How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters

Size: px
Start display at page:

Download "How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters"

Transcription

1 environmental failure analysis & prevention health technology development How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters Kevin L. Ong, Ph.D., P.E. Managing Engineer, Medical Device Consultant Exponent Inc., Philadelphia, PA A leading engineering & scientific consulting firm dedicated to helping our clients solve their technical problems.

2 2 There s never time to do it right, but there s always time to do it over

3 3 Overview Where do design controls fit in? Elements of design controls Design controls and the FDA ISO and Medical Devices Directive Managing risk against product liability claims

4 4 Medical Device Development Technology and Business Forecasting and Planning Design Control Safety/Efficacy Research, Development Production, Design Cycle, Performance Testing

5 5 Areas of Risk Risk to safety of patients, users, handlers Business Regulatory Product liability

6 6 Quality System Regulation QSR, 21 CFR Part (eff. 1997) replaced the 1978 Good Manufacturing Practices (GMP) for medical devices Ensure that good QA practices are used for medical device design and that they are consistent with quality system requirements worldwide Applies to finished medical devices intended to be commercially distributed for human use: some exemptions for Class I devices Harmonized with efforts by Global Harmonization Task Force

7 21 CFR Part Quality System Regulation Subpart A General Provisions Scope Definitions Quality System Subpart B Quality System Requirements Management Responsibility Quality audit Personnel Subpart C Design controls Design controls Subpart D Document controls Document controls Subpart E Purchasing controls Purchasing controls Subpart F Identification and traceability Identification Traceability Subpart G Production and process controls Production and process controls Inspection, measuring and test equipment Process validation Subpart H Acceptance activities Receiving, in-process and finished device accept Acceptance status Subpart I Nonconforming product Nonconforming product Subpart J Corrective and preventative action Corrective and preventative action Subpart K Labeling and packaging control Device labeling Device packaging Subpart L Handling, storage, distribution and installation Handling Storage Distribution Installation Subpart M Records General requirements Device master record Device history record Quality system record Complaint files Subpart N Servicing Servicing Subpart O Statistical techniques Statistical techniques 7

8 8 Significance Of Design Controls FDA found that approximately 44% of the quality problems that led to voluntary recall actions during this 6-year period were attributed to errors or deficiencies that were designed into particular devices and may have been prevented by adequate design controls. Federal Register, October 7, CFR Parts 808, 812, and 820 Medical Devices; Current Good Manufacturing Practice (CGMP); Final Rule

9 Warning Letters FDA issued 89 Warning Letters to medical device companies for QS/GMP deficiencies QS subsystem Corrective and preventive actions Production & process controls Design controls Management controls Document controls Source: CDRH # warning letters with cite % warning letters with cite

10 10 Most Frequent Quality Systems Warning Letter Cites CFR (a) CFR (a) CFR (a) CFR CFR CFR (c) CFR (i) CFR (a) CFR (g) CFR (general) 17 Source: CDRH Design changes Design validation

11 11 Design Control Warning Letter Cites CFR (i) - Design Changes CFR (g) - Design Validation CFR (a) - General CFR (j) - Design History File 6 21 CFR (c) - Design Input 4 21 CFR (f) - Design Verification 4 21 CFR (e) - Design Review 3 21 CFR (d) - Design Output 3 21 CFR CFR (b) - Design & Development Planning 1

12 12 FDA Design Control Guidance Assist manufacturers in understanding quality system requirements concerning design control Make systematic assessment of the design an integral part of development Increase the likelihood that the design transferred to production will translate into a device that is appropriate for its intended use Applies to the design of medical devices and associated manufacturing processes; new designs as well as modifications or improvements to existing device designs

13 13 Application of Design Controls User Needs Review Design Input Design Process Validation Intended use is fulfilled Verification Requirements are fulfilled Design Output Medical Device

14 14 Elements of Design Controls CFR Establishment, maintenance, and documentation of: Design and Development Planning Design Input (Requirements) Design Output (Specifications) Design Review Design Verification (Meets specifications) Design Validation (Meets clinical needs) Design Transfer (Moves from design to manufacturing) Design Changes Design History File

15 15 Medical Device Classification Class I - General Controls Minimal potential harm to the user, e.g., elastic bandages, examination gloves Prohibit adulteration and misbranding; allow for bans and restriction of sale and distribution; notification of risk and of repair, replacement or refund; registration of manufacturers and listing of devices Most are exempt from the premarket notification and/or GMP regulation Class II General Controls with Special Controls Those for which general controls alone are insufficient to assure safety and effectiveness, e.g. infusion pumps, surgical drapes Special controls may include special labeling requirements, mandatory performance standards, patient registries, postmarket surveillance Few Class II devices are exempt from the premarket notification Class III - General Controls and Pre-Market Approval Those for which insufficient information exists to assure safety and effectiveness solely through the general or special controls, e.g., implantable pacemaker, artificial disc replacement Pre-Market Approval (PMA) application scientific review Clinical data may be necessary - IDE/HDE

16 16 Design Control Scope (a) Design control requirements apply to: All Class II and Class III devices The following class I devices: Devices automated with computer software Tracheobronchial suction catheters Surgeon's gloves Protective restraints Manual radionuclide applicator system Radionuclide teletherapy source

17 17 Design & Development Planning (b) Product concept is usually outlined in a set of marketing requirements and include the wish list for device features and functionality Is there a market opportunity for the device? Where will the new device be used and who will use it? How will the new device be used? How long will the new device be used? Once the project moves forward, a project plan is drawn up to guide and control the design and development of the device Implementation of design controls is initiated

18 18 Design & Development Planning (b) Establish and maintain plans that: Describe or reference design and development activities Define responsibility for implementation Identify or describe interfaces with different groups or activities Review, update and approve plans as design and development evolves

19 19 Design & Development Planning (b) Elements include: Major tasks Deliverables Responsible individuals/units, including interface with suppliers, contractors Resources Project timeline, milestones Major reviews and decision points Review teams Design documentation Planning method: Flow diagram or spreadsheet Gantt chart

20 20 Design Input (c) Design input means the physical and performance requirements of a device that are used as a basis for device design Establish and maintain procedures to assure that device design requirements: Are appropriate Address the intended use of the device Include in procedures a mechanism for addressing requirements that are: Incomplete Ambiguous Conflicting

21 21 Design Input (c) Starting point for product design Usually fall into three categories: Functional requirements specify what the device does, focusing on operational capabilities and processing of inputs and resultant outputs. Performance requirements specify how much or how well the device must perform, addressing speed, strength, response times, accuracy, limits of operation, etc. Also includes device reliability and safety requirements Interface requirements specify characteristics critical to compatibility with external systems; specifically, those characteristics which are mandated by external systems and outside the control of the developers. An example is important in every case is the user and/or patient interface

22 22 Types of Design Inputs Intended Use User Needs Physical/chemical characteristics Performance characteristics Safety Reliability Environmental limits User interfaces Regulatory requirements Labeling Human factors Maintenance Compatibility with other devices Sterilization Energy source Toxicity and biocompatibility

23 23 Sources of Design Inputs Customer input through focus groups, surveys, etc. Comparison testing of competitor product for specific performance criteria or other benchmarking activities Review of similar product histories to include production data for scrap, rework, testing Review of complaints, MDRs, FDA Enforcement Reports, recall of similar products or earlier generations of the product Performance requirements stipulated by the FDA, voluntary standards, or other regulatory agencies Literature review Input from cross-functional teams (R&D, Quality, Regulatory, Marketing, Manufacturing, etc.)

24 24 Design Output (d) Design output means the results of a design effort at each design phase and at the end of the total design effort. The finished design output is the basis for the device master record. The total finished design output consists of the device, its packaging and labeling, and the device master record Establish and maintain procedures for Defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements Containing or referencing acceptance criteria Ensuring design outputs essential for the proper functioning of the device are identified

25 25 Design Output (d) Design output are the design specifications which should meet design input requirements, as confirmed during design verification and validation and ensured during design review The output includes the device, its labeling and packaging, associated specifications and drawings, and production and quality assurance specifications and procedures

26 26 Types of Design Output Production specifications and other descriptive materials that define and characterize the design Production specifications used to procure components, fabricate, inspect, install, maintain and service the device Other materials that are necessary for establishing conformance, but are not used in device production: Biocompatibility test results Bio-burden test results

27 27 Design Review (e) Documented, comprehensive, systematic examination of a design to evaluate adequacy of requirements, capability of design to meet requirements, and to identify problems Establish and maintain procedures for design reviews Plan and conduct formal documented design reviews of the design results at appropriate stages Include at each design review Representatives of all functions concerned Independent reviewer for the stage being reviewed Any specialists needed

28 28 Considerations for Design Review Formal process to evaluate design, resolve concerns, and implementation of corrective actions Required for phase approval Formal design review prior to release for manufacture and distribution Responsibility and authority for decision-making Determine frequency of review and design phases for review Controlling documentation and record keeping Competency of reviewers in terms of background and experience

29 29 Design Verification (f) vs. Design Validation (g) Design Verification Did I make the product right? i.e. specified requirements have been fulfilled Design Validation Did I make the right product? i.e. device specifications conform with user needs and intended use(s)

30 30 Design Verification (f) Establish and maintain procedures for verifying the device design Confirm design output meets design input requirements Types of verification activities include: Inspections Tests Analyses

31 31 Examples of Design Verification Activities Evaluation of physical characteristics Mechanical/chemical/electrical testing Failure modes and effects analysis Package integrity testing Biocompatibility testing Sterility testing Comparison against predicate device with successful history

32 32 Design Validation (g) Establish and maintain procedures for validating the device design Perform design validation Conform to defined user needs and intended uses Under actual or simulated use conditions Under defined operating conditions On initial production units, lots, or batches or their equivalents Includes software validation and risk analysis, where appropriate

33 33 Examples of Design Validation Activities Clinical studies Consumer preference testing Bench testing under simulated use conditions Review of labeling and packaging Environmental testing (e.g. stress during shipping and installation) Testing that distinguishes between customers, users, and patients

34 34 Design Transfer (h) Establish and maintain procedures to ensure that the device design is correctly translated into production specifications Design moves from R&D to manufacturing Specifications ensure that manufactured devices are repeatedly and reliably produced within product and process capabilities

35 35 Types of Production Specifications Assembly drawings Component procurement specifications Workmanship standards Manufacturing instructions Inspection and test specifications Computer-aided manufacturing programming files Manufacturing jigs/fixtures

36 36 Design Transfer: Elements Product risk documentation shows what failure modes, components, parts, etc. should be monitored and trended Process validations indicate critical control points for manufacturing and design

37 37 Design Changes (i) Establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation Establish design change control procedures, determine the risk presented by the change and implement design control requirements to the extent appropriate for that risk

38 38 Considerations for Document Control Enumeration of design documents, and tracking their status and revision history Maintain master list or document index to define product and/or process Maintain history of document revisions Periodically inventory controlled documents to confirm that contents are up to date Develop procedure for removal and deletion of obsolete documents

39 39 Considerations for Change Control Document decisions for change requests identifies possible design matter which the requester believes should be corrected Document change order following acceptance of change request identifies relevant document(s) and revision of document content May also include assignment to designers to further evaluate the problem and develop corrective action Review and assess impact of design change on design input requirements and intended uses

40 40 Design Change Review The effect of the changes on products already in commercial or clinical use Whether prior design verification or validation results are impacted Changes are implemented in such a manner that the original problem is resolved and no new problems are created; or if new problems are created, they are also tracked to resolution

41 41 Design History File (j) Establish and maintain a design history file for each type of device Include in the DHF or reference records necessary to demonstrate that the design was developed in accordance with the design plan and CFR 820 requirements Design history file (DHF) means a compilation of records which describes the design history of a finished device

42 42 Design History File Tells the story of a device or device family s development from beginning to end Contains the documentation necessary to assure changes to design or process for manufacture of the device do not adversely impact the device design and lead to improvement in design and process Maintains knowledge base after design decisions have been made, employees leave, company reorganizes, or subcontractors unavailable etc.

43 43 Elements of Design History File Design and development plan Design input documents Risk analyses documents Design output Pre-production design change control records Engineering/laboratory notebooks V&V protocols and results Design review records Design transfer records Copies of controlled documents used during the design process, e.g. SOPs etc. Initial DMR

44 44 Benefits of Design Controls Improve the product development process by producing higher quality devices Reduce liability exposure Reduce design related product recalls Shorten development time Recognize problems earlier and make corrections Lower manufacturing costs Meet customer expectations (user and patient needs)

45 45 Other Quality System Requirements Medical Devices Directive (MDD 93/42/EEC): developed in 1993; lists requirements regarding medical device design within EU Revised in 2007 (2007/47/EC) Compliance became mandatory in March, 2010

46 46 Medical Devices Directive All devices require clinical data Of the actual device or a similar device Scientific literature of a similar device Human factors considered a key factor

47 47 ISO Standards on Quality ISO (Medical devices - Quality management systems - Requirements for regulatory purposes): standard that recommends procedures in order to have a quality system that complies with Medical Devices Directive (MDD 93/42/EEC) Based on ISO 9001, but excludes some of the requirements that are not appropriate for regulatory requirements

48 48 ISO Similar elements to ISO Section 7.3 Design and development Design and development planning Design and development inputs Design and development outputs Design and development review Design and development verification Design and development validation Control of design and development changes

49 49 ISO Provide framework within which experience, insight and judgment are applied systematically to manage the risks Risk management is the systematic application of Management policies Procedures and practices for analyzing, evaluating, and controlling risk

50 50 ISO Specifies process through which the manufacturer can identify hazards associated with a medical device, estimate and evaluate the risks associated with these hazards, control these risks, and monitor the effectiveness of that control

51 51 Quality System Inspection Technique (QSIT) Guidance for auditing medical device manufacturers against QS Regulation (21 CFR Part 820) and related regulations Focuses on four major subsystems that are the basic foundation of the QS to evaluate and states the purpose and importance of each subsystem nspectionguides/ucm pdf

52 52 What are the four main subsystems? Design Controls Material Controls Corrective & Preventive Actions Management Records, Documents, & Change Controls Production & Process Controls Equipment & Facility Controls

53 53 What does FDA look for in the Design Control Subsystem? Design procedures and plan established? Design inputs or requirements for device identified? Design outputs or specifications for device identified? Design verification conducted? Design validation conducted? Software validation completed? Risk analysis carried out? Design reviews conducted? Design transfer to manufacturing completed successfully?

54 54 Warning Letters - Examples do not have design control procedures no documentation of design validation or design change controls no documentation of risk analysis or structural (b)(4) testing of the embedded software

55 55 Warning Letters - Examples unable to demonstrate that the validation for the redesign project was validated using initial production units the verification study for the coating which is applied to the was conducted after the validation study was completed failed to demonstrate bench testing of the [X] had been completed, prior to their installation into [Y] for design validation studies failed to establish acceptance specifications prior to conducting verification and validation studies

56 56 Warning Letters - Examples Design validation was not executed or documented in order to ensure that the devices conform to defined user needs and intended uses A design and development plan that describes or references the design and development activities and define responsibility for implementation was not established or maintained A design history file was not established or maintained by your firm

57 57 Warning Letters - Examples Design Input for [X] contains no inputs for the [Y] which are to be used with the [X] Because you did not include [Y] inputs in your [X] design verification/validation, your [X] Risk Analysis/Failure Modes and Criticality Analysis, document (b)(4) is incomplete in that you failed to identify and evaluate potential hazards involved with use of the [Y]

58 58 Product Liability/Negligence Claims Design Defects Foreseeable risks? Could have been reduced or avoided? Omission of reasonable, safer, and technologically feasible alternative design? Risks outweigh benefits?

59 59 Key Documents to Consider During Litigation DHF Regulatory documents (510(k), memo to file, PMA) DFMEA PFMEA Specifications Lot/device history records Manufacturing procedures Design drawings CAPA reports MDR/complaint files IFU Surgical brochures Marketing brochures/documents Audit reports

60 60 Impact on Litigation Design changes Design controls for line extensions Memo to file justification Questions to consider: After a design change or line extension, are there additional risks for the device? Is additional design verification or validation needed? Are the specifications clinically relevant? How were they developed? Are the specifications consistent? Were the inputs translated from marketing inputs into engineering inputs? Were finished products considered? Was validation completed before verification? Was there compliance with own requirements/procedures?

61 61 Final Thoughts The design control regulations provide a comprehensive and systematic framework for device development The design control process is iterative Design control regulations are a regulatory requirement, but when used properly serve as good business practice

62 Design control does not end with the transfer of a design to production. Design control applies to all changes to the device or manufacturing process design, including those occurring long after a device has been introduced to the market. This includes evolutionary changes such as performance enhancements as well as revolutionary changes such as corrective actions resulting from the analysis of failed product. The changes are part of a continuous, ongoing effort to design and develop a device that meets the needs of the user and/or patient. Thus, the design control process is revisited many times during the life of a product. 62 Design Control Guidance for Medical Device Manufacturers; March 11, 1997; FDA, CDRH

63 63 Reading Materials FDA Guidance Document ( GuidanceDocuments/ucm pdf) Medical Devices Directive ( ISO ( ISO ( Quality System Inspection Techniques ( 8.pdf) AdvaMed Points to consider when preparing for an FDA inspection under the QSIT design controls subsystem (

64 64 Thank you! Questions? Kevin Ong, Ph.D., P.E

Comparison between FDA QSR and ISO 13485

Comparison between FDA QSR and ISO 13485 Comparison between FDA QSR and ISO 13485 Most countries in the world including the Europe, for the conformity assessment of medical devices to be used by their countrymen, assess not only whether the product

More information

IVD Regulation Overview. Requirements to Assure Quality & Effectiveness

IVD Regulation Overview. Requirements to Assure Quality & Effectiveness IVD Regulation Overview Requirements to Assure Quality & Effectiveness CLIAC Jan. 2002 Statutory and Regulatory Requirements Statute: Food, Drug, and Cosmetic Act Food and Drugs Act of 1906 Food and Drug

More information

ORACLE CONSULTING GROUP

ORACLE CONSULTING GROUP ORACLE CONSULTING GROUP 9 Golder Ranch Rd., Ste. 1 Tucson, Arizona 9 Web Site: E-mail: 20-2-0 20-2-0 (FAX) CONSULTING MEMORANDUM QUALITY SYSTEM INSPECTION TECHNIQUE

More information

Quality System: Design Control Procedure - Appendix

Quality System: Design Control Procedure - Appendix Quality System: Design Control Procedure - Appendix Page 1 of 10 Quality System: Design Control Procedure - Appendix CORP Medical Products Various details have been removed, indicated by [ ] 1. Overview

More information

Corrective and Preventive Action Background & Examples Presented by:

Corrective and Preventive Action Background & Examples Presented by: Corrective and Preventive Action Background & Examples Presented by: Kimberly Lewandowski-Walker Food and Drug Administration Division of Domestic Field Investigations Office of Regulatory Affairs Overview

More information

Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC

Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC Overview of Medical Device Controls in the US By Nandini Murthy, MS, RAC 18 controls are a regulatory requirement for medical devices. In the US, compliance with the design controls section of 21 Code

More information

GxP Process Management Software. White Paper: Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations

GxP Process Management Software. White Paper: Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations GxP Process Management Software : Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations Most FDA violations involve one of the following: Not having procedures in a regulated area

More information

2014 Annual FDA Medical Device Quality System Data. FDA Form 483 Observations and Warning Letter Citations

2014 Annual FDA Medical Device Quality System Data. FDA Form 483 Observations and Warning Letter Citations 2014 Annual FDA Medical Device Quality System Data FDA Form 483 Observations and Warning Letter Citations Why is FDA making these data available? In support of the FDA s Transparency and Case for Quality

More information

Combination Products. Presented by: Karen S. Ginsbury For: IFF March 2014. PCI Pharma

Combination Products. Presented by: Karen S. Ginsbury For: IFF March 2014. PCI Pharma Combination Products Presented by: Karen S. Ginsbury For: IFF March 2014 Types of products Biological and medical device (freeze dried + syringe dual volume) Medical device and plasma devised product (syringe)

More information

Medical Product Software Development and FDA Regulations Software Development Practices and FDA Compliance

Medical Product Software Development and FDA Regulations Software Development Practices and FDA Compliance Medical Product Development and FDA Regulations IEEE Orange County Computer Society March 27, 2006 Carl R. Wyrwa Medical Product Development and FDA Regulations Introduction Regulated FDA Overview Medical

More information

MEDICAL DEVICE QUALITY SYSTEMS MANUAL: A SMALL ENTITY COMPLIANCE GUIDE. First Edition

MEDICAL DEVICE QUALITY SYSTEMS MANUAL: A SMALL ENTITY COMPLIANCE GUIDE. First Edition HHS Publication FDA 97-4179 MEDICAL DEVICE QUALITY SYSTEMS MANUAL: A SMALL ENTITY COMPLIANCE GUIDE First Edition (Supersedes the Medical Device Good Manufacturing Practices Manual) Andrew Lowery, Judy

More information

Company Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature

Company Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature Far West Technology, Inc. ISO 9001 Quality Manual Document No.: QM Revision: 0 Issue Date: 27 August 1997 Approval Signatures President/CEO Executive Vice President Vice President/CFO Change Record Rev

More information

ORACLE CONSULTING GROUP

ORACLE CONSULTING GROUP ORACLE CONSULTING GROUP An Official United States Agent Firm for Foreign Establishments CONSULTING MEMORANDUM: DEALING WITH A MEDICAL DEVICE IN THE U.S. 5398 Golder Ranch Rd., Ste. 1 Tucson, Arizona 85739

More information

Empowering the Quality and Regulatory Compliance Functions

Empowering the Quality and Regulatory Compliance Functions Empowering the Quality and Regulatory Compliance Functions Management must take steps to ensure that regulatory and quality compliance is everyone s responsibility. By: J. Glenn George, Kenneth Imler,

More information

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers.

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers. Supplier Quality Standard 1.0 Purpose The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers. These expectations

More information

ISO 13485:201x What is in the new standard?

ISO 13485:201x What is in the new standard? ISO 13485:201x What is in the new standard? Eric Finegan, Quality Mgr, BTE Technologies, Inc. 2015-09-10 1 Presentation Slides This slide deck is the presentation performed on 2015-09-10. A more detailed

More information

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to INTRODUCTION This book offers a systematic, ten-step approach, from the decision to validate to the assessment of the validation outcome, for validating configurable off-the-shelf (COTS) computer software

More information

CORPORATE QUALITY MANUAL

CORPORATE QUALITY MANUAL Corporate Quality Manual Preface The following Corporate Quality Manual is written within the framework of ISO 9001:2008 Quality System by the employees of CyberOptics. CyberOptics recognizes the importance

More information

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE Company Name Street Address City, State, Zip code Phone Number Fax Company Website Email Address ORGANIZATION NAME PHONE NUMBER EMAIL ADDRESS President/CEO General Manager Engineering Manager Production

More information

Specialties Manufacturing. Talladega Castings & Machine Co., Inc. ISO 9001:2008. Quality Manual

Specialties Manufacturing. Talladega Castings & Machine Co., Inc. ISO 9001:2008. Quality Manual Specialties Manufacturing Talladega Castings & Machine Co., Inc. ISO 9001:2008 This document is the property of TMS and may not be reproduced, wholly, or in part, without the express consent of TMS. Rev.

More information

Inspections, Compliance, Enforcement, and Criminal Investigations

Inspections, Compliance, Enforcement, and Criminal Investigations Page 1 of 7 Home Inspections, Compliance, Enforcement, and Criminal Investigations Compliance Actions and Activities Warning Letters 2014 Inspections, Compliance, Enforcement, and Criminal Investigations

More information

Table of Contents 1. SCOPE... 3 2. APPLICABLE DOCUMENTS... 4 3. TERMS AND DEFINITIONS... 4 4. QUALITY MANAGEMENT SYSTEM...4-8

Table of Contents 1. SCOPE... 3 2. APPLICABLE DOCUMENTS... 4 3. TERMS AND DEFINITIONS... 4 4. QUALITY MANAGEMENT SYSTEM...4-8 Table of Contents 1. SCOPE... 3 2. APPLICABLE DOCUMENTS... 4 3. TERMS AND DEFINITIONS... 4 4. QUALITY MANAGEMENT SYSTEM...4-8 5. MANAGEMENT RESPONSIBILITY...8-9 6. RESOURCE MANAGEMENT... 10 7. PRODUCT

More information

QUALITY POLICY MANUAL Document: 01-090000 Revision: E Effective Date: January 15, 2010

QUALITY POLICY MANUAL Document: 01-090000 Revision: E Effective Date: January 15, 2010 Section i INTRODUCTION QUALITY POLICY STATEMENT: The purpose of our business is to provide continuity of electrical power to keep businesses in business. We do this by helping clients solve their power

More information

Surgi Manufacturing Quality Manual

Surgi Manufacturing Quality Manual Surgi Manufacturing Page 1 of 18 Approvals: QA: Eng. Mgt. : A Date: 18Aug98 1. Introduction... 4 1.1 Scope... 4 1.2 Purpose... 4 1.3 Authority... 4 1.4 Issue of the Manual... 4 1.5 Amendments... 4 1.6

More information

Correspondence between ISO 13485:2003 and the US Quality System Regulation

Correspondence between ISO 13485:2003 and the US Quality System Regulation Correspondence between ISO 13485:2003 and the US Quality System Regulation Correspondence between ISO 13485:2003 and the US Quality System Regulation 1 Scope 1.1 General This International Standard specifies

More information

Medical Device Training Program 2015

Medical Device Training Program 2015 Medical Device Training Introduction Supplementary training and education is often overlooked by medical device professionals until it is triggered by an upcoming FDA or Notified Body and/or ISO 13485

More information

CAPA - the importance of data analysis

CAPA - the importance of data analysis CAPA - the importance of data analysis Presented by: Sue Jacobs QMS Consulting, Inc. 1 847 359 4456 sue@qmsconsultant.com QMS Consulting, Inc. 2007 1 Topics Regulatory Requirements Design Controls and

More information

Documents, Records and Change Control

Documents, Records and Change Control Documents, Records and Change Control Lori S. Lawless Medical Device Specialist Food and Drug Administration Baltimore District Office Lori.Lawless@fda.hhs.gov (410) 779-5442 May 13-14, 2008 Baltimore

More information

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 globalsales@maetrics.com

More information

Medical Device Software

Medical Device Software Medical Device Software Bakul Patel Senior Policy Advisor 1 Overview Medical devices and software Oversight principles and Current approach Trends, Challenges and opportunities Addressing challenges 2

More information

Quality Management System Manual

Quality Management System Manual Quality Management System Manual Assurance ISO / AS Manual Quality Management System EXCEEDING ALL EXPECTATIONS Since our inception in 1965, Swiss-Tech has supplied the medical, aerospace, hydraulic, electronic

More information

Quality Management Systems Manual

Quality Management Systems Manual Washington Division Quality Management Systems Manual ISO 13485:2003 DOCUMENT NUMBER 01-QM-MED REVISION LEVEL 01 PAGE 1 of 40 This manual describes the quality management systems structure at GM Nameplate

More information

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE 1.1 1 AMENDMENT RECORD 1.2 2 SCOPE 2.0 3 EXCLUSIONS 2.1 3

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE 1.1 1 AMENDMENT RECORD 1.2 2 SCOPE 2.0 3 EXCLUSIONS 2.1 3 TABLE OF CONTENTS DESCRIPTION SECTION PAGE INTRODUCTION 1.0 1 APPROVAL SIGNATURE PAGE 1.1 1 AMENDMENT RECORD 1.2 2 SCOPE 2.0 3 EXCLUSIONS 2.1 3 CORPORATE POLICY 3.0 3 QUALITY MANAGEMENT SYSTEM 4.0 4 GENERAL

More information

Why do Project Fail? 42% 37% 27% 26% 24% 24% 0% 10% 20% 30% 40% 50% IBM Software Group Rational software. Source: AberdeenGroup, August 2006

Why do Project Fail? 42% 37% 27% 26% 24% 24% 0% 10% 20% 30% 40% 50% IBM Software Group Rational software. Source: AberdeenGroup, August 2006 Why do Project Fail? Unclear or continually changing product definitions Product does not meet customer or market requirements 37% 42% Unrealistic schedule expectations Projects not adequately staffed

More information

GE Healthcare MAR 1 2013

GE Healthcare MAR 1 2013 K130155 Pagel1 of 2 GE Healthcare MAR 1 2013 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: Submitter: Primary Contact Person: Secondary Contact Person:

More information

Product Liability Risk Control Checklist

Product Liability Risk Control Checklist Product Liability Risk Control Checklist Appendix C1 This material will help you evaluate your products liability exposure. It is designed to enable you to focus on areas that need to be addressed to safeguard

More information

MEDFORD FABRICATION CSC, INC. Quality System Manual. Date of issue: 03/25/2010 Revision : F

MEDFORD FABRICATION CSC, INC. Quality System Manual. Date of issue: 03/25/2010 Revision : F MEDFORD FABRICATION CSC, INC Quality System Manual Date of issue: 03/25/2010 Revision : F Table of Contents System Description 1.0 Introduction 2.0 Company Quality Policy 3.0 Organization Charts 4.0 Quality

More information

Q(K SVJM~jPagelIof 3

Q(K SVJM~jPagelIof 3 K13 0460 Q(K SVJM~jPagelIof 3 JUL 1 12013 This 51 0(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k)

More information

Title:: Effective GMP AUDITS for APIs and Formulation Pharma Companies By G.Sundar-Director/Consultant PharmQA Compliance solutions

Title:: Effective GMP AUDITS for APIs and Formulation Pharma Companies By G.Sundar-Director/Consultant PharmQA Compliance solutions WELCOME Title:: Effective GMP AUDITS for APIs and Formulation Pharma Companies By G.Sundar-Director/Consultant PharmQA Compliance solutions Contents: Introduction GMP Audit GMP Audit plan GMP Auditing

More information

Quality Assurance Manual

Quality Assurance Manual Quality Assurance Manual Revised April 30, 2014 2004, 2005, 2006, 2007, 2008, 2009, 2010, 2011, 2012, 2013, 2014 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound

More information

Quality Manual ISO9001:2008

Quality Manual ISO9001:2008 Quality Manual ISO9001:2008 Connor Winfield Corp. Date: 4/26/10 Revision: 3 Quality Manual Quality Manual Connor Winfield GENERAL Section 0.1 Section Rev.: 2 Rev. Date: 4/26/10 Section Page 1 Index and

More information

Mapping Your Success Top Non-Conformities Requiring a CAPA Process

Mapping Your Success Top Non-Conformities Requiring a CAPA Process Mapping Your Success Top Non-Conformities Requiring a CAPA Process Agenda FDA Statistics for 2012 BSI Statistics for 2012 EQMS industry benchmark data About CAPA Traits of Software Solution for CAPA 2

More information

ALL PRODUCTS MFG & SUPPLY

ALL PRODUCTS MFG & SUPPLY ALL PRODUCTS MFG & SUPPLY 618 ANDERSON DRIVE ROMEOVILLE, IL 60446 PHONE: 877-255-8700 FAX: 877-255-8701 WWW. APGASKET.COM QUALITY MANAGEMENT SYSTEM MANUAL DATE: 11/20/12 REVISION 9.1 UNCONTROLLED COPY

More information

PROPOSED DOCUMENT. Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change

PROPOSED DOCUMENT. Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change AHWP/WG3/P001:2013 PROPOSED DOCUMENT Title: Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change Author: AHWP Work Group 3 Date: 13 November

More information

February 5, 2015. Dear Kristin Pabst,

February 5, 2015. Dear Kristin Pabst, DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 5, 2015, Inc.

More information

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Quality Management System MANUAL SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Doc. No. G5500 Rev. 9.1 Status : APPROVED Effective: 12/11/2014 Page 2 of 23 Quality Manual Table of Contents

More information

Setting Up A Complaint Handling System. Presented by Sue Jacobs Principal Consultant, QMS Consulting, Inc. Complaint Systems

Setting Up A Complaint Handling System. Presented by Sue Jacobs Principal Consultant, QMS Consulting, Inc. Complaint Systems Setting Up A Complaint Handling System Presented by Sue Jacobs Principal Consultant, QMS Consulting, Inc. January 2008 ~ Orlando, Florida QMS Consulting, Inc. 2008 1 Complaint Systems Written Procedures

More information

Product Lifecycle Management in the Medical Device Industry. An Oracle White Paper Updated January 2008

Product Lifecycle Management in the Medical Device Industry. An Oracle White Paper Updated January 2008 Product Lifecycle Management in the Medical Device Industry An Oracle White Paper Updated January 2008 Product Lifecycle Management in the Medical Device Industry PLM technology ensures FDA compliance

More information

Systems Engineering for FDA QSR Compliance

Systems Engineering for FDA QSR Compliance Systems Engineering for FDA QSR Compliance James H. Jones Optants Documented Decision Support Company (ODDSCO) 297 Casitas Bulevar, Los Gatos, CA 95032-1119 www.optants.com jjones@optants.com Copyright

More information

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Quality Management System MANUAL SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Doc. No. G5500 Rev. 9.2 Status : APPROVED Effective: 9/25/2015 Page 2 of 23 Quality Manual Table of Contents

More information

Quality Assurance QUALITY ASSURANCE PLAN

Quality Assurance QUALITY ASSURANCE PLAN Revision 2 Page 1 of 40 QUALITY ASSURANCE PLAN PLAN APPROVALS: Jeff Shouse Signature on File DIRECTOR OF QUALITY ASSURANCE DIRECTOR OF QUALITY ASSURANCE (signature) DATE Rodney Baltzer Signature on File

More information

QUALITY MANAGEMENT SYSTEM Corporate

QUALITY MANAGEMENT SYSTEM Corporate Page 1 of 12 4 Quality Management System 4.1 General Requirements The Peerless Pump Quality Management System shall include: Documented statements of a quality policy and of quality objectives; A quality

More information

Software Development for Medical Devices

Software Development for Medical Devices Overcoming the Challenges of Compliance, Quality and Cost An MKS White Paper Introduction Software is fast becoming the differentiator for manufacturers of medical devices. The rewards available from software

More information

Aligning IT investment and Business

Aligning IT investment and Business IBM Software Group Aligning IT investment and Business The role of requirements management, portfolio management and enterprise architecture Productivity, Governance, Innovation Dr Tariq Aslam 2009 IBM

More information

January 12, 2016. Dear Amy Yang:

January 12, 2016. Dear Amy Yang: DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 12, 2016 Amy

More information

FINAL DOCUMENT. Implementation of risk management principles and activities within a Quality Management System. The Global Harmonization Task Force

FINAL DOCUMENT. Implementation of risk management principles and activities within a Quality Management System. The Global Harmonization Task Force GHTF/SG3/N15R8 FINAL DOCUMENT Title: Implementation of risk management principles and activities within a Quality Management System Authoring Group: GHTF Study Group 3 Endorsed by: The Global Harmonization

More information

April 7, 2015. Acclarent, Inc. Mr. James Patrick Garvey II Sr. Manager, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025

April 7, 2015. Acclarent, Inc. Mr. James Patrick Garvey II Sr. Manager, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025 DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 7, 2015 Acclarent,

More information

Harmonizing Change Control Processes Globally

Harmonizing Change Control Processes Globally Quality & Compliance Associates, LLC Harmonizing Change Control Processes Globally President Quality & Compliance Associates, LLC When We Deal In A Global Environment, How Do We Design A System That Addresses

More information

Quality Agreement Template

Quality Agreement Template Quality Agreement Template Prepared by Dan O Leary Ombu Enterprises, LLC 3 Forest Ave. Swanzey, NH 03446 603-209-0600 This document is intended to form the basis for a Supplier Agreement. The document

More information

Corrective and Preventive Actions

Corrective and Preventive Actions Corrective and Preventive Actions A Five Step Approach Topics to Be Covered What is CAPA? Governing authority Five steps to a good CAPA process Where companies have difficulty Example citations Recap What

More information

DEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation

DEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation Digital Internal Use Only DEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation DOCUMENT IDENTIFIER: A-DS-EL00017-02-0000 Rev A, ABSTRACT: This standard

More information

How To Validate Software

How To Validate Software General Principles of Software Validation; Final Guidance for Industry and FDA Staff Document issued on: January 11, 2002 This document supersedes the draft document, "General Principles of Software Validation,

More information

CDRH Regulated Software

CDRH Regulated Software CDRH Regulated Software An Introduction John F. Murray Jr. CDRH Software Compliance Expert CDRH Regulates Software in the following areas Medical Devices Automation of Production Systems Automation of

More information

Making Recall Determinations

Making Recall Determinations Making Recall Determinations AdvaMed January 23, 2008 Orlando, FL Edward C. Wilson, Jr., Esq. Partner Hogan & Hartson LLP (202) 637-5839 ecwilson@hhlaw.com www.hhlaw.com Michael S. Heyl, Esq. Associate

More information

Revision Date Author Description of change. 10 07Jun13 Mark Benton Removed Admin. Manager from approval

Revision Date Author Description of change. 10 07Jun13 Mark Benton Removed Admin. Manager from approval Page 2 of 15 Document Revision History Revision Date Author Description of change 10 07Jun13 Mark Benton Removed Admin. Manager from approval 12Feb13 Mark Benton 08 01Oct12 Mark Benton 07 8/30/2012 Refer

More information

Product Liability. Risk Control

Product Liability. Risk Control Product Liability Risk Control With tort reform failing to stop the steady increase in product liability claim awards, manufacturers and enterprises like yours must take defensive action to protect their

More information

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions Document issued on: March 27, 2006 For questions regarding this document, contact Everette Beers, Ph.D. at 240-276-4200

More information

Case Study: A Mock-QSIT Audit of Ford & Firestone By Tim Wells

Case Study: A Mock-QSIT Audit of Ford & Firestone By Tim Wells Case Study: A Mock-QSIT Audit of Ford & Firestone By Tim Wells Published in ASQ Biomedical Feedback, Volume 30, No. 1, Winter 2001 Background In December the Firestone Company released a report on the

More information

Quality Management System Manual Revision L

Quality Management System Manual Revision L This Page 1 of 35 of the Quality Management System Manual If issued as a controlled copy, the serial number of this copy is Quality Management System Manual Certified to AS9100 Revision C Printed copies

More information

Mossfiel Electrical & Safety Management Pty Ltd

Mossfiel Electrical & Safety Management Pty Ltd ISO 9001:2008 QUALITY MANAGEMENT PLAN Be warned. Any reproduction of this quality management system or associated documents without the express permission of the company and

More information

Soft Computer Consultants, Inc. 4/30/15

Soft Computer Consultants, Inc. 4/30/15 Soft Computer Consultants, Inc. 4/30/15 Department of Health and Human Services Public Health Service Food and Drug Administration Florida District 555 Winderley Place, Suite 200 Maitland, Florida 32751

More information

Simpler Implant Solutions Inc. 2930 Arbutus St. Vancouver, BC V6i 3Y9. haroldbergmanl~ogmail.com 604 736 9890 Telephone 604 736 9747 Fax K080115

Simpler Implant Solutions Inc. 2930 Arbutus St. Vancouver, BC V6i 3Y9. haroldbergmanl~ogmail.com 604 736 9890 Telephone 604 736 9747 Fax K080115 APR 5 2013 Summary: K121281 Submitter: Company Address Contact Person: Simpler Implant Solutions Inc. 2930 Arbutus St. Vancouver, BC V6i 3Y9 Dr. Harold Bergman haroldbergmanl~ogmail.com 604 736 9890 Telephone

More information

March 20, 2015. Dear Mr. Chen:

March 20, 2015. Dear Mr. Chen: DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 March 20, 2015 Biomate

More information

ISO 13485:2003 U.S. QSR (21 CFR 820) Quality Systems Manual

ISO 13485:2003 U.S. QSR (21 CFR 820) Quality Systems Manual Insert Your Company Name Here ISO 13485:2003 U.S. QSR (21 CFR 820) Quality Systems Manual Street Address City, State Zip This manual can be used as a template in developing your ISO 13485 Quality Manual.

More information

ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan)

ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan) ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan) Part 1: Project-Specific Quality Plan Part 2: Company Quality Manual Part 3: Submittal Forms Part 4:

More information

September 17, 2014. Keeler Instruments Inc. Mr. Eugene R. Van Arsdale Marketing Manager 459 Parkway Broomall, PA 19008

September 17, 2014. Keeler Instruments Inc. Mr. Eugene R. Van Arsdale Marketing Manager 459 Parkway Broomall, PA 19008 DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 September 17, 2014 Keeler

More information

Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices

Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices Experts in medical device quality systems, compliance, regulatory affairs, auditing and training www.nsf.org The Right People.

More information

Guidance for Industry and FDA Staff: Current Good Manufacturing Practice Requirements for Combination Products

Guidance for Industry and FDA Staff: Current Good Manufacturing Practice Requirements for Combination Products Guidance for Industry and FDA Staff: Current Good Manufacturing Practice Requirements for Combination Products DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Comments

More information

SQF Level 2 Proposed Preventive Controls Comparison Modules 2 & 11

SQF Level 2 Proposed Preventive Controls Comparison Modules 2 & 11 http://leavittpartners.com/global-food-solutions Contact: david.acheson@leavittpartners.com April 2013 SQF Level 2 Proposed Preventive Controls Comparison Modules 2 & 11 Introduction SQF Level 2, which

More information

Micro Plastics, Inc. Quality Manual

Micro Plastics, Inc. Quality Manual ISO 9001:2008 11 Industry Lane Flippin, Arkansas 72634 QM-001-2008-F Page 2 of 39 Introduction Micro Plastics, Inc. developed and implemented a Quality Management System in order to document the company

More information

Cartel Electronics. AS 9100 Quality Systems Manual

Cartel Electronics. AS 9100 Quality Systems Manual Cartel Electronics AS 9100 Quality Systems Manual 1900 C Petra Lane Placentia, California 92870 Introduction Cartel Electronics, as a global supplier to the aviation, space, and space industries, has developed

More information

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? CEMO Congress, Paris, February 3rd, 2011 Dr. Martine Dehlinger-Kremer Vice President Global Regulatory Affairs Agenda

More information

Center for Devices and Radiological Health. This Guidance relates to FDA 21 CFR 820.30 and Sub-clause 4.4 of ISO 9001

Center for Devices and Radiological Health. This Guidance relates to FDA 21 CFR 820.30 and Sub-clause 4.4 of ISO 9001 Center for Devices and Radiological Health DESIGN CONTROL GUIDANCE FOR MEDICAL DEVICE MANUFACTURERS This Guidance relates to FDA 21 CFR 820.30 and Sub-clause 4.4 of ISO 9001 March 11, 1997 FOREWORD To

More information

Example of a food company quality

Example of a food company quality Appendix A manual Example of a food company quality Contents Date: 13/03/95 RME-QLMN-OO Page 1 of 3 Section Title ISO 9001 reference 01 In trod uction 02 Purpose 03 Scope 04 Definitions 05 Management responsibility

More information

Clinical evaluation Latest development in expectations EU and USA

Clinical evaluation Latest development in expectations EU and USA Clinical evaluation Latest development in expectations EU and USA Medical Devices: staying ahead of regulatory developments Gert Bos BSI Israel 22 April - Herzliya Copyright 2012 BSI. All rights reserved.

More information

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES A WHITE PAPER ON THE REQUIREMENTS, REGULATIONS AND OPPORTUNITIES CONTAINED IN EC DIRECTIVE 98/79/EC AND A BRIEF COMPARISON

More information

Exar Corporation Quality Manual

Exar Corporation Quality Manual Exar Corporation Quality Manual Exar Corporation 48720 Kato Road Fremont, CA 94538 Tel: (510) 668-7000 Fax: (510) 668-7001 Executive Quality Improvement Team Commitment This quality manual provides a statement

More information

Guidance for Industry Quality Systems Approach to Pharmaceutical CGMP Regulations

Guidance for Industry Quality Systems Approach to Pharmaceutical CGMP Regulations Guidance for Industry Quality Systems Approach to Pharmaceutical CGMP Regulations U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER)

More information

Today s Best Practices in Global Product Surveillance Systems

Today s Best Practices in Global Product Surveillance Systems Today s Best Practices in Global Product Surveillance Systems Presented by: Nancy Singer Compliance-Alliance Richard J. DeRisio, M.S. Kinetic Concepts, Inc. 1 Management of Complaint and Service Experience

More information

Implementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes.

Implementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes. Implementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes.. www.pharmout.net Page 1 of 15 Version-02 1. Scope 1.1. Purpose This paper reviews the implementation of the ANSI/AAMI/IEC

More information

CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES

CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES 1.0 SCOPE CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES This document establishes the minimum quality program requirements for a contractor providing equipment, material, and construction

More information

QUALITY MANUAL ISO 9001. Quality Management System

QUALITY MANUAL ISO 9001. Quality Management System Page 1 of 20 QUALITY MANUAL ISO 9001 Quality Management System Printed copies are not controlled unless marked "CONTROLLED". Upon receipt of this document, discard all previous copies. Page 2 of 20 Approval

More information

Quality Assurance Manual for Low Level Radioactive. Waste Disposal Facility

Quality Assurance Manual for Low Level Radioactive. Waste Disposal Facility WM 07 Conference, February 25 March 1 2007, Tucson, AZ Quality Assurance Manual for Low Level Radioactive Waste Disposal Facility Yasser T. Mohamed Hot Laboratories and Waste Management Center, Egyptian

More information

Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables

Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables (1987) Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables 1. General 1.1 Introduction 1.1.1 The present guidelines are to serve as a supplement to the IACS

More information

Implementation of Risk Management in the Medical Device Industry

Implementation of Risk Management in the Medical Device Industry San Jose State University SJSU ScholarWorks Master's Theses Master's Theses and Graduate Research 2010 Implementation of Risk Management in the Medical Device Industry Rachelo Dumbrique San Jose State

More information

Software Development for Medical Devices

Software Development for Medical Devices Software Development for Medical Devices Overcoming the Challenges of Compliance, Quality and Cost Software is fast becoming the differentiator for manufacturers of medical devices. The rewards of software

More information

FINAL DOCUMENT. Global Harmonization Task Force

FINAL DOCUMENT. Global Harmonization Task Force FINAL DOCUMENT Global Harmonization Task Force Title: Quality management system Medical Devices Guidance on corrective action and preventive action and related QMS processes Authoring Group: Study Group

More information

Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP)

Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP) Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP) 1 Quality Systems Regulation (QSR) 2 Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP) Sets of checks

More information

DICOM Grid, Inc. January 25, 2016. Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington Street, Suite 100 ALEXANDRIA VA 22314

DICOM Grid, Inc. January 25, 2016. Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington Street, Suite 100 ALEXANDRIA VA 22314 DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 DICOM Grid, Inc. Senior

More information