relevant, a statement of compliance/non compliance with requirements/specifications c) Estimated uncertainty of measurement; need when: -relevant to validity or application of results, -a customer requirement, -uncertainty affects compliance to specification limit d) Where appropriate and needed, opinions and interpretations (see 5.10.5) e) Additional information, required by specific methods, customers or groups of customers. 5.10.3.2 In addition to 5.10.2 and 5.10.3.1, results include the following, for interpretation of results: a) The date of sampling b) Unambiguous identification, manufacturer s name, model or type of designation, serial numbers c) The location of sampling, including any diagrams, sketches or photographs. d) A reference to the sampling plan and procedures used e) Environmental conditions during sampling that may affect the interpretation of the test results. f) Standard/other specification for sampling methods, deviations, additions, exclusions. Issue No.: 1 Page 10 of 12 Date of Issue: 2010-06-15

Additional / General Comments (This space may also be used to expand on comments in specific sections) Issue No.: 1 Page 11 of 12 Date of Issue: 2010-06-15

5.10.4 Calibration certificates 5.10.4.1 In addition to 5.10.2, do certificates include the following, for interpretation of calibration results: a) Conditions (e.g. environmental) that have an influence on the measurement results. b) Measurement uncertainty/statement of compliance with metrological specification/clauses thereof. c) Evidence that the measurements are traceable? (see note 2 in 5.6.2.1.1) 5.10.4.2 Relate only to quantities and the results of functional tests. Statement of compliance with specification, identify which clauses of specification are met/not met Compliance statement omitting results & uncertainties, record maintain results for future reference Statements of compliance made, the uncertainty of measurement taken into account. 5.10.4.3 Adjusted or repaired, calibration results before and after adjustment/repair, if available, reported. 5.10.4.4 Not recommend calibration intervals unless agreed with customer, superseded by legal regulations 5.10.5 Opinions and interpretations: Document basis upon which they have been made. Clearly marked Notes Std as such in a test report. 5.10.6 Testing and calibration results obtained from subcontractors Test results clearly identified. The sub-contractor reports the results in writing or electronically. Calibration laboratory performing work issues the calibration certificate to contracting laboratory. 5.10.7 Electronic transmission of results: Requirements of this international standard met (see also 5.4.7). 5.10.8 Format of reports and certificates: Designed to accommodate each type of test or calibration Notes Std carried out and to minimize the possibility of misunderstanding or misuse. 5.10.9 Amendments to test reports & calibration certificates: Further document/data transfer include Supplement to Test Report/Calibration Certificate, identification, / equivalent form of wording Meet all the requirements of this international standard When complete new report/certificate, uniquely identified, contain reference to original replaced. General/additional comments and matters to follow up at next assessment Signed Lead/Technical Assessor: Date: Issue No.: 1 Page 12 of 12 Date of Issue: 2010-06-15