BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Similar documents
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Update on ACR Digital Mammography QC Manual

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Contract: FIGM-CT Measures for Optimising Radiological Information and Dose in Digital Imaging and Interventional Radiology (DIMOND III)

Digital Breast Tomosynthesis QC Requirements

X-ray absorptiometry of the breast using mammographic exposure factors: application to units featuring automatic beam quality selection

Quality Control of Full Field Digital Mammography Units

Needle crystal detector technology in mammography further dose reduction and clinical image quality with

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Patient Exposure Doses During Diagnostic Radiography

Mammography Applications for MAMMOMAT 1000/3000 Nova

Digital Mammography Update: Design and Characteristics of Current Systems

Breast Biphasic Compression versus Standard Monophasic Compression in X-ray Mammography 1

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Concepts for High-Resolution Low-Dose CT of the Breast

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

American College of Radiology Mammography Accreditation Program

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Development of High-sensitivity and High-resolution Digital Mammography System AMULET

Sustaining a High-Quality Breast MRI Practice

Hologic Selenia Dimensions C-View Software Module. October 24, 2012

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Digital Mammogram National Database

Assessing Radiation Dose: How to Do It Right

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

MAMMOGRAPHY GOALS AND OBJECTIVES

CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME

Infrared Thermography Not a Useful Breast Cancer Screening Tool

Breast Sonography general goal. Optimizing Breast Sonography. BUS indications -- all. Breast Sonography specific goals.

Certification protocol for breast screening and breast diagnostic services

Worldwide Quality Assurance networks for radiotherapy dosimetry

for breast cancer detection

How To Compare Mammography To A 3D System

Medicare Part B. Mammograms - Updated Billing Guide for Screening and Diagnostic Tests

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Personalized Breast Screening Service

DIAGNOSTIC REFERENCE LEVELS IN MEDICAL IMAGING: REVIEW AND ADDITIONAL ADVICE

Multi-slice Helical CT Scanning of the Chest

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Performance testing for Precision 500D Classical R/F System

College of Radiology, Academy of Medicine of Malaysia Recommendations & Guidelines for Quality Assurance in MAMMOGRAPHY

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Breast Imaging Made Brief and Simple. Jane Clayton MD Associate Professor Department of Radiology LSUHSC New Orleans, LA

Atomic and Nuclear Physics Laboratory (Physics 4780)

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

Q: What differentiates a diagnostic from a screening mammography procedure?

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

MIN-R 2000 Plus Film

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program

Current Status and Problems of Breast Cancer Screening

Diagnostic Exposure Tracking in the Medical Record

Clinical Commissioning Policy: Proton Beam Radiotherapy (High Energy) for Paediatric Cancer Treatment

Low-dose CT Imaging. Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

DICOM metadata-mining in PACS for computed radiography X-Ray exposure analysis: a mammography multisite study

Mammography. What is Mammography?

Recommendations for a technical quality control program for diagnostic X-ray equipment

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS

GE Healthcare Clinical Education. Mammography. Course catalogue

GE Healthcare. Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together

Update on MQSA and ACR Mammography Accreditation

Credibility and Pooling Applications to Group Life and Group Disability Insurance

Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

SOURCES AND EFFECTS OF IONIZING RADIATION

Supports screening, diagnostic, and multimodality workflows

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Physics testing of image detectors

Breast Cancer. Presentation by Dr Mafunga

Failure Modes and Effects Analysis (FMEA)

D. FREQUENTLY ASKED QUESTIONS

Screening Mammography for Breast Cancer: American College of Preventive Medicine Practice Policy Statement

ACR BI-RADS ATLAS MAMMOGRAPHY MAMMOGRAPHY II. REPORTING SYSTEM. American College of Radiology 121

How To Decide If You Should Get A Mammogram

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

The Exposure Index and Its Standardization

VI. FREQUENTLY ASKED QUESTIONS CONCERNING BREAST IMAGING AUDITS

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Goals and Objectives: Breast Cancer Service Department of Radiation Oncology

Transcription:

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department of Radiology, KU Leuven, Leuven, Belgium c Telecom ParisTech, CNRS LTCI, Paris, France Email address of main author: nausikaa.geeraert@student.kuleuven.be Abstract Average Glandular Dose (AGD) is the quantity generally accepted for dose monitoring of X ray mammographic examinations. A survey of breast characteristics (compressed breast thickness, peak breast density) and AGD values was carried out on a database describing the acquisition conditions of 147,487 routine mammographic images acquired in Europe, North America and Asia-Pacific on GE Healthcare Senographe DS and Essential mammographic equipment. These systems compute the breast density in the densest part of the image and the related AGD using normalized glandular dose values interpolated from published tables. The correlation between peak breast density and compressed breast thickness was determined for the different geographical zones, as well as the distributions of the AGD. Breasts in Asia-Pacific were on average significantly thinner and denser than in Europe and North America. Breasts in North America were on average significantly thicker and denser than in Europe. The average AGD per image was 1.42mGy in North America and Asia-Pacific, and 1.48mGy in Europe. We computed the difference in AGD values between an individual density correction factor based on the peak breast density and a population-averaged density correction factor, as routinely used for dosimetry surveys in Europe. The resulting collective AGD values derived from the two distributions were similar (difference <5%), while the individual AGD values differed by up to 30%. Breast dosimetry can be improved by replacing the population-averaged breast density by the individual volumetric breast density. 1. INTRODUCTION In many countries, large-scale breast X ray imaging is used for early detection of breast cancers and for diagnostic purposes. The limited exposure to X rays must allow a good image quality and produce mammographic images with high sensitivity and specificity for the detection and characterization of suspicious lesions. It is generally accepted that risk or detriment from mammographic examinations can be limited to the effects of the exposure to the breast. Based on the assumption that only the glandular cells are radiosensitive, the Average Glandular Dose (AGD) has been defined to quantify the risk from breast irradiation and has been widely accepted for regulations (e.g. MSQA [1]) and guidelines (e.g. European Guidelines [2]). Since the introduction of digital mammography systems it is easier to acquire and collect automatically large amounts of data worldwide. In this paper we study for different geographical regions the distribution of the peak breast density as a function of the compressed breast thickness, as well as the distribution of the doses. 2. MATERIAL AND METHODS 2.1. Methods to compute AGD The two most often used methods to calculate the AGD are based on Monte Carlo simulations [3, 4]. The breast is modeled by a semicircular phantom with a homogeneous content and surrounded by a layer of skin-representing tissue. The attenuation coefficient of the content corresponds to the breast density, the ratio of the amount of gland to total breast. The methods compute the AGD as the product of the entrance skin dose and a conversion factor from exposure to AGD, called normalized

GEERAERT, KLAUSZ, MULLER, BLOCH AND BOSMANS dose. Wu et al. [3, 5] compute values of this normalized dose (D gn ) as a function of relevant parameters (spectrum, breast thickness and breast density). Dance et al. [4, 6] factorize the normalized dose in three factors, a conversion factor for a 50% density breast with a Mo/Mo anode/filter combination, one for the spectral differences related to the anode/filter combinations, and one for the breast density. In their model the density correction factor can be calculated for the nominal breast density or it can be retrieved from population-averaged values of breast density as a function of compressed breast thickness and age. The latter approach is based on two UK film-screen mammography surveys [7, 8] (1545 women) where the density was computed from a manually positioned AEC sensor. 2.2. Collected data We obtained the technical data from acquisitions on Senographe DS (FOV:19 23cm) and Essential (FOV: 24 31cm and 19 23cm) (GE Healthcare, Chalfont, UK). These GE mammographic systems have an automatic exposure control (automatic optimization of parameters, AOP). From a pre-exposure image the densest area of the breast is found and an attenuation-equivalent thickness is computed from a calibrated model. The density in this area (peak breast density) is computed as a function of equivalent and compressed breast thickness [9] and recorded in the DICOM-header. The system computes the AGD applying the method of Wu et al. [5] with the effective values of D gn interpolated from the published tables using the peak breast density. The compressed breast thickness, the peak breast density and the AGD were collected from different geographical zones: Europe, North America and Asia-Pacific. Neither individual patient data, such as ID and age, nor the images were accessible. Only data acquired in contact mode with a fullformat rigid paddle in CC (cranio-caudal) and MLO (medio-lateral oblique) angulations were retained. We computed the distributions of the breast densities as a function of the compressed breast thickness for the geographical zones and superimposed the population-averaged distributions synthesized earlier by Dance [6]. For each individual exposure we computed the difference between the AGD using the individual breast density and using the population-averaged density from Dance s method [6]. Comparison of all characteristics was done with tailed t-tests. 3. RESULTS AND DISCUSSION 3.1 Population statistics The number of cases retained and their distribution over different exposure modes are represented in Table 1. The different possible modes were Manual and three AOP modes (Dose, STD, CNT). TABLE I. OVERVIEW OF THE DISTRIBUTION OF THE DATA Geographical zone Total number of cases Number of systems Manual AOP Dose AOP STD AOP CNT Europe 49847 105 1.10% 10.50% 85.30% 2.10% North America 43514 121 4.30% 7.70% 83.70% 4.10% Asia-Pacific 54136 31 0.80% 1.30% 63.60% 34.30% 3.2 Breast characteristics The histogram of compressed breast thickness as a function of peak density is shown in Figure 1 for the different geographical zones together with the distributions used by Dance [6] (dotted line: 40-49y, full line: 50-64y). Negative peak densities were due to errors on the compressed breast thickness, to which the density computation, used by the systems, is very sensitive [9]. 2

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X-RAY MAMMOGRAPHY A LARGE SAMPLE WORLDWIDE SURVEY The Asia-Pacific population has significantly denser, thinner breasts (p<0.0000001) than the European and North-American populations. The North-American population has significantly denser (p=0.0005) and thicker (p<0.0000001) breasts than the European population. FIG. 1. Compressed breast thickness as a function of breast density on GE systems (lines: Data from Dance full line: 50-64y, dotted line 40-49y) The statistics on the differences in AGD are presented in Table 2. The average AGD over the population, computed with the peak breast density was slightly higher than when using the 40-49 year population-averaged correction factor (up to 4.4%) and slightly lower than when using the 50-64 year population-averaged correction factor (up to 4.8%). The difference in AGD was less than 10% for 68% of the exposures and less than respectively 23% and 30% for 99.7% of exposures. TABLE II. THE DIFFERENCES BETWEEN THE AGD COMPUTATIONS USING DANCE WITH THE PEAK BREAST DENSITY AND THE POPULATION-AVERAGED DENSITY Geographical 40-49 year 50-64 year zone Average St dev +3 St dev Average St dev +3 St dev Europe 4.4% 11.0% 31.9% -2.1% 10.2% 23.6% North America 3.3% 11.4% 30.8% -3.0% 10.6% 23.6% Asia-Pacific 1.1% 10.1% 26.6% -4.8% 9.7% 17.2% 3.3 Dosimetric data The distribution of the dose per image per zone computed from the system is presented in Figure 2 and statistics are presented in Table 3. The secondary peak around 2mGy on the dose distribution of Asia-Pacific can be explained by the higher proportion of exposures taken using the AOP CNT mode. 3

GEERAERT, KLAUSZ, MULLER, BLOCH AND BOSMANS FIG. 2. AGD distributions for the different zones TABLE III. STATISTICS FOR THE AGD DISTRIBUTIONS COMPUTED FROM THE GE SENOGRAPHE DS AND ESSENTIAL (in mgy) Geographical zone Average Median <25% <75% <95% Europe 1.48 1.4 1.2 1.6 2.4 North America 1.42 1.4 1.2 1.6 2.1 Asia-Pacific 1.42 1.3 1.1 1.7 2.3 5. CONCLUSION The average AGD per image (CC and MLO) was between 1.4 and 1.5mGy for all geographical zones, almost 12% lower than the results from Hendrick et al. [10] for the GE Senographe 2000D. The average difference between the doses computed with the two density correction factors from Dance [6] was found negligible (<5%), making the population-averaged correction factor fully applicable for collective dose evaluation. However individual doses were up to 30% higher than the populationbased predicted values. In both cases, breast density is estimated in a small area (AEC sensor), and further improvement towards patient specific dosimetry should come from a true estimation of volumetric breast density instead. REFERENCES [1] US Code of Federal Regulations 21CFR900.12 (MQSA). [2] EUROPEAN COMMISSION, European guidelines for quality assurance in breast cancer screening and diagnosis, Fourth edition, ISBN 92-79-01258-4, Office for Official Publications of the European Communities, Luxembourg (2006). [3] WU, X., BARNES, G.T., TUCKER, D.M., Spectral dependence of glandular tissue dose in screen-film mammography, Radiology 179 (1991) 143-148. [4] DANCE, D.R., Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose, Phys. Med. Biolog. 35 (1990) 1211-1219. [5] WU, X., GINGOLD, E.L., BARNES, G.T., TUCKER, D.M., Normalized average glandular dose in molybdenum target-rhodium filter and rhodium target-rhodium filter mammography, Radiology 193 (1994) 83-89. [6] DANCE, D.R., SKINNER, C.L., YOUNG, K.C., et al., Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol, Phys. Med. Biolog. 45 (2000) 3225-3240. [7] YOUNG, K.C., RAMSDALE, M.L., BIGNELL, F., Review of dosimetric methods for mammography in the UK screening programme, Rad. Prot. Dosim. 80 (1998) 186-187. [8] BECKETT, J.R., KOTRE, C.J., Dosimetric implications of age related glandular changes in screening mammography, Phys. Med. Biolog. 45 (2000) 801-813. 4

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X-RAY MAMMOGRAPHY A LARGE SAMPLE WORLDWIDE SURVEY [9] DESPONDS, L., KLAUSZ, R., Automatic estimation of breast composition with mammographic X-ray systems, in 80 th Scientific Assembly and Annual Meeting, Radiological Society of North America, Nov. 27 -Dec. 2, 1994, Radiology, 193 P Suppl. (Nov. 1994). [10] HENDRICK, R.E., PISANO, E.D., AVERBUKH, A., et al., Comparison of acquisition parameters and breast dose in digital mammography and screen-film mammography in the American College of Radiology Imaging Network Digital Mammographic Imaging Screening Trial, Am. J. Roentgen. 194 (2010) 362-369. 5