In vitro diagnostic reagent, calibrator and control material stability



Similar documents
References to Computer and Software in FDA s CGTPs (Final Rule and Draft Guidance) for HCT/Ps Preamble to the CGTP Final Rule

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC

Arrhenius Equation Demystified History, Background, and Common Usage in the Accelerated Aging of Packaging for Medical Devices

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

ISO/IEC QUALITY MANUAL

Content Sheet 5-1: Overview of Sample Management

TRI-LEVEL CARDIAC CONTROL (CRD CONTROL 1, 2, 3)

CHAPTER 7 QUALITY ASSESSMENT

Content Sheet 10-1: Overview of External Quality Assessment (EQA)

Measuring instruments in industry.

New York State Department of Health (NYSDOH) Bureau of Immunization Vaccine Storage and Handling Plan

How To Know If A Mobile App Is A Medical Device

MEDICAL DEVICE GUIDANCE

Reflection paper for laboratories that perform the analysis or evaluation of clinical trial samples

Journal of Chemical and Pharmaceutical Research

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

Good Clinical Laboratory Practice (GCLP) An international quality system for laboratories which undertake the analysis of samples from clinical trials

New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01)

FINAL DOCUMENT. Global Harmonization Task Force. Title: Label and Instructions for Use for Medical Devices

EMA Clinical Laboratory Guidance - Key points and Clarifications PAUL STEWART

How To Inspect A Blood Bank

A Stability Program for the Distribution of Drug Products

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

HIV Rapid Test Kits USAID Approval and Technical Guidance June 17, 2013 Revision

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Example of a food company quality

URINE COLLECTION, PREPARATION AND HANDLING

Mayo Central Laboratory for Clinical Trials A Division within Mayo Clinical Trial Services

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

ORACLE PROCESS MANUFACTURING QUALITY MANAGEMENT

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Validation of measurement procedures

Procedure for Equipment Calibration and Maintenance

WORKPLACE HEALTH AND SAFETY AUDITING GUIDELINES

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1

Generic CMMS Quality Assurance Plan

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

Content Sheet 3-1: Equipment Management Overview

Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing

Content Sheet 16-1: Introduction to Documents & Records

CDC-Lipid Standardization Program. Laboratory Data Collection System (LDCS) Instruction Manual

Presented by Rosemarie Bell 24 April 2014

Best-Practice Guide Pharmaceutical-Chain Temperature Control and Recording

Prepublication Requirements

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Procedure for Procurement and Receipt

MEDICAL DEVICES INTERIM REGULATION

SMF Awareness Seminar 2014

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

Unit 205 Order, receive and store pharmaceutical stock

Validation and Calibration. Definitions and Terminology

Working Party on Control of Medicines and Inspections. Final Version of Annex 16 to the EU Guide to Good Manufacturing Practice

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

IMMUNOASSAY CONTROL - LEVEL 1 (IA CONTROL 1)

OVERVIEW OF STANDARD 7E & STANDARD 20

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

Protocol for ordering, storing and handling vaccines. March 2014

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

GE Power Electronics Business Total Quality Management for Suppliers

PUBLICATIONS. Introduction

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

INTEROPERABILITY UNIT

What is Functional Safety Management?

BSI: An In Vitro Diagnostics Notified Body. A guide to the In Vitro Diagnostic Directive....making excellence a habit.

CHAPTER 6 PROCUREMENT AND SUPPLY OF PHARMACEUTICAL PRODUCTS IN THE PUBLIC AND PRIVATE MEDICAL SECTORS

2015 No. 575 FINANCIAL SERVICES AND MARKETS. The Solvency 2 Regulations 2015

Care and Monitoring of Refrigerators and Freezers

The Facts on Equine Drug Testing

FDA Regulation of Whole Slide Imaging (WSI) Devices: Current Thoughts

Checklist. Standard for Medical Laboratory

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers.

Quality Assurance for the Clinical Laboratory

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day

DATA MANAGEMENT IN CLINICAL TRIALS: GUIDELINES FOR RESEARCHERS

MEDICAL DEVICES SECTOR

Guidance for Industry

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations

IMMUNOASSAY PREMIUM PLUS LEVEL 3 (IA PREMIUM PLUS 3)

Oligonucleotide Stability Study

QUESTIONS FOR YOUR SOFTWARE VENDOR: TO ASK BEFORE YOUR AUDIT

MATRIX GEMINI LIMS AT WORK IN THE FOOD INDUSTRY

Standards of Practice for Pharmacists and Pharmacy Technicians

Content Sheet 4-1: Purchasing and Inventory Overview

Measuring equipment in a laboratory.

Quality Management Systems Manual

Blood Glucose Monitoring: The Facts about Accuracy

New changes to cleanroom & clean air device classifications: ISO & 2

Office for Nuclear Regulation

Guidelines for Specimen Identification and Labeling for Power Chart and Lifepoint Users

2. Contact Person: Garo Mimaryan, MS., RAC 7 Technical Regulatory Affairs Specialist III 3. Phone Number: (914)

OPERATIONAL STANDARD

Vaccine Storage Practices Appropriate Vaccine and Diluent Storage Conditions

Transcription:

In vitro diagnostic reagent, calibrator and control material stability Pauline Armstrong Randox Laboratories, Ltd. Crumlin The stability of in vitro medical diagnostic devices whether they are reagents, calibrators or quality control materials with respect to laboratory accreditation and current international and European quality standards will be discussed in this presentation, which is divided into two main parts. In the first part, stability issues affecting laboratory accreditation will be discussed followed by a review of what current international and European standards have to say about the stability of an in vitro medical diagnostic device whether it is a reagent, a reagent product, a calibrator or a control material. In the second part the general requirements for stability testing and the generation of stability data will be considered. In the draft version of the standard BS-EN 13640 Stability Testing of in vitro Diagnostic Medical Devices the following definition of stability is offered: stability is "the ability of an in vitro diagnostic reagent, when kept under specified conditions, to retain throughout the shelf life its characteristics and/or performance within limits specified by the manufacturer". There are some important concepts in this definition. Firstly, stability is related to specified storage conditions for example, refrigerated (2 C-8 C), room temperature (15 C-25 C) and indeed frozen ( -20 C). Secondly, stability is related to the characteristics and performance limits established by the manufacturer. Performance limits may include recovery / accuracy, precision, sensitivity etc. Most clinical pathology laboratories purchase their in vitro diagnostic reagents, calibrators and quality control materials from a variety of national and international in vitro medical diagnostic devices manufacturers. The manufacturer, as an integral part of product development determines the stability characteristics of the product. The stability claims are usually quoted on the product labelling. The end user is not therefore required to conduct any type of stability testing. In their Standards for the Medical Laboratory, Clinical Pathology Accreditation (UK) Ltd. state that it is essential to have proper management of reagents, calibration and quality control materials and that these materials must be correctly identified with the date of receipt, lot number, first use and expiry. Some automated clinical chemistry analysers such as the Dade-Behring

Dimension are now capable of assimilating reagent shelf life information frequently from a barcode label attached to the reagent, calibrator or control. This feature prevents the inadvertent use of product that has passed its quoted expiry date. However, many analysers, for example the COBAS MIRA and the Hitachi 717, cannot identify product shelf life details from the product label and so the responsibility of using only within date product falls on the laboratory staff. Consequently, for laboratory accreditation purposes, procedures must be in place to ensure that only within date product is used. Whilst a wide spectrum of in vitro medical diagnostic devices may be purchased from their manufacturers, not all the reagents, calibrators and controls required are available in this way. Therefore, it is essential for all laboratory staff to understand the concept of stability testing whether or not they are required to put it into practice. Since this symposium is being held specifically to discuss certification and accreditation of laboratories, the question "why does a laboratory require certification or accreditation?" may be asked. The answer in some cases is very simple. In the absence of accreditation the laboratory cannot operate legally to produce, report and offer interpretation of the results obtained from the analysis of a patient's blood or urine sample. Whilst this may be the reality the whole ethos of laboratory accreditation is to ensure the quality of the product or service. In the United Kindom, Clinical Pathology Accreditation (UK) Ltd. state that accreditation is the process by which an organisation gains recognition that its activities and products have a guaranteed high quality. The accreditation of a laboratory usually involves an external audit to assess the ability of the laboratory to provide a service of high quality. How may a service of high quality be defined? ISO 9000:2000 offers the following definition of quality as "the degree to which a set of inherent characteristics fulfils requirements". David Burnett in his book Understanding Accreditation in Laboratory Medicine offers a definition from an anonymous source that is perhaps more pertinent to laboratory medicine, "a blood glucose result measured on the correct specimen, that is accurate, timely and properly interpreted is a quality product or service". By considering this definition and referring to the definition of stability it can be seen how inherent instability or compromised stability (due, for example, to inappropriate storage of the in vitro diagnostic reagent) may prevent delivery of a quality service by an accredited laboratory. To ensure the accuracy and consistency of the results being reported the stability of the in vitro diagnostic reagent is paramount. It is important to appreciate that the in vitro medical diagnostic devices Directive 98/79/EC does not impose any shelf life requirements on an in vitro medical diagnostic device. For example, there is no general statement that all in vitro medical diagnostic devices must have a shelf life of two years. Neither does the

International Standard ISO 9001 in either its 1994 or recently released 2000 versions. Nor does the Food and Drug Administration (FDA) in the United States in its 510 (K) Premarket Notification or Premarket Approval Systems of product registration. Rather, the manufacturer or organisation is required to make stability claims for a particular product, including in vitro medical diagnostic devices and to substantiate these claims with stability data. For any in vitro medical diagnostic device that has been declared compliant with the Directive 98/79/EC and is consequently "CE-Marked" details of stability testing and the stability data to substantiate the claims are held in the product technical file. It is recommended that stability assessment studies are carried out with reference to the draft standard BS-EN 13640. So, what types of stability study should be carried out? Shelf life Firstly, determination of reagent shelf life. The manufacturer must establish and inform the customer, of the expiry date of an in vitro diagnostic reagents kit, reagent, calibrator or control. The period of time from the date of manufacture to the expiry date is termed the shelf life. Shelf life may also be defined as the period of the time during which the performance of the product is considered acceptable. Where possible, shelf life should be based on real time testing. This is not always realistic, as instabilities in an in vitro diagnostic reagents are often only detectable after considerable storage under "normal" conditions. Consequently, most manufactures opt to use data derived from accelerated, heat-stressed testing of the product in its final container, in conjunction with their experience gained from similar in vitro diagnostic reagents to place a product on the market. The results from this type of stability testing can be extrapolated to normal storage conditions. If a reagent is placed on the market with stability claims based on accelerated testing these data should be verified by real-time testing. a) Accelerated stability testing Accelerated testing is usually based on exposure of the reagent, calibrator or control to different elevated temperatures, typically, 25 C, 37 C or 40 C depending on the initial storage temperature recommendation (2 C-8 C or 15 C-25 C). Product is stored at the specified elevated temperature for a predefined period of time, typically, 1 week up to 8 weeks. Samples of the product are then removed and tested according to a predefined protocol. Comparison is made to results obtained from reagent stored at the recommended temperature. In addition, the product may be stored under specified conditions until a significant deterioration of the product is seen. The Arrhenius Equation may then be used to predict the shelf life of the product. b) Real-time stability testing

For real-time stability testing sufficient test material is stored at the recommended storage temperature in the final reagent container to last through the complete stability study. Where possible, sufficient calibrators and controls should also be stored to allow completion of the study without changing the lot of calibrator or control. If assessing calibrator or control materials sufficient reagent should be stored to minimise lot changes during the stability study. Product is tested at predefined time intervals using a predefined protocol. Results are analysed according to predefined acceptance criteria. Data may be summarised as a Levy-Jennings plot. The study should be completed to the end of the target shelf life but may be continued until significant degradation in performance can be seen. Reconstituted Stability Testing This type of stability testing applies to lyophilised product. The material is reconstituted, stored appropriately and tested at predefined intervals, for example, between 7 and 28 days inclusive using a pedefined test protocol. The results are compared to those obtained with freshly reconstituted material. Conclusions are made using predefined acceptance criteria. On-Board / Open Vial Stability This is the determination of the stability of an in vitro diagnostic reagent after first opening the primary container and may be assessed either on-board a clinical chemistry analyser in the refrigerated compartment or in a refrigerator at 2 C-8 C. In both cases the reagent is stored in the final reagent vial for up to 28 days, being tested after 7, 14, 21, and 28 days storage according to a predefined protocol. A comparison is made to the results obtained with reagents stored in closed bottles at 2 C-8 C at each time point. Conclusions are made using predefined acceptance criteria. Routine stability testing To ensure that commercially available in vitro medical diagnostic devices meet their respective stability claims, manufacturers must assess the stability of reagents selected at random. This type of testing may be termed 'routine stability testing' and the studies are conducted as for 'real-time' stability testing. Verification of the stability of an in vitro diagnostic reagent after modification This type of stability testing is designed to ensure that any modification to an in vitro diagnostic reagent does not affect the stability of the product. The testing protocol usually involves accelerated, real-time, on-board and shipping stability studies. Conclusions are made to predefined acceptance criteria.

Stability assessment during transportation This type of stability study is designed to assess whether or not special shipping requirements exist for an in vitro diagnostic reagent. During shipment the product may be exposed to elevated or sub-zero temperatures. Storing the product at either -20 C or 37 C for predefined periods of time may simulate these conditions. The results are compared to those obtained with product stored under normal temperature conditions. Conclusions are made using predefined acceptance criteria. In conclusion, whilst none of the current European and international quality standards impose any shelf life requirements on an in vitro medical diagnostic device, the stability of the in vitro diagnostic reagent, calibrator or control is paramount to the accuracy and consistency of the result being reported. Citació recomanada per a aquest document: Armstrong P. In vitro diagnostic reagent, calibrator and control material stability. In vitro veritas 2001;2, art.22: <http://www.acclc.cat>