Indirect Blood Pressure Measurement



Similar documents
Urethral Catheterization in the Female Dog Made Easy Jamie M. Burkitt-Creedon, DVM, Dipl. ACVECC, Critical Consultations

Section Two: Arterial Pressure Monitoring

MEASURING AND RECORDING BLOOD PRESSURE

How to measure blood pressure using digital monitors

Placement of Epidural Catheter for Pain Management Shane Bateman DVM, DVSc, DACVECC

Blood Pressure Assessment Program Screening Guidelines

Section Four: Pulmonary Artery Waveform Interpretation

OEM MAXNIBP Frequently Asked Questions

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Plank 1 Tool: Hypertension Medical Assistant Training (Providence Medical Group)

Senior pets are not unlike senior citizens.

Placement of an indwelling urinary catheter in female dogs

Reading Assignment: Tasks for the Veterinary Assistant, pages , Appendix L, O and P

Responses to Static (Isometric) Exercise

Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions

HEALTH MANAGEMENT PLAN PROGRAMME

PTA 210 PTA Technique. Vital Signs

Manual Blood Pressure Measurement

CRITICAL CARE MONITORING: WHAT, WHEN, AND HOW? Susan G. Hackner, BVSc.MRCVS.DACVIM.DACVECC.

Work Instructions. Title: Rat Restraint Techniques 1. GENERAL GUIDELINES

NASAL OXYGEN ADMINISTRATION Bernie Hansen DVM MS North Carolina State University College of Veterinary Medicine

Measurement of Blood Pressure

An Overview of Transradial Patient Set-up. Susan R. Cooney RN,BSN,CCRN Durham VAMC Cardiac Catheterization Lab Duke University School of Nursing

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Central Venous Catheters for Veterinary Technicians

Dialysis Professional

MEASURING AND RECORDING BLOOD PRESSURE

DOG AND CAT HANDLING AND RESTRAINT

MEASURING VITAL SIGNS TRAINING CHECKLIST

Measuring Your Blood Pressure at Home. A Review of the Research for Adults

Body cavities. Body Planes

Patient Possible differentials Recommended diagnostics Puppy or kitten with a soft systolic murmur

Accurate Air Flow Measurement in Electronics Cooling

CFMRI Physiological Monitoring System. Operator s Manual

Automatic Wrist Blood Pressure Monitor with

Blood pressure measurement devices. December 2013

Circulation Stations

Blood Pressure Guidelines and Screening Techniques

Lunette 2 Series. Curved Fixed Frame Projection Screen. User s Guide

Normal & Abnormal Intracardiac. Lancashire & South Cumbria Cardiac Network

Subclavian Steal Syndrome By Marta Thorup

EQUIPMENT SET UP RECURVE BOW

Operating Manual FINGERSIM SET /5.11. for SECULIFE NIBP in Combination with SECULIFE OX1

Upper Arm Blood Pressure Monitor SC Instructions for use

IMPORTANT INFORMATION - PLEASE READ. Table of Contents. Introduction. General Information

Normal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network

PetSafe Spray Bark Control Collar

24 Hour Blood Pressure Monitoring A guide to the test

Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003

HAND-ARM VIBRATION SYNDROME ASSESSMENT TESTS

KAWAI VERTICAL PIANO REGULATION MANUAL

Chapter 11. Everting skin edges

Arterial pressure monitoring Direct arterial pressure monitoring permits continuous measurement of systolic, diastolic, and mean pressures and allows

HEART MONITOR TREADMILL 12 LEAD EKG

D-Luciferin In Vivo Protocol

Assignment 3 - Instructional Goal Subordinate Analysis

Dog Kennel WORLDWIDE

Instructions for Drawing Blood Samples

Fellow TEE Review Workshop Hemodynamic Calculations Director, Intraoperative TEE Program. Johns Hopkins School of Medicine

GUIDE TO A PERFECT MEN S SHIRT FIT

Policies and Procedures. Related to. IABP Therapy

Project 4.2.1: Heart Rate

Equine Sedation, Anesthesia and Analgesia

Vasc Health Risk Manag. 2011; 7: Published online 2011 November 30. doi: /VHRM.S27193 PMCID: PMC

Time: Depending on the size and behavior of the dog, this procedure will take no longer than approximately two hours.

Vtial sign #1: PULSE. Vital Signs: Assessment and Interpretation. Factors that influence pulse rate: Importance of Vital Signs

BT-3. Bark Control Collar. Please Read All Instructions Carefully.

Measuring central venous pressure

INSTRUCTION MANUAL. Ultra Premium Blood Pressure Monitor featuring ComFit TM Cuff. Model HEM-775

How can I be sure it will be performed to a high standard?

Electroconvulsive Therapy (ECT)

INSTRUCTION MANUAL. 7 series Blood Pressure Monitor with ComFit TM Cuff. Model BP760 ESPAÑOL ENGLISH

PRACTICE GUIDELINE TITLE: INTRAVENOUS LINE INSERTION: PERIPHERAL AND CENTRAL

TENS and Electrodes Guide

X-Plain Varicose Veins Reference Summary

SURGICAL EXTRACTIONS: TECHNIQUE AND CAUTIONS By Tony M. Woodward, DVM, AVDC

Heart Rate and Physical Fitness. Evaluation copy

AMBULATORY BLOOD PRESSURE MONITORING QUICK START GUIDE

CARDIAC REHABILITATION PROGRAM

Signal Extraction Technology

Medical Section. acmedical@aircanada.ca. Fax : (toll-free) or

Fact sheet Exercises for older adults undergoing rehabilitation

REARWARD- & FORWARD-FACING USER MANUAL ECE R GROUP WEIGHT AGE 0+/ kg 6m-4y

HEART MURMURS. Information Leaflet. Your Health. Our Priority.

Anesthesia Technician

Cardiovascular Physiology

Preparation guidelines for your Child s Sleep Study

Child in Car Safety Leaflet 1/8/07 8:54 am Page 1 Child In Car Safety

OPERATING P L AYA B I L I T Y P R E C I S I O N D E S I G N

The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology

Practical class 3 THE HEART

Cold Agglutination Titer detecting Cold Reacting Antibodies

P R E S S U R E P O I N T S S E R I E S : Measuring your blood pressure at home

Setting Up and Using the Omnisense 8000S

STANDARD OPERATING PROCEDURE #201 RODENT SURGERY

Transcription:

P r o c e d u r e s P r o C A R D I O L O G Y Peer Reviewed Indirect Blood Pressure Measurement Maintenance of appropriate systemic arterial blood pressure is vital for survival. Because many common situations and conditions (eg, general anesthesia, renal or cardiovascular disease) place animals at risk for dangerously high or low blood pressure, blood pressure monitoring is a fundamental necessity. In clinical practice, blood pressure is usually measured indirectly. The two most common methods of indirect blood pressure measurement are Doppler ultrasound using a sphygmomanometer and oscillometry. Both techniques use an inflatable cuff placed around an extremity. Highdefinition oscillometry, a newer indirect technique, is not used often in clinical practice. DOPPLER ULTRASOUND MEASUREMENT Doppler ultrasound measurement is likely the most common blood pressure measurement technique used in small animal practice. The procedure is straightforward, and the equipment is relatively inexpensive and reliable. WHAT YOU WILL NEED For Doppler Ultrasound Blood Pressure Measurement Doppler ultrasound with attached Doppler probe Headphones, if desired Sphygmomanometer Tape measure with metric scale Cuff of appropriate size Clippers with clean blade Ultrasound transmission gel For Standard Oscillometry Blood Pressure Measurement Oscillometer with pressure tubing attached Tape measure with metric scale Cuff of appropriate size 26...NAVC Clinician s Brief / May 2012 / Procedures Pro

Jamie M. Burkitt Creedon, DVM, DACVECC Red Bank Veterinary Hospital Network Cherry Hill, New Jersey Doppler ultrasound reliably measures only systolic blood pressure in dogs and cats; it cannot be used to measure mean or diastolic blood pressure. STANDARD OSCILLOMETRY MEASUREMENT In clinical practice, oscillometers are often built into multiparameter monitors and usually have a stat feature that allows continual, repeated measurements. Thus oscillometers are commonly used for critically ill or anesthetized patients that may require more intensive monitoring. Oscillometers report systolic, diastolic, and mean blood pressures; however, they are unreliable in patients with cardiac dysrhythmias because measurement relies on rhythmic arterial pulsations. GENERAL TECHNIQUES While actual blood pressure measurement varies between the techniques described here, some important points apply to both. Patient Acclimation & Restraint Ideally, blood pressure should be measured in a quiet, comfortable environment after the patient has become acclimated (ie, without disturbance) for at least a few minutes, but acclimation is rarely possible for acutely or critically ill patients. The patient should remain still during measurements to optimize accuracy and, if necessary, can be physically restrained gently. However, some patients cannot be restrained in lateral recumbency because of agitation, respiratory distress, or other reasons. Chemical restraint should be avoided, as it often alters results. Cuff Height To acquire the most accurate blood pressure values, cuff height should be as close to the level of the right atrium as possible (Step 1A, page 28). The cuff can be placed on the antebrachium, distal pelvic limb, or tail. If the vertical distance between the cuff site and the right atrium exceeds 10 cm, a correction factor must be integrated to interpret blood pressure measurement results (see Correcting for Cuff & Right Atrium Level Disparity, page 28). When the cuff is much lower than the right atrium, pressure values are falsely high because of the weight of the blood column between the atrium and cuff site. Multiple Measurements The first blood pressure measurement should be discarded. At least 3 additional measurements should be obtained for a minimum of 3 to 7 consecutive, consistent measurements with less than 20% variability. Once these measurements have been obtained, they should be averaged to yield the blood pressure measurement. C O N T I N U E S All photos of Beezus courtesy of Dr. Burkitt Creedon Procedures Pro / NAVC Clinician s Brief / May 2012...27

P r o c e d u r e s P r o C O N T I N U E D STEP-BY-STEP DOPPLER ULTRASOUND BLOOD PRESSURE MEASUREMENT STEP 1 Gather supplies and select appropriate cuff size. The numeric cuff size (ie, width in centimeters) should be 40% of the circumference of the cuff site in dogs and 30% to 40% of the circumference of the cuff site in cats. Measure the site using a centimeter-marked tape (A) or estimate using the cuff itself (B). A cuff that is too small or tight results in falsely elevated readings, and a cuff that is too large or loose results in falsely low readings. A B CORRECTING FOR CUFF & RIGHT ATRIUM LEVEL DISPARITY If the height difference between the cuff and the right atrium exceeds 10 cm, subtract 0.8 mm Hg for every 1 cm the cuff sits below the right atrium. In lateral recumbency, the sternum can generally be used as the reference. In sternal recumbency the right atrium level is at about 40% the height of an imaginary line between the sternum and the dorsal spinous process at the caudal edge of the scapula. STEP 2 Clip the hair overlying the Doppler probe s target artery. The target artery must lie distal to the cuff site. When using the antebrachium for cuff placement, clip over the palmar aspect of the metacarpals (shown). When using the distal pelvic limb, clip from the dorsal or plantar surface of the metatarsals for access to the dorsal pedal artery or branch of the saphenous artery, respectively. When placing the cuff at the tail base, clip from the tail s ventral midline just distal to the intended cuff site. Here, the midpoint of the cuff sits approximately 27 cm below the right atrium. If the measured systolic pressure at this level were 142 mm Hg, correct the measurement as follows to yield a more accurate reading: 27 cm 0.8 mm Hg/cm = 21.6 mm Hg 142 mm Hg 21.6 mm Hg = 120.4 mm Hg 28...NAVC Clinician s Brief / May 2012 / Procedures Pro

STEP 3 Apply the cuff securely and snugly around the appendage, and attach it to the sphygmomanometer. The cuff should not be occlusively tight, and there should be no room to insert a finger between the cuff and appendage. The sphygmomanometer pressure should read 0 mm Hg. STEP 4 Apply copious acoustic gel to the concave surface of the Doppler probe, and press the concave surface gently but firmly over the target artery. Turn the Doppler unit on and listen for the pulsatile whooshing sound of the artery. With the patient in lateral or sternal recumbency, it is usually possible to hold the limb comfortably at approximately the level of the right atrium (see Step 5). Consider using headphones to minimize ambient noise and to avoid frightening the patient. STEP 5 Use the sphygmomanometer to inflate the cuff to approximately 20 mm Hg greater than the expected systolic blood pressure. The cuff should occlude the artery at this pressure, and whooshing arterial pulsations should be inaudible. Slowly deflate the cuff and note the pressure at which arterial pulsation is again audible; this marks the systolic arterial blood pressure. Most commercial cuffs have Velcro closures that can loosen during inflation. Medical tape can be placed to secure the cuff s closure, but be certain not to further tighten the cuff with the tape, which would invalidate the measurement. This is true for both techniques. Deflate the cuff only a few mm Hg/sec in order to hear the first pulse wave passing under the cuff, heralding systolic pressure. Opening the valve too quickly results in falsely low readings because the true systolic pressure is missed. C O N T I N U E S Procedures Pro / NAVC Clinician s Brief / May 2012...29

P r o c e d u r e s P r o C O N T I N U E D STEP-BY-STEP STANDARD OSCILLOMETRY BLOOD PRESSURE MEASUREMENT STEP 1 Note: Same as Step 1 for Doppler Ultrasound Blood Pressure Measurement. Gather supplies and select the appropriate cuff size. The numeric cuff size (ie, width in centimeters) should be 40% of the circumference of the cuff site in dogs and 30% to 40% of the circumference of the cuff site in cats. Measure the site using a centimetermarked tape (A) or estimate using the cuff itself (B). A cuff that is too small or tight results in falsely elevated readings, and a cuff that is too large or loose results in falsely low readings. A B STEP 2 Place the cuff securely and snugly around the appendage. It should not be occlusively tight, but there should be no room to insert a finger between the cuff and the appendage. Attach the cuff to the oscillometer. STEP 3 Turn on the oscillometer and depress the measurement button while holding the cuff at approximately the height of the right atrium. See Aids & Resources, back page, for references & suggested reading. Support the dog s limb from underneath while restraining the dependent limb. Many dogs resent having their paws handled, and this technique minimizes patient movement and tension in the limb, which can adversely affect results. 30...NAVC Clinician s Brief / May 2012 / Procedures Pro