CeSpace Titanium / PEEK



Similar documents
Aesculap Spine activ C

ANTERIOR CERVICAL DISCECTOMY AND FUSION. Basic Anatomical Landmarks: Anterior Cervical Spine

Aesculap Spine activ L

Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach.

Aesculap Spine activ L

Patient Information. Lateral Lumbar Interbody Fusion Surgery (LLIF).

ANTERIOR LUMBAR INTERBODY FUSION (ALIF) Basic Anatomical Landmarks: Anterior Lumbar Spine

Aesculap Spine activ L

TABLE OF CONTENTS. Surgical Technique 2. Indications 4. Product Information Patient Positioning and Approach 2

Options for Cervical Disc Degeneration A Guide to the M6-C. clinical study

Patient Information. Anterior Cervical Discectomy and Fusion Surgery (ACDF).

ZERO-P VA. Variable angle zero-profile anterior cervical interbody fusion (ACIF) device SURGICAL TECHNIQUE

coligne treatment technology

A Patient s Guide to Artificial Cervical Disc Replacement

activ C Aesculap Cervical Disc Prosthesis Surgical Technique Aesculap Spine

Minimally Invasive Spine Surgery

Minimally Invasive Lumbar Fusion

Options for Cervical Disc Degeneration A Guide to the Fusion Arm of the M6 -C Artificial Disc Study

AVS Anchor-C Cervical Cage System Surgical Technique. Large Graft Volume Self-Locking Screws Angled and Flexible Instrumentation

visualized. The correct level is then identified again. With the use of a microscope and

4052 Slimplicity Tech final_layout 1 6/29/15 3:29 PM Page 2 Surgical Technique

Each year, hundreds of thousands of adults are diagnosed with Cervical Disc Degeneration, The Cervical Spine. What is the Cervical Spine?

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail

Degenerative Spine Solutions

Does the pain radiating down your legs, buttocks or lower back prevent you from walking long distances?

Patient Guide to Neck Surgery

Direct Lateral Interbody Fusion A Minimally Invasive Approach to Spinal Stabilization

Thoracic Spine Anatomy

SPINAL FUSION. North American Spine Society Public Education Series

TABLE OF CONTENTS. Indications and Contraindications 3. Features and Benefits 4. Ease of connection 4 Stable fixation concept 5

LUMBAR LAMINECTOMY AND DISCECTOMY. Basic Anatomical Landmarks: Posterior View Lumbar Spine

Biomechanics of the Lumbar Spine

CORNERSTONE-SR. Instrumentation Surgical Technique SOFAMOR DANEK. as described by:

Presented by Zoran Maric, M.D. Orthopaedic Spine Surgeon May 22, 2010

R/F. Applications and Present Issues of Tomosynthesis in Spine Surgery. 87th Annual Meeting of the Japanese Orthopaedic Association Evening Seminar 4

Advances In Spine Care. James D. Bruffey M.D. Scripps Clinic Division of Orthopaedic Surgery Section of Spinal Surgery

Information for the Patient About Surgical

Implant materials. Learning outcomes. Implant materials in trauma. How to use this handout? Functions of implants. Types of materials

Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.

1 REVISOR (4) Pain associated with rigidity (loss of motion or postural abnormality) or

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery.

MatrixNEURO CMF Matrix

Minimally Invasive Spine Surgery What is it and how will it benefit patients?

Integra. Subtalar MBA and bioblock Implant SURGICAL TECHNIQUE

Disc Replacement or Fusion: which is best for me?

Anterior Cervical Discectomy and Fusion

If you or a loved one have suffered because of a negligent error during spinal surgery, you will be going through a difficult time.

Spine Surgery Coding

Case Study: Reduction and Stabilization of Grade III L5-S1 Dysplastic Spondylolisthesis in 15-Year-Old Female Using Posterior Approach

Dental Implant Options in Atrophic Jaws

Scout Vessel Guard. A hydrogel membrane with exceptional handling characteristics.

Spinal Arthrodesis Group Exercises

(English) NEXUS SPINE SPACER SYSTEM

A review of spinal problems

Aesculap. YASARGIL Aneurysm Clip System. From experienced hands for experienced hands! Aesculap Neurosurgery

Program Aesculap Spine Days, Tuttlingen Perspectives in Cervical Spine Surgery. April 25-26, 2016 Aesculap Academy, Tuttlingen / Germany

Technology Breakthrough in Spinal Implants (Technical Insights)

NCB Distal Femur System. Surgical Technique

Patient Guide to Lower Back Surgery

Zimmer FuZion Instruments. Surgical Technique (Beta Version)

Getting to the Backbone of Spinal Coding in ICD-10-PCS

ATLANTIS Anterior Cervical Plate System ATLANTIS Translational Plate. 510(k) Summary. February 2007

Titanium Wire with Barb and Needle. For canthal tendon procedures.

Spine Clinic Neurospine Specialists, Orthopaedics and Neurosurgery

V-TEK IVP System 2.7 System 4.0

Lumbar Spine Anatomy. eorthopod.com 228 West Main St., Suite D Missoula, MT Phone: Fax: info@eorthopod.

How To Fix A Radial Head Plate

Zimmer Small Fragment Universal Locking System. Surgical Technique

Exeter X3 RimFit. Acetabular Cup Surgical Protocol

MEDACTA UNCONSTRAINED SCREW TECHNOLOGY. Surgical Technique. Hip Knee Spine Navigation

Technique Guide. Screw Removal Set. Instruments for removing Synthes screws.

Magnetic Resonance Imaging

CPT Changes in Spine 2012

MEDICAL TECHNOLOGY IN IMPRESSIVE 3D

Cervical Spondylotic Myelopathy Associated with Kyphosis or Sagittal Sigmoid Alignment: Outcome after Anterior or Posterior Decompression

Cranial/Neurosurgical Implants

TwinFix Cannulated Compression Screw

Modular Sternal Cable System. Flexibility and strength in sternal closure and repair.

Lentur Cable System. Surgical Technique

.org. Fractures of the Thoracic and Lumbar Spine. Cause. Description

Implants of high-performance ceramic. vk-lange :44 Uhr Seite 1. One-piece precision implant. Made in Germany

Standard Internal Hex

Achilles Tendon Repair, Operative Technique

Why an Exactech Hip is Right for You

Foot and Ankle Technique Guide Proximal Inter-Phalangeal (PIP) Fusion

ORTHODONTIC MINI IMPLANTS Clinical procedure for positioning. Orthodontics and Implantology

Posterior Referencing. Surgical Technique

AFFINITY. Anterior Cervical Cage System Surgical Technique. as described by: Thomas A. Zdeblick, M.D. University of Wisconsin Madison, Wisconsin

ZEPHIR Anterior Cervical Plate System

The Petrylaw Lawsuits Settlements and Injury Settlement Report

ICD-10-PCS Documentation and Coding for Spinal Procedures October 22, 2015

A Patient s Guide to Diffuse Idiopathic Skeletal Hyperostosis (DISH)

James A. Sanfilippo, M.D. CONSENT FOR SPINAL SURGERY PATIENT: DATE:

Selection & Use of Surgical Needles

Corporate Brochure. Aesculap, Inc. Aesculap Implant Systems, LLC

March 20, Dear Mr. Chen:

BRYAN. Cervical Disc System. Patient Information

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

SPINAL STENOSIS Information for Patients WHAT IS SPINAL STENOSIS?

Transcription:

Aesculap Anterior Cervical Interbody Fusion System CeSpace Titanium / PEEK Aesculap Spine

CeSpace Content A Foreword 3 B Implant Material 4 C Implant Features 6 D Surgical Technique 8 E Ordering Information E1 Implants 12 E2 Instruments 14 2

Foreword A CeSpace is a spacer used for cervical interbody fusion. It is indicated for the treatment of degenerative diseases of the cervical disc and instabilities in the C3 to C7 region. The design of the CeSpace implant allows a maximum contact area between implant and vertebral endplates. CeSpace implants are available in PEEK and Titanium, depending on the preference of the surgeon. CeSpace stands for Primary stability Restoration of the natural disc height and lordosis and Long-term maintenance of the spinal balance. Combined with reliable instrumentation, CeSpace a solution for cervical interbody fusion. Stabilization with CeSpace Titanium. Stabilization with CeSpace PEEK. 3

CeSpace CeSpace Titanium B The heart of this implant is a solid titanium alloy core (Ti6Al4V / ISO5832-3). The core is mantled with the proven Plasmapore coating to increase the area of contact between implant and endplate. Plasmapore is a pure titanium coating (Ti / ISO5832-2) which offers an optimal foundation for the ingrowth of bone due to its balanced relationship between pore depth, porosity and roughness. Using a special manufacturing procedure, the implant surface is sprayed with pure titanium powder. Molten titanium particles settle on the core of the implant where they cool rapidly, building a firm form-lock between coating and core. In this way, each layer of the coating is built up and an optimal surface for bone ingrowth is created. Aim of the Plasmapore coating: Primary Stability The increased surface roughness of the Plasmapore coating ensures immediate stability of the motion segment. Secondary Stability Bone growth into the coating is ensured over a short period due to the optimal features of Plasmapore. Bone fusion between vertebrae and implant is achieved in this way. The coating concept, which has been proven as a result of many years of use in the field of hip prosthetics, has now become a new standard in spinal surgery. The bony integration of Plasmapore cages has been radiologically proven in lumbar fusion by Kroppenstedt S et al. Fig. 1 Fig. 2 4

CeSpace PEEK The material used is biocompatible PEEK-OPTIMA, which was introduced by Invibio in 1999. PEEK stands for PolyEtherEtherKetone. PEEK-OPTIMA polymer complies with ISO 10993-1, USP Class VI and ASTM F2026 for use as a medical implant material. The use of PEEK-OPTIMA as an orthopedic device material enjoys increased popularity in recent years due to the material s unique combination of characteristics. It s properties include radiolucency, high mechanical strength, biocompatibility and compatibility with standard sterilization methods. Of particular interest is the modulus of elasticity of PEEK- OPTIMA of 3.6 GPa, which is similar to that of cortical bone. This specific stiffness encourages load sharing between implant material and natural bone, thereby stimulating bone healing activity. The material provides excellent strength and rigidity. PEEK-OPTIMA also exhibits high fatigue resistance and a low wear factor. Extensive investigations of the biocompatibility of PEEK- OPTIMA have proven that the material is suitable for the use as a long-term implant. The intrinsic radioscopic transparency of the material provides permeability on X-rays and CT scans, allowing to visualize bone growth adjacent to the implant. This allows quick and simple assessment of the bone structure and progress towards bone fusion. To verify the position of PEEK implants on radioscopic images, non-radiolucent tantalum marker were integrated serving as location marker (Fig. 1). Fig. 1 Lateral view: AP view: 5

CeSpace Titanium Implant Features CeSpace Titanium C Plasmapore coating: rapid and safe osteointegration High primary stability due to a rough surface High secondary stability due to a fast migration of bone cells into the Plasmapore structure Intelligent implant design Fixation crown for an exact implant fit and high primary stability Optimized ratio between contact area and opening Option of filling with bone or bone substitute to enhance bone bridging Implant variety Adequate range of sizes providing the right implant to fit the patient Minimum 4 mm height Thought-out instruments Simple in handling Reliable and safe Clearly arranged 6

CeSpace PEEK Implant Features CeSpace PEEK Position verification despite X-ray transparency PEEK-OPTIMA allows quick and simple assessment of the bone structure and progress towards bone fusion Rod style marker for easy and exact implant positioning and localization Intelligent implant design Anatomical shape and serrated profile for an exact implant fit and high primary stability Optimized ratio between contact area and opening Option of filling with bone or bone substitute to enhance bone bridging Implant variety Adequate range of sizes providing the right implant to fit the patient Minimum 4 mm height Thought-out instruments Simple in handling Reliable and safe Clearly arranged 7

CeSpace Surgical Technique D 8

Fig. 2 Fig. 1 CASPAR Cervical Retractor System Patient Positioning The patient is placed in the supine position with the head slightly reclined (Fig. 1) and stabilized in a head holder. Once the lordotic cervical spine has been supported, the thorax may be placed on a pillow to emphasize the reclination of the cervical spine. The arms are fixed along the sides of the body. Exposure of the Intervertebral Space After the skin incision and preparation, the CCR retractor is applied. The blades are available in PEEK and Titanium. A counter retractor can be used (Fig. 2). The subcutaneous tissue is separated from the platysma cranially, caudally and medially, and the platysma is also separated following the direction of its fibres. The margins of the platysma can be held apart with the retractor or with two surgical forceps. Now the medial edge of the sternocleidomastoid muscle is located and prepared with the index finger in the connective tissue space over the ventral surface of the cervical spine and under lateralization of the vascular nerve bundle and medialization of the trachea, esophagus and thyroid gland. After the Langenbeck hooks have been inserted, the ventral surface of the cervical spine, still covered by a thin prevertebral layer of connective tissue, is revealed. This layer can now be exposed by either a blunt scissor or alternatively through bipolar coagulation in order to expand the tissue cranially and caudally using a swab. A wire is set under X-ray monitoring to mark the intervertebral disc space. 9

CeSpace Surgical Technique Fig. 3 Fig. 5 CASPAR Vertebral Body Distractor CASPAR Distraction Screws Fig. 4 Trial Implants CeSpace Titanium FJ164R-FJ167R/ FJ174R-FJ177R Trial Implants CeSpace PEEK: FJ474R-FJ478R/ FJ484R-FJ488R Distraction / Discectomy / Preparation of the Endplates The distraction screws are placed in position and the CASPAR distractor is applied following the CASPAR technique (Fig. 3). Complete discectomy is performed using various rongeurs, rectangular curettes and bone curettes (Fig. 4). While using a high speed drill to remove the posterior rim and/or dorsal osteophytes, care must be taken to avoid damaging the vertebral body endplates. Please note: Excessive preparation of the endplates may weaken the construct and cause subsidence of the CeSpace implant. Implant Selection The correct implant size can be established using the trial implants (Fig. 5). Due to the different geometry of the CeSpace Titanium and CeSpace PEEK implants specific trials for the respective system are needed. Laser markings on the handle as well as on the trial itself indicate the cranial and caudal side of the trial. Determination of Implant Size of CeSpace Titanium The height of the CeSpace Titanium trials is inclusive of the fixation crown. Determination of Implant Size of CeSpace PEEK The CeSpace PEEK trials regard the anatomical shape and serrated profile of the CeSpace PEEK implant. 10

Fig. 6 Fig. 7 Inserter CeSpace Titanium FJ100R Inserter CeSpace PEEK FJ415R/FJ497R CeSpace PEEK Packing Block / Punch FJ413P/FF914R CeSpace Insertion The Titanium implant is held securely and firmly onto the CeSpace inserter by means of a screw joint. The flexible sheath on the inserter has a stop at the front end which prevents the implant from being inserted too deeply into the intervertebral disc compartment. The CeSpace PEEK inserter has a clamp mechanism and is available with or without safety stop. Laser markings indicate the cranial and caudal side of the instrument. The implant should be inserted centrally in AP and with a distance of approximately 1-2 mm to both the anterior and posterior rim (Fig. 7). If indicated an Aesculap cervical plate should be used for additional stabilization. Once CeSpace is attached to the inserter, it can be introduced into the intervertebral space using image converter monitoring (Fig. 6). 11

CeSpace Titanium Ordering Information Implants E1 Diameter Height Angle Depth Art. no. Description Height Diameter Depth Angle FJ134T CeSpace Titanium 4 mm 14 mm 11.5 mm 5 FJ135T CeSpace Titanium 5 mm 14 mm 11.5 mm 5 FJ136T CeSpace Titanium 6 mm 14 mm 11.5 mm 5 FJ137T CeSpace Titanium 7 mm 14 mm 11.5 mm 5 FJ144T CeSpace Titanium 4 mm 16 mm 13.5 mm 5 FJ145T CeSpace Titanium 5 mm 16 mm 13.5 mm 5 FJ146T CeSpace Titanium 6 mm 16 mm 13.5 mm 5 FJ147T CeSpace Titanium 7 mm 16 mm 13.5 mm 5 Implant materials ISOTAN F Titanium forged alloy (Ti6Al4V / ISO 5832-3) Plasmapore Pure titanium (Ti / ISO 5832-2) 12 The specified height for the lordotic implant refers to the average height, which means the anterior section of the implant is higher than the posterior section. All CeSpace implants are individually sterile packed.

CeSpace PEEK Ordering Information Implants Diameter Angle Height Depth Art. no. Description Height Diameter Depth Angle FJ404P CeSpace PEEK 4 mm 14 mm 11.5 mm 5 FJ405P CeSpace PEEK 5 mm 14 mm 11.5 mm 5 FJ406P CeSpace PEEK 6 mm 14 mm 11.5 mm 5 FJ407P CeSpace PEEK 7 mm 14 mm 11.5 mm 5 FJ408P CeSpace PEEK 8 mm 14 mm 11.5 mm 5 FJ424P CeSpace PEEK 4 mm 16 mm 13.5 mm 5 FJ425P CeSpace PEEK 5 mm 16 mm 13.5 mm 5 FJ426P CeSpace PEEK 6 mm 16 mm 13.5 mm 5 FJ427P CeSpace PEEK 7 mm 16 mm 13.5 mm 5 FJ428P CeSpace PEEK 8 mm 16 mm 13.5 mm 5 Implant material PEEK-OPTIMA (PolyEtherEtherKeton) Registered trademark of Invibio Ltd, Lancashire FY5 4QD, UK The specified height for the lordotic implant refers to the average height, which means the anterior section of the implant is higher than the posterior section. All CeSpace implants are individually sterile packed. 13

CeSpace Titanium Ordering Information Implantation Instruments E2 FJ170 CeSpace Titanium instrumentation consisting of: Art. no. Description Handle Colour Recommended FJ164R Trial implant, 5, 14 x 4 mm blue 1 FJ165R Trial implant, 5, 14 x 5 mm blue 1 FJ166R Trial implant, 5, 14 x 6 mm blue 1 FJ167R Trial implant, 5, 14 x 7 mm blue 1 FJ174R Trial implant, 5, 16 x 4 mm green 1 FJ175R Trial implant, 5, 16 x 5 mm green 1 FJ176R Trial implant, 5, 16 x 6 mm green 1 FJ177R Trial implant, 5, 16 x 7 mm green 1 FJ100R Inserter 1 FJ171R Perforated tray with holding and storage elements 1 14 Recommended container for tray FJ171R: JK440 (without perforation) / JN440 (with perforation) Recommended lid: JK489

CeSpace PEEK Ordering Information Implantation Instruments FJ005 CeSpace PEEK instrumentation consisting of: Art. no. Description Handle Colour Recommended FJ474R Trial implant, 5, 14 x 4 mm blue 1 FJ475R Trial implant, 5, 14 x 5 mm blue 1 FJ476R Trial implant, 5, 14 x 6 mm blue 1 FJ477R Trial implant, 5, 14 x 7 mm blue 1 FJ478R Trial implant, 5, 14 x 8 mm blue 1 FJ484R Trial implant, 5, 16 x 4 mm green 1 FJ485R Trial implant, 5, 16 x 5 mm green 1 FJ486R Trial implant, 5, 16 x 6 mm green 1 FJ487R Trial implant, 5, 16 x 7 mm green 1 FJ488R Trial implant, 5, 16 x 8 mm green 1 FJ413P CeSpace PEEK packing block 1 FF914R Punch 1 FJ415R Inserter 1 FJ497R Safety stop 1 FJ499R Revision 1 FJ411P CeSpace PEEK tray 1 Recommended container for tray FJ411P: JK440 (without perforation) / JN440 (with perforation) Recommended lid: JK489 15

The main product trademark Aesculap and the product trademark CeSpace are registered trademarks of Aesculap AG. Aesculap AG Am Aesculap-Platz 78532 Tuttlingen Germany Phone +49 (0) 74 61 95-0 Fax +49 (0) 74 61 95-26 00 www.aesculap.com Aesculap a B. Braun company Subject to technical changes. All rights reserved. This brochure may only be used for the exclusive purpose of obtaining information about our products. Reproduction in any form partial or otherwise is not permitted. Brochure No. O16202 1111/1.5/6