Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom



Similar documents
IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE

DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION

DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR

PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level

TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON

Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)

IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

DEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS

Lake at IPEN location

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

PERSONNEL MONITORING AND DOSIMETRY POLICIES

Ra, 232 Th and 40 K analysis in sand samples from some beaches of Great Vitória, Espírito Santo, Brazil: Preliminary results.

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

centre for radiation protection

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

Bonn Call-for-Action Joint Position Statement by the IAEA and WHO

A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS

Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?

Measuring Patient Exposure in Interventional Radiology

Introduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection

Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health

THE EFFECT OF THE HDDR PROCESS ON THE PRODUCTION OF Pr-Fe-Co-B-Nb SINTERED MAGNETS

Principles of dosimetry The ionization chamber

Radiation survey meters used for environmental monitoring

Lessons learned from the radiological accident in Mayapuri, New Delhi, India

Education and training activities

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

NUCLEAR ENERGY EDUCATION SCENARIO AROUND THE WORLD

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Dedicated computer software to radiation dose optimization for the staff performing nuclear medicine procedures

STUDY OF MAGNETISM IN GDTIN (T = NI, PD, CU) COMPOUNDS BY PERTURBED GAMMA-GAMMA ANGULAR CORRELATION SPECTROSCOPY

Training in Emergency Preparedness and Response

Natural and Man-Made Radiation Sources

RAD 114 RADIATION PROTECTION THREE LEVELS OF REGULATION NORMAL PROCESS OF CREATING LAWS

Sterile Cleanroom Management

BiO-dOSimetric Tools for triage to Responders. 14 th International Congress of Radiation Research Warsaw, Poland, August 28 - September 1, 2011

3rd International Symposium of the System of Radiological Protection Seoul, Korea, October 2015

The Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS)

notes MERRTT IS-302 Radiological Survey Instruments and Dosimetry Devices

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

National Laboratory for Metrology of Ionizing Radiation (LNMRI) 1

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry

Environmental Radiation Monitoring in Taiwan

Radiation Safety Characteristics of the NOMAD Portable X-ray System

THE PRINCIPLE OF JUSTIFICATION

Overview of a University and Affiliated Hospital Radiation Safety Program

VIRTUAL INSTRUMENTATION ON MOBILE DEVICES FOR DEPLOYMENT IN NUCLEAR INSTALLATIONS

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

Chapter XIII. PDD TAR TMR Dose Free Space

A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials

INTRODUCTION. A. Purpose

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Postgraduate Educational Course in Radiation Protection and the Safety of Radiation Sources Standard Syllabus

The Calibration Chain: Role of BIPM,

High-Energy Electron Beam Therapy Dosimetry with Ionisation Chambers

Incident and Emergency Centre

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control

HYDROGEN STORAGE AND MICROSTRUCTURE INVESTIGATIONS OF La 0.7- Av. Prof. Lineu Prestes, 2242, ZIP , São Paulo, Brazil.

Lung Cancer and Exposure to Ionizing Radiation

Simone Alberigi and Brigitte Roxana Soreanu Pecequilo

Nuclear Power Plant Instrumentation and Control in MEXICO

Incoming Letter. 01/10/1994 Final No Comments 8/18/00. 01/01/1994 Final No Comments 10/25/ /15/1994 Final No Comments 08/18/2000

José Egidio M.Tondin 1, Mario O. de Menezes 1, Marina B.A.Vasconcellos 1 and João A.Osso Jr. 1 ABSTRACT 1. INTRODUCTION

RN-222 RELEASE TO THE ENVIRONMENT: COMPARISON BETWEEN DIFFERENT GRANITE SOURCES

International Action Plan On The Decommissioning of Nuclear Facilities

DOE Policy- Reporting of. Update

Radiation and Environmental Protection at the CNSC

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

arxiv: v1 [physics.ins-det] 14 Jan 2014

DOSE RATES EVALUATION ON ENVIRONMENT AROUND ALMIRANTE ALVARO ALBERTO NUCLEAR POWER STATION AFTER ANGRA 2 NUCLEAR POWER PLANT S OPERATION

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY

Study of the Natural Circulation Phenomenon for Nuclear Reactors

Transcription:

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, IPEN-CNEN/SP, Av. Prof. Lineu Prestes, 2242, 05508-000, Brazil fbnonato@ipen.br,ctcescon@ipen.br,lcaldas@ipen.br Abstract. Thirty one direct reading dosimeters (pen dosimeters) were calibrated and tested in standard beams of gamma radiation, with and without the use of a phantom. The calibration was performed with a 60 Co source and tested with a 137 Cs source. The dose-response curves of the pen dosimeters and their calibration factors for a 60 Co source, with and without the use of a phantom were obtained. The results show the need to calibrate the pen dosimeters with a phantom. 1 Introduction Persons who are occupationally exposed to radiation have to be monitored in accordance to the recommendation of ICRP 60; in Brazil, the national standard "Basic Guidelines for Radiological Protection" determined by the Brazilian Commission of Nuclear Energy (CNEN) is used too. The individual monitoring controls the exposures, and estimates the doses received by the radiation workers. The ICRP recommends the exposure limits that ensure the health of the workers [1-2]. The individual dosimeters may be used to measure the exposures of the whole body or of the extremities. One of them is an additional dosimeter, the pen dosimeter. This instrument has a size of a pen pocket with an ionization chamber. These kinds of detectors are still very used in Brazil to measure accumulated doses or quantities of X and gamma radiation [3]. The objective of this work was to study the influence of the use of a phantom during the calibration of pen dosimeters with gamma radiation sources.

2 Materials and Methods Thirty one pen dosimeters of three different manufactures and four different models were calibrated and tested with and without the use of a phantom in standard gamma radiations. Their characteristics are presented in Table 1. They were divided per use time: old and new pens. Table 1. Characteristics of the pen dosimeters calibrated in this work Manufacturer Model Code of old pens Code of new pens Arrow-tech 138 1-6 14-31 Victoreen 541L 7-8 - Phy Seq-5 9-11 - Seq-6 12-13 - The instruments were initially calibrated with and without the presence of a slab phantom, using a 60 Co source of the gamma irradiator IPEN, serial number IR 003, named A, shown in Figure 1.a. Afterwards, they were tested with a 137 Cs source from another gamma irradiator, manufactured by Steuerungstechnik & Strahlenschutz GmbH, model OB85, named B, shown in Figure 1.b. The beam dosimetry of the gamma radiation 137 Cs source was performed using an ionization chamber from Physikalisch-Technische Werkstätten (PTW), model W32002 035, calibrated at the National Laboratory for Ionizing Radiation Metrology (LNMRI), Rio de Janeiro, Brazil. The characteristics of the gamma irradiators are presented in Table 2. The slab phantom, with dimensions of 300 mm x 300 mm x 150 mm, was developed at IPEN to represent the human torso (used for the calibration of whole body dosimeters) [4]. The calibration procedure of pen dosimeters with and without phantom is shown in Figures 2.a and 2.b. Fig. 1. The gamma irradiators IPEN, named A (a) and from Steuerungstechnik & Strahlenschutz GmbH, named B (b)

Table 2. Characteristics of the gamma radioactive sorces utilized in this work Irradiator Radiation Source Activity Calibration Date A gamma 60 Co 108.90 GBq 01/2002 B gamma 137 Cs 470.00 GBq 05/1995 Fig. 2. Calibration of pen dosimeters with (a) and without (b) phantom The measurements were taken with and without the presence of the slab phantom, and dose-response curves were obtained to compare the results. Afterwards, the calibration factors were obtained in a 60 Co beam, with and without the presence of the slab phantom. The calibration method was the use of known fields. The pen dosimeters were aligned with the radiation field in each case. The calibration factor (N) is determined by the ratio between the conventional true value (R) and the measured (M) value by the radiation detector, according to [4]: R N =. M The uncertainties of the measurements were determined in accordance with the recommendations of Gonçalves and Souza [5]. (1) 3 Results The results for the old pens were not totally satisfactory. Many of them did not show any differences at all on their responses when using or not the phantom. Most of the new pens showed good results. In this work, some results (dose-response curves) from old and new pens were chosen for presentation in Figures 3 and 4 for 60 Co and 137 Cs

sources respectively. The exposure unit of all these pen dosimeters is the old one (mr). The curves obtained from the new pens, with the 60 Co source, show that the difference in relation to the use or not of the phantom is smaller compared with the case of old pens. Moreover, all results with the 137 Cs source showed higher differences in relation to the use or not of the phantom, for both old and new pens. The calibration factors of the pen dosimeters were obtained for the 60 Co source, and their result intervals are presented in Table 3 with the presence or not of the slab phantom. Of the thirty one pen dosimeters, 80.6% of them presented calibration factors with phantom greater than without phantom; 9.7% presented calibration factors with phantom lower than without phantom; and 9.7% presented the same calibration factors to both cases. Then, most of the pen dosimeters presented different calibration factors depending on the presence of the phantom. The results show the need to calibrate these instruments with phantom.

Fig. 3. Dose-response curves of old pens (# 5 and 10) and new pens (# 21, 29 and 31), for 60 Co gamma radiation

Fig. 4. Dose-response curves of old pens (# 3, 7 and 17) and new pens (# 22 and 30), for 137 Cs gamma radiation

Table 3. Interval of calibration factors with (N 1 ) and without (N 2 ) phantom, for 60 Co radiation. Variation of calibration factor Pen Code N 1 N 2 1-6 0.92-1.04 0.97-1.09 7-8 0.96-1.04 1.06-1.09 9-11 0.82-0.96 0.90-0.99 12-13 0.87-1.00 0.85-1.00 14-31 0.89-1.93 0.91-1.11 4 Conclusions This work presents the dose-response curves of the pen dosimeters for 60 Co and 137 Cs sources, and with and without the presence of the phantom. The results show differences between them. Therefore, it is necessary to calibrate the pen dosimeters with the phantom, taking into consideration the backscattering. The new pens presented better results than the old pen dosimeters. Acknowledgments The authors are thankful to CNPq, FAPESP, CAPES and MCT (INCT for Radiation Metrology in Medicine), Brazil, for partial financial support. References 1. ICRP, International Commission on Radiological Protection, 1990 Recommendations of the International Commission on Radiological Protection, ICRP publication 60, 1991. 2. CNEN, Comissão Nacional de Energia Nuclear, Basic Guidelines for Radiation Protection (in Portuguese), CNEN NN 3.01, 2005. 3. Data sheet of pen dosimeter arrow-tech: http://www.arrowtechinc.com/pdf%20files/catalog%20at%205hr%20dosimeter%20info.pd f, accessed on March 5, 2011. 4. IAEA, International Atomic Energy Agency, Calibration of Radiation Protection Monitoring Instruments, Safety Report Series No. 16, 2000. 5. Gonçalves, A.A.Jr.; Sousa, A.R, Fundamentals of Scientific and Industrial Metrology, Barueri, Manole, (in Portuguese), 2008.