EGS4 APPLICATION TO DESIGN OF PARALLELL-PLATE FREE-AIR IONIZATION CHAMBER FOR HIGH-ENERGY SYNCHROTRON RADIATION

Similar documents
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Atomic and Nuclear Physics Laboratory (Physics 4780)

Production of X-rays and Interactions of X-rays with Matter

GAMMA-RAY SPECTRA REFERENCES

Introduction to the Monte Carlo method

Vacuum Evaporation Recap

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Lectures about XRF (X-Ray Fluorescence)

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

Polarization Dependence in X-ray Spectroscopy and Scattering. S P Collins et al Diamond Light Source UK

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, Bologna, Italy

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

Monte Carlo Simulations in Proton Dosimetry with Geant4

Introduction to Geiger Counters

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

ACCELERATORS AND MEDICAL PHYSICS 2

Appendix A. An Overview of Monte Carlo N-Particle Software

Information about the T9 beam line and experimental facilities

Chapter XIII. PDD TAR TMR Dose Free Space

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues

University of California at Santa Cruz Electrical Engineering Department EE-145L: Properties of Materials Laboratory

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Heating & Cooling in the Interstellar Medium

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Manual for simulation of EB processing. Software ModeRTL

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

2 Absorbing Solar Energy

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi

1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

How To Understand Light And Color

VARIANCE REDUCTION TECHNIQUES FOR IMPLICIT MONTE CARLO SIMULATIONS

Principles of dosimetry The ionization chamber

Meeting the Grand Challenge of Protecting an Astronaut s Health: Electrostatic Active Space Radiation Shielding for Deep Space Missions

AAPM protocol for kv x-ray beam dosimetry in radiotherapy and radiobiology

Diagnostic x-ray imaging relies on the attenuation of

World-first Proton Pencil Beam Scanning System with FDA Clearance

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:

Proton tracking for medical imaging and dosimetry

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Monte Carlo (MC) Model of Light Transport in Turbid Media

A Quick primer on synchrotron radiation: How would an MBA source change my x-ray beam. Jonathan Lang Advanced Photon Source

ENERGY LOSS OF ALPHA PARTICLES IN GASES

The Existence of a Neutron

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Cathode Ray Tube. Introduction. Functional principle

Damping Wigglers in PETRA III

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Refractive Index Measurement Principle

Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Undulators and wigglers for the new generation of synchrotron sources

On the first Townsend coefficient at high electric field

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

where h = J s

Composite Electromagnetic Wave Absorber Made of Permalloy or Sendust and Effect of Sendust Particle Size on Absorption Characteristics

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

Lawrence Berkeley National Laboratory Lawrence Berkeley National Laboratory

ELECTRON-CLOUD EFFECTS IN THE TESLA AND CLIC POSITRON DAMPING RINGS

Clinical Photon Beams

Coating Thickness and Composition Analysis by Micro-EDXRF

Radiation Detection and Measurement

Determination of Molecular Structure by MOLECULAR SPECTROSCOPY

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

How To Use A Proton For Radiation Therapy

The accurate calibration of all detectors is crucial for the subsequent data

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

Variance reduction techniques used in BEAMnrc

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Infrared Spectroscopy: Theory

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Status of Radiation Safety System at

FDTD Analysis of Site Free Space VSWR in Test Site Used for Disturbance Measurement above 1 GHz

A Guide to Acousto-Optic Modulators

EDXRF of Used Automotive Catalytic Converters

Fast Varifocal Lenses Based on KTa 1-x Nb x O 3 (KTN) Single Crystals

Coating Technology: Evaporation Vs Sputtering

TIME OF COMPLETION NAME SOLUTION DEPARTMENT OF NATURAL SCIENCES. PHYS 3650, Exam 2 Section 1 Version 1 October 31, 2005 Total Weight: 100 points

Magnetic Field of a Circular Coil Lab 12

7üVNORD. TÜV NORD Röntgentechnik

Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector

Rn, Rn and their progeny concentrations in o$ces in Hong Kong

TOF-PET DETECTOR CONCEPT BASED ON ORGANIC SCINTILLATORS

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Acceleration of germination of tomato seed by applying AC electric and magnetic "elds

Transcription:

Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.19-24 EGS4 APPLICATION TO DESIGN OF PARALLELL-PLATE FREE-AIR IONIZATION CHAMBER FOR HIGH-ENERGY SYNCHROTRON RADIATION N. Nariyama 1, N. Kishi 2, T. Itoh 2,S.Taniguchi 1 and T. Moriya 2 1 Japan Synchrotron Radiation Research Institute Kouto 1-1-1, Sayo, Hyogo 679-5198 Japan 2 Kyoto University Yoshida, Sakyo-ku, Kyoto 606-8501 Japan Abstract Energy deposition in two-sized parallell-plate free-air ionization chambers was calculated for 30-200 kev photons with EGS4 to estimate the electron escape fraction for the chambers. For the low-energy x-ray ionization chamber of 5-cm plate separation, the escape increased with energy to 70 kev, decreased from 70 to 100 kev, and above 150 kev rose again. For the highenergy x-ray ionization chamber of 8.5-cm plate separation, the escape became almost below 1% and the charge collection eciencies calculated based on the Boag's equation were over 99% for the exposure rate below 1 R/s at an applied voltage of 10 kv. Inuence of linear polarization was found to be negligible. 1 Introduction At SPring-8 of 8-GeV synchrotron radiation facility in Japan, monoenergetic photons of the energies up to at least 190 kev are available at the bending magnet beamlines. The exposure rate is at most several tens mgy/s, at which a parallell-plate free-air ionization chamber is expected to be applicable. For high-energy x rays, however, size of plate separation becomes extremely large, which lower the electric eld slope. For example, electron range at 100 kev in air is 13.5 cm, which corresponds to 27-cm plate separation necessary. Actually, ionization chambers for up to 250 kv x rays have had 20-cm and 30-cm plate separation at NBS and NPL, respectively[1]. In Japan, as the standard free-air chamber for hard x rays at Electrotechnical Laboratory, 24-cm plate separation has been employed[2]: these facilities have been used for the national standard, in which extremely high accuracy is required. On the other hand, Attix and DeLaVergne have calculated electron ionization losses for various plate separations in a free-air chamber and found that to obtain < 1% loss a mere 12-cm separation is necessary even for 200 kv x rays[3]. Boag has pointed out that the reason is that the proportion of the total ionization due to photoelectrons declines rapidly with energy, and the Compton recoil electrons have much lower energies[4]. Thus if the high accuracy such as required at the national standard facilities is needless and even 1% accuracy is sucient, the large plate separation appears not to be necessary. For the detail estimation, Monte Carlo simulation is indispensable. In this study, energy deposition in ionization chambers was calculated using EGS4 to estimate electron energy loss from the collection plates. Moreover, charge collection eciency was calculated based on Boag's equation[4] at a voltage applied for various exposure rates. 1

2 Compton and Photoelectrons Figure 1 shows the photon attenuation coecients in air for Compton and photoelectric eect, respectively. Above 30 kev, inuence of Compton eect is becoming gradually dominant. Figure 2 shows the energy spectra of Compton electrons for 50- and 100-keV photons, which were calculated based on Klein-Nishina equation. The maximum energies are much smaller than those of photoelectrons. Angular distributions of Compton and photoelectrons for 50- and 100-keV photons are indicated in Figs.3 and 4. The angle of photoelectrons was calculated based on the Sauter distribution[5]. Compton electrons are all ejected below 90 degrees, while some photoelectrons over 90 degrees. From these results, it is expected that in the energy region of 50 to 100 kev where Compton interaction is dominant, plate separation of parallell-plate free-air ionization chamber will not have to be comparable to the electron range of photoelectrons. 3 EGS4 Calculations In EGS4 calculations, 50- to 200-keV photons were incident to parallell-plate ionization chambers and energy deposition between the collection electrodes was calculated. Beam size used was 5 mm square and cuto energies of 1 kev and 5 kev were used for photons and electrons, respectively. Statistical errors were all below 1%. Deposited energies obtained were compared with those calculated using the energy absorption coecients of air. Two kinds of chambers were used: one for 40 kev[6] and the other for > 40keV. Plate separation, guard electrode length and collection electrode width were 5 cm, 2 cm and 7 cm for 40 kev, and 8.5 cm, 15 cm and 20 cm for > 40keV, respectively. Collection electrode length was 1 cm for both chambers. Moreover, inuence of linear polarization was examined. 4 Results and Discussion 4.1 Ionization chamber for low-energy x rays Figure 5 shows the comparison between experiment and calculation of the ionization chamber for the low-energy photons. The experimental values are ratios of absorbed doses in the chamber to those in a Si photodiode[7]. The dierence of unity and the values on the vertical axis denotes the electron escape fraction. The calculated values decreased to 0.82 at 60 kev, reversely increased to 0.91 at 100 kev and decreased again to 0.56 at 200 kev. Compared with experiments, almost agreement was obtained for BL20B2 beamline, while not agreed for BL38B1. This reason can be attributed to the large beam size of 1 cm by 2 cm used at BL20B2. Figure 6 shows the result of build-up thickness, that is, the guard electrode length adjusting to the photon energy. Compared with Fig.5, the calculated values signicantly increased, which showed the large inuence of the buildup region. To examine the inuence of plate separation, guard electrode length and collection electrode width, each length was changed and energy deposition was calculated for 80-keV photons. The result is tabulated in Table 1. In the same way asabove, eect of guard electrode length was found to be the greatest, which is 11%. Consequently, the dierence of experiment and calculation at BL38B1 in Fig.5 is principally attributed to the modelling of the build-up air layer that exists outside of the chamber. 4.2 Ionization chamber for high-energy x rays Figure 7 indicates the EGS4 calculation result of ionization chamber for the high-energy x rays. Under 150 kev, the electron energy losses became below 1%. At 200 kev, however, electron escape 2

fraction rises to 4% because even the Compton electron energies become large and then longer buildup layer is required. Inuence of scattering from the chamber wall was not found owing to an EGS4 calculation. Charge collection eciency f of the present model was calculated based on the Boag's equation[4]: 1 f = 1+ 1 6 2 p = m( d2 q V ) m=36.7 d: Plate separation (cm) q: Ionization rate (esu/cm 3 /s) V: Applied voltage (V) The equation and parameters are applicable in the range f>0.7. Figure 8 shows the eciencies calculated at the applied voltage of 10 kv. An eciency of > 99% is obtained up to 1 R/s, which is adequate exposure rate at the bending-magnet beamlines. Angle of secondary electrons ejected in the photon interaction depends on the direction of linear polarization. Photoelectrons are easy to be ejected in the plane including photon electric eld[5]. On the other hand, Compton electrons tend to be strongly scattered vertically to the plane: both eects compensate each other. Consequently, owing to EGS4 calculations considering linear polarization, large inuence was not also found clearly. 5 Conclusions Calculations with EGS4 have showed that the electron escape fraction is below 1-2% at a plate distance of 8.5 cm for 50- to 150-keV photons. The charge collection eciencies at 10 kv were also over 99% owing to a theoretical calculations. The distance of 8.5 cm is much shorter than the twice of the electron range of 26.5 cm at 150 kev. References [1] G.H. Aston and F.H. Attix, \An intercomparison of the roentgen standards of Great Britain and U.S.A.", Acta. Radiol. 46, 747-752(1956). [2] K. Matsumoto and J. Naoi, \Resettings of exposure standard for medium and hard x rays", Electrotechnical Laboratory Ihou 47 (1983)825 [in Japanese]. [3] F.H. Attix and L. DeLaVergne, \Plate separation requirements for standard free-air ionization chambers", Radiology 63, 853-866(1954). [4] J.W. Boag, \Ionization chambers", Radiation Dosimetry II ed. F.H. Attix, 1-72 (1966). [5] C.M. Davisson and R.D. Evans, \Gamma-ray absorption coecients", Rev. Med. Phys. 24, 79-107(1952). [6] S. Ban, et al., \Measurement of photon energy absorption coecient of air, nitrogen and argon at 30 kev", Appl. Radiat. Isot. 44, 769-772(1993). [7] T. Itoh, \Development of beam ux monitors for intense undulator X rays", Kyoto University Nuclear Engineering Master Thesis (2001) [in Japanese]. 3

Table 1: Inuence of guard electrode length, width and plate separation at 80 kev. Energy deposition was normalized at the maximum. Guard electrode Width Plate separation Deposited length (cm) (cm) (cm) energy 2 7 5 0.815 9 7 5 0.917 9 18 5 0.935 9 18 9 0.987 9 18 18 1.000 Figure 1: Proportion of Compton and photoelectric eect in air. Figure 2: Energy of Compton electrons. 4

Figure 3: Angular distribution of Compton electrons. Figure 4: Angular distribution of photoelectrons. Figure 5: Comparison of experiments and calculation for low energy x-ray ionization chamber. 5

Figure 6: Calculations for build-up layer of adequate thickness. Figure 7: Calculations for high energy x-ray ionization chamber. Figure 8: Analytical calculation of charge collection eciency based on Boag's equation for high energy x-ray ionization chamber at 10 kv. 6