Suggestions for Care and Handling of Fine Surgical Instrumentation

Similar documents
care & cleaning guide

Profiles for floors of same height Cerfix Projoint

GENERAL MAINTENANCE MANUAL

& CLEANING GUIDE MAINTAIN BEAUTY & PERFORMANCE

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised

Metal Rescue Rust Remover Technical Data

Installation Instructions

Keeping Your Swimming Pool and Spa Healthy Environmental Health Guide

Thermo Scientific Rotor Care Guide. Protect your rotor investment

White rust on galvanized and galvanized pre-painted steel. Technical Information

THE BASICS OF DECONTAMINATION

Introducing an All New Line of Cleaning Chemicals

Test Methods for the Colour Fastness of Leather & Dyes. TFL Leather Technology Ltd. Quality and Environment February 2004 Version 1.

Austin Peay State University Department of Chemistry CHEM Empirical Formula of a Compound

KODAK Developer System Cleaner and Neutralizer

WATER CHEMISTRY AND POOL WATER BALANCE

The Owners Coated Coil, Cleaning and Maintenance Service Manual.

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION

ACO Cleaning principles for drainage

Operator Quick Guide EC SENSOR

Tractor Painting Tips

OXIDATION-REDUCTION TITRATIONS-Permanganometry

DR/4000 PROCEDURE. CHLORINE, Free

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL

CHARISMA SHOWER/TUB DOOR INSTALLATION INSTRUCTIONS

MAINTAINING YOUR SPA WATER

Never use a spa or hot tub alone. Do not allow children to use a spa without supervision.

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)

THE BLUE WASHING BOOK

Detergents in CSSD. Wim Renders

Scaletec Plus. Descaling Overview, Guidelines, Cleanup Tips, And Question & Answers

STERILIZATION د STERILIZATION

BUTTERFLY SHOWER DOOR INSTALLATION INSTRUCTIONS

Experiment 12- Classification of Matter Experiment

IMO Approved Tank Cleaning Chemicals

CLEANING AND SANITIZING

Sugar or Salt? Ionic and Covalent Bonds

A Guide to Trouble-Free Cooling Towers

Operating and Installation Instructions

Care and Cleaning for KOHLER Products

A Primer for Protecting Your Investment in Surgical Instru m e n t s. Official Guide: Maintaining and Cleaning Surgical Instruments

Infection Control in the Dental Practice through Proper Sterilization

Experiment 7: Titration of an Antacid

Oxivir Tb and Accelerated Hydrogen Peroxide (AHP) Frequently Asked Questions

Instruction Manual. Image of SP-3015 & SP Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

Bathroom Installation Guide

Sulfites for Oxygen Control

Pilkington Reflite Handling and Processing Guidelines Solar Control

Water Treatment & Purification Chemicals

RBS detergents for laboratories

Laboratory Glassware Cleaning

Waste and Cost Reduction Techniques for Small Parts Cleaners

Safety, Cleaning, and Chemical Disposal Procedures

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Instruction manual. VWR unstirred water baths

Mixtures and Pure Substances

Instruction Manual. AWK-115S

Chem 100 Lab Experiment #9 - ACID/BASE INDICATORS

A. Work on this project shall consist of, but is not limited to, the following:

Physical and Chemical Changes

Healthcare Environments Cleaning, Care, and Maintenance

The Chemical Adjustment Reference Cheat Sheet

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN or ).

How to prepare the Green Cleaning Recipes for a Healthy Home pages:

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

Bathroom faucets. 10 Year Limited Warranty INCLUDED

PART 1 - INTRODUCTION...

Hands-On Labs SM-1 Lab Manual

A Guide to Stains, Rings, Discoloration, Mineral Deposits: What They Are & How To Take Care Of Them

conductivity emulsification saponification wetting peptizing dispersion rinsability controlling mineral deposits Sequestration chelation

Recovery of Elemental Copper from Copper (II) Nitrate

Heat Transfer. Condensation. Revaporization

12. Inspection and Maintenance

INSTALLATION MANUAL. Glass Vessel Sink. I toll free: I Kraus USA Inc.

Acids, Bases, and ph

Protecting Galvanized Steel Sheet Products from Storage Stain Rev 2.1 Aug-03

CARE AND MAINTENANCE GUIDELINES FOR LAMINATE & WOOD CABINETS AND THEIR SURFACES. Roller and Ball-Bearing Slide Systems 7

Care and Handling of Rotors

Private Water Supplies Sampling Manual. A Field Guide

STAINLESS STEEL SINKS

ph Measurements of Common Substances

Non-polar hydrocarbon chain

Physical and Chemical Properties and Changes

Chapter 7 Equipment and Utensil Cleaning and Sanitization

A Volumetric Analysis (Redox Titration) of Hypochlorite in Bleach

VISION SHOWER DOOR & TUB DOOR INSTALLATION INSTRUCTIONS

8 Chemicals from water treatment and distribution

Paper Chromatography: Separation and Identification of Five Metal Cations

USE, CLEANING AND MAINTENANCE SILESTONE AND ECO BY COSENTINO

Steam Sterilization Cycles for Lab Applications

DETERGENTS. Application Recommendations. Application/ Concerns. Recommended Detergent. Articles Cleaned/ Soil Removed. Cleaning Method.

Portable Steam Cleaner


The National Board of Boiler and Pressure Vessel Inspectors 1055 Crupper Avenue Columbus, Ohio

THE ROWA Anti-limescale Unit MANUAL No chemicals No electricity Maintenance-free Catalytic granulate for limescale stabilising

Thermo Scientific Rotor Care Guide. Protect your rotor investment

WATER-SOLUBLE METALWORKING FLUIDS

DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included.

Espresso/Cappuccino/Single Serve Coffee Maker. Instruction Manual HY YEAR LIMITED WARRANTY

CLEANING AND SANITIZING

Transcription:

Suggestions for Care and Handling of Fine Surgical Instrumentation WATER Tap water should not be used for cleaning surgical instruments as it frequently contains significant quantities of foreign ions that can damage instruments. High concentrations of chlorides, for example, can lead to pitting and stress-related fractures. Also, the presence of minerals in the cleaning water can discolor the instrument. (These discolorations can be removed by using the appropriate cleaning products.) To prevent these problems, it is recommended that only salt-free, demineralized water be used for instrument cleaning. PREPARATION FOR CLEANING Instruments should be cleaned immediately after use. If that is not possible, they should be placed in a basin of distilled water or a solution of neutral ph detergent until such time as cleaning is possible. However, instruments should not be stored in aqueous solutions for prolonged periods. If they are placed in a basin prior to cleaning, the basin should be drained and cleaned daily. MANUAL CLEANING SUGGESTIONS Carefully follow all manufacturers instructions regarding solution temperatures, soaking times and mixture concentration. If powdered cleaning solutions are used, be sure that the mixture has had adequate time to dissolve. We recommend using a neutral ph cleaner. Products other than neutral ph can result in instrument damage. Change the cleaning solution daily. Evaporation can increase the concentration and cause corrosion. Dirty solutions do not clean as effectively. For manual cleaning, use lint-free, soft textile cloth, paper cloth, soft plastic brushes (toothbrush) or water spray guns. After manual cleaning, instruments should be rinsed twice, first in tap water and then in distilled water. This two-step rinsing will remove any remnants of surgical debris as well as remove any contaminants that may have been in the tap water. After rinsing, instruments should be thoroughly dried. If it is possible, forced air should be used for drying. If that is not possible, gently shake the excess water from the instruments and place them on clean, dry, lint-free towels to dry. Page 1 of 5 25-0001 0712

ULTRASONIC CLEANERS As in all cleaning methods, the solutions used for cleaning must be prepared according to manufacturer's instructions. Failure to do so will result in either an acidic or alkaline ph, both of which can cause corrosion and breakage. We recommend using a neutral ph cleaner. Products other than neutral ph can result in instrument damage. Breakage resulting from corrosion is called "stress corrosion cracking." This usually occurs around the box lock of needle holders and hemostats or on the screws of scissors. It appears as a buildup of what looks like rust. If seen, it should be cleaned away immediately with a soft toothbrush. STEAM STERILIZATION Complete an "empty" cycle every day prior to sterilizing the instruments. To prevent corrosion damage to the instruments, it is important that the steam be completely free of foreign substances. To avoid excessive condensation, stay with the manufacturer's recommended load for the sterilizing unit. If heavy sets are unavoidable, instruments should be spread out to reduce the condensation buildup. Care must be taken with delicate scissors to prevent tip breakage. Once sterilized, instruments must be dried completely before being stored away. Without proper drying times, "water spotting" can result. LUBRICANTS AND DETERGENTS Besides lubricating moving parts, lubricants inhibit rust, corrosion and discoloration. As part of a regular instrument care program, proper lubrication helps reduce instrument wear. A one minute soaking prior to autoclaving is adequate to insure protection. Neutral ph detergents are ideal for surgical instruments. The neutral ph will not damage stainless steel or tungsten carbide inserts. The solution is gentle enough for manual as well as ultrasonic cleaning. Page 2 of 5 25-0001 0712

INSTRUMENT CARE SUGGESTIONS 1. Before cleaning, inspect instruments for damage. (Don't autoclave chrome plated and stainless steel instruments together.) 2. After each use, clean instruments thoroughly. a. Use only neutral ph solutions for the pre-soak. b. When cleaning by hand or with ultrasonic cleaner, instruments should be in the open position. c. To remove stubborn stains, use a non-metallic brush (toothbrush). Abrasive cleaning solutions or scouring pads should NOT BE USED. d. DO NOT USE bleach for cleaning. e. Follow the mixture rations precisely for ALL solutions that will touch your instruments. Failure to monitor the acidic (PH 0-6) or alkaline (PH 9-14) balance will lead to instrument failure. Even though instruments are passivated, serious damage can result if there is exposure to any of the following: Hydrochloric acid Ferric chloride Dilute suplhuric acid Iodine Bleach Whenever possible, the following substances should be kept away from instruments: Aluminum chloride Lysol Dalkin's solution Barium Chloride Mercury Chloride Ferrous Chloride Biochloride of mercury Phenol Stannous chloride Calcium Chloride Carbolic acid Tartaric acid Potassium Permanganate Potassium thiocynate Chlorinated lime Sodium hypochlorite HANDLING NEW INSTRUMENTS Following the procedures outlined previously, clean and rinse new instruments before their first sterilization. Instruments should not be stored in a cabinet with any chemical that produces a corrosive vapor. MANUAL CLEANING SUGGESTIONS 1. Check instrument for cracks and chipped areas. 2. Clean instruments immediately after use with a neutral ph solution. Be sure to follow manufacturer's instructions for soak time and mixture ratio. 3. When cleaning, use a soft brush, NOT a steel brush or scouring pad. Clean in open position and pay special attention to removing debris from box locks. 4. Rinse instruments thoroughly after cleaning. Use distilled water, not tap water. S. Dry instruments with clean, dry, lint-free towels. 6. Before sterilizing, soak the instruments in lubricant for one minute. Do not use dry heat sterilizers. 7. Autoclave according to manufacturer's instructions. Page 3 of 5 25-0001 0712

ULTRASONIC CLEANING SUGGESTIONS 1. Check instrument for cracks and chipped areas. 2. Clean instruments immediately after use with a neutral ph solution. Be sure to follow manufacturer's instructions for soak time and mixture ratio. 3. When loading ultrasonic cleaner, do not stack instruments on top of each other and be sure instruments are in the open position. 4. Follow manufacturer's instructions for cleaning cycle time. Remove instruments promptly when cycle is complete and rinse thoroughly in distilled water. 5. Dry instruments with clean, dry, lint-free towels. 6. Before sterilizing, soak the instruments in lubricant for one minute. Do not use dry heat sterilizers. 7. Autoclave according to manufacturer's instructions. Page 4 of 5 25-0001 0712

Cleaning and Sterilization Trouble-Shooting Guide Problem Cause Prevention Spotting Mineral deposits left by slow or improper 1. Check operation of autoclave. 2. Use chloride-free solutions for drying sterilizing, disinfecting, rinsing, and cleaning. (Distilled or mineral free water.) Rust (Corrosion) Film left by steam 1. Check purity of water supply 2. If water softeners are used, check for composition. 3. Purge Steam pipes, especially new installations. 4. 4. Clean inside of sterilizer regularly according to manufacturer s specifications. Rust (Corrosion) Deposit 1. Do not mix stainless steel with other metals, especially where there is evidence of defective plating. 2. Rinse with distilled water (important where tap water may have high metallic content). 3. Remove debris from box lock. 4. Dry all instruments thoroughly. Use fulltime cycle Pitting Chemical and electronic attack of surfaces (important when instruments are wrapped). 1. Rinse instruments thoroughly immediately after use. 2. Avoid long exposure to chlorides and acids. 3. Do not use detergents having high ph levels. 4. Do not mix metals in ultrasonic cleaners. Black to purple stains Ammonia 1. Avoid exposure to ammonia in solutions and cleaning compounds. 2. Rinse instruments thoroughly (distilled water preferred). Brown stains Minerals 1. Check water supply in sterilizer. 2. Check cleaning compounds and detergents. 3. Clean inside of sterilizer regularly according to manufacturer s specifications. Page 5 of 5 25-0001 0712