Fundamentals of Breast Tomosynthesis



Similar documents
Hologic Selenia Dimensions C-View Software Module. October 24, 2012

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Mammography. What is Mammography?

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

How To Compare Mammography To A 3D System

Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1

Digital Mammography Update: Design and Characteristics of Current Systems

Concepts for High-Resolution Low-Dose CT of the Breast

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

A Guide to Breast Imaging: The Latest Technology for Screening and Detecting Breast Cancer

VI. FREQUENTLY ASKED QUESTIONS CONCERNING BREAST IMAGING AUDITS

Digital Breast Tomosynthesis QC Requirements

Q: What differentiates a diagnostic from a screening mammography procedure?

Breast Density Legislation: Implications for primary care providers

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Breast Ultrasound: Benign vs. Malignant Lesions

UNITED STATES OF AMERICA FOOD AND DRUG ADMINISTRATION RADIOLOGICAL DEVICES PANEL October 24, :00 a.m.

Patient Prep Information

Quality Control of Full Field Digital Mammography Units

Infrared Thermography Not a Useful Breast Cancer Screening Tool

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

White Paper. Study to Demonstrate Efficacy of 8MP Color Display - Mammography 1.INTRODUCTION PURPOSE METHODS RESULTS...

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

MAMMOGRAPHY GOALS AND OBJECTIVES

Personalized Breast Screening Service

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Physics testing of image detectors

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

Mammography Applications for MAMMOMAT 1000/3000 Nova

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

The Principles of Contrast Mammography

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Low-dose CT Imaging. Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital

Breast Sonography general goal. Optimizing Breast Sonography. BUS indications -- all. Breast Sonography specific goals.

How To Write A Draft Report On Breast Cancer Screening

Digital Mammogram National Database

A Breakthrough in Breast Specimen Imaging

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

X-ray (Radiography) - Bone

Sustaining a High-Quality Breast MRI Practice

im3d S.p.A. La Ricerca fa Impresa

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

The File-Card-Browser View for Breast DCE-MRI Data

How To Improve Your Ct Image Quality

Breast Imaging Made Brief and Simple. Jane Clayton MD Associate Professor Department of Radiology LSUHSC New Orleans, LA

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

National Medical Policy

Automated Breast Volume Scanning 3D Ultrasound of the Breast

Cardiac CT for Calcium Scoring

for breast cancer detection

Guideline for the Imaging of Patients Presenting with Breast Symptoms incorporating the guideline for the use of MRI in breast cancer

A PACS-Aware DICOM Image Object

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

Update on ACR Digital Mammography QC Manual

Data. microcat +SPECT

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

Coronis 5MP Mammo. The standard of care for digital mammography

State-of-the-Art Technology in Cardiac CT

Needle crystal detector technology in mammography further dose reduction and clinical image quality with

BREAST IMAGING. Developed by the Ad Hoc Committee on Resident and Fellow Education of the Society of Breast Imaging

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Great Strides MARKEY COMPREHENSIVE BREAST CARE CENTER: SETTING A NEW STANDARD OF CARE

Supports screening, diagnostic, and multimodality workflows

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

OBJECTIVES By the end of this segment, the community participant will be able to:

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Page: 1 of 6 Page: 1 of 6

BREAST IMAGING FOR SCREENING AND DIAGNOSING CANCER

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Purchasing a cardiac CT scanner: What the radiologist needs to know

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Nicole Kounalakis, MD

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Understanding astigmatism Spring 2003

Computerized Tomography, X-rays, and the Radon Transform

D. FREQUENTLY ASKED QUESTIONS

Breast MRI Quality Control

Recommendations for cross-sectional imaging in cancer management, Second edition

Stereotactic Breast Biopsy

BREAST IMAGING SOLUTIONS. Integrated Breast Imaging Workflow Solutions

Update on MQSA and ACR Mammography Accreditation

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

Twelve. Figure 12.1: 3D Curved MPR Viewer Window

DENTAL Cone beam 3D X-RAY SYSTEM with

Transcription:

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 1 Fundamentals of Breast Tomosynthesis Improving the Performance of Mammography Andrew Smith, Ph.D. This white paper is one in a series of research overviews on advanced technologies in women s healthcare. For copies of other white papers in the series please contact womenshealth@hologic.com Introduction Breast tomosynthesis is a three-dimensional imaging technology that involves acquiring images of a stationary compressed breast at multiple angles during a short scan. The individual images are then reconstructed into a series of thin high-resolution slices that can be displayed individually or in a dynamic ciné mode. Reconstructed tomosynthesis slices reduce or eliminate the problems caused by tissue overlap and structure noise in single slice two-dimensional mammography imaging. Digital breast tomo synthesis also offers a number of exciting opportunities including improved diagnostic and screening accuracy, fewer recalls, greater radiologist confidence, and 3D lesion localization. Hologic has conducted a multi-center, multi-reader clinical trial to measure the clinical performance of tomosynthesis in a screening environment. This paper outlines the theory of tomosynthesis, its expected clinical benefits, and summarizes the results of the clinical trial. Selenia Dimensions Breast Tomosynthesis System* Theory of Tomosynthesis Conventional x-ray mammography is a two-dimensional imaging modality. In conventional mammography, pathologies of interest are sometimes difficult to visualize because of the clutter of signals from objects above and below. This is because the signal detected at a location on the film cassette or digital detector is dependent upon the total attenuation of all the tissues above the location. Tomosynthesis 1,2,3,7,8 is a three-dimensional method of imaging that can reduce or eliminate the tissue overlap effect. While stabilizing the breast, images are acquired at a number of different x-ray source angles. Objects at different heights in the breast display differently in the different projections. In Figure 1, two objects (a spiculated lesion and ellipse) superimpose when the x-rays are at 0º, but the off-axis acquisitions shift the objects shadows relative to one another in the images. The final step in the tomosynthesis procedure is reconstructing the data to generate images that enhance objects from a given height by appropriate shifting of the projections relative to one another. An example is shown in Figure 2 where we reconstruct a cross sectional slice at one specific height. In this example, the images are summed, shifting one relative to another in a specific way that reinforces the spiculated lesion object and reduces the contrast of the ellipsoidal object by blurring it out. Note that additional acquisitions are not required to enhance the visibility of objects at any given height one set of acquired data can be reprocessed to generate the entire 3D volume set. Performing the Acquisition The geometry of tomosynthesis is shown in Figure 3. The breast is compressed in a standard way. While holding the breast stationary, the x-ray tube is rotated over a limited angular range. A series of low dose exposures are made every degree or so, creating a series of digital images. Typically, the tube is rotated through 10-20 degrees and 10-20 exposures are made every 1 or so during a total scan of 5 seconds or less. The individual images are projections through the breast at different angles and these are what are reconstructed into slices. * Caution. Investigational device in the U.S. FDA clearance pending

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 2 Figure 1: Tomosynthesis Can Reduce or Eliminate Tissue Overlap Figure 3: Tube Head Swings Taking Multiple Exposures in a Few Seconds Incident x-rays Images as seen on cassette or digital detector With tomosynthesis imaging, images acquired from different angles separate structures at differing heights. Conventional mammography would acquire only the central image Figure 2: Tomosynthesis Enhances Objects at a Given Height Individual Images Synthesized Image Individual Image Projection 1 Individual Image Projection 2 Individual Image Projection 3 1 + 2 + 3 = 1mm Slice By shifting and adding the acquired projections, 3-D imaging increases the visibility of objects by blurring out objects from other heights Normally the breast would be placed in the MLO or the CC view, although the tomosynthesis system should support the ability to acquire images in any desired orientation. One consideration in the design of tomosynthesis systems is the motion of the x-ray source during acquisition. The x-ray tube can move in a continuous or step-and-shoot motion. With continuous motion x-ray exposures must be short enough to avoid image blurring due to focal spot motion. If step-and-shoot motion is employed, the gantry must come to a complete stop at each angular location before turning on the x-rays, otherwise vibration will blur the image. The most important criteria is that the total scan time be as short as Tomosynthesis acquisition geometry showing the motion of the x-ray source. possible, to reduce the possibility of patient motion which will degrade the visibility of small microcalcifications and small spiculations. The angular range and number of exposures acquired during the scan are additional variables that need to be optimized. In general, more exposures will allow reconstructions with fewer artifacts. This must be balanced against the fact that for a given total examination dose, more exposures will mean smaller signals for each of the individual shots. For sufficiently small exposures, imager receptor noise will dominate the image and degrade reconstructed image quality. Increased number of exposures also increase raw data size and reconstruction times. In regards to angular range, a larger angular range gives increased reconstructed slice separation, where smaller angular ranges keep more structures in focus in a given slice. Increased separation theoretically might be desired for resolving two closely lying structures, but could greatly impair the appreciation of a cluster of microcalcifications by having individual calcifications appear in different slices, or the appearance of spiculations lying in more than one narrow plane. Tomosynthesis System Requirements Detector Efficiency and Dose Tomosynthesis imaging consists of a series of low dose exposures, with every acquisition about 5-10% of a normal single-view mammogram. Because each exposure is low dose, it is essential that the image receptor have a high quantum efficiency and low noise. Because images are being acquired at a rate of several images per second, rapid imaging is another

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 3 requirement. Selenium-based image receptors, with their high Detective Quantum Efficiency (DQE), greater than 95% x-ray absorption at mammographic energies, and rapid readout capabilities, are an ideal detector for tomosynthesis systems. Using a selenium detector, one is able to perform a tomosynthesis examination with a total radiation dose similar to conventional mammography. Modes of Acquisition The tomosynthesis system must be capable of performing all existing 2D digital mammography examinations in addition to the tomosynthesis acquisitions. Tomosynthesis images must be able to be taken in all standard orientations, not just CC and MLO. The system should also be able to take a normal 2D mammogram and the tomosynthesis examination in the same compression. To facilitate this, automated grid retraction is a requirement, so the system can rapidly and automatically switch between 2D and 3D imaging modes. Image Reconstruction In Figure 4 the tomosynthesis reconstruction process consists of computing high-resolution images whose planes are parallel to the breast support plates. Typically, these images are recon structed with slice separation of 1 mm, thus a 5 cm compressed breast tomosynthesis study will have 50 recon structed slices. Rapid reconstruction time is essential, especially when tomo synthesis is being considered as part of an interventional study, and for this reason it is important to keep post-acquisition processing to 10 seconds or less. Display Methodology The reconstructed tomosynthesis slices can be displayed similarly to CT reconstructed slices. The operator can view the images one at a time or display them in a ciné loop. The original projections are identical to conventional projection mammograms, albeit each one is very low dose, and these can be viewed as well, if desired. If the system acquired a 2D and a 3D mammogram in the same compression, images from these two modalities are completely co-registered. Workstation user interfaces that allow rapid switching between the two modes will facilitate image review, and allow rapid identification of lesions in one modality with the corresponding lesion in the other modality. Figure 5 shows an example of selected reconstructed tomo synthesis slices in a breast. Figure 4: Tomosynthesis Takes Multiple Angle Breast Views and Reconstructs Them Into Cross-sectional Slices Left image shows three of the 15 projection images acquired through the breast at different angles. Right image shows three of the 1-mm cross-sectional reconstructed slices

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 4 Figure 5: Reconstructed Tomosynthesis Slices Reconstructed tomosynthesis slices through the breast from breast platform up to compression paddle reveal objects lying at differing heights in the breast, such as cysts and calcifications shown by arrows Figure 6: Superior Low Contrast Visibility When the ACR phantom is placed on a cadaver breast and imaged (left), visibility of low contrast objects are reduced. Even at 4 a conventional dose the digital mammogram (middle) shows inferior low contrast visibility to a tomosynthesis image (right) using ¼ the dose

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 5 Figure 7: Tomosynthesis Offers Clearer Images Tissues that overlap in conventional mammography (left image) and hide pathologies (blue lesion) are less likely to be obscured using tomosynthesis (right image) Potential Clinical Benefits Reduced Recalls Fewer Biopsies Improved Cancer Detection Tomosynthesis should resolve many of the tissue overlap reading problems that are a major source of the need for recalls and additional imaging in 2D mammography exams. The biopsy rate might also decrease through improved visualization of suspect objects. Some pathologies that are mammographically occult will be discernible through the elimination of structure noise and tomosynthesis may therefore allow improved detection of cancers. Tissue Localization Because the location of a lesion in a tomosynthesis slice completely determines its true 3D coordinate within the breast, biopsy tissue sampling methods can be performed using the tomosynthesis generated coordinates. Clearer Images Because the images are presented with reduced tissue overlap and structure noise, objects are expected to be visualized with improved clarity. This will likely lead to more confident readings. Figure 7 demonstrates why we expect improved confidence with 3D tomosynthesis imaging. In conventional mammography breasts are compressed so as to reduce tissue overlap. Tomosynthesis is able to provide good visibility of lesions because of the reduction of structure noise. In this figure, the pathology, shown in blue, is obscured in the 2D image from the overlapping tissues shown in white. The appropriate crosssectional 3D slice, shown on the right, allows clear visualization of the lesion. The result is improved confidence by the radiologist in their assessments. One versus two views In the early development of tomosynthesis it was suggested 4 that tomosynthesis imaging might only require acquisitions in the MLO view, because the 3D nature of the tomosynthesis images allow viewing the breast from multiple angles. Current indications are that this is not true and that tomosynthesis will require both the MLO and the CC view. This is not surprising, because tomosynthesis differs from other 3D imaging modalities such as CT in that one cannot generate orthogonal multi-planar reconstructions such as sagittal and coronal views from the transverse tomosynthesis image sets. Pathologies that are elongated, planar, or non-spherical in shape may well be better visualized when imaged in one orientation than another. A recent scientific presentation 6 found that 9% of cancers in their study were seen in the CC tomosynthesis view but not visible in the MLO tomosynthesis view. Tomosynthesis Clinical Trials Hologic has completed a multi-center, multi-reader trial investigating the performance of tomosynthesis. 5 The purpose of the study was to compare radiologists cancer detection rate and screening recall rate using conventional digital mammography (2D) plus breast tomosynthesis (3D), to the cancer detection rate and recall rate observed when using 2D alone. In the study, 1,083 women from 5 clinical centers underwent 2D and 3D imaging of both breasts. Cases were collected from a screening population and enriched with patients from diagnostic mammography. Both 2D and 3D imaging consisted of CC and MLO images of both breasts. The CC and MLO 3D images were performed using the Hologic Selenia tomo - synthesis prototype. Three hundred sixteen imaging data sets were randomly chosen to be reviewed by 12 radiologists. The 2D images were scored first, and then the readers reviewed and scored the 2D and 3D exams together. For all 12 readers, clinical performance was superior for 2D plus 3D imaging compared with 2D alone, as measured using the area under the ROC curve. Figure 8 shows the ROC curve generated from averaging the individual 12 ROC curves. The mean area under the ROC curve for the readers increased from 0.83 to 0.90 using a forced BIRADS scoring, showing an increase of 0.07, a highly significant increase with a p-value of 0.0004. Using 2D plus 3D versus 3D alone, sensitivity, defined as cancers scored as BIRADS 4 & 5, improved from 66% to 76%; specificity, defined as non-cancers with BIRADS scores of 1-3, increased from 84% to 89%; and a mean reduction in the recall rate of 43% was observed. In this multi-center, multi-reader study, radiologist performance improved significantly when using 2D combined with 3D compared with using 2D alone.

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 6 Figure 8: Tomosynthesis Improved Performance Compared to Mammography ROC curves for 2D and for 2D+3D imaging performance, averaged over 12 readers, shows a significant improvement in clinical performance using tomosynthesis Conclusions Breast tomosynthesis provides a 3D imaging capability that allows the more accurate evaluation of lesions by enabling better differentiation between overlapping tissues. A lower recall rate, higher positive predictive value for a biopsy recommendation, higher cancer detection rates, fewer recalls, fewer biopsies, and improved radiologist confidence are expected to result from the use of this technology. Breast tomosynthesis should be valuable in both screening mammography and diagnostic mammography. References 1 Dobbins JT III and Godfrey DJ. Digital x-ray tomosynthesis: current state of the art and clinical potential, Phys. Med. Biol. 2003 Oct 7; 48(19)R65-106. 2 Newman, L. Developing technologies for early detection of breast cancer: a public workshop summary. Washington, D.C.: Institute of Medicine and Commission on Life Sciences National Research Council, 2000. 3 Niklason LT, Christian BT, Niklason LE, Kopans DB, et al. Digital tomosynthesis in breast imaging. Radiology. 1997 Nov; 205(2): 399-406. 4 Rafferty EA, Kopans DW, Wu T, Moore RH. Breast Tomosynthesis: Will a Single View Do? Presented at RSNA 2004, Session SSM02-03 Breast (digital mammography). 5 Rafferty EA, Niklason L, Halpern E et al. Assessing Radiologist Performance Using Combined Full-Field Digital Mammography and Breast Tomosynthesis Versus Full-Field Digital Mammography Alone: Results of a Multi-Center, Multi-Reader Trial. Presented at RSNA 2007, Session SSE26-02 Late Breaking Multicenter Clinical Trials. 6 Rafferty EA, Niklason L, Jameson-Meehan L. Breast Tomosynthesis: One View or Two? Presented at RSNA 2006, Session SSG01-04 Breast Imaging (digital tomosynthesis.) 7 Smith A, Hall PA, Marcello DM. Emerging technologies in breast cancer detection. Radiology Management. July-August 2004; 16-27. 8 Smith AP, Ren B, DeFreitas K et al. Initial Experience with Selenia Full Field Digital Breast Tomosynthesis. In: Proceedings of IWDM Durham North Caroline, June 2004, Ed. Etta Pisano. Glossary 2D 3D BIRADS CC Cesium iodide CT DM DQE Forced BIRADS MLO ROC Selenium Tomosynthesis Acknowledgments Two-dimensional Three-dimensional Breast Imaging Reporting And Data System Craniocaudal A radiation detection material used in indirectconversion x-ray image receptors Computed Tomography Digital Mammography Detective Quantum Efficiency. A measure of the dose efficiency of a detector BIRADS scoring allowing values of 1 to 5 only, i.e. no recalls Mediolateral oblique Receiver Operating Characteristics. A graphical plot of the true positive fraction (sensitivity) vs. false positive fraction (1-specificity) for a binary classifier system as its discrimination threshold is varied A radiation detection material used in direct-conversion x-ray image receptors Tomosynthesis combines digital image capture and processing with simple tube/detector motion as used in conventional radiographic tomography. Although there are some similarities to CT, it is a separate technique I want to thank the following institutions for supplying the images shown here: Dartmouth Hitchcock Medical Center, Lebanon, NH USA Magee Women s Hospital, Pittsburgh, PA USA Massachusetts General Hospital, Boston, MA USA Netherlands Cancer Institute Antoni Van Leeuwenhoek Hospital, Amsterdam, Holland University of Iowa Health Care, Iowa City, IA USA Yale University School of Medicine, New Haven, CT USA Andrew Smith, PhD, is principal scientist at Hologic, Inc. in Bedford, Mass, where he is involved in research and development in digital imaging systems. He attended the Massachusetts Institute of Technology, where he received a bachelor and doctoral degrees in physics.

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 7 Clinical Examples Tubulolobular adenocarcinoma not seen well in the conventional digital mammogram (left image) is much better appreciated in the slice from the tomosynthesis data set (right image) Suspicious areas in digital mammography (left image) can be resolved with tomosynthesis (right image) Infiltrating lobular carcinoma and multifocality not seen well in the conventional digital mammogram (left image) are clearer in the slice from the tomosynthesis data set (right image) Mammographically occult cancers (left image) is visible with tomosynthesis (right image)

78967-12_WP-00007_FundmntlsTomo 10/3/12 3:47 PM Page 8 Hologic, Inc. 35 Crosby Drive Bedford, MA 01730 U.S.A. T: 781.999.7300 www.hologic.com womenshealth@hologic.com WP-00007 June 08