Guidance on Notified Body s Tasks of Technical Documentation Assessment on a Representative Basis



Similar documents
Medical Device Software Do You Understand How Software is Regulated?

CE-Marked Product List (MDD)

Reporting Changes to your Notified Body

Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System

Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

GUIDELINES ON MEDICAL DEVICES

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

GUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA : A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

Konformitätsbewertung 3.9 B 17. Guidance for Notified Bodies auditing suppliers to medical device manufacturers

Registration of Class B Medical Devices

PLM: Privat Label Manufacturer (Customer of the OEM-PLM relation) OEM: Original Equipment Manufacturer (Supplier of the OEM-PLM relation)

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

ALL Medical Devices regardless of Classification

Conformity assessment procedures of breast implants

Want to know more about the Notified Body?

Basic Information about the European Directive 93/42/EEC on Medical Devices

GUIDELINES ON MEDICAL DEVICES

NBOG s Best Practice Guide

GUIDELINES ON MEDICAL DEVICES

Guidelines for conformity assessment of In Vitro Fertilisation (IVF) and Assisted Reproduction Technologies (ART) products

Medidée Services SA. Nano-Tera.ch. 05 February 2015 part 12. Special Issues Q&A Session. Michael Maier

How To Know If A Mobile App Is A Medical Device

Conformity assessment certification

(OJ L 169, , p. 1)

GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

Medical Device Training Program 2015

Manual for ITC's clients, Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC

Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH

BSI Unannounced Audits

CE Marking: Your Key to Entering the European Market

ANNEX F. Organic agriculture (Art. 11) ARTICLE 1. Objectives

Emerging Device Topics for Regulatory Consideration.. Janine Jamieson May 2015

BSI Road Show: September 8 th to 15 th, 2014

Med-Info. Malaysia Medical Device Regulations. TÜV SÜD Product Service GmbH. International expert information for the medical device industry

Frequently Asked Questions. Unannounced audits for manufacturers of CE-marked medical devices. 720 DM a Rev /10/02

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC

Clinical evaluation Latest development in expectations EU and USA

Medical Devices: CE Marking Step-by Step

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device

Medical Device Software Standards for Safety and Regulatory Compliance

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)

ISO 13485:201x What is in the new standard?

Safeguarding public health The Regulation of Software as a Medical Device

QMS for Software as a Medical Device [SaMD] Lessons Learned from a Quality Perspective

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002

Medical Device Classification

ETSI SR V1.1.2 ( )

Navigating the New EU Rules for Medical Device Software

RAPS ONLINE UNIVERSITY

End of consultation (deadline for comments) 14 October Adoption by Committee for advanced therapies 15 October 2010

Risk Management and the Impact of EN ISO 14971:2012 Annex Z

Software within the medical device regulatory framework in the EU

CONSOLIDATED VERSION IEC Medical device software Software life cycle processes. colour inside. Edition

DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL

Global Regulatory Requirements for Medical Devices

TEXTUAL PROPOSAL TECHNICAL BARRIERS TO TRADE (TBT) Article 1 Objective and Scope

Guidance on the preparation of an Annex XV dossier for the identification of substances of very high concern

Guidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC. GNB-CPD position paper from SG17 - EN :2009

EA Document on. Accreditation. For Notification Purposes

Application of ISO/IEC for the Accreditation of Food Safety Management Systems (FSMS) Certification Bodies

NOTICE TO MANUFACTURERS OF RADIOTHERAPY MEDICAL DEVICES

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994

An introduction to the regulation of apps and wearables as medical devices

(ACCORDINGLY TO ARTICLE 11º OF DECREE-LAW N.º 145/2009) INTRODUCTION page 2. ACESS TO THE SYSTEM page 3. NOTIFICATION MANAGEMENT page 4

Processes for the Development of Healthcare Applications. Christian Johner

The Medical Products Agency s Working Group on Medical Information Systems. Project summary

FINAL DOCUMENT. Guidelines for Regulatory Auditing of Quality Management Systems of Medical Device Manufacturers Part 1: General Requirements

Network Certification Body

Overview of Key Obligations Under Regulation (EC) No. 1272/2008 on the Classification, Labelling and Packaging of Substances and Mixtures (CLP)

What is ATEX? The European Regulatory Framework for Manufacture, Installation and Use of Equipment in Explosive Atmospheres

DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL

TG TRANSITIONAL GUIDELINES FOR ISO/IEC :2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES

xxxxx Conformity assessment Requirements for third party certification auditing of environmental management systems - competence requirements

Information supplied by the manufacturer with medical devices

ETSI EN V2.2.2 ( )

PUBLICATIONS. Introduction

Scheme rules for thermal insulation products for building equipment and industrial installations - VDI certification scheme

ISO 9001: 2008 Boosting quality to differentiate yourself from the competition. xxxx November 2008

A Revised EMC Directive from Europe

DIRECTIVE 2006/95/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL. of 12 December 2006

Official Journal of the European Union. (Acts whose publication is obligatory)

COMMISSION REGULATION (EU)

Medical Devices. IRELAND Arthur Cox

Regulatory Strategy for Drug-Device Combinations By John Worroll

Raad voor Accreditatie (Dutch Accreditation Council RvA) Assessment of Conformity Assessment Schemes

PROPOSED DOCUMENT. Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change

DISCUSSION NON-PAPER. How to put ideas for cooperation under TTIP into practice a few examples

CHECKLIST ISO/IEC 17021:2011 Conformity Assessment Requirements for Bodies Providing Audit and Certification of Management Systems

Guidelines. of

JUDGMENT OF THE COURT 28 September 2015

NOTICE. Re: GD210: ISO 13485:2003 Quality Management System Audits Performed by Health Canada Recognized Registrars

RECOMMENDATION 2013/473/EU ON NOTIFIED BODIES AUDITS

Transcription:

NBOG s Best Practice Guide applicable for AIMDD, MDD, and IVDD 2009-4 Guidance on Notified Body s Tasks of Technical Documentation Assessment on a Representative Basis 1 Introduction The Directive 93/42/EEC concerning medical devices (MDD) contains possible conformity assessment procedures in order to CE mark devices. Up to now, there has been inconsistency in the way Notified Bodies (NB) have performed the assessment of technical documentations for Class lla and Class llb products following quality system conformity assessment routes. The Directive 2007/47/EC, covering the revision of the MDD, includes additional requirements for review of technical documentation for all Class IIa and Class IIb medical devices covered by quality system assessment routes. 2 Scope This guideline has been prepared for NBs on how to assess the technical documentation on a representative basis according to the Directive 2007/47/EC. Attention was given to the sampling size and depth of the assessment. It will also be useful for Designating Authorities in monitoring the performance of their NBs in this area. 3 Definitions 3.1 Device subcategory (Directive 2007/47/EC, Art. 2, Paragraph 1 (iii), point (l) and Directive 93/42 EEC Art 1 Paragraph 2 (l)) A set of devices having common areas of intended use or common technology. 3.2 Generic device group (Directive 2007/47/EC, Art. 2, Paragraph 1 (iii), point (m) and Directive 93/42 Art. 1 Paragraph 2 (m)) A set of devices having the same or similar intended uses or commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics. 4 Directive Requirements 4.1 Class IIa devices For Class IIa devices the requirements are covered in Directive 2007/47/EC at Annex ll, Paragraph 2 (i), Paragraph 5 (f), and Paragraph 6 (e). The Directive 93/42/EEC is amended and Annex II 7.2, Annex V 6.2 and Annex VI 6.2, states that For devices in Class IIa the notified body shall assess... the technical documentation for at least one representative sample for each device subcategory for compliance with the provisions of this Directive. NBOG BPG 2009-4 Page 1 of 5

4.2 Class IIb devices For Class IIb devices the requirements are covered in Directive 2007/47/EC at Annex ll, Paragraph 2 (i). The Directive 93/42/EEC is amended and Annex II 7.3 states that For devices in Class IIb the notified body shall assess, as part of the assessment in Section 3.3, the technical documentation as described in Section 3.2 (c) for at least one representative sample for each generic device group for compliance with the provisions of this Directive. 4.3 Concept of the reviews Directive 2007/47/EC, Recital (22) introduces the concept of the reviews of Class lla and llb devices: The depth and extent of this review should be commensurate with the classification of the device, the novelty of the intended treatment, the degree of intervention, the novelty of the technology or construction materials, and the complexity of the design and/or technology. This review can be achieved by taking a representative example of design documentation of one or more type(s) of devices from those being manufactured. Further review(s), and in particular the assessment of changes to the design that could affect conformity with the essential requirements, should be part of the surveillance activities of the notified body. Directive 93/42/EEC is amended and in Annex II 7.4, Annex V 6.3 and Annex VI 6.3, there are rationales provided for the sampling: In choosing representative sample(s) the notified body shall take into account the novelty of the technology, similarities in design, technology, manufacturing and sterilisation methods, the intended use and the results of any previous relevant assessments (e.g. with regard to physical, chemical or biological properties) that have been carried out in accordance with this Directive. The notified body shall document and keep available to the competent authority its rationale for the sample(s) taken. 4.4 Requirements for surveillance assessments Directive 2007/47/EC Annex ll, Paragraph 2 (i), Paragraph 5 (f), Paragraph 6 (e) and Paragraph 6.4 introduces requirements for surveillance assessments. The Directive 93/42/EEC is amended and in Annex II 7.5, Annex V 6.4 and Annex VI 6.4 it is stated: Further samples shall be assessed by the notified body as part of the surveillance assessment.... 5 Timing Directive 2007/47/EC Article 4 states Member States shall apply those measures from 21 March 2010. Following the adoption of Directive 2007/47/EC, from 21 March 2010 at the latest, for Class IIb devices following the Annex II assessment route and Class IIa devices following the Annex II, V or VI assessment route there is a specific requirement for technical documentation to be reviewed by the NB on a sample basis. At any rate and at the latest, the first surveillance following 21 March 2010 will need to address this element. After the 21 March 2010 new certificates can be issued only according to the new requirements. For existing certificates issued prior to this date please see Interpretative document of the Commission's services Implementation of Directive 2007/47/EC amending Directives 90/385/EEC, 93/42/EEC and 98/8/EC [1] for details. In this document the Commission s services call upon the member states to require Notified Bodies to ensure that surveillance audits are performed at least once a year as specified in the standard EN ISO /IEC 17021 [2]. NBOG BPG 2009-4 Page 2 of 5

6 Depth of Assessment The assessment of each set of technical documentation selected, as a minimum, should consist of a review of the intended use including confirmation that the product is a medical device and it s correct classification, the validity of the essential requirements checklist, especially when harmonized standards have not been applied in full, risk management file, pre-clinical data (studies in animal models, biocompatibility, technical performance tests etc.), clinical evaluation in accordance with 93/42/EEC Annex X (Note the NB review should be as per MEDDEV 2.7.1 [3]), information supplied by the manufacturer (label, instructions for use), declaration of conformity or the draft thereof, other technical documentation based on risk, and should be done according to the respective parts of NBOG BPG 2009-1 [4]. Consequently, there will be one part of the assessment performed within the normal course of the audit but in addition, file reviews will be necessary. The time spent on the review and the level of expertise used should be proportionate to the risk and complexity of the devices in question. For the more complex devices this may require a detailed technical review. If such an in depth assessment is not considered necessary then a justification should be available. Depending on the risk and complexity of the device(s), different experts may be involved. Therefore, NBs may not be able to perform the complete assessment on site. 7 Reporting For those elements covered the report produced for each technical documentation reviewed should be in line with the NBOG guidance on Design Dossier Report Content [4]. This could be either as part of the audit report or separately as a supplement. Nonconformities/Corrective Action Requests should be raised in the normal manner, including those raised against technical documentation under conformity assessment procedures according to Annex V or Annex VI. 8 Sampling For initial audits of devices in Class IIb the Directive states that the NB shall assess the technical documentation for at least one representative sample for each generic device group (i.e. a set of devices having the same or similar intended uses or commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics). This is the same description as for generic device groups in EN ISO 15225 [5], which specifies the GMDN system. For initial audits of devices in Class IIa the Directive states that the NB shall assess the technical documentation for at least one representative sample for each device subcategory (i.e. a set of devices having common areas of intended use or common technology). Since no similar description exists for subcategory/ies like for generic device groups, the grouped collective NBOG BPG 2009-4 Page 3 of 5

terms described in NBOG BPG 2009-3 [6], which define scope expressions for NBs, could be used for this. However, in practice it is expected that the NB identifies the complete range of Class IIa and IIb products manufactured and define a suitable sampling plan for initial audits. This should be done based on the numbers of generic device groups and grouped collective terms. It is considered that if large numbers of samples are required to be taken then the following plan should be used: For Class IIa devices: a sample of each group of scope expressions according to NBOG BPG 2009-3 [6]. For Class IIb devices: a sample of generic device groups, referenced by different GMDN preferred terms, to the following plan: Up to 2 groups: a sample from each group Up to 10 groups: a sample from 3 of these groups Up to 20 groups: a sample from 5 of these groups Up to 30 groups: from 7 groups a sample N > 30 groups: from N/10 + 5 groups a sample For surveillance audits the NB should identify the complete range of Class IIa and IIb devices and group them according to preferred terms and grouped collective terms (scope expressions). For each certification period, and at least every 3 years, the NB should identify the technical documentations to be reviewed in line with the above sampling plan. As long as there is still a technical documentation not yet reviewed, at least one technical documentation should be sampled each year. This activity may or may not be done in conjunction with routine surveillance assessments. During subsequent certification periods appropriate plans should be drawn up to ensure that in the long run the majority, if not all, of the technical documentations will be assessed. 9 Changes/New Devices Manufacturers introducing a new subcategory of Class IIa device or a new generic Class IIb device need to inform their NB who need to review relevant technical documentation prior to the products being placed on the market. For changes to existing designs the normal change requirements apply. Reference Directive 93/42/EEC Annex II, V and VI Directive 2007/47/EC Sources [1] Interpretative document of the Commission's services: Implementation of Directive 2007/47/EC amending Directives 90/385/EEC, 93/42/EEC and 98/8/EC, 5 June 2009 [2] EN ISO/IEC 17021 : 2006 Conformity assessment Requirements for bodies providing audit and certification of management systems [3] MEDDEV 2.7/1 Evaluation of clinical data a guide for manufacturers and notified bodies NBOG BPG 2009-4 Page 4 of 5

[4] NBOG BPG 2009-1 Guidance on Design-Dossier Examination and Report Content [5] EN ISO 15225 : 2000/A1 : 2004 Nomenclature Specification for a nomenclature system for medical devices for the purpose of regulatory data exchange [6] NBOG BPG 2009-3 NBOG Guidance for Designation Authorities to Define the Notification Scope of a Notified Body Conducting Medical Devices Assessments Keywords conformity assessment, device subcategory, generic device group, sampling basis, technical documentation Date of issue July 2009 NBOG BPG 2009-4 Page 5 of 5