MEDICAL DEVICE SEMINAR ISO 13485:2016 & MDR. May 2016

Similar documents
ISO 13485:201x What is in the new standard?

Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)

ISO 9001:2000 Gap Analysis Checklist

DNV GL Assessment Checklist ISO 9001:2015

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

Internal Quality Management System Audit Checklist (ISO9001:2015) Q# ISO 9001:2015 Clause Audit Question Audit Evidence 4 Context of the Organization

Revision Date Author Description of change Jun13 Mark Benton Removed Admin. Manager from approval

ISO 9001:2000 AUDIT CHECKLIST

ISO 9001:2008 Audit Checklist

QUALITY MANUAL ISO 9001:2015

RECALLS in EUROPE. Past, present, near & further future. Gert Bos BSI Medcon May 2012

ISO 9001 (2000) QUALITY MANAGEMENT SYSTEM ASSESSMENT REPORT SUPPLIER/ SUBCONTRACTOR

Row Manufacturing Inc. Quality Manual ISO 9001:2008

ISO 9001:2008 Quality Management System Requirements (Third Revision)

ISO/TS 16949:2002 Guidance Document

Want to know more about the Notified Body?

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

ISO-9001:2000 Quality Management Systems

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

ISO 9001 Quality Systems Manual

Changes to Medical Device Regulations

(OJ L 169, , p. 1)

How To Know If A Mobile App Is A Medical Device

QUALITY MANAGEMENT SYSTEM REQUIREMENTS General Requirements. Documentation Requirements. General. Quality Manual. Control of Documents

Preparing for Unannounced Inspections from Notified Bodies

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

The Prophotonix (UK) Ltd Quality manual

FINAL DOCUMENT. Guidelines for Regulatory Auditing of Quality Management Systems of Medical Device Manufacturers Part 1: General Requirements

ISO 9001:2015 Internal Audit Checklist

ISO 9001 : 2008 QUALITY MANAGEMENT SYSTEM AUDIT CHECK LIST INTRODUCTION

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE AMENDMENT RECORD SCOPE EXCLUSIONS 2.1 3

ISO 9001: 2008 Boosting quality to differentiate yourself from the competition. xxxx November 2008

GUIDELINES ON MEDICAL DEVICES

PROPOSED DOCUMENT. Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change

Network Certification Body

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

BSI Road Show: September 8 th to 15 th, 2014

Clinical evaluation Latest development in expectations EU and USA

This is Document Schedule 5 Part 1 referred to in this Contract SCOTTISH MINISTERS REQUIREMENTS SCHEDULE 5 PART 1 QUALITY MANAGEMENT SYSTEM

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

Quality Management System Manual

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

Quality Systems Manual

Table of Contents INTEGRATED MANAGEMENT SYSTEM MANUAL

ISO 9001:2015 Overview of the Revised International Standard

Title: Rio Tinto management system

Medical Device Training Program 2015


Quality Management System Manual

QUALITY MANUAL ISO Quality Management System

Quality Manual. UK Wide Security Solutions Ltd. 1 QM-001 Quality Manual Issue 1. January 1, 2011

Frequently Asked Questions. Unannounced audits for manufacturers of CE-marked medical devices. 720 DM a Rev /10/02

Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device

TG TRANSITIONAL GUIDELINES FOR ISO/IEC :2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES

14620 Henry Road Houston, Texas PH: FX: WEB: QUALITY MANUAL

Reporting Changes to your Notified Body

FSSC Certification scheme for food safety systems in compliance with ISO 22000: 2005 and technical specifications for sector PRPs PART II

CENTRIS CONSULTING. Quality Control Manual

Medical Device Software Do You Understand How Software is Regulated?

NOTICE. Re: GD210: ISO 13485:2003 Quality Management System Audits Performed by Health Canada Recognized Registrars

Micro Plastics, Inc. Quality Manual

ALL Medical Devices regardless of Classification

What is ATEX? The European Regulatory Framework for Manufacture, Installation and Use of Equipment in Explosive Atmospheres

Karas Engineering AS9100 QUALITY MANAGEMENT SYSTEM MANUAL

John Tighe ISO 9001 Lead Auditor & Scheme Champion. Alongside CD1, the ballot for 3 specific questions agreed in Sept 2013.

Your essential guide to CE Marking

BSI Unannounced Audits

Implementation of a Quality Management System for Aeronautical Information Services -1-

Introduction to AS 9100

Vigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September Copyright 2015 BSI. All rights reserved.

AUDITOR GUIDELINES. Responsibilities Supporting Inputs. Receive AAA, Sign and return to IMS with audit report. Document Review required?

Vigilant Security Services UK Ltd Quality Manual

Guidance on the preparation of an Annex XV dossier for the identification of substances of very high concern

ISO/IEC QUALITY MANUAL

12040/1/15 REV 1 LES/ns 1 DG B 3B

QMS for Software as a Medical Device [SaMD] Lessons Learned from a Quality Perspective

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994

OUTSOURCING STRATEGIES & RISKS FROM AN ISO PERSPECTIVE. Amera-Veritas. Amera-Veritas. Experts in in Quality, Safety & Supply Chain Assurance

QUALITY MANUAL 3 KENDRICK ROAD WAREHAM, MA FAX

How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC

DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers.

AVNET Logistics & EM Americas. Quality Manual

QUALITY MANAGEMENT SYSTEM Corporate

PUBLICATIONS. Introduction

GOOD DISTRIBUTION PRACTICE FOR MEDICAL DEVICES (GDPMD)

Quality Management System Manual ISO9001:2008

Quality & Safety Manual

Contents. Management Policy Manual SEM USA Page 2 of 12

Konformitätsbewertung 3.9 B 17. Guidance for Notified Bodies auditing suppliers to medical device manufacturers

Certification Procedure of RSPO Supply Chain Audit

Quality Manual Printed copy valid for 24 hours from time of printing unless stamped CONTROLLED COPY in red. Page

NORTH AMERICA OPERATIONS. (Fairmont and Montreal Facilities) QUALITY MANUAL. Prepared to comply with the requirements of ISO 9001:2008

TÜV UK Ltd Guidance & Self Evaluation Checklist

Transcription:

MEDICAL DEVICE SEMINAR ISO 13485:2016 & MDR May 2016

Presenters Fergal O Byrne Head, Business Excellence, NSAI Susan Murphy European Medical Device Operations Manager Mary Murphy Lead Auditor, Medical Device Department Aishling Owens Lead Auditor, Medical Device Department Lisa Mee Lead Auditor, Medical Device Department

Certification Overview

Welcome to NSAI - Aims of the morning A time of change for many Certification Management Standards Understanding changes & their impacts Management Standards general ISO 13485:2016 Medical Devices -QMS @NSAI_Standards

Why change ISO Standards? ISO Review and Revise to support Continual Improvement and Best Practice the HLS ISO has supported the mantra of Integration for many years the HLS. See ISO guide NSAI audits, reports and audit teams delivered an integrated service the HLS in practice Standards needed to catch up 32 HLS MS by now

Management Systems Certification Main Certification schemes a time of great change All HLS ISO 9001:2008 to 2015, Quality ISO 14001:2004 to 2015, Environment OHSAS 18001:2007 to ISO 45001:2016/17 Occupational Health & Safety Three year transition period period of co-existence Common Language Common Structure - HLS Ease of integration Reduced: work, costs, less disruption, better understanding Aligned with the business - a Business Standard

Order out of Chaos

High level structure 1 Scope 2 Normative references 3 Terms and definitions 4 Context of the organization 5 Leadership 6 Planning 7 Support 8 Operation 9 Performance evaluation 10 Improvement

High level structure (1) Context (4) Leadership (5) Improvement (10) Planning (6) Performance Evaluation (9) Support (7) Operations (8)

ISO 13485:2016 odd one out ISO 13485:2016 revised but not in line with the HLS ISO TC 210 ducked the HLS for 3 and possibly 5 years ISO 9001:2015 and ISO 13485:2016 no longer in line Need to map the gaps guidance to follow

NSAI Dedicated NSAI team Jackie Mateer Medical Devices E: jackie.mateer@nsai.ie T:01 8073861 Gwen Thornberry Medical Devices E: gwen.thornberry@nsai.ie T:01 8073929 Jennifer Casey General Management Systems E: jennifer.casey@nsai.ie T: 01 8073817 Patricia Whelan General Management Systems E: patricia.whelan@nsai.ie T: 01 8073993 Events: A Clients perspective ISO 9001:2016 & ISO 14001:2015 Autumn 2016 QI Conference November 2016 ISO 45001:2016/17 Understanding the Changes

ISO 13485:2016

Are you ready for ISO 13485:2016? New Technologies New Regulations New Era

What s New in ISO 13485:2016? ISO 13485:2016 did NOT follow ISO 9001:2015 into the Higher Level Structure format ISO 9001:2015 now has 7* QS core Processes ISO 13485:2016 retains 5* QS core Processes Annex B of ISO 13485:2016 provides a handy cross-reference between ISO 9001:2015 and ISO 13485:2016 normative requirements. *(excluding: Scope, Normative References, Terms & Definitions)

What s New In 2016? Introduction - General The standard can be used by an organization involved in one or more stages of the life-cycle of a medical device And, by suppliers and other external parties providing product to such organizations

What s New In 2016? Section 1 Scope 1.2 - Any requirement of Clause 6, 7, or 8 can be deemed not applicable to the organization (with documented justification)

What s New In 2016? Section 1.2 Scope Of Particular Note in 1: The processes required by this International Standard that are applicable to the organization, but are not performed by the organization, are the responsibility of the organization and are accounted for in the organization s quality management system by monitoring, maintaining, and controlling the processes.

What s New In 2016? Section 4 - Quality Management System 4.1 General (software): New Technologies 4.1.6 Requires documented procedures for the validation of software used in the Quality Management System 4.1.6 Such software shall be validated prior to initial use, and as appropriate, after changes

What s New In 2016? Section 4 - Quality Management System *4.2.3 - Medical Device File: For each medical device type or medical device family, the organization shall establish and maintain a file(s) either containing or referencing documents generated to demonstrate conformity with the requirement of this International Standard and compliance with applicable regulatory requirements. Regular references to the regulatory requirements (*New sub-clause)

What s New In 2016? Section 6 - Resource Management 6.2 Human Resources/ Competence The organization shall document the process(s) for establishing competence, providing needed training, and ensuring awareness of personnel 6.3 Infrastructure: The organization shall document the requirements for the infrastructure needed to achieve conformity to product requirements, prevent product mix-up and ensure orderly handling of product

What s New In 2016? Section 7.3 - Design and Development 7.3.2 Design and Development Planning The organization shall plan and control the design and development of product During design and development planning, the organization shall document: the methods to ensure traceability of design and development outputs to design and development inputs; and the resources needed including necessary competence of personnel

What s New In 2016? Section 7.3 - Design and Development 7.3.6 Design and Development Verification The organization shall document verification plans If the intended use requires that the medical device be connected to, or have an interface with, other medical device(s), verification shall include confirmation that the design outputs meet design inputs when so connected or interfaced Verification shall be completed prior to release for use of the product to the customer

What s New In 2016? Section 7.3 - Design and Development 7.3.7 Design and Development Validation The organization shall document validation plans Design validation shall be conducted on representative product Validation shall be completed prior to release for use of the product to the customer

What s New In 2016? Section 7.3: Design and Development *7.3.8 Design and Development Transfer The organization shall document procedures for transfer of design and development outputs to manufacturing. Results and conclusions of the transfer shall be recorded (see 4.2.5) (*New sub-clause)

What s New In 2016? Section 7.3 - Design and Development 7.3.9 Control of Design and Development Changes The organization shall document procedures to control design and development changes. The organization shall determine the significance of the change to function, performance, usability, safety and applicable regulatory requirements for the medical device and its intended use

What s New In 2016? Section 7.3 - Design and Development 7.3.9 Control of Design and Development Changes (cont d) The review of design and development changes shall include evaluation of the effect of the changes on constituent parts and product in process or already delivered, inputs or outputs of risk management and product realization processes

What s New In 2016? Section 7.3: Design and Development *7.3.10 Design and Development Files The organization shall maintain a design and development file for each medical device type or medical device family (*New sub-clause)

What s New In 2016? Section 7.4 - Purchasing 7.4.1 Purchasing Process The organization shall establish criteria for the evaluation and selection of suppliers The organization shall plan the monitoring and reevaluation of suppliers Non fulfillment of purchasing requirements shall be addressed with the supplier proportionate to the risk associated with the purchased product and compliance with applicable regulatory requirements

What s New In 2016? Section 7.4 Purchasing 7.4.2 Purchasing Information Purchasing information shall include, as applicable, a written agreement that the supplier notify the organization of changes in the purchased product prior to implementation of any changes that affect the ability of the purchased product to meet specified purchase requirements

What s New In 2016? Section 7.4 Purchasing 7.4.3 Verification of Purchased Product The extent of verification activities shall be based on the supplier evaluation results and proportionate to the risks associated with the purchased product When the organization becomes aware of any changes to the purchased product, the organization shall determine whether these changes affect the product realization process or the medical device

What s New In 2016? Section 7.5 - Production & Service Provision 7.5.4 Servicing Activities The organization shall analyze records of servicing activities carried out by the organization or its supplier: a. to determine if the information is to be handled as a complaint b. as appropriate, for input to the improvement process

What s New In 2016? Section 7.5 - Production & Service Provision 7.5.6 Validation of Processes The specific approach and activities associated with software validation and revalidation shall be proportionate to the risk associated with the use of the software including the effect on the ability of the product to conform to specifications

What s New In 2016? Section 7.5 - Production & Service Provision 7.5.8 Identification The organization shall identify product status with respect to monitoring and measurement requirements throughout product realization If required by applicable regulatory requirements, the organization shall document a system to assign unique device identification to the medical device

What s New In 2016? Section 7.5 - Production & Service Provision 7.5.11 Preservation of Product The organization shall protect product from alteration, contamination or damage when exposed to expected conditions and hazards during processing, storage, handling, and distribution

What s New In 2016? Section 8 - Measurement, Analysis & Improvement 8.2.1 Feedback The organization shall document procedures for the feedback process. This feedback process shall include provisions to gather data from production as well as post-production activities The information gathered in the feedback process shall serve as potential input into risk management

What s New In 2016? Section 8 -Measurement, Analysis & Improvement *8.2.2 Complaint Handling The organization shall document procedures for timely complaint handling in accordance with applicable regulatory requirements (*New sub-clause)

What s New In 2016? Section 8 - Measurement, Analysis & Improvement *8.2.3 Reporting to Regulatory Authorities If applicable regulatory requirements require notification of complaints that meet specified reporting criteria of adverse events or issuance of advisory notices, the organization shall document procedures for providing notification to the appropriate regulatory authorities (*New sub-clause)

What s New In 2016? Section 8 - Measurement, Analysis & Improvement 8.3 Control of Nonconforming Product Now contains 4 separate sub-clauses: 1) 8.3.1 General 2) 8.3.2 Actions in response to nonconforming product detected before delivery 3) 8.3.3 - Actions in response to nonconforming product detected after delivery 4) 8.3.4 Rework (NOTE: same requirements as 2003 version)

What s New In 2016? Section 8 - Measurement, Analysis & Improvement 8.5.2 Corrective Action The organization shall take action to eliminate the cause of nonconformities in order to prevent recurrence. Any necessary corrective actions shall be taken without undue delay The organization shall document a procedure to define requirements for: Verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device

Summary A. Regulatory Requirements B. Competence C. Design controls D. Purchasing controls E. Software validation

Comparison between ISO 13485:2016 & ISO 9001:2015 ISO 9001:2015 ISO 13485:2016 Structure Follow the HLS with 7 processes Retains 5 processes Scope Meet customer requirements Meet customer requirements & regulatory requirements Focus Personnel Continual Improvement to enhance customer satisfaction Continuing suitability, adequacy and effectiveness of QMS & safety and performance of the Medical Device Retains need for Management Rep See Annex B to see the mapping of 9001 to 13485 (and visa versa) Use this as a basis for your company gap analysis. www.nsaiinc.com

Comparison between ISO 13485:2016 & ISO 9001:2015 Clauses not directly mapped in ISO 9001 are: Cleanliness of product Installation Servicing Particular requirements for sterile medical devices Particular requirements for validation of processes for sterilisation & sterile barrier systems *New to ISO 13485:2016 Medical Device File

ISO 13485:2016 Transition Policy

Current Situation ISO 13485:2003 EN ISO 13485:2012

Regulatory requirements ISO 13485:2003 CMDCAS Australia Japan MDSAP EN ISO 13485:2012 EU

Co-existence period ISO 13485:2016, replaces ISO 13485:2003 and incorporates the Technical Corrigendum ISO13485:2003/Cor. 1:2009

Co-existence period ISO 13485:2003 & ISO 13485:2016 will co-exist for a 3 year period Effective March 1 st 2018, ISO 13485:2003 certificates will no longer be issued by NSAI

Co-existence period Effective March 1 st 2019 all ISO 13485:2003 certificates will become invalid Effective until March 1 st 2018 any new, modified or revised ISO 13485:2003 issued by NSAI will have an expiry date not to exceed 1 st March 2019

Manufacturer s responsibility Carry out a gap analysis between ISO 13485:2003 &ISO 13485:2016 Update QMS to the requirements of ISO 13485:2016 Request registration to ISO 13485:2016 from your registrar

Registrar s responsibility Develop a checklist to help the manufacturers meet compliance to ISO 13485:2016 Carry out upgrade audits either at the next surveillance or re-assessment audit Two years following publication, new registrations will be issued only to ISO 13485:2016 Three years following implementation of ISO 13485:2016, ISO 13485:2003 will be withdrawn

EN ISO 13485:2016 Refer to the Official Journal (OJ) for updates In the meantime EN ISO 13485:2012 remains valid

Changes to the Regulatory framework ARE YOU READY?

Current situation Both regulations are in the final stages of the legislative procedure and are estimated to finish sometime in 2016, allowing them to come into effect by the end of 2016 or early 2017 Some time will be needed to polish the agreed text and have it translated into the official EU Languages

Why Change from Directives to Regulations? High profile vigilance cases with hips Pelvic floor meshes exposé by British Newspapers Discovery of fraud in PIP breast implants using low quality industrial grade silicone oil

Why Change from Directives to Regulations? Outcome More focus on Notified Bodies through the; Commission recommendation 2013-473 Annex I product assessment Annex II on QMS assessment Annex III Unannounced audits Implementing regulation 920/2013 on the designation and monitoring of Notified Bodies

New Monitoring mechanism of Notified Bodies All Notified Bodies are to be re-designated per the requirements of 920/2013 The audit team comprise representatives of Food and Veterinary Office (FVO) NB s Competent Authority 2 other CA s The designation is valid for 5 years unless the NB wants to extend their scope All results available on the NANDO website: ec.europa.eu/growth/tools-databases/nando

What the FVO found Summary of audit findings Insufficient evidence that staff employed for conformity assessment activities were appropriately qualified and experienced for the task Thoroughness of NB s review of manufacturer s clinical evaluations The sampling and depth of review of technical files for Class IIA & Class IIB Devices The overall documentation and traceability of the certification process was not always deemed appropriate

Output from FVO Some Notified Bodies have been de-designated Some Notified Bodies have elected to stop certification activities due to constraints of the new system Some Notified bodies have scope reductions / limitations to their operation Exclusion of medical devices in class III Subset of devices within a NANDO designation Medical devices including the Machinery Directive to include active medical devices only

What lies ahead?

Chapter 1 Definitions Significant changes ahead Products currently not classified as medical devices under the MDD will be included in the MDR Accessories will now include devices that specifically or directly assist another device in its intended purpose Custom made excludes mass produced by industrial means Inclusion of products containing non-viable human cell or tissue derivatives Importantly non medical devices with a risk profile similar to medical devices e.g. aesthetic implants, cosmetic laser products and contact lenses (Annex XV )

Common Specifications (CS) Where No Harmonised Standard exists Current standards are not sufficient Pertain to safety, performance and risk with a prominent focus on clinical data Applicable in Annex 1, safety and performance Annex II technical documentation Annex XIII, Clinical evaluation and PMCF Annex XIV Clinical evaluation Annex XV Design requirements

Next steps Aesthetic products per Annex XV Standalone software? What do I need to do? Accessories Custom made?

Annex XV, Products without an intended medical purpose Review CS CE mark devices per Annex XV Identify Annex XV devices Develop technical docs Update QMS

Chapter II Making available of devices Obligations of economic operators Reprocessing CE Marking Free movement

Changes Responsible person for regulatory compliance QMS Requirements Changes Liability at all levels, including authorised reps Each person in the supply chain will have specific regulatory responsibility

Under Discussion Product liability insurance for both manufacturers and Authorised Reps Under discussion Reprocessing of single use devices (art 15)

Next steps Reprocessing Recall criteria? Items to review OBL Access to Technical files New QMS requirements?

Next steps Product liability within the chain Person responsible for regulatory compliance Items to review AR contracts Reporting obligations of importers and distributers

Next steps Implant cards, (information to be provided to the patient with an implanted device) Items to review Parts manufacturers vs adverse events Update DOC, MDR model available

Chapter III Eudamed Traceability Registration Publication of device performance UDI

Changes Summary of safety and clinical performance Class 3 & implants Summary to include Intended purpose Device indications & contra-indications Reference to standards and CS Summary of clinical evaluations Profile and training of users

Next steps Manufacturer, authorised rep, economic operator Prepare for implementation of EU UDI Traceability /UDI Apply to the device and higher levels of packaging for all devices Single UDI for manufacturers, AR & importers

Chapter V Classification Up classification Nanotechnology Spinal disk replacements Life saving AMD Substance based devices

Changes Conformity assessment New Essential safety and performance requirements Mandatory tech file structure and content Clinical evaluation for all devices (not new) Scrutiny procedure for the most complex high risk medical devices where no CS have been developed

Next steps Gap analysis of all approved devices on the market Classification Review and implement new classification requirements Adapt new conformity assessment where required

Chapter VI New concepts relating to Article 42 relates to following Annex VIII (full QA ) for all implantables, (with the exception of sutures) Clinical evaluation and Clinical investigation processes are better defined Clinical investigations will be required for implantable and Class 3 devices with few exemptions Mandatory PMCF & PSUR (periodic safety update reports)

Chapter VI High Risk devices and permanent implants Summary of clinical evaluation will be publically available Provided in layman s terms Use of clinical data form equivalent devices must be robust, clinically evidenced and scientifically justified

Next steps Gap analysis on current data Mandatory PMCF Clinical evaluations & investigations Update clinical evaluations in line with new requirements New requirements for claiming equivalence

Next steps Generate clinical evidence Clinical evaluation and PMCF Implement new requirements

Next steps Perform gap analysis Clinical investigations (derogation) Implement for new and existing devices

Chapter VII PMS To become a continuous evaluation and improvement loop Linking to continuous reviews of risk management & Annual update of public summary of safety and performance Clinical evaluation Vigilance New reporting requirements Implement trend reporting

Next steps Implement new plan for ongoing PMCF & PMS requirements Implement trend reporting PMS/Vigilance Gather PMS data as early as possible Implement new Vigilance reporting criteria

Chapter VIII Centralised Governance Development of Common Specifications Developed by expert groups Focus heavily on clinical data sets and minimum patient outcomes for specific device categories

Chapter IX Penalties regime Implementation of market funded market surveillance Companies need to prepare for this fund

Chapter X Final provisions Transition period not yet fixed What happens in the interim? It is possible to comply with the new rules during the transition period Need to find a Notified Body that is designated under the regulation

Final Provisions Investment required People and time Grandfathering is not an option All device must transition Appropriate supporting evidence will be required Additional clinical evidence, updated technical files Declarations of Conformity must be updated