Corneal Thickness Measurement with Optical and Ultrasonic Pachymetry



Similar documents
Comparison to other anterior segment imaging devices

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

Comparison of central corneal thickness measurements using Pentacam and ultrasonic pachymetry in post-lasik eyes for myopia

Page: 1 of 6. Corneal Topography/Computer-Assisted Corneal Topography/ Photokeratoscopy

ALL-IN-ONE Optical Biometry, Dual Scheimpflug Tomography and Placido Topography

Total Corneal Power Estimation: Ray Tracing Method versus Gaussian Optics Formula PATIENTS AND METHODS

Comparison of Residual Stromal Bed Thickness and Flap Thickness at LASIK and Post-LASIK Enhancement in Femtosecond Laser-Created Flaps

PARALLEL TO THE DEVELOPMENTS OF SURGICAL

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

Ectasia Risk Factors: Lessons Learned

Cataract Testing. What a Patient undergoes prior to surgery

Cornea and Refractive Surgery Update

OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction

Common Co-management Questions

EFFECT OF MYOPIC LASIK ON RETINAL NERVE FIBER LAYER THICKNESS- IS IT SAFE OR UNSAFE?

How do we use the Galilei for cataract and refractive surgery?

How do you adjust for IOP in a LASIK patient?

Insert to. January Comfort and confidence for all IOL calculations

Complications of Combined Topography-Guided Photorefractive Keratectomy and Corneal Collagen Crosslinking in Keratoconus

Validation of a New Scoring System for the Detection of Early Forme of Keratoconus

OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment

Case Reports Post-LASIK ectasia treated with intrastromal corneal ring segments and corneal crosslinking

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

Keratoconus Detection Using Corneal Topography

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

The Cornea, The Lens & Refractive Surgery. John S. Berestka, M.D. Timothy J. Ehlen, M.D. Matthew R. Ramsey, M.D. Stanley D. Walker, M.D.

Technicians & Nurses Program

Simple regression formula for intraocular lens power adjustment in eyes requiring cataract surgery after excimer laser photoablation

~Contributing to Eye Health~

Trends in Ophthalmology a Market Perspective. Matthias Karl, Michael Kempe

CONSENT FORM. Procedure: Descemet s Stripping Automated Endothelial Keratoplasty (DSAEK)

OCULUS PENTACAM BASIC CLASSIC HR

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

IOL Power Calculations for Postrefractive Surgery Eyes

Post LASIK Ectasia. Examination: Gina M. Rogers, MD and Kenneth M. Goins, MD

INTRODUCTION. Trans Am Ophthalmol Soc 2006;104:

IOLMaster 500. Excellence in Biometry. Precision. Simplicity. Outcomes.

Corneal topography recent advances

Refractive surgeons routinely face the challenge of

Premium Lenses in Problematic Patients

REFRACTIVE SURGERY NIGHTMARES Dr.ATHIYA AGARWAL

Update on Post- Refractive Surgery IOL Calculations

Program Requirements for Fellowship Education in Cornea, External Diseases & Refractive Surgery*

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

Diego Fernando Suárez Sierra, MD Fellow Cornea and Refractive Surgery Fellow Lens and Ocular Surface Vejarano Laser Vision Center

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery

Excimer Laser Refractive Surgery

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Corporate Medical Policy Implantation of Intrastromal Corneal Ring Segments

IOLMaster with Software Version V 3.0

Short and long term complications of combined. Protocol) in 412 keratoconus eyes (2 7 years follow up)

Our Commitment To You

Pentacam. The. The gold standard in anterior segment tomography. Supplement to February 2011

Kensington Eye Center 4701 Randolph Road, #G-2 Rockville, MD (301)

Alexandria s Guide to LASIK

COMPARISON OF INTRAOCULAR LENS POWER CALCULATION METHODS IN EYES THAT HAVE UNDERGONE LASER-ASSISTED IN-SITU KERATOMILEUSIS

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life

THE GUIDE TO REFRACTIVE LENS EXCHANGE SEE CLEARLY.

Medicare and Corneal Surgery: Cosmetic versus Functional

Collagen Cross-linking combined with PRK and LASIK

How To Implant A Keraring

Surgical Advances in Keratoconus. Keratoconus. Innovations in Ophthalmology. New Surgical Advances. Diagnosis of Keratoconus. Scheimpflug imaging

Case Report Laser Vision Correction on Patients with Sick Optic Nerve: A Case Report

MAZAHERI LASIK METHOD FOR VISUAL ENHANCEMENT TECHNICAL FIELD OF THE INVENTION. [0001] The present invention is directed, in general, to

Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda

Overview of Refractive Surgery

OCULUS Pentacam Pentacam HR

Laser Vision Correction: A Tutorial for Medical Students

Richard S. Hoffman, MD. Clinical Associate Professor of Ophthalmology Oregon Health & Science University

Maximizing Your Cataract Surgery Outcomes in Corneal Disease

INTRODUCTION MATERIAL AND METHODS

TABLE OF CONTENTS: LASER EYE SURGERY CONSENT FORM

CustomVue Treatments for Monovision in Presbyopic Patients with Low to Moderate Myopia and Myopic Astigmatism


Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

Consumer s Guide to LASIK

PATIENT INFORMATION BOOKLET

Long-Term Outcomes of Flap Amputation After LASIK

The LenSx Laser System. Discover the assurance of bladeless cataract surgery

How To Know If You Can See Without Glasses Or Contact Lense After Lasik

KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,

5/24/2013 ESOIRS Moderator: Alaa Ghaith, MD. Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD

Minimally Invasive Surgery: Femtosecond Lasers and Other Innovative Microsurgical Techniques

Alain Saad, MD, Alice Grise-Dulac, MD, Damien Gatinel, MD, PhD

Ophthalmic Equipment Guide

Femtosecond Lasers in LASIK Surgery

Corneal intrastromal implantation surgery for the treatment of moderate and high myopia

Financial Disclosure. LASIK Flap Parameters IntraLase Microkeratome 6/9/2008. Femtosecond LASIK Flaps: What Could We Customize Yesterday?

Educational Goals & Objectives for Ophthalmology Residents at Parkland Memorial Hospital (PMH)

Transcription:

Corneal Thickness Measurement with Optical and Ultrasonic Pachymetry Hassanain M. Amed Ophthalmology Unit, Surgery Dept., College of Medical, Kufa University. M J B Abstract Background: Evaluation of central and peripheral corneal thickness is of great clinical importance in the ophthalmology. The main aim of this study is to evaluate the optical and ultrasound pachymetric measurements by comparing the central corneal thickness measurements obtained by the (SP 3000 pachymeter from TOMEY) and Allegro Oculyzer (pentacam ) pachymeter. Patients and methods: 28 patients attending the photorefractive surgery unit at IBSAR LASIK center in ALnajaf were collected. The patients were grouped into three major groups according to the refractive errors they got and the central corneal thickness for each patient was measured by two methods:1. using the Allegro Oculyzer (pentacam) as a non contact optical pachymeter.2. using the ultrasonic pachymeter SP 3000 pachymeter from TOMEY, all measurments were done by the same surgeon. Results: the findings of this study reported that the ultrasound pachymeter significantly overestimates the corneal thickness by about 2.85 micron and the eyes with refractive errors ranging from 0 to -6D in our study have higher optical central corneal thickness values than ultrasonic values but the differences were statistically non significant. Conclusion: Ultrasound pachymetry overestimates the corneal thickness measurements compared to optical pachymetry and they should not be used interchangeably during sequential checking of this ocular parameter الخالصة قياس سمك القرنيو الوسطي والطرفي لو اىميو سريريو, تيدف ىذه الد ارسو الى مقارنو قياس سمك القرنيو الوسطي بطرقتين مختمفتين ىما الطريقو البصريو واالمواج فوق الصوتيو.اجريت الد ارسو عمى 28 شخصا في مركز ابصار لمعيون في محافظة النجف.اظيرت الد ارسو اني طريقو االمواج فوق الصوتيو تعطي ق ارئو اعمى من الطريقو الصوتيو بصوره ممحوظو بحوالي 2,85 مايكروميتر. Introduction E xamination of the cornea and anterior segment of the eye, especially evaluation of central and peripheral corneal thickness, is important in various fields of ophthalmology.[1] The corneal thickness is an important indicator of the corneal function and is routinely determined in several clinical settings. Its evaluation reflects the endothelial function, offers an insight to the cornea`s adaptation in contact lens wear, is of postoperative prognostic value for penetrating keratoplasty, can reveal local or systemic metabolic disorders, and is important when diagnosing and treating patients with glaucoma [2]. Pachymetry, measurements of the corneal thickness, has gained increasing importance in ophthalmic practice because of the growing interest in keratorefractive surgery and extended-wear contact lens. The design and outcome of refractive surgical procedures rely on the accuracy of pachymetry measurements.[1,3] Hassanain M. Amed 772

Corneal thickness measured by pachymetry is a sensitive indicator of corneal health. One of the most common approaches to corneal pachymetry is ultrasound technology. Slit lamp mounted optical pachymetry is the other method for obtaining measurements of corneal thickness. Specular microscopic pachymetry,especially the newly introduced automated non contact specular pachymetry, is also a widely used and accepted method to measure this important parameter that reflects corneal health and is considered essential for diagnosing multiple corneal diseases or planning treatment for some problems.[3,4] The principle behind obtaining this measurement,whether in optical or ultrasound pachymetry, depends on the reflection of light or ultrasound from the anterior and posterior surface of the cornea.[3] Corneal thickness has been evaluated by various methods, and a wide array of literature on the topic is available.[5] Ultrasound pachymetry has been the gold standard for estimating the corneal thickness for the past few decades. Now modern pachymetric techniques are commercially available, ultrasound pachymetry is the reference against which the rest are currently tested[.3,6] Different studies report different results while comparing the two methods for assessing the corneal thickness and most of the results showed that optical pachymetry tends to underestimate corneal thickness compared to ultrasonic pachymetry because the former only measures between Descemet s membrane and Bowman s layer.[7] In other studies a slight tendency for ultrasound methods to underestimate corneal thickness was noted especially if it was carried out prior to optical pachymetry, because of the changes in tissue hydration secondary to the ultrasound speed through the cornea.[4] The aim of this study is to evaluate the optical and ultrasound pachymetric measurements by comparing the central corneal thickness measurements obtained by the (SP 3000 pachymeter from TOMEY) and Allegro Oculyzer (pentacam) pachymeter. Patients and Methods The study was conducted on 56 undiseased eyes of 28 patients attending the photorefractive surgery unit at IBSAR LASIK center in ALnajaf. All patients were subjected to thorough ocular examination as a part of their preparation for photorefractive surgery. The spherical equivalents for their refractive errors range from (+6)D to (-15.25D). The patients were grouped into three groups according to the refractive errors they got; 1. The hypermetropes, those having spherical equivalent of refractive error ( >0). 2. The myopes with spherical equivalent range between (0 to -6D) 3. The myopes with spherical equivalent of refractive error more than (-6D) The central corneal thickness for each patient was measured by two methods: 1. Using the Allegro Oculyzer (pentacam ) as a non contact optical pachymeter 2. Using the ultrasonic pachymeter(sp 3000 pachymeter from TOMEY) All measurments were done by the same surgeon. The readings of both methods and the differences in the two readings for each eye were tabulated.mean and standard deviation for the readings in each group were calculated and the Hassanain M. Amed 773

statistical significance was calculated using the T test with a P value <0.5 considered as significant. Results The patients age range was 17-54 year with a mean of 32.78±8.77 year. Male to female ratio was 3:4. Ultrasound and optical pachymetric measurements means and standard deviations for whole studied patients are shown in table 1. Student T test applied on the paired observations of the studied groups revealed the P values shown in table 1. Table 1 Mean and SD of the Pachymetric readings for the studied patients all patients Hypermetrops Myopes <6D Myopes>6D No. of eyes 56 8 31 17 Mean CCT (US readings) in micron Mean CCT(optical readings)in micron 520.98±31.72 561±32 505.35±24.32 530.64±29.62 518.39±30 552.37±31.12 505.54±23.95 525.82±31.22 Difference of the 2.85±8.29 8.62±10.71-0.19±8.44 4.82±5.67 means(us-o) in microns P value <0.01* <0.01* >0.5 <0.001* (*) statistically significant Discussion Accurate assessment of corneal thickness is important in many clinical situations such as the diagnosis of corneal ecstatic conditions and corneal physiology, contact lens research, or refractive surgery procedures.[4] Ophthalmic biometry should provide rapid, objective, and accurate measurements of ocular dimensions. Optical coherence tomography is a high resolution technique that can create precise cross-sectional images of the anterior and posterior eye segments.[3] The physical basis of imaging depends on differences in optical infrared reflectivity between different tissue microstructures within the eye. Thus it is the optical analog of ultrasound B scan, although it provides images with higher resolution in both axial and lateral dimensions and does not require direct contact with the eye for qualitative and quantitative assessments. Optical pachymeters depend on this basis.[3] At present, ultrasound is the most frequently used pachymetry technique and is considered the gold standard because of its high degree of interobserver, intraobserver, and interinstrument reproducibility.[1,8] However, the required probe-cornea contact is associated with patient discomfort, the need for topical anesthesia, the risk for epithelial lesions and transmission of infections leading to the preference of the non contact methods such as the optical coherence tomography, partial coherence tomography, low coherence reflectometry, specular microscopy, scanning slit pachymetry and corneal confocal microscopy.[1] Hassanain M. Amed 774

Most of the published studies report differences between the readings of the central corneal thickness obtained by the optical and the ultrasound pachymeters. Ultrasound based systems of pachymeters tends to overestimate the readings with variable significance. This could be attributed to large probe-tip size, probable probe tilting, the averaging US mode, epithelial edema from local anesthesia, and most importantly the calibration or sound velocity factors because the sound velocity is likely to vary between different layers of the cornea. On the other hand compression of the tear film during contact US pachymetry is expected to give lower values than non contact other methods, thus complicate the relationship. The current study demonstrates, after comparing the paired data, that ultrasound pachymeter significantly overestimates the corneal thickness by about 2.85 micron. But upon reviewing the data of the different groups of refractive errors, the moderately myopic eyes didn`t show the same result. The US measurements in eyes with refractive error ranging between 0 and -6D were below the optical ones, however the difference was statistically non significant. Table 1 Prospero et al reported, in a series of one hundred sixty-three eyes, that ultrasound pachymetry measurements were consistently higher OCT measurements (mean difference 7.5 +/- 1.4 micron) ( P<.0005) in normal eyes and post-lasik eyes (P<.0005).[9] Amano et al showed that the mean corneal thicknesses were comparable between ultrasonic pachymetry optical pachymetry. The measurements had significant linear correlations with one another and both methods had highly satisfactory measurement repeatability. [10] The mean difference of readings measured by US pachymetry with those measured by optical pachymetry was 2.3 µm in Karimian`s study.[11] Christopher K. demonstrated thinner corneas when measured optically (visnate OCT) than ultrasonically but he found comparable results when he used another model of optical pachymeter (the SLOCT- OCT).Suggesting that different devices with different algorithm for image analysis display different results.[12] There is an evident agreement of our statistically significant levels of differences between the paired data with most of the published studies that emphasized on the US readings to be higher than the optical readings, however none of the reviewed published articles mentioned that the level of refractive errors or corneal curvature may play a role as a factor affecting the dynamics of corneal thickness measurements. Eyes with refractive errors ranging from 0 to -6D in our study have higher optical central corneal thickness values than ultrasonic values but the differences were statistically non significant. This finding needs to be further evaluated thoroughly in a larger population study. Furthermore, other factors, most importantly the age of the patients, should also be studies as a probable factor affecting the interchangeability of central corneal thickness measurements by different devices. In conclusion, ultrasound pachymetry overestimates the corneal thickness measurements compared to optical pachymetry and they should not be used interchangeably during sequential checking of this ocular parameter. Careful attention is needed when interpreting data obtained from different imaging systems, different models and even different versions of analysis software. Better Hassanain M. Amed 775

standardization is needed for new devices and instruments. References 1.Ramin k,tanja M,Gerd U.Central and peripheral pachymetry measurements accordind to age using the pentacam rotating Scheimpflug camera.j Cataract Refract Surg 2007;33:830-836 2.Christopher W, Christian S, Hans H, et al.noncontact corneal pachymetry with slit lamp adapted optical coherence tomography.am J Ophthalmol 2002;133:444-450. 3.Laszlo M, Achim L, Berthold S. Corneal thickness measurements with contact and noncontact specular microscopic and ultrasound pachymetry. Am J Ophthalmol 2001:132;517-521. 4. Jose M, Alejandro C, Eva Y, et al.central and peripheral corneal thichness measurement with Orbscan II and topographical ultrasound pachymetry. J Cataract Refract surg 2003 ; 29:125-132. 5.Parafita M, Yebra E, Giraldez M, et al.optical methods for corneal thickness measurement;a review. In:Pandalai SG,ed,Recent ResearchDevelopment in Optics,vol 2.India, Trivandrum Research Signpost, 2002;35-51. 6.Yaylali V, Kaufman S, Thompson H. Corneal thickness measurements with the Orbscan topography system and ultrasonic pachymetry. J Cataract Refract Surg 1997 ; 23: 1345-1350. 7.Argus W. Ocular hypertention and central corneal thickness. Ophthalmology 1995;102:1810-1812. 8.Bohnke M, Chavanne P, Gianotti R. continous non contact corneal pachymetry with a high speed reflectometer. J Cataract Refract Surg 1998;14;140-146. 9. Prospero P, Rocha K, Smith S, et al.central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and postlaser in situ keratomileusis eyes.j Cataract Refract Surg. 2009 Jun;35(6):1055-62. 10. Amano S, Honda N, Amano Y, et al. Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.ophthalmology. 2006 Jun;113 (6):937-41. 11. Karimian F, Feizi S, Faramarzi A, et al. Evaluation of corneal pachymetry measurements by Galilei dual Scheimpflug camera. Eur J Ophthalmol. 2011 Sep 21:65. 12.Christopher K. Corneal Thickness, an author reply. Ophthalmology 2008; 115(No.12 ): 2316. Hassanain M. Amed 776