LifeASSURE PFS Series Filter Cartridge Integrity Testing



Similar documents
TECHNICAL BRIEF. Microfluor II Filter Cartridge Integrity Testing. Introduction. Forward flow integrity test. 1. Definition

LifeASSURE PFS. Series Capsules & Cartridges Sterilising Grade PTFE Membrane Filters. 3M Purification. Product brochure

Assembly and Installation Procedures

Application Note USTR 805. Steam Sterilization of Pall Filter Assemblies Utilizing Replaceable Filter Cartridges

Lenntech Tel Fax

TR40 Sterilizing Filtration of Gases A comparison with TR26 Sterilizing Filtration of Liquids

Scientific & Technical Report

Validation Guide. USTR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges

3M Housings DS Series

Reduced Pressure Principle Assembly 3 Valve Procedure

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use

Air Eliminators and Combination Air Eliminators Strainers

CUNO TM Series NB Bags for Liquid Filtration

Durapore. High Throughput Flow Rates. Durapore 0.2 µm hydrophilic

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series

etape Continuous Fluid Level Sensor Operating Instructions and Application Notes

Installation, Maintenance, & Repair Series 4000B/LF4000B

Cartridge Filter Application Notes

AUTOMATIC CHANGEOVER MANIFOLD Series 915

Installation Manual DIAPHRAGM WELL TANK

How single-use connections advance aseptic processing: Increased process flexibility and reliability, reduced costs

Gas Custody Transfer Calibration

Installation Manual DIAPHRAGM WELL TANK

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / Instruction Manual. Page 1 of 9

Hydrophobic filter Integrity Testing

LifeASSURE 020SP. Series Capsules & Cartridge Filters. Sterilising Grade Nylon 6,6 Membrane Filter

el-720 Carbon Dioxide Gas Leak Detector User s Guide

C. starting positive displacement pumps with the discharge valve closed.

SANITARY GRAVITY SEWER LINE TESTING. B. Section CCTV Inspection of Gravity Sewer Lines

Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS

In-Line Filters and Coalescers with Automatic Drains

Lenntech Tel Fax

LifeASSURE BLA. Series cartridge fi lters

NITROUS TRANSFER PUMP INSTRUCTIONS

COUNTERBALANCE VALVES

Filtering Samples to On-Line Process Analyzers. Guidelines for Proper Filtration and Delivery of Samples to On-Line Process Analyzers

FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Standard Hydraulic Piping Flushing Procedure

Scotchgard Multi-Layer Protective Film 1004

OIL-X EVOLUTION Compressed Air Filters ENGINEERING YOUR SUCCESS. TECNI-AR Ltda - Tel: w.tecni-ar.com.br

Operating Instructions & Maintenance Guidelines for Precision Vaporizers

Rexroth Hydraulic Pump A10VO Series User Manual

INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES

3M Water Filtration Products. CFS01 Series, CFS02 Series, CFS11 Series, and CFS12 Series Housings. Installation and Instruction Manual For:

POLICY BULLETIN Tecumseh Compressor Company Compressor Group

CRITICAL MATERIALS HANDLING INVUE INTEGRATED FLOW CONTROLLERS NT6510, NT6520

Flushing and Cleaning the A/C System

3 Single-Coated Foam Tapes

3 Scotch-Seal TM. Tamper Proof Sealant Technical Data Issue No. 5

81000 Series Aluminum Gate Valve Installation and Repair Manual

Density Measurement. Technology: Pressure. Technical Data Sheet INTRODUCTION. S min =1.0 S max =1.2 CONSTANT LEVEL APPLICATIONS

FOAM CHAMBER MODEL - FCA & FCA-S HD FIRE PROTECT PVT. LTD. TECHNICAL DATA FEATURES APPLICATION. Carbon Steel Construction FCA-65, FCA-80 & FCA-100

ENGINE COOLING SYSTEM

Product Manual. CVS Type 67AFR Filter Regulator. Introduction. Installation. Description

Crystal Clear Sealant

Scotch-Weld. Low-Odor Acrylic Adhesives DP810 DP810 NS. Technical Data June, 2004

KNC Model 3666 Automatic Pressure Calibration System

Heating Water by Direct Steam Injection

Sample Exam Questions and Recommended References for ABC s Backflow Prevention Assembly Tester Exam

Installing a Probe. Typical Installation

Equivalents & Conversion Factors 406 Capacity Formulas for Steam Loads 407 Formulas for Control Valve Sizing

Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN

Dryaire Desiccant System

Refrigerant Changeover Guidelines R-22 to R-407C. Leading the Industry with Environmentally Responsible Refrigerant Solutions

SELECTION, APPLICATION AND MAINTENANCE

Recharge Procedure for Telemark TVP-2000 and TVP-3500

Soil Suction. Total Suction

MODEL G300 BRAKE BLEEDER

STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE

VANGUARD A OPERATOR S MANUAL PORTER. FM-952 Rev. E 7/07

Calibrating Differential Pressure Sensors

Overview. Optimizer Bulk

FORMATION DAMAGE AND WELL TREATMENT SYSTEM

Customer Application Brief. Filtration Processes Applied in Therapeutic Monoclonal Antibody Production. Bioprocess, Biologicals, & Pharmaceutical

This equipment has been certified by ITS/ETL to meet or excees SAE J1990 and J1991 for AR512 series and SAE J2099 and J2210 for AR500 series

UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE ABSORPTION OF CARBON DIOXIDE INTO WATER

User s Guide. Oreck Air Purifier with HEPA Filtration. Important! Read this manual carefully, and keep for future reference.

FILTRATION SOLUTIONS PhARmAceUTIcAL manufacturing FILTRATION SPecIALISTS

3M Thermally Conductive Adhesive Transfer Tapes

Installation, Maintenance, & Repair Series 909 and LF909 Reduced Pressure Zone Assemblies

3 Scotch-Weld TM. Structural Adhesive Films AF 6 AF 10 AF 13. Technical Data July, 2001

BERMAD Fire Protection

WHAT YOU DON T KNOW ABOUT ACCUMULATORS CAN KILL YOU!

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

BY MCI SOLUTIONS SECONDARY CONTAINMENT ADAPTOR

Scotch-Weld PUR Easy 250 Adhesive Applicator

Vacuum Control. Vacuum Applications. Vacuum Regulators. Vacuum Breakers (Relief Valves) Vacuum Control Devices. Vacuum Terminology

Element D Services Plumbing

ThermoStar Refrigeration Air Dryers

Submersion/Insertion Toroidal Sensor

Pipette Tips PACKARD BIOSCIENCE MultiPROBE II COMPATIBLE

CPX/HT Series. REFRIGERATED AND HIGH TEMPERATURE DRYER 12 to 2400 CFM. High-performance products. Designed for you!

Decaffeination of Raw, Green Coffee Beans Using Supercritical CO 2

3 Thermal Transfer Acrylate Label Material 3921

B150 Series Commercial Brewer. Direct Plumb Kit

Pressure Relief and Regulating Valves

Self Purging Liquid Trap (SPLT ) User s Manual

Transcription:

LifeASSURE PFS Series Filter Cartridge Integrity Testing Introduction Integrity testing is the end user s method of confirming the structural integrity of a membrane filter before and after use. It is a nondestructive testing that has been correlated to bacteria retention and validates the performance of the filter cartridge. An in specification result confirms the porosity of filter membrane and that the cartridge is structurally integral. Compressed gas and tank vent filters are typically sterilized in-place as part of the production process system to ensure the equipment has maintained sterility after in situ sterilization and that the equipment is ready to receive sterile product. Compressed gas and tank vent filters are non-product contact filters and are usually used over an extended period of time as compared to liquid filters that are typically discarded after a single product batch has been filtered. Hydrophobic compressed gas and tank vent filters are typically subjected to multiple sterilizations and; therefore, integrity tested multiple times as compared to filters used in liquid filtration applications. Filter integrity can be determined by any of the following four test methods. 1. Forward Flow Integrity Test (FFIT) 2. Bubble Point Test (BPT) 3. Hold Test (PHT) 4. Water Intrusion Test (WIT) Unlike the typical hydrophilic filter medium filters used for liquid filtration applications, compressed gas and tank vent filters are fabricated with hydrophobic filter medium. Hydrophobic filter medium prevents the bulk flow of aqueous based product through the filter at differential pressures below the intrusion pressure of the medium, while permitting the free passage of gas through the filter. Integrity testing of hydrophobic filters by three of the four conventional integrity test methods listed previously (FFIT, BPT and PHT) require the use of wetting agents, such as alcohol or an alcohol and water mixture, to thoroughly wet the filter membrane prior to integrity testing. All three of these tests must be performed off-line to prevent the wetting fluid from contaminating the sterile downstream side of the filter membrane. The fourth integrity test (WIT) can be performed in situ using sterile water. This eliminates the use of wetting agents and does not require the penetration of the filter membrane by the wetting fluid. All four tests can be conducted either manually or with an automated integrity test device, such as the CUNOCheck 2 Automatic Integrity Tester. The following procedures provide the instructions for integrity testing 10" cartridges. Contact 3M Purification Inc. Technical Service for details on how to integrity test other cartridge and capsule configurations. Forward Flow Integrity Test 1. Definition According to Fick s Law of Diffusion, when a differential gas pressure exists across a wetted membrane, the gas molecules will diffuse through the wetting fluid filling the pores of the membrane. The rate of passage is proportional to the solubility of the gas in the wetting fluid, the surface tension of the wetting fluid, the differential pressure, the thickness of the membrane, and the surface area of the membrane. The diffusion rate is measured at the minimum completely retentive membrane bubble point pressure. The Forward Flow Integrity Test (FFIT) may be employed with multi-cartridge housings. The FFIT is the preferred test method to determine if the membrane and cartridge are structurally integral. The diffusion rate is typically measured at a test pressure at or near the minimum completely retentive bubble point pressure. The FFIT has been shown to be both a quantitative measurement and to correlate to bacteria retention. The bubble point of the membrane can be a subjective determination and does not always correlate to bacteria retention. The bubble point of a membrane containing small defects will not correlate to bacteria retention due to the limited sensitivity of the Bubble Point Test (BPT) method to detect these small defects. Only the bubble point of a completely membrane will correlate to bacteria retention. Selecting a test pressure for the FFIT at the industry standard practice of 80% of the initial membrane bubble point does not ensure the integrity or the retention capability of the filter. This leaves a region between the test pressure and the membrane bubble point that

is not tested and; therefore, the integrity of the membrane can not be ensured in this region. This is analogous to testing at 80% of the expected membrane bubble point. Therefore, membrane filters should be tested at a more aggressive test pressure to ensure both the integrity and the correct pore size of the membrane. It is, therefore, recommended that the FFIT be conducted at a test pressure at or near the minimum completely retentive membrane bubble point. Integrity testing can be performed manually or using the CUNOCheck 2 Automated Integrity Tester. For more information about using the CUNOCheck 2 Automated Integrity Tester to perform an integrity test, refer to CUNOCheck 2 Operator Manual (70-0201-8694-9). 2. Procedure (Manual Testing Figure 1) A. Configure the system as shown in Figure 1. B. Install the filter in the housing and flow wetting fluid (25% / 75% TBA Tertiary butyl alcohol / water (v/v) @ 20 C, 60% / 40% IPA Isopropyl alcohol / water (v/v) @ 25 C, or 70% / 30% IPA Isopropyl alcohol / water (v/v) @ 25 C) through the filter at a rate of approximately 1 gpm/10" cartridge for a minimum of 2 minutes. Valid integrity limits will be obtained when the wetting fluid temperature is maintained within ± 2 C and concentration within ± 2% v/v. C. Close inlet and outlet valve to isolate the housing. all the wetting fluid from the upstream side of the filter by positioning to direct flow to drain. Adjust the pressure regulator to deliver 3 5 psig of air or nitrogen (DO NOT USE CO 2 ). Slowly open. Close when wetting fluid discharge is no longer evident. Gauge Graduated Cylinder Figure 1. FFIT (manual method). D. Connect one end of a flexible tube to the outlet port and submerse the other end in a container of water*. Position to direct flow to the container of water. * NOTE: It is important to make an aseptic connection and to close the outlet port immediately after the system has been integrity tested to prevent contamination. E. Open and slowly adjust the pressure regulator to pressurize the system to the specified FFIT pressure value and allow the system to equilibrate for a maximum of 2 minutes, or until steady bubbling is seen from the submersed end of the tube. F. Place the opening of the tube under an inverted graduated cylinder or burette (calibrated in milliliters or cubic centimeters) of an appropriate size filled with and submersed under sterile water. G. Measure the air flow for 3 minutes. Calculate the diffusion rate in cc/min. H. When the test is complete, compare the result to the cartridge FFIT specification. I. After the test is complete, flush the wetting fluid from the assembly with clean, filtered water and drain the water from the housing or adjust the pressure regulator to deliver 3 5 psig of air or nitrogen. Slowly open. Close when wetting fluid discharge is no longer evident. Dry the cartridge before placing the cartridge in service. 2

J. If the diffusion rate is higher than the specification, consider the following questions and re-test if necessary: Was the filter completely wetted? Was the correct pore size filter installed? Was the temperature of the wetting and filter ambient? Was the stabilization time adequate? Was the test time adequate? Was the filter seated correctly in the housing and were the O-rings undamaged? 3. Procedure (Automated Testing Figure 2) A. Configure the system as shown in Figure 2. B. Install the filter in the housing and flow wetting fluid (25% / 75% TBA Tertiary butyl alcohol / water (v/v) @ 20 C, 60% / 40% IPA Isopropyl alcohol / water (v/v) @ 25 C, or 70% / 30% IPA Isopropyl alcohol / water (v/v) @ 25 C) through the filter at a rate of approximately 1 gpm/10" cartridge for a minimum of 2 minutes. C. Close inlet and outlet valve to isolate the housing. all the wetting fluid from the upstream side of the filter by positioning to direct flow to drain. Adjust the pressure regulator to deliver 3 5 psig of air or nitrogen (DO NOT USE CO 2 ). Slowly open. Close when wetting fluid discharge is no longer evident. Shut-off Valve Vent CUNOCheck 2 Gauge Filter Housing D. Close the housing inlet valve and connect the Figure 2. FFIT (automated method). CUNOCheck 2 Automated Integrity Tester at and initiate the automated FFIT protocol. When prompted, program a stabilization time of 2 minutes and a test time of 3 minutes. E. When the test is complete, compare the measured Forward Flow Integrity value against the acceptable limit for the filter cartridge under test. F. After the test is complete, flush the wetting fluid from the assembly with clean, filtered water and drain the wetting fluid from the housing or adjust the pressure regulator to deliver 3 5 psig of air or nitrogen. Slowly open. Close when wetting fluid discharge is no longer evident. Dry the cartridge before placing the cartridge in service. G. Disconnect the CUNOCheck 2 Automated Integrity Tester from the housing. LifeASSURE PFS Series Grade FFIT psi (bars) FFIT Specifi cation (cc/min) per 10" equivalent PFS020 16 (1.1) 8.7 (25%/75% TBA/Water) @ 20 C 35.5 (60%/40% IPA/Water) @ 25 C 54.0 (70%/30% IPA/Water) @ 25 C NOTE: Valid integrity limits will be obtained when the wetting fluid temperature is maintained within ± 2 C and concentration is maintained within ± 2% v/v. 3

Bubble Point Test 1. Definition The bubble point is the minimum gas pressure required to overcome the surface tension holding a wetting fluid in a membrane filter s largest pore. The bubble point pressure measurement is only recommended for single 10" cartridge filters or smaller. When more filter area is on line, it becomes difficult to distinguish diffusional flow from the true bulk flow which occurs at the bubble point pressure. For assemblies with 2 or more 10" equivalent cartridges, diffusion flow or pressure hold measurements are recommended. The CUNOCheck 2 Automated Integrity Tester can be used to perform a BPT. When using the CUNOCheck 2 Automated Integrity Tester, follow the installation instructions for connecting the unit to the upstream valve as shown in Figure 4. Consult the appropriate 3M Purification Inc. literature for the filter cartridge bubble point value for the filter to be used and program this value into the CUNOCheck 2 Automated Integrity Tester when requested during the test set-up. 2. Procedure (Manual Testing Figure 3) A. Configure the system as shown in Figure 3. Gauge B. Install the filter in the housing and flow wetting fluid (25% / 75% TBA Tertiary butyl alcohol / water (v/v) @ 20 C, 60% / 40% IPA Isopropyl alcohol / water (v/v) @ 25 C, or 70% / 30% IPA Isopropyl alcohol / water (v/v) @ 25 C) through the filter at a rate of approximately 1 gpm/10" cartridge for a minimum of 2 minutes. C. Close inlet and outlet valve to isolate the housing. all the wetting fluid from the upstream side of the filter by positioning to direct flow to drain. Adjust the pressure regulator to deliver 3 5 psig of air or nitrogen (DO NOT USE CO 2 ). Slowly open. Close when wetting fluid discharge is no longer evident. D. Connect a tube from the outlet port to a container of sterile water*. Figure 3. BPT (manual method). * NOTE: It is important to make an aseptic connection and to close the outlet port immediately after the system has been integrity tested to prevent contamination. E. Open. Using the pressure regulator, slowly pressurize the system with air or nitrogen (DO NOT USE CO 2 ), raising the pressure 5 psi (0.34 bar) per minute. When within 5 psi (0.34 bar) of the expected bubble point pressure, make only very gradual 1 psi (0.07 bar) increases allowing 5 10 seconds between pressure increases to observe evidence of bubbling. F. Observe any air flow from the tube connected to the downstream port. A modest flow of small bubbles is diffusion flow only. When a continuous flow of large bubbles appears, the filter s bubble point has been reached. G. When the test is complete, compare the measured Bubble Point value against the acceptable Bubble Point value for the filter cartridge under test. 4

H. After the test is complete, flush the wetting fluid from the assembly with clean, filtered water and drain the water from the housing or adjust the pressure regulator to deliver 3 5 psig of air or nitrogen. Slowly open. Close when wetting fluid discharge is no longer evident. Dry the cartridge before placing the cartridge in service. I. If the bubble point is less than the recommended value, consider the following questions and re-test if necessary: Was the filter completely wetted? Was the correct pore size filter installed? Was the temperature of the wetting fluid and filter ambient? Was the filter seated correctly in the housing and were the O-rings undamaged? 3. Procedure (Automated Testing Figure 4) A. Configure the system as shown in Figure 4. B. Install the filter in the housing and flow wetting fluid (25% / 75% TBA Tertiary butyl alcohol / water (v/v) @ 20 C, 60% / 40% IPA Isopropyl alcohol / water (v/v) @ 25 C, or 70% / 30% IPA Isopropyl alcohol / water (v/v) @ 25 C) through the filter at a rate of approximately 1 gpm/10" cartridge for a minimum of 2 minutes. Shut-off Valve Vent CUNOCheck 2 Gauge Filter Housing C. Close inlet and outlet valve to isolate the housing. all the wetting fluid from the upstream side of the filter by positioning to direct flow to drain. Adjust the pressure regulator to deliver 3 5 psig of air or nitrogen (DO NOT USE CO 2 ). Slowly open. Close when wetting fluid discharge is no longer evident. Figure 4. BPT (automated method). D. Disconnect the wetting fluid supply line and allow the wetting fluid to drain from the housing. Leave or open to atmosphere. E. Close and connect the CUNOCheck 2 Automated Integrity Tester at and initiate the automated BPT protocol. F. When the test is complete, compare the measured Bubble Point value against the acceptable limit for the filter cartridge under test. G. After the test is complete, flush the wetting fluid from the assembly with clean, filtered water and drain the water from the housing or adjust the pressure regulator to deliver 3 5 psig of air or nitrogen. Slowly open. Close when wetting fluid discharge is no longer evident. Dry the cartridge before placing the cartridge in service. H. Disconnect the CUNOCheck 2 Automated Integrity Tester from the housing. LifeASSURE PFS Series Grade Minimum Bubble Point psi (bars) PFS020 16 (1.1) 5

Hold Test 1. Definition A variation of the FFIT is the Hold Test (PHT). Instead of measuring the diffusion rate of gas across the membrane, the PHT uses a sensitive pressure gauge to measure the decay of pressure in a closed volume on the upstream side of the membrane as the gas diffuses through the membrane. PHT values are dependent on the volume of the specific filter housing employed, less the volume of the installed cartridges. Therefore, they must be determined on a case by case basis. Please contact 3M Purification Inc. Technical Service for assistance, if necessary. The CUNOCheck 2 Automated Integrity Tester can be used to perform a PHT. When using the CUNOCheck 2, follow the installation instructions for connecting the unit to the upstream valve as shown in Figure 6. Consult the appropriate 3M Purification Inc. literature for the test pressure and PHT value for the filter cartridge to be used or calculate the maximum allowable value using the following equation. Program these values into the CUNOCheck 2 Automated Integrity Tester when requested during the test set-up. The following equation can be used to calculate the PHT value: ΔP T = D(Pa) V hsg D = Manufacturer s maximum allowable diffusion rate for all the installed filters in cc/min (see FFIT specifications) T = Time (typically 3 minutes) Pa = Atmospheric pressure V hsg = Upstream housing volume (cc) less the volume occupied by the cartridge(s) ΔP = Allowable pressure loss 2. Procedure (Manual Testing Figure 5) A. Configure the system as shown in Figure 5. Gauge B. Install the filter in the housing and flow wetting fluid (25% / 75% TBA Tertiary butyl alcohol / water (v/v) @ 20 C, 60% / 40% IPA Isopropyl alcohol / water (v/v) @ 25 C, or 70% / 30% IPA Isopropyl alcohol / water (v/v) @ 25 C) through the filter at a rate of approximately 1 gpm/10" cartridge for a minimum of 2 minutes. C. Close inlet and outlet valve to isolate the housing. all the wetting fluid from the upstream side of the filter by positioning to direct flow to drain. Adjust the pressure regulator to deliver 3 5 psig of air or nitrogen (DO NOT USE CO 2 ). Slowly open. Close when wetting fluid discharge is no longer evident. D. Disconnect the wetting fluid supply line and allow the wetting fluid to drain from the housing. Leave or open to atmosphere. Figure 5. PHT (manual method). 6

E. Open and slowly adjust the pressure regulator to pressurize the system to the specified PHT pressure value and allow the system to equilibrate for a maximum of two minutes. F. Using a stopwatch, measure the pressure decay for 3 minutes. Calculate the pressure decay over the 3 minute period in psi/min. G. When the test is complete, compare the measured Hold value against the acceptable pressure hold value for the filter cartridge under test H. After the test is complete, flush the wetting fluid from the assembly with clean, filtered water and drain the wetting fluid from the housing or adjust the pressure regulator to deliver 3 5 psig of air or nitrogen. Slowly open. Close when wetting fluid discharge is no longer evident. Dry the cartridge before placing the cartridge in service. I. If the pressure decay is higher than the specification, consider the following questions and re-test if necessary: Was the filter completely wetted? Was the correct pore size filter installed? Was the temperature of the wetting fluid and filter ambient? Was the stabilization time adequate? Was the test time adequate? Was the filter seated correctly in the housing and were the O-rings undamaged? Are there any leaks on the upstream side of the filter? 3. Procedure (Automated Test Figure 6) A. Configure the system as shown in Figure 6. B. Install the filter in the housing and flow wetting fluid (25% / 75% TBA Tertiary butyl alcohol / water (v/v) @ 20 C, 60% / 40% IPA Isopropyl alcohol / water (v/v) @ 25 C, or 70% / 30% IPA Isopropyl alcohol / water (v/v) @ 25 C) through the filter at a rate of approximately 1 gpm/10" cartridge for a minimum of 2 minutes. Shut-off Valve Vent CUNOCheck 2 Gauge Filter Housing C. Close inlet and outlet valve to isolate the housing. all the wetting fluid from the upstream side of the filter by positioning to direct flow to drain. Adjust the pressure regulator to deliver 3 5 psig of air or nitrogen (DO NOT USE CO 2 ). Slowly open. Close when wetting fluid discharge is no longer evident. Figure 6. PHT (automated method). D. Disconnect the wetting fluid supply line and allow the wetting fluid to drain from the housing. Leave or open to atmosphere. E. Close and connect the CUNOCheck 2 Automated Integrity Tester at and initiate the PHT protocol. Close the housing inlet valve and connect the CUNOCheck 2 Automated Integrity Tester at and initiate the automated FFIT protocol. When prompted, program a stabilization time of 2 minutes and a test time of 3 minutes. F. When the test is complete, compare the measured Hold value against the acceptable limit for the filter cartridge under test. 7

G. After the test is complete, flush the wetting fluid from the assembly with clean, filtered water and drain the water from the housing or adjust the pressure regulator to deliver 3 5 psig of air or nitrogen. Slowly open. Close when wetting fluid discharge is no longer evident. Dry the cartridge before placing the cartridge in service. H. Disconnect the CUNOCheck 2 Automated Integrity Tester from the housing. Water Intrusion Test 1. Definition The WIT is only used with hydrophobic filters. The pores of a hydrophobic filter will resist the passage of water up to a certain pressure. The pressure at which water will pass through the pores of a hydrophobic filter is inversely proportional to the size of the pore. The WIT offers several advantages over the previously described conventional integrity tests that are usually conducted with an alcohol or alcohol water mixture. Safety no flammability issues Contamination no wetting fluid or alcohol contamination of product Time no drying of the filter required before placing back into service The CUNOCheck 2 Automated Integrity Tester can be used to perform a WIT. When using the CUNOCheck 2 Automated Integrity Tester, follow the installation instructions for connecting the unit to the upstream valve as shown in Figure 7. Consult the appropriate 3M Purification Inc. literature for the test pressure and WIT value for the filter cartridge to be used and program these values into the CUNOCheck 2 Automated Integrity Tester when requested during the test set-up. 2. Procedure (Automated Testing Figure 7) A. Configure the system as shown in Figure 7. Gauge B. Install the filter(s) in the housing and close all valves. C. Open. Fill the upstream side of the system with water through either or. When water emerges from, stop the water flow and close whichever valve was used to introduce the water to the system. Shut-off Valve Vent CUNOCheck 2 Filter Housing D. When water stops flowing from close. E. Connect the CUNOCheck 2 Automated Integrity Tester to the pressurized gas source, open the shut-off valve, and, using the regulator, adjust the pressure to the CUNOCheck 2 Automated Integrity Tester to the necessary level for the test being conducted. Figure 7. WIT (automated method). F. Close the housing inlet valve and connect the CUNOCheck 2 Automated Integrity Tester at and initiate the automated WIT protocol. When prompted, program a stabilization time of 5 minutes, a test time of 5 minutes and set the volume calculation to type 3. G. When the test is complete, compare the test result to the WIT specification for the filter being tested. H. After the test is complete, drain the water from the housing and place the cartridge in service. 8

I. Disconnect the CUNOCheck 2 Automated Integrity Tester from the housing. J. If the WIT value is higher than the specification, consider the following questions and re-test if necessary: Was the filter dry and free of solvents or surfactants prior to starting the test? Was the correct pore size filter installed? Was the temperature of the water and filter ambient? Was the temperature stable during the test period? Was the water free of solvents and/or surfactants? Was the stabilization time adequate? Was the test time adequate? Was the filter seated correctly in the housing and were the O-rings undamaged? Are there any leaks on the upstream side of the filter? LifeASSURE PFS Series Grade WIT psig (barg) WIT Specifi cation (cc/min) per 10" equivalent PFS020 40 (2.76) 0.59 9

Important Notice The information described in this literature is accurate to the best of our knowledge. A variety of factors, however, can affect the performance of the Product(s) in a particular application, some of which are uniquely within your knowledge and control. INFORMATION IS SUPPLIED UPON THE CONDITION THAT THE PERSONS RECEIVING THE SAME WILL MAKE THEIR OWN DETERMINATION AS TO ITS SUITABILITY FOR THEIR USE. IN NO EVENT WILL 3M PURIFICATION INC. BE RESPONSIBLE FOR DAMAGES OF ANY NATURE WHATSOEVER RESULTING FROM THE USE OF OR RELIANCE UPON INFORMATION. It is your responsibility to determine if additional testing or information is required and if this product is fi t for a particular purpose and suitable in your specifi c application. 3M PURIFICATION INC. MAKES NO REPRESENTATIONS OR WARRANTIES, EITHER EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OF ANY OTHER NATURE HEREUNDER WITH RESPECT TO INFORMATION OR THE PRODUCT TO WHICH INFORMATION REFERS. Limitation of Liability 3M Purifi cation Inc. will not be liable for any loss or damage arising from the use of the Product(s), whether direct, indirect, special, incidental, or consequential, regardless of the legal theory asserted, including warranty, contract, negligence or strict liability. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you. 3M Purification Inc. 400 Research Parkway Meriden, CT 06450, U.S.A. Tel (800) 243-6894 (203) 237-5541 Fax (203) 630-4530 www.3mpurification.com 3M is a trademark of 3M Company. LifeASSURE and CUNOCheck are trademarks of 3M Company used under license. 2010 3M Company. All rights reserved. Please recycle. Printed in U.S.A. 70-0201-8804-4 REV 1110