UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES



Similar documents
Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day

Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC

How To Know If A Mobile App Is A Medical Device

BSI: An In Vitro Diagnostics Notified Body. A guide to the In Vitro Diagnostic Directive....making excellence a habit.

Medical Devices: CE Marking Step-by Step

Med-Info Regular Information Bulletins for the Medical Device Industry

2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

Reporting Changes to your Notified Body

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

Medical Device Training Program 2015

GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

Want to know more about the Notified Body?

Safeguarding public health The Regulation of Software as a Medical Device

RECALLS in EUROPE. Past, present, near & further future. Gert Bos BSI Medcon May 2012

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

CE Marking: Your Key to Entering the European Market

Medical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011

Quality Management System Certification. Understanding Quality Management System (QMS) certification

Registration of Class B Medical Devices

Quality Management System Certification. Understanding Quality Management System (QMS) certification

ISO 27001: Information Security and the Road to Certification

Medical Device Marketing Authorization System (MDMA) MDMA User Manual

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

BSI Unannounced Audits

Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)

BSI Road Show: September 8 th to 15 th, 2014

SMF Awareness Seminar 2014

Comparative analysis between the possible regulatory approaches to GMP compliance TITOLO PRESENTAZIONE

MEDICAL DEVICES INTERIM REGULATION

Conformity assessment certification

Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994

Regulatory Considerations for Medical Device Software. Medical Device Software

(OJ L 169, , p. 1)

Navigating ISO 9001:2015

Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau

Manual for ITC's clients, Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC

Navigating the New EU Rules for Medical Device Software

Clinical evaluation Latest development in expectations EU and USA

Frequently Asked Questions. Unannounced audits for manufacturers of CE-marked medical devices. 720 DM a Rev /10/02

Initial appraisal of a European Commission Impact Assessment

Medical Device Software Do You Understand How Software is Regulated?

GUIDELINES ON MEDICAL DEVICES

Basic Information about the European Directive 93/42/EEC on Medical Devices

Medical Technologies Evaluation Programme Methods guide

IAF Mandatory Document for the Application of ISO/IEC in Medical Device Quality Management Systems (ISO 13485)

MEDICAL DEVICE LICENSING REQUIREMENTS Life Sciences British Columbia NRC-Industry Research Assistance Program

DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL

Global Regulatory Requirements for Medical Devices

Presented by Rosemarie Bell 24 April 2014

Final Document. Software as a Medical Device (SaMD): Key Definitions. Date: 9 December Despina Spanou, IMDRF Chair. IMDRF/SaMD WG/N10FINAL:2013

Preparing for Unannounced Inspections from Notified Bodies

Diagnostic Tests. Brad Spring Director, Regulatory Affairs

Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System

How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters

Changes to Medical Device Regulations

Konformitätsbewertung 3.9 B 17. Guidance for Notified Bodies auditing suppliers to medical device manufacturers

2.0 Rationale, Purpose and Scope Definitions General Principles... 7

GLOBAL FUND QUALITY ASSURANCE POLICY FOR DIAGNOSTICS PRODUCTS. (Issued on 14 December 2010, amended on 5 February 2014)

EA Document on. Accreditation. For Notification Purposes

EU DIRECTIVE ON GOOD CLINICAL PRACTICE IN CLINICAL TRIALS DH & MHRA BRIEFING NOTE

DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL

Revision of the Directive 98/79/EC on In Vitro Diagnostic Medical Devices. Response from Cancer Research UK to the Commission August 2010

Med-Info. Malaysia Medical Device Regulations. TÜV SÜD Product Service GmbH. International expert information for the medical device industry

The Regulation of Medical Devices in the UK

RAPS ONLINE UNIVERSITY

Software within the medical device regulatory framework in the EU

HOW CAN YOU REASSURE YOUR CUSTOMERS ABOUT THE ORIGINS OF YOUR WOOD & PAPER PRODUCTS?

Medical Device Regulatory Requirements for Taiwan

Guideline on good pharmacovigilance practices (GVP)

Auditing as a Component of a Pharmaceutical Quality System

Memorandum of Understanding

DRAFT GUIDANCE DOCUMENT ON THE LOW VOLTAGE DIRECTIVE TRANSITION

Guideline on good pharmacovigilance practices (GVP)

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Principles of In Vitro Diagnostic (IVD) Medical Devices Classification

Medical Device Approvals in Brazil: A Review and Update

NOTICE. Re: GD210: ISO 13485:2003 Quality Management System Audits Performed by Health Canada Recognized Registrars

A GUIDE FOR THE QUALITY ASSURANCE

Medical Devices. IRELAND Arthur Cox

UNDERSTANDING THE FSSC FOOD SAFETY SYSTEM CERTIFICATION STANDARD

Mobile Medical Application Development: FDA Regulation

PLM: Privat Label Manufacturer (Customer of the OEM-PLM relation) OEM: Original Equipment Manufacturer (Supplier of the OEM-PLM relation)

Understanding Principles and Concepts of Quality, Safety and Environmental Management System Graham Caddies

Adoption by GCP Inspectors Working Group for consultation 14 June End of consultation (deadline for comments) 15 February 2012

Boost the Success of Medical Device Development With Systematic Literature Reviews

US & CANADA: REGULATION AND GUIDELINES ON MEDICAL SOFTWARE AND APPS OR

12040/1/15 REV 1 LES/ns 1 DG B 3B

Medical Device Classification

Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device

EPF Position Statement on the European Commission s proposal for a Regulation on In Vitro Medical Devices

Introduction to Q10 Pharmaceutical Quality System

Vigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September Copyright 2015 BSI. All rights reserved.

EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below.

Accreditation in Europe

INTERNAL MARKET SCOREBOARD

Transcription:

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES A WHITE PAPER ON THE REQUIREMENTS, REGULATIONS AND OPPORTUNITIES CONTAINED IN EC DIRECTIVE 98/79/EC AND A BRIEF COMPARISON TO OTHER GLOBAL REGULATORY SCHEMES FOR IVD MEDICAL DEVICES FEBRUARY 2010 AUTHOR Sharon Williams Global IVD Certification Manager, SGS

ABSTRACT The purpose of this white paper is to provide an introduction to the European Union s EC Directive 98/79/EC on in vitro diagnostic medical devices (IVDs). This document is not intended to provide every possible detail on the European regulatory requirements for IVDs or their implementation, due to the level of complexity involved. Instead it aims to advance understanding of the key requirements that manufacturers of IVDs need to comply with in order to sell their products in Europe, and briefly compares European regulations with the medical device regulatory frameworks of the United States, Canada and Japan. CONTENTS I. EXECUTIVE SUMMARY 1 II. REGULATORY CHALLENGE FOR IVD MANUFACTURERS 2 III. KEY REQUIREMENTS OF EC DIRECTIVE 98/79/EC ON IVDs 4 IV. EC DIRECTIVE 98/79/EC ON IVDs COMPARISON WITH OTHER KEY SCHEMES 8 V. CONCLUSIOn 9 I. EXECUTIVE SUMMARY Even with an increasing move toward global harmonisation between the major medical device regulatory schemes of Europe, North America and Asia, there is still a need for a detailed understanding of these regional schemes, especially for those responsible for product compliance with regulatory affairs. The European Union s regulatory scheme for IVD medical devices (EC Directive 98/79/EC on IVDs) 1 has been in place as a mandatory legal requirement since 2003, with the EU s product conformity marking (CE Marking) now permitting regulatory access to the 30 European nations that make up the European Economic Area (EEA) and Switzerland, a significant market region. EC Directive 98/79/EC (hereafter abbreviated as 98/79/EC) represents a robust and workable regulatory framework based on ISO 13485 for quality system compliance. This can integrate into a manufacturer s existing system and processes to meet the additional regulatory requirements and ensure device compliance. IVD manufacturers need the knowledge to make decisions on the conformity route they take to comply with 98/79/EC and, where necessary, the choice of the regulatory third party, the Notified Body that they work with to successfully navigate that process. 1. Official Journal of the European Communities L 331 vol. 41, 7 December 1998. 1

II. REGULATORY CHALLENGE FOR IVD DEVICE MANUFACTURERS The global market for IVDs is, unsurprisingly, an extremely regulated one. Safety for users and end users (the patient) of IVDs is of critical importance, which includes acceptable clinical performance for the intended use. The challenge for IVD manufacturers is to understand the complex array of regulations for the markets they sell into and how it applies to their specific products. By doing this, they can ensure that their products are compliant for efficient entry into that market and to remain on the market, maximising their commercial potential. Knowledge of the similarities and, sometimes even more importantly, the differences between global regions is crucial to meet regulatory requirements efficiently. Within Europe, the responsibility for ensuring the safety and health of the population with regard to IVD devices resides with individual EU member state governments, but is legally transposed from 98/79/EC and is implemented by each national Competent Authority 1. In order to maintain a consistent level of safety and performance throughout the region, these regulations apply to any IVD placed on the European market, regardless of where it was manufactured. Compliance with the provisions of 98/79/EC is demonstrated by the CE Marking displayed on the device, allowing free movement of the product within the region. The current extent of this European market includes the EU member states (27 at present) as well as those additional members of the European Economic Area (EEA) and Switzerland. Further EU candidate countries are expected to join in the future. 1. List of National EU Competent Authorities http://ec.europa.eu/enterprise/sectors/ medical-devices/files/list-of-contact-points-within-the-national_en.pdf 2

EC DIRECTIVE 98/79/EC HISTORICAL TIMELINE The stated aim of the European Commission in enacting 98/79/EC was, To ensure the highest level of patient safety while promoting the innovation and the competitiveness of this sector, allowing for market access, international trade and regulatory convergence. 1957 Treaty of Rome establishes the European Economic Community (EEC). 1985 Council of the European Communities Resolution on New Approach to Technical Harmonisation and standards. 1 1990 Council of the European Communities Directive on active implantable medical devices (AIMD) 90/385/EEC. 1993 Council of the European Communities Directive on Medical Devices (MDD) 93/42/EEC. 1993 European Union (EU) established by the Treaty of Maastricht, an economic and political union of member states, replacing and built upon the foundations of the EEC. 1994 Agreement on European Economic Area (EEA), which includes three of the four members of the European Free Trade Association (EFTA) Norway, Iceland and Liechtenstein and brings them into the EU Internal Market. 1998 Council of the European Union Directive on In Vitro Diagnostic medical devices 98/79/EC. 1999 Switzerland (fourth EFTA member) completes bilateral agreements with the EU across a wide range of areas, including movement of persons, transport and technical barriers to trade. 2003 ISO 13485 issued as a quality management system standard for medical device and IVD manufacturers and is recognised as a harmonised standard for the Medical Device Directives. 2007 Council of the European Union Directive 2007/47/EC, technical revision to AIMD and MDD. 2008 European Commission consults with stakeholders on the revision of the legal framework for medical devices a recast of all three Directives. 2 1. Official Journal of the European Communities C 136, 4.6.1985, p.1. 2. Recast of the Medical Devices Directives - Summary of the responses of the Public Consultation, Brussels, 5 December 2008 ENTR/F/3/D (2008) 39582. 3

III. KEY REQUIREMENTS OF EC DIRECTIVE 98/79/EC ON IVD DEVICES SCOPE AND LEGAL DEFINITIONS The scope of 98/79/EC applies to IVD medical devices and accessories, which can include not just reagents and kits but also instruments and software. Article 1 of 98/79/EC gives the legal definitions which confirm that IVDs are a specific category of medical devices with particular differences and are therefore controlled under this separate Directive. It builds upon the basic medical device definition to state that: An IVD medical device is any medical device which is a reagent, reagent product, calibrator, control material, kit, instrument, apparatus, equipment or system, whether used alone or in combination, intended by the manufacturer to be used in vitro for the examination of specimens, including blood and tissue donations, derived from the human body, solely or principally for the purpose of providing information: concerning a physiological or pathological state; or concerning a congenital abnormality; or to determine the safety and compatibility with potential recipients; or to monitor therapeutic measures. This also includes specimen receptacles, but general laboratory equipment is excluded. The first step for any manufacturer is to determine that their device actually falls under the above definition, with a clear medical purpose. ESSENTIAL REQUIREMENTS (ER) AND HARMONISED STANDARDS 98/79/EC, because it has to apply to the whole spectrum of IVDs, does not define specific technical factors but instead identifies broad based essential safety requirements to be met by all devices. More detailed technical solutions to compliance are given by the specifically developed or adopted European harmonised standards, which give a presumption of conformity to particular areas of the Essential Requirements (ER) in Annex I of the Directive. ISO 13485 is recognised as the harmonised standard for regulatory quality system compliance, and all such standards are published in the EU s Official Journal. 4

CLASSIFICATION The risk presented by a device determines the classification, and therefore the level of control and regulatory review required. Annex II of the Directive identifies specific device types that are categorised as either high risk (List A) or moderate risk (List B), so only the particular devices meeting that description are classified as such, for example only reagents for the tumour marker PSA are List B but no other equivalent cancer marker. Self-test IVDs, because of the greater risk associated with being used by untrained lay users, have special requirements while all other devices not classified as either List A, List B or self-test are regarded as general IVDs. TECHNICAL DOCUMENTATION AND PERFORMANCE EVALUATION A key obligation is for all IVDs to have Technical Documentation, as described in Annex III, to provide the supporting evidence for their compliance with the Essential Requirements. Central to this documentation will be the performance evaluation of the device, demonstrating that the performance meets the claims and intended use specified by the manufacturer. For List A devices the Common Technical Specification (CTS) give the criteria for the expected state-of-the-art performance evaluation. CONFORMITY ASSESSMENT ROUTES AND NOTIFIED BODY ROLE As a New Approach Directive, 98/79/EC aims to give manufacturers different options to reach conformity with the requirements. These different conformity routes are dependent on the classification of the device and hence are linked to the perceived risk of that product. Under the EU s regulatory scheme, each national Competent Authority has responsibility for the implementation of the Directive in their country, but pre-market approval is delegated to specifically designated independent third parties called Notified Bodies. For higher risk devices (List A, List B and Self-Test) this means that the manufacturer must gain independent certification by a Notified Body in order to complete the conformity route process, apply for CE Marking and be able to place the device on the European market. For the very highest risk devices in List A, there are additional measures required to address this high risk. Under the Full QA conformity route of Annex IV, each List A device must undergo a detailed review of the technical documentation called the EC Design Examination. Once approved for CE Marking, by either of the permitted routes, every batch of List A product has on-going batch verification before it is released into the market to ensure the continued consistent performance of the device. 5

Class of IVD Available Conformity Routes Type of Route Additional Requirements as per Route Notified Body Required Annex IV Full quality assurance (QMS - design and manufacturing) EC Design Examination + Batch Verification Yes List A Annex V and Annex VII EC Type Examination (product testing) and Production quality assurance (QMS - manufacturing) + Batch Verification Yes Annex IV Full quality assurance (QMS - design and manufacturing) n/a Yes List B Annex V and Annex VII EC Type Examination (product testing) and Production quality assurance (QMS - manufacturing) n/a Yes Increasing level of device risk Self-Test IVDs General IVDs Annex V and Annex VI Same as for List B plus: Annex III section 6 Annex III EC Type Examination (product testing) and EC Statistical Verification (product testing) EC Design examination for self-test products only EC Declaration of Conformity (self declaration by manufacturer) n/a n/a n/a Yes Yes No 6

REGISTRATION AND AUTHORISED REPRESENTATIVES A remit for a European databank to share critical data between competent authorities is described in 98/79/EC, but currently the project is still underway. When in place, it will mean that only a single registration in the manufacturer s home country or where their Authorised Representative (AR) is located is necessary, but in the meanwhile an interim situation requires that initial registration but then also a notification to other member state s competent authorities. Manufacturers need to research the specific implementation requirements for each member state to confirm if they require notification if they are a potential market country. Non-EU manufacturers need only one AR within the EU region, to act on their behalf and be a legal contact point, but they need to ensure they have a suitable agreement or contract with the AR that clearly defines the AR s duties. LABELLING AND CE MARKING Essential Requirements (ER 8.1 8.7) define the type and minimum content of information to be provided to the user, where the labelling includes both the labels and Instructions for Use (IFU) of the device. This includes clearly identifying the legal manufacturer and the intended use of the IVD device. Additional criteria apply to self-test devices, focussing on the clarity of the information for non-professional users. There are national language requirements for labelling, which are obviously mandatory for self-test IVDs where the user has to have instructions in their own language but there may be some flexibility regarding professional use devices on acceptable alternative languages, such as English. Manufacturers need to research for each member state it plans to sell their product in, to ensure they are compliant. The IVD manufacturer is responsible for any translations and must treat outsourced translation services as subcontracting and control it appropriately. CE Marking, in the correct format defined in Annex X, is displayed on labelling along with details of an AR if one is involved, and allows free movement of the device throughout the EU, the EEA and Switzerland. VIGILANCE & POST-MARKET SURVEILLANCE While Notified Bodies are involved in the pre-market approval process, the Competent Authorities (CA) concentrate on post market surveillance. Manufacturers are required to set up an ongoing systematic process to review experience gained for their device on the market and to have a vigilance procedure to immediately inform relevant Competent Authorities of incidents involving their device (98/79/EC article 11). This is a critical area and European Commission guidance for manufacturers is available 1, which was last updated in 2009. Each member state CA still has the ultimate power to invoke the safeguard clause (98/79/EC article 8) to remove a device that they believe is unsafe from their national market. 1. MEDDEV 2.12/1 rev.6 on medical device vigilance system, dated Dec 2009. 7

IV. EC DIRECTIVE 98/79/EC ON IVD DEVICES COMPARISON WITH OTHER GLOBAL SCHEMES The technology and prevalence of IVDs has changed over the years, and the corresponding regulations to control these products have also developed with ever more regions worldwide defining their own regulations for IVDs. 98/79/EC has many concepts in common with other key global regulatory systems such as: classification based on device risk; pre-market approval; basic safety criteria; QMS requirements (often based on ISO 13485); post market surveillance; and risk management throughout the life of the device. However, it is the differences in the specific details between these schemes that require particular attention. The first key area is consideration of the actual definition of an IVD product. Under US, Canadian and Japanese regulations, IVDs are just considered to be types of medical devices and come under their general medical device regulations, compared to the specific regulation of the IVD Directive. Under 98/79/EC, device classification is determined by a listing system of individual device types while Canada uses a rule based classification and in US and Japan there are defined product categories, so the same device may not have the same classification in each region as the method of determining it varies. Future developments of 98/79/EC will most likely see a change to a rule based classification for IVDs, in line with the Global Harmonisation Task Force (GHTF) guidance, and general medical devices under the Medical Devices Directive 93/42/EEC. The responsibilities of the national Competent Authorities in each EU country as well as the role and remit of the Notified Bodies is clearly defined in the 98/79/EC. In practice this can allow a faster and more predictable pre-market approval route, when compared to certain other global regulatory schemes, which benefits the manufacturer. In other regions the proportion of involvement of third party Conformity Assessment Bodies (CAB) varies, from quality system assessment by Accredited Persons in the US and Recognised Registrars in Canada, to technical documentation review for moderate risk devices by Registered Certification Bodies in Japan. EU USA Canada Japan Regulatory Authority EU National Competent Authorities US Food and Drug Administration (FDA) Health Canada Japanese Ministry of Health, Labor and Welfare (MHLW) Regulatory scheme 98/79/EC (IVDs only) Code of Federal Regulations Title 21 (21CFR) (medical devices and IVDs) Canadian Medical Device Regulations (MDR) (medical devices and IVDs) Japanese Pharmaceutical Affairs Law (JPAL) (medical devices and IVDs) Conformity Assessment for pre-market approval 3rd Party Notified Bodies for high and moderate risk devices Registration with CA only for low risk devices FDA process: pre-market approval (PMA) for high risk, 510(k) and exemption for some low risk devices Site inspection by FDA or 3rd Party under Accredited Persons Program for Quality System Requirements (QSR) Health Canada pre-market approval of technical documentation 3rd Party Recognised Registrar ISO 13485 certification under CMDCAS (Canadian Medical Device Conformity Assessment Scheme) for II, III and IV only Pharmaceutical and Medical Devices Agency (PMDA) approval for high risk products 3rd Party Registered Certification Body (RCB) assessment of moderate risk devices Notification of low risk devices with PMDA Classification Specific device lists in Annex II List A, List B, plus self-test devices (risk based) 21 CFR 862, 864, 866 medical speciality panels: based on intended use of device and also indications for use: Classes I, II, III Separate risk based set of rules for IVD devices Classes I, II, III, IV Risk based classification using product groups, based on GMDN*. Classes General, Controlled and Specially Controlled (I, II, III & IV) * GMDN - Global Medical Device Nomenclature system (http://www.gmdnagency.org/). 8

V. Conclusion 98/79/EC is a good example of state-of-the-art regulation, aiming to keep pace with product innovation and new clinical knowledge, and with future changes and improvements as part of a defined programme. It has harmonised regulatory requirements and allows for the free movement of goods across a large economic region of in Europe, giving clear responsibilities to manufacturers and the key entities within the regulatory framework (Notified Bodies and National Competent Authorities) with the ultimate aim of patient safety. The use of designated and monitored Third Party Notified Bodies permits the manufacturer choice and transparency in the regulatory process, allowing them to contract with a Notified Body they feel most compatible with and who is focussed on service delivery. In the wider global regulatory scene, working with a third party Conformity Assessment Body that is able to deliver regulatory certification and support across the global market regions that the manufacturer needs is critical. 9

ABOUT THE AUTHOR Sharon Williams Global IVD Certification Manager, SGS Sharon Williams has over eight years experience working in quality assurance and regulatory affairs roles within the IVD industry sector, and has spent the last ten years working for a UK Notified Body as an IVD Directive Specialist. She is now responsible for the technical development of the SGS s IVD sector certification services, including ISO 13485 and 98/79/EC, to further enhance the current portfolio of medical device regulatory certification schemes that SGS are recognised for globally. ABOUT SGS SGS is the world s leading inspection, verification, testing and certification company. Recognised as the global benchmark for quality and integrity, we employ over 59 000 people and operate a network of more than 1 000 offices and laboratories around the world. As your partner in medical device certification, SGS helps ensure your medical devices, including IVDs, are safe, legal and that they get to the market in the minimum amount of time and remain in the market allowing you to maximise their market potential. SGS has a global network of qualified medical device and pharmaceutical auditors able to provide auditing and training services from more than 40 countries. This allows SGS to support those customers to meet their regulatory obligations with recognised certification across a range of regulatory schemes that are essential for marketing their products around the world, including North America, the EU and Asia Pacific. In addition, SGS is a Notified Body under 98/79/EC with an initial scope of designation which includes all LIST B devices and Self-Test IVD devices. This scope will be further expanded to include all List A devices during the latter part of 2010. For more information, visit www.sgs.com/medicaldevices or email medicaldevices@sgs.com COPYRIGHT NOTICE The information contained in this document represents the current view of SGS SA on the issues discussed as of the date of publication. Because SGS must respond to changing market conditions, it should not be interpreted to be a commitment on the part of SGS, and SGS cannot guarantee the accuracy of any information presented after the date of publication. This White Paper is for informational purposes only. SGS makes no warranties, express, implied or statutory, as to the information in this document. Complying with all applicable copyright laws is the responsibility of the user. Without limiting the rights under copyright, no part of this document may be reproduced, stored in or introduced into a retrieval system, or transmitted in any form or by any means (electronic, mechanical, photocopying, recording, or otherwise), or for any purpose, without the express written permission of SGS. SGS may have patents, patent applications, trademarks, copyrights, or other intellectual property rights covering subject matter in this document. Except as expressly provided in any written license agreement from SGS, the furnishing of this document does not give you any license to these patents, trademarks, copyrights, or other intellectual property. ANY REPRODUCTION, ADAPTATION OR TRANSLATION OF THIS DOCUMENT WITHOUT PRIOR WRITTEN PERMISSION IS PROHIBITED, EXCEPT AS ALLOWED UNDER THE COPYRIGHT LAWS. SGS SA 2010. ALL RIGHTS RESERVED. 10

SGS SA 2010. ALL RIGHTS RESERVED. WWW.SGS.COM/MEDICALDEVICES