Speciation of Arsenic in Rice Flour Used as a Capsule Void Filler in Dietary Supplements

Similar documents
Chromium Speciation in Drinking Water by LC-ICP-MS

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Table 1. Toxicology of Arsenic Species. Arsenic Species

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

How To Test For Contamination In Large Volume Water

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Sample Analysis Design Isotope Dilution

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Opiates in Urine by SAMHSA GC/MS

Vitamin C quantification using reversed-phase ion-pairing HPLC

Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Tamsulosin Hydrochloride Capsules

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

SUCRALOSE. White to off-white, practically odourless crystalline powder

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

A Complete Solution for Method Linearity in HPLC and UHPLC

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Analysis of Various Vitamins in Multivitamin Tablets

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

Guidance for Industry

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Elemental Analyses by ICP-AES

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

Appendix 5 Overview of requirements in English

How To Develop A Recipe Card

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

How To Use The Nexion 300

4.2 Bias, Standards and Standardization

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

Implementing New USP Chapters for Analytical Method Validation

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Oasis HLB Cartridges and 96-Well Plates

ERDOSTEINE - MONOGRAPH.

Innovative vs Traditional

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

Analytical Test Report

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

MOISTURE (Karl Fischer, Buffered)

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Supplementary Materials and Methods (Metabolomics analysis)

Thermo Scientific Dionex AS-AP, AS-DV, and AS-HV Autosamplers. Automation, sample preparation. Autosamplers for Ion Chromatography

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Analytical Test Method Validation Report Template

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Drospirenone and Ethinyl Estradiol Tablets. Type of Posting. Revision Bulletin Posting Date. 31 July 2015 Official Date

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Laboratory Procedure Manual

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Transcription:

APPLICATION NOTE LC-ICP-MS Authors: Helmut Ernstberger PerkinElmer, Inc. Shelton, Connecticut James Neal-Kababick Flora Research Laboratories, LLC Grants Pass, Oregon Speciation of Arsenic in Rice Flour Used as a Capsule Void Filler in Dietary Supplements Introduction The contamination of a variety of foods with arsenic has received significant attention recently. For food, regulatory limits have been introduced for rice which distinguish inorganic from organic arsenic forms in order to take into account their significant toxicity differences 1. Rice flour, when used as a capsule filler material, may also constitute a major proportion of dietary supplements. Regulations that limit the amount of inorganic arsenic in dietary supplements are going to become effective under U.S. Pharmacopeia (USP) General Chapter <2232>. Inorganic arsenic in rice is a reasonably anticipated contaminant (RAC) and must be evaluated in rice powder or flour used as an excipient in dietary supplements to comply with the cgmps for dietary supplements under 21 CFR Part 111.

In this paper, we evaluate the speciation of arsenic (As) in several types of rice flour products marketed as excipients for the manufacture of dietary supplements. Any procedure used to assess the speciation of arsenic must be validated for suitability for the particular sample matrix investigated. We validate the methodology presented by analysis of a standard reference material (SRM) and using spike recoveries for the extracted excipient samples. In addition, we widened the range of tested materials to include samples which may not exclusively contain rice. This allows assessing method robustness for coping with sample matrices deviating from the norm. Furthermore, the efficiency of the extraction was assessed by comparison to total arsenic values obtained after sample digestion. Experimental Sample Preparation for LC-ICP-MS Samples (1 g) were extracted in 50 ml Sarstedt tubes with 10 ml of 0.28 M HNO 3 (GFS Chemicals, Veritas grade) on a PerkinElmer SPB 50-24 hotblock at 95 C for 90 min 2. Two distinguishing features of this extraction method are that it has been thoroughly validated for rice, and that it maintains species integrity for extracted species 3. The cooled extracts were centrifuged at 6000 rpm for 5 min. A portion of the supernatant was taken up with a 20 ml NormJect all-plastic syringe and filtered through a 0.45 µm, 25 mm nylon syringe filter unit. The filtrate (3-7 ml) was collected in 15 ml Sarstedt tubes. The extracts were further diluted 10-fold (0.5 ml to 5 ml) with ultrapure water (ELGA PureLab 18.2 MΩ-cm). Instrumental Conditions The analysis was performed with a PerkinElmer Altus HPLC and NexION 350D ICP-MS, controlled by Waters Empower 3 Software. The instrument conditions are given in Table 1. Table 1. Instrument conditions. LC Mobile Phase Altus HPLC 50 mm NH 4 NO 3, 2 mm OSA, 2 mm malonic acid, ph 4.0, 1% MeOH Column Shiseido Capcell PAK, MG, C18, 4.6 x 250 mm, 5 µm Column Temperature 50 C Vials 1.5 ml plastic Injection Volume 10 µl ICP-MS NexION 350D Nebulizer Glass concentric Spray Chamber Glass cyclonic RF Power 1600W Plasma Gas 15 L/min Auxiliary Gas 1.2 L/min Isotope 75 As Mode Standard mode Dwell Time 500 ms The mobile phase was prepared using OSA, 1-octanesulfonic acid, sodium salt (98%, Sigma-Aldrich ) and malonic acid (99%, Acros Organics ). The desired NH 4 NO 3 level was achieved using 3.125 ml/l nitric acid (Veritas grade, GFS Chemicals ) and addition of sufficient ammonium hydroxide (Optima grade, Fisher Scientific ) to achieve the final ph 4.0. Finally, methanol (Optima grade, Fisher Scientific ) was blended into the mobile phase. Quantitation Standards were prepared from 1000 ppm stock solutions for As3 (Spex Certiprep ) and As5 (PerkinElmer). DMA, dimethylarsinic acid (98%) was purchased from Sigma and MMA, monosodium methylarsonate (99%) was obtained from Chem Service. For the organoarsenicals, standards were based on total weight of the chemicals (excluding crystal water for MMA). For reporting purposes, organoarsenical results were converted to elemental As. Standards for LC-ICP-MS were prepared in mobile phase and, for comparison, additionally in the diluted extract matrix (0.028M HNO 3 ). Both matrices give the same arsenic response (Table 2). However, standards in 0.028M HNO 3 are not consistently stable. Species interconversion (As3 oxidation to As5) was observed in that matrix on repeated analysis of a check standard (1 ppb). Arsenic species proved to be stable in mobile phase (ph 4), consistent with previous observations 4. The standards stability over time spanning longer than one day was not investigated in detail. We recommend, as a best practice, to prepare standards fresh each day for best accuracy, considering the low levels (ppt) involved. Table 2. Comparison of 1 ppb standard response (ppb) in mobile phase vs. 0.028M HNO 3 calibration was carried out with standards in mobile phase. Sample Injection As5 MMA As3 DMA 1 ppb 0.028M HNO 3 1 0.98 1.03 1.03 1.01 1 ppb 0.028M HNO 3 2 0.97 0.96 1.04 1.02 1 ppb Mobile Phase 1 1.02 0.98 1.05 1.02 1 ppb Mobile Phase 2 1.05 1.03 1.02 1.02 Samples The tested excipient materials are identified in Table 3. Together with the SRM, NIST 1568b Rice Flour, their extracts were analyzed unspiked and spiked with mixed As standard. Table 3. Sample identification of tested capsule filler materials. ID# Label Description FRL-2 Gluten-Free Mixed Flour Gluten-free mix with rice flour representing complex solid blend 1 FRL-3 Gluten-Free Mixed Flour Gluten-free mix with rice flour representing complex solid blend FRL-4 White Rice Flour Basic white rice flour fill used in industry FRL-5 Brown Rice Flour Basic brown rice flour fill used in industry FRL-8 White Rice Flour Basic white rice flour fill used in industry FRL-12 Rice Concentrate Rice extract from China marketed as capsule filler (emerging material) 2

Total As After Microwave Digestion For totals analysis, 0.25 g of sample material was digested with 10 ml HNO 3 using standard 75 ml vessels in the PerkinElmer Titan MPS Microwave Sample Preparation System, according to the temperature program given in Table 4. Table 4. Microwave digestion program for total arsenic analysis. Step Temperature ( C) Pressure Limit (bar) Ramp (min) Hold (min) Power (W) 1 160 30 5 10 90 2 180 30 3 20 100 3 50 30 1 15 0 Digests were clear for all samples except FRL-12, which contained a substantial particulate residue. The digests were transferred to 50 ml Sartstedt tubes and filled to 50 ml with deionized water. Analysis was carried out after further 5-fold dilution with deionized water, addition of 2 ppb Ir as internal standard, and filtration through 0.45 µm syringe filters. Standards for totals analysis were prepared in 4% HNO 3 in the range 20 ppt - 1 ppb (six standards), and analysis was done in Reaction mode using O 2 as reaction gas (0.6 ml/min; RPq = 0.5). Results and Discussion The instrument was calibrated in the range 50 ppt - 5 ppb (corresponding to 5-500 ppb in the solid) to cover the expected range of arsenic concentrations for the unspiked and spiked samples (2 ppb spike level). The 5-point calibrations were linear with r 2 > 0.999 for all four species. An example chromatogram for a mid-range standard prepared in mobile phase is shown in Figure 1, displaying adequate peak resolution within a short run time. Separation of the species is complete within 2.5 min. The benefits of a short run time include higher throughput and lower detection limits, the result of minimal dispersion experienced at short times (i.e. taller and narrower peaks). As5 As3 Overlaid in Figure 1 are example chromatograms for one sample before and after spiking. Retention times for samples were identical to those observed for standards, and neither peak shape nor resolution was negatively affected by the sample/extraction matrix. Materials consisting not exclusively of rice flour (FRL-2, 3, and 12) behaved identically, demonstrating that the method is robust in terms of coping with some variation in the sample type. To validate the methodology, the NIST 1568b reference material was analyzed. The NIST 1568b results show good agreement with certified values (Table 5). Inorganic As is recovered to 98%, proving reliability of the method for toxicity assessments. Table 5. Standard reference material results in ppb (NIST 1568b) analysis results are reported as elemental As and are corrected to dry weight (µg/kg). Sample As5 As3 Total As Inorganic MMA DMA Sum of Species NIST-E1 31.4 61.8 93.2 11.2 166 270 NIST-E2 32.9 55.1 87.9 10.6 161 260 Certified - - 92 11.6 180 285 Recovery (Avg) - - 98% 94% 91% 93% In order to examine the time sensitivity of the analysis, diluted NIST extracts were reanalyzed after storage for two days at room temperature. Deviations are small and mostly random, averaging ± 1% change for most species (Table 6). For MMA, relative deviations were higher, which may be largely attributable to the low concentration of that species (compare Table 5), and absolute deviations are still small. Nevertheless, this could indicate degradation of DMA to MMA in very small amounts. Diluted NIST extracts, both unspiked and spiked with 1 ppb, were stored for seven days at room temperature and reanalyzed, resulting in spike recoveries of 98%, 103%, 97%, and 91% for As5, MMA, As3, and DMA, respectively (average of duplicates), demonstrating long-term species stability. The excipient materials were analyzed on the day of extract preparation and again on the following day with no major deviations observed; deviations averaged 6%, 2%, 0%, and 2% for As5, MMA, As3, and DMA, respectively, indicating that the diluted extracts of excipient samples are stable for one day. MMA DMA Table 6. Aging of diluted NIST extracts for two days with changes reported as percentage relative to the initial value. Sample As5 MMA As3 DMA Sum of Species NIST-E1 10x -2% 7% 2% -5% -3% NIST-E2 10x 0% 14% -1% 4% 3% Avg -1% 11% 1% -1% 0% Figure 1. Separation of arsenic species achieved using ion interaction chromatography. Overlay of 1 ppb standard in mobile phase, and FRL-5 sample extract (brown rice flour) with and without 2 ppb spike. 3

The speciation results for samples are reported in Figure 2. Results for individual species range from 2-287 ppb with sharp differences in the distribution patterns observed. The dominant individual species could either be organic (FRL-3 and FRL-8) or inorganic (remaining FRL samples). Within the group of inorganic species, As3 was mostly dominating over As5, with one notable exception (FRL-12). This observation demonstrates that it is valuable to collect information on the individual inorganic species. Whilst a summary of inorganic concentration is mostly required for toxicity assessment, information on the nature of the sample would be lost without the ability to distinguish between As3 and As5. In this case, the more detailed information provided by the distinction between As3 and As5 enabled singling out FRL-12 as an unusual sample in line with its identification as concentrate rather than flour (Table 3). As for the organic species, DMA was the dominant form in all cases, with MMA being mostly absent. To further validate the accuracy of the speciation analysis, a spike recovery experiment was carried out. Spike recoveries on diluted extracts averaged 99-107% for individual species, confirming that species were recovered quantitatively and demonstrating that species were stable and did not interconvert (Table 7). Table 7. Spike recoveries obtained on diluted extracts of samples and NIST reference material (extracted in duplicate) spiked with 2 ppb. Sample As5 MMA As3 DMA FRL-2 100% 112% 101% 102% FRL-3 96% 107% 98% 98% FRL-4 102% 113% 99% 104% FRL-5 101% 109% 99% 106% FRL-8 99% 107% 105% 102% FRL-12 99% 103% 101% 102% NIST-E1 97% 103% 99% 101% NIST-E2 94% 104% 97% 99% Avg 99% 107% 100% 102% As [µg/kg] The effectiveness of the extraction of arsenic species was assessed by measuring total arsenic by ICP-MS after microwave digestion of the samples. Analysis of the NIST SRM 1568b digest showed close agreement with the certified value (101% recovery), confirming the accuracy of the analysis. Arsenic totals obtained after digestion are compared with the sum of As species obtained by LC-ICP-MS in Figure 4. The close agreement between the two methods confirms the quantitative extraction of arsenic species. Figure 2. As species distribution for excipient samples showing sharp differences in the distribution of individual species. Inorganic arsenic dominated As speciation for most samples (Figure 3). For two samples (FRL-3 and FRL-8), only half of the arsenic was in inorganic form. Those samples benefit most from toxicity assessment via speciation, as opposed to assessment by total arsenic content. Furthermore, speciation analysis clearly identifies FRL-12 as being of concern with its high arsenic level being mostly attributable to inorganic forms. As in solid [µg/kg] Figure 4. Comparison of sum of species obtained by LC-ICP-MS with totals obtained after microwave digestion. As [µg/kg] Figure 3. Contribution of inorganic arsenic to the total As content of excipient samples. 4

Replicate performance was assessed on two extracted samples analyzed in duplicate. The results show good agreement between the replicates (Table 8). Table 8. Duplicate analysis shows adequate replicate performance. Values are reported as elemental As (µg/kg). Sample As5 MMA As3 DMA Inorganic As Sum of Species % Inorganic As FRL-8 a 3.7 < 2.0 56.8 59.3 60.5 120 51 FRL-8 b 4.5 < 2.0 54.1 56.6 58.6 115 115 FRL-12 a 284 2.0 34.0 13.6 318 334 95 FRL-12 b 287 3.4 36.0 16.6 323 343 94 Conclusion This study examined the performance of the PerkinElmer Altus HPLC coupled to the NexION 350D ICP-MS for the analysis of arsenic species in rice flours used as filler material in dietary supplement capsules. The method employs an acidic extraction with subsequent analysis by ion interaction chromatography. Good agreement of the obtained speciation data with totals measured after microwave digestion confirms that the species are extracted quantitatively. The toxicologically important sum of both inorganic arsenic species was in close agreement with the certified value, as were the other data obtained for the reference material (sum of species, totals is digests). Those data together with spike-recovery experiments confirm that quantitation was accurate in the matrices examined, and that the species are stable once extracted. Distinction between the inorganic arsenic forms proved useful for sample characterization based on differences in the species distribution, highlighting the usefulness in employing a procedure which maintains the integrity of the extracted inorganic species. Samples were stable at least for one day after extraction. The sample/extract matrix did not impact the chromatography as evidenced by unaffected retention times and good peak shape throughout. Performance was equally good for samples off the norm, demonstrating method robustness. In addition to those benefits, the analysis by ion interaction chromatography yields desirable low detection limits and fast analysis times. References 1. Commission Regulation (EU) 2015/1006 Amending Regulation (EC) No 1881/2006 as regards Maximum Levels of Inorganic Arsenic in Foodstuffs. 2. Huang, J. H.; Ilgen, G.; Fecher, P. Quantitative Chemical Extraction for Arsenic Speciation in Rice Grains. J. Anal. At. Spectrom. 2010, 25, 800 802. 3. Huang, J. H.; Fecher, P.; Ilgen, G.; Hu, K. N.; Yang, J. Speciation of Arsenite and Arsenate in Rice Grain - Verification of Nitric Acid Based Extraction Method and Mass Sample Survey. Food Chem. 2012, 130 (2), 453 459. 4. Ernstberger, H.; Neubauer, K. Accurate and Rapid Determination of Arsenic Speciation in Apple Juice; PerkinElmer Application Note; 2015. Consumables Used Component Part Number 0.45 μm Nylon Syringe Filter Units 02542905 Column: C18, 4.6 x 250 mm, 5 μm N8145326 Autosampler Vials, clear, 1.5 ml (package of 100, with caps) N9301736 PEEK Tubing, 0.007 ID x 1/16 OD (5 feet) N9302678 PEEK Finger Tight Fittings 09920513 PEEK Solvent Filter, 10 μm N8122249 PEEK In-line Filter, 10 μm N8122250 Nebulizer Connector for HPLC WE024372 Connector for Peristaltic Pump Tubing to PEEK Tubing N8122258 Finger Tight Connector for 1/16 OD PEEK Tubing 09920513 50 ml Sarstedt Tubes N8145124 15 ml Sarstedt Tubes N0777471 PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2016, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 012553_01 PKI