Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered



Similar documents
Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation safety in dental radiography

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Radiation safety in dental radiography

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

X-Mind. Instinct for perfection

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Production of X-rays and Interactions of X-rays with Matter

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

RADIATION SAFETY REQUIREMENTS FOR USERS OF X-RAY IN HEALING ARTS OR SERVICERS OF X-RAY EQUIPMENT

X-ray Imaging Systems

Personal Dosimetry in Dental Radiology

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

Agency for Health Care Administration

JB-70 Dental X-Ray System

Elements of required physical infrastructures: space, schielding, and patient flow..

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Radiation Safety Issues for Radiologic Technologists

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

portable x-ray survey report

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

Staff Doses & Practical Radiation Protection in DEXA

X-RAY REGULATORY GUIDE

Radiation Protection Update

Digital Breast Tomosynthesis QC Requirements

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Cutting Edge Digital Radiography Technology in Daily NDT Use

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

The University of Southern Maine. XRF X-Ray Radiation Protection Program

Physics testing of image detectors

Quality Control and Maintenance Programs

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Required RS Training Info

Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health

WARNING: We recommend that you consult the Safety, Regulatory and Technical Specification User Guide before using the CS 2200.

Introduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

X-ray Safety Manual 2013

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Radiation Safety In Dental Practice

RADIATION SAFETY STANDARD

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

X-ray Imaging Systems

PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS

Radiation Safety Committee College of William and Mary

Diagnostic X-ray facilities as per quality control performances in Tanzania

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

Radiation Safety In Dental Practice

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Analytical X-ray Safety Refresh Training. Radiation Safety Officer Department of Environmental Health and Safety

Portable X-ray System for Intraoral Radiographic Imaging USER MANUAL

CODE OF PRACTICE & SAFETY GUIDE

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Radiation protection principles observance in Iranian dental schools

The Nuclear Safety Council, in exercise of the functions conferred to

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Dose Standards and Methods for Protection. Radiation and Contamination

LOUISIANA STATE UNIVERSITY SCHOOL OF DENTISTRY INSTITUTIONAL POLICY FOR THE DIAGNOSTIC USE OF IONIZING RADIATION

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Atomic and Nuclear Physics Laboratory (Physics 4780)

Radiation Dose Measurements in Routine X ray Examinations

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

Checklist for submission of application form for Registration

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Performance testing for Precision 500D Classical R/F System

Quality Control of Full Field Digital Mammography Units

Safety and Radiation Protection Office Working with X-Ray Equipment

Digital Dental X-ray. Midmark Animal Health Products

AAPM protocol for kv x-ray beam dosimetry in radiotherapy and radiobiology

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

INTRODUCTION. A. Purpose

EQUIPMENT REQUIREMENTS AND QUALITY CONTROL FOR MAMMOGRAPHY

PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT TECHNICAL MANUAL /11 LAT LAT SIN DEX. kv kv. mas. mas CTL

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

Transcription:

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC John Ludlow, DDS School of Dentistry, University of North Carolina *Consultant to Aribex, Inc., Orem, Utah Presented at the 41st Annual Conference on Radiation Control Conference of Radiation Control Program Directors Columbus, Ohio May 18-21, 2009 ABSTRACT A hand-held, battery-powered dental intraoral x-ray system (60 kv, constant potential output) was compared to a conventional, wall-mounted intraoral x-ray system (70 kvp, self-rectified output) in terms of image quality, and patient and staff radiation doses. The image quality comparison included quantitative measurements of image sharpness (resolution) and contrast. Patient doses were compared using the FDA dental phantom and adjusting the radiation dose to obtain the same density on intraoral dental films. Staff radiation doses were measured using personal dosimetry badges for dental facilities before and after introduction of the hand-held x-ray system allowing accurate comparison of staff doses with both systems under similar workloads and operating conditions. Results for both image quality, and dose to patients and staff are provided. 1

Apologies Our apologies for using the brand name of a commercial product so frequently At onset of this project we were referring to this device as a hand-held device However, other hand-held devices are coming into the market place these new devices may vary significantly from the device evaluated INTRODUCTION The authors apologize for the use of a brand name of a commercial product so frequently. At the onset of this project we were referring to this device as a handheld device. However, other hand-held devices are coming into the market place. These new devices may vary significantly from the device evaluated. 2

Purpose Compare Nomad intraoral x-ray x systems to wall-mount systems in terms of Image resolution Contrast Half Value Level (HVL) Patient dose Scattered radiation X-ray tube leakage Dose to operators INTRODUCTION This paper compares the Nomad (Aribex, Orem, Utah) intraoral x-ray system to conventional, wall-mounted systems in terms of image resolution and contrast, the half-value layer (HVL) of the beam, patient dose, scattered radiation, x-ray tube leakage, and dose to the operators. MATERIALS AND METHODS SYSTEMS COMPARED The hand-held, intraoral x-ray system evaluated was the Nomad. For image quality and patient radiation dose purposes this unit was compared to Gendex GX-770 (Gendex Dental Systems, Lake Zurich, Illinois). The specifications for these systems are provided in Slide 8. The Nomad is specifically designed as a hand-held x-ray device. Consequently, special design features have been incorporated including: Increased shielding around the x-ray tube Built-in, integral leaded acrylic shield to protect the user from backscattered radiation Shielded position indicating device (PID), or collimator. 3

Background Hand-held units are coming into use in North America Several firms are manufacturing hand-held held overseas and beginning to market these in the U.S.A. Over 3000 units are in use today in the U.S. in dental radiography, veterinary medicine, forensic, military, and research applications Regulatory concerns have been expressed about these devices INTRODUCTION Hand-held, battery-powered x-ray units are coming into use in North America. Over 3,000 units are in use today in the U.S. in dental radiography, veterinary medicine, forensic, military, and research applications. 4

Concerns Image quality Perception of lower kvp Patient dose Operator exposure from the x-ray x tube and scattered radiation INTRODUCTION Regulatory concerns have been expressed about the use of these devices including issues about image quality, patient dose, operator radiation dose from x-ray tube leakage and scatter, and the perception of these units using lower kilovoltage. 5

Nomad Intraoral Dental System Nomad is designed as hand-held held x-ray x device Special design features include Increased x-ray x shielding around the x-ray tube Built-in in integral leaded acrylic shield to protect operator from backscatter Shielded position indicating device (PID) or cone SYSTEMS COMPARED The hand-held, intraoral x-ray system evaluated was the Nomad. For image quality and patient radiation dose purposes this unit was compared to Gendex GX-770 (Gendex Dental Systems, Lake Zurich, Illinois). The specifications for these systems are provided in Slide 8. The Nomad is specifically designed as a hand-held x-ray device. Consequently, special design features have been incorporated including: Increased shielding around the x-ray tube Built-in, integral leaded acrylic shield to protect the user from backscattered radiation Shielded position indicating device (PID), or collimator. 6

Comparison Units Hand-held system Nomad Aribex, Inc. Orem, Utah Wall-mount system GX-770 Gendex Dental Systems Lake Zurich, Illinois SYSTEMS COMPARED [Continued] 7

Kilovoltage Specifications Nomad 60 (Constant Potential) Wall- Mount 70 (Single phase) ma 2.3 7 Typical exposure time (F-Speed Film) in sec 0.34 0.17 PID diameter (cm) 6 7 Source-to-cone tip (in) 8 8 Focal spot size (mm) 0.4 0.6 SYSTEMS COMPARED [Continued] 8

X-Ray Waveforms Conventional Alternating voltage output is 70 kvp Average energy approximately 56 kv Constant potential generators (CPG), also known as DC, provide the same, constant kilovoltage 60 kv is 60 kv SYSTEMS COMPARED [Continued] METHODS Calculated x-ray waveforms were compared. The kilovoltage waveform for the Nomad was modeled as a constant-potential waveform (Slide 10) with the average kilovoltage being that specified by the manufacturer, i.e., 60 kv. The kilovoltage waveform was modeled as a single-phase waveform for the conventional system (Slide 10) with the peak kilovoltage of 70 kvp. A filter was added to the beam to produce a filtered waveform similar to that found in clinical practice. Image resolution was measured using a Nuclear Associates 0.025-mm thick lead test pattern (#71412) with frequencies ranging from 1.5 to 20 c/mm. Kodak Insight E-F speed film was used for all images. The contrast (density difference) was determined using the FDA dental phantom and measuring the density difference between two areas of the phantom. The radiation exposure and HVL measurements were made with a Radcal Model 9010 dosimetry system with either a 6 cm3 or 180 cm3 ionization chamber. HVL measurements were made using type 1100 aluminum. Scattered radiation was measured using a typical one-gallon milk jug filled with water to simulate the human head. Staff dosimetry measurements were obtained for 18 facilities resulting in 422 reports for Nomad users and 122 reports for users of conventional x-ray equipment for a total of 546 individual staff dosimetry reports. Dose comparisons were carried out in four different ways: a) the percentage of dosimeters showing no measurable radiation; b) the average of all dosimeter readings; c) the average of all non-zero reading dosimeters; and d) the average of paired dosimeter measurements. Paired dosimeter measurements means that staff dosimetry data was obtained from staff using a conventional, wall-mounted system before the introduction of the Nomad. Subsequently staff dosimetry from the same operators was obtained after the introduction of the Nomad. This resulted in 42 paired measurements, i.e., measurements for the same staff using different, conventional intraoral x-ray systems and then the Nomad. The data were from facilities using either D-speed film or digital imaging. 9

Comparison of Waveforms Ave = 43 kv Ave = 56 kv Ave = 60 kv RESULTS The average kilovoltage for the unfiltered, 70 kvp single phase waveform is 43 kv. With filtration added to meet FDA requirements (1.5 mm HVL at 70 kvp) the average kilovoltage is 56 kv. Consequently, the 60 kv average kilovoltage of the Nomad is higher than that of a filtered 70 kvp beam. 10

Image Resolution (cycles/mm) Contact 1 inch 2 inches 3 inches Nomad > 20 14 10 6 Wall-mount > 20 13 6.5 6 At Various Object-to-Image Distances RESULTS Image resolution is reduced for greater object to image distances, for a fixed focal spot to image distance. In other words, as the object is moved away from the film the resolution is decreased as in clear in Slide 11. When the object, or resolution test pattern, is in contact with the film the resolution is limited by the image receptor (film or digital receptor). The Nomad exhibits better resolution than the conventional system due to the differences in focal spot size (0.4 mm vs 0.6 mm, respectively). It should be noted that the Nomad was hand-held for these exposures, i.e., it was not mounted on a tripod or similar device. 11

Contrast (Density Difference) Density Difference Nomad 0.55 Wall-mount 0.47 RESULTS The contrast or density difference was significantly higher (better) for the Nomad as compared to the conventional x-ray system. 12

Half-Value Layer (HVL) HVL (mm of Al) Nomad 1.92 Wall-mount 2.25 FDA Limit is 1.5 mm RESULTS The HVLs of the two systems met the FDA minimum values of 1.5 mm Al at 70 kvp, with the conventional x-ray system having a slightly higher HVL. 13

Patient Entrance Exposure Exposure (mr) RAP* (R-cm 2 ) Nomad 153 4.3 Wall-mount 126 4.9 For Kodak Insight (E-F Speed) Film Exposure difference = 18%, within typical variability Exposure differences due to HVL differences RAP = Roentgen-area-product, proportional to effective dose RESULTS Patient radiation doses were 153 mrad for the Nomad compared to 126 mrad for the conventional system. This 18% difference is well within the variability of patient doses from one unit to another regardless of type of x-ray system. The slightly lower dose for the conventional system is probably due to the slightly higher HVL. However, it should be stressed that the dose-area product (roentgen-area product, RAP) is lower for the Nomad at 4.3 R-cm2 compared to 4.9 R-cm2 for the conventional system. This indicates that the absorbed radiation dose to the patient for the Nomad will be 14% lower than for the conventional system due to the fact that the irradiated area is smaller for the former compared to the latter. 14

Scattered Radiation % of Entrance Exposure Nomad 0.089% Wall-mount 0.153% 90 scatter at 10 cm from 1 gallon milk container Hand-held/Wall-mount = 0.58 RESULTS The scattered radiation measured at 90 and 10 cm from the water-filled milk jug was lower for the Nomad at 0.089% of the entrance exposure compared to 0.153% for the conventional system. In other words, the scattered radiation for the Nomad is 58% of that for the conventional system, primarily due to the small irradiated area. 15

X-Ray Tube Leakage FDA maximum allowable leakage 100 mr/hr at 1 meter NOMAD leakage Not measurable at 1 meter (With 180 cm 3 chamber) RESULTS The maximum leakage radiation is specified by the FDA as 100 mr/hr at 1 meter. 16

X-Ray Tube Leakage Measurements made at 60 kv, 0.99 s*, at 5 cm (distance of hand on grip from x-ray x tube) 25 µr R per exposure Maximum permissible exposure to hand is 50,000 mr or equivalent of 2 million x-ray x exposures *Typical exposure 0.35 s RESULTS However, most x-ray tube leakage is on the order of 25 mr/hr or less at 1 meter. 17

How Does Leakage Compare? FDA maximum 100 mr/hr at 1 meter Assume no more than 25 mr/hr at 1 meter Leakage radiation for Nomad is 0.00015 that of wall-mount (based on 25 mr/hr) RESULTS The measured leakage for the Nomad is 0.015% (0.00015) of that for an x-ray tube producing 25 mr/hr. 18

Dose to Operators Retrospective dosimetry study 423 dosimetry reports for Nomad 122 dosimetry reports for wall-mount Included 42 Paired reports, i.e., reports for same staff using wall-mount and then Nomad All readings converted to monthly values RESULTS The results of the staff radiation dose measurements are shown in Slides 20-22. The percentage of dosimeters showing no measurable radiation (M, or 0 mrem) for Nomad users was 94.3% while that for the users of conventional x-ray systems was 77.9%. 19

All Dosimeter Readings n % with m (0 mrem) Average All (mrem) SEM Nomad 423 94.3 0.051 0.016 Wall-mount 122 77.9 0.604 0.138 RESULTS The average monthly dose for all dosimeters from Nomad users was 0.051 mrem and 0.604 for users of conventional equipment. In other words, Nomad users received about 8% of the radiation dose received by users of conventional equipment. 20

Non-Zero Dosimeter Readings n Average (mrem) SEM Nomad 24 0.901 0.216 Wall-mount 27 2.73 0.418 RESULTS The average staff dose for those with the non-zero dosimeter readings was 0.901 mrem for the Nomad compared to 2.73 mrem for users of conventional intraoral x- ray systems. In this case, the average staff dose for Nomad users was 33% of that for users of conventional equipment. 21

Paired Dosimeter Readings n Average (mrem) SEM Nomad 42 0.028* 0.004 Wall-mount 42 0.786* 0.123 *Statistically significant difference at p = 0.01 level RESULTS The comparison of staff dosimetry for paired measurements removes many of the variables, e.g., work load, etc., from the data. The average monthly, paired staff doses for the Nomad users was 0.028 mrem compared to 0.786 mrem for those using conventional intraoral x-ray systems, a statistically significant difference at the p = 0.01 level. In other words, the average monthly dose for the Nomad users was 3.6% of that for users of conventional dental x-ray systems. 22

Conclusions Nomad Wall-Mount Resolution Contrast HVL Meets FDA Meets FDA Patient Dose Scattered Radiation Leakage Radiation = Superior Performance CONCLUSIONS In conclusion, this study indicates that the resolution and contrast for the Nomad are superior to the Gendex x-ray system. In addition, the leakage and scattered radiation are lower for the Nomad compared to conventional, wall-mounted intraoral dental systems. The HVL meets the FDA requirements, with the Gendex having a slightly higher HVL than the Nomad. Both entrance radiation doses and the dose-area products for the two systems are similar. 23

Conclusions Occupational doses lower with Nomad than with wall-mount due to X-ray tube shield design Integral scattered radiation shield Shielded PID CONCLUSIONS Occupational doses are lower with the Nomad than with conventional intraoral x- ray systems. This is probably due to the tube shielding design (the Nomad is designed to be hand held and has significantly more shielding around the x-ray tube than a conventional system), the Nomad integral shield to protect the user from scattered radiation, and the shielded position indicating device collimator). 24

Conclusions Use of Nomad dental intraoral x-ray x system results in lower staff doses compared to wall-mount systems Additional measures, e.g., use of stands or lead aprons, are not warranted CONCLUSIONS Based on the results of this study, use of the Nomad dental intraoral x-ray system results in significantly lower staff doses compared to wall-mounted systems. Consequently, additional measures, e.g., use of lead aprons or stands, are not warranted. 25

Contact Information Edgar Bailey, MSEHE, CHP edbaileychp@msn.com Ph 512-934 934-2357 Edgar Bailey, M.S.E.H.E., C.H.P. 2804 Misty Shore Lane Pflugerville, Texas 78660-7744 e-mail edbaileychp@msn.com Joel Gray, Ph.D. DIQUAD, LLC 222 Lakeview Court Steger, Illinois 60475 John Ludlow, D.D.S. School of Dentistry University of North Carolina Chapel Hill, North Carolina 27599 26