Scripps Proton Therapy Center: Configuration and Implementation



Similar documents
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

IBA Proton Therapy. Biomed days Vincent Bossier. System Architect Protect, Enhance and Save Lives

ADROTERAPIA A MEDAUSTRON

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Quality Assurance in Stereotactic. Radiotherapy

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

HADRON THERAPY FOR CANCER TREATMENT

Proton Therapy - Frequently Asked Questions For Clinicians

World-first Proton Pencil Beam Scanning System with FDA Clearance

QA of intensity-modulated beams using dynamic MLC log files

How To Treat Cancer With Proton Therapy

Use of 3D Printers in Proton Therapy

Worldwide Quality Assurance networks for radiotherapy dosimetry

HITACHI Proton Beam Therapy System

Quality Assurance of Radiotherapy Equipment

THE DOSIMETRIC EFFECTS OF

Radiotherapy in Hungary: present status and future needs. Tibor Major, PhD National Institute of Oncology Radiotherapy Department Budapest, Hungary

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry

Secondary Neutrons in Proton and Ion Therapy

2015 ASTRO INVESTOR MEETING. October 20, 2015

Current Status of Proton Clinical Activities at PTC H

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

Status and perspective of emission imaging techniques for ion beam therapy in Lyon

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

Overview of Proton Beam Cancer Therapy with Basic Economic Considerations

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

The Science behind Proton Beam Therapy

First Three Years After Project Proton Therapy Facility:

Proton Therapy. What is proton therapy and how is it used?

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Proton Therapy Essentials

Current Status and Future Direction of Proton Beam Therapy

Quality Assurance for Treatment Planning Systems

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

THE POWER AND PRECISION OF PROTON BEAM THERAPY IS WITHIN REACH

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

In 1946 Harvard physicist Robert Wilson ( ) suggested:

Digital vs. Analogue Control Systems

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy

Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group

Quality Assurance of accelerators; the technologists responsibility

Proton beam. Medical Technical Complex Joint Institute for Nuclear Research, Dubna, Russia

A Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock

Clinical Education A comprehensive and specific training program. carry out effective treatments from day one

Proton therapy, yesterday, today and tomorrow

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Proton tracking for medical imaging and dosimetry

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

Proton therapy out of reach? Think again!

ACCELERATORS AND MEDICAL PHYSICS 2

CONQUERING CANCER SAVING LIVES. Mayo Clinic Proton Beam Therapy Program

ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic

Northern Illinois. proton therapy facility

Intensity-Modulated Radiation Therapy (IMRT)

Proton Radiotherapy. Cynthia Keppel Scientific and Technical Director Hampton University Proton Therapy Institute. Lead Virginia.

Proton Therapy Center at the Institute of Nuclear Physics in Kraków - from the Eye Treatment to the Scanning Gantry

PROTON THERAPY FREQUENTLY ASKED QUESTIONS

Challenges in small field MV photon dosimetry

CBCT for Prone Breast. Todd Jenkins, MS, DABR Nash Cancer Treatment Center

Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.

Failure Modes and Effects Analysis (FMEA)

Our Department: structure and organization

Design and Scheduling of Proton Therapy Treatment Centers

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

HDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer

Proton therapy at Kraków, Poland: Fighting cancer with accelerator technologies

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Therapy with protons and ion beams

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

The feasibility of a QA program for ISIORT Trials

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA USA phone NOMOS

Medical Physics Residency Program Department of Radiation and Cellular Oncology The University of Chicago Chicago, IL 60637

Key words: treatment planning, quality assurance, 3D treatment planning

Proton Therapy: Cutting Edge Treatment for Cancerous Tumors. By: Cherilyn G. Murer, JD, CRA

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS

Safety Risk Management in RT: A Software Manufacturer Perspective

RADIATION THERAPY guide. Guiding you through your treatment

How To Use A Proton For Radiation Therapy

How To Treat Cancer With Proton Therapy

Transcription:

Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA

Facility Configuration

Scripps Proton Therapy Center ***All Rooms : PBS Only 360 o Gantry Fixed Horizontal Beamline Cyclotron Max E : 250 MeV

Scripps Proton Therapy Center Gantry-mounted dual x-ray imaging system Full CBCT function Internal and external imaging console TrueBeam-like treatment console Robotic, 6-degree of freedom patient positioner

Scripps Proton Therapy Center Designed treatment capacity: 2400 pts per year Planned patient mix: Prostate Cancer: 800 patients/year Lung Cancer: 200 patients/year. Head & Neck Cancers: 200 patients/year Pediatric Cancer: 100 patients/year Breast Cancer: 200 patients/year CNS Tumors: 400 patients/year Other tumor sites

Scripps Proton Therapy Center Physicians (4) Carl Rossi Huan Giap, Ryan Grover, Andrew Chang Physicists (7) Lei Dong Annelise Giebler, Anthony Mascia, Yongbin Zhang, Franko Piskulich, Richard LePage, Luis Perles Physicist Assistants To be hired: January 2014 Dosimetrists (2) First cohort started: July 2013 Radiation Therapists (5) First cohort started: August 2013

Quality Assurance

Daily Machine QA Duration 15 minutes; performed by RTTs In short, an end-to-end test using daily QA detector Rotate proton energy daily, Monday Through Friday PBS volumetric irradiations (i.e. not single spots) Courtesy PTW Ding X, Zheng Y, Zeidan O, Mascia A, Hsi W, Kang Y, Ramirez E, Schreuder N, Harris B. A novel daily QA system for proton therapy. J Appl Clin Med Phys. 2013 Mar 4;14(2):4058

Weekly Machine QA Summary Comprehensive spot pattern tests Initially performed: weekly Deliver a fixed spot pattern at two gantry angles, two energies This spot pattern is benchmarked at commissioning The physical measurement results and their comparison to benchmark are recorded 30 minutes per room Test Parameters Position location of spot on Lynx versus planned location Shape ratio of x-axis and y-axis Size sigma of x-axis and y-axis Output output constancy Courtesy IBA Dosimetry

Monthly Machine QA The Daily + Weekly QA programs are comprehensive and reviewed daily/weekly; detailed trend analysis performed as part of monthly QA Monthly QA tests a wider spectrum of the beam delivery system in one session (e.g. more ranges, more doses, etc) Precise quantification of the treatment couch using Winston-Lutz style tests performed at cardinal gantry angles Image quality assessment of the imaging system

Annual QA As opposed to constancy measurements, the annual QA re-measures / re-validates baseline/commissioning data using commissioning caliber equipment Detectors: Water tank + Bragg peak chamber Water tank + Farmer chamber Lynx or Logos scintillator detectors OctaviusXDR ion chamber array Absolute Calibration: IAEA TRS 398

Patient Specific QA Every field for every patient undergoes quality assurance procedure prior to treatment Procedure highlights: For each field, three transverse profiles, each at different depth, measured with 2D ion chamber array; compared to calculation in Eclipse For each field, at least one point dose measurement is made using cross-calibrated ion chamber (i.e. IBA Dosimetry CC04) Measurement setup: 1. Gantry 0 degrees: all field recalculated to 0 degrees; all fields measured at 0 degrees 2. Planned Gantry Angle : all fields measured at planned gantry angles and compared to calculation Passing criteria is being developed Starting point: 3% / 3mm with 90% passing Evalute and determine during commissioning and validation

Novel Monthly Machine QA Technique Courtesy PTW Use OctaviusXDR or StarCheckMaxi + BQ Check phantom Using single detector, measure: spot position, dose constancy, proton range/energy, field uniformity (symmetry/flatness) Additional setup (e.g. imaging bb + Lynx mounted to gantry) required for Winston Lutz tests Testing/development with Varian, Rinecker Proton Therapy Center and Scripps Proton Therapy Center

Novel QA Implementation Logos Systems Inc XRV device: Conical scintillator-based detector Alignment: laser, imaging device, and beam isocenter. Isofocus Beam Summary Beam Center X Y Z Average: 0.22 0.09 75.71 Max: 0.49 0.41 76.33 Min: -0.03-0.23 74.84 StDev: 0.18 0.23 0.43 FD= 69.44 FE= 79.45 Laser @ 0.64 of DE: Laser @ 75.8464 X (mm) Y (mm) Z (mm) Laser Error 0.22 0.09-0.14

Novel Patient QA Technique Aluminum mount interfaces to Varian nozzle OctaviusXDR array rigidly held in place 4 x Slide Nuts securely hold up to 25.0cm of solid water in front of the ion chamber array Allows transverse plane at any water equivalent depth (up to 25.0cm) Because Varian nozzle holder translates, also allows measurements upstream and down stream of isocenter Thanks to Glen Mounce, machinist

Acknowledgements SPTC Colleagues, especially the physics group: Lei Dong Franko Piskulich Richard LePage Annelise Giebler Yongbin Zhang Luis Perles Our partners at Advanced Particle Therapy (APT) Our partners at Varian Medical Systems, Particle Therapy Group