Technical guidelines for magnetic resonance imaging for the surveillance of women at increased risk of developing breast cancer



Similar documents
SITE IMAGING MANUAL ACRIN 6698

Recommendations for cross-sectional imaging in cancer management, Second edition

Guideline for the Imaging of Patients Presenting with Breast Symptoms incorporating the guideline for the use of MRI in breast cancer

CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME

Breast Cancer. Presentation by Dr Mafunga

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010

The File-Card-Browser View for Breast DCE-MRI Data

NOTTINGHAM UNIVERSITY HOSPITAL NHS TRUST NOTTINGHAM BREAST INSTITUTE BREAST AND OVARIAN FAMILY HISTORY GUIDELINES

Musculoskeletal MRI Technical Considerations

Breast MRI Quality Control

Sustaining a High-Quality Breast MRI Practice

Mammography. What is Mammography?

Breast MRI: Imaging and Intervention. Jaroslaw Nicholas Tkacz, M. D.

Patient Prep Information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

Breast Imaging Made Brief and Simple. Jane Clayton MD Associate Professor Department of Radiology LSUHSC New Orleans, LA

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

OBJECTIVES By the end of this segment, the community participant will be able to:

A Guide to Breast Imaging: The Latest Technology for Screening and Detecting Breast Cancer

7/22/2011. Breast MRI: Pulse Sequences, Acquisition Protocols, and Analysis. Objectives. Challenges in DCE Breast Imaging

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

Nicole Kounalakis, MD

D. FREQUENTLY ASKED QUESTIONS

Certification protocol for breast screening and breast diagnostic services

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

Infrared Thermography Not a Useful Breast Cancer Screening Tool

Prepublication Requirements

Digital Mammogram National Database

Q: What differentiates a diagnostic from a screening mammography procedure?

Coronis 5MP Mammo. The standard of care for digital mammography

MRI for Paediatric Surgeons

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

Automated Breast Volume Scanning 3D Ultrasound of the Breast

THE ROYAL COLLEGE OF RADIOLOGISTS

Surgical guidelines for the management of breast cancer

Medicare Part B. Mammograms - Updated Billing Guide for Screening and Diagnostic Tests

III. REPORTING SYSTEM

This vision does not represent government policy but provides useful insight into how breast cancer services might develop over the next 5 years

Evaluation and Management of the Breast Mass. Gary Dunnington,, M.D. Department of Surgery Internal Medicine Ambulatory Conference December 4, 2003

The Role of the MDT Coordinator. Laura Throssell

How To Be A Breast Consultant Radiographer

Office of Population Health Genomics

In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer

Patient sample criteria for the OPEIR Measures Group are all patients regardless of age, that have a specific CT procedure performed:

Digital Image Processing with DragonflyPACS

Phyllodes tumours: borderline malignant and malignant

Hologic Selenia Dimensions C-View Software Module. October 24, 2012

MRI of the Breast. What is MRI of the Breast?

Invasive lobular breast cancer

MRI QA Technologist s Tests

Ductal carcinoma in situ (DCIS)

NHS breast screening Helping you decide

Breast Screening Explained. We can supply this information in other languages, in large print, on audio or in Braille.

MRI SEQUENCES. 1 Gradient Echo Sequence

Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD

There are many different types of cancer and sometimes cancer is diagnosed when in fact you are not suffering from the disease at all.

Focussing on Inequalities: Variation in breast cancer outcomes with age and deprivation

Breast cancer affects one in eight Australian women. It is the most common cancer for Victorian women, with almost 3,700 diagnoses in 2012.

OPTIMIZING PATIENT EXPOSURE TO IONIZING RADIATION (OPEIR) MEASURES GROUP OVERVIEW 2015 PQRS OPTIONS FOR MEASURES GROUPS:

VI. FREQUENTLY ASKED QUESTIONS CONCERNING BREAST IMAGING AUDITS

Official reprint from UpToDate UpToDate

Case Report: Whole-body Oncologic Imaging with syngo TimCT

Hereditary Breast and Ovarian Cancer (HBOC)

How can I be sure it will be performed to a high standard?

Provincial Quality Management Programs for Mammography, Colonoscopy and Pathology in Ontario

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute

BreastScreen and You. Information about mammographic screening

Breast Ultrasound: Benign vs. Malignant Lesions

GE Healthcare. Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together

Current Industry Neuroimaging Experience in Clinical Trials Jerome Barakos, M.D.

Segmentation of Breast Tumor from Mammographic Images Using Histogram Peak Slicing Threshold

Introduction Breast cancer is cancer that starts in the cells of the breast. Breast cancer happens mainly in women. But men can get it too.

Common Breast Complaints:

Measuring Size and Distance in Digital Mammography

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Voxar 3D TM. A suite of advanced visualization and analysis software tools

RADIOLOGY HOUSE STAFF MANUAL

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Department of Veterans Affairs VHA DIRECTIVE Veterans Health Administration Washington, DC February 8, 2011

Quantitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

How To Compare Mammography To A 3D System

5 Factors Affecting the Signal-to-Noise Ratio

Supporting people with learning disabilities to take care of their breasts - a guide for supporters. Your breasts, your health

Breast Cancer Screening

Protocol for the safe administration of iodinated contrast media in diagnostic radiology

Mammography as you ve never seen it before

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Personalized Breast Screening Service

CAR Standards for Teleradiology

MRI SAFETY JEOPARDY (Technologist Edition) Questions

GRID COMPUTING FOR DIGITAL MAMMOGRAPHY

Improvements in quality and quantification of 3D PET images

Guideline for the Follow Up of Patients Following Treatment for Breast Cancer

Getting to Know Avreo

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Radiology Department. Local Procedure

Guidelines for. Quality Assurance in Cervical Screening. Second Edition

Transcription:

Technical guidelines for magnetic resonance imaging for the surveillance of women at increased risk of developing breast cancer NHSBSP PUBLICATION No 68 DECEMBER 2009 1

EDITORS Dr Barbara JG Dall Consultant Radiologist, United Leeds Teaching Hospitals Prof Fiona Gilbert Roland Sutton Chair of Radiology and Head of Imaging, Institute of Medical Sciences, University of Aberdeen Prof Martin Leach Professor of Imaging, Institute of Cancer Research Dr Robin Wilson Consultant Radiologist, Breast Radiology, King s College Hospital NHS Foundation Trust PUBLISHED BY NHS Cancer Screening Programmes Fulwood House Old Fulwood Road Sheffield S10 3TH Tel: 0114 271 1060 Fa: 0114 271 1089 Email: info@cancerscreening.nhs.uk Website: www.cancerscreening.nhs.uk NHS Cancer Screening Programmes 2009 The contents of this document may be copied for use by staff working in the public sector but may not be copied for any other purpose without prior permission from the NHS Cancer Screening Programmes. The report is available in PDF format on the NHS Cancer Screening Programmes website. Version Control Technical guidelines for magnetic resonance imaging for the surveillance of women at increased risk of developing breast cancer Electronic publication date December 2009 Review date (1) June 2010 Review date (2) Comments May be sent to anny.jones@cancerscreening.nhs.uk in readiness for review. 2

CONTENTS 1. INTRODUCTION 4 1.1 AIMS OF THE GUIDANCE 4 1.2 BACKGROUND 4 1.3 PROVISION OF MRI AND MAMMOGRAPHY SURVEILLANCE 4 2. STANDARDS FOR EQUIPMENT AND IMAGE ACQUISITION 5 2.1 MRI SYSTEM 5 2.2 BREAST COIL 5 2.3 BREAST MOVEMENT 5 2.4 CONTRAST 5 2.5 TIMING OF EXAMINATIONS 5 2.6 SCANNING PARAMETERS 5 2.7 DICOM COMPLIANCE AND DATA IMPORT 6 2.8 VIEWING TOOLS 7 3. REPORTING MRI EXAMINATIONS 7 3.1 READING REQUIREMENTS 7 3.2 ASSESSMENT OF ABNORMALITIES 8 3.3 MRI GUIDED BIOPSIES 8 3.4 RECALL RATE 8 4. QUALITY CONTROL AND QUALITY ASSURANCE 9 4.1 QUALITY CONTROL 9 4.2 ACCREDITATION 9 4.3 MRI AUDIT DATABASE 10 REFERENCES 10 APPENDIX 2: QUALITY ASSURANCE STANDARDS FOR MRI BREAST SURVEILLANCE 14 3

1. INTRODUCTION 1.1 Aims of the guidance The NHS Breast Screening Programme (NHSBSP) is responsible for setting and monitoring the standards for imaging surveillance of women below the screening age who are assessed as being at increased risk of developing breast cancer. This may be because of a family history of breast cancer or previous radiotherapy treatment for Hodgkin s disease. The guidance sets out the technical standards required for the provision of breast MRI as part of the imaging surveillance of women at increased risk, and describes the arrangements for reporting MRI eaminations and for quality assurance of MRI surveillance. The guidance is based on the results of the magnetic resonance imaging breast study (MARIBS) 1 and has been compiled by a UK working party made up of members of the MARIBS steering group, representatives from industry including manufacturers of MRI equipment, and software companies producing tools for breast MRI analysis. 1.2 Background The Cancer Reform Strategy 2 recognised that the opportunities and protocols for surveillance of women identified as being at increased risk of developing breast cancer have been variable across the country. The strategy recommended that all women identified as being at increased risk should be offered the opportunity to have their risk formally assessed and, if appropriate, to discuss their management options, in accordance with guidelines published by the National Institute for Health and Clinical Ecellence (NICE). 3 The NHSBSP is responsible for managing this surveillance to national standards and ensuring that women receive a consistent and high quality service. Surveillance with digital -ray mammography and magnetic resonance imaging should be provided where appropriate and previous guidance issued for surveillance of women treated for Hodgkin s disease with mantle radiotherapy 4 should be taken into account. 1.3 Provision of MRI and mammography surveillance It is epected that 2500 women in England and Wales will be assessed by medical geneticists as being at high risk of breast cancer as a result of their family history and will qualify for imaging surveillance by MRI and mammographic screening. Women in this category will be screened only at breast screening centres that satisfy the criteria set out in these guidelines. It is anticipated that the risk assessment process will identify women in whom MRI or mammography is contraindicated. Centres that provide breast imaging using MRI must meet the national minimum standards relating to equipment, image acquisition and data storage. They must follow national protocols for reporting MRI eaminations and should participate in the audit of MRI surveillance. It is anticipated that symptomatic eaminations will be performed to the same high standard. The provision of MRI for surveillance will be monitored as part of breast screening quality assurance (QA) activities. Three demonstration sites have been set up to develop the clinical and administrative protocols for MRI surveillance of women at high risk. These protocols will be circulated separately. 4

2. STANDARDS FOR EQUIPMENT AND IMAGE ACQUISITION 2.1 MRI system High field modern MRI machines should be used. Very high field strengths (3 Tesla and above) can be used, provided images are of diagnostic quality and B1 transmit homogeneity is uniform Minimum standard: at least 1 Tesla Epected standard: 1.5 Tesla 2.2 Breast coil Dedicated bilateral breast coils must be used (open or closed) with uniform signal homogeneity across the coils. 2.3 Breast movement Breast movement should be minimised during the procedure in order to obtain the best quality dynamic data. 2.4 Contrast Pump injection of at least 0.1mmol/kg contrast is recommended, with a 3cc/sec flow rate and 20ml bolus saline. The contrast pump should be set to start after the pre-contrast scans, so that the contrast s arrival in tissue coincides with the centre of the k-space (approimately 20 seconds from the start of injection). UK recommendations on avoiding contrast reactions and nephrogenic systemic fibrosis should be followed. 5 2.5 Timing of eaminations The eamination should be carried in the mid portion of the menstrual cycle to reduce normal parenchymal tissue enhancement. Minimum standard: eamination in day 6-16 of the menstrual cycle 2.6 Scanning parameters Time: Total eamination time <30 mins Orientation: Both breasts should be eamined. This can be achieved by scanning in the aial (preferred) or coronal plane, or by using an interleaved sagittal approach. The choice of scanning plane will be determined by the viewer s preference and the need to optimise resolution. (See below.) Resolution o Spatial resolution High spatial resolution is needed to assess the lesion s morphology. Slice thickness (as defined by the specific slice profile) 2mm In plane resolution (as defined by the specific resolution) <1.0 mm o Temporal resolution High temporal resolution is needed for the dynamic sequences, so that lesion kinetics can be fully assessed. Dynamic scan time < 120s. Dynamic scan repeated out to 5-7 minutes after contrast is administered. 5

o Optimisation Spatial and temporal resolution should be optimised. If parallel imaging is available it should be used. The direction of parallel imaging may dictate the scanning orientation. Time saving techniques should be adopted: for eample, reducing the field of view in the phase encoding direction when scanning in the coronal plane. Sequences o T2-weighted fast spin echo Provides high resolution for characterizing lesions and improves specificity. Requires no fat suppression. o T1-weighted spoiled gradient echo Enables a dynamic set of sequences to be obtained before contrast is injected and then repeated as rapidly as possible for 5 to 7 mins after a rapid intravenous bolus of a contrast agent containing gadolinium. Provides 3D or 2D (multislice) images. A fat suppression technique should be used to improve the lesion s conspicuity. This can be obtained either by integral frequency based fat suppression (such as frequency-selective fat saturation or water ecitation) of the dynamic sequence, or by using a subtraction technique. If integral frequency based fat suppression is applied this sequence may be run prior to the dynamic set with no fat suppression to help characterize lesions. Applying a 3D non-rigid registration may help to minimise the production of artefacts in the subtraction images if the patient moves during the eamination. o T1-weighted spoiled gradient echo with fat suppression Produces an ultra high resolution post contrast scan for assessing the lesion s morphology. Should give a 50% improvement in voel size compared with the dynamic scan (unless an in-plane resolution of 0.6mm is already achieved). Includes an integral frequency based fat suppression, such as frequency-selective fat saturation or water ecitation. 2.7 DICOM compliance and data import It is epected that all modern MRI machines and workstations will be DICOM compliant. Minimum standard: uncompressed DICOM 3.0 Data should be written to a CD (or equivalent) directly from the scanner or workstation and not via PACS, as some centres report that data have been rescaled by PACS. If using PACS, then a quality control (QC) process should be in place to ensure that the source data have not been modified. 6

2.8 Viewing tools The following tools should be available on the reporting workstation o Subtraction, multiplanar reconstruction (MPR), and minimum intensity projection (MIP) software o signal enhancement analysis package with tools for selecting regions of interest (ROIs) o hotspot analysis for diagnosis (maimum signal intensity, guidance on scaling, presence of air or fat) o software package to support image analysis image registration features the ability to change how images are displayed o pharmacokinetic modelling o piel by piel analysis of ROIs with wash-in and wash-out maps. 3. REPORTING MRI EXAMINATIONS 3.1 Reading requirements It is epected that the majority of women in the high risk group will undergo both -ray mammography and breast MRI. Some of these may have MRI surveillance only, for eample because of their young age, while a few may have mammography only because MRI is contraindicated. Mammograms must be acquired in full-field digital format. If possible mammography should be performed on the same day as the MRI. The minimum standard is that mammography should be performed within two weeks of the MRI eamination. The mammograms and the MRI images should be read together and must be independently double read. At least one of the MRI readers should be a breast radiologist who satisfies the NHSBSP standards as a breast screening film reader (among them, conducting a minimum of 5000 mammograms per year). 5 The MRI readers should each report a minimum of 100 breast MRI eaminations (including screening and symptomatic cases) per year. The results of a client s previous screening MR and mammography eaminations should be available for comparison. Standard MRI screening reporting categories should be used MRI 1: normal MRI 2: benign MRI 3: indeterminate MRI 4: suspicious MRI 5: malignant MRI screening and assessment forms have been developed for the national breast screening system (NBSS). These will be used to record the MRI opinion and also the 7

outcome of the assessment of MRI detected abnormalities. These forms are shown at Appendi 1. Protocols for their use by MRI units and breast screening centres are being developed by the three demonstration sites for MRI surveillance of women at high risk, and will be circulated separately. 3.2 Assessment of abnormalities Abnormalities detected by MRI and/or digital mammography should be assessed by the local NHSBSP screening team and preferably by the radiologist who has read the MRI eamination. This assessment need not take place in an NHSBSP screening assessment unit, provided the assessing team satisfies the NHSBSP requirements. Initial assessment should be with conventional imaging (further mammography and ultrasound). The reporting forms at Appendi 1 should be used to record data on the national breast screening system (NBSS). Indeterminate MR abnormalities (MRI 3: enhancing mass < 5mm or area of enhancement <10mm) detected by conventional imaging should trigger a repeat MR eamination within si months. Suspicious MR abnormalities (MRI 4 and 5: enhancing mass lesions > 5mm or areas of enhancement > 10mm or lesions with suspicious morphology) detected by conventional imaging should be considered for MRI guided biopsy. 3.3 MRI guided biopsies MRI guided biopsies should only be performed at a designated MRI biopsy screening centre. Minimum standards o designated screening MRI centres must perform a minimum of 12 MRI guided breast biopsies per year o all MRI guided biopsies should be performed using vacuum assisted mammotomy (VAM) in a centre that has recognised epertise in VAM and carries out at least 50 image guided VAM procedures per year. Audits should be undertaken to ensure that reporting accuracy and biopsy rates are at an acceptable standard. It is epected that the great majority of significant abnormalities detected by MRI will be identified at assessment and biopsied using conventional imaging, and that MRI guided biopsy will rarely be required. Standards for radiological eperience are set out at Appendi 2. Standards for reporting accuracy and biopsy rates will be developed in due course in the light of operational eperience. 3.4 Recall rate The number of women recalled for further imaging or biopsy should be kept to a minimum. Minimum standard <10% Epected standard <7% 8

4. QUALITY CONTROL AND QUALITY ASSURANCE 4.1 Quality control There must be regular quality control checks to ensure that MRI screening centres meet the NHSBSP minimum standards for equipment, image acquisition, and data storage. Annual checks The following checks are to be carried out by a physicist with the required MRI epertise. They should be performed at least annually and after all major software or hardware upgrades 1. Check that screening sequences meet the specified criteria 2. Evaluate the performance of the system and screening sequences a) spatial and slice resolution, using a test object b) contrast performance over a range of T1s c) signal to noise performance d) uniformity of ecitation e) homogeneity of scanner f) effectiveness of fat suppression sequences g) effectiveness of shimming h) accuracy of image data recording, on output and storage i) archiving process j) targeting and alignment of biopsy apparatus and process, where applicable k) a range of tests on the evaluation software to ensure that it reads and analyses image data accurately. Routine weekly quality control checks The following simple QC checks should be performed on site by a competent radiographer or physicist using a phantom in the breast coil provided for the purpose. These should monitor a) signal to noise b) contrast sensitivity at fied T1s c) fat suppression 4.2 Accreditation As part of the routine commissioning process, commissioners should ensure that all MRI screening centres meet the minimum standards for o equipment and measurement set up o performance of routine QC checks o radiological training and eperience. The relevant standards are summarised at Appendi 2. 9

4.3 MRI audit database In consultation with the NHS National Office, a database of all MRI measurements will be established for each MRI screening centre. National Office will support this new development, the training of MRI readers, and evaluation of the effectiveness of the surveillance process. The MRI physicist will be responsible for image data archiving and analysis. Standards for image acquisition, reading and assessment are summarised at Appendi 2. REFERENCES 1. Leach MO, The MARIBS Advisory Group, Staff and Collaborating Centres. The UK national study of magnetic resonance imaging as a method of screening for breast cancer (MARIBS). Journal of Eperimental and Clinical Cancer Research, 2002, 21 (Suppl. 3): 107-114. 2. Cancer Reform Strategy. Department of Health, 2007. Available at http://www.dh.gov.uk/en/publicationsandstatistics/publications/publicationspolic yandguidance/dh_081006. 3. Familial breast cancer: the classification and care of women at risk of familial breast cancer in primary, secondary and tertiary care. National Institute for Health and Clinical Ecellence (NICE), 2006 (Clinical Guideline CG41: partial update of CG14). Available at http://www.nice.org.uk/article.asp?a=30600. 4. Increased risk of breast cancer after radiotherapy for Hodgkin s Disease: patient notification eercise. Updated Toolkit. Department of Health, Letter and attachment, dated 31 October 2003. 5. Gadolinium-based contrast media and Nephrogenic Systemic Fibrosis. Royal College of Radiologists, 2007. Available at http://www.rcr.ac.uk/docs/radiology/pdf/bfcr0714_gadolinium_nsf_guidance Nov07.pdf 6. Quality Assurance Guidelines for Breast Cancer Screening Radiology. NHS Cancer Screening Programmes, 2005 (NHSBSP Publication No 59) 7. Consolidated Guidance on Standards for the NHS Breast Screening Programme (Version 2). NHS Cancer Screening Programmes, 2005 (NHSBSP Publication No 60) 10

APPENDIX 1: MRI SCREENING FORMS AND ASSESSMENT FORM FORM A: BSP MRI screening request (To be sent to the MRI service undertaking the procedure) For MRI use Date received: Reference: For BSP use Location: Hospital code: R (Affi return address label here) Client details Surname: Forenames: Date of birth: Hospital number: SX number: NHS number: Address: Telephone No 1: Telephone No. 2: Ethnic origin code: Year mastectomy performed (right): (left): GP name: GP address: MRI appointment Appointment date: Time: Appointment note: Status: Booked Attended Attended not screened Not attended MRI contraindicated Status comment: (Use this space to enter free tet about the appointment) 11

FORM B: BSP MRI screening request (to be completed by MRI radiographer) Client details (may be completed in advance by referring screening office) Surname: Forenames: Date of birth: Hospital number: SX number: NHS number: MRI screening results (MRI radiographer may note here any breast problems reported by the client or observed. No clinical eamination should be undertaken) Date of MRI eamination: MRI scanner (make and model) Clinical findings Right Left Recent lump Distortion/change in breast shape Nipple discharge Nipple eczema Recent nipple inversion Skin tethering or dimpling Other relevant breast information (eg surgery details) Right: Left: Reporting (choose Clinical if the MRI was normal but client needs to be clinically recalled; choose N/A if that breast was absent) Date reported: Right Clinician name/ code Normal Abnormal Clinical N/A Normal Abnormal Clinical N/A 1 2 Overall Left 12

FORM C: BSP MRI screening request (to be used if results were abnormal and further eamination of the MRI is needed) Client details Surname: Forenames: Date of birth: Hospital number: SX Number: NHS number: MRI assessment results (complete a separate form for each lesion, as multiple lesions will be recorded by NBSS) Date performed: Location name or code: Clinician name or code:. Side Assessed: Right Left Lesions and abnormalities: Lesion or abnormality Background enhancement: Widespread Moderate Low Homogenous Heterogenous Rim Central Segmental enhancement: Segmental Non segmental Enhancement curve: Type I Type II Type III Mass: Benign Intermediate Suspicious None Morphology: Smooth Well defined Lobulated Spiculate Size (mm): Ailla Normal: Normal Abnormal MRI Lesion notes: MRI Lesion description: Type of site: Single Multiple Opinion: G1 Normal G2 Benign G3 Uncertain G4 Suspicious G5 Malignant (NB Opinion codes are consistent with NBSS. G is used because NBSS already uses M for mammography.) 13

APPENDIX 2: QUALITY ASSURANCE STANDARDS FOR MRI BREAST SURVEILLANCE Objective Criteria Minimum standard Equipment To optimise diagnostic image quality Field strength Minimum standard: 1 Tesla Epected standard: 1.5 Tesla To optimise diagnostic image quality To optimise diagnostic image quality To ensure accurate data storage Number of elements Image acquisition MRI sequences Dynamic sequence Total eamination time Fat suppression Breast movement Eamine both breasts with isotropic voels Uniform ecitation and received signal homogeneity across image Slice thickness In-plane resolution Timing of eaminations Data storage All machines and workstations to be DICOM compliant Minimum standard: 4 channel Epected standard: 7 channel Minimum standard: 60 sec Epected standard: 45 sec < 30 mins Minimum standard: subtraction Epected standard: integrated fat suppression Support of breasts to minimise motion 2 mm < 1.0mm Day 6-16 of the menstrual cycle Uncompressed DICOM 3.0 14

Objective Criteria Minimum standard Radiological eperience To ensure that MRI readers are suitably eperienced To ensure that screening MRI centres have sufficient eperience To ensure comparison with digital screening mammograms To ensure that standard MRI reporting categories are used To minimise aniety for women who are awaiting the results of their screening imaging 7 Reader 1 No of mammograms No of MRI eaminations Reader 2 No of mammograms per year No of MRI eaminations per year Number of MRI guided breast biopsies per centre Number of image guided VAM procedures per centre Reading and assessment Availability of digital screening mammograms Reporting categories used Percentage of women who are sent their result within two weeks 5000 per year (screening and/ or symptomatic) 100 per year (screening and/ or symptomatic) Only reader 1 is required to meet above standard 100 per year(screening and/or symptomatic) At least 12 per year At least 50 per year Always available Standard MRI reporting categories must used Minimum standard: 90% Target: 100% To minimise the number of women who are recalled for further tests To minimise the interval from the screening imaging to assessment 7 Number of women recalled for further imaging or biopsy Percentage of women who attend an assessment centre within three weeks of attendance for their screening imaging. Minimum standard < 10% Epected standard <7% Minimum standard: 90% Target: 100% 15