Quality Assurance/Quality Control in Acid Deposition Monitoring



Similar documents
6. QUALITY ASSURANCE AND QUALITY CONTROL

1 Quality Assurance and Quality Control Project Plan

Validation and Calibration. Definitions and Terminology

ASSURING THE QUALITY OF TEST RESULTS

APPENDIX N. Data Validation Using Data Descriptors

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

9. Quality Assurance 9.1 INTRODUCTION 9.2 FIELD SAMPLING QUALITY ASSURANCE 9.3 ANALYTICAL QUALITY ASSURANCE Internal Quality Control

APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN

DRY CLEANING PROGRAM QUALITY ASSURANCE MANAGEMENT PLAN

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

CHAPTER ONE TABLE OF CONTENTS 1.0 INTRODUCTION... 1

Control Charts and Trend Analysis for ISO Speakers: New York State Food Laboratory s Quality Assurance Team

GUIDELINES FOR USE BY CLASS A LICENSEES IN MEETING SNP REQUIREMENTS JULY 1996

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Ecology Quality Assurance Glossary

LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037

DIVISION OF ENVIRONMENT QUALITY MANAGEMENT PLAN PART III: SUPERFUND (MACA) PROGRAM QUALITY ASSURANCE MANAGEMENT PLAN

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

How to Verify Performance Specifications

Terms concerned with internal quality control procedures

Data Quality Objectives and Quality Assurance Project

QUALITY ASSURANCE and QUALITY CONTROL DATA STANDARD

Data Management Implementation Plan

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10

Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs

4.2 Bias, Standards and Standardization

Procedure for Writing Technical Procedures

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

Assessment of Accuracy and Precision

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Sample Management SOP 2v8 Page 1 STANDARD OPERATING PROCEDURE NO. 2 SAMPLE MANAGEMENT REVISION LOG. Revision Number Description Date

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Minnesota Pollution Control Agency. Quality Assurance Project Plan Guidance

USE OF REFERENCE MATERIALS IN THE LABORATORY

QUALITY ASSURANCE PROJECT PLAN FOR BROWNFIELDS/VOLUNTARY CLEANUP PROGRAM SITES

Medical Cannabis Laboratory Approval Program

Continuous Monitoring Manual

Commonwealth of Kentucky. Wastewater Laboratory Certification Manual

QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis

EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM. LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007)

DATA QUALITY OBJECTIVES

NADP AMoN Operations Manual, Version 1.3. Ammonia Monitoring Network Site Operations Manual

Analytical Test Method Validation Report Template

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES

Procedure for Equipment Calibration and Maintenance

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

PERFORMANCE-BASED EVALUATION OF LABORATORY QUALITY SYSTEMS An Objective Tool to Identify QA Program Elements that Actually Impact Data Quality

A.1 Sensor Calibration Considerations

Quality Assurance (QA) & Quality Control (QC)

Guidance for Industry

Quantifying Bacterial Concentration using a Calibrated Growth Curve

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

Canadian Environmental Technology Verification Program Applicant Information Package CONTENTS

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

ALACC Frequently Asked Questions (FAQs)

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

COMMISSION DIRECTIVE 2003/94/EC

Glossary. Acceptable Value - An observed or corrected value that falls within the acceptable range.

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

Calibration Verification

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Measurement and Calibration

Dissolved and precipitated oxalate

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

REDUCING UNCERTAINTY IN SOLAR ENERGY ESTIMATES

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

4 QUALITY ASSURANCE 4.1 INTRODUCTION

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

DIVISION OF ENVIRONMENT QUALITY MANAGEMENT PLAN VOLUNTARY CLEANUP AND PROPERTY REDEVELOPMENT PROGRAM QUALITY ASSURANCE MANAGEMENT PLAN

LAB 37 Edition 3 June 2013

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

Emission monitoring, reporting, and verification (MRV) for U.S. cap and trade programs. Jeremy Schreifels US EPA

IAEA-TECDOC Development and use of reference materials and quality control materials

EPA GUIDANCE FOR QUALITY ASSURANCE PROJECT PLANS EPA QA/G-5

ESS Method 130.1: General Auto Analyzer Procedures

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3

NC SBI QUALITY ASSURANCE PROGRAM

MIDLAND AREA SOIL SAMPLING MIDLAND, MICHIGAN

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Laboratory Information Management System (LIMS) in EMRL

The Analytical Balance

Experiment 1: Measurement and Density


Quality Assurance Manual

Texas Commission on Environmental Quality Page 1 Chapter Control of Air Pollution from Nitrogen Compounds

Quality Assurance For Environmental Laboratories

Transcription:

Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial distributions and temporal trends, -research into wet deposition processes, -research into aquatic and terrestrial effects of wet deposition, -the development and evaluation of long range transport computer models 1

Why QA/QC? There has been a strong demand by the scientific community for the quantification of data quality, particularly data uncertainty. The relatively new field of `quality assurance/quality control for wet deposition measurements' have been evolved. Many of the philosophies, principles and procedures in this area have been adapted from QA/QC practices for air pollution measurements and analytical laboratories. The difference between QA and QC The US EPA differentiates between QC and QA in the following way: ** QC is a `system of activities to provide a quality product ' and QA is a `system of activities to provide assurance that the QC system is performing adequately. In other words, QA is QC for QC'.** QC ensure that the analytical system is 'in control '(i.e. operating within specified limits). QA ensure that the quality control procedures are working correctly. 2

The five QA/QC elements QA programs generally address five QA/QC elements, namely: 1. Accuracy -the degree of agreement of a measurement with an accepted reference or true value. 2. Precision - a measure of the mutual agreement among individual measurements of the property, usually under prescribed similar conditions. 3. Completeness - a measure of the amount of valid data obtained from a measurement system compared to the amount that was expected to be obtained under correct normal conditions. 4. Representativeness -the degree to which data accurately and precisely represent a characteristic of a population or parameter variations at a sampling point, a process condition, or an environmental condition. 5. Comparability -the confidence with which one data set may be compared to another 3

1. Accuracy Because the absolute accuracy of wet deposition measurements cannot be defined, most QA program do not asses `overall accuracy'. In stead, they break the measurement system into component parts and determine the measurement uncertainty with each. Methods for assessing uncertainty of component parts, they include: -submission of field blanks for defining errors due to sampling and sample handling procedures, -routine collection of field information for use in validating chemistry data, -special studies using standard reference solutions to determine errors induced by evaporation and dry deposition, -routine audits of the field, laboratory and data management systems, -submission of blind samples to the laboratory, -analysis of ultra pure water samples by the laboratory. 4

2. Precision Overall precision, i.e. precision of the entire measurement system, is measured by colocating two precipitation chemistry collectors. The two sets of instruments are operated independently and the precipitation samples are analyzed separately at the laboratory. The laboratory precision is measured by running duplicate analyses of the same samples, both within and between analytical runs. 3. Completeness There are four major data completeness issues of concern to wet deposition sampling. They are: (1) the portion of the data summary period(season or year) with complete records of the amount of precipitation that fell (%PCL: Percent precipitation coverage length), %PCL = ((Number of days in the summary period)-(number of days with missing or unknown precipitation))/(number of days in the summary period)*100 (2)the potion of the precipitation associated with valid chemical analysis and valid sample collection (%TP: Percent total precipitation), %TP = (Sum of precipitation amounts for samples with valid sample component measurements)/(sum of precipitation amounts for all samples)*100 (3)the fraction of the total number of sampling periods(daily, weekly, monthly etc.) having valid concentration data (%VSMP), and (4)the percentage of time in the summary period represented by samples having valid concentrations (%VSL). 5

4. Representativeness Site representativeness can be assured and controlled by auditing the sites and their surrounding at regular intervals. This ensures that the program's siting criteria are not being violated without the knowledge of the data users. Timely modification or relocation of sites can be used to correct violations of the siting criteria. Sample representativeness can be assured by minimizing sample contamination and maximizing collection efficiency of the precipitation chemistry collectors. Quality control measurers such as wearing gloves during sample changes and adjusting collector sensors to maximize sensitivity can be used. 5. Comparability Comparability is often referred to as `relative accuracy'. There are two methods used to address comparability in wet deposition measurement programs. One is to carry out co-located sampling of several measurement programs at the same site, and the other is carry out laboratory inter-comparison studies. The former focuses on overall measurement comparability while the latter addresses laboratory comparability. 6

QA/QC for Field Operations -Sample Collection and Handling (1) Collection vessel preparation -Cleanness of collection funnel, tubes and bottles should be noted, e.g. conductivity tests on rinse water, wearing plastic gloves when placing new sampling bottles. Sample collection -Samples should be collected from the precipitation chemistry collector without contamination, e.g. approaching the collector from the downwind side, keeping fingers out of collector vessels. Sample Collection and Handling (2) Sample handling -Samples should be handled and stored on-site without contamination or chemical change. Field data reporting -Complete and accurate data reporting are important. Sample shipping-shipping procedures are also important to arrive at the laboratory within specified times and without spilling and chemical change, e.g. the use of curie systems to minimize travel time and the use of ice-pack. 7

QA/QC for Laboratory Operations QA/QC for Laboratory Operations can be broken down into next functions: -laboratory support for field operations -laboratory sample handling (e.g. weighing samples upon arrival to check the field weight), -chemical analysis (e.g. calibration checks, replicate analyses, blank checks, control charting), -sample keeping, -corrective action, -data reporting, -inter laboratory inter comparisons, and -quality assurance reporting. Major QC in laboratories Many laboratory quality control procedures exist. Some are qualitative and some are quantitative. 8

The qualitative procedures Include: -Good laboratory practice(glp) -Good sample handling practice -Definition of standard operating procedures(sops) -Preventive maintenance of instruments -Analyst training and upgrading -Appropriate facilities and instrumentation -Proper documentation -Laboratory inspections or audits -Laboratory safety -Well-defined chain-of-custody -Traceability of standards The quantitative procedures Control charting (1) A statistical technique designed to ensure that a laboratory analysis system is in statistical control. Control charts are plots on which quality control data are plotted (ideally on a real-time basis) against historically-determined statistical limits which identify 'in-control', 'warming', and 'out-of-control' situation. In most cases, control chart data come from the analysis of calibration control solutions, within-run and between-run replicate samples, reagent blanks, and so on. 9

Control charting (2) The control limits are determined from the mean(xbar) and standard deviation(σx) of repeat measurements of these samples. The warning limits at Xbar ±2σ The control limits at ±3σ This is done by plotting the QC sample results against the predetermined control limits after each analytical run. If the control limits are exceeded, then the analytical run is repeated. Instrument calibration and calibration checks (1) It should be kept in mind, -making all calibration solutions from ultra-pure or reagent grade chemicals traceable to reference standards -checking that the slope, intercept, and correlation co-efficient of the calibration curve are within acceptable range 10

-using calibration standards that cover a concentration range of at least 95% of the expected sample concentrations, and using more standards in non-linear section of the calibration curve; -calibrating at the beginning of each analytical run, and at the end of each run; -analyzing and control-charting one or more calibration check solutions in each batch of samples (note that these check solutions are normally secondary in-house reference standards) Instrument calibration and calibration checks (2) Calibration checks by standard reference material Instrument calibrations are quality controlled by analyzing standard reference materials(srms) at regular intervals. The accuracy of the analytical system is normally reported as the deviation from the true value, i.e. %Error=(Xbar-X srm )/X srm, where Xbar: mean value of repeated measurements of the SRM, X srm :certified value of SRM 11

Replicate sample analyses Replicate analysis of samples provides a measure of analytical precision. Two types of replicate analyses are carried out -within-run and between-run. Both within-run and between-run precision are calculated as the standard deviation(sd) of replicate analyses X and Y of samples i from 1 to n. When the standard deviation is independent of concentration, it is calculated as SD={Σ(Xi-Yi) 2 /2n} 1/2 Precision is frequently expressed as the coefficient of variation of the replicate analyses, i.e. the SD/mean concentration. The frequency of replicate analyses varies from laboratory to laboratory, but normally represents about 4 to 5 % of the sample load. Major QA in laboratories Quality assurance procedures ensure that the quality control system is in working correctly. Two QA procedures are of particular note: Blind audits Blind audits are programs in which standard reference solution(or precipitation samples of known concentration) are submitted to the chemical laboratory without the knowledge of the laboratory staff (i.e. blindly). Laboratory inter comparison studies This studies are used to determine the comparability (or relative accuracy) of several laboratories. 12

Quality Assurance Reporting The final deliverable of every quality assurance program is the assessment and reporting of data quality. This is achieved in many different ways, one of the most common of which is the publication of formal Quality Assurance Reports. The second methods of reporting is the journal or report publication that focuses on a specific aspect of quality. Summary In without quality control procedures in place, laboratories cannot hope to remain in-control, and without quality assurance procedures in place, they cannot ensure that they are in-control. Also, without formal reporting of the QA/QC results, laboratories cannot demonstrate to data users that they are producing high quality analytical data. It is therefore very important that Quality Assurance Programs adopt QA/QC reporting as a regular laboratory function. 13