Battery Issues from Standpoint of Healthcare Technology Managers (HTM)

Similar documents
IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS. For use with a Power Wheels vehicle that requires a 12-Volt 9.5 Amp/Hr lead acid battery.

The Role of Key Professionals in Improving Patient Outcome through Technology Life Cycle Management Clinical Engineer

Battery Replacement Instructions First Alert 148 & 168, Vista- 10, Vista- 20, DSC 1555 and DSC 1616 Residential Control Panels. Battery Safety Warning

User Manual (English)

Healthcare Technology Management (HTM)

Help ensure the integrity of your lifesaving system.

Medical Equipment Management Plan

Clinical Alarms Improvement Initiative

Software-based medical devices from defibrillators

Dear Neuton Owner, This is a helpful hint to guide you in getting the most out of your NEUTON Battery and Charger.

Technical Update TAA.TU Rev. 1

3 WATT LED SPOTLIGHT Model No. SLM

BC-5000 OPERATIONS MANUAL BATTERY CAPACITY TESTER COFKO LLC.

TT TOYS TOYS - OWNER S MANUAL. We are delighted that you have decided to purchase this TT Toys Toys children s car.

Scheduled Maintenance

How To Use A Power Supply Unit (Upu)

PROPOSED BASALT QUARRY

St Peter s College ELECTRICAL SAFETY. B. University Safety Office Policy Statement S4/10, Working safely with Electricity.

The Growing Concern Surrounding Medical Alarm Fatigue

Energizer Non-Rechargeable Batteries: Frequently Asked Questions

Chapter 3 Using a Risk Based Assessment for Establishing a Medical Equipment Maintenance Program

The Growing Concern Surrounding Medical Alarm Fatigue

Ride-on Car OWNER S MANUAL KT1151TG. Keep instructions for future reference

CappController Pipette Controller. Operation Manual Version:

User s Manual AURORA 1.2K/2.2K

Typical water pump claims and causes

Comparison of CMS Preventive Maintenance Regulations

Customer Support Agreements

How to conduct risk management & vulnerability assessments of medical devices using current ISO/IEC standards as a guidance

EXTENDED SERVICE AGREEMENTS

User Manual HomeGrid 3000 Solar Generator

Installation Instructions for Alarm Module Kit A043F059

Porsche Approved Certified Pre-Owned. Limited Warranty

Operating instructions in the back. PD1020L

2013 VTech Printed in China US

10000HXL31/15000HXL31/20000HXL31

safe, smart, protected

BroadBand PowerShield. User Manual

Emergency Power System Services Industrial UPS, Batteries, Chargers, Inverters and Static Switches

car + boat Box Contents x2 x _DC16_CAR_BOAT_OPSMAN.indd 1 29/7/08 15:27:01

Optimizing Total Cost of Ownership (TCO)

USER S MANUAL. MaxPower UPS. Uninterruptible Power System 28-2MAXPO0018

Match. GE Digital Energy. Uninterruptible Power Supply VA. Technology for the Digital World. Match UPS. GE Digital Energy.

How To Get A Great Service From Swersey Electric

INSPECTION AND TESTING OF EMERGENCY GENERATORS

National- Spencer Inc.

Corrections and Removals 21 CFR Part 806

Connection method for charging systems a key element for electric vehicles

Instruction manual of pumps WQ, WQF, SN, Furiatka, Furia, KRAKEN, BIG, IP, IPE, SWQ, CTR

FDA Perspectives on Human Factors in Device Development

Voltmaster Trash Pumps Model TSP2, TSP3 and TSP4

user s manual Battery Case model #: SPB3200 Battery Case Charger for Samsung Galaxy S 4

Risk Management for Medical Devices

Kia 7 Year Warranty. Terms and Exclusions

1 / 4 IN. FFL x 4 FT. 9 IN. 1 / 2 IN. ACME x 4 FT. 9 IN. Gauge Accuracy +/-2% Set includes Quick Disconnect Valves

Operating Instructions

"#$%&'(!)*+,-! "./0,!%&,!*12%(*3%!'4!%&,!5'3$+,6%!&,(,7!.&,!*12%(*3%!82!%/083*99/!*!2&'(%!2$++*(/!'4!%&,! 3'6%,6%2!'4!%&,!5'3$+,6%7-!

Human Factors and Medical Device Accidents. Presented by Frank R. Painter, MS, CCE University of Connecticut

RCM++ Report Sample. The analysis shown in the report is based on the SAE JA1012 A Guide to the Reliability-Centered (RCM) Standard guidelines.

Current Monitoring Kit QUICK START GUIDE

QUESTIONS TO ASK WHEN REPLACING YOUR HVAC TOGETHERWESAVE.COM

Austin Independent School District Police Department Policy and Procedure Manual

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

Keep Moving.

Enterprise Remote Monitoring

Optimization of Clinical Engineering in Private Healthcare

Charging what can be more simple? SAE J1772

Current Monitoring Kit

NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

Your flexible care options from Foster

3.5 mm Audio Jack. Manual innoxplore ix B24 A2DP Stereo Bluetooth Audio Adapter Dongle 3.5mm Audio Jack [ix B24] Contents

400 Amp Rechargeable Jump Start System RAC-HP082

TERMS AND CONDITIONS 1 CONTRACT INFORMATION

Back-UPS 650 VA 120 V with AVR (BX650CI-LM)

KOOLKARE CAR & COMMERCIAL AIR CONDITIONING SERVICE SOLUTIONS

Facility Safety and Emergency Management (FSE)

The risk assessment of fire damper installations

Medical Equipment Management. Medical Equipment Management Activities (EC and EC )

Healthcare Technology in Non- Clinical Environments: Challenges and Opportunities for Home Care

GALSON LABORATORIES, INC. New Account Information. (Please submit a Sales Tax Exemption Certificate if applicable.) Purchase Order Required via

uptimeplus.co.uk predictive maintenance strategy galileo Supported by

Public relations activities of Kansai Electric Safety Inspection Association concerning safe use of electricity

IMPORTANT SAFETY NOTICE

Contract hire driver s guide

VOLVO PENTA DIESEL AND GASOLINE ENGINES LAYING-UP AND LAUNCHING GUIDE

INDUSTRY WIDE LABOR-MANAGEMENT SAFETY COMMITTEE SAFETY BULLETIN #23

Procurement s Role in Plant Reliability and Safety. Reduce Instrumentation Costs Without Sacrificing Reliability and Safety

Operating Instructions

National Equipment Management Policy Country

Slashing Imaging Equipment Maintenance Costs While Increasing Uptime and Reliability

Leak Detection System

APC Step-Down Transformer. User s Manual

Technician High Pressure Pump Guide for the 7.3 Power Stroke Engine

CMMS & SCADA. SCADA Intro CMMS Intro. SCADA/CMMS Interfaces Real World Applications

Inspection, Testing and Maintenance

Kia 7 Year Warranty. Terms and Exclusions

Industrial Process Pump Safety Manual IMPORTANT SAFETY NOTICE

The Third Rail System User Manual

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

SUBJECT: SAFETY MANAGEMENT PLAN REFERENCE #1003 PAGE: 1 DEPARTMENT: HOSPITALWIDE OF: 1 EFFECTIVE:

Transcription:

Battery Issues from Standpoint of Healthcare Technology Managers (HTM) Alan Lipschultz CCE PE CSP President HealthCare Technology Consulting LLC Wilmington, Delaware Alan@HCTC.pro www.hctc.pro 302-584-1050 1

About AAMI (Association for the Advancement of Medical Instrumentation) AAMI is a diverse community of nearly 7,000 healthcare technology professionals united by one important mission supporting the healthcare community in the development, management, and use of safe and effective medical technology. AAMI is the primary source of consensus and timely information on medical instrumentation and technology. 2

Who do I represent?? I will try to represent the viewpoint of Healthcare Technology Managers (HTM) AKA Clinical Engineers (CE) or Biomedical Engineers (BME) or Biomedical Equipment Technicians (BMET) Director of Clinical Engineering for 37 years Started consulting firm in 2011 Very active in AAMI for 39 years 3

General Outline of Talk Survey on battery issues sent out to HTM to get data from many institutions Data from Aramark Healthcare Technologies Overall Conclusions One polling question for this group 4

HTM Survey on Battery issues 216 responses Does your HTM program routinely replace batteries at specified intervals, without waiting for failure? 84% YES, of these: 6% said All medical devices with rechargeable batteries 49% listed all Life Support 64% listed debrillators 33% listed IV pumps 26% listed UPS systems 20% listed transport physiologic monitors Others listed IABP, battery operated suction, transport incubators, etc. Others said only sealed lead acid batteries A few said they tested all, and replace bad ones. 5

HTM Survey on Battery issues 216 responses Do you routinely purchase replacement medical device batteries from a 3rd party distributor? 91% YES, of these: 8% still bought OEM batteries for some equipment types, like defibrillators or life support equipment 3% specified same battery manufacturer 4% tested batteries upon arrival. Some on random basis, some only for new battery types Most said that they only used one or more trusted vendors 6

Are there other significant medical device battery issues that should be raised at this workshop? I got 147 comments Need better batteries & chargers Cost & reliability User education plug in battery How much time before device fails? Battery alarm strategy Battery enclosures - battery leakage & swelling, access to change battery Confusion over battery types, how to handle & store. Recommended a guidance document from UK 7

Guidance on the Safe and Effective Use of Batteries and Chargers for Medical Devices This UK report covers a variety of issues including: disastrous failures, where the battery overheats, leaks acid or explodes unexpected battery exhaustion during use inadequate battery and charger maintenance incorrect charging of rechargeable batteries use of replacement batteries that do not meet the required specification poor training of staff in battery usage, maintenance, charging or replacement requirements battery chargers and their leads and connectors overheating use of battery chargers that are incompatible with the batteries in the medical device. 8

Evidence-Based Maintenance: Comparison of Equipment Failure Rates Using OEM versus non-oem Batteries Salil Balar, MS, MBA

Data Analyzed Each equipment listed represents a certain brand & model Aramark routinely changes batteries on a periodic basis, based on experience with brand & model Each hospital uses either OEM-supplied or non-oem supplied batteries EQUIPMENT Defib A Defib B Defib C Doppler PCA Pump Inf Pump B Ventilator Inf Pump A BATTERY #HOSPITALS #UNITS SOURCE OEM 5 182 non-oem 4 132 OEM 4 145 non-oem 13 290 OEM 4 134 non-oem 13 471 OEM 9 147 non-oem 8 113 OEM 2 120 non-oem 4 126 OEM 1 531 non-oem 6 1,926 OEM 18 323 non-oem 5 90 OEM 7 2,798 non-oem 12 3,070 TOTAL 10,598 10

Failure Cause Codes for Equipment - SM Failures MAINTENANCE TYPE Scheduled maintenance (SM) including inspection, calibration, and preventive maintenance CAUSE CODE EF HF PF NPF DESCRIPTION Evident failure, i.e., a problem that can be detected--but was not reported--by the user without running any special tests or using specialized test/measurement equipment. Hidden failure, i.e., a problem that could not be detected by the user unless running a special test or using specialized test/measurement equipment. Potential failure, i.e., a failure that is either about to occur or in the process of occurring but has not yet caused the equipment to stop working or problems to patients or users. No problem found. 11

Failure Cause Codes for Equipment - CM Failures MAINTENANCE TYPE Corrective maintenance (CM), including repairs performed for failures detected during SM CAUSE CODE UPF USE PPF SIF CND FFPM DESCRIPTION Unpreventable failure, evident to user, typically caused by normal wear and tear but is unpredictable. Failures induced by use, e.g., abuse, abnormal wear & tear, accident, or environment issues. Does NOT include use error (typically no equipment failure) Preventable and predictable failure, evident to user. Service-induced failure, i.e., failure induced by corrective or scheduled maintenance that was not properly completed or a part that was replaced and had premature failure ( infant mortality ). Cannot duplicate. Includes use errors. Same as NPF. Failure found during PM (to avoid duplication of codes) 12

Failure Cause Codes for Peripheral Failures MAINTENANCE TYPE CAUSE CODE DESCRIPTION CM or SM BATT Battery failure, i.e., battery(ies) failed before the scheduled replacement time. ACC Accessory (excluding batteries) failures NET Failure in or caused by network, The BATT code is what was used for this study. Each BATT is counted as a failure and failure rate is computed as: #BATT coded reports/#units, in the period of one year. 13

ARAMARK RESULTS 60% 100% Battery-Induced Equipment Failure Rate OEM 80% non-oem Annualized Failure Rate 40% 20% 60% 40% 20% 0% Defib A Defib B Defib C Doppler PCA Pump Inf Pump B Ventilator 0% Inf Pump A 14

ARAMARK CONCLUSIONS 1. The cause of premature battery failures seems to be a mismatch between the OEMs expectations and clinical users understanding and ability to care for the batteries. 2. This is a human-factors issue similar to the use errors and alarm fatigue challenges; numerous medical devices are involved. 3. Strict regulation of medical device batteries is unlikely to solve this issue. Quite the contrary, regulatory approach is only likely to increase costs and waste resources. 4. Recommend that multidisciplinary committee(s) to be created to establish standard practices acceptable to both OEMs and clinical users, as well as HTM professionals. 15

Overall Conclusions 1. Multiple battery technologies present challenges for HTM 2. Safety risks device doesn t work; or physical hazards from batteries 3. Batteries need frequent replacement 4. Disposal issues are significant 5. Costs for battery replacement are sigificant 6. Need for industry wide group to tackle battery issues collaboratively 16

Yes No Do you agree that an outcome of this workshop should be a: Recommendation that multidisciplinary committee(s) to be created to establish standard practices regarding rechargeable battery care used in medical equipment; acceptable to OEMs, clinical users and HTM professionals? A. Yes B. No 85% 15%

Please contact one of us for questions or suggestions Alan Lipschultz, PE, CCE, CSP President, HealthCare Technology Consulting LLC Telephone: 302-584-1050 Email: Alan@HCTC.pro Salil Balar, MS, MBA Director, Technology Management ARAMARK Healthcare Technologies Telephone: 704-948-5719 Email: balar-salil@aramark.com 18