LOAD CELL CABLING GENERAL. REVERE TRANSDUCERS APPLICATION NOTE 03/4-02/01 Page 1

Similar documents
Subminiature Load Cell Model 8417

Load Cell Troubleshooting

PART 1 - INTRODUCTION...

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

MAGTROL. ZS Series Load Cell. ZS Data Sheet. Features. Description

Digital inductive conductivity transmitter

Mounting Instructions. Torque Transducer. A en

Microcontroller to Sensor Interfacing Techniques

T0118 T2118 T3118. Instruction Manual

ADC-20/ADC-24 Terminal Board. User Guide DO117-5

Mounting instructions. Force transducer. A en

HARDWARE MANUAL. Ixx24-A 1/20. Val Controls A/S Limfjordsvej 3 DK-6715 Esbjerg N Tel vc@valcontrols.com

Procon Engineering. Technical Document PELR TERMS and DEFINITIONS

Resistance, Ohm s Law, and the Temperature of a Light Bulb Filament

Contents. Document information

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples

Loop Calibration and Maintenance

Mounting instructions. Junction box VKK2R-8 DIGITAL. A en

Application Note 142 August New Linear Regulators Solve Old Problems AN142-1

INSTRUCTION MANUAL. SIGNAL AMPLIFIER 4-20 ma

BSNL TTA Question Paper-Instruments and Measurement Specialization 2007

Instruction Manual. T3111 T3111L T3111P T3111Ex

Section 3. Sensor to ADC Design Example

RF data receiver super-reactive ASK modulation, low cost and low consumption ideal for Microchip HCS KEELOQ decoder/encoder family. 0.

Electricity. Confirming Coulomb s law. LD Physics Leaflets P Wie. Electrostatics Coulomb s law

A99B Series Temperature Sensors

Basic RTD Measurements. Basics of Resistance Temperature Detectors

PMP 4000 Series. Druck Amplified Output Pressure Transducers. GE Sensing. Applications. Features

CONNECTOR AMPLIFIER FOR PROPORTIONAL VALVES (4-20 ma Input Version)

Pressure monitoring equipment for oil-sf 6. bushings, type GOEK

Alphasense Application Note AAN DESIGNING A POTENTIOSTATIC CIRCUIT

Environmental Monitoring with Sensors: Hands-on Exercise

Lab 3 - DC Circuits and Ohm s Law

Kit Watt Audio Amplifier

An Overview of Calibration Methods and Procedures for Process and Inventory Weigh Systems

Vehicle Brake System Monitoring 2-wire, 4-20 ma option; Intrinsic Safety Approval to E Exia IIC T4 (Tamb=60 C) BASEEFA, CENELEC EN50-020

Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter

Position Transducers with Restoring Spring 10, 25, 50, 75, 100 mm TR, TRS Series

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE

The Do s and Don ts of Pressure Transducers

Physics 3330 Experiment #2 Fall DC techniques, dividers, and bridges R 2 =(1-S)R P R 1 =SR P. R P =10kΩ 10-turn pot.

ATL Fuel Level Sender Probes

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

FAQ s on Siemens Magmeters

Clear Glass Sensor. Ordering Information E3S-CR67/62. Optimum Sensor for Detecting Transparent Glass and Plastic Bottles

PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P

Single Channel Loop Detector

THE STRAIN GAGE PRESSURE TRANSDUCER

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS

Data Sheet PS1B Pressure Switch

LEVEL GAUGES WITH MAGNETIC JOINT L 14 - L 22 - L 34

Detection of Pipe Leaks Connection of Pulse Monitoring System in Thermal Utilities

Digital Energy ITI. Instrument Transformer Basic Technical Information and Application

Solid Core and Split Core Adjustable Current Status Switches CSS-O, CSS-C; CSP-O, CSP-C

Application Note. So You Need to Measure Some Inductors?

RX-AM4SF Receiver. Pin-out. Connections

Ampere's Law. Introduction. times the current enclosed in that loop: Ampere's Law states that the line integral of B and dl over a closed path is 0

ALARMLINE ANALOGUE Linear Heat Detection System

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES

Measuring Temperature withthermistors a Tutorial David Potter

Floodlight Range Installation Guide

Vario-Series Accessories

PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller

Electronic Pressure Switch EDS 300

Objectives 200 CHAPTER 4 RESISTANCE

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

ABB i-bus EIB / KNX Analogue Input AE/S 4.2

Displacement encoders

CPA PC-Programmable Current/Voltage and RTD/Thermocouple Limit Alarm Trips

Check out our website for this handbook and more helpful information on load cells and weigh modules

Calibration with bridge measurements

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

Electrical Fundamentals Module 3: Parallel Circuits

Inductors in AC Circuits

Modular I/O System Analog and Digital Interface Modules

ABB Drives. User s Manual. Pulse Encoder Interface Module RTAC-01

APPLICATION NOTES OF SIGNAL ISOLATORS

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

S300 Series Dust Monitors

Current Probes. User Manual

J TB/TW Temperature limiter, temperature monitor

Load Cell Amplifier Module. Instruction Manual LCA

Supplementary Certificate of Approval NMI S669

Temp-485-Pt100. A temperature sensor (Pt100 or Pt1000) communicating over the RS-485 bus with a simple communication protocol

Application Note - Connecting an Electricity Meter to SolarEdge Devices (Europe and APAC)

MINI MCR-SL-R-UI(-SP)

ELECTRIC HEATER COMPANY Operating and Maintenance Manual For Immersion Heating Elements

HLC A1... HLC B1... HLC F1... Data Sheet. Load cells. Special features. B en

T1-40 T1-40BF T1-40V T1-38 T1-38BF T1-38V

SWISS MADE. Series 320 XB MASTER. Analytical. Precision

Series and Parallel Circuits

Experiment #5, Series and Parallel Circuits, Kirchhoff s Laws

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

Sense it! Connect it! Bus it! Solve it! EncoderS

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin

Mounting instructions. Acceleration Transducer B12. B 26.B12.10 en

Fig. 1 Analogue Multimeter Fig.2 Digital Multimeter

LM134-LM234 LM334 THREE TERMINAL ADJUSTABLE CURRENT SOURCES. OPERATES from 1V to 40V

ACTIV-8 PLUS Insulated Conductor Bar

Transcription:

LOAD CELL CABLING GENERAL Most industrial load cells are used in multiple load cell weighing systems. Load cells should be electrically connected in such a way that the signal (output) lines, excitation (power supply) and sense (when present) lines are in parallel. Usually the connection is not made at the indicator, but in a separate housing, a so called junction box, located adjacent to the weighing system. Junction box 4 2 2 4 Indicator : +Excitation 2: -Excitation : +Output 4: -Output 5: +Sense 6: -Sense 5 2 4 6 Junction box 5 6 2 4 Indicator The above figures represent the two basic configurations, using four- or six-wire ( sense ) load cells. With 4-wire extension cables the load cell output should be connected to pairs of diagonally opposite wires. Sometimes it is necessary to trim the output of each individual load cell to avoid corner load differences, which are caused by: # the parallel connection. Each load cell will be loaded with the resistance of the other load cells. As a result, the individual load cell output tolerances will be increased by the individual output resistance tolerance. # unequal load distribution. Trimming can be done by placing resistors or variable resistors into the excitation (Excitation trim) or output (signal trim) lines. All load cells should be placed on the same horizontal level. Check for mechanical unequal load conditions before trimming the load cells! REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page

EXCITATION TRIM Excitation trim is the oldest and still most used method of trimming each individual load cell. The output is trimmed by placing series resistance into the supply circuit to reduce the excitation voltage, resulting in the reduction in sensitivity of the individual load cells to equal the lowest one. The figure opposite shows a diagram of three excitation trimmed load cells. A temperature-independent variable resistor or potentio-meter of typically 0Ω is inserted in the + excitation lead of each load cell. There are two methods of trimming with excitation. The first method is to adjust the potentiometers by trial and error whilst moving calibration weights around from corner to corner. All potentiometers should be adjusted to have zero ohm resistance at the start of this procedure. Then once the lowest output corner is located, the other cells are trimmed to match this lowest output. This method can be very time consuming, particularly for high capacity scales, or vessel scales where the use of test weights at the corners is not practical. In these cases the second and best method is "pretrimming" the potentiometers with the use of a sensitive voltmeter ( at least 42 digit ). The following procedure can be used: ) Determine the exact mv/v rating of each load cell, which can be found on the load cell's calibration certificate. 2) Determine the exact excitation voltage supplied by the indicator by measuring this voltage with the voltmeter ( for example 0.05 V ). ) Multiply the lowest mv/v-value ( step ) by the excitation voltage ( step 2 ). 4) Divide the trimming-factor found in step by the mv/v-value for the remaining load cells. 5) Measure and adjust the excitation voltage of the three remaining load cells with each respective potentiometer. Verify the results and make the final adjustment while moving a testload from corner to corner. Output step step 4 step 5 # 2.995 2.995 * 0.05 Do not adjust #2.00 0.0 /.00 0.00 #.00 0.0 /.00 0.02 #4 2.998 0.0 / 2.998 0.040 WARNING: The reduction of sensitivity of one load cell will cause a change in zero of all load cells. Therefore adjust in small steps and always verify the results of each adjustment! REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page 2

SIGNAL TRIM The principle of signal trim is based on placing a relatively high parallel resistance (476220 KΩ) b- etween the two output leads of each load cell, to shunt a small part of its signal. Because of the low voltage and current values in the signal circuit, it takes a large resistance change to accomplish a small output change. The main benefit of signal trim is a neglectible interaction between span and zero and excellent temperature stability. The figure opposite shows a diagram of three signal trimmed load cells. To avoid an interaction problem between cells in a multiple load cell system, a series resistor (Rs) of typically 2500Ω is placed in each of the output leads. Conversely to excitation trim; the larger the shunt resistor (Rt), the more signal will be available. The following procedure should be followed for "pretrimming" of the potentiometers: ) Determine the exact mv/v rating of each load cell, which can be found on the load cell's calibration certificate. 2) Measure the output impedance of each load cell with a sensitive ohmmeter ( at least 42 digit ). ) Calculate the trimming-factor (TF) for each cell by dividing the lowest output value by each of the others. 4) Calculate the appropriate parallel resistance values (R) for each cell by dividing the load cell output impedance by (-TF). 5) Connect only the load cell output lines. 6) Remove the output leads from one of the load cell signal inputs and measure the resistance across those inputs with an ohmmeter, adjusting the trim potentiometer until the resistance equals the value of "R" calculated for that particular cell. When finished, replace the load cell output leads and repeat the process for each load cell except for the one with the lowest output, for which trim should be disabled. Output Output impedance step step 4 # 2.995 50. Ω Disable #2.00 50.4 Ω 2.995 /.00 75258 Ω #.00 50. Ω 2.995 /.00 49 Ω #4 2.998 50.5 Ω 2.995 / 2.998 50266 Ω Verify the results and make the final adjustment whilst moving a testload from corner to corner. WARNING: The resistors "Rs" should be very well matched (" 0.%) and should have a small temperature coefficient (#0 PPM), else the system will be non-linear! REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page

4-WIRE / 6-WIRE LOAD CELLS A load cell may have a cable with four or six wires. A six-wire cable, besides having +/- excitation and +/- signal lines also has + and - sense lines. It is a common misconception that the possibility to sense the actual voltage at the load cell is the only difference between 4-wire and 6-wire load cells. A load cell is compensated to perform within specifications over a certain temperature range ( usually -0 6 +40 EC ). Since cable resistance is a function of temperature, the cable response to temperaturechanges must be eliminated. The 4-wire cable is part of the temperature compensating system of the load cell. The load cell is calibrated and compensated with a certain amount of cable attached. Never cut a 4-wire load cell cable. The 6-wire cable is not part of the temperature compensating system of the load cell. The sense lines are connected to the sense terminals of the indicator, to feed back the actual voltage at the load cells. The indicator either adjusts its output voltage or adjusts its amplifier to compensate for any resistance change in the cable. The advantage of using this "active" system is the possibility to cut ( or extend ) the 6-wire load cell cable to any length. A 6-wire load cell will not perform within specifications if the sense lines are not used. TEMPERATURE EFFECTS ON EXTENSION CABLES In many cases the degraded performance of a weighing system caused by temperature effects on the extension cable is very small. However, for long cable runs or high accuracy applications, this can be a significant factor. The resistivity of any pure metal increases with its temperature. The figure opposite shows the effect of temperature on the resistivity of Zinc some metals. Over a limited range of temperature the graphs are approximately linear, hence a temperature coefficient of resistivity can be specified for the metal concerned. Aluminium If the resistance of a wire "R " at a temperature "T " is known then the resistance "R 2 " at temperature "T 2 " can be calculated by: R 2 =R (+α*(t 2 -T )) Copper where α is the temperature coefficient of resistivity ( K - ) 0 200 400 600 800 Temperature / K REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page 4

The resistance of a wire at temperature T can be measured with an accurate ohmmeter or calculated with: R =σ*l / A σ Resistivity at 20EC (Ωm), for copper,75.0-8 Ωm l Length (m) A Cross-section (m 2 ) The error to the system ( which is a typical span error ) caused by a temperature change of the extension cable can be calculated by: -((R ti +2R ) / (R ti +2R 2 ))*00% where R ti is the combined input resistance ( R i ) of multiple load cells ( N ) placed in parallel. R ti =R i / N The current flow through the signal ( and sense ) lines is neglectible, therefore a small change in resistance in these lines will not effect the system. The calculations are only applied on the excitation part of the circuit. For example: Four load cells type SSB with an input resistance of 50Ω are connected in parallel. An extension cable with a length of 50m and a cross-section of 0,75mm 2 is connected between the junction box and the indicator. The cable is made of copper with α = 4,.0 - K -. The span-error over a temperature change of 0EC ( 20 6-0 EC ) will be: ) R =,75.0-8 * (50/0,75.0-6 ) =,67Ω 2) R 2 =,67 * ( + 4,.0 - * (-0-20)) =.06Ω ) R ti = 50 / 4 = 87,5Ω 4) Error = ( - (87,5 + 2,4)/(87,5 + 2,02)) * 00% = -0,4% The figure below shows the correct diagram (for high accuracy or Weights and Measures applications) when 4-wire load cells are used in conjunction with an extension cable of considerably length. The sense lines are used to compensate for any change in resistance in the extension cable. Junction box 5 2 4 6 Extension Cable 5 6 2 4 Indicator : +Excitation 2: -Excitation : +Output 4: -Output 5: +Sense 6: -Sense REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page 5

GROUNDING AND SHIELDING Proper grounding and shielding can be critical to the successful application of load cells which are generating low level signals ( <5µV / scale division!). The basic rule: Avoid continuous ground loops; a system should not be grounded at multiple points. This may occur, for example, if the shield of the load cell cable is connected to earth at both ends. Revere Transducers load cell cables are provided with a braided shield which provides protection from electrostatic interference when properly used. This shield is floating ( not connected ) at the load cell avoiding the inadvertent creation of a "ground loop". A basic system configuration is represented in the figure below: Junction box Indicator The load cell case and junction box are grounded by mechanical attachment to the structure to which they are mounted. Usually this structure is grounded. The braided shield enclosing the load cell leads is grounded at the indicator, which is grounded through the power cord or housing. In order to avoid interference, load cell cables and extension cables should be kept away from power circuits, with a recommended distance of at least one meter. Power supply cables should be crossed at right angles. CABLE REQUIREMENTS The selection of a certain extension cable is frequently determined by the specific requirements of the installation. Not only the environmental conditions such as the presence of chemicals, sunlight, vibration etc. must be considered but also the cable's capacitance (µf). The cable's capacitance plays an important role with long extension cables and should be as low as possible! REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page 6

OUTPUT/INPUT CAULATIONS The total output signal of a multiple load cell system will be approximately the arithmetic mean value of the individual load cell outputs. The combined output can be calculated by: U o =((DL+AL) / (E max *N))*S*U e U o DL AL E max N S U e Output voltage (mv) Deadload or tareweight Applied load Load cell capacity Number of load cells Rated output (mv/v) Excitation voltage For example A scale is built with four load cells type SHBxM, capacity 200 kg and rated output 2 mv/v. The deadload or tare weight of the construction equals 75 kg and the maximum scale capacity equals 600 kg. With an excitation voltage of 5 V, the total load cell output at full load will be: Uo = ((75 + 600) / (200 * 4)) * 2 * 5 = 8,44 mv The total load cell output in a non-loaded situation will be: Uo = (75 / (200 * 4)) * 2 * 5 = 0,94 mv If the weighing range is divided in 000 divisions of 0,2 kg each, then the output for each division will be: (8,44-0,94) / 000 = 2,50 mv The maximum current flow in the excitation or input circuit can be calculated by dividing the excitation voltage by the combined resistance of load cells and extension cable ( when present ): I=U e / (R ti +2R ) ( see "Temperature effects on extension cables" to calculate R ti and R ). The above calculations can be used to verify the scale and indicator suitability. The resolution in microvolts per division is one of the most important features of a system which must be maximised for high accuracy applications ( Weights and Measures ). WELDING Avoid electric welding after installation of the load cells. If welding is necessary and the load cells can not be removed then disconnect each individual load cell cable from the junction box or measuring instrument. Place the clamp earthing electrode of the welding apparatus in the close proximity of the weld to avoid a current path through the load cells. Further, connect a flexible copper lead of at least 6mm 2 cross-section between the vessel and foundation over each load cell. REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page 7

CABLE JUNCTION BOXES A junction box should be protected to at least IP54 ( DIN.40.050 ), but for outdoor use we strongly recommend a junction box protected to IP65. The degree of protection can only be achieved if the junction box is installed correctly: # Select the right location based on the environmental conditions, NOT on the ease of installation. # Put on the cover-screws according to the manufacturer's specifications. # Place the cable-entries downwards and provide a dripping-loop in the cable. # Avoid moisture during the installation, also ensure that no moisture enters the load cell cable before and during installation. Moisture could decrease the insulation resistance of the circuit and might cause unstable readings. A bag with drying agent ( silicagel ) should be enclosed to absorb condensation. This bag should never make contact with any not-insulated wiring in the junction box. For Weights and Measures applications the junction box must be provided with a sealing capability, for example two sealing screws with a wire and lead seal. : Glass fibre reinforced polyester junction box. 2: Stainless steel junction box. : Correct installation, with cable entries placed downwards and dripping loops. Customer support: The Revere Transducers group combines fifty years of load cell manufacturing with fifty years of application know how. For any further information, please contact our manufacturing operation or any one of our regional sales offices. Revere Transducers Europe Ramshoorn 7 Postbus 6909, 4802 HX Breda The Netherlands Tel. (+)76-5480700 Fax. (+)76-542854 Regional offices in Germany, France and United Kingdom REVERE TRANSDUCERS APPLICATION NOTE 0/4-02/0 Page 8