10A Micro-Ohmmeter Model 6240

Similar documents
Multi-Function Ground

Digital Ground Resistance Tester. Models 4620 & Digital Ground Resistance Tester

5000V Graphical Megohmmeter Model 5070

Digital Transformer Ratiometer DTR Model 8500

Model POWER QUALITY ANALYZERS, METERS & LOGGERS Power Quality Meter

Loc-1Tx/Loc-5Tx/Loc-10Tx Data Sheet V3.3

Hand-held thermometer Model CTH7000

MIT510/2, MIT520/2 and MIT1020/2

DC Electronic Loads 8500 series

JNIOR. Overview. Get Connected. Get Results. JNIOR Model 310. JNIOR Model 312. JNIOR Model 314. JNIOR Model 410

Handheld portable appliance testers

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

User Guide. 4 wire Earth Resistance Tester Model GRT300

User Guide. Model Insulation Tester / Megohmmeter

ACL 395 Resistivity Meter

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

Manual Ranging MultiMeter

Uninterruptible Power Supply

DET3 and 4 Contractor Series Earth/Ground Resistance Testers

1.44 kw Programmable DC Power Supplies XLN Series

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

W O R L D L E A D E R I N p r e c i s i o n i n s t r u m e n t s

Simple Logger II Data Loggers

TEGAM Test Leads Work With Most Manufacturers Four-wire Ohmmeters

TORKEL 820 Battery Load Unit

Voltech DC1000. Precision DC Bias Current Source

Emergency lighting units EM powerled

HCS-3300/3302/3304 USB Remote Programmable Laboratory Grade Switching Mode Power Supply

Electronics Ltd. Presenting. Power, Intelligence... with a sense of feeling

Ron 3025 Crane Scale with 1"/25mm Display

Industrial-Grade UPS System Heavy-duty power protection for harsh industrial environments

SIMATIC Flat Panel Monitors

Dielectric Withstand and Ground Bond Testers For Production Line Safety Agency Compliance Testing

How To Use A Power Supply Unit (Upu)

Solar Controller / Battery Charger User s Manual

User Manual 1K/2K/3K Online UPS

Power-monitoring unit PowerLogic System. Technical data sheet. Power Meter Series 800

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

User s manual FLIR VP50/VP52 Non-contact AC voltage detector

Insulation resistance testing. A complete solution for every person.

1650 Series Multifunction Installation Testers

PRODUCT PROFILE FEATURES

ACL 800 MEGOHMMETER OPERATION MANUAL

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

Ground Resistance Clamp On Tester

USB-500 Series Stand-Alone, Low-Cost Data Loggers

T5000 T000 LCD Digital Fan Coil Thermostat

User Manual. EtherUSB

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE

Zebra RW 420 Mobile Battery Eliminator Kits

Troubleshooting Guide, Freedom and Fleet Power Inverter/Chargers

MEGGER Insulation & Low Resistance Tester

Owner s Manual AWM910 JENSEN AWM910 COMPACT DISC PLAYER RADIO CD COMPACT MUSIC SYSTEM MUTE AUX BAND AUX IN PUSH PUSH PWR VOL ALARM T/F AUD SPK A SPK B

Thyristor-controlled power supplies and battery chargers

Selection Guide to Clamp-On Current Probes

Power Meter Series 700

User's Guide. True RMS Industrial Multimeter

AutoRanging Digital MultiMeter

FIREDEX Conventional Fire Panels

CM705B - Universal Expander Module CM707B - Plug On Zone Expander Security Systems

Density/Specific Gravity/ Concentration Meter DMA 500

PAT120, PAT150 and PAT150R Handheld portable appliance testers

T Substation Maintenance and Commissioning Test Equipment

Everything you need for protection scheme testing

Galleon 6-10K Online UPS. Voltronic Power Technology Corporation

Programmable Room Thermostat 7 Day (5-2 Day) Models: / Power Supply: Battery / Mains

Auto-ranging Digital Multimeter INSTRUCTION MANUAL

T0118 T2118 T3118. Instruction Manual

Process modules Digital input PMI for 24 V DC inputs for 120 V AC inputs

INSTRUCTION MANUAL PLEASE READ ALL THE INSTRUCTIONS COMPLETELY BEFORE USE AND SAVE THIS MANUAL FOR FUTURE REFERENCE

USB-500 Series Low-Cost Data Loggers and Accessories

REGULINE 600VA / 1000VA REGULATOR USER MANUAL

USER MANUAL Online UPS 1K/2K/3K Rack

APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out

How To Use A Cdm250 Digital Multimeter

ExStik TM ph Waterproof Meters

HP UPS R1500 Generation 3

Name Description Model Number. Parameters Min. Typ. Max. Note. Vaux Voltage 9.8 V 12 V 13.2 V Auxiliary Supply Voltage

Nucleus Freedom BTE Rechargeable Battery (A25) and Nucleus Freedom BTE Recharger

Online UPS. PowerWalker VFI 1000R/1U

HP ProDisplay P19A LED Backlit Monitor. User Guide

Current Monitoring Kit QUICK START GUIDE

Product Information Sheet UH28CS-12kV Hi Pot Tester

NEULOG HEART RATE AND PULSE LOGGER SENSOR GUIDE

User Manual for CH-PFC76810

Features. Display. Measurements. Intelligent. Accuracy. Models. Installation DEIF A/S. Multi-instrument D

Easy Installation All five standard monitors fit into shallow mounting areas, requiring about 3" of depth. The XT1502, SXT1811, and SXT2011 models hav

Operation Guide VIP-45-WIFI. Video Inspection Probe

12 Volt 30 Amp Digital Solar Charge Controller

33.6W Power over Ethernet Waterproof Adapter PoE Plus Single Port Injector for Outdoor Application

TX OPEN RS232/485 module

Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry

Industrial Tower Uninterruptible Power Supply Specifications Bulletin Numbers 1609-P

SNMP Web Management. User s Manual For SNMP Web Card/Box

PC Personal 120V 600VA 375W Standby UPS with Pure Sine Wave Output, Tower, 6 Outlets

USB-500/600 Series Low-Cost Data Loggers and Accessories

4-Channel Thermometer / Datalogger

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS

Type: EASY719 DC RC Article No.: Ordering information Relay outputs Quantity 6 Power supply V DC 24 V DC. Description

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

Transcription:

10A Micro-Ohmmeter Model 6240 Selectable test current and resistance ranges 0.25% accuracy and 1µΩ resolution Simple one button press-to-start operation Software temperature correction for accurate measurements Automatically takes multiple measurements by simply moving the test probes from point to point Front panel buttons for polarity reversal, data storage, data recall and display options Stores up to 99 test results Large easy-to-read backlit electroluminescent display Data storage, analysis and report generation through DataView software included

Model 6240 checks contact resistance on a ground bus. The 10A Micro-Ohmmeter Model 6240 is a rugged, low resistance tester designed for both plant and field use. It is a cost effective, feature rich instrument designed for measuring contact resistance, bond resistance and other low resistance connections. The Micro-Ohmmeter Model 6240 accurately measures very low resistance with 0.25% accuracy utilizing a four-lead Kelvin method of testing. An internal NiMH battery pack is rechargeable from 110/230V, 50/60Hz. A fully charged battery can perform up to 850, 10A tests. Resistance measurements are automatically calculated from 1µΩ to 400Ω with resolutions down to 1µΩ. The Model 6240 has selectable test currents of 10A, 1A, 100mA and 10mA and includes a polarity reversal function. The Model 6240 is designed to conduct tests on both resistive and inductive material. Polarity reversal is easily accomplished with a push of a button. The resistance measurements of both positive and negative test results, as well as the average, are displayed. A large easy-to-read multi-line backlit display indicates all test result information including the resistance measurement, test current and test voltage. An automatic test mode feature provides for repetitive tests by simply moving the test probes from point to point, displaying the test results and automatically storing the reading in the instrument s internal memory for later recall and report generation. Up to 99 measurements can be stored and printed directly to a printer or downloaded to a PC for analysis and report generation. An optically isolated USB port provides quick and easy downloads while protecting the PC from any electrical interference. The DataView software included allows you to: download data, temperature compensate the test results, store data and print reports using predefined templates or custom templates you create. Features Reliable low resistance measurements Four-terminal Kelvin resistance measurement eliminates test lead resistance 10A test current at up to 4000µΩ 0.25% basic accuracy 1µΩ resolution Direct reading, easy to operate Six selectable resistance ranges Polarity reversal button Overload and input fuse protection Manufactured to international safety and environmental standards Automatic scaling and zeroing Large terminals accept banana plugs and spade lugs Rechargeable NiMH battery with internal charger (110/230V) Rugged, double insulated waterproof case Includes DataView software for instrument configuration, data storage, analysis and report generation applications Aerospace metallic coating resistance measurement Bonding verification on earth/ground systems Weld joint integrity verification Contact resistance measurement of breakers and switchgear Aircraft and rail bonding checks Wire to terminal connections and resistance checks Battery strap resistance checks Cable joint and bus bar connection checks Mechanical bond tests

Specifications Model 6240 ELECTRICAL Range 4000µΩ 40mΩ 400mΩ 4000mΩ 40Ω 400Ω Accuracy 0.25% ± 2cts 0.25% ± 2cts 0.25% ± 2cts 0.25% ± 2cts 0.25% ± 2cts 0.25% ± 2cts Resolution 1µΩ 10µΩ 0.1mΩ 1mΩ 10mΩ 100mΩ Test Current 10A 1A 1A 100mA 10mA 10mA Measurement Mode Single test start/stop or auto test and store of multiple measurements Power Source Rechargeable 6V, 8.5 Ah NiMH batteries Battery Life Up to 850, 10A tests Battery Charging Internal 110/230V, 50/60 Hz charger Power Management User selectable power save mode Low Battery Indication is displayed when battery voltage is low Memory Stores up to 99 test results Data in memory can be reviewed on the instrument display, on a PC or via direct printout MECHANICAL Dimensions 10.70 x 9.76 x 7.17" (270 x 250 x 180mm) Weight 10 lbs. (4.5kg) Index of Protection IP53 (case open) IP54 (case closed) DISPLAY Display Type 4000-count LCD Size 4 x 2.25" (102 x 57mm) Backlight Blue electroluminescent Color Safety yellow case; gray faceplate UL94 V-O COMMUNICATION Interface Port Optically isolated USB ENVIRONMENTAL Operating Temperature +14 to +131 F (-10 to 55 C); 10 to 85% RH Storage Temperature -40 to +158 F (-40 to +70 C); 10 to 90% RH Safety Safety Rating EN61010-1 300V Cat. II; 50V Cat. III Double Insulation Yes CE Mark Yes Catalog No. 2129.80 ACCESSORIES Kelvin probes, set of two: (1A - spring loaded) 10 ft, Cat. #2118.73 (1A- spring loaded) 20 ft, Cat. #2118.52 Kelvin clips, set of two: (1A - spring loaded) 10 ft Cat. #1017.83 Kelvin clips, set of two: (10A - Hippo) 10 ft, Cat. #1017.84 (10A - Hippo) 20 ft, Cat. #2118.70 DataView is included with the Model 6240. ORDERING INFORMATION catalog NO. 10A Micro-Ohmmeter Model 6240...Cat. #2129.80 Includes set of two 10ft Kelvin clips (10A - Hippo), NiMH rechargeable battery pack, optical USB cable, US 115V power cord, one pad of measurement results forms, DataView software, extra large tool bag, two spare fuses (12.5A), product warranty and registration card and a user manual. Accessories (Optional) Kelvin Clips (1A), set of two, 10 ft...cat. #1017.83 Kelvin Clips (10A - Hippo), set of two, 10 ft...cat. #1017.84 Kelvin Probes (1A - spring loaded), set of two, 20 ft...cat. #2118.52 Kelvin Clips (10A - Hippo) set of two, 20 ft...cat. #2118.70 Kelvin Probes (1A - spring loaded), set of two, 10 ft...cat. #2118.73

CONSTRUCTION AC line recharging receptacle USB Communication port Rotary selection switch Kelvin terminals Large multi-line backlit liquid crystal display Start/Stop button * Five function buttons * Function buttons allow direct access to data storage and retrieval, polarity reversal, auto/manual measurement and displayed values. FUNCTIONAL DISPLAY Power save Battery charge Caution Polarity Overheating Reading hold Average when polarity reversal is in use Active comport Memory s Memory use s Measurement units Primary - Resistance measurement Measurement units Memory location Secondary - test current, voltage, stored measurements

SOFTWARE Features Configure from your PC Customize views, templates and reports to your exact needs Data Analysis and Reporting Software Create and store a library of configurations that can be used with the Model 6240 as needed Display test results Print reports using standard included templates or custom templates you design All set up functioning available from one screen. Statistical presentation of measurements. Clean, easy-to-read report of test results. Easily documents statistics about user and test site to be stored with the data.

www.aemc.com For all your power quality needs... POWER QUALITY Technical Hotline: (800) 343-1391 www.aemc.com For full product specifications, please visit our website at www.aemc.com sales@aemc.com (800) 343-1391 (508) 698-2115 Fax (508) 698-2118 Chauvin Arnoux One Source 950.BR-6240_0108 Printed in the USA Chauvin Arnoux, Inc. d.b.a. AEMC Instruments 200 Foxborough Blvd. Foxborough, MA 02035