Two-Column Volar Distal Radius Plate. For fragment-specific fracture fixation with variable angle locking technology.

Similar documents
Technique Guide. 2.4 mm LCP Distal Radius System. A comprehensive plating system to address a variety of fracture patterns.

Fractures around wrist

Distal Radius Fractures. Lee W Hash, MD Affinity Orthopedics and Sports Medicine

VariAx Distal Radius Locking Plate System. Anatomical & Universal Volar Plates Dorsal Plates Fragment Specific Plates

Wrist Fractures. Wrist Defined: Carpal Bones Distal Radius Distal Ulna

How To Fix A Radial Head Plate

PERIPROSTHETIC IMPLANTS

Zimmer Periarticular Elbow Locking Plate System

Wrist and Hand. Patient Information Guide to Bone Fracture, Bone Reconstruction and Bone Fusion: Fractures of the Wrist and Hand: Carpal bones

INJURIES OF THE HAND AND WRIST By Derya Dincer, M.D.

Citation International Orthopaedics, 2011, v. 35 n. 3, p Creative Commons: Attribution 3.0 Hong Kong License

Zimmer Periarticular Proximal Tibial Locking Plate

TriLock 1.5 Implants for the Phalanges

Malleolar fractures Anna Ekman, Lena Brauer

Volar Fixation for Dorsally Displaced Fractures of the Distal Radius: A Preliminary Report

TwinFix Cannulated Compression Screw

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Abdominal Pedicle Flaps To The Hand And Forearm John C. Kelleher M.D., F.A.C.S.

.org. Distal Radius Fracture (Broken Wrist) Description. Cause

Technique Guide. 2.7 mm/3.5 mm Variable Angle LCP Ankle Trauma System. Part of the Synthes variable angle locking compression plate (VA-LCP) system.

Zimmer Small Fragment Universal Locking System. Surgical Technique

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery.

Volar Plate Fixation for the Treatment of Distal Radius Fractures: Analysis of Adverse Events

RADIOGRAPHIC EVALUATION

NCB Distal Femur System. Surgical Technique

Rigid Internal Fixation of Displaced Distal Radius Fractures

EXTENSOR POLLICIS TENDONITIS SYNDROME

KnifeLight. Carpal Tunnel Ligament Release. Operative Technique

Radial Head Fracture Repair and Rehabilitation

LCP Compact Foot/Compact Hand.

THE WRIST. At a glance. 1. Introduction

ASSOCIATED LESIONS COMPLICATIONS OSTEOARTICULAR COMPLICATIONS

Most active and intricate part of the upper extremity Especially vulnerable to injury Do not respond well to serious trauma. Magee, pg.

Common wrist injuries in sport. Chris Milne Sports Physician Hamilton,NZ

Technique Guide. DHS/DCS Dynamic Hip and Condylar Screw System. Designed to provide stable internal fixation.

WRIST EXAMINATION. Look. Feel. Move. Special Tests

Arthroscopy of the Hand and Wrist

The 10 Most Common Hand Pathologies In Adults. 1. Carpal Tunnel and Cubital Tunnel

Systemic condition affecting synovial tissue Hypertrohied synovium destroys. Synovectomy. Tenosynovectomy Tendon Surgery Arthroplasty Arthrodesis

ASOP Exams PO Box 7440 Seminole, FL The Manual of Fracture Casting & Bracing Exam 80% Passing ID # Name Title. Address. City State Zip.

Technique Guide. 3.5 mm and 4.5 mm Curved Locking Compression Plates (LCP). For minimally invasive osteosynthesis.

Technique Guide. Large Fragment LCP Instrument and Implant Set. Part of the Synthes locking compression plate (LCP) system.

Upper Limb QUESTIONS UPPER LIMB: QUESTIONS

Chapter 30. Rotational deformity Buddy taping Reduction of metacarpal fracture

LCP Olecranon Plate. The anatomical fixation system with angular stability for olecranon and proximal ulnar fractures.

Achilles Tendon Repair, Operative Technique

Fracture Care Coding September 28, 2011

Mini TightRope CMC Surgical Technique

OptiLock Periarticular Plating System For Proximal Tibial Fractures. Pre-Launch Surgical Technique

Chapter 7 The Wrist and Hand Joints

Ulnar sided Wrist Pain

Adult Forearm Fractures

Examination of the Elbow. Elbow Examination. Structures to Examine. Active Range of Motion. Active Range of Motion 8/22/2012

PATIENT SPECIFIC PLATES FOR MANDIBLE: CUSTOM MADE PLATES FOR TRAUMA AND RECONSTRUCTION

Wrist Fractures: What the Clinician Wants to Know 1

The wrist and hand are constructed of a series of complex, delicately balanced joints whose function is essential to almost every act of daily living.

Technique Guide. 4.5 mm LCP Proximal Femur Plates. Part of the Synthes Periarticular LCP Plating System.

V-TEK IVP System 2.7 System 4.0

Cable System. For Orthopaedic Trauma Surgery.

Scaphoid Non-union. Dr. Mandel Dr. Gyomorey. May 3 rd 2006

Technique Guide. Orthopaedic Cable System. Cerclage solutions for general surgery.


DIAGNOSING SCAPHOID FRACTURES. Anthony Hewitt

Wrist and Hand Injuries Keep Your Edge: Hockey Sports Medicine 2015 Toronto, Canada August 28-30

Pediatric Sports Injuries of the Wrist and Hand. Sunni Alford, OTR/L,CHT Preferred Physical Therapy

Semmelweis University Department of Traumatology Dr. Gál Tamás

Treatment of Distal Radius Fractures With a Nonbridging Cross-Pin Fixator (The CPX System)

NOW PLAYING THE WRIST. David Costa, OTR/L October 20, 2007

Elbow & Forearm H O W V I T A L I S T H E E L B O W T O O U R D A I L Y L I V E S?

ANTERIOR CERVICAL DISCECTOMY AND FUSION. Basic Anatomical Landmarks: Anterior Cervical Spine

Titanium Wire with Barb and Needle. For canthal tendon procedures.

Technique Guide. Screw Removal Set. Instruments for removing Synthes screws.

MINI FRAGMENT SYSTEM. Instruments and implants for 1.5 mm, 2.0 mm, and 2.4 mm plate fixation PRODUCT OVERVIEW

The Hand Exam: Tips and Tricks

3. Be able to perform a detailed clinical examination of the forearm and wrist.

Posttraumatic medial ankle instability

LISS PLT. The less invasive stabilization system for proximal tibia fractures.

Distal Radius Fractures

Scaphoid Fracture of the Wrist

The Emergent Evaluation and Treatment of Hand and Wrist Injuries

Hand and Upper Extremity Injuries in Outdoor Activities. John A. Schneider, M.D.

Surgical technique. End Cap for TEN. For axial stabilization and simultaneous protection of soft tissue.

Musculoskeletal Trauma of the Wrist

Total Elbow Arthroplasty and Rehabilitation

Injuries to Upper Limb

Calcaneus (Heel Bone) Fractures

Coding for Difficult Tendon Repairs and Transfers. Current Procedural Terminology 2013 American Medical Association. All Rights Reserved.

9 DISTAL RADIUS AND ULNA FRACTURES

Hand and Wrist Injuries and Conditions

Wrist Fractures. Wrist Injuries/Pain. Upper Extremity Care in an Aging Population. Objectives. Jon J. Cherney, M.D. Fractures of the Distal Radius

Median Nerve Injuries in, Fractures in the Region of the Wrist

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail

Elbow, Forearm, Wrist, & Hand. Bony Anatomy. Objectives. Bones. Bones. Bones

Anatomic Percutaneous Ankle Reconstruction of Lateral Ligaments (A Percutaneous Anti ROLL)

Knotilus TM. Anchor Instability Repair. Technique Guide

NCB Proximal Humerus System. Surgical Technique

MatrixNEURO CMF Matrix

Transcription:

2.4 mm Variable Angle LCP Two-Column Volar Distal Radius Plate. For fragment-specific fracture fixation with variable angle locking technology. Tips and Tricks This material was created in cooperation with Nemocnice České Budějovice Instruments and implants approved by the AO Foundation

2.4 mm Variable Angle LCP Two-Column Volar Distal Radius Plate. For fragment-specific fracture fixation with variable angle locking technology. Features and Benefits Anatomic fit close to the volar ridge with highly polished smooth edges for minimized soft tissue irritation Two columns for increased stability with superb fine-contouring properties Dedicated screws for fixation of radial styloid (blue) and support of lunate facet and DRUJ (green) Kirschner wire holes for preliminary plate fixation Oblongue hole for plate positioning and radius length adjustment 2 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

Variable angle Holes allow up to 15 off-axis screw angulation in all directions Variable Angle Locking Screws (VA-LCP) Threaded, rounded head locks securely into threaded VA-LCP holes to provide angular stability at angles determined by the surgeon. Also securely locks into standard locking shaft holes (LCP) of the plate at the pre-defined angle (use of threaded drill guide required). 15 15 Unique drill guides Allowing to use ±15 angulation and pre-defined screw angles Easy x-ray control in pre-defined angle mode and 20 inclined lateral view Synthes 3

Introduction Three Column Theory The treatment of distal radius fractures requires a meticulous reconstruction of the joint surface, as well as stable internal fixation and early functional post-operative treatment. Extraarticular fractures require both the restoration of the volar tilt and radial length to reduce the possibility of displacement. Any malalignment may result in limitations of movement, changes of load distribution, mid-carpal instability as well as the increased risk of osteoarthritis in the radiocarpal joint. Intra-articular fractures with articular displacement of more than 2 mm in the radiocarpal joint inevitably result in osteoarthritis and functional impairment. The distal radius and distal ulna form a three-column biomechanical construction 3 : The ulnar column is the distal ulna, the triangular fibrocartilage and the distal radio-ulnar joint. The intermediate column is the medial part of the distal radius, with the lunate fossa and the sigmoid notch. The radial column is the lateral radius with the scaphoid fossa and the styloid process. VA-LCP Two-Column Volar Distal Radius Plate allows both fixation and buttressing of the two columns of the distal radius A dorsally displaced fracture of the distal radius indicates not only dorsiflection in the sagittal plane, but also radial deviation in the frontal plane and supination in the transverse plane. Following reduction, stabilization requires optimal fixation of the intermediate column as well as the radial column. In the case of a fractured distal ulna that compromises the distal radio-ulna joint, the ulnar column must be stabilized as well. Radial Column Intermediate Column Ulnar Column 3 Rikli DA, Regazzoni P (1996) Fractures of the distal end of the radius treated by internal fixation and early function. A preliminary report of 20 cases. J Bone Joint Surg [Br] 78 (4):588 592 4 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

Clinical Case Case Male patient, 29 years, AO23C3.2 fracture, fall from a 3-meter height Preoperative Peroperative Volar aspect Dorsal aspect Functional results: 10 months after surgery Synthes 5

Tips and Tricks General principles Indications Most distal radius fractures requiring surgical treatment can be accessed from the volar aspect. Combined approach is sometimes used in high-energy fractures, burst fractures and significantly dislocated fractures. In this case, volar approach is first created to reduce volar fragments and after achieving reduction, dorsal approach is used to fix dorsal fragments (mainly the key dorso-ulnar fragment) with the LCP DRP 2.4 plating system. At present, locking implants are indicated for distal radius fractures in case of unstable fractures, i.e. extra-articular fractures with comminuted shaft zone classified by AO as type 23A3, fractures partially affecting the articular area classified as type 23B3 (volar approach) and rare type 23B2 fractures with split dorsal edge (dorsal approach). The ORIF is also indicated for most completely articular, unstable, dislocated, type 23C1 3 fractures. Locking implants are suitable for biologically young patients with osteoporotic bones. The shape of modern implants for distal radius fractures respects anatomical structures of distal radius epiphysis: the edges of implants do not protrude over anatomical landmarks. (See the image.) Operation timing In distal radius fractures, the ORIF is usually implanted 5 7 days post-injury, after soft tissue swelling subsides. This approach is preferred to acute surgeries performed on the day of the injury as it helps to avoid further trauma to soft tissue (the acute carpal tunnel syndrome as a result of swellinginduced compression of the median nerve) when performing invasive treatment shortly after the trauma. most protruding part of distal radius in the volar aspect the watershed line pronator quadratus muscle line volar radius ridge, protrusion In significantly dislocated fractures, urgent reduction and temporary plaster-splint fixation is preferred, or, if possible, external fixator is used to allow better control of soft tissue condition before performing the final surgery and better control of the alignment after primary reduction. 6 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

Approach 1 Make an incision Make a 7 8 cm incision approximately in the axis of the 2 nd metacarpal, over the FCR tendon palpable in the volar radial aspect of distal forearm. To avoid formation of a retracting scar, the incision should be performed in an oblique direction towards the horizontal lines on the wrist. 2 Expose the FCR tendon Continue the dissection over or close to the FCR tendon sheath in the radial aspect, in a longitudinal direction, towards the distal radius. Synthes 7

Approach 3 Create access over the flexor carpi radialis (FCR) muscle sheath 4 Expose the belly and part of the tendon of the FPL Perform ulnar retraction of the flexor pollicis longus (FPL) tendon, exposing the pronator quadratus muscle. It is useful to partially separate the FPL belly from its insertion located on the radial side of the radial shaft. 8 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

5 Expose the pronator quadratus (PQ) muscle 6 Cut through the pronator quadratus muscle perpendicularly to the muscle fibre direction Cut across the PQ fibres approx. 1 1.5 cm from the radial insertion of the muscle. Synthes 9

Approach 7 Cut through the pronator quadratus muscle along the watershed line At PQ distal insertion, cut the PQ edge in the area of the watershed line (a line marked by the most volar portion of the distal radius). Although the distal edge of the VA LCP TCP plate reaches the watershed line, it doesn t cross it thanks to the anatomical shape of the plate, i.e. the plate doesn t protrude in the direction of finger flexor tendons. 8 Separate the PQ on the radial and ulnar side 11 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

9 Expose the bone When the PQ is sufficiently loosened, Hohmann retractors can be introduced behind the radial and ulnar edge of the radius shaft to achieve a sufficient view of the distal radius and the volar aspect of the fracture. Synthes 11

Reduction 10 Intermedial column Dorsally dislocated fragments can be reduced through ligamentotaxis. Alternatively, small chisel or raspatory can be inserted in the fracture line to anatomically reduce the fragments of the intermedial column. 11 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

11 Radial column One of the strongest dislocation forces acting on the radial column is the traction of the brachioradialis muscle, whose insertion is located in the processus styloideus radii. The insertion in the styloid process of the radius can be used to reduce the radial column with a hook carefully inserted between the bone and the tendon, retracting it in the direction of the axis of the extremity, thus reducing the radius styloid process and restoring the ulnar inclination of the distal radius articular surface. After reduction, the fragment can be temporarily fixed by inserting Kirschner wire or reduction forceps (see the peroperative image). Synthes 11

Osteosynthesis 12 Place and fix the plate Place the plate and fix it to the proximal fragment with a 2.7 mm cortical screw inserted in the elongated hole. This will allow for a sufficient correction of the position of the distal edge of the plate towards the watershed line. 11 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

13 Fix the reduced distal fragments to the plate with Kirschner wires Before inserting VA LCP screws, distal fragments can be temporarily fixed with 1.25 mm Kirschner wires inserted in up to six holes in the plate (most often, only two distal holes are used to fix articular surface fragments). Synthes 11

Osteosynthesis 14 Find optimal VA screw trajectories using conical guide As VA LCP screws allow insertion in multiple directions, they can be inserted in subchondral bone as close to the articular surface as possible to ensure optimal support of distal fragments. 11 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

15 Complete the osteosynthesis The lay-out of distal holes allows you to insert a sufficient number of screws ensuring fixation of individual distal radius columns. At least 1 2 LCP screws should be inserted in the diaphyseal part of the plate, mainly in patients with osteoporotic bones to prevent secondary bone shortening or loosening of the plate from the radius shaft. Synthes 11

Closure 16 Close the pronator quadratus muscle using interrupted stitches Start closing the surgical site by suturing the pronator quadratus muscle using interrupted stitches. You will thus separate the plate from finger flexor tendons. 17 Close the incision 11 Synthes VA-LCP Two-Column Volar Distal Radius Plate Tips and Tricks

Follow-up care In current medical practice, a slim suction drain is inserted in the wound after each open reduction performed via volar approach and is left it in situ for 24 hours. The drain is placed on the pronator quadratus muscle after it has been sutured. To complete the treatment, wrist brace is used. Patients are recommended to wear the brace for 6 weeks. In case of osteosynthesis in significantly osteoporotic bones or if a patient suffers concurrent distal ulnar fracture, the extremity is fixed with plaster splints reaching above the elbow line to avoid forearm rotation for 4 6 weeks. Rehabilitation starts immediately, on the first day after the operation. Patients begin exercising fingers and after postoperative pain subsides (2 4 days after operation), physiotherapist-led rehabilitation may be initiated up to approx. ⅓ ½ ROM in the first 3 4 weeks. The beginning and the length of rehabilitation is always decided according to a particular case, based on the nature of the fracture, osteosynthesis stability and bone quality. Increase of the movement range up to full ROM is usually permitted after 4 weeks. Extraction At present, extraction of plates after distal radius fractures is generally not done. Implants are only extracted in situations where secondary reconstruction surgery is expected or in the course of such operation and also in case of increased prominence of the plate in the radial direction (this applies mainly to older-type plates, 3.5 and 2.4 LCP). Synthes 11

About the author Petr Toufar, MD., has been working at the Department of Traumatology and Plastic Surgery at the Traumacenter of Nemocnice České Budějovice, a.s. for 15 years. Under the supervision of the head physician Pavel Kopacka, MD. he could introduce the state-of-the-art methods in the department, gaining skill and experience with new implants manufactured by various companies. Each year, approximately 2,000 injuries are surgically treated at the Department of Traumatology with around 350 polytrauma injuries and multiple serious injuries and the total of 670 patients with distal radius fracture treated with the ORIF. Thanks to this the department ranks among the prominent sites in the Czech Republic. Petr Toufar has been involved in the treatment of distal radius fractures for over 10 years, participating in giving physicians specialized trainings within the AO group in AO courses organized both in the Czech Republic and abroad. Synthes GmbH Eimattstrasse 3 CH-4436 Oberdorf www.synthes.com All technique guides are available as PDF files at www.synthes.com/lit 036.000.TCP Synthes and author 3/2010 Synthes, LCP and Stardrive are trademarks of Synthes, Inc. or its affiliates