Using Equotip Hardness Test Blocks Abstract 1. Introduction 2. Verification of Leeb hardness testers according to standards



Similar documents
Portable Hardness Tester

Unmatched Metal Hardness Testing

Portable Hardness Testing Leeb Rockwell UCI

MENG 302L Lab 1: Hardness Testing

LEEB HARDNESS TESTER TH130/132/134

Description of mechanical properties

MANUAL TR-110 TR-110

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Procedure for Equipment Calibration and Maintenance

Tubing Data. Contents. Tubing Selection. Tubing Handling. Tubing Material. Tubing Outside Diameter Hardness. Tubing Wall Thickness

VeMet, Utrecht, NL «Solution in Wear Protection» Dipl.-Ing. Wolfgang Leichnitz. Quit

Technical Data. 7. Bearing Fits. 7.1 Interference. 7.2 Calculation of interference F B LLLLLLLLL( A-54

WOOD WEAR TESTING USING TRIBOMETER

FALEX Four-Ball Extreme Pressure Test Machine

Surface roughness. Picture 1: Friction coefficient as a function of surface roughness Curve 1: Roughness R z 0,5µm Curve 2: Roughness R z 5 µ

AISI O1 Cold work tool steel

How To Test A Rockwell Type Hardness Tester

STAVAX SUPREME. Stainless tool steel

PRACTICAL HARDNESS TESTING MADE SIMPLE. Table of Contents 1. GENERAL 1 2. INTRODUCTION 3 3. BRINELL HARDNESS TESTING 9 4. VICKERS HARDNESS TESTING 14

ACL 395 Resistivity Meter

FLAME CuTTIng te la P is utting B c

Temperature. LumaSense Technologies, Inc. supplies a unique selection of very precise calibration

peel adhesion 2 breaking load 4 rolling ball 6 shear adhesion 7 carton test 9 summary test methods 11

Penetration Testing for Material Hardness

SURFACE PREPARATION STANDARDS

VERZUS 750CCD Universal Hardness Tester

Specification for Rotary Drill Stem Elements

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

Hardness Testing at Elevated Temperatures. SJ Shaffer, Ph.D. Bruker-TMT

High speed machining and conventional die and mould machining

RAMAX S Prehardened stainless holder steel

Calibration Certificate

Making A Mirror Grinding Tool By Allen Malsburry

7.3 Fit selection. Inner ring: Rotating. Outer ring: Stationary. Inner ring: Stationary. Outer ring: Rotating. Inner ring: Stationary

Bearing Failure: Causes and Cures

PHOTO-ETCHING OF TOOL STEEL

Mobile Hardness Testing

Engine Bearing Materials

To measure an object length, note the number of divisions spanned by the object then multiply by the conversion factor for the magnification used.

Proper use of the Rebound Hammer Updated to reflect the changes to ASTM C805

METHOD OF TEST FOR UNIT WEIGHT OF FRESH CONCRETE

Material data sheet. EOS StainlessSteel GP1 for EOSINT M 270. Description, application

White Paper IK code: Impact protection in line with the IEC standard. By Sven Schnautz

User checks and maintenance of laboratory balances

The Determination of Uncertainties in Charpy Impact Testing

Pedestrian slip resistance testing: AS/NZS : 1993 For Resene Paints Ltd. Aquapoxy semi-gloss

LABORATORY EXPERIMENTS TESTING OF MATERIALS

1.02 DESCRIPTION A. This Section specifies furnishing and installing Cast in Place Detectable/Tactile Warning Surface tiles where indicated.

1 WATERFLUX - Comparisonon COO Rectangular Magmeter in water applications

Guide to adhesively mounting accelerometers

Flocking of textiles. Flocked shirt 13. flocking drying cleaning. adhesive. application. creation

SFI SPECIFICATION 61.1 EFFECTIVE: FEBRUARY 26, 2014

SITEMA PowerStroke. Technical Information TI-P11. 1 Function. 2 Applications. Mould Closing Devices series FSK. Contents

Impact testing ACTIVITY BRIEF

Measuring equipment in a laboratory.

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Check-list for auditors. Audit visit at candidate registered laboratories

What Are the Differences?

PNEUMATIC - HIGH PRESSURE FLAT-FORM HEAVY LIFTING BAGS

Group. The precision company SURFACE TECHNOLOGIES INTERNATIONAL. Hard chrome. Surface solutions by STI Group. English

PRELIMINARY BROCHURE. Uddeholm Ramax HH

TechCut 4 Precision Low Speed Saw

Automotive Electronics Council Component Technical Committee

Testing Your Laboratory Balance

NEW SIX-SEALS CUTTING RING. INTERNATIONAL INDUSTRIAL PATENT Nr of the 10/03/99 FLANKS AND DOES NOT REPLACE THE STANDARD RING CURRENTLY IN USE

STATIC COEFFICIENT OF FRICTION MEASUREMENT USING TRIBOMETER. Static COF Time(min) Prepared by Duanjie Li, PhD

Lapping and Polishing Basics

Test equipment for temperature and its reduction

North American Stainless

Geometry and dimensional tolerances of engine bearings

Understand, Maintain and Service Your Torque Wrench

Peel Adhesion of Pressure Sensitive Tape

62 Max/62 Max + Infrared Thermometer

TIGER Drylac U.S.A., Inc Swenson Avenue St Charles, IL T F office.us@tiger-coatings.us

Decoding Heat Cure Denture Base Polymers. By Tom Zaleske

Standard Test Procedures Manual

(This report is endorsed) Industrivej 20, 9900 Frederikshavn, Danmark

Designed for comfort and built to last.

Construction. 3-part thixotropic epoxy patching mortar. Product Description. Tests

Measuring instruments in industry.

LiftAlloy Chain Slings

Roughness measurements of stainless steel surfaces

Contents BUSHING S ELECTION P RODUCT S PECIFIC I NFORMATION TECHNICAL DATA A SSORTMENT. World s Widest Bushing Stock Assortment 2

SCENIC VIEW SV 10 SECTION SOLAR CONTROL FILMS

THRUST NEEDLE ROLLER AND CAGE ASSEMBLIES Assembly Dimensions, Load Ratings, Thrust Washer and Piloting Dimensions

Measuring thermal drift of machine tools

UDDEHOLM ELMAX SUPERCLEAN

ISO A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven

NORSOK STANDARD M-650 Edition 4. Qualification of manufacturers of special materials

Series 8146/5-V37 and Series 8150/5-V37

CMMs and PROFICIENCY TESTING

X15TN TM. A high hardness, corrosion and fatigue resistance martensitic grade CONTINUOUS INNOVATION RESEARCH SERVICE.

Brush Plating of Nickel-Tungsten Alloy for Engineering Application

FRAUNHOFER INSTITUTE FOR PRODUCTION TECHNOLOGY IPT PROCESS STRATEGIES FOR DEFECT-FREE, POLISHED STEEL SURFACES

Drill Pipe Hard-facing

TECHNICAL DATA SHEET GRILON BG-15 S

Transcription:

Using Equotip Hardness Test Blocks Heinz-Horst Pollok, consultant for material testing and instrument electronics (Cologne, Germany) Dr. Ralph T. Mennicke, product manager for material testing, Proceq SA (Schwerzenbach, Switzerland) Abstract Section 1 motivates the use of hardness test blocks. Section 2 introduces the requirements concerning the verification procedure for Leeb testers and the specifications of hardness test blocks according to standards. Section 3 gives an overview over the range of Equotip hardness test blocks. Section 4 describes how the daily verification of Leeb instruments is carried out. Section 5 summarizes this article. 1. Introduction For all hardness test methods for metallic materials and even more generally the reliability of test results depends on the stability of the measuring device. Due to staining, abrasion or damage of the indenter, misalignment or damage of the instrument mechanics, failure of the sensor system or evaluation electronics or due to other effects, measuring results could change over time, either slowly or abruptly. In order to detect such changes in due time, it is common to verify the operation of the testing device using calibrated hardness reference blocks. Corresponding national and international standards as well as individually agreed acceptance test procedures demand such checks more or less mandatorily. This also applies to hardness testing according to Leeb, which is standardized in ASTM A956 and DIN 50156 [1,2]. A more general overview over hardness reference material is given in [4]. 2. Verification of Leeb hardness testers according to standards With the launch of the first Equotip Leeb hardness tester by Proceq, a periodic verification of the instrument was introduced analogous to the standard practice in traditional hardness testing. In the two most commonly used Leeb standards, ASTM A956 and DIN 50156, the periodic verification has also been included. 1. Proceq s instructions (since 1975): Proceq recommends considering an instrument as successfully verified when the arithmetic mean value of >3 readings does not differ by more than 6 HL from the reference calibration value. The >3 impacts shall be randomly distributed over the test surface of an Equotip test block. Since 2009, Proceq has been increasingly urging to users to select a test block whose hardness is similar to the expected hardness of the samples under test. Users who conduct the sample test according to a national standard are advised to adhere to the instructions therein for daily verification. 2. ASTM A956 (first published in 1996, revised in 2002 and 2006): The periodic verification is specified in section 7 and described in more detail in sections 13 to 15 of the standard. The instrument shall be verified prior to each work shift, work period of use, and following a period of extended continuous use (1000 impacts), as well as when parts of the test equipment are replaced. At least in the latter case, the use of multiple test blocks is strongly recommended: the softest block should be softer than the minimum expected sample hardness, one block hardness should fall well within the expected sample range, and the hardest block should exceed the maximum expected sample hardness. Two impact indentations shall be made on a calibrated Leeb hardness test block. The instrument is successfully verified if neither reading differs from the reference set-point value by more than 6 HL.

3. DIN 50156 (first published in 2007): Appendix A of Part 1 (out of 3) deals with the periodic verification. For each impact device type, three s are specified. For the most common types D and G, these are as follows: type D < 500 HLD 500 700 HLD > 700 HLD type G < 450 HLG 450 600 HLG > 600 HLG Tab. 1: Hardness ranges for the daily verification of Leeb impact devices types D and G according to the test standard DIN 50156. The performance check of the instrument should be carried out prior to use on each day the instrument is used, at approximately each hardness level and for each scale that is to be used. At least 3 impacts should be made on a calibrated Leeb hardness test block. The instrument is considered as verified satisfactory if the arithmetic mean value and the reference calibration value do not differ by more than 15 HL. From this, it seems ASTM requires a closer matching of the measurement and the real value than DIN 50156. However, this is put into a different perspective when considering that DIN refers to certified reference material traceable to a national etalon held by Physikalisch-Technische Bundesanstalt (PTB) in Germany. This fixes the Leeb hardness scales according to DIN 50156. In contrast, the ASTM standard is not based on a standardized national Leeb scale, but test blocks are calibrated with Leeb testers certified by the instrument manufacturers. Different testers may hold systematic deviations between each other that cannot be revealed through checks on the same manufacturer s blocks. Only a test on a genuine Equotip block would permit determination of the bias (and thus the uncertainty) with respect to the inventor s genuine Leeb scale and other manufacturers in-house standards. 3. Range of Leeb hardness test blocks The standard specifications of Leeb hardness test blocks in DIN 50156-3 and ASTM A956, section 17 are very similar. The key parameters are listed in Tab. 2. type of impact device D/DC, DL, C, S, E / G diameter of test block 90 mm / 120 mm) thickness of test block 54 mm / 70 mm) roughness of calibrated test surface ASTM: R a 0.4 µm DIN: R a 0.025 µm roughness of other surfaces ASTM: R a 7 µm DIN: not specified homogeneity of calibrated test ASTM: R 1) < 13 HL surface DIN: V 2) 2.0 % (< 450 HL) V 1.5 % (450 to 750 HL) V 1.0 % (> 750 HL) reference calibration value average from 10 readings, randomly- and evenly-distributed over the test surface two-sided calibration of test block permitted 1) R: range, difference between the lowest and the highest reading out of the 10 readings 2) V: variation coefficient, standard deviation of 10 readings, divided by their average Tab. 2: Requirements on Leeb hardness test blocks according to ASTM A956 and DIN 50156-3. The ASTM homogeneity requirement is generally slightly tighter than the DIN specification. The most striking difference between Leeb test blocks and blocks for static hardness tests (such as HRC or HB) is the size of the blocks. The larger block size is important as the Leeb method assumes that the dynamic Leeb test impact is carried out on a sufficiently large and heavy sample.

Reference [5] reasons that the range of readings (min/max), or alternatively the variation coefficient, is not a genuine measure of the homogeneity of the test block surface. In addition, readings often decrease slightly from the center to the edge of the Leeb test blocks due the measurement principle. Proceq offers the widest range of Leeb hardness test blocks worldwide. 1. Hardness test blocks of diameter ~90 mm, thickness ~55 mm, weight ~2.8 kg: impact device type low (<225 HV / <220 HB) mid (~35 HRC / ~335 HV / ~325 HB) high (~630 HV / ~56 HRC) very high (~800 HV / ~63 HRC) D/DC 1) < 500 HLD ~600 HLD ~775 HLD DL <710 HLDL ~780 HLDL ~890 HLDL C <565 HLC ~665 HLC ~835 HLC S 2) 2) ~815 HLS ~875 HLS E 2) 2) ~740 HLE ~810 HLE 1) both two- and one-sided test blocks available 2) additional calibration available on test blocks for the other device types 3) not recommended for use in this due to increased wear of impact body Tab. 3: Proceq s range of Equotip hardness test blocks for the daily verification of Leeb hardness testers types D/DC, DL, C, S and E. 2. Hardness test blocks of diameter ~120 mm, thickness ~90 mm, weight ~6.3 kg: impact device type low (<200 HB) mid (~340 HB) G < 450 HLG ~570 HLG D/DC, DL, C, S, E 1) 1) 1) additional calibration available on test blocks for the other device types Tab. 4: Proceq s range of Equotip hardness test blocks for the daily verification of Leeb hardness testers type G.

Hardness test blocks calibrated by Proceq are supplied along with calibration certificates. The certificates contain the same data as such issued by accredited test laboratories. The calibration procedure is carried out as specified in DIN 50156-3 and ASTM A956. In particular, Proceq s plantinternal primary reference measuring device is identical in construction to the German national standard Leeb measuring device of Physikalisch-Technischen Bundesanstalt (PTB) in Germany and National Institute of Metrology (NIM) in China. In addition, the reference values are engraved in the test surface and quoted on barcode labels on the block circumference. The validity of Proceq test block certificates is not bindingly limited in time. However, they contain a reference to DIN 50156-3, according to which the validity should be limited to 5 years. In theory, it should be highlighted that the hardness of the material can change with time, particularly when speaking of very hard and very soft material. From experience, however, Equotip hardness test blocks have proven stable even over much longer periods of time provided proper storage. Supplemental to Proceq s calibration certificate, additional calibrations from accredited institutes are available for Equotip test block surfaces: Leeb hardness scales HLD, HLE, HLS and HLG according to DIN 50156-3 Brinell hardness scale HB according to ISO 6506-3 Vickers hardness scale HV according to ISO 6507-3 Rockwell hardness scales HRC, HRB according to ISO 6508-3 The reference measurement equipment used by the accredited institute for these calibrations is traceable to national hardness etalons. 4. How to carry out periodic instrument checks According to DIN 50156-1, the periodic check of Leeb hardness testers shall be carried out at ambient air temperatures between 10 and 35 C. ASTM A956 limits the test piece temperature to 4 to 38 C. If this is not possible, the user should report the relevant temperature in the test report. Remove the protective sticker or alternative corrosion protection from the test surfaces of the hardness test block. Carefully clean the test surface with acetone (or any other acid-free non-polar cleaning agent). Clean the ball indenter of the impact body with acetone. Place the test block on a plane solid support. The support must not move or vibrate during the measurement. A plane and smooth steel plate of at least 200 x 200 mm and a thickness of more than 25 mm is ideally suited. To ensure good contact and avoid scratching of the backside test surface, sandwich a plastic film between support and test block. As described further above, a number of Leeb impacts shall be distributed evenly across the test surface. Take care that the distance between the edges of two adjacent indentations exceeds three times the indentation diameter. Otherwise, higher hardness values will be measured: around existing indentations, the material has been cold work hardened [3]. The measurements shall be compared against the reference set-point value as per instructions given in the referred standards. If the periodic check fails, carefully clean the impact device with the supplied cleaning brush, and then repeat the check. It is recommended to keep records of the results of the check, quoting at least the serial number of the device, the serial number of the test block, the results, the date of the check, and the name of the inspector. In addition, any circumstances should be reported that can have affected the test result (such as an ambient temperature outside the limits specified in the relevant test standard). When finished, clean both test surfaces of the block with acetone. Even slight hand perspiration under finger prints may cause rapid corrosion of the surfaces! Protect the test surfaces against corrosion (e.g. with protective stickers or acid-free oil). Hardness test blocks must be stored in a dry place that does not undergo excessive temperature variations.

With an increasing number of indentations on the surface of a test block, the measured hardness values will increase and scatter more [4]. If there are too many indentations on a surface to adhere to the minimum distance of 3 indentation diameters between indentation edges, the test surface must not be used anymore. Neither must test surfaces be used if they are corroded, (heavily) scratched or have other obvious damages. Never recycle test surfaces by grinding off the upper skin. On the one hand, the hardness may be affected from the machining. On the other hand, remainders of the cold work hardened zones of the ground indentations will lead to higher hardness values and larger scatter of the values. And finally, the block parameters specified in Tab. 2 will not normally be met anymore. Conclusions As for any other hardness test devices, the accuracy and stability of Leeb hardness testers should periodically be verified using calibrated Equotip hardness test blocks. The test standards ASTM A956 and DIN 50156 specify the requirements on the hardness test blocks and provide very useful guidelines for instrument checks. Proceq offers a portfolio of Leeb hardness test blocks that covers the requirements of these standards. Also traceability to primary standards is secured. Periodic instrument checks with Equotip hardness test blocks are pretty much straightforward. When a few simple rules are followed, there should hardly emerge any problems. For further information please contact your Proceq representative or visit http://www.proceq.com and http://www.equotip.com. References [1] ASTM A956, Standard Test Method for Leeb Hardness Testing of Steel Products, 2006. [2] DIN 50156, Metallische Werkstoffe Härteprüfung nach Leeb, 2007. [3] K. Herrmann et al., Härteprüfung an Metallen und Kunststoffen, Expert Verlag, 2007. [4] T. Yamamoto, Role and use of standard hardness blocks, 2002. [5] T. Yamamoto, M. Yamamoto, K. Miyahara, Accuracy for Standard Blocks for Hardness and Uncertainty of Hardness, XIX IMEKO World Congress, 2009.