VALIDATION OF STERILIZATION EQUIPMENTS

Similar documents
Steam Sterilization and the 2007 Revision of PDA Technical Report 1

or

Guidance on safe use of Autoclaves

POROUS BURNER - A New Approach to Infrared

Sterilization methods and equipment Lab 1-2

Steam Sterilization Principles

BSL3 and BSL4 Autoclaves. Laboratory Line

Heat Transfer. Condensation. Revaporization

Steam Sterilization Cycles for Lab Applications

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Validation and Management of Heat Sterilization (Autoclave and Dry Heat Oven)

Terminal Sterilization. vs. Aseptic Processing

Partnership. Medical Waste Autoclaves. Medical Waste Autoclaves

Fact Sheet on Steam Sterilizers at Stanford University

Top 10 considerations when validating an autoclave

FDA and the Compounding Pharmacy

Laboratory- and sterilisation equipment HMC EUROPE

Effective Heat Sterilization in CO 2 Incubators

Technical Specification for Model RSK2503 (Z)-06A IR Belt Solar Cell Firing Furnace

Biological Indicators and

Cooking at the Speed of light!

Laboratory Line. Vertical and Bench-Top Life Science Autoclaves

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/ Ø90x160mm W200xH385xD (Net:28)

Vertical and Benchtop Autoclaves for Life Sciences. Laboratory Line

AUTOCLAVES. Table Top. STE-TEN, 1 Liter Autoclave

STERILIZATION AND DISINFECTION

Sterile Cleanroom Management

LABOKLAV. Autoclaves Made in Germany.

Air Removal from the Sterilization Chamber

Particle Monitoring Requirements in Pharmaceutical Cleanrooms

Understanding pasteurisation and sterilisation processes

Scottish Health Technical Memorandum 2010 (Part 5 of 6)

PSAC/SAIT Well Testing Training Program Learning Outcomes/Objectives Level B

Medical Waste Autoclaves. Medical Waste Autoclaves

BB-18 Black Body High Vacuum System Technical Description

TEACHER BACKGROUND INFORMATION THERMAL ENERGY

AUTOCLAVE STEAM STERILIZERS

Model Voltage (V/Hz) Power (W) Capacity (L) Chamber size Outside size(mm) Weight (kg) STE-TEN 220/ Ø90x160mm W200xH385xD (Net:28)

Specific Volume of Liquid (Column 7). The volume per unit of mass in cubic feet per pound.

Critical review on qualification of sterilization equipment in aseptic processing

Large Steam Sterilizers for the Life Sciences. Laboratory Line

Determination of Thermal Conductivity of Coarse and Fine Sand Soils

STEAM STERILIZERS. DLOV & DLOM series. more than fifty years of experience in sterilization. DE LAMA s Quality Management System is certified:

Chapter 7: Oil Vapor Diffusion Pumps

German engineering under high pressure

Heating Water by Direct Steam Injection

Packaging Validation according to ISO 11607

Fundamental Autoclave Techniques

CONTENTS. ZVU Engineering a.s., Member of ZVU Group, WASTE HEAT BOILERS Page 2

A/C refrigerant system, overview

Incubator with mechanical convection

EUROPEAN STANDARDS AND MARKINGS FOR HAND AND ARM PROTECTION

T5 Watt-Miser Linear Fluorescent lamps

Laboratory & Scientific Line. Bulk Steam Sterilizers. Since

Die Ovens. This Chapter Sponsored by CASTOOL Tooling Solutions

COMMISSIONING OF HVAC SYSTEMS

Environmental Monitoring of Clean Rooms

THE BASICS OF STERILIZATION

Temperature Calibration; Depths of Immersion

FUNDAMENTALS OF ENGINEERING THERMODYNAMICS

Rusty Walker, Corporate Trainer Hill PHOENIX

HVAC Design Requirements

DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 4 QUALITY MANAGEMENT IN STERILIZIATION

FACT SHEET : Using Autoclaves Safely

Application Note USTR 805. Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges

Soil Suction. Total Suction

High Capacity Hot Air Dryer

National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook

WHITE PAPER April 18, 2016

Cleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9

Sterile E Online NetworkTutorials

Laminar Flow Cabinets

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

KGI MEDICAL WASTE MANAGEMENT PLAN

OVERVIEW OF STANDARD 7E & STANDARD 20

Overview of Waste Heat Recovery for Power and Heat

T5 LongLast. GE Lighting DATA SHEET

Heating oven with mechanical control

GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS

UCCS Biosafety Management Policy

Heat and cold storage with PCM

Type Testing Procedure for Crankcase Oil Mist Detection and Alarm Equipment

SECTION TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION

5.2. Vaporizers - Types and Usage

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

Water for Injection (WFI) and Pure Steam Production in a Lean Manufacturing Environment. Larry Zanko Project Manager STERIS Corporation

CONDENSATION IN REFRIDGERATED BUILDINGS

Accurate Air Flow Measurement in Electronics Cooling

Horizontal Sterilizer MU 012.B / I (09/07) Page 2/46. 1 Index... 2

Chapter 18 Temperature, Heat, and the First Law of Thermodynamics. Problems: 8, 11, 13, 17, 21, 27, 29, 37, 39, 41, 47, 51, 57

A GUIDE TO SELECTING YOUR INDUSTRIAL OVEN

F0 a technical note. What it means How to calculate it How to use it for adjustment, control and validation of moist-heat sterilization processes

Microbiological Testing of the Sawyer Mini Filter. 16 December Summary

Clinic Washer-Disinfectors. Washer-Disinfector Line

Condensing Boiler Efficiency

High-ohmic/high-voltage resistors

Peltier Application Note

TABLE OF CONTENT

Chapter 12 - Liquids and Solids

Overview. Introduction Cooling Tower Basics Principles of Operation Types of Cooling Towers Common Applications Design Considerations

Transcription:

Slide: 1/51 VALIDATION OF STERILIZATION EQUIPMENTS Aseptic Area Validations Pharmaceutical & Chemical Industry Research and Development Foundation

Slide: 2/51 Content: Definition of Sterilization and Depyrogenation Microbiological aspects of Sterilization and Depyrogenation, Lethality calculation, D- Value, F H & F 0 Values Z- Value and use of microbiological indicators.

Slide: 3/51 Content: - Dry Heat Ovens - Dry Heat Sterilization Tunnels - Steam Sterilizator (Autoclaves) 1. Design Qualification 2. Installation Qualification 3. Operational Qualification 4. Performance Qualification 4.1. Thermodynamical aspects of Sterilization 3.2. Temperature Distribution and Heat Penetration studies.

Slide: 4/51 Definitions: 1. Sterilization: Validated process used to render a product free of living microorganisms including bacterial endospores. 2. Depyrogenation: Removal or inactivation of bacterial endotoxin.

Slide: 5/51 Sterilization Only: The cycle is designed to assure that the probability of survival of the native microflora is no greater than one cell in one million units of the commodity. (10-6 probability of nonsterility) Dry Heat Sterilization, Theoretical requirement: 170 0 C, 32 min. Steam Sterilization Theoretical requirement: 121 0 C, 15 min.

Slide: 6/51 Sterilization - Overkill The overkill approach provides assurance of sterilization well in excess of the 10-6 probability of non- sterility. For example an F H provided by an overkill cycle may produce a 12- log reduction of a biological indicator that exhibits a high resistance to dry heat.

Slide: 7/51 Sterilization & Depyrogenation Applies to the cycles where the purpose is both sterilization and depyrogenation. Whenever depyrogenation is a desired end point, relatively high temperatures and/or extended heating times are necessary. Thus, microbial lethality delivered by these cycles provides a margin of safety far in excess of a 10-6 probability of nonsterility. Dry Heat Depyrogenation Theoretical requirement: 250 0 C-30 min.

D - Value : Time required for one log (or 90%) reduction of microorganism population at base temperature. Microbial Death Curve Slide: 8/51 Log numb. of Survivors 1000000 100000 10000 1000 100 10 0,1 1 0,01 0,001 0,0001 0,00001 0,000001 0,0000001 D-Value= 1.0 min. 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Minutes at Base Temp.

Slide: 9/51 Determination of Z - Value: Determine the D - value of an organism at min. three different temperatures. Construct a Thermal Death Curve by plotting the logarithm of the D- Value versus temperature.

Slide: 10/51 Z-Value: Death Rate Constant Assesment of Z Value 100 D 130 0 C : 10 min. Log D Value Z Value = 20 0 C 10 D 0 150 C : 1.0 min. 1 120 130 140 150 160 170 0,1 Temperature 0 C

Slide: 11/51 Z-Value: In general, for Dry Heat sterilization, Z- Value may be assumed as 20 0 C. And for Steam Sterilization as 10 0 C. However, it will be appropriate to verify for the biological indicators when they are used to measure the integrated lethality of a dry heat or steam sterilization cycle.

Slide: 12/51 LETHALITY RATE: Also defined as : F H For Dry Heat Sterilization F o For Steam Sterilization The equivalent sterilization time spent at the base temperature. T b : 170 o C (For Dry Heat Sterilization) T b : 121 o C (For Steam Sterilization)

LETHALITY CALCULATION Patashnik Method Slide: 13/51 Lethality Rate : 10 (T-Tb)/Z F H = t x Lethality Rate t : Cycle time T : Actual Cycle temperature T b : Base Temperature Z : Microbial Death Rate Constant

Slide: 14/51 LETHALITY CALCULATION Example: Determination of F H of a 3 min. dry heat sterilization cycle at 175 0 C t = 3 min T = 175 0 C F H = 4 x 10 (175-170)/20 Tb = 170 0 C F H = 5.31 z = 20 0 C Sterilization at 175 0 C for 3 min. is equivalent to 5.31 min. at 170 0 C.

Slide: 15/51 Lethality in Dry Heat Sterilization Time Temperature Lethality Rate (min) ( 0 C) min. at 170 0 C 5 105 0,0006 10 110 0,0010 15 120 0,0032 20 135 0,0178 25 150 0,1000 30 165 0,5623 35 170 1,0000 40 172 1,2589 45 174 1,5849 50 174 1,5849 55 174 1,5849 60 175 1,7782 65 165 0,5623 70 150 0,1000 75 140 0,0316 80 130 0,0100 85 110 0,0010 90 105 0,0006 Σ of Lethal Rates : 10.1912 F H = t x Σ of Lethal Rates t = 5 min. F H = 5 x 10.192 F H = 50.961 min. at 170 0 C.

Slide: 16/51 PART-1 DRY HEAT STERILIZATION AND DEPYROGENATION VALIDATION

Slide: 17/51 DRY HEAT STERILIZATION & DEPYROGENATION Dry heat is often the agent of choice for sterilizing items which will tolerate high temperatures. Dry heat sterilization processes are generally less complicated than steam processes, although higher temperature and/or longer exposure times are required because microbial lethality associated dry heat is much lower than that for saturated steam at the same temperature.

Thermodynamical Aspects of Heating Process: 1. Convection Heating Process: Slide: 18/51 The heat transfer through a medium by motion of it s parts. Natural convection is a result of differences in density caused by temperature gradients in the fluid mass. Forced convection heating is effected by the action of a mechanical device.

Thermodynamical Aspects of Heating Process: Slide: 19/51 2. Conduction Heating Process: Conduction is accomplished ether by a molecular interaction from higher energy level to a lower energy level or by free electrons. Thus, the ability of solids to conduct heat varies directly with the free electron concentration. Pure metals are best conductors and non- metals are the poorest.

Thermodynamical Aspects of Heating Process: Slide: 20/51 3. Radiant Heating Process: Radiant heating is the process which energy flows from high temperature body to a lower temperature. The geometry of both source and the exposure section of the unit will affect the uniformity of the radiation density in a unit.

Slide: 21/51 Dry Heat Sterilization Equipment Validation Batch Sterilizers Dry Heat Ovens Continuous Sterilizers Sterilization Tunnel The basics of the Batch and Continuous sterilizers are mainly the same. Since the continuous (Tunnel) sterilizer validation is more complicated, the topics will concentrate on the Convection continuous process qualification.

Slide: 22/51 Batch & Continuous Processing 1. BATCH PROCESSING: Washing Manuel Transfer Sterilization Dry Heat Oven (Double Door) Filling Manuel Transfer 1. CONTINUOUS PROCESSING: Washing Continuous Transfer Sterilization (Tunnel) Continuous Transfer Filling Filling

Dry Heat Sterilization Validation 1. Design Qualification: Slide: 23/51 Facility layout, decision of batch or continuous process. Utility requirements and specifications. Pressure differential requirements. Required capacity of the sterilizer. Type of materials to be sterilized. Any requirements for presterilization.

Dry Heat Sterilization Validation 2. Installation Qualification: Slide: 24/51 The equipment should comply with the original purchase specifications. Exceptions should be appropriately documented. The structural installation like; Leveling, insulation, and air flow requirements should meet manufacturer s specifications.

Dry Heat Sterilization Validation 2. Installation Qualification: Slide: 25/51 All utility connections such as electrical and HVAC should meet the design specifications. Materials of construction of both the sterilizer and the facility should meet the design specifications.

Dry Heat Sterilization Validation Slide: 26/51 2. Installation Qualification- CALIBRATIONS: The following pieces of equipment should be calibrated by removing or in situ: Temperature sensors and recording devices Temperature Controllers (in situ) Pressure gauges Belt speed controller and recorder Cycle set point switches Velometers

Dry Heat Sterilization Validation 3. Operational Qualification: Slide: 27/51 The actual operational performance of the electro/mechanical components should be verified and documented. Electrical Logic: Ensure that each step is in the correct sequence and it s repeatable. Cycle Set Point Adjustability: Limit Switch sequencing should be verified.

Dry Heat Sterilization Validation 3. Operational Qualification: Slide: 28/51 Overload interlocks: Should not allow excess commodity build- up during processing. Gasket Integrity: Zone to zone leak rate should be within the limits at all panel gaskets. Air Balance Ability: Check that, the baffle/linkage mechanisms can be adjusted for balance.

Dry Heat Sterilization Validation 3. Operational Qualification: Slide: 29/51 Blower Rotation: Check that the blowers rotated in the specified direction and speed. Vibration Analysis:Check the dynamic balancing of the blowers to minimize the vibration in each phase. Air Balance: Check that the P is positive with respect to the preparation side of the tunnel.

Slide: 30/51 Tunnel Sterilizer Pressure Differential

Dry Heat Sterilization Validation 3. Operational Qualification: Slide: 31/51 Heater Elements: Check that all the heater elements are properly operating. Belt Speed: Check that the belt and belt speed recorder are operable. HEPA Filters: Verify the integrity of the filters.

Dry Heat Sterilization Validation 4. Performance Qualification: Slide: 32/51 In a conductive dry heat sterilization and depyrogenation process, significant variations may occur depending on the load configuration. Initial load temperature, specific heat of the load components, and the load variations should be tested for delta temperature and slowest to heat zone.

Dry Heat Sterilization Validation 4. Performance Qualification: Slide: 33/51 Temperature Distribution: External monitoring and recording instruments shall be calibrated before and after the OQ/PQ studies (3 point calibration, ± 0.5 0 C tolerance). Uniformity of the temperature distribution in case of Min&Max. loading should be verified by using Thermocouples with 3 replicates. T/C (Thermocouple) placement shall be documented on a diagram.

Dry Heat Sterilization Validation 4. Performance Qualification: Slide: 34/51 Temperature Distribution: Min. 10-12 T/C s shall be used and they should not be inserted in the load. Data should be recorded during the whole cycle at 1 min. intervals. At least one T/C shall be placed adjacent to the equipment temperature controller. Location of the cold spot should be determined and documented.

Dry Heat Sterilization Validation 4. Performance Qualification: Slide: 35/51 Heat Penetration- Acceptance Criteria: Thermocouples should be inserted into the load. At least three biological indicators and T/C s shall be placed around the cold spot. External T/C readings should comply with manufacturer s specifications (with Max ± 3 0 C difference) Biological indicator inactivation results should assure 6-log reduction for Bacillus Subtilis and 3-log reduction for endotoxin. Lethality calculation should verify the Equivalent F H value for defined cycle.

Slide: 36/51 HEAT PENETRATION STUDIES

Slide: 37/51 HEAT PENETRATION STUDIES LOAD CONFIGURATION S7 a13 b14 7 b13 a14 S4 a7 b8 4 b7 a8 S1 a1 b2 1 b1 a2 S1-S9 : Thermocouples S8 a15 b16 8 b15 a16 S5 a9 b10 5 b9 a10 S2 a3 b4 2 b3 a4 a1-a18: Bio- Indicators (Bacillus Subtilis) S9 a17 b17 a11 b11 a5 b5 9 6 S6 3 b18 a18 b12 a12 S3 b6 a6 END MIDDLE START b1-b18: Endotoxin challenged containers.

Slide: 38/51 PART-2 STEAM STERILIZATION VALIDATION

Slide: 39/51 STEAM STERILIZATION Various types of steam sterilzers are commercially available; Saturated Steam Water Immersion Water Cascade System Air-Steam Mixtures Gravity Air Displacement (unpacked materials sterilization) Vacuum air Displacement (Packed materials)

Slide: 40/51 STEAM STERILIZATION Microbiological aspects of Steam Sterilization and Dry Heat Sterilization are basically the same; D- Value is determined in the same way Z- Value = 10 0 C T b = 121 0 C Lethality (F 0 ) can be calculated in the same way.

Slide: 41/51 Lethality in Steam Sterilization Time Temperature Lethality Rate (min) ( 0 C) min. at 121 0 C 5 100 0,0080 6 103 0,0160 7 106 0,0320 8 109 0,0630 9 112 0,1260 10 115 0,2510 11 118 0,5010 12 121 1,0000 13 121 1,0000 14 121 1,0000 15 118 0,5010 16 115 0,2510 17 112 0,1260 18 109 0,0630 19 106 0,0320 20 103 0,0160 21 100 0,0080

Steam Sterilization Validation 1. Design Qualification : Slide: 42/51 Facility layout. Utility requirements and specifications. Required capacity of the sterilizer. Type of materials to be sterilized (Liquids, wrapped,hollow or porous materials) Requirement for Gravity and/or Prevacuum cycles.

Slide: 43/51 Steam Sterilization Validation 2. Installation Qualification- CALIBRATIONS: The following pieces of equipment should be calibrated by removing or in situ: Pressure Gauges Timing Devices Temperature Recording Devices Verification of safety Systems and Devices (EN- 61010 Part- 1and EN- 61010-2-041)

Steam Sterilization Validation 3. Operational Qualification: Slide: 44/51 The actual operational performance of the electro/mechanical components and utilities should be verified and documented. Clean Steam Generator (Free from non- condensables EN- 285) Air Filtration Systems and compressed air Power Source Heat Exchanger, Cooling Water

Steam Sterilization Validation 4. Performance Qualification: Slide: 45/51 Temperature Distribution: External monitoring and recording instruments shall be calibrated before and after the OQ/PQ studies (3 point calibration, ± 0.5 0 C tolerance). Uniformity of the temperature distribution in case of Min&Max. loading should be verified by using Thermocouples with 3 replicates. T/C placement shall be documented on a diagram.

Steam Sterilization Validation 4. Performance Qualification: Slide: 46/51 Temperature Distribution: At least one T/C shall be placed located in the steam exhaust line or adjacent to the equipment temperature controller. Min. 10-12 T/C s shall be used and they should not be inserted in the load. Data should be recorded during the whole cycle at 1 min. intervals. Location of the cold spot should be determined and documented.

Steam Sterilization Validation 4. Performance Qualification: Slide: 47/51 Heat Penetration- Acceptance Criteria: Thermocouples should be inserted into the load. At least three biological indicators and T/C s shall be placed around the cold spot. External T/C readings should comply with manufacturer s specifications (with Max ± 1 0 C difference) Biological indicator (bacillus stearothermophilus) results should ensure the 6-log reduction and Lethality calculation should verify the Equivalent F 0 (15 min. at 121 0 C) value for defined cycle.

Steam Sterilization Validation 4. Performance Qualification: Slide: 48/51 AIR REMOVAL TEST: The ability of the pre-vacuum autoclaves to effectively remove the air and noncondensable gases should be tested. If the air is not effectively removed, air pockets will occur in the chamber and sterilization conditions will not be attained. Bowie-Dick or DART Test pack, the uniformity of the colour change on the indicator sheet should be checked. (3.5 min. at 134 0 C )

Steam Sterilization Validation 4. Performance Qualification: Slide: 49/51 LEAK RATE TEST: The presence of air prevents proper penetration of of the load by steam and thus inhibits sterilization. Air leaking from outside into the chamber at the end of sterilization cycle will contaminate the load. A leak rate equivalent to a rate of change in pressure of 1 mm Hg/min. over a period of 10 min. after stabilization is the maximum permitted rate.

Slide: 50/51 STERILIZATION VALIDATION-GENERAL CHANGE CONTROL AND REVALIDATION Any changes to the sterilization equipment and/or related utilities should be evaluated by a Change Control Procedure. Typical Changes Requiring Revalidation Any changes in operating cycle (i.e:temperature, time, belt speed, chamber pressure) Change in load configuration. Change in sterilized materials. Major maintenance work on critical instruments/elements or utilities.

Slide: 51/51..The END.. Thanks For Your Attention, Any Questions Please?.