Much more about sterilization process monitoring. than you ever thought possible



Similar documents
Biological Indicators and

1/22/2012. When is it a good time to learn about sterile processing in your facility? Continuing Education Contact Hours. Learning Objectives

or

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

THE BASICS OF PACKAGING

3 Attest Biological Monitoring System

Sterile E Online NetworkTutorials

THE BASICS OF STERILIZATION

Because You Can t See Sterile.SM

Best Practices for Loaner Surgical Instruments

Body Art Facility Infection Prevention And Control Plan Guideline

Heat Transfer. Condensation. Revaporization

Steam Sterilization Cycles for Lab Applications

CENTRAL SERVICE REGULATIONS FOR AMBULATORY SURGERY CENTERS

Air Removal from the Sterilization Chamber

Sterilization A REVIEW OF THE BASICS. By Gayla Tilton and Myrna Kauffman 66 MANAGING INFECTION CONTROL

PRINTED: 10/12/2012 FORM APPROVED CENTERS FOR MEDICARE & MEDICAID SERVICES OMB NO (X2) MULTIPLE CONSTRUCTION A.

DISINFECTION & STERILIZATION INFECTION CONTROL GUIDELINES SECTION 4 QUALITY MANAGEMENT IN STERILIZIATION

Comparison Study of the Growth Promotion Capabilities of Self Contained Biological Indicator Culture Medium in Extended Steam Sterilization Cycles

STATIM G4 Cassette Autoclaves From start to dry in the blink of an eye.

The Supplier Perspective: Sterilization. Presented by: Thad Wroblewski STERIS Isomedix Services

Chemical Indicators & Accessories from Albert Browne International Ltd. Assurance through Technology

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use

Network Planning and Analysis

Spandrel Glass Types and Recommendations

Reprocessing. Critical & Semi-Critical Equipment

Water should be removed as soon as possible once the safety of the structure has been verified.

Electronics and Soldering Notes

Infection Control in the Dental Practice through Proper Sterilization

Caring for Your PleurX Pleural Catheter

Packaging Validation according to ISO 11607

INTRODUCTION TO MACHINE-VISION INSPECTION

BESTPRACTICE. HTM01-05 Q&A: You can now follow us on Twitter! We have launched our HTM01-05 questions and answers service on Twitter.

Infectious Waste Management Plan

Lab 3 - DC Circuits and Ohm s Law

WEEKLY SAFETY MEETING All Euramax Subsidiaries LOCKOUT / TAGOUT. Safety Meeting Contents. Meeting Notice. Leaders Guide.

How to Have a Pee-Free Home

Accurate Convenient Affordable Results in 24 Hours. Mail-to-Lab Spore Testing. Mail-to-Lab Spore Testing

STANDARD OPERATING PROCEDURES Safe Autoclave Operations

CARING FOR WATER DAMAGED PAINTINGS

How Does a Doctor Test for AIDS?

Autoclave Testing Service, Inc.

Infection Prevention + Control

PLANT TECHNICIAN SKILLS AND ABILITIES PRACTICE TEST

AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP

Cidex OPA Solution Training Checklist and Competency for Semi-Critical Medical Devices

Connie LeFevre. 22 Texas Home & Living September/October 2009

LOCK-OUT / TAG-OUT SAFETY TRAINING

FACT SHEET : Using Autoclaves Safely

Chemical reaction (slow): Enzyme-catalyzed reaction (much faster):

KGI MEDICAL WASTE MANAGEMENT PLAN

Measuring Soil Moisture for Irrigation Water Management

Sterilization of Surgical Implants: Did You Know

Horizontal Steam Sterilizers. Healthcare Line. Since

MEDICAL DEVICE GUIDANCE

Cutting Vinyl Effectively

The design planning implications of HTM 01-05: decontamination in primary care dental practices

Fundamental Autoclave Techniques

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Guidelines. for Miniscrew Implant Sterilization and Use What orthodontists do and don t need to do when placing MSIs

Laboratory Biosafty In Molecular Biology and its levels

AST Standards of Practice for Packaging Material and Preparing Items for Sterilization

Background on Biodegradable Additives

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

Acids and Bases. AND a widemouth container of the following solids:

Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.

Steam Sterilization and the 2007 Revision of PDA Technical Report 1

PT AVENUE GUIDE OVERVIEW

Laboratory 5: Properties of Enzymes

Corden Pharma Latina S.p.A. 5/20/16

Aseptic Technique. A GMP/GTP Training Module

Carpet Maintenance Guideline Broadloom Carpet with

Operating Manual Color Changing Fountains

Soldering. Resources and methods for learning about these subjects (list a few here, in preparation for your research):

(Refer Slide Time: 2:03)

Supplier Quality Agreements

WEEKLY SAFETY MEETING All Euramax Subsidiaries MATERIAL SAFETY DATA SHEETS (MSDS) Safety Meeting Contents. Meeting Notice.

When you want to explain why something happens, you can use a clause of

KODAK PROFESSIONAL D-76 Developer

Sterilization of Dental Instruments Dental Clinical Guidance

Colours to stand the test of time. paint range

Steam Sterilization Principles

position statement INFECTION CONTROL INTRODUCTION POSITION STATEMENT PERFORMANCE EXPECTATIONS

Lesson 4. Temperature change

A Fishy Tale. Observing the Circulatory System of a Goldfish with a Compound Light Microscope

Look out for the reviews before you happen to finalize the siding replacement service

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised

OA4-13 Rounding on a Number Line Pages 80 81

25 Integers: Addition and Subtraction

AHRI White Paper: Reports of R-134a Contaminated with R-40 and Other Refrigerants

The faecal occult blood (FOB) test

600 Series Camera. User Guide

Lab Exercise 3: Media, incubation, and aseptic technique

FINAL DOCUMENT. Global Harmonization Task Force. Title: Label and Instructions for Use for Medical Devices

Decomposing Numbers (Operations and Algebraic Thinking)

Oxivir Tb and Accelerated Hydrogen Peroxide (AHP) Frequently Asked Questions

INSTALLATION RECOMMENDATIONS. Outdoor Rubber Tiles

THERAPEUTIC USE OF HEAT AND COLD

MAKING A BETTER MARK WITH INK JET As substrates, industry regulations and customer demands change, fluids are also evolving

Safe Dry Goods Storage

Transcription:

Much more about sterilization process monitoring than you ever thought possible

Welcome to your official guide to sterilization process monitoring. Shrinking health care budgets and growing concerns about health care liability have made a complete and successful monitoring program more important than ever. So you couldn t have ordered this guide at a better time. Within these pages you ll find all you need to create a sterility assurance program that s not only effective, but easy to follow. And for easy reference, we ve put a chart in the back that includes the various categories of sterilization process monitoring. So, without further delay, we offer you much more about sterilization process monitoring than you ever thought possible.

Table of Contents Introduction............................................2 3M Sterilization Assurance Program........................4 Load Control...........................................6 Pack Control..........................................10 Equipment Control.....................................14 Exposure Control......................................18 Record Keeping.......................................22 Wrap Up.............................................24 Sterilization Process Chart...............................25 1

2 Introduction

Clean Sterilization Process Prep/Pack Sterilize Store Issue/Use here are five basic steps in the sterilization process itself: Clean, Prep/Pack, Sterilize, Store and Issue/Use. Sterilization process monitoring impacts just two of these steps, Sterilize and Issue/Use. Monitoring tools verify the outcome of Sterilize and supply safeguards to the Issue/Use step so that no nonsterile medical devices can go beyond this point. The threat of postoperative infections caused by nonsterile medical devices makes sterilization process monitoring extremely important. 3

3M Sterilization Assurance Program n undetected sterilization process failure can put patients, staff and the financial health of your facility at risk. Additional costs associated with postoperative infections and other health care liabilities, make the implementation of a sterilization process monitoring program an extremely critical practice for all health care facilities. The 3M Sterilization Assurance Program is a comprehensive and practical approach to sterilization monitoring procedures and methods that you can count on to reduce your risk of undetected sterilization process failure. The 3M Sterilization Assurance Program helps you control and monitor sterilization procedures. It provides you with an easy-to-follow set of guidelines to help ensure safety and product sterility throughout your facility. The 3M Sterilization Assurance Program consists of five separate, but interrelated steps: Load Control, Pack Control, Equipment Control, Exposure Control and Record Keeping. These five 4

steps monitor every aspect of the sterilization process and help you establish, manage and maintain a consistent protocol for sterilization in your facility. Load Control Pack Control Health care professionals worldwide have come to trust the complete line of 3M sterilization monitoring products to help them monitor all stages of the sterilization process. 3M sets today s standard for sterilization process monitoring with dependable, high-performance products, an effective, easy-to-follow program, customer service and technical support. Equipment Control Exposure Control Record Keeping 5

3M Sterilization Assurance Program Load Control 6 t s no accident that Load Control is the first level of control on the list. It is the foundation of a successful sterilization process monitoring program. So, what is it? Load Control is the process by which a load is monitored and released based on the result of a biological indicator (BI) in a test pack. Only a BI can detect the actual killing of microbial spores inside the sterilizer. This is why Load Control remains the most reliable level of testing you can use. Because if all spores die inside the biological indicator, you have assurance that other infectious organisms have also died inside the sterilizer. Because it detects the killing of microbial spores, a BI yields information more valuable than any other sterilization process monitor. Other types of indicators may be used to help assess sterilizer performance. However, to reliably validate the performance of the load in the routine sterilization process, you still need to use a BI. In fact, only BIs are currently recognized for use to release loads

In the long run, regular use of a rapid readout BI is more cost effective because it reduces the risk of recall, the release of contaminated loads, and postoperative patient infections. The sterilization reliability that only a biological indicator can deliver

Load Control with implantables. And, when other monitors indicate a possible sterilizer problem, it s good practice to revalidate the sterilizer with a BI. Given its many applications, you can see why a BI in every load can be the most practical approach to sterilization process monitoring. Frequent use of a BI is the most reliable way to control loads. Here s another way to look at it: If Load Control is done only once daily, you might assume other loads run during the day were as effectively sterilized as the load that was monitored with a BI. This assumption can lead to a false sense of security, and could allow medical devices that weren t successfully sterilized to be released. In addition, if you get a positive result with a BI, you need to recall all medical devices since your last negative BI. However, if you re using a BI in every load, you need to recall and reprocess only the one load. It s also important to remember that BIs yield faster results today than ever before, some as quickly as one hour. Given its reliability and the quality of the monitoring information received, regular use of a BI leads to better patient outcomes and improves the performance of the sterilization process. 8

3M LOAD CONTROL PRODUCTS Accepted practices state that a BI should be used in every load of an EO sterilizer and at least once a day in a steam sterilizer. However, to maintain the highest level of effectiveness, a BI in every load of a steam sterilizer is highly recommended. In addition, BIs should always be used in loads with implantables and during installation testing, ongoing quality assurance, after major repairs, after a positive BI result and during product testing. With rapid readout technology, every load can be monitored and efficiently quarantined until the BI result is determined. This will do away with recall and credibility issues. 9

The inside counts in sterilization process monitoring First, trust your biological indicator for Load Control, then trust your chemical indicator for Pack Control.

3M Sterilization Assurance Program Pack Control ow that you re a believer in the value of Load Control using a BI, we can move on to pack control. It s defined as the use of chemical indicators for internal monitoring of packs, trays, containers and peel pouches. Internal pack monitoring validates that the sterilant has penetrated to the point of placement in the pack and confirms that sufficient exposure conditions have been met. Sounds good, but what does it mean in English? The basic idea behind Pack Control is this: By placing a multi-parameter or integrating chemical indicator inside every pack, you can detect local problems that sometimes occur due to human error or mechanical malfunction. You see, even with a negative BI result, sometimes the sterilant may not penetrate individual packs. This can be caused by a number of things, depending on your method of sterilization. There could be an air pocket in your steam sterilizer or a small leak in the vacuum system, or the pack itself may be wrapped too densely or the load packed too tightly for the sterilant to 11

Pack Control penetrate. And even a short distance apart, chemical indicators can respond differently than biological indicators. Again, the advantage of Pack Control is that it allows you to single out individual Chemical integrators should be used in conjunction with BIs. packs that were not exposed to sufficient sterilization conditions. So if your BI for Load Control is negative but your chemical indicator for a specific pack indicates a problem, you know you have to recall and reprocess only the pack. * The trick is to always trust the indicator that indicates a problem and to remember that Pack Control monitoring serves as a companion tool to Load Control, with the results of Load Control taking dominance over Pack Control monitoring. There are two basic types of chemical indicators for internal pack monitoring in steam and EO sterilizers. The most common is the color strip, which is coated with a sensitive chemical that goes through a color change when exposed to the right conditions for sterilization. These are most commonly multi-parameter indicators, meaning they monitor at least two of the critical parameters of the process (i.e., time and temperature). 12

TM 3M PACK CONTROL PRODUCTS A higher order of chemical indicators are the integrating indicators, generally referred to as chemical integrators. These chemical monitors are designed to approximate the response of biological indicators to all the critical parameters of the sterilization process. Chemical integrators are movingfront style indicators which use special color chemistry that advances across a window to accept or reject markings. Both types of chemical indicators provide important information about conditions of exposure inside packs the essence of Pack Control. Comply TM SteriGage TM DARK BAR MUST PASS THIS POINT Steam Chemical Integrator REJECT ACCEPT 1243 Comply TM SteriGage EO Chemical Integrator REJECT ACCEPT 1244 Blue bar must enter ACCEPT area Comply TM 00117 Comply TM Comply 1250 Steam Chemical Indicator Stripe turns dark with processing 12 Comply Steam Chemical Indicator - Plus Strip 270 0 F Accept only if both bars are green EO 1251 EO Chemical Indicator Stripe turns green with processing STEAM Steam Chemical Indicator - Plus Strip 250 0 F 00119 Accept only if both bars are green STEAM Comply TM 00118 Steam Chemical Indicator - Plus Strip Accep 12 Comply TM 00117 Steam Chemical Indicator - Plus Strip 270 0 F Accept only if both bars are green STEAM *Remember, if the BI results are positive, every pack in the load must be recalled and reprocessed. Indicator bar turns as dark or darker than color standard during steam sterilization; may turn silver or grey when overexposed. Color Standard Comply TM Steam Chemical Indicator Strip 00109 STEAM Comply TM EO Chemical Indicator Strip 00152 EO 1 Indicator bar turns blue when exposed to ethylene oxide sterilizing conditions

3M Sterilization Assurance Program Equipment Control quipment Control is a way to find out whether or not your sterilizer is doing its job properly. To do this for vacuum-assisted sterilizers, you begin with an air removal test pack such as a Bowie-Dick test. When using vacuum-assisted cycles, it is critical that you know your steam sterilizer is removing air efficiently before you start your daily sterilization routine. If your sterilizer has a weak or leaking vacuum, air pockets may form inside the sterilizer and prevent proper steam penetration of some packs in the load. This would likely compromise sterility. To detect air pockets in your sterilizer, you should run either a facility-prepared Bowie- Dick towel pack or a commercially available Bowie-Dick type disposable test pack at the beginning of each day. Both use a chemical indicator inside a challenge pack that will not develop properly if air remains trapped. 14

Equipment Control reveals if the sterilizer is operating to set conditions of time, temperature, pressure, air removal, moisture conditioning and sterilant exposure. Keeping the mechanics of sterilization process monitoring running smoothly 1

Equipment Control If the Bowie-Dick test indicates a problem, you must take the sterilizer out of service until the sterilization problem is identified and serviced, or until the unit is replaced. If the Bowie-Dick test shows no problem, you can begin your daily sterilization routine using the other types of Equipment Control: the use of mechanical monitors and chemical integrator test packs. Mechanical monitors allow you to watch the gauges on the sterilizer and record Remember to record your results. time, temperature and pressure so you can identify an equipment malfunction before it becomes a problem. Incorrect gauge readings, or readings that vary from set conditions, indicate a malfunction and you should consider the load nonsterile. The sterilizer should be taken out of service until you make adjustments or the problem is corrected. Another means of Equipment Control in the daily sterilization routine is the use of chemical integrator test packs. The test packs utilize chemical integrators contained inside a porous challenge pack. Designed to approximate the response of biological indicator test packs, chemical integrator test packs should be used in every load that does not contain a biological indicator. Test packs with chemical integrators can serve as a monitoring bridge between BI-monitored loads by providing immediate results on sterilizer performance before the load is released. 16

3M EQUIPMENT CONTROL PRODUCTS These Equipment Control tests provide information on how the sterilizer is operating to set conditions of time, temperature and pressure as well as its ability to remove air, promote sterilant penetration in packages and condition the load with moisture (EO). Other sources of failure, such as improper packaging and loading, can be caught with the use of Pack Control mentioned earlier. So you re set. Well, almost set. There are just a couple more things. 17

Seeing is believing in sterilization process monitoring Exposure Control monitors may not be multi-parameter or integrating chemical indicators and must not be used as internal Pack Control monitors. 1

3M Sterilization Assurance Program Exposure Control ince Pack Control for sealed packs is mainly for user departments once the sealed packs have been distributed, there must be another way for sterilizer operators to know whether sealed packs have been exposed to the sterilization process. Enter Exposure Control. Exposure Control monitoring is a way to identify processed from unprocessed medical devices at a glance. It assures the operator handling the processed items that the pack has been exposed to the sterilization process without the need to open the pack or check Load Control records. Exposure Control involves the use of chemical indicators that are externally visible. They can be on a tape used to seal packs, on paper strips that can be placed into peel pouches and unwrapped trays, or on load cards. The one most commonly used for Exposure Control is the indicator tape. 19

Exposure Control Chemical indicators for Exposure Control tell you only whether they were exposed to the sterilization process. The tape, strip and card undergo a color change when they ve been exposed to the An external chemical indicator should be used with every package. sterilant. The nice thing about tape is its dual purpose. It s used as an indicator and as a means to secure packages for sterilization. Indicator tapes come in a variety of formats. Some are made for disposable wraps and contain stronger adhesive. Others are made for reusable wraps. Their adhesive is strong enough to hold the wrap, but light enough to leave no adhesive residue on the wrap when it s cleaned. Externally visible chemical indicators for Exposure Control should be used with every package. For unwrapped trays or peel pouches, Pack Control monitoring may eliminate the need for exposure indicators since their result is externally visible. As with other chemical indicators, Exposure Control indicators are available for steam and EO. 20

3M EXPOSURE CONTROL PRODUCTS Though not as reliable, Exposure Control for sterilization operators is like Pack Control for end users. Steam Tape Steam Tape for Disposable Wraps EO Tape 21

3M Sterilization Assurance Program Record Keeping 22 he final touch to a complete and successful sterilization process monitoring system is Record Keeping. This level of control documents the materials that have been processed and their monitoring control evidence. It s a way to keep track of the process using record keeping envelopes, forms and labels. If you ve ever run into a recall situation that forced you to investigate the source of the problem, you know how valuable this can be. For example, suppose a sealed pack from a load is delivered to a user department. Several days or even weeks later, it is opened and, as indicated by the Pack Control chemical indicator, it wasn t exposed to sufficient conditions for sterilization. The pack is recalled so it can be reprocessed. Now suppose that several more packs are recalled within a short period of time and a pattern of failure begins to develop. You need to find the origin of the problem. If you ve been keeping good records, you should be able to trace each pack backward through the levels of monitoring control to the sterilization event itself. This will make it much easier to diagnose the problem. It s important to note that by using a BI for Load Control in every load, tracing your steps becomes much less complicated. Yet another good reason to get in the habit of using a rapid readout BI in every load and quarantining until the BI results are available.

Documenting the evidence in sterilization process monitoring Each pack should be labeled so it can be traced back through records in the event of a problem. 1

Wrap Up This is a lot of information to digest. So, to sum up, here are the basics of what you need to do to maintain the most complete and successful sterilization process monitoring program, followed by a chart you can use as a handy reference. We hope this has been helpful. THE BASICS For the best results, use a rapid readout biological indicator in each load (Load Control); a chemical indicator inside each pack (Pack Control); a chemical indicator outside each pack, tray, container or peel pouch (Exposure Control); a Bowie-Dick test pack each day in each vacuum assisted sterilizer, read and record the mechanical monitors on the sterilizer, use a chemical integrator test pack in each load not monitored with a biological indicator (Equipment Control); and record all the monitoring results in a record keeping system (Record Keeping). 24

Sterilization Process Clean Process Monitoring Load Control The sterilization reliability that only a biological indicator can deliver. Prep/Pack Every load. Sterilize Monitored by load, equipment and exposure control. How Often? Store With What? Issue/Use Monitored by exposure and pack control. Rapid readout or standard biological indicator. Pack Control The inside counts in sterilization process monitoring. Equipment Control Keeping the mechanics of sterilization process monitoring running smoothly. Exposure Control Seeing is believing in sterilization process monitoring. Record Keeping Documenting the evidence in sterilization process monitoring. Every pack. Air removal: Beginning of each day, after major repairs or for installation testing. Mechanical monitoring: Every load. Integrator test pack: Every load that does not have a BI test pack. Every pack. Label each pack so there is a traceable path to monitoring control records. Multi-parameter chemical indicators or chemical integrators. Disposable Bowie-Dick test packs, test sheets for AAMI Bowie-Dick test packs and chemical integrator test packs. Internal indicator rolls, indicator strips or indicator tapes. Load record cards, load labels, record keeping binders and log books. 3M Health Care 3M Center, Building 275-4E-01 St. Paul, MN 55144-1000 USA 1-800-228-3957 3M 1997 70-2008-9732-3 25

3M is a world-wide leader in sterilization assurance products, helping you to ensure safety and product sterility. Combined with our comprehensive product line as well as 3M s excellence in innovation, experience and service, we stand ready to put our leadership, products and service to work for you. For more information about 3M s Sterilization Assurance Program and the full line of monitoring products, to place an order, or for additional copies of this booklet, call 1 800 563-2921 (English), 1 800 361-4488 (Atlantic Canada), or 1 800 361-2650 (French) or call the 3M representative for your area. 3 Health Care 3M Canada Post Office Box 5757 London, Ontario N6A 4T1 1 800 563-2921 English 1 800 361-2650 French 1 800 361-4488 Atlantic Canada Printed in Canada. 3M 1997 9805 MS07843 E