LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037



Similar documents
State of Illinois Illinois Emergency Management Agency. Radon Quality Assurance Program Guidance

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

1 Quality Assurance and Quality Control Project Plan

Home Safe Home Inspections

Ecology Quality Assurance Glossary

4 QUALITY ASSURANCE 4.1 INTRODUCTION

Control Charts and Trend Analysis for ISO Speakers: New York State Food Laboratory s Quality Assurance Team

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10

Quality Assurance/Quality Control in Acid Deposition Monitoring

Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs

APPENDIX N. Data Validation Using Data Descriptors

OPERATIONAL STANDARD

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN

US EPA REGION III QUALITY MANAGEMENT PLAN REVIEW CHECKLIST

Ramon 2.2 Radon-Monitor

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

A.1 Sensor Calibration Considerations

QA Transactions and Reports

Best Practices for Radon Measurement in Minnesota Schools and Commercial Buildings

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

Quality Assurance For Environmental Laboratories

- 2 - Protocols for Radon and Radon Decay Product Measurements in Homes. Office of Air and Radiation (6609J) EPA 402-R , June 1993.

FREQUENTLY ASKED QUESTIONS

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

National Radon Proficiency Program. Guidance on Quality Assurance

American Association for Laboratory Accreditation

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3

How to Verify Performance Specifications

DRY CLEANING PROGRAM QUALITY ASSURANCE MANAGEMENT PLAN

DIVISION OF ENVIRONMENT QUALITY MANAGEMENT PLAN PART III: SUPERFUND (MACA) PROGRAM QUALITY ASSURANCE MANAGEMENT PLAN

ASSURING THE QUALITY OF TEST RESULTS

Quality Assurance Testing of AtomLab Dose Calibrators

US Environmental Protection Agency Office of Pesticide Programs

STUDY #2 IN A SERIES OF REPORTS ON RADON IN BC HOMES. Prince George: COMMUNITY-WIDE RADON TESTING RESULTS AS PART OF ITS PROGRAM

ETV Joint Verification Statement

Analytical Test Method Validation Report Template

Minnesota Pollution Control Agency. Quality Assurance Project Plan Guidance

Air Quality Performance Test Guidelines. Kansas Department of Health and Environment Bureau of Air and Radiation

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

PERFORMANCE EVALUATION AUDIT CHECKLIST EXAMPLE. EIIP Volume VI

Data Quality Objectives and Quality Assurance Project

Chapter Seven. Multiple regression An introduction to multiple regression Performing a multiple regression on SPSS

Implementing New USP Chapters for Analytical Method Validation

PROCEDURES FOR HANDLING OOS RESULTS

Examples: collecting the right sample on the right patient; labeling the sample accurately;

QUALITY ASSURANCE and QUALITY CONTROL DATA STANDARD

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

Working Copy - 1,- / t&,1 r. Responsibilities of the Laboratory Quality Assurance Officer. EffectiveDate: 7/ltl/ l'rritials: tuwt

Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software

Procedure for Equipment Calibration and Maintenance

STANDARD OPERATING PROCEDURE NO. 66 GAS ANALYZER

U. S. Department of Energy Consolidated Audit Program Checklist 5 Laboratory Information Management Systems Electronic Data Management

ISO/IEC QUALITY MANUAL

Conducting Effective OOS or OOT Investigations for Unexpected Results from the BET Assay

Third Party Data Validation: What you need to know and do!

Validation and Calibration. Definitions and Terminology

Data Management Implementation Plan

Continuous Monitoring Manual

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

To avoid potential inspection

Texas Commission on Environmental Quality Page 1 Chapter Control of Air Pollution from Nitrogen Compounds

FAILURE INVESTIGATION AND ROOT CAUSE ANALYSIS

Standard Deviation Estimator

1.0 INTRODUCTION 1.1 Overview

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

QUALITY ASSURANCE PROJECT PLAN FOR BROWNFIELDS/VOLUNTARY CLEANUP PROGRAM SITES

Maine Home Buyer s and Seller s Guide to Radon

PROCEDURE 1. QUALITY ASSURANCE REQUIREMENTS FOR GAS CONTINUOUS EMISSION MONITORING SYSTEMS USED FOR COMPLIANCE DETERMINATION

Dealing With RADON in Real Estate Transactions

OH&S MANAGEMENT SYSTEM CHECKLIST - AS 4801:2001 (STATUS A = Acceptable; N = Not Acceptable; N/A = Not Applicable)

CHAPTER ONE TABLE OF CONTENTS 1.0 INTRODUCTION... 1

STATE OF MISSOURI, DEPARTMENT OF NATURAL RESOURCES MISSOURI AIR CONSERVATION COMMISSION PERMIT TO CONSTRUCT

Experiment #1, Analyze Data using Excel, Calculator and Graphs.

Quality Assurance Procedures for Construction and Materials Sampling and Testing 14

MIDLAND AREA SOIL SAMPLING MIDLAND, MICHIGAN

HOW ACCURATE ARE THOSE THERMOCOUPLES?

A Citizen s Guide To Radon

North American Development Bank. Bid Evaluation Procedures

Transcription:

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting Method Number: ELAP method code 7037 SOP Number: Revision Number: SOP Date: Personnel / Data Records observed: CRM Model Number(s): General LLD: < 1 pci/l Page 1 of 7

RADON IN AIR Page 2 of 7 NYSDOH Environmental Laboratory Approval Program Radon Checklist 1. Are both of the following EPA publications available at the lab: a. Protocols for Radon and Radon Decay Product Measurements in Homes? (EPA 402-R92-003) N002 b. Indoor Radon and Radon Decay Product Measurement Device Protocols? (EPA 402-R92-004) N003 2. Does the facility monitor Radon levels in air for a minimum of 48 continuous hours, process the collective data and interpret the result N004 for the client (rather than merely distributing devices)? 3. Is a written standard operating procedure (SOP) available? (Duplication of the protocols will not suffice.) N005 a. Does the analyst follow the manufacturer s instructions, the US EPA protocols and the lab s SOP? N072 4. Is a written quality assurance project plan (QAPP) appropriate to each device available? N006 5. Does the data collection log include: a. Date and time of deployment? N007 b. Date and time of removal? N008 c. Condition of the devices? N009 d. Attainment of closed building conditions? N010 1. Delayed start, if implemented N010A e. Exact location of device (i.e., building, room and sampling position) within the property being monitored? N011 f. Serial number, model number, and manufacturer of the detector? N012 g. Unique identifier of client and client s address (if different from testing location)? N013 h. Condition of crawl space vents? N014 i. Performance of a Routine Instrument Check (to include, but not limited to, zeroing the integrator, ensuring correct date N253 and time indications, and fully charged battery)? 6. Are monitoring devices deployed and retrieved by trained employees of the radon laboratory? N015 7. When monitoring devices are set out and picked up by different N016

persons, is this noted in the record? 8. Is a qualifying statement used if the lab does not control the monitoring process (i.e., placed and/or retrieved by persons other N017 than trained lab personnel)? 9. Are all devices deployed for 48 continuous hours if the measurement is used for remedial action or to determine the need for further measurements? N018 a. In cases where closed conditions have not been met, is there a 12-hour delayed start or extended monitoring period of at least 60 hours after closed conditions have been implemented? 10. Is the technical director of the laboratory aware of the device s Lower Limit of Detection (LLD), and has the background level been determined? FIELD DATA N021 N019 1. Is the field data record (instrument field sheet) maintained? N166 QUALITY CONTROL 1. Is the device an US EPA approved unit? N251 Record CRM Model number. 2. Is the device capable of detecting at least 1 pci/l? N252 3. Have all elements of the QC protocol been addressed in the QA Plan? a) background reading b) accuracy relative to a known reference standard via annual calibration c) precision of duplicates d) cross-checks SYSTEM CALIBRATION and CROSS CHECKS N171 1. Has either the manufacturer or the user (through an approved calibration service provider) calibrated the complete system in a N177 radon chamber before being put into service and after any repairs or modifications? 2. Has the calibration been performed at least annually? N178 Page 3 of 7

3. Is a record of annual calibrations available? N179 a. Does the calibration account for instrument adjustments, background measurements, and use of traceable standards? 4. Is the detector cross checked semi-annually against another recently calibrated instrument? a. Is a record of the semi-annual cross check available? 5. Do all detectors have individual calibration (correction) factors? N179A N180 N180A N182 a. Are calibration (correction) factors applied correctly to each radon monitoring job? BACKGROUND MEASUREMENTS 1. Is the background count measured in a low radon environment every 20th 48-hour measurement? N182A N185 2. Is the background measured below 2 pci/l? N184 DUPLICATE MEASUREMENTS 1. If two or more detectors are available, are duplicate measurements taken? a. Is the side-by-side measurement performed at a frequency of one (1) in ten (10) analyses? b. Is the side-by-side measurement available for review? 2. Are duplicate measurements made in at least 10% of the total number of measurement locations, or 50 each month, whichever is smaller? a. If field duplicates are taken, are the field locations selected for duplicate measurements distributed systematically throughout the entire population of samples? 3. Is the precision of the duplicate data from radon levels measured at 4 pci/l or above within 10% Relative Standard Deviation (RSD: for a long-term series of duplicate measurements), or 14% Relative Percent Difference (RPD: for a single pair of duplicates)? N187 N181 N181A N188 N193 N189 Page 4 of 7

Page 5 of 7 NYSDOH Environmental Laboratory Approval Program Radon Checklist 4. Are the data from duplicates available (e.g., control charts)? N190 5. Are the acceptance limits for duplicates determined? N191 RECORD KEEPING 1. Are the following record(s) kept for 5 years by the measurement organization: a. A copy of the final report including measurement results and the statement outlining any recommendations concerning N237 retesting or mitigation provided to the building occupant or agent? b. The address of the building monitored including zip code? N238 c. The exact locations of all deployed measurement devices? (Examples: room sketches, detailed written description) N239 d. Exact start and stop dates and times of the measurement period required for analysis? N240 e. A description of the device used including the device identification number and serial number, if applicable? N241 f. A description of the condition of any permanent vents? N242 g. The name of the service organization used for calibrations and the certificates of calibration? h. The name of the individual who conducted the test along with their associated training records? i. A description of any variations from or uncertainties about standard measurement procedures, closed building conditions, or other factors that may affect the measurement result? j. A description of any non-interference controls and copies of signed non-interference agreements? k. A record of any quality control measurements associated with the test? l. The written authorization from the client enabling the lab to report results to a third party (if applicable)? REPORTING 1. Is the US EPA average reported to the client? N260 First four (4) hours is omitted. N243 N244 N245 N246 N247 N254

a. Are results reported to the client with % uncertainty? N261 2. Is the %Uncertainty determined by the following calculation: %Uncertainty = 100% X [1/(SQRT of counts)], where counts are N250 determined as the [cph per 1pCI/L)] x (radon level measured expressed in pci/l x EPA avg hrs of test) a. Are reported results having uncertainties greater than 10% qualified? N255 NXXXNXXX b. Do reported values reflect the minimum reportable concentration? (example: For Sun Nuclear #1027 and #1028, this level would be 0.9 pci/l for the EPA average of a 48 hour sample.) 3. Does the report contain the advisory of 10 NYCRR 16.130 for properties tested within New York State having measured radon gas levels >20 pci/l or 0.1 working level? 4. For labs within NYS, does the laboratory report all of its radon monitoring data to the State according to 10 NYCRR 16.130? 5. Are field duplicate readings, if available, reported in an acceptable manner? (See EPA 402R92-003). N256 N257 N258 N259 10 NYCRR 16.130 Radon testing and reporting. (b) General requirements. (2) When any radon screening or long-term testing result exceeds 20 pci/l or 0. 1 working level as defined in section 16.2(a)(145) of this Part, the radon testing firm shall advise the customer, if a resident of this State, in writing to contact the New York State Department of Health, Bureau of Environmental Radiation Protection, for further technical advice and assistance. (3) A radon mitigation firm shall report to the department in writing, within 30 days following the end of a reporting period, a summary of the number of homes mitigated in the State during that reporting period. Reporting periods shall be from January 1st to June 30th and July 1st to December 31st of each year. The report shall include the dates for which the report is made and the number of buildings for which mitigation was performed for the reporting period in each county or ZIP code area in the State. **The relative sensitivities per 1 pci/l for commonly encountered Rn monitors are as follows: 2.5 cph for Sun Nuclear Model #1027 and #1028 3.0 cph for Sun Nuclear Model #1030 (to be confirmed) 18 cph for Femtotech, 69 cph for Pylon CRMs Page 6 of 7

NOTES: (The following paragraphs are from the updated EPA 402R92-004 Indoor Radon and Radon Decay Product Measurement Device Protocols ) Users of continuous monitors must perform sufficient instrument background measurements to establish a reliable instrument background and to act as a check on instrument operation. The two previous editions of these protocols (U.S. EPA 1986, 1989a) used the value coefficient of variation (COV), defined as the standard deviation divided by the mean, as the expression used for the goal (at 4 pci/l or 0.02 WL) of 10 percent for precision. The COV should decrease with increasing concentration. This edition explains that there is a variety of ways to calculate and express precision, including the COV and the relative percent difference, defined as the difference between two duplicates divided by their mean. It is important to monitor precision over the entire range of radon levels that are encountered routinely in the measurement program, and that a systematic and documented method for evaluating changes in precision be part of the standard operating procedures. While a limited precision error is desirable (e.g., COV of < 10% at 4 pci/l), it is most important to maintain the total error of any individual device (including both errors in precision and accuracy) to within ± 25 percent of the "true" radon or decay product concentration for concentrations at or above 4 pci/l). It is important to emphasize to the technical director of the Rn lab the DIFFERENT MEANINGS between the 10% Rel. Std. Deviation (or COV as mentioned in the paragraph above) and the simple 14% RPD. The 10% RSD limit is calculated from many measurements (20 minimum) taken over an extended period of time. It is NOT appropriate to try to evaluate this number based upon the readings of a single pair of duplicate measurements. In real terms, it is a number used for evaluating the LONG TERM PRECISION, taken over AT LEAST 20 MONITORING EVENTS, using co-located devices. In mathematical terms, it is the typical or standard deviation calculated from ALL THE DIFFERENCES (may use raw difference or relative percent difference) BETWEEN ALL THE INDIVIDUAL PAIRS of duplicate measurements. The standard deviation is therefore derived from many events, at many different Rn levels, using duplicate measurements taken by more than one device during several deployments. The 14% RPD limit is a simple calculation based solely upon the difference between ONE PAIR of readings taken during ONLY ONE monitoring event. Page 7 of 7