MiSeq Reagent Kit v2 Reagent Preparation Guide

Similar documents
Illumina Security Best Practices Guide

Decode File Client User Guide

TruSeq Custom Amplicon v1.5

MiSeq Reporter Generate FASTQ Workflow Guide

Sequencing Library qpcr Quantification Guide

iscan System Quick Reference Guide

Rapid Aneuploidy and CNV Detection in Single Cells using the MiSeq System

RNA Express. Introduction 3 Run RNA Express 4 RNA Express App Output 6 RNA Express Workflow 12 Technical Assistance

MiSeq System User Guide FOR RESEARCH USE ONLY

qpcr Quantification Protocol Guide

Illumina Sequencing Technology

Data Processing of Nextera Mate Pair Reads on Illumina Sequencing Platforms

TruSeq DNA Methylation Library Preparation Guide

Illumina Q Financial Results. July 21, 2015

Illumina. LIMS Project Manager Guide. For Reasearch Use Only. ILLUMINA PROPRIETARY Catalog # SW Part # Rev. C

GenomeStudio Data Analysis Software

Absolute Quantifi cation of Gene Expression using SYBR Green in the Eco Real-Time PCR System

Preparing Samples for Sequencing Genomic DNA

Consistent Assay Performance Across Universal Arrays and Scanners

GenomeStudio Data Analysis Software

Go where the biology takes you. Genome Analyzer IIx Genome Analyzer IIe

PicoMaxx High Fidelity PCR System

Nextera XT Library Prep: Tips and Troubleshooting

HiSeq Analysis Software v0.9 User Guide

BacReady TM Multiplex PCR System

Paired-End Sample Preparation Guide

Supplies 1. Toner cartridges. Printing. Using color. Paper handling. Maintenance. Troubleshooting. Administration. Index

BioResearch. RAFT 3D Cell Culture Kit Protocol

QIAsymphony RGQ Protocol Sheet

FOR REFERENCE PURPOSES

Paired-End Sequencing Sample Preparation Guide

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN or ).

GenScript BloodReady TM Multiplex PCR System

MystiCq microrna cdna Synthesis Mix Catalog Number MIRRT Storage Temperature 20 C

Updating the QIAcube operating software

Aptima Specimen Transfer Kit

One Shot TOP10 Competent Cells

TruSeq RNA Sample Preparation v2 Guide FOR RESEARCH USE ONLY

HiPer RT-PCR Teaching Kit

ab StayBlue/AP Plus Stain Kit Instructions for Use An immunohistochemical chromogen substrate for staining tissue sections

QIAGEN Supplementary Protocol

Notice. DNA Sequencing Module User Guide

ProteinChip Energy Absorbing Molecules (EAM)

Protocol. Introduction to TaqMan and SYBR Green Chemistries for Real-Time PCR

Genomic DNA Extraction Kit INSTRUCTION MANUAL

Human Free Testosterone(F-TESTO) ELISA Kit

The fastest spin-column based procedure for purifying up to 10 mg of ultra-pure endotoxin-free transfection-grade plasmid DNA.

Araldite LY 3505* / Hardeners XB 3403* / Hardener XB * / Aradur 3405*

Using Illumina BaseSpace Apps to Analyze RNA Sequencing Data

empcr Amplification Method Manual - Lib-A

Path-ID Multiplex One-Step RT-PCR Kit

ID kit. imegen Anchovies II. and E. japonicus) DNA detection by. User manual. Anchovies species (E. encrasicolus. sequencing.

Data Sheet. PD-L1:B7-1[Biotinylated] Inhibitor Screening Assay Kit Catalog # Size: 96 reactions

bcl2fastq2 Conversion Software User Guide

Malondialdehyde (MDA) ELISA

LM135-LM235-LM335. Precision temperature sensors. Features. Description

UltraClean Soil DNA Isolation Kit

Follicle Dermal Papilla Cell

Sequencing Analysis Software User Guide

Accelerate genomic breakthroughs in microbiology. Gain deeper insights with powerful bioinformatic tools.

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #:

ncounter Gene Expression Assay Manual Total RNA and Cell Lysate Protocols

Statement of Work. Next Business Day LCD, Monitor & Television Exchange Service

SEQUENCING. From Sample to Sequence-Ready

Agencourt RNAdvance Blood Kit for Free Circulating DNA and mirna/rna Isolation from μL of Plasma and Serum

[ Care and Use Manual ]

Cell Sorting. S3e Cell Sorter Quick Guide

ab Hi-Fi cdna Synthesis Kit

ISO INTERNATIONAL STANDARD

Global Economic Briefing: Global Inflation

SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting

Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit

16S Metagenomic Sequencing Library Preparation Preparing 16S Ribosomal RNA Gene Amplicons for the Illumina MiSeq System

APPENDIX N BULK MILK TANKER SCREENING TEST FORM

HUMULIN 70/30 KwikPen

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

DeCyder Extended Data Analysis (EDA) Software

Experiment 10 Enzymes

Trace Dissolved Oxygen Sensor

xerox Xerox D95/D110/125/136 Warranty

Blood Collection and Processing SOP

Bovine Vitamin B12 (VB12) ELISA Kit

Image Lab Software for the GS-900 Densitometer

CompleteⅡ 1st strand cdna Synthesis Kit

Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit

Cluster Generation. Module 2: Overview

RT-PCR: Two-Step Protocol

Thermo Scientific Barnstead TII Pure Water Systems. Pure water. Pure research. Total confidence.

Statement of Work. Next Business Day Axim Exchange Service (Axim Exchange Service)

QuarkXPress 8.01 ReadMe

Arcturus PicoPure RNA Isolation Kit. User Guide

RT rxns. RT rxns TRANSCRIPTME Enzyme Mix (1) 40 µl 2 x 50 µl 5 x 40 µl

Tipping Bucket Rain Gauge Models 6011-A 6011-B

Taq98 Hot Start 2X Master Mix

BD238. Low voltage PNP power transistor. Features. Applications. Description. Low saturation voltage PNP transistor

Applied Biosystems 7500/7500 Fast Real-Time PCR Systems. SUBJECT: Memory Upgrade Requirements for 7500 Software v2.0

BD241A BD241C. NPN power transistors. Features. Applications. Description. NPN transistors. Audio, general purpose switching and amplifier transistors

Illumina TruSeq DNA Adapters De-Mystified James Schiemer

Transcription:

MiSeq Reagent Kit v2 Reagent Preparation Guide FOR RESEARCH USE ONLY Revision History 3 MiSeq Reagent Kits 4 Kit Contents and Requirements 6 Flow Cell Types 7 Software and Flow Cell Compatibility 8 Reagent Cartridge Overview 9 Preparing Reagents 11 What's Next 13 Technical Assistance ILLUMINA PROPRIETARY Part # 15034097 Rev. B October 2012

This document and its contents are proprietary to Illumina, Inc. and its affiliates ("Illumina"), and are intended solely for the contractual use of its customer in connection with the use of the product(s) described herein and for no other purpose. This document and its contents shall not be used or distributed for any other purpose and/or otherwise communicated, disclosed, or reproduced in any way whatsoever without the prior written consent of Illumina. Illumina does not convey any license under its patent, trademark, copyright, or common-law rights nor similar rights of any third parties by this document. The instructions in this document must be strictly and explicitly followed by qualified and properly trained personnel in order to ensure the proper and safe use of the product(s) described herein. All of the contents of this document must be fully read and understood prior to using such product(s). FAILURE TO COMPLETELY READ AND EXPLICITLY FOLLOW ALL OF THE INSTRUCTIONS CONTAINED HEREIN MAY RESULT IN DAMAGE TO THE PRODUCT(S), INJURY TO PERSONS, INCLUDING TO USERS OR OTHERS, AND DAMAGE TO OTHER PROPERTY. ILLUMINA DOES NOT ASSUME ANY LIABILITY ARISING OUT OF THE IMPROPER USE OF THE PRODUCT(S) DESCRIBED HEREIN (INCLUDING PARTS THEREOF OR SOFTWARE) OR ANY USE OF SUCH PRODUCT(S) OUTSIDE THE SCOPE OF THE EXPRESS WRITTEN LICENSES OR PERMISSIONS GRANTED BY ILLUMINA IN CONNECTION WITH CUSTOMER'S ACQUISITION OF SUCH PRODUCT(S). 2012 Illumina, Inc. All rights reserved. Illumina, illuminadx, BaseSpace, BeadArray, BeadXpress, cbot, CSPro, DASL, DesignStudio, Eco, GAIIx, Genetic Energy, Genome Analyzer, GenomeStudio, GoldenGate, HiScan, HiSeq, Infinium, iselect, MiSeq, Nextera, NuPCR, SeqMonitor, Solexa, TruSeq, VeraCode, the pumpkin orange color, and the Genetic Energy streaming bases design are trademarks or registered trademarks of Illumina, Inc. All other brands and names contained herein are the property of their respective owners.

Revision History Part # Revision Date Description of Change 15034097 B October 2012 15034097 A July 2012 Added the following information: Kit descriptions of the MiSeq Reagent Micro Kit v2 and MiSeq Reagent Nano Kit v2 Flow cell descriptions and output for micro and nano flow cells Software and flow cell compatibility Initial release. Revision History MiSeq Reagent Kit Reagent Preparation Guide 3

MiSeq Reagent Kits To perform a run on the MiSeq, you need one single-use MiSeq Reagent Kit, which is available in three kit types. Each type of MiSeq reagent kit includes a kit-specific flow cell type and reagents required for performing one run. Kit Type Flow Cell Type Available Kit Sizes MiSeq Reagent Kit v2 Includes a standard (14 tile) flow cell 500 cycles 300 cycles 50 cycles MiSeq Reagent Micro Kit v2 Includes a micro (4 tile) flow cell 300 cycles MiSeq Reagent Nano Kit v2 Includes a nano (2 tile) flow cell 500 cycles 300 cycles The flow cell, PR2 bottle, and reagent cartridge provided in the kit use radiofrequency identification (RFID) for accurate consumable tracking and compatibility. You must use the reagent cartridge associated with your flow cell type. If the reagent cartridge is not compatible, a message appears on the screen during the run setup steps. To proceed, you must load a compatible reagent cartridge. Reagent Kit Sizes Reagent kit size is measured by the number of cycles that can be performed with one kit. Possible sizes are 500 cycles, 300 cycles, and 50 cycles. Kit Size 500-Cycle Kit 300-Cycle Kit 50-Cycle Kit Number of Cycles Provides reagents for up to 525 cycles of sequencing, which is sufficient for up to a 251-cycle paired-end run plus two eight-cycle index reads. Provides reagents for up to 325 cycles of sequencing, which is sufficient for up to a 151-cycle paired-end run plus two eight-cycle index reads. Provides reagents for up to 75 cycles of sequencing, which is sufficient for up to a 26-cycle paired-end run plus two eight-cycle index reads. 4 Part # 15034097 Rev. B

MiSeq Reagent Kit v2 Consumable Catalog # MiSeq Reagent Kit v2, 500 Cycles Catalog # MS-102-2003 MiSeq Reagent Kit v2, 300 Cycles Catalog # MS-102-2002 TG MiSeq Reagent Kit v2, 300 Cycles Catalog # TG-142-1003 MiSeq Reagent Kit v2, 50 Cycles Catalog # MS-102-2001 MiSeq Reagent Micro Kit v2 Consumable Catalog # MiSeq Reagent Micro Kit v2, 300 Cycles Catalog # MS-103-1002 TG MiSeq Reagent Micro Kit v2, 300 Cycles Catalog # TG-142-1002 MiSeq Reagent Kits MiSeq Reagent Nano Kit v2 Consumable Catalog # MiSeq Reagent Nano Kit v2, 500 Cycles Catalog # MS-103-1003 MiSeq Reagent Nano Kit v2, 300 Cycles Catalog # MS-103-1001 TG MiSeq Reagent Nano Kit v2, 300 Cycles Catalog # TG-142-1001 MiSeq Reagent Kit Reagent Preparation Guide 5

Kit Contents and Requirements All MiSeq reagent kits are packaged in two boxes. As soon as you receive your kit, promptly store the kit components at the indicated temperature to ensure proper performance. Box 1: Store at -15 to -25 C Quantity Component Storage Description 1 Reagent Cartridge -15 to -25 C Single-use pre-filled cartridge 1 HT1-15 to -25 C 5 ml tube, Hybridization Buffer Box 2: Store at 2 to 8 C Quantity Component Storage Description 1 PR2 Bottle 2 to 8 C 500 ml bottle, Incorporation Buffer 1 MiSeq Flow Cell 2 to 8 C Single-use PE flow cell Preparation Requirements A B C D Flow Cell Must be thoroughly cleaned prior to use. See the MiSeq System User Guide, Part # 15027617. Reagent Cartridge Must be thawed and inspected prior to use. PR2 Bottle No preparation is required. HT1 (not shown) Must be thawed and chilled prior to use. 6 Part # 15034097 Rev. B

Flow Cell Types The flow cell is a single-lane glass-based substrate on which clusters are generated and the sequencing reaction is performed. During the sequencing run, the single lane of the flow cell is imagined in small imaging areas called tiles. The flow cell is immersed in storage buffer in a flow cell container. The flow cell container cap color indicates the flow cell type: } Clear cap Cap color for a standard (14 tile) flow cell. } Green cap Cap color for a micro (4 tile) flow cell. } Yellow cap Cap color for a nano (2 tile) flow cell. Flow cell types differ in output and number of tiles imaged. Dual-surface imaging, which is possible on the standard and micro flow cells, doubles the number of tiles imaged. Flow cell output depends on the number of reads possible from each flow cell. Reads are measured as the number of clusters passing filter. Flow Cell Types Flow Cell Tiles Flow Cell Tiles Imaging Surface Total Tiles Imaged Standard Flow Cell 14 tiles Top and bottom 28 tiles total Micro Flow Cell 4 tiles Top and bottom 8 tiles total Nano Flow Cell 2 tiles Top only 2 tiles total Flow Cell Output Flow Cell # of Reads Kit Size (Cycles) Output (Max.) 2 x 75 Output 2 x 150 Output 2 x 250 Output Standard Flow Cell 15 M 50, 300, 500 7.5 Gb 2.25 Gb 4.5 Gb 7.5 Gb Micro Flow Cell 4 M 300 1.2 Gb 600 Mb 1.2 Gb -- Nano Flow Cell 1 M 300, 500 500 Mb 150 Mb 300 Mb 500 Mb MiSeq Reagent Kit Reagent Preparation Guide 7

Software and Flow Cell Compatibility Previous versions of the MiSeq Control Software (MCS) cannot process micro or nano flow cells. Additionally, previous versions of the software image the standard flow cell in 12 tiles instead of 14 tiles. You must use a compatible version of MCS to read the RFID of the flow cell and determine how many tiles and surfaces to image. Flow Cell Standard Flow Cell Micro Flow Cell Nano Flow Cell Required MCS Version MCS v2.0 or later MCS v2.1 or later MCS v2.0 or later For more information including requirements for dual-surface flow cell imaging, see Version Compatibility and Requirements in the MiSeq System User Guide, Part # 15027617. 8 Part # 15034097 Rev. B

Reagent Cartridge Overview The MiSeq reagent cartridge is a single-use consumable consisting of foil-sealed reservoirs pre-filled with clustering and sequencing reagents sufficient for sequencing one flow cell. Each reservoir on the cartridge is numbered. Prior to beginning the sequencing run, sample libraries are loaded onto the cartridge in position 17, which is labeled Load Samples. Figure 1 Reagent Cartridge Reagent Cartridge Overview Table 1 Reagent Cartridge Contents Position Reagent Name Description 1 IMS Incorporation Mix 2 SRE Scan Mix 4 CMS Cleavage Mix 5 AMS1 Amplification Mix, Read 1 MiSeq Reagent Kit Reagent Preparation Guide 9

Position Reagent Name Description 6 AMS2 Amplification Mix, Read 2 7 LPM Linearization Premix 8 LDR Formamide 9 LMX1 Linearization Mix 10 LMX2 Read 2 Linearization Mix 11 RMF Resynthesis Mix 12 HP10 Read 1 Primer Mix 13 HP12 Index Primer Mix 14 HP11 Read 2 Primer Mix 15 PW1 Laboratory-grade water 16 PW1 Laboratory-grade water 17 Empty Load Samples (Reserved for sample libraries) 18 Empty Optional use for custom Read 1 primer 19 Empty Optional use for custom Index Read primer 20 Empty Optional use for custom Read 2 primer 21 PW1 Laboratory-grade water 22 Empty Empty NOTE For more information about using custom primers on the MiSeq reagent cartridge, see the MiSeq System User Guide, Part # 15027617. 10 Part # 15034097 Rev. B

Preparing Reagents The following instructions describe how to thaw reagents using a room temperature water bath. This method requires approximately one hour. NOTE Alternatively, you can thaw reagents overnight in 2 to 8 C storage. Reagents are stable up to one week when stored at this temperature. 1 Remove the tube of HT1 (Hybridization Buffer) from -15 to -25 C storage and set aside at room temperature to thaw. When thawed, store at 2 to 8 C until you are ready to dilute denatured libraries. (This step is not required for TruSeq Amplicon libraries.) For more information, see the MiSeq System User Guide, Part # 15027617. 2 Remove the reagent cartridge from -15 to -25 C storage. 3 Place the reagent cartridge in a water bath containing enough room temperature deionized water to submerge the base of the reagent cartridge up to the water line printed on the reagent cartridge. Do not allow the water to exceed the maximum water line. 4 Allow the reagent cartridge to thaw in the room temperature water bath for approximately one hour or until completely thawed. 5 Remove the cartridge from the water bath and gently tap it on the bench to dislodge water from the base of the cartridge. Dry the base of the cartridge. Make sure that no water has splashed on the top of the reagent cartridge. Preparing Reagents Inspect the Reagent Cartridge 1 Invert the reagent cartridge ten times to mix the thawed reagents, and then visually inspect that all positions are thawed. 2 Visually inspect the reagent in position 1 to make sure that it is fully mixed and free of precipitates. 3 Gently tap the cartridge on the bench to reduce air bubbles in the reagents. NOTE The MiSeq sipper tubes go to the bottom of each reservoir to aspirate the reagents, so it is important that the reservoirs are free of air bubbles. MiSeq Reagent Kit Reagent Preparation Guide 11

4 Place the reagent cartridge on ice or set aside at 2 to 8 C until you are ready to set up your run. For best results, proceed directly to loading your sample and setting up the run. WARNING This set of reagents contains formamide, an aliphatic amide that is a probable reproductive toxin. Personal injury can occur through inhalation, ingestion, skin contact, and eye contact. Dispose of containers and any unused contents in accordance with the governmental safety standards for your region. For more information, see the MSDS for this kit, at http://www.illumina.com/msds. 12 Part # 15034097 Rev. B

What's Next After you have properly thawed the reagent cartridge, you are ready to proceed to the following steps described in the MiSeq System User Guide, Part # 15027617: 1 Denature and dilute your libraries for cluster generation and sequencing (unless you are sequencing TruSeq Amplicon libraries). 2 Load your libraries onto the reagent cartridge in the reservoir labeled Load Samples. 3 Using the MiSeq Control Software (MCS) interface, follow the run setup steps to load the flow cell and reagents, and then start the run. Visit the MiSeq support pages on the Illumina website for access to documentation, software downloads, and frequently asked questions. To view a comprehensive list of MiSeq training courses, go to www.illumina.com/training/miseq. What's Next MiSeq Reagent Kit Reagent Preparation Guide 13

Notes

Technical Assistance For technical assistance, contact Illumina Technical Support. Table 2 Illumina General Contact Information Illumina Website http://www.illumina.com Email techsupport@illumina.com Table 3 Illumina Customer Support Telephone Numbers Region Contact Number Region Contact Number North America 1.800.809.4566 Italy 800.874909 Austria 0800.296575 Netherlands 0800.0223859 Belgium 0800.81102 Norway 800.16836 Denmark 80882346 Spain 900.812168 Finland 0800.918363 Sweden 020790181 France 0800.911850 Switzerland 0800.563118 Germany 0800.180.8994 United Kingdom 0800.917.0041 Ireland 1.800.812949 Other countries +44.1799.534000 Technical Assistance MSDSs Material safety data sheets (MSDSs) are available on the Illumina website at http://www.illumina.com/msds. Product Documentation You can obtain PDFs of additional product documentation from the Illumina website. Go to http://www.illumina.com/support and select a product. To download documentation, you will be asked to log in to MyIllumina. After you log in, you can view or save the PDF. To register for a MyIllumina account, please visit https://my.illumina.com/account/register. MiSeq Reagent Kit Reagent Preparation Guide

Illumina Headquartered in San Diego, California, U.S.A. +1.800.809.ILMN (4566) +1.858.202.4566 (outside North America) techsupport@illumina.com www.illumina.com