Molybdate, Low Range Pocket Colorimeter II

Similar documents
Phosphorus, Reactive (Orthophosphate)

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Quaternary Ammonium Compounds DOC

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate

DR/4000 PROCEDURE. CHLORINE, Free

Color, True and Apparent

Color, True and Apparent

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Thermo Scientific Orion AQUAfast Powder and Tablet Reagent Chemistry User Guide

Experiment 12- Classification of Matter Experiment

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Mixtures and Pure Substances

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

2014 Service Technician Training Program Closed Loop Systems

Neutralizing an Acid and a Base

Recovery of Elemental Copper from Copper (II) Nitrate

Physical and Chemical Properties and Changes

DEIONIZATION IN A "NUT SHELL"

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

ALS TRIBOLOGY Engine Coolant Reference Guide

Experiment 8 - Double Displacement Reactions

Chapter 16: Tests for ions and gases

Experiment 18: ph Measurements of Common Substances. Experiment 17: Reactions of Acids with Common Substances

Water Softening for Hardness Removal. Hardness in Water. Methods of Removing Hardness 5/1/15. WTRG18 Water Softening and Hardness

Removing Heavy Metals from Wastewater

Chapter 5, Lesson 3 Why Does Water Dissolve Salt?

FT FIELD MEASUREMENT OF RESIDUAL CHLORINE

Measurement by Ion Selective Electrodes

The Molar Mass of a Gas

Ion Exchange Determination of Na+ by Displacement and Zn 2+ Using Preconcentration. Reading: Harris pp , 699

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

SODIUM CARBOXYMETHYL CELLULOSE

Syllabus OC18 Use litmus or a universal indicator to test a variety of solutions, and classify these as acidic, basic or neutral

Odor (Threshold Odor, Consistent Series)

Solids, Volatile Dissolved and Fixed Dissolved

MOLES AND MOLE CALCULATIONS

Appendix D lists the Field Services Standard Operating Procedures. Appendix E lists the Biological Monitoring Standard Operating Procedures.

MOISTURE (Karl Fischer, Buffered)

IB Chemistry. DP Chemistry Review

Calculation of Molar Masses. Molar Mass. Solutions. Solutions

HOMEWORK 4A. Definitions. Oxidation-Reduction Reactions. Questions

Properties of Acids and Bases

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Adjusting Chemical Levels in a Swimming Pool

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Chloride and Salinity

Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation

UltraClean Soil DNA Isolation Kit

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES

NALCO WT-1000 Closed System Corrosion Inhibitor

CHEMICAL REACTIONS OF COPPER AND PERCENT YIELD KEY

OXIDATION-REDUCTION TITRATIONS-Permanganometry

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin

Preparation of frequently used solutions

USEPA Electrode Method Method 8156

Texas Commission on Environmental Quality

Bluelab Soil ph Meter Instruction Manual

1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l.

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION

LACHAT METHOD NUMBER D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately g CN/L g CN/L. 09e

Experiment 16-Acids, Bases and ph

DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included.

Writing and Balancing Chemical Equations

Reagents also compatible in Hach

ph and Acidity in Wine and Fruit Juice

Carolina s Solution Preparation Manual

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

NITRIC OXIDE and NITROGEN DIOXIDE 6014

WATER CHEMISTRY AND POOL WATER BALANCE

PECTINS. SYNONYMS INS No. 440 DEFINITION DESCRIPTION. FUNCTIONAL USES Gelling agent, thickener, stabilizer, emulsifier CHARACTERISTICS

WHAT IS IN FERTILIZER OTHER THAN NUTRIENTS?

Experiment #8 properties of Alcohols and Phenols

HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY

Phosphates (ortho- and total)

Calculating Molar Mass of a Compound

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit

Complexometric Titrations

FIVE-DAY BIOCHEMICAL 7.0 OXYGEN DEMAND

The Determination of an Equilibrium Constant

Standard methods in water analysis

Total Suspended Solids Total Dissolved Solids Hardness

THE ACTIVITY OF LACTASE

18 Conductometric Titration

Standard Methods for the Examination of Water and Wastewater

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

THE USE OF OZONE IN COOLING TOWERS

ENCM Environmental Impact Assessment and Environmental Monitoring. Determination of nitrate and phosphate levels in well water/surface water

Transcription:

Analytical Procedure Molybdate, Low Range Pocket Colorimeter II Ternary Complex Method (0.10 to 5.00 mg/l MoO4 ) Tips and Techniques The Pocket Colorimeter II is designed to measure solutions contained in sample cells. DO NOT dip the meter in the sample or pour the sample directly into the cell holder. The results can be expressed as mg/l molybdenum (Mo) or mg/l sodium molybdate (Na 2 MoO 4 ) by multiplying the mg/l molybdate (MoO 4 ) result by 0.600 or 1.287, respectively. After adding reagent, a green color will develop if molybdate is present in the sample. Analyze samples immediately after collection. Filter turbid samples through a 0.45-micron filter membrane apparatus before analysis. (The membrane filter apparatus consists of: Code L6485 filter, Code L773 filter holder, and Code L775 plastic syringe.) Wipe the outside of the sample cells before each insertion into the instrument cell holder. Use a damp towel followed by a dry one to remove fingerprints or other marks. Highly buffered samples or extreme sample ph may exceed the buffering capacity of the reagent and require sample pretreatment. For increased accuracy, determine a reagent blank value for each new lot of reagent. Follow the procedure using deionized water in place of the sample. Subtract the reagent blank value from the final results. Wear safety glasses when working with chemicals. Results outside the working range of this procedure are invalid. Sample temperature must be between 20 25 C (68 77 F) for accurate results. Find a contact near you by visiting www.ge.com/water and clicking on Contact Us. * Trademark of General Electric Company; may be registered in one or more countries. 2008, General Electric Company. All rights reserved..doc Aug-08

Procedure 1. Fill a 25-mL mixing graduated cylinder to the 20-mL mark with sample. Note: Filter turbid samples. See Tips and Techniques for details. 2. Add the contents of one Molybdenum TC1 Reagent Powder Pillow to the cylinder. Stopper the cylinder and invert or gently shake until all particles are dissolved. 3. Split the sample by filling two 10-mL sample cells to the 10-mL mark. Cap one of the cells. (This is the blank.) Note: Exact measurement is not critical for the blank. The other cell must contain 10 ml of sample. 4. Using the calibrated dropper, add 0.5 ml of Molybdenum TC2 to the sample in the uncapped 10- ml cell. Cap the cell and swirl to mix. (This is the prepared sample.) 5. Wait two minutes for full color development. Note: A green color will develop if molybdate is present in the sample. 6. Press the POWER key to turn the meter on. The arrow should indicate the low range (LR). Note: See the Pocket Colorimeter II manual for information on selecting the correct range. 7. Wipe the sample cells free of liquid and fingerprints. 8. Place the blank in the cell holder with the diamond mark towards the keypad. Cover the blank with the instrument cap. Page 2

9. Press ZERO/SCROLL. The meter will show ----, then 0.00. Remove the blank from the cell holder. Note: If the meter reads 0.0, see the Pocket Colorimeter II manual for information on selecting the correct range. 10. Place the prepared sample in the cell holder with the diamond mark towards the keypad. 11. Cover the sample cell with the instrument cap. 12. Press READ/ENTER. The display will show ---- followed by the results in mg/l Molybdate (MoO 4 ). Note: If the display flashes, the result is above the working range of the test. Dilute the sample and repeat the test. Page 3

Interferences Interference studies were conducted by preparing a molybdate standard solution (3.34 mg/l MoO 4 ) as well as a solution of the potential interfering ion. When the standard solution concentration changed by ± 5% with a given ion concentration, the ion was considered an interference. Tables 1 3 list the details of these studies. Table 1 Substances that Cause a Negative Interference Interfering Substance Alum Aluminum AMP (Phosphonate) Bicarbonate Bisulfate Borate Chloride Chromium Copper Diethanoldith-iocarbamate EDTA Ethylene Glycol Iron Lignin Sulfonate Nitrite Orthophosphate Phosphonohydroxyacetic Acid Interference Levels and Treatments Greater than 7 mg/l Greater than 2 mg/l Greater than 15 mg/l Greater than 5650 mg/l Greater than 3300 mg/l Greater than 5250 mg/l Greater than 1400 mg/l Greater than 4.5 mg/l* Greater than 98 mg/l Greater than 32 mg/l Greater than 1500 mg/l Greater than 2% (by volume) Greater than 200 mg/l Greater than 105 mg/l Greater than 350 mg/l Greater than 4500 mg/l Greater than 32 mg/l In system waters, the presence of quaternary ammonium compounds can cause a negative interference based on the level of the compound. If the presence of a quaternary ammonium compound is suspected, or a green color develops after adding Molybdenum TC1 Reagent: 1. Add two brass dippers (Code L113) of Molybdenum Interference Eliminator Powder (Code L1974) to 20 ml of deionized water. Mix thoroughly. Quaternary Ammonium Compounds 2. Add 0.5 ml of this solution to 25 ml of sample. If Spectrus* BD1507 is being used, add 1 ml of the prepared solution to a 25 ml sample to counter interference from up to 50 ppm BD1507. 3. Shake the solution for two minutes. 4. Filter the 25 ml of treated samples through a 0.45 micron filter membrane apparatus (Codes L6485, L773, and L775). HEDP (Phosphonate) 5. Use the filtered sample as the test sample in step 1. The presence of the phosphonate HEDP at concentrations up to 30 mg/l will increase the apparent molybdate concentration reading by approximately 10% (positive interference). Multiply the value obtained in step 12 on page 2 by 0.9 to obtain the actual MoO4 concentration. Sulfite Greater than 6500 mg/l * Read the molybdate concentration immediately after the 2-minute reaction period. Page 4

Table 2 Substances that Cause a Positive Interference Interfering Substance Interference Levels and Treatments Benzotriazole Biocides Carbonate Highly buffered samples or extreme sample ph Morpholine Phosphonate HEDP Silica Greater than 210 mg/l Large interferences are caused by some biocides used in cooling tower samples. Test the ternary complex procedure on molybdate standard containing the specific biocide in use to determine if the ternary complex method will work with these samples. Greater than 1325 mg/l May exceed the buffering capacity of the reagent and require sample pretreatment. Use a ph meter (Code L6606) or ph paper (Code L1255) to measure ph. Adjust the sample ph between 3 and 5 by adding (by drops) an appropriate amount of 1.0 N Sulfuric Acid Standard Solution (Code L283.1000) or 1.0 N Sodium Hydroxide Standard Solution (Code L255.1000). If significant quantities of acid or base are used, a volume correction should be made. Divide the total volume (sample + acid + base) by the original volume. Multiply the test result by this number. Greater than 6 mg/l Positive interference of about 10% up to 30 mg/l. As the concentration increases above 30 mg/l, a decrease in the molybdate concentration reading occurs (negative interference). Greater than 600 mg/l Table 3 Non-interfering Substances Interfering Substance Bisulfite Calcium Chlorine Magnesium Manganese Nickel PBTC (phosphonate) Sulfate Zinc Interference Levels and Treatments 9600 mg/l 720 mg/l 7.5 mg/l 8000 mg/l 1600 mg/l 250 mg/l 500 mg/l 12,800 mg/l 400 mg/l Page 5

Accuracy Check To prepare a 3.34 mg/l molybdate standard solution (as MoO 4 ): 1. Pipet 10 ml from a 10-mg/L Molybdenum (16.7 mg/l MoO 4 ) Standard Solution (Code L2038.0500) into a 50-mL graduated mixing cylinder (Code L2631). 2. Dilute the cylinder to a final volume of 50 ml with Deionized Water (Code L243.4000). Mix thoroughly. 3. Use the standard in place of the sample in the procedure. Note - Use a safety bulb (Code L1575) when pipetting. Method Performance Precision Standard 95% Confidence Limits of Distribution 3.34 mg/l MoO 4 3.27 3.41 mg/l MoO 4 Description Quantity/Test Units Code Required Chemicals* Molybdenum TC1 Reagent, 20-mL sample size 1 100/pkg L2358 Molybdenum TC2 Reagent Solution 0.5 ml 50 ml L2359 Required Apparatus Cylinder, graduated mixing, 25-mL 1 each L2377 Sample Cells, 2.5-cm, 10-mL, with cap 2 6/pkg L2379.0006 Pocket Colorimeter II, Molybdate Kit 1 each L2888 * Wear safety glasses when working with chemicals. Page 6