Title: Standard Operating Procedure Measurement of Wind Speed and Direction Anemometer & Vane System



Similar documents
METEOROLOGICAL INSTRUMENTS

APPENDIX A. Bay Area Air Quality Management District

Installation Guide. WSD-100 Wind Speed and Direction Sensor For XR5 Data Loggers. February, 2011

Wind Direction Smart Sensor (S-WDA-M003)

User manual data files meteorological mast NoordzeeWind

WEATHER STATION FOR SOLAR FARM MONITORING

Series 6000 Torque measured metal bellow coupling

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

EDUMECH Mechatronic Instructional Systems. Ball on Beam System

Silicon Pyranometer Smart Sensor (Part # S-LIB-M003)

TR SUPPLEMENTARY REQUIREMENTS FOR THE ACCREDITATION OF CONTINUOUS AMBIENT AIR QUALITY MONITORING STATIONS

Servo Info and Centering

NSW Air Quality Monitoring Network: Quality Assurance Procedures

Wind Sensor W-RS485 with RS485 Interface

Accurate Air Flow Measurement in Electronics Cooling

A.1 Sensor Calibration Considerations

Description. Dimensions. Features. precision works better

LDG M-7700 External Meter for Icom IC-7700

Mounting Tripod Kit Installation Manual

Sense it! Connect it! Bus it! Solve it! EncoderS

WEB log. Device connection plans

WEB log. Device connection plans

Solution: Angular velocity in consistent units (Table 8.1): Velocity of a point on the disk: Rate at which bits pass by the read/write head:

Current Loop Tuning Procedure. Servo Drive Current Loop Tuning Procedure (intended for Analog input PWM output servo drives) General Procedure AN-015

Autonomous Mobile Robot-I

AXIS T81B22 DC 30W Midspan

Basic Principles of Inertial Navigation. Seminar on inertial navigation systems Tampere University of Technology

QR12 (1.22 ) Diameter Optical Encoder

* DISCLAIMER: Contents. How to Use This Guide: COMMERCIAL INSTALL GUIDE 2

Micrio WS1 Replacement Wind Speed Sensor and WC1 Replacement Wind Compass Sensor for Raymarine ST50 and ST60 Wind Instruments. Rev 4.

Reaction Torque Sensor

Advantages of Auto-tuning for Servo-motors

Electric Motors and Drives

RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G

TELESCOPE Motorized Lighting Suspension

Solar Tracking Application

FIRERAY 2000 Installation Guide

T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE

, User s Manual

Model SRMD Setra Remote Monitoring Display

MoTeC USA GPS. Part # M GPS BL Available in 10 Hz or 20 Hz. USER MANUAL Version 1.4

Fan Coil EC Motor Control

EVALUATION OF ZEPHIR

Mercury TM 1500P PCB-Mount Digital Encoders

Cross-beam scanning system to detect slim objects. 100 mm in

Agilent Automotive Power Window Regulator Testing. Application Note

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Mobile field balancing reduces vibrations in energy and power plants. Published in VGB PowerTech 07/2012

Air Monitoring Directive Summary of Feedback and Responses for Chapter 6 (Ambient Data Quality)

The Fan-Aspirated ISS includes these components: Metric Rain Adapter. The hardware shown here is provided for assembly and mounting:

Series 7500 Torque Sensor

The self-starting solar-powered Stirling engine

Technical Guide No High Performance Drives -- speed and torque regulation

What Is an Electric Motor? How Does a Rotation Sensor Work?

Electronic Power Control

Pump Skid Fabrication for Magnetic Coupling. Rick Soltis Chief Mechanic City of Bedford

Using Arduino Microcontrollers to Sense DC Motor Speed and Position

THE MECHANICS OF TENSION CONTROL By Jeff Damour CONVERTER ACCESSORY COPORATION Wind Gap, PA USA

Operation Manual For NDJ-8S Digital Rotary Viscometer

Electric Rotary Actuator - On/Off or programmable control actuator

Research-Grade Research-Grade. Capture

K-Type Thermocouple Sensor User s Guide

Typical ECM/PCM Inputs

Radiant Temperature Sensor TY7321

ABB Drives. User s Manual. Pulse Encoder Interface Module RTAC-01

ENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013

Three Channel Optical Incremental Encoder Modules Technical Data

TRIMBLE ATS TOTAL STATION ADVANCED TRACKING SYSTEMS FOR HIGH-PRECISION CONSTRUCTION APPLICATIONS

D.C. Motors. Products and specifications subject to change without notice.

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

American Society of Agricultural and Biological Engineers

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

Figure 1: Multiple unsynchronized snapshots of the same sinusoidal signal.

G-PORTER. Portable GPS Tracker GP-102 User s Manual

IP Box Camera ACM Ver Hardware User s Manual

Subminiature Load Cell Model 8417

Chen. Vibration Motor. Application note

Agilent U1610/20A Handheld Digital Oscilloscope Quick Start Guide

Length measuring instruments

BLWR23MDA Series. 24V, 15A Brushless Controller / Motor. User s Guide. 910 East Orangefair Lane, Anaheim, CA info@anaheimautomation.

FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases

Installation Guide. Wyse D Class 4-GB RAM Upgrade Option Kit. Products: D90D7, D90DW. Issue: PN: L Rev. A

Wireless Data Logger Monitoring Systems

Servo Motors (SensorDAQ only) Evaluation copy. Vernier Digital Control Unit (DCU) LabQuest or LabPro power supply

Successful Implementation of an Alternative Co-located Transfer Standard Audit Approach: Continuous Deployment of CTS Wind Sensors on a Tall Tower

Application case - Solar tracking system

Trigno/Vicon System Integration

Electrical Resonance

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Once you know the volume of air and the static pressure of the system to be cooled, you can determine the fan specifications for your product.

OPL BASIC. Dosing System for Professional Laundry machines. Contents

Introduction to Linear Actuators: Precision Linear Motion Accomplished Easily and Economically

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002

Digital vs. Analogue Control Systems

Selecting and Sizing Ball Screw Drives

SFS81 Series. Servo Finishing Stand PUSHCORP, INC. Dallas, Texas

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Meteorological SPACE WEATHER SPECIAL! BRITISH ANTARCTIC SURVEY The meteorological capabilities and work of the BAS explained

T0118 T2118 T3118. Instruction Manual

Transcription:

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 1 of 8 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, meteorological and climate trend reporting and to meet the requirements for inclusion of the data in provincial and national meteorological databases, a continuous method of measuring Wind Speed and Direction is used. This method is capable of measurement updates at a rate of once every five minutes or faster. Readings from instruments of this method enables the calculation of hourly averaged measurements of Wind Speed and Direction. Commercially available anemometer and vane Wind Speed and Direction sensors are used in the method. This method is applicable to the measurement of Wind Speed and Direction in ambient air in the range of 160 kilometers per hour (kph) or 50 meters per second (m/s) and 0 to 360. This method adheres to the requirements of the current Air Monitoring Directive (AMD) drafted by Alberta Environment in 1979. In some cases the limits and specifications exceed the requirements of the current AMD and subsequent amendments. It should be considered that the current and any future amendments or drafts of the AMD will be used as the benchmark for requirements and criteria for ambient air monitoring practices conducted in the Province of Alberta. Information used to write this procedure was also taken from sources identified in the reference section. 2. PRINCIPLE OF THE METHOD The method used to determine Wind Speed for this procedure is rotation of a vertical shaft connected to an anemometer. The anemometer is typically configured as three cups mounted to a hub. The other end of the vertical shaft is connected to a photochopper which generates an electronic signal directly proportional to wind speed. Horizontal air movement pushes on the cups of the anemometer, which causes the shaft to rotate generating an electrical signal with frequency proportional to the speed of air movement. The signal is introduced to a translator which converts the frequency to an analog signal, typically voltage. The translator can be configured from the factory to provide an output in either meters per second, kilometers per hour or miles per hour. The method used to measure Wind Direction for this procedure is the flat plate vane which is a direct sensing method. The flat plate vane is connected via a vertical shaft to a precision wire-wound potentiometer capable of rotating through 360 as the sensing element. In operation, a reference voltage is applied across the resistance element of the potentiometer and wind direction becomes a function of voltage signal of the wiper of the potentiometer. The output signal is directly proportional to the azimuth angle. The

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 2 of 8 wind direction sensing device must be adjusted to compensate for inclination or declination from magnetic north. 3. MEASUREMENT RANGE AND SENSITIVITY The Wind Speed and Direction sensors used in this method are commercially available models. The measurement range of wind speed is typically 0 to 160 kilometers per hour (kph) and wind direction is 0 to 360 (degrees). The detection limit of the wind speed and direction sensors is specified by the manufacturer. For the wind speed sensor, a threshold determination is conducted on the size and design of the anemometer used in a wind tunnel. This threshold is the amount of wind speed required to start the anemometer in motion from a calm condition. For wind direction the process is similar, however, the threshold for the vane is much lower due to the physics of the sensor configuration. Typically the thresholds are 0.8 kph for wind speed and 1.0 degrees for wind direction. 4. EQUIPMENT AND APPARATUS The following are commercially available sensors suitable for use in this method which are currently used in the AENV Network: Blue Sky Instruments Model 857 Wind Speed and Direction Sensor Athabasca Research Model 540-E1 Wind Speed and Direction Sensor Met One Model 024A and 034B RM Young Model 5103V 5. INTERFERENCES Interferences with measurement accuracy can be caused by improper location of the sensors relative to objects which affect airflow. Bearing friction reduces accuracy and threshold detection. Extreme cold and ice and snow build up can cause inaccuracy of measurements or render the sensors totally unresponsive.

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 3 of 8 6. PRECISION AND ACCURACY The measurement precision is generally considered to be the repeatability of the measurement. Precision is established by the manufacturer, but confirmed during calibration checks. The accuracy of the sensor is generally considered the deviation from true. This means how close it is to what it should be. The benchmark of what it should be is provided by Alberta Environment. In calibrating the sensor, accuracy is also dependent on the accuracy of the calibration transfer standard used to calibrate the wind speed sensor and its translator. The determination of uncertainty of measurements has been done using the Type A protocol specified by the Canadian Association for Environmental Analytical Laboratories (CAEAL) to comply with the requirements of CAN- P-4D (ISO/IEC 17025). The uncertainty of measurement estimate is the result the statistical analysis in which the root sum of squares (square root of the sum of the squares) of the variables is determined. For Wind Speed the variables are: Transfer standard precision 1.0 kph Sensor precision 0.8 kph The resultant uncertainty is 1.0 kph For Wind Direction the variables are: Transfer standard precision 2 (Degrees) Sensor precision 3 (Degrees) The resultant uncertainty is 3.6 (Degrees) 7. SITE REQUIREMENTS Site location of the Wind Speed and Direction monitoring station should be determined according to the intended application of the monitoring data. The detailed requirements for selection of sites for monitoring ambient Wind Speed and Direction in Alberta can be found in the Air Monitoring Directive. Wind Speed and Direction sensors are sited and operated in accordance with the guidelines outlined by the AES Guidelines for Co-operative Climatic Autostations Guide 89-1.

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 4 of 8 8. INSTALLATION REQUIREMENTS All the installation requirements as specified by the manufacturer in the installation procedures as well as the general requirements below must be followed. 8.1 Trees Trees and vegetation can cause disturbance of airflow patterns which affect Wind Speed and Direction measurements. To minimize this effect, the distance between the sample inlet probe of the Wind Speed and Direction sensors and the drip line of the tree must be at least twice the height of the tree above the inlet probe or at least 10 metres away whichever is greater. 8.2 Other spacing requirements for the wind sensors such as height above ground, distance from obstructions and distance from horizontal or vertical support are as follows: Height above ground 10 to 15 metres Distance from support structure > 2 metres Distance from any air flow obstacle, i.e. buildings > 2 height of obstacle above the sensor Airflow obstructions unobstructed airflow in all directions for more than 2 metres. 8.3 A data acquisition system is required to record or download the signal output from the translator. 9. OPERATIONAL REQUIREMENTS The following table of activities must be performed when operating continuous Wind Speed and Direction sensors in the AENV network. The following documentation must be available to the operators on site: operational and maintenance manual(s), quality system documentation and station site documentation.

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 5 of 8 Table 2 Analyzer Operational Requirements Action Set sensor and data system range (WS 120 kph and WD 360 ) Verification of operational parameters Sensor maintenance Calibration Time and Frequency After installation Procedure Documentation Action by As per manufacturer s manual Entry to log book Station operator Weekly Visual Check Entry to log book Station operator As recommended by manufacturer After installation or repair and as required thereafter Yearly As per manufacturer s manual Remove sensor and translator and calibrate Entry to log book and Work Order Entry to log book; and Work Order Instrument Technician Instrument Technician 10. CALIBRATION Calibrations of the entire wind system must be completed annually at a minimum. This procedure involves removing the sensor from the tower or mast where it is installed, verifying the mechanics are functional and completing a calibration by spinning the anemometer shaft at set frequencies, and moving the wind direction shaft to set increments throughout the 360 degrees of travel. The system manuals will provide necessary information on; corresponding wind speeds for frequency set points, system ranges, and calibration procedures. The minimum required procedure is outlined below.

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 6 of 8 10.1 Before any hardware is disturbed, visually verify the operation of the wind speed and direction sensors by comparison to the data collected by the data system. Ensure all cables and connections from the sensors to the translators and data collection system are secure. Prior to removing the wind head from the tower, the alignment of the head must be verified. This can be completed using either a magnetic compass or the solar noon method. To use a magnetic compass, adjust the bezel of the compass to the appropriate declination depending on the co-ordinates of the sensor. Check the alignment of the crossarm to true north, not magnetic north. If it is not aligned correctly, document the deviation from true north so that collected data can be adjusted. 10.2 The following equipment is required for a wind system calibration: Wind speed sensor calibrator DC motor with optical feedback Wind direction sensor calibration template. Torque wheel 10.3 Flag the channels collecting the data for the wind speed and direction system as down for calibration. Arrange to record the data from the calibration using the data system or strip chart recorder. If the data system is used to collect the calibration data, it must be set to record in a small enough increment, typically 30 seconds or less. Also prepare a calibration form to record all parameters of the calibration. A sample form is attached below. 10.4 Verify that the starting wind speed is within the tolerance by attaching the torque wheel to the wind speed shaft. Holding the sensor horizontally with the screw on the appropriate space, the wheel should fall freely to the bottom. If it does not, the bearings should be replaced. This procedure is repeated for the wind direction shaft to ensure the bearings are within specs. 10.5 Connect the wind speed motor to the wind speed shaft using the appropriate connector. Spin the shaft at predetermined setpoints within the range of the sensor. The setpoints can be determined using the RPM vs WS tables found in the operations manual. If there are no tables available, an equation to calculate WS for RPM would be available. Each setpoint should be maintained for a length of time long enough to verify stability of the system. 5 to 10 minutes is typical.

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 7 of 8 10.6 Connect the wind direction template to the outer housing of the wind direction sensor. Then connect the pointer to the wind direction shaft. While monitoring the data system used to collect the data in real time, align the pointer and plate so that the wind direction is reading 360 degrees and secure both in place. Then rotate the shaft and pointer to achieve predetermined set points. Each setpoint should be maintained for a length of time long enough to verify stability of the system. 5 to 10 minutes is typical 10.7 Document the results of the calibration using the appropriate calibration form. 10.8 Attach a label at a visible location on the instrument (typically the translator so that it is sheltered and will not weather) indicating the date of most recent calibration. 10.9 Return the wind head to the tower insuring the wind head is aligned to true north as described above. 11. APPLICABLE DOCUMENTS EM-018a Met One Model 010C Wind Speed Sensor Operating Manual EM-018b Met One Model 020C Wind Direction Sensor Operating Manual EM-018c Environment Canada AES Guidelines for Co-operative Climatic Autostations: Guide 89-1 June 30, 1989 12. LITERATURE REFERENCES Lodge. 1991. Methods of Air Sampling and Analysis, 3 rd edition. Guidance Document on Achievement Determination-Canada Wide Standards for Particulate Matter and Ozone. ISBN: 1-896997-41-4 PN 1330, October 2002, Canadian Council of Ministers of the Environment. Quality Assurance and Quality Control Guidelines - National Air Pollution Surveillance Network Report Series No. AAQD 2004-1 Blue Sky Instrument Model 805 Air Quality Wind System 8059001 June 1995

Procedure No: SOP-018 Revision No: 0 (new document) Page No.: 8 of 8 13. REVISION HISTORY Revision 0 (new document) Reviewed December 29, 2010 14. APPROVAL Approved by: Harry Benders Date: June 20, 2008 Title: Air Monitoring Team Leader