Hydrogen Peroxide Detection by Ion Chromatography and Electrochemical Detection

Similar documents
What is Eluent Generation?

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Optimize Your Process Operations by Improving Process Monitoring

A Complete Solution for Method Linearity in HPLC and UHPLC

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

Determination of Calcium and Magnesium in Brine

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Environmental Analysis Using the ICS-3000 Ion Chromatography System

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

How To Control A Record System

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Dionex IonPac AS19-4µm

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities

columns IonPac CS18 Cation-Exchange Column

Thermo Scientific ConFlo IV Universal Interface

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Carbohydrate in Coffee: AOAC Method vs a New Fast Ion Chromatography Method

Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Earning Customer Loyalty

Anion chromatography using on-line recycled eluents

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

Thermo Scientific Barnstead TII Pure Water Systems. Pure water. Pure research. Total confidence.

Dissolved and precipitated oxalate

Automation of Sample Preparation in Ion Chromatography

Foreign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund

ProteinChip Energy Absorbing Molecules (EAM)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

FT-NIR for Online Analysis in Polyol Production

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Thermo Scientific Dionex AS-AP, AS-DV, and AS-HV Autosamplers. Automation, sample preparation. Autosamplers for Ion Chromatography

INTRODUCTION. The principles of which are to:

Gas Chromatography Liner Selection Guide

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

[ Care and Use Manual ]

Determination of Benzenesulfonic Acid Counterion in Amlodipine Besylate by Ion Chromatography

Determination of Caffeine in Beverages HPLC-1

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Application Note USD3013. Monoclonal Antibody Production in Suspension CHO Culture Using Single-Use PadReactor Mini Bioreactor

Thermo Scientific SIEVE Software for Differential Expression Analysis

Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795

Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA

Report on Government Information Requests

Thermo Scientific ExactFinder Software

Quality Assurance for the Analysis of Steel by Gas Component Analysis

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

QIAsymphony RGQ Protocol Sheet

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

LIFECYCLE. Thermo Scientific. Enterprise Solutions

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Keysight Technologies Using Fine Resolution to Improve Thermal Images. Application Note

Dionex ICS-1100 Ion Chromatography System

WATERS QUANTITATIVE ANALYSIS solutions

[ Care and Use Manual ]

Global Economic Briefing: Global Inflation

Thermo Scientific Qtegra Intelligent Scientific Data Solution Software. Delivering quality. Driving productivity

CNE Progress Chart (CNE Certification Requirements and Test Numbers) (updated 18 October 2000)

Thermo Scientific Syncronis HPLC Columns. Technical Manual

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

HAS BRAZIL REALLY TAKEN OFF? BRAZIL LONG-RUN ECONOMIC GROWTH AND CONVERGENCE

Accuracy counts! SENSORS WITH ANALOG OUTPUT

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Pall Aria Integrated MF/RO Systems for Cost-Effective Treatment of Sea and Brackish Waters

Thermo Scientific Dionex ICS-1100 Ion Chromatography System

Report on Government Information Requests

Report on Government Information Requests

How many students study abroad and where do they go?

OCTOBER Russell-Parametric Cross-Sectional Volatility (CrossVol ) Indexes Construction and Methodology

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Get the benefits of Norgren s unique range of Online services

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

AN886 APPLICATION NOTE

Agilent Television Power Consumption Testing. Application Note

Research for leachables of nuclear grade cation exchange resin

Determination of Trace Sodium in Diesel and Biodiesel Fuel

[NUGENESIS SAMPLE MANAGEMENT ] AMPLE IMPROVING LAB EFFICIENCY, ANAGEMENT ACCELERATING BUSINESS DECISIONS. bigstock.com $69

Image Lab Software for the GS-900 Densitometer

Thermo Scientific Capitol Vial Drugs of Abuse Products. The tools to create and maintain. a safe and drug free environment

Transcription:

Hydrogen Peroxide Detection by Ion Chromatography and Electrochemical Detection Sheetal Bhardwaj, Rong Lin, Kannan Srinivasan, and Christopher Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Overview Purpose: To develop a new ion chromatography (IC) method using the Thermo Scientific Dionex CarboPac PA2 Carbohydrate Column, potassium hydroxide eluent, and electrochemical detection (ECD) for analysis of hydrogen peroxide. Methods: A Reagent-Free IC (RFIC ) system and direct injection without derivatization of the samples was used in this application. Results: Analysis of hydrogen peroxide from a variety of samples was successfully demonstrated. Introduction Hydrogen peroxide is a clear and colorless liquid with a slight pungent odor. It is a strong oxidizing agent and can be a reducing agent for strong oxidizers. Because the decomposition products of hydrogen peroxide are water and oxygen, hydrogen peroxide is preferred for many applications. When hydrogen peroxide is used as a disinfectant in water purification and wastewater treatment, it decomposes in the presence of heat or light. Therefore, there is a need to carefully monitor various water samples for hydrogen peroxide. The presence of hydrogen peroxide in air degrades air quality. Measurement of atmospheric hydrogen peroxide also requires analytical methods to monitor its fate. Beverage samples, such as cola drinks, are also known to contain hydrogen peroxide and need to be monitored. Toothpastes and whitening kits also contain hydrogen peroxide and need to be analyzed. Various chromatographic methods have been adapted for hydrogen peroxide analysis. An indirect method for detecting hydrogen peroxide was pursued using IC and conductivity detection. This approach involved addition of bisulfite to the sample prior to analysis, and exploited a reaction between hydrogen peroxide and sulfite, detecting the resulting sulfate as a means of detecting hydrogen peroxide. Another method exploited an enzymatic reaction of hydrogen peroxide with catalase methanol to form formaldehyde, with subsequent derivatization of this species to achieve fluorescence detection. 2 Another publication reported two methods for analyzing hydrogen peroxide. 3 The first was based on a derivatization reaction with fluorescence detection. The second used a direct ECD method with a sodium acetate eluent which was adjusted to ph.5. Both methods exhibited excellent performance. The ECD method was simpler since it was a direct injection of the sample. Methods Conditions Columns: Eluent: Dionex CarboPac PA2 Guard (3 3 mm) Dionex CarboPac PA2 Analytical (3 5 mm) 5 mm KOH Eluent Source: Thermo Scientific Dionex EGC-KOH Eluent Generator Cartridge with Thermo Scientific Dionex CR-ATC 5 Continuously Regenerated Anion Trap Column Flow Rate:.5 ml/min Temperature: 3 C Detection: Pulsed amperometric detection, disposable gold electrode ECD is a well established method of choice for carbohydrate and amino acid detection. The disposable gold electrode provides reproducible analysis while maintaining good peak shapes and excellent sensitivity. Time-consuming, labor-intensive electrode polishing steps are eliminated. Here, the disposable gold electrode was used for the detection of hydrogen peroxide. Carbohydrate Waveform for the ECD: Time (s) Potential (V). +..2 +..4 +..4-2..42-2..43 +.6.44 -..5 -. 2 Hydrogen Peroxide Detection by Ion Chromatography and Electrochemical Detection

Sample Preparation In this work, hydrogen peroxide analysis was pursued using standards without any sample pretreatment. Analysis of samples containing hydrogen peroxide, such as toothpaste and mouthwash, was pursued with the following sample preparation procedure. Dilute approximately gm of the sample (toothpaste or mouthwash) into gm of deionized water. Place the sample on a vortex to fully disperse and then filter using a.45 µm Supor (PES) Membrane Disc Filter. Results Analysis of hydrogen peroxide was pursued with minimal sample preparation and without any derivatization reactions. The detection used a gold electrode and ECD. Figure 2 shows the detection of a 2 ppm hydrogen peroxide standard. Excellent peak shape and sensitivity was observed using the present method. A response versus concentration study was pursued using the present method in two ranges: a high range (. mg/l), and a low range (. mg/l). Figure 3 shows the response versus concentration for the high range, demonstrating that the data fit to a quadratic curve. Figure 4 shows the response versus concentration for the low range, demonstrating that the data was linear. Analysis of unknown samples was pursued using the low range. The RFIC system used in this work provided significant ease of use, automating eluent generation and reliably providing consistent eluent. The disposable electrodes facilitated reproducible analysis that was fast and complete within 5 minutes. Overall, the present method was simple and easy to implement. FIGURE. Thermo Scientific Dionex ED Electrochemical Detector Cell configuration. FIGURE 2. Analysis of a 2 mg/l standard of hydrogen peroxide. 5 375 25 25-25 -25-375 -5 2 3 4 5 6 7 8 9 Thermo Scientific Poster Note PN7967_PITTCON_24_E_3/4S 3

Analysis of hydrogen peroxide was demonstrated (above) under isocratic conditions with good peak shape. The analysis was complete in 5 minutes. FIGURE 3. Response versus concentration curve for the high range of hydrogen peroxide sample (. mg/l). Concentration FIGURE 4. Response versus concentration curve for the low range of hydrogen peroxide sample (. mg/l). 6 4 2 8 6 4 2 2 4 6 8 Concentration 4 Hydrogen Peroxide Detection by Ion Chromatography and Electrochemical Detection

FIGURE 5. Analysis of a commercial toothpaste sample containing hydrogen peroxide. 8 6 4 2-2 -4-5 2 3 4 5 6 7 8 9 Excellent peak shape for the hydrogen peroxide can be inferred from the above chromatogram. There were no other interfering components in this toothpaste sample using the present method. It is clear that the run time can be adjusted to 5 minutes. FIGURE 6. Analysis of a commercial mouthwash sample containing hydrogen peroxide. 5 375 25 25-25 -25-375 -5 2 3 4 5 6 7 8 9 Excellent peak shape for the hydrogen peroxide can be inferred from the above chromatogram. There were no other interfering components in this mouthwash sample using the present method. It is clear that the run time can be adjusted to 5 minutes. Thermo Scientific Poster Note PN7967_PITTCON_24_E_3/4S 5

FIGURE 7. Comparison of a sample containing a 2 ppm hydrogen peroxide standard versus a toothpaste sample containing hydrogen peroxide and one that did not contain hydrogen peroxide. 5 2 ppm Hydrogen Peroxide Sample -5 5 Toothpaste Sample Containing Hydrogen Peroxide 25 2 Toothpaste Sample Containing No Hydrogen Peroxide -5 2 3 4 5 6 7 8 9 The above figure shows a comparison of a standard containing 2 mg/l of hydrogen peroxide with a toothpaste sample containing hydrogen peroxide. A control run was also pursued with a toothpaste sample that did not contain hydrogen peroxide and no peaks for hydrogen peroxide were detected as shown. Conclusion A simple, reliable method for hydrogen peroxide analysis was demonstrated in this work using an RFIC system. It was a direct injection method with minimal sample pretreatment. The combination of an RFIC system with ECD with disposable gold electrodes made the overall method fully automated, easy to use, and reliable. The Dionex IonPac PA2 column provided excellent separation of hydrogen peroxide and fast analysis with a run time of less than 5 minutes, as demonstrated in this work. References. Jenke, D. R. Indirect Determination of Hydrogen Peroxide by Ion Chromatography. J. Chromatogr. Sci. 986, 24 (8), 352 355. 2. Hamano, T.; Mitsuhashi, Y.; Yamamoto, S. Determination of Hydrogen Peroxide in Beverages by High-Performance Liquid Chromatography with Fluorescence Detection. J. Chromatogr. 987, 4, 423 429. 3. Tarvin, M.; McCord, B.; Mount, K.; Sherlach, K.; Miller, M. L. Optimization of Two Methods for the Analysis of Hydrogen Peroxide: High Performance Liquid Chromatography with Fluorescence Detection and High Performance Liquid Chromatography with Electrochemical Detection in Direct Current Mode. J. Chromatogr., A. 2, 27, 7564 7572. 6 Hydrogen Peroxide Detection by Ion Chromatography and Electrochemical Detection

www.thermoscientific.com 24 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Supor is a registered trademark of Pall Corporation. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 333 5 34 Australia +6 3 9757 43 Austria +43 8 282 26 Belgium +32 53 73 42 4 Brazil +55 373 54 Canada + 8 53 8447 China 8 8 58 (free call domestic) 4 65 58 Denmark +45 7 23 62 6 Europe-Other +43 333 5 34 Finland +358 9 329 2 France +33 6 92 48 Germany +49 63 48 4 India +9 22 6742 9494 Italy +39 2 95 59 Japan +8 6 6885 23 Korea +82 2 342 86 Latin America + 56 688 87 Middle East +43 333 5 34 Netherlands +3 76 579 55 55 New Zealand +64 9 98 67 Norway +46 8 556 468 Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9:28 Certified. Russia/CIS +43 333 5 34 Singapore +65 6289 9 Sweden +46 8 556 468 Switzerland +4 6 76 77 Taiwan +886 2 875 6655 UK/Ireland +44 442 233555 USA + 8 532 4752 PN7967_E 3/4S