Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System



Similar documents
TGA/SDTA851 e Module. Thermal Analysis. Technical data Quality Service

SDT 2960 Simultaneous DSC-TGA including DTA capabilities

QGA Quantitative Gas Analyser

Bridging the analytical gap

COVER SHEET. This is the author version of article published as:

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA ABSTRACT

Unit 6 The Mole Concept

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)

CATALYSIS & THERMAL ANALYSIS

Thermo Scientific ConFlo IV Universal Interface

ESTEC-8. Peter J. Hatton & Barry W.L. Southward

Pesticide Analysis by Mass Spectrometry

Chapter 1: Moles and equations. Learning outcomes. you should be able to:

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemical Proportions in Compounds

Amount of Substance.

Mass Spectrometers. for Gas Analysis Applications. Gas Analysis

Pore size and BET surface area. analysis at your fingertips. SA Gas Adsorption Analyzer

MOLES AND MOLE CALCULATIONS

Mass Spectrometers. for Gas Analysis Applications. Gas Analysis

Chapter 3: Stoichiometry

FT-NIR for Online Analysis in Polyol Production

IB Chemistry. DP Chemistry Review

Chemical Calculations: The Mole Concept and Chemical Formulas. AW Atomic weight (mass of the atom of an element) was determined by relative weights.

Characterization of Electronic Materials Using Thermal Analysis

Functional Data Analysis of MALDI TOF Protein Spectra

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

13C NMR Spectroscopy

Module 5: Combustion Technology. Lecture 33: Combustion air calculation

TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES

The Mole Concept. The Mole. Masses of molecules

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Stoichiometry. 1. The total number of moles represented by 20 grams of calcium carbonate is (1) 1; (2) 2; (3) 0.1; (4) 0.2.

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Limiting Reagent Worksheet #1

= atm. 760 mm Hg. = atm. d. 767 torr = 767 mm Hg. = 1.01 atm

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

CONE CALORIMETER A TOOL FOR MEASURING HEAT RELEASE RATE

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis

atm = 760 torr = 760 mm Hg = kpa = psi. = atm. = atm. = 107 kpa 760 torr 1 atm 760 mm Hg = 790.

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

Chapter 8 How to Do Chemical Calculations

4.2 Bias, Standards and Standardization

Unit 2: Quantities in Chemistry

Honors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1. A chemical equation. (C-4.4)

The Mole and Molar Mass

F321 MOLES. Example If 1 atom has a mass of x g 1 mole of atoms will have a mass of x g x 6.02 x = 7.

Mass Spectrometry. Overview

Stoichiometry. Lecture Examples Answer Key

Balancing Chemical Equations

AN INVESTIGATION ON THE FORMATION OF CALCIUM NAPHTHENATE FROM COMMERCIAL NAPHTHENIC ACID SOLUTIONS BY THERMOGRAVIMETRIC ANALYSIS

DETERMINING THE ENTHALPY OF FORMATION OF CaCO 3

Chapter 1 The Atomic Nature of Matter

Chemical Composition. Introductory Chemistry: A Foundation FOURTH EDITION. Atomic Masses. Atomic Masses. Atomic Masses. Chapter 8

Moles. Balanced chemical equations Molar ratios Mass Composition Empirical and Molecular Mass Predicting Quantities Equations

Other Stoich Calculations A. mole mass (mass mole) calculations. GIVEN mol A x CE mol B. PT g A CE mol A MOLE MASS :

ATOMS. Multiple Choice Questions

CHEMISTRY II FINAL EXAM REVIEW

Quality Assurance for the Analysis of Steel by Gas Component Analysis

Summer Holidays Questions

AS1 MOLES. oxygen molecules have the formula O 2 the relative mass will be 2 x 16 = 32 so the molar mass will be 32g mol -1

AMD Analysis & Technology AG

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2

FP628. Nitrogen / Protein Determination by Combustion. Rev 1-6/8/11

Better DSC Isothermal Cure Kinetics Studies Using Power Compensation DSC

Trace Gas Exchange Measurements with Standard Infrared Analyzers

Calculation of Molar Masses. Molar Mass. Solutions. Solutions

Characterization of Polymers Using TGA

SCH 4C1 Unit 2 Problem Set Questions taken from Frank Mustoe et all, "Chemistry 11", McGraw-Hill Ryerson, 2001

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

The Mole. Chapter 10. Dimensional Analysis. The Mole. How much mass is in one atom of carbon-12? Molar Mass of Atoms 3/1/2015

vii TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION DEDICATION ACKNOWLEDGEMENT ABSTRACT ABSTRAK

W1 WORKSHOP ON STOICHIOMETRY

We know from the information given that we have an equal mass of each compound, but no real numbers to plug in and find moles. So what can we do?

Q1. A student studied the reaction between dilute hydrochloric acid and an excess of calcium carbonate.

Review - After School Matter Name: Review - After School Matter Tuesday, April 29, 2008

In semiconductor applications, the required mass flows

Molecules, Atoms, Grams and Mole Calculation Practice

10 The Mole. Section 10.1 Measuring Matter

Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston

Chemical Calculations: Formula Masses, Moles, and Chemical Equations

Analyzing Small Molecules by EI and GC-MS. July 2014

CHEMICAL REACTIONS AND REACTING MASSES AND VOLUMES

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2

Chapter 3. Chemical Reactions and Reaction Stoichiometry. Lecture Presentation. James F. Kirby Quinnipiac University Hamden, CT

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

FTIR and DSC of polymer films used for packaging: LLDPE, PP and PVDC

An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry

Chemical Equations & Stoichiometry

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

Lecture 35: Atmosphere in Furnaces

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Stoichiometry: Baking Soda and Vinegar Reactions

Chapter 3! Stoichiometry: Calculations with Chemical Formulas and Equations. Stoichiometry

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Transcription:

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Carlton G. Slough TA Instruments, 109 Lukens Drive, New Castle DE 19720, USA ABSTRACT The use of mass spectrometry for the identification of evolved gas species in thermogravimetric (TG) experiments is well known. The mass spectrometer (MS) can be attached directly to the outgas port of the thermogravimetric (TGA) instrument. Typically the mass spectrometer is used for identification only. However, there are situations where it would be desirable to obtain more information. For example, when two or more species are evolved during a weight transition, the total weight loss is known from the TGA, but the weights for the individual components are unknown. Using a simple calibration technique, trend data from a mass spectrometer can be used to semiquantitatively determine the amount of material evolved for low mass-to-charge (m/e) ratio species such as water and carbon dioxide. INTRODUCTION It is very common in the field of thermogravimetry to analyze the gases evolved during experiments by techniques such at mass spectrometry or Fourier transform infrared spectrometry (FTIR) (1). Such hyphenated techniques are typically used for detection and identification purposes only. However, it is beneficial if further information is derived. For example, when multiple species are released during a single weight loss event in a TGA, only the total weight loss is known. It is useful to know how the weight loss is distributed among the various components. Also, if reactions occur between the evolved gases and the purge gas, information concerning the masses involved is valuable. If a mass spectrometer is attached to the TGA, then it is possible to analyze the ion current data for a specific m/e species during a weight transition and convert it to a mass for that species. The technique involves first calibrating the TG/MS system by running a sample that evolves only the gas species of interest. For example, if water is the component of interest, then calibration involves collecting TG/MS data from a sample that only gives off water during a weight transition. EXPERIMENTAL A TA Instruments Q50 TGA was attached to a Pfeiffer Vacuum ThermoStar mass spectrometer. The Q50 TGA has a quartz-lined furnace that is ideal for evolved gas experiments. Evolved gas components from the TGA are fed via a capillary heated to 1 TA306

200 C to the mass spectrometer. The input gas is ionized by electron impact. The ions are then sent through a quadrapole mass filter and finally impacted onto a secondary electron multiplier (SEM) that produces an output current proportional to the ion current. Powder samples of calcium oxalate monohydrate (CaC 2 O 4 H 2 O) from Fisher Scientific and 99.97 % pure calcium carbonate (CaCO 3 ) from Alfa-Aesar were run in 100 µl platinum pans. The pans were cleaned and tared prior to sample loading. Figure 1 Simultaneous TGA/MS Thermal Curves of Calcium Oxalate Monohydrate RESULTS Figure 1 shows a typical thermogravimetry thermal curve of calcium oxalate monohydrate in a helium atmosphere. As shown, CaC 2 O 4 H 2 O decomposes in an inert atmosphere in three steps (2). 1. CaC 2 O 4 H 2 O CaC 2 O 4 + H 2 O (loss of water) 2. CaC 2 O 4 CaCO 3 + CO (loss of carbon monoxide) 3. CaCO 3 CaO + CO 2 (loss of carbon dioxide) If oxygen is present, a reaction with the CO produced in step 2 occurs: 4. CO + ½ O 2 CO 2 + O Thus, in the presence of low amounts of oxygen, both CO and CO 2 are produced simultaneously during the second weight loss transition. Figure 1 also shows the mass spectral trend data captured for the calcium oxalate run. In a trend scan only certain m/e species are tracked over time by the mass spectrometer. In this example data on m/e species 16, 17, 18, 28, 32 and 44 are collected and plotted. The typical ions associated with these m/e numbers are O, HO, H 2 O, CO, O 2 and CO 2 respectively. The data shows that during the second weight loss both CO and CO 2 are detected. This is not unusual, as carbon dioxide will fragment into smaller mass species, including CO, when subjected to electron impact ionization. However, the data also shows that the O 2 level drops during 2 TA306

the second weight loss. This is a definitive indication that reaction 4 is also proceeding. The total weight loss for the second transition from Figure 1 is 1.13 mg. This is all from CO, but it would be interesting to know how much of the CO is converted to CO 2, and how much CO 2 is thus produced. In order to answer this question, the MS system was calibrated to measure weight of evolved CO 2. Figure 2 Simultaneous TGA/MS Thermal Curves of Calcium Carbonate Calcium carbonate thermally decomposes according to step 3 outlined above. The presence of oxygen at small levels does not influence this reaction (3). Three different masses of calcium carbonate were run in the TG/MS system. Figure 2 shows an overlay of TG and MS trend data from a typical run. One weight loss transition occurs and CO 2 (m/e = 44) is released. In this trend scan, increases in CO (m/e = 28) and O (m/e = 16) are also detected, but these are expected fragments of CO 2. The area under the CO 2 curve can be related to the weight loss detected in the TGA. Figure 3 shows how the area of the CO 2 peak grows with increasing mass of sample. Table 1 lists the relevant measurements from the three runs and Figure 4 plots the area under the CO 2 curve versus weight loss as seen in the TGA. A linear relationship exists between the curve area and the weight loss. Results from a linear fit are also displayed on the graph. The correlation is high and the Y intercept is near 0, as should be the case. Table 1 Replicate Determinations of CaCO 3 MTGA/MS Calibration Initial Weight Weight Loss Weight Loss CO 2 Peak Area 0.437 mg 0.204 mg 46.7 % 0.939 µa s 2.43 mg 1.10 mg 45.2 % 5.327 µa s 12.3 mg 5.42 mg 44.2 % 28.24 µa s 3 TA306

This working curve can now be used to estimate the amount of CO 2 formed during the second transition in the calcium oxalate data. Integration of the CO 2 peak from the data in Figure 2 gives an area of 39.6 na min. Evaluating this data using the linear equation obtained in Figure 4 gives a calculated amount of 0.46 mg of CO 2 formed. Multiplying this by the ratio of the molecular weight of CO to the molecular weight of CO 2 gives a value of 0.29 mg of CO converted. Figure 3 Increase in CO 2 Mass Spectral Data with Initial Mass of Calcium Carbonate It is difficult to measure the accuracy of the technique. An alternate measure by a separate technique should be employed. However, as an approximate check of the accuracy of the method, the peak area of the CO 2 can be converted to a weight for the third weight loss event in calcium oxalate. Here only CO 2 is being released. The calculation gives a value of 1.36 mg as compared to a value of 1.74 mg by the TGA. These values are within 22 % of each other. DISCUSSION The calibration technique also works for other gases. Figure 5 shows the resultant plot for water. The first weight loss in calcium oxalate, which only releases water, was used to produce this curve. Again, three different masses of calcium oxalate were run and the results graphed. As with the graph for CO 2, the data is highly linear with a Y intercept near 0. Accuracy checks of this calibration were performed using cupric sulfate (hydrate), which loses water in two well-separated transitions. Calculations gave results that were within 15 % and 20 % of the TGA measurements of these two weight losses respectively To ensure that the best accuracy is obtained the several points concerning experimental procedure should be observed. The calibration must be done in the same purge gas as the sample being run. Changing purge gases could potentially shift baseline levels of the gases of interest. The calibration is sensitive to the baseline level and if the baseline is not the same during the experimental run as during the calibration run the 4 TA306

results will be in error. The error is roughly proportional to the ratio of the baseline levels. Figure 4 Correlation of TGA Weight Loss with Mass Spectral Peak Area of CO 2 As a calibration material, powders are the best choice. Preliminary data show that use of chunk samples can produce a highly non-linear relation between peak area and weight loss. Finally, the assumption is made that a gas fragments in the mass spectrometer always in the same proportions. For example, CO 2 will fragment into CO and O (and other fragments) always in the same proportions. Statistically, for large numbers of molecules, this assumption should be valid. Comparing the ratio of CO 2 to O for the three calibration runs can test this assumption. The ratio was the same in all three runs. Figure 5 Correlation of TGA Weight Loss with Mass Spectral Peak Area for Water 5 TA306

CONCLUSIONS A simple calibration technique for converting areas under mass spectral data curves to weights of the detected species has been discussed. Results for low m/e species such as H 2 O and CO 2 were presented with accuracies within 25% obtained. While better accuracy is desirable, baseline shifts in the mass spectral data that occur during the collection probably make it difficult. The technique may be extended to higher m/e ratio ions. REFERENCES 1. Edith A. Turi (Ed.), Thermal Characterization of Polymeric Materials, 2 nd Edition, 1997, 1, pp. 49-73. 2. Clement Duval, Inorganic Thermogravimetric Analysis, 2 nd Edition, 1963, pp. 281-282. 3. Wesley W. Wendlandt, Thermal Methods of Analysis, 2 nd Edition, 1974, p.16. KEYWORDS Compositional analysis, decomposition, inorganics, mass spectrometry, thermogravimetry TA Instruments United States, 109 Lukens Drive, New Castle, DE 19720 Phone: 1-302-427-4000 Fax: 1-302-427-4001 E-mail: info@tainst.com United Kingdom Phone: 44-1-372-360363 Fax: 44-1-372-360135 E-mail: uk@tainst.com Spain Phone: 34-93-600-9300 Fax: 34-93-325-9896 E-mail: spain@tainst.com Belgium/Luxembourg Phone: 32-2-706-0080 Fax: 32-2-706-0081 E-mail: Belgium@tainst.com Netherlands Phone: 31-76-508-7270 Fax: 31-76-508-7280 E-mail: Netherlands@tainst.com Germany Phone: 49-6023-9647-0 Fax: 49-6023-96477-7 E-mail: germany@tainst.com France Phone: 33-1-304-89460 Fax: 33-1-304-89451 E-mail: france@tainst.com Italy Phone: 39-02-27421-283 Fax: 39-02-2501-827 E-mail: italia@tainst.com Sweden/Norway Phone: 46-8-594-69-200 Fax: 46-8-594-69-209 E-mail: Sweden@tainst.com Japan Phone: 813 5479 8418) Fax: 813 5479 7488 E-mail: nurayama@taij.po-jp.com Australia Phone: 613 9553 0813 Fax: 61 3 9553 0813 E-mail: steve_shamis@waters.com To contact your local TA Instruments representative visit our website at www.tainst.com 6 TA306