Methods and Equipment for Calibration of Variable Area Meters (Rotameters)



Similar documents
ACFM vs. SCFM vs. ICFM Series of Technical White Papers from Ohio Medical Corporation

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES

QUESTIONS THERMODYNAMICS PRACTICE PROBLEMS FOR NON-TECHNICAL MAJORS. Thermodynamic Properties

Meriam Instrument LFE Data Sheet

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Barometric Effects on Transducer Data and Groundwater Levels in Monitoring Wells D.A. Wardwell, October 2007

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *

Equivalents & Conversion Factors 406 Capacity Formulas for Steam Loads 407 Formulas for Control Valve Sizing

2 1/2 Pipe. 40 = height. the gauge pressure inside the vessel from the gauge pressure at the nozzle inlet as shown:

Technical Information

The Fundamentals of Gas Flow Calibration

Air Eliminators and Combination Air Eliminators Strainers

User s Manual Laminar Flow Elements

Pump Formulas Imperial and SI Units

The Precharge Calculator

= 800 kg/m 3 (note that old units cancel out) J 1000 g = 4184 J/kg o C

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection.

Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document

Energy Recovery Ventilation Equipment

For Multi-Parameter Meters see mvx

Instrument Loop Diagrams

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

Water Technologies Wallace & Tiernan Flow Measurement Equipment. Purge, Flow and Varea-Meter Products

PSAC/SAIT Well Testing Training Program Learning Outcomes/Objectives Level B

BaseCal 3. BaseCal, the web-based performance calculation tool for Diaphragm Seals

Application of the Orifice Meter for Accurate Gas Flow Measurement page 1. Application of the Orifice Meter for Accurate Gas Flow Measurement.

APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS

Laminar Flow Elements

Vogt Valves The Connection Bulletin for Forged Steel Flow Control Valves CB 15

Specific Volume of Liquid (Column 7). The volume per unit of mass in cubic feet per pound.

Technical & Sizing Bulletin for Meriam LFEs. Laminar Flow Elements

Energy Systems Engineering Technology

The Use of Control Valve Sizing Equations with Simulation Based Process Data

Viscosity - Flow Cups. elcometer.com

Installation, Operation and Maintenance Manual

Selecting a Centrifugal Pump to Handle a Viscous Liquid


ANSI/IEC Degrees of Protection Provided by Enclosures (IP Code) (identical national adoption)

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class # 2010

Establishment of the Gas Flow Calibration Facility at CSIR-NPL, India

Flow Measurement. Reference - Flow Measurement FUNDAMENTALS. ρ = fluid density (lb/ft3 or kg/m3)

Grade 8 Science Chapter 9 Notes

Technical Bulletin 144. Air Volume Conversions. by:

Turbine Flowmeters for Gas Applications

CONTROL VALVE PRESSURE DROP AND SIZING

SPE Copyright 1999, Society of Petroleum Engineers Inc.

Examination syllabuses for Manual Stack emissions monitoring. Environment Agency Version 9 January 2015

Preferred SI (Metric) Units

HEAVY OIL FLOW MEASUREMENT CHALLENGES

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid.

IN LINE FLOW-METER CALIBRATION ADVANCE METHODS AND SOLUTIONS

SIZING AND CAPACITIES OF GAS PIPING

LAB #3: MEASURING SPECIFIC GRAVITY AND DENSITY. Set-up and Materials for Experiment

Flow Measurement in Pipes and Ducts. Flow Measurement in Pipes and Ducts, Course #503. Presented by:

NOTE: FOR PROJECTS REQUIRING CONTRACTOR MIX DESIGN, THE DESIGN PROCEDURES ARE SPECIFIED IN THE SPECIAL PROVISIONS OF THE CONTRACT.

Variable V Area Flowmeter Rotameter The Original

THE HUMIDITY/MOISTURE HANDBOOK

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Validation of process gas systems

Pressure and Temperature Controls

ME Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS

Liquid level measurement using hydrostatic pressure and buoyancy

CO MPa (abs) 20 C

FLUID FLOW Introduction General Description

Thermal Dispersion Mass Flow

Experiment 3 Pipe Friction

Commercial refrigeration has been in the environmental. Refrigerant. as a. Basics Considerations PART 1:

Volume Correction Factor Calculation Development in American Petroleum Institute

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER

An Overview of Calibration Methods and Procedures for Process and Inventory Weigh Systems

Understanding Pressure and Pressure Measurement

Inventory Control. Inventory Control Tank & Silo Weighing

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

p atmospheric Statics : Pressure Hydrostatic Pressure: linear change in pressure with depth Measure depth, h, from free surface Pressure Head p gh

HVAC and REFRIGERATION

The Unique Accelabar Flow Meter

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

American National Standard For Electrical Rigid Steel Conduit (ERSC)

Calibration of Volumetric Glassware

American National Standard For Electric Power Systems and Equipment Voltage Ratings (60 Hertz)

Centrifugal Fans and Pumps are sized to meet the maximum

CHAPTER 2: LIQUID VISCOSITY MEASUREMENT

STEPS TO COMPRESSOR SELECTION & SIZING

EXPERIMENT 9 Evaluation of the Universal Gas Constant, R

Pipe Sizes For Water Distribution System Design

Theory overview of flow measurement using differential pressure devices based on ISO-5167 standard.

Heat Exchangers - Introduction

QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08

Rosemount 1199 Submersible Seal

TELEDYNE HASTINGS INSTRUMENTS

Custody Transfer Measurement. with the V-Cone Flowmeter

Gas Line Sizing. Z Maximum gas demand shall be determined by adding all of the equipment Btu ratings from appliances connected on the system.

_CH06_p qxd 1/20/10 9:44 PM Page 69 GAS PROPERTIES PURPOSE

Archimedes Principle. Biological Systems

DuPont Kapton HN. polyimide film

Ambient Pressure = and Pressure, collectively

OPTIMASS 1400 Technical Datasheet

Transcription:

RECOMMENDED PRACTICE ISA RP16.6 1961 Methods and Equipment for Calibration of Variable Area Meters (Rotameters) Approved 30 June 1961

ISA RP16.6 1961, Methods and Equipment for Calibration of Variable Area Meters (Rotameters) ISBN 0-87664-345-4 Copyright 1961 by the Instrument Society of America. All rights reserved. Printed in the United States of America. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any means (electronic, mechanical, photocopying, recording, or otherwise), without the prior written permission of the publisher. ISA 67 Alexander Drive P.O. Box 12277 Research Triangle Park, North Carolina 27709

Preface This Recommended Practice has been prepared as a part of the service of ISA toward a goal of uniformity in the field of instrumentation. To be of real value this report should not be static, but should be subject to periodic review. Toward this end the Society welcomes all comments and criticisms, and asks that they be addressed to the Standards and Practices Board Secretary, ISA, 67 Alexander Drive, P.O. Box 12277, Research Triangle Park, North Carolina 27709, e-mail: standards@isa.org. This report was prepared by the 8D-RP16 Committee of the Production Processes Standard Division on Variable Area Meters. NAME COMPANY L.N. Combs, Chairman E.I. du Pont de Nemours & Company S. Blechman Brooks Instrument Co., Inc. W. F. Dydak Schutte & Koerting Company R. W. Eberly Schutte & Koerting Company H. V. Mangin Fischer & Porter Company T. E. Quigley E.I. du Pont de Nemours & Company R. L. Shapcott Daystrom, Inc., Weston Instrument Div. The assistance of those who aided in the preparation of this Recommended Practice by answering questionnaires, offering suggestions, and in other ways is gratefully acknowledged. The following have reviewed the Report and served as Board of Review: NAME COMPANY S. A. Cole Wallace & Tiernan, Inc. W. A. Crawford E.I. du Pont de Nemours & Co., Inc. R. C. Kimball American Viscose Corporation H. S. Kindler ISA H. L. Knight E.I. du Pont de Nemours & Co., Inc. N. L. Massey Badger Manufacturing Company Approved for Tentative Publication by the Standards & Practices Board in June, 1961 NAME COMPANY E. A. Adler United Engineers & Constructors, Inc. Standards & Practices Department Vice President F. E. Bryan Douglas Aircraft Company Aerospace Standards Division Director R. E. Clarridge General Electric Company Intersociety Standards Division Director W. J. Ladniak Minneapolis Honeywell Reg. Company Nuclear Standards Division Director F. H. Winterkamp E.I. du Pont de Nemours & Co., Inc. Production Processes Standards Division Director E. J. Minnar ISA Headquarters Staff Standards & Practices Board Secretary ISA-RP16.6-1961 3

Contents 1 Purpose... 7 2 History and development... 7 3 Scope... 7 4 Methods and equipment... 7 ISA-RP16.6-1961 5

1 Purpose This Recommended Practice is intended to provide a useful guide for the general considerations important to the calibration of variable area meters (rotameters). 2 History and development The need for this Recommended Practice was established by a survey of instrument users sent out initially in September, 1954, to about two-hundred-and-fifty meter users and manufacturers. 3 Scope This Recommended Practice has been prepared to describe the methods and equipment used for calibrating the glass and metal metering tube area meters (rotameters) covered in RP16.1.2.3. Nomenclature and terminology for extension devices are included in ISA-RP16.4. Installation, operation and maintenance instructions are included in ISA-RP16.5. 4 Methods and equipment When calibrating a rotameter, certain techniques for determining the proper equipment and procedures must be considered. Care should be taken in selecting equipment for controlling flows, measuring volumes and time intervals, etc., with acceptable accuracies of measurement. The use of the rotameter dictates the accuracy of calibration required. When the rotameter is used as a flow indicating device and repeatability of readings is of paramount importance the rotameter can usually be calibrated with a reference liquid such as water and corrections for density can be made using the formulas presented in Paragraph 4.1.1. To determine whether the viscosity of the metering liquid will permit calibration with a reference liquid, refer to Chart, Figure RP16.6.1. If absolute accuracy of measurement is most important, the rotameter should be calibrated with the process fluid that it will eventually handle or kinematic simulation should be employed to assure highest absolute accuracy. (Refer to Paragraph 4.2) 4.1 Three basic methods are employed in the calibration of rotameters. They are volumetric, gravimetric and comparison. ISA-RP16.6-1961 7

4.1.1 Volumetric In these methods the volume of fluid flowing is accurately measured and timed as it passes through the rotameter into the collecting chamber at a controlled rate. Instruments in this group include gasometers, burettes, and stand pipe systems which provide measurements directly in basic volumetric terms. Liquids For liquid calibrations, the collecting chamber may vary from a small Bureau of Standards certified burette for calibrating very small rotameters to large accurately calibrated receiving tanks for calibrating larger instruments. (Refer to Figure RP16.6.2) Formula used with Figure RP16.6.2 Liquid Calibration Q m = V ---------- c 60 Sec ( U f U m )U --------------------------- c ( U f U c )U m Where: Q m = Volumetric flow rate of liquid to be metered in units per minute V c = Volume of calibrating liquid collected in units consistent with Q m Sec = Collection time in seconds = Density of metering float in grams/cc U f U m U c = Density of liquid to be metered in grams/cc = Density of calibrating liquid in grams/cc Liquid to Gases Rotameter sizes 1/2" and larger may be calibrated volumetrically using water as the flowing medium. GPM of liquid, specific gravity 1.0 times 4.15 equals SCFM gas, specific gravity 1.0, when metered at 0 psig and 70 F. Gases For calibrating rotameters with gases, a gasometer as illustrated in Figure RP16.6.3 is recommended. In this method, gas at controlled flow rate, temperature, and pressure flows through the rotameter and is collected in the inverted bell. Formulas used with Figure RP16.6.3 Gas Calibration To determine the calibration of the rotameter undergoing test, use the following formulas: 1. QStd= V ---------- g Sec ---------------------------------------------------------------- 60 S c ( P b + P g ) 530 S o 14.7 T g V c ----- V 1 2. QAct= V ---------- g Sec ----------------------------------------------------------------------------- 60 S c ( P b + P g ) 530 V 1 V 1 14.7 T g 13.35 V c ----- V 1 3. W= V g ---------- Sec 60 0.075 Sc ( P b + P g ) 530 ------------------------------------------------------------------------------------ 14.7 T g V c ----- V 1 8 ISA-RP16.6-1961

Where: QStd = Std. CFM at 14.7 psia & 70 F. QAct = Actual CFM (volume at metering temperature and pressure) W = PPM Flow Rate Vg = Volume gas collected in cu. ft. Sec = Collection time, seconds T g = Gasometer temperature, degrees Rankine (460 + F.) P b = Barometric pressure at calibration (psia) P g = Static pressure of Gasometer (psi) (in. H 2 O x.0361) S c S o = Specific Gravity of calibrating gas at 14.7 psia and 70 F. (Air=1.0) = Specific Gravity of gas to be metered at 14.7 psia and 70 F. *V c = Specific Volume of calibrating gas at calibration temperature and pressure, cu. ft./lb. *V 1 = Specific Volume of gas to be metered at metering temperature and pressure, cu. ft./lb. *For perfect gases Where: = metering pressure at calibration, psia P c V c = 13.35 ---------- 14.7 P c T c --------- 530 1 ----- S c T c = metering temperature at calibration, degrees Rankine (460 + F.) Where: = operating metering pressure, psia P 1 V 1 = 13.35 ---------- 14.7 P 1 T 1 --------- 530 1 ----- S o T 1 = operating metering temperature, degrees Rankine (460 + F.) NOTE: Standard conditions in rotameter industry usually stated as 14.7 psia and 70 F. 4.1.2 Gravimetric In this method the liquid at controlled flow rates passes through the rotameter under test and is collected in a receiving tank. The weight of liquid is accurately measured by a precision scale or other weighing device. Flowing time is accurately recorded and flow rate in gravimetric units computed directly by deciding scale readings by time. Automatic operation may be achieved by means of photocell or mercury contact timing devices. Equipment for this method of measurement weighs the liquid passing through the rotameter under test and provides flow rate readings in gravimetric or weight units. Unlike volumetric systems, the readings taken from the gravimetric systems are independent of density or viscosity of the fluid. (Refer to Figure RP16.6.4) ISA-RP16.6-1961 9

Formula used with Figure RP16.6.4 Liquid Calibration W W c 60 m = ------------------- Sec ( U f U m )U ----------------------------- m ( U f U c )U c Where: W m = Mass flow rate of fluid to be metered in pounds per minute W c Sec U f U m U c = Weight of calibrating fluid collected in pounds = Collection time in seconds = Density of metering float = Density of liquid to be metered = Density of calibrating liquid 4.1.3 Comparison In this method, liquid and gas calibrations are made by comparing the rotameter under test against an accurately calibrated flow measuring instrument (See Figure RP16.6.5). Calibration accuracy will depend entirely on the accuracy of the calibrated flow measuring instrument serving as a secondary standard. 4.2 Kinematic simulation When it is impractical to calibrate rotameters directly because of viscosity considerations, hazards to the operator or equipment, or for economic reasons, kinematic simulation of fluid properties is employed. Through the use of a variety of fluid media and by maintaining controlled temperature conditions, viscosities and densities over a wide range can be achieved to match those physical properties of the fluid for which the rotameter is designed. Formulas used are the same as those under 4.1. Kinematic simulation follows the equation: Centistokes cal. fluid = Centistokes Metering Fluid Fluid density metering fluid -------------------------------------------------------------------------- Fluid density calibration fluid 4.3 Field check Any of the above methods may be used to check rotameters. Tank gaging may be used for checking liquid handling rotameters and liquid displacement methods may be used for rotameters handling gases. Actual results will be commensurate with the care employed in the calibration and the precision of the facilities. 10 ISA-RP16.6-1961

Figure RP16.6.1 ISA-RP16.6-1961 11

Figure RP16.6.2 12 ISA-RP16.6-1961

Figure RP16.6.3 ISA-RP16.6-1961 13

Figure RP16.6.4 14 ISA-RP16.6-1961

Figure RP16.6.5 ISA-RP16.6-1961 15

Developing and promulgating technically sound consensus standards, recommended practices, and technical reports is one of ISA's primary goals. To achieve this goal the Standards and Practices Department relies on the technical expertise and efforts of volunteer committee members, chairmen, and reviewers. ISA is an American National Standards Institute (ANSI) accredited organization. ISA administers United States Technical Advisory Groups (USTAGs) and provides secretariat support for International Electrotechnical Commission (IEC) and International Organization for Standardization (ISO) committees that develop process measurement and control standards. To obtain additional information on the Society's standards program, please write: ISA Attn: Standards Department 67 Alexander Drive P.O. Box 12277 Research Triangle Park, NC 27709 ISBN: 0-87664-345-4